SKF TKTI 20 - Reliability Direct

36
Instructions for use SKF TKTI 20

Transcript of SKF TKTI 20 - Reliability Direct

Page 1: SKF TKTI 20 - Reliability Direct

Instructions for use

SKF TKTI 20

Page 2: SKF TKTI 20 - Reliability Direct

2 SKF TKTI 20

EC Declaration of conformity ....................................................................................3

Safety recommendations ..........................................................................................4

1. Description .......................................................................................................5

2. Getting started .................................................................................................52.1 Contents.......................................................................................................................... 52.2 Charginginternalbattery............................................................................................. 62.3 Replacingtheinternalbattery.................................................................................... 7

3. Basic features of the SKF Thermal Camera TKTI 20 ..........................................83.1 Hardware........................................................................................................................ 83.2 SwitchingontheTKTI20............................................................................................. 93.3 Focusing.......................................................................................................................... 93.4 Usingthelaserpointer...............................................................................................103.5 Fieldofview..................................................................................................................113.6 Span,Level,Range......................................................................................................11

4. Using the direct access buttons .......................................................................124.1 Directaccessbutton1................................................................................................124.2 Directaccessbutton2................................................................................................154.3 Directaccessbutton3................................................................................................174.4 Directaccessbutton4................................................................................................17 4.4.1 MenuInfraredSettings.................................................................................18 4.4.2 MenuMeasurementOptions........................................................................21 4.4.3 MenuCameraSettings.................................................................................24 4.4.4 MenuAudioSettings.....................................................................................25 4.4.5 MenuImageBrowser....................................................................................26 4.4.6 MenuDate&TimeSettings.........................................................................27 4.4.7 MenuLanguageSelection............................................................................27 4.4.8 MenuDisplaySettings..................................................................................29

5. Saving and annotating pictures.......................................................................305.1 Savingapicture...........................................................................................................305.2 Addingavoicemessage.............................................................................................305.3 AddingaCaption.........................................................................................................31

6. Technical specifications ...................................................................................32

Table of contents

Original instructions

Page 3: SKF TKTI 20 - Reliability Direct

3SKF TKTI 20

EC Declaration of conformity

We, SKFMaintenanceProducts Kelvinbaan16 3439MTNieuwegein TheNetherlands

herewithdeclarethatthefollowingproduct:

SKF Thermal Camera TKTI 20

Conformstothefollowingstandards;

• EMCDirective2004/108/ECasamended“CouncilDirectiveontheapproximationofthelawsoftheMemberStatesrelatingtoelectromagneticcompatibility”asoutlinedinharmonizednorms:EN61000-6-2:2005ImmunityforindustrialenvironmentsEN61000-6-3:2007Emissionstandardforresidential,commercialandlight-industrialenvironments

• CFR47:2009CodeofFederalregulations:Pt15subpartB–RadioFrequencyDevices–UnintentionalRadiators

Thelaserisclassifiedinaccordancetothe21CFR1040.10and1040.11exceptfordeviationspursuanttolasernoticeNo.50datedJune24th2007andcomplieswithIEC/EN60825-1(2001).

Nieuwegein,TheNetherlands,March2011

SébastienDavidManagerProductDevelopmentandQuality

Page 4: SKF TKTI 20 - Reliability Direct

4 SKF TKTI 20

Safety recommendations

Thisproductcomplieswith21CFR1040.10and1040.11exceptfordeviationspursuanttolasernoticeNo.50

datedJune24th2007complieswithIEC/EN60825-1(2001).

Warning• TheequipmentdescribedinthisdocumentusesaClass2laser.

Donotlookdirectlyintothelaserbeamorthelaserbeamexitaperture,irreversibledamagetotheeyemayoccur.Thelasershouldnotbeoperatedwhentherearepersonnelinthecamera’sfieldofview.

Caution• Useofcontrolsoradjustmentsorperformanceofproceduresotherthan

thosespecifiedinthisdocumentmayresultinhazardouslaserradiationexposure.

• Tohelpminimizeburnhazards,beawarethatthermalimagesofhighlyreflectiveobjectswillshowlowerthanactualtemperaturemeasurements.

• Donotuseinamannernotspecifiedinthisinstructionforuse.• Donotuseinexplosiveatmospheres.• Alwaysusethelanyardprovidedtohelpminimisedamagetothe

thermalimagerandpersonalinjuryaccidents.• OnlyusegenuineSKFsparesandaccessories.Othersparesand

accessoriesmaydamagethethermalcamera.• Thethermalcameracontainsnouserserviceableparts.Openingthe

imagercasingvoidswarranty.AllserviceworkmustbecarriedoutbyanauthorisedSKFworkshop.

635nm <1mW <0.2mr

Page 5: SKF TKTI 20 - Reliability Direct

5SKF TKTI 20

1. DescriptionTheSKFTKTI20isadigitalcamerawithadvancedthermalimagingcapabilities.ItisidealforuseasaPredictiveMaintenance(PdM)tool.Thermographyisanon-invasivetechniqueandiswellsuitedformechanicalandelectricalinspections,energyconservationandplantsafetyactivities.TheTKTI20hasalargebacklit31/2”colorscreen,whichgivesaclear,sharpimageusinganyoftheeightcolorpalettes.Designedforeasyoperation,multipletemperaturescanbemeasuredandthedifferencesbetweenthemdisplayed.Otheradvancedmeasurementoptionsarealsoavailable.ImagescanbestoredasradiometricdataanddigitalimagesonthemicroSDcard(supplied).PowerfulPCsoftwaresuppliedwiththeTKTI20,allowstheimagestobefurtheranalyzedandtobeoptimizedforeaseofinterpretation.

2. Getting started

2.1 Contents

12 5 6

4 3

Figure 1. Contents of the case

1 SKFThermalCameraTKTI202 Powersupplyunit(4-6V)and4Internationaladaptors3 CD-Softwareandinstructions(MP5387)4 USBcable(cameratoPC)5 QuickStartguideSKFThermalCameraTKTI20(MP5384)6 CertificateofCalibrationandConformance

Inspectalltheitems.Ifanyitemisdamagedormissing,pleasenotifyyourdealerorlocalSKFofficeimmediately.

Page 6: SKF TKTI 20 - Reliability Direct

6 SKF TKTI 20

2.2 Charging internal battery

• TheTKTI20’sbatterycanbechargedinthecamerabyconnectingthesupplied4-6VpowersupplyunitintotheMiniUSBportinthehatchofthethermalcamera(seefigure2).

• Thebatteryconditionindicatorappearsonthescreenwhentheimagerisswitchedon(seefigure3).

• AgreenLEDwillappearwhilechargingandturnredoncethechargeiscomplete(seefigure2).Afullychargedbatterywillapproximatelylast5hours.

MiniUSBPort

Red=FullGreen=Charginginprogress

Figure 2. Connection ports

TemperatureSpanLevel

Spot Delta Emissivity Reflected Batterytemperature temperature temperature condition(hot&cold)

StatusofDirectaccessbuttonsThewhiteborderediconistheactivefunction.

Figure 3. Screen overview

Thebottommenuwilldisappearafter10seconds,pressanybuttontomakeitreappear.Thegreenrectangleshowstheareawheretemperatureismeasurable.

Page 7: SKF TKTI 20 - Reliability Direct

7SKF TKTI 20

2.3 Replacing the internal battery

TheTKTI20hasafieldreplaceablebattery.

Figure 4: Battery replacement

Openthelockonthebatterycoverandremovethecover,removethebattery,insertanewoneensuringthecontactpadsaretowardsthefrontofthecamera.Replaceandlockthecover.

Page 8: SKF TKTI 20 - Reliability Direct

8 SKF TKTI 20

3. Basic features of the SKF Thermal Camera TKTI 20

3.1 Hardware

TheTKTI20isdesignedforuseasahandheldthermalcamera.ThermalimagesarestoredonamicroSDmemorycard(suppliedsizeis2GB).TheseimagescanbetransferredtoaPCusingtheUSBcableprovidedorviaamicroSDcardreaderconnectedtothePC.PCsoftwareSKFTKTI20ThermalCameraSuiteisincludedforviewingandanalysisofsavedthermalanddigitalimages.

HatchOn/offbutton

4navigationbuttons&

CentretogglebuttonScreen

Directaccessbuttons1to4

Figure 5. Main camera controls

Connectionports:-MiniUSB(Power+PC)

-MicroSD-Headset

Savebutton

Thermalcameralens

Focusring

Digitalcameralens

Light

Figure 6. Front panel

Page 9: SKF TKTI 20 - Reliability Direct

9SKF TKTI 20

3.2 Switching on the TKTI 20

• TheTKTI20isswitchedonandoff bypressingandholdingdownthe powerbutton.

• TheTKTI20takesapproximately 20secondstobefullyoperational.

Figure 7. Camera on/off button

3.3 Focusing

• Sharpentheimagebyrotating thefocusring

Figure 8. Focusing

Focus:• Isthemostimportantsteptoobtainagoodqualityimage• Influencestheaccuracyofthetemperaturereading• IstheonlysettingthatcannotbemodifiedwhenviewingtheimagesonaPCItisadvisedtostayina100%thermalimagemodeforfocusing(seechapter4.2).Somepalettesallowyoutomakeabetterfocus.Makesureyouareinautomaticspanandlevel.

Rotatingthefocusringinaclockwisedirection(fromuserpointofview)focusestheimagesatshorterdistancesdowntoaminimumof30cm.Rotatingthelensintheoppositedirectionfocusestheimageratlongerdistancesuptoinfinity.

Rotatethelensuntiltheimagehasasharpdefinition,withthesharpestcontrastatobjectedges.

Page 10: SKF TKTI 20 - Reliability Direct

10 SKF TKTI 20

Focused Not focused

Figure 9. Focusing

Whenfocusing,afocusbarappearsonthescreenshowingthefocuseddistance.REMEMBERTOREMOVETHELENSCAP

Figure 10. Focus bar

3.4 Using the laser pointer

WhentheTKTI20isswitched‘ON’,thelasermaybeactivatedbypressingandholdingdownthelaserbutton(seefigure11)onthehandle.

Laserpointerbutton

Figure 11. Laser pointer button

Thelaserpointerisusedtoilluminateandidentifyfeaturesintheimage.

Note: Thelasercentreisalignedtoobjectsatadistanceof3m.

Page 11: SKF TKTI 20 - Reliability Direct

11SKF TKTI 20

3.5 Field of view

Figure 12. Field of view

TheTKTI20hasa20˚x15˚fieldofview,anda160x120pixeldetector.Thedetectorresolutioncanbeinterpolatedto320x240(seesection4.4.1;Interpolation).Thefollowingtablegivesanindicationoftheminimumdimensionofanobjectyoucanmeasurethetemperatureatagivendistance.

Distance (m) 0,5 1 5 10 15

Object (cm) 0,3 0,7 3,3 6,6 10

Theobjectsizeisproportionaltothedistance.TheIFOVis2,2mRad.

3.6 Span, Level, Range

TheTKTI20canoperatefrom-10°Cto+250°C;thisistherangeofthecamera.Thespanisdefinedbytheminimumandmaximumtemperaturesonthescaleontherightsideofthescreen.Awiderspanwillgiveyoulessthermaldetails,whereasanarrowerspanwillgiveyougoodthermaldetailsbutalsoincreasesthenoiseintheimage.Youcanadjustitspositionwithintherange,thisisthelevelsetting.

Page 12: SKF TKTI 20 - Reliability Direct

12 SKF TKTI 20

4. Using the direct access buttons

TheDirectaccessbuttonpositionsareshowninthebelowfigure.Theiconwithawhitebordershowstheselectedfunctionthatcanbeadjusted(seealsofigure5):

1 2 3 4

5

6

6

66

Figure 13. Direct access buttons

1 Directaccessbutton1 4 Directaccessbutton42 Directaccessbutton2 5 Centretogglebutton3 Directaccessbutton3 6 4Navigationbuttons

4.1 Direct access button 1

WithDirectaccessbutton1youcanaccess:

a. Alignment (ifvisibleon :seeDirectaccessbutton2).

Asthevisibleandthermalcameraarenotco-axialthevisibleandthermalimageoftenneedtobealigned.Thisisusuallyrequiredwheninspectingobjectsatdifferentdistances.Itispossibletoalignthethermalandvisible(digital)imagesonlyon“Live”mode.Furtheradjustmentscanbedoneonthesavedimageswiththesoftware.

1. PressDirectaccessbutton2tolocatethecameraon/off optionandtogglethecentrebuttontoturnthecameraoptionon.

2. PressDirectaccesbutton1tolocatethealignment option.

3. Alignmentpresetsareaccessiblewithcentretoggle button.Youcanalsousethenavigationbuttonsforfinaladjustment.

Page 13: SKF TKTI 20 - Reliability Direct

13SKF TKTI 20

Align up

Align left Fully aligned Align right

Align down

Figure 14. Examples of alignment

Note: Onlythedigitalimageismoving.

b. Measurement options

TobedisplayedwhilenavigatingwithDirectaccessbutton1,theseoptionsneedtobeenabledintheMeasurementOptionmenu.

