rasio neuro pati

5
692 A Pocket-size Device to Detect Autonomic Neuropathy Niels Ejskjaer, M.D., Ph.D., 1  Jesper Fleischer, M.Sc., BME, 2  Jacob Fleischer, M.D., 3 Poul Erik Jacobsen, M.D., 4  Per Logstrup Poulsen, M.D., DMSci, 1  Hans Nygaard, DMSci 5 Author Afliations: 1 Aarhus University Hospital, Department of Medicine M (Diabetes & Endocrinology) and Clinical Research Units, Aarhus University Hospital, Aarhus, Denmark, 2 Medicus Engineering Ltd, The Engineering College of Aarhus, Aarhus, Denmark, 3 Soendersoe General Practice, Soendