ExoMars PanCam Planetary Protection Alex Rousseau.

18
ExoMars PanCam Planetary Protection Alex Rousseau

Transcript of ExoMars PanCam Planetary Protection Alex Rousseau.

Page 1: ExoMars PanCam Planetary Protection Alex Rousseau.

ExoMars PanCam Planetary Protection

Alex Rousseau

Page 2: ExoMars PanCam Planetary Protection Alex Rousseau.

PanCam isolation of accountable surfaces

• Use a combination of seals & HEPA filter on the OB to exclude the complete internal volume of the OB from bioburden accountable surfaces:– WACs (including filter wheels)– HRC– PIU electronics (including all associated harnessing).

• MSSL will follow “Excluding bioburden accountable surfaces for Instruments” EXM-PL-TNO-ESA-00001, procedures.

Page 3: ExoMars PanCam Planetary Protection Alex Rousseau.

PanCam isolation of accountable surfaces (Cont.)

Page 4: ExoMars PanCam Planetary Protection Alex Rousseau.

Identification of isolation zones on the PanCam OB

1. OB HEPA – Metallic fibre2. HRC optical port – Metallic or non-metallic.3. WACs optical ports (x2) – Metallic or non-

metallic seals.4. Connector feedthroughs – On OB – hermetic

type connector feedthroughs.

Page 5: ExoMars PanCam Planetary Protection Alex Rousseau.

PanCam PP Philosophy• Hot ultrasonic at unit level.• Frequent solvent wipe during AIT. (not for BB

reduction but bulk BC removal from surfaces)• Hydrogen Peroxide sterilisation at instrument

level.• For PanCam AIT phase, WACs, HRC and PIU

handled, assembled in ISO Class 5 & 6 environments.

• During AIT PanCam will be processed using standard flight hardware handling methods (strict CC practices).

Page 6: ExoMars PanCam Planetary Protection Alex Rousseau.

PanCam DHMR & H2O2

• Two subsystems on PanCam are DHMR incompatible:– WACs detectors.– PIU

• Lowest DHMR T at 110°C is a concern (PIU thermally stressed to unacceptable levels)

• H2O2 sterilisation method will need to be qualified as an alternative to DHMR.

Page 7: ExoMars PanCam Planetary Protection Alex Rousseau.

PanCam Hardware Assay philosophy

• Assay frequently during AIT phases:– After unit level ultrasonic cleaning.– After unit level vacuum bake out. (internal &

external)– After HRC/OB integration. (Internal & external).– After WACs/PIU/OB integration. (internal &

external).– Before and after instrument level sterilisation

Page 8: ExoMars PanCam Planetary Protection Alex Rousseau.

PanCam Hardware Assay flow diagram

Page 9: ExoMars PanCam Planetary Protection Alex Rousseau.

MSSL PanCam facilities

• MSSL ISO class 5 cleanroom facility has two zones:1. ISO class 5 PanCam hardware processing and

assembly area.2. ISO Class 5 aseptic area.

• Aseptic area includes ISO Class 4 laminar flow bench.

Page 10: ExoMars PanCam Planetary Protection Alex Rousseau.

MSSL PanCam facilities

Page 11: ExoMars PanCam Planetary Protection Alex Rousseau.

MSSL PanCam AIT test facilities

• Cleanroom – MSSL (Assembly, processing)• Vibration – MSSL (ISO Class 7 air shower)• TV/TB – MSSL (ISO Class 6)• Shock – Offsite (Portsmouth, classification TBC)• EMC – Offsite (Location & classification TBC).• PanCam instrument calibration – At MSSL or at

NPL (If at MSSL ISO Class 6, if at NPL classification TBC)

Page 12: ExoMars PanCam Planetary Protection Alex Rousseau.

MSSL PanCam AIT test flow diagram

Page 13: ExoMars PanCam Planetary Protection Alex Rousseau.

MSSL Cleanroom facility bioburden reduction

• Start with an empty cleanroom and clean all surfaces. (walls, floor and ceiling).

• Process equipment needed for PanCam AIT. Cleaning includes, ultrasonic, vacuum bake out and solvent wet wiping methods.

• Once equipment in place in cleanroom, further process equipment/tooling required for aseptic area.

• During AIT phase PanCam hardware processing and assembly area will be cleaned at the start of each day before activities begin.

Page 14: ExoMars PanCam Planetary Protection Alex Rousseau.

MSSL Cleanroom facility bioburden monitoring

• ISO Class 5 aseptic area will be bioburden monitored using:– Slit-agar air sampler– Swab & wipe assay

• ISO Class 5 cleanroom will be verified at (or just after CDR) using assay procedure 2.

• Frequent assays will be done during PanCam AIT phase of the ISO Class 5 aseptic area.

Page 15: ExoMars PanCam Planetary Protection Alex Rousseau.

ISO Class 5 cleanroom upgrade

• ISO Class 5 cleanroom to be upgraded and finished at CDR.

• Cleanroom commissioning and verification after CDR.

• Cleanroom ready for normal operations at PanCam EQM phase

Page 16: ExoMars PanCam Planetary Protection Alex Rousseau.

PDR RIDs responseRID #: PANCAM-1

Note that although the RID references the KT document (EXM-PL-PCHRC-KT-PL-0003 PP and CC_Plan), the issues raised are applicable to the MSSL PP, CCP & PPIP.

Estimate of internal and external surface area:• MSSL response: The PanCam PPIP will be updated with the internal and external surface areas

after PDR. Estimate of volume of non-metallic materials:• MSSL Response: The PanCam PPIP will be updated with the non-metallic materials volume

after PDR. Identification of items that cannot be wiped or assayed with damps swabs:• MSSL response: The PanCam PPIP will be updated with a list identifying areas that cannot be

wiped or assayed. Temperature tolerance for DHMR:• MSSL response: The PanCam PPIP will be updated to include temperature tolerances of the

PanCam instrument including detector subsystems. Preliminary organic materials list:• MSSL response: The PanCam PPIP will be updated to include an organic materials list. UV bioburden control:• MSSL response: The DML will be updated to exclude UV as a bioburden reduction method.

Page 17: ExoMars PanCam Planetary Protection Alex Rousseau.

PDR RIDs responseRID #: PANCAM-2

Note that although the RID references the KT document (EXM-PL-PCHRC-KT-PL-0003 PP and CC_Plan), the issues raised are applicable to the MSSL PP, CCP & PPIP.

Who will be responsible for the OB HEPA and for the request for approval according to PPL-PP-ALL-0070 and TN EXM-PL-TNO-ESA-00001?:

• MSSL response: MSSL will request approval for the OB HEPA according to PPL-PP-ALL-0070. It is envisaged that the HEPA will be subjected to qualification tests at ESA facilities.

• Responsibility of the OB HEPA is MSSLs.

Page 18: ExoMars PanCam Planetary Protection Alex Rousseau.

PDR RIDs responseRID #: PANCAM-3

There is no clear need/requirement identified for a final DHMR of the entire assembly integrated on the mast.

• MSSL response: MSSL will contact TAS-I and clarify the issues with regard to the PanCam instrument and the DHMR process. As mentioned in the PPIP, there are two PanCam subsystems that are not compatible with DHMR. Ways in which this can be dealt with and responsibilities of PP after delivery of the PanCam instrument will be discussed with TAS-I and the relevant PP & PPIP sections will be updated to reflect this after PDR.