ToaccesstheMeasurementOptionmenu

pressDirectaccessbutton4:

andselectthecorrecticon:

Page 14: SKF TKTI 20 - Reliability Direct

14 SKF TKTI 20

Directaccessbutton1candisplaythefollowingicons:

CursoroneMenu>Measurementoptions>Cursors:1

CursortwoMenu>MeasurementOptions>Cursors:2

CursorthreeMenu>MeasurementOptions>Cursors:3

CursorfourMenu>MeasurementOptions>Cursors:4

IsothermlowtempMenu>MeasurementOptions>Isotherms:(Low,High&Low)

IsothermhightempMenu>MeasurementOptions>Isotherms:(High,High&Low)

AreaMenu>MeasurementOptions>Area:On

Hot/coldtrackingMenu>MeasurementOptions>Tracking:(High,Low,High&Low)

Align

(IfCameraOn:seeDirectAccessButton2)

LowtemprangeMenu>MeasurementOptions>Temp.Alarm:(Low,High&Low)

HightemprangeMenu>MeasurementOptions>Temp.Alarm:(High,High&Low)

HorizontalprofileMenu>MeasurementOptions>Temp.Profile:Horizontal

VerticalprofileMenu>MeasurementOptions>Temp.Profile:Vertical

Seefordetaileddescriptionofthefunctionalityoftheseoptionssection4.4.2:MeasurementOptions.

Page 15: SKF TKTI 20 - Reliability Direct

15SKF TKTI 20

4.2 Direct access button 2

Whenapictureisdisplayedonthescreen,thisbuttongivesaccessto3differentoptionswith2settingsforeach(toggleusingcentretogglebutton):

1. Camera on/off

Enable/disablethedisplayofadigitalpictureblendedwith thermalimage.

How to set Blending:

1. PressDirectaccessbutton2tolocatethecameraon/offoption andtogglethecentrebuttontoturnthecameraoptionon.

2. Blendingbetweenthedigitalandthermalimagebypressingthe up/downbuttons

3. Thiswillblendthetwoimagesonapercentagescale0%,25%,50%75%,100%andPictureandPicture(PiP).

4. BlendingisalsopossibleinPiPmode. Youcanadjustthepercentagebypressingtheleft/rightbutton.

100% thermal 25% visible 75% thermal 50% visible 50% thermal

75% visible 25% thermal 100% visible Picture in Picture

2. Light on/off

Enable/disablehighpowerLEDinthefrontpanel, helpfultotakepicturesinadarkenvironment(seefigure6).

Page 16: SKF TKTI 20 - Reliability Direct

16 SKF TKTI 20

3. Temperature range scale

Auto

Automaticallyadjuststheimagetoallowforthehighestandlowest temperatureinthescene.Usefulwhenstartinganimagerandswitchingtodifferentpartsofascene.

Manual

Theuserdefinestherangeoftemperaturestobedisplayed.Manualsettingisusefulwhentheuserwantstoexaminevariouspiecesofsimilarequipment.Inafixedtemperaturerange,anomaliesordiscrepanciescanbepreciselyhighlighted.

Setthespanusingtherightandleftnavigationbuttonandthelevel usingtheupanddownnavigationbutton.

Persistant Manual

Thespanandlevelsettingsareeitherselectedautomaticallybythecameraormanuallybytheuser.Ifthismodeisselected,thecamerastoresthelevelandspansettings.OnpoweringupthedefaultmodeisalwaysAuto,howeverselectingthepersistentmanualmodesetsthelevelandspantothelaststoredvalues.

Setthespanusingtherightandleftnavigationbuttonandthelevel usingtheupanddownnavigationbutton.

Page 17: SKF TKTI 20 - Reliability Direct

17SKF TKTI 20

4. Play audio (only in picture viewing mode) Playspreviouslyrecordedvoiceannotationsavedwiththeimage.

4.3 Direct access button 3

Thisbuttontogglesbetweenliveimageandafrozenimage

4.4 Direct access button 4

Thisbuttonselectsthecameramenus.

Infraredsettings

Measurementoptions

Camerasettings

Image browser

Date &Timesettings

Languageselection

Audio settings

Display settings

Page 18: SKF TKTI 20 - Reliability Direct

18 SKF TKTI 20

Menu selection

1. PressDirectaccessbutton4toenterthemenuselectionscreen

2. Navigatetothedesiredoptionusingthenavigationbuttons. (Theiconwiththeyellowbordershowtheselectedoption)

3. PressDirectaccessbutton3orthecentretogglebuttontoenter theselectedoption.

Menu items

1. Arrowupanddowntothedesireditem

2. Usetheleft/rightarrowstochangethevalueofthisitem3. SelectDirectaccessbutton4toexitorDirectaccessbutton1 togobacktothemainmenu.

4.4.1 Menu Infrared Settings

Emissivity

Differentsurfacescanradiatedifferentamountsofinfraredenergyatthesametemperature.Thisdifferenceintemperaturemeasurementcanbecorrectedbysettingtheemissivityvalue.Theamountofinfraredradiationemittedbyasurfacedependsonbothitstemperatureanditsemissivity.Surfacesthataregoodreflectors(e.g.polishedmetal)arepooremitters,andsurfacesthataregoodemitters(e.g.humanskin)arepoorreflectors.Ablackbodyisdefinedasanobjectthatabsorbsallradiationfallingonitanditisaperfectemitterofradiation.

Theemissivityofasurface(usuallywrittenε)istheratiooftheenergyradiatedbythatsurfacetotheenergyradiatedbyablackbodyatthesametemperature.Foraccuratetemperaturemeasurements,theemissivityofthesurfacebeingmeasuredmustbeenteredintothecamera.Thisisdonebyenteringanumberintherange0.10(forpolishedchromium)to1.00(forablackbody).Anemissivitylookuptableisprovided,whichliststheemissivitiesofarangeofcommonmaterials.

Page 19: SKF TKTI 20 - Reliability Direct

19SKF TKTI 20

Itisnotrecommendedthattemperaturemeasurementsbeattemptedwhenemissivityvalueslowerthan0.70arerequired,becauselargeerrorsarelikelyduetoreflectedradiationfromsurroundingobjects.

Setemissivityvaluebetween0.10and1.00formeasuringtemperature.Werecommendausualvalueof0.95.Emissivitycanbechangedafterwardinthesavedpicture.

PressingDirectaccessbutton3(εTable)givesatableofemissivityvaluesofcommonmaterialsfromwhichaselectioncanbemade.Ifmorethanonecursorisselected,itispossibletosetadifferentemissivityforeachcursorbypressingDirectaccessbutton3(Cursorε),pressingthesamebuttonagainwillgiveaccesstotheemissivitytable.

Palette (8 options)

Differentcolourpalettesareusefulinviewingobjectsandscenesfordifferentrequirements.Asasimpleruleofthumb,paletteswithalotofdifferentcoloursaremorehelpfulwhenlookingforhotspotsorcoldspots,whereaspaletteswithfewerandgraduallychangingcoloursaremoreusefulinviewingchangesoftemperaturesinasceneorobject.Werecommendthatyoufindthepalettewhichsuitsyoubest.ItispossibletochangethepaletteonthePCsoftware.

1 2 3 4 5 6 7 8

1. White Hot Mono2. Ironbow3. Rainbow4. Isotherm Style

5. Hot Metal6. High Contrast7. 16 step Rainbow8. Black Hot Mono

Page 20: SKF TKTI 20 - Reliability Direct

20 SKF TKTI 20

Reflected TemperatureThis is only applicable if an emissivity of less than 1 is selected.

Someoftheinfraredenergyseenbythecamerafromasurfacewithanemissivityoflessthan1.00isenergyreflectedbythebackground.Ifthereisahotobjectinthebackground,thiscanhaveasignificanteffectonthetemperaturemeasured.Byenteringareflectedtemperaturevalue,thecameracancorrectfortheeffectofthisreflectedbackgroundenergy.Usuallysettotheambienttemperature.

Temperature unitsChoosebetween˚Cand˚F.

Integration. (1 low to 4 high.)Thecameranormallyoperatesataframerateof8Hz(i.e.theimageisupdated8timespersecond).Forviewingscenesinwhichthereisverylittletemperaturevariation,theimagemaybeimprovedbyintegratingoverseveralframestoreducethenoise.

Chooseanintegrationperiodfrom1(fast)to9(slow).Howeveritisrecommendedtokeepthevalueat1fornormaluse,highervaluescanbeusedwhenthecameraisheldsteadilyorfixedonatripod.

Interpolation. (On/Off.)Interpolationisonbydefaultasitsmoothestheimagetoprovideabettervisualimage.Turningitoffgivesamorepixelatedimage.Someusersmaywishtoobservethescenewiththeactualresolutionofthedetector.Theprocessorcalculatesthetemperatureofsub-pixelsandthereforedisplaysmorepixelsonthescreen.Thissettingdoesn’tchangeanythingtothetemperaturemeasurement.Actualresolutionis160x120.Byturninginterpolationontheresolutionisinterpolatedto320x240.

Page 21: SKF TKTI 20 - Reliability Direct

21SKF TKTI 20

4.4.2 Menu Measurement Options

Thismenuenablestheselectionofoptionsfortemperaturemeasurements.ThesymbolsshownbelowindicatetheiconsshownforDirectaccessbutton1whenthevariousoptionsareselected.

Cursors

Uptofourmeasurementcursorscanbeselected.Whentwocursorsarechosen,thetemperaturesofbothcursorsandthetemperaturedifferencebetweenthemwillbedisplayed.WhenoneofthecursorsisselectedbyDirectaccessbutton1,itcanbemovedaroundonthedisplaybythenavigationbuttons.Iftwoormorecursorsareselectedthenindividualemissivityvaluescanbeassignedtoeachcursor.Whenanemissivitylessthan1.00isselectedforacursor,thenthatcursorflashesonthescreen.Theemissivityofcursor1isalwaysdisplayed.

Four tabular cursors Three tagged cursors

Page 22: SKF TKTI 20 - Reliability Direct

22 SKF TKTI 20

Tracking

Select“High”,“Low”,or“High&Low”inordertotrackandmeasurethefollowing:• thehottestpartofthescene• thecoldestpartofthescene• boththehottestandcoldestpartofthescene.

Cursors temperature Cursor temperature difference

Tracking hot in red and cold in blue

Area

Areaanalysisisusefulwhenwantingtoknowthehighestandlowesttemperatureinapartofthescene.Italsoshowstheaveragetemperature.Therearethreeboxsizestochoosefrom.

Maximal temperature Minimal temperature Average temperature

Area analysis

Page 23: SKF TKTI 20 - Reliability Direct

23SKF TKTI 20

Isotherms

Select“High”,“Low”,or“High&Low”inordertohighlightareasofthescenewithtemperatureswithinahigh(red)orlow(blue)temperatureband.Thetemperaturebandsareadjustablebymeansofdirectaccessbutton1andthenavigationkeys.Useleft/rightnavigationbuttontosetthespanandupanddownnavigationbuttontosettheleveloftheisotherm.

Red isotherm Blue isotherm Combined isotherms

Temperature profiles

Select“Horizontal”or“Vertical”toenableahistogramoftemperaturevaluesalongahorizontalorverticalcrosssectiontobedisplayedontherighthandsideorthebottomofthedisplay.ThepositionofthecrosssectionisindicatedbysmallarrowsattheleftandrightortopandbottomoftheimageandcanbeadjustedbymeansofHotkey1andthenavigationbuttons.

Vertical profile Horizontal profile

Page 24: SKF TKTI 20 - Reliability Direct

24 SKF TKTI 20

Temperature alarms

Select“High”,“Low”,or“High&Low”.Visualandaudio(ifenabledinAudiosettingmenu)alarmswillbetriggeredifeithercursororapointwithinthedesignatedareaishigherorlowerthanasettemperature.ThehighandlowsettemperaturesmaybeadjustedbymeansofDirectaccessbutton1andthenavigationbuttons.

4.4.3 Menu Camera Settings

CaptionModeSelect“On”toenabletheadditionofatextcaptionwhensavinganimage.Optionswillthenbedisplayedwhensavinganimage,tobeselectedbymeansofHotkey2andHotkey3.

Auto OffSelect“5Mins”,“10Mins”,“20Mins”toallowthecameratoswitchitselfoffafteradefinedperiodofinactivityinordertosavepower.AlwaysonmeansthattheAutoofffunctionisdisabled.

Camera ResetSelectDirectaccessbutton3torestorethefactorysettings.

Sequence recordThisoptionallowstheusertosaveimagestothemicroSDcard.Whensequencerecordingisselected,itispossibletosaveimagesevery05seconds,10seconds,20seconds,30seconds,1minute,2minutes,5minutes,10minutes,20minutes,30minutesand1hour.Thealarmsettingworksinconjunctionwiththetemperaturealarmfunction.Whenthetemperatureinthesceneexceedstheuserdefinedhighalarmthresholdtemperatureanimageissaved.Equallywhenthetemperatureinthescenegoesbelowtheuserdefinedlowalarmthresholdtemperatureanimageissaved.

Page 25: SKF TKTI 20 - Reliability Direct

25SKF TKTI 20

Sequences CapturesOncesequencerecordhasbeenselectedtheusercanchoosehowmanyimagestosavefromachoiceof:10,50,100,500or1000.

4.4.4 Menu Audio Settings

Imager SoundsSelect“Off”tomuteallsounds.

Voice AnnotationSelect“Session”toaddavoicemessageatthestartofasetofimages(Asessionendswhentheimagerisswitchedoff).Select“Individual”toaddavoicemessagetoeachsavedimage.Select“Combined”toaddacommonvoicemessageatthestartofasetofimagesandaddadditionalcommentsforeachimage.Ifsessionorcombinedisselected,thevoicemessageisrecordedintheaudiosettingsbypressingDirectaccessbutton3.RecordingisstoppedbypressingDirectaccessbutton3again.Directaccessbutton2canbeusedtoplaybacktherecordedmessage.Directaccessbutton3canbeusedtore-recordifnecessary.

Voice PlaybackSelect“Speaker”or“Headset”forthedesiredmethodofaudibleoutputs.

VolumeSelectthevolumeoftheaudibleoutputsfrom1to9.

Page 26: SKF TKTI 20 - Reliability Direct

26 SKF TKTI 20

4.4.5 Menu Image Browser

Thesavedimagesareshownonthescreenwiththemostrecentlysavedimagefirst.

Selectthedesiredimagebymeansofthenavigationkeys.TodisplaytheselectedimagepressDirectaccessbutton3.TodeletetheselectedimagepressDirectaccessbutton2,toconfirmdeletionpressDirectaccessbutton3.

Whenastoredimageisdisplayed,pressDirectaccessbutton3toreturntoliveimaging.

Page 27: SKF TKTI 20 - Reliability Direct

27SKF TKTI 20

4.4.6 Menu Date & Time Settings

a) Usetheleft/rightbuttonstonavigateinthismenu,theitemthatcanbechangedishighlightedinred.Inthepictureabovetheday(DD)14ishighlighted.

Thedatedisplayedformatcanbechangedbypressingdirectaccesskeynumber2

b) Usetheup/downbuttonstochangethevalue.

• Minute.(00to59.)• Hour.(00to23in24hourformat,amorpmin12hourformat.)• 12houror24hourclock.• Day.(01to31.)• Month.(01to12.)• Year.(2000to2099.)• Format(DD-MM-YYYY,MM-DD-YYYY,YYYY-MM-DD).

4.4.7 Menu Language Selection

Inthismenuyoucanselectthelanguageinwhichyouwanttousethecamera.

Page 28: SKF TKTI 20 - Reliability Direct

28 SKF TKTI 20

Theiconwithayellowbordershowstheselectedlanguage.

PressDirectaccessbutton3toselect.

Languages

French Korean

German Japanese

Italian Turkish

Spanish Russian

Portuguese Czech

Chinese

Page 29: SKF TKTI 20 - Reliability Direct

29SKF TKTI 20

4.4.8 Menu Display Settings

LCD brightnessSelectfrom1(low)to9(high)tocontrolthescreenbrightness.Alowsettingwillcontributetosavebatterylife

ZoomAllowstheusertodigitallyzoomintotheimage.OptionsareX1,X2andX4zoom.

Cursor TypeChoosetodisplayatabularCursororatagnexttothecursor.

εChoosewhetherornottodisplaythereflectedtemperature.Onlyapplicablewhenemissivityisselectedtobelessthan1.

Temp. Units.Choosewhetherornotthetemperatureunits(°Cor°F)aredisplayedonthescreen.

Colour ScaleChoosewhetherornotthecolourscaleistobedisplayed.

Icon TimeoutTheIconsabovethe4directaccesskeyscanbedisplayedeithercontinuouslyorfor10secondsafterthelastkeyispressed.

Page 30: SKF TKTI 20 - Reliability Direct

30 SKF TKTI 20

5. Saving and annotating pictures

5.1 Saving a picture

Tosaveapicturepressthesavebutton(seefigure6).Afterpressingyoucanadd‘voicemessage’and‘caption’tothepicture.Note:microSDcardshouldbeinsertedforsavingapicture(seefigure2).

5.2 Adding a voice message

WhensavinganimagewithIndividualVoiceAnnotationturnedon(see4.4.4).Thereyoucansaveavoicemessagewitheachimage.Thevoicerecordingcanbeplayedbackwhenviewingthesavedimagesinthebrowser(seechapter4.4.5).

a) YoucanattachavoicemessagetothissavedimagePressDirectaccessbutton2forno.PressDirectaccessbutton3foryes.PressDirectaccessbutton4tocancelandreturnto‘live’

b) Startvoicerecording?

PressDirectaccessbutton3foryes.PressDirectaccessbutton4tocancelandreturnto‘live’

c) StopRecording

PressDirectaccessbutton3tostop.PressDirectaccessbutton4tocancelandreturnto‘live’

d) OptiontoRe-recordthevoicemessagePressDirectaccessbutton1forreplayingyourvoicemessagePressDirectaccessbutton2forYesPressDirectaccessbutton3forNoPressDirectaccessbutton4tocancelandreturnto‘live’

Page 31: SKF TKTI 20 - Reliability Direct

31SKF TKTI 20

5.3 Adding a Caption

WhensavinganimagewithCaptionModeturnedon(see4.4.3),youcanattachatextcaptiontoeachimage.

a) YoucanattachavoicemessagetothissavedimagePressDirectaccessbutton2forno.PressDirectaccessbutton3foryes.PressDirectaccessbutton4tocancelandreturnto‘live’

1. Usetheup/downarrowbuttonstocyclethroughlettersandnumbersuntiltheonerequiredappears.Theavailablesymbolsare:ABCDEFGHIJKLMNOPQRSTUVWXYZabcdefghijklmnopqrstuvwxyz0123456789

2. Usetheleft/rightarrowbuttonstomovetothenextspaceandrepeattheabovestepuntilthecaptioniscompleted.

3. PressDirectaccessbutton2toclearthewholemessage.

4. PressDirectaccessbutton3tosavetheimageandcaption

Note: Captionscannotbeviewedwiththesavedimagesinthebrowser.ThecaptionscanbeviewedusingthePCsoftware(formoredetailedinformationseeIFUTKTI20ThermalCameraSuiteSoftware(MP5386)).

Imagewithvoicemessageisbeingsaved.

Boththermalandvisibleimagesarenowsavedwithannotationsandcaptionifrecorded.

Page 32: SKF TKTI 20 - Reliability Direct

32 SKF TKTI 20

6. Technical specifications

Performance

Fieldofview(FOV) 20ºx15º

SpectralResponse 8to14µm

Sensitivity NETD≤80mK(0,08ºC)@23ºCambientand30°Cscenetemperature

Detector 160x120pixelarray(interpolatedto320x240)uncooledmicrobolometer

Framerate 8Hz

FocalRange Manualfocusfrom0,3m(12in)toinfinity

Image Storage

Number Upto2500imageson2GBmemorycardsupplied

Medium MicroSDCard

Display 31⁄2”colourLCDwithLEDBacklight.8colourpalettes.Thermalimagesorvisibleimagesormixedthermalandvisibleimagesincludingpictureinpicture.

LaserPointer AbuiltinClass2laserissuppliedtohighlightthereferencepixel

Measurement

Temperaturerange -10ºCto+250ºC(+14ºFto+482ºF)

Radiometry Fourmoveabletemperaturemeasurementcursorsgivingautomatictemperaturedifferencemeasurementandautolockingontohottestandcoldestpoints

EmissivityCorrection Userselectable0,1to1,0instepsof0,01withreflectedambienttemperaturecompensation.Thefourmeasurementcursorscanhaveindividualemissivityvaluesassignedtothem.

Accuracy Thegreaterof±2°Cor±2%ofreadingin°Cfortheoperatingtemperaturerangeof-15°Cto+45°C.

Page 33: SKF TKTI 20 - Reliability Direct

33SKF TKTI 20

Camera Power Supply

Battery RechargeableLithium-Ionfieldreplaceable

Operationtime Upto5hourscontinuousoperation

ACoperation USBACpoweradaptorsupplied

Mechanical & Environment

Housing ImpactResistantPlasticwithover-mouldedsoftplastic

Dimensions 130x95x220mm(5.1x3.74x8.66in)

Weight 0.80kg(1.76lbs)

Temperatureoperating&storageRange

-15°Cto50°C(5°Fto122°F)

Humidity 10%to90%noncondensing/-20°Cto70°C(-4°Fto158°F)

IPrating IP54

Software SKFTKTI20ThermalCameraSuite:Advancedimageranalysisandreportwritingsoftware

ComputerRequirements

PCwithoneofthefollowingoperatingsystems:WindowsXP,VISTA,orWindows7

Page 34: SKF TKTI 20 - Reliability Direct

34 SKF TKTI 20

Page 35: SKF TKTI 20 - Reliability Direct

35SKF TKTI 20

Page 36: SKF TKTI 20 - Reliability Direct

SKF Maintenance Products

The contents of this publication are the copyright of the publisher and may not be reproduced (even extracts) unless prior written permission is granted. Every care has been taken to ensure the accuracy of the information contained in this publication but no liability can be accepted for any loss or damage whether direct, indirect or consequential arising out of use of the information contained herein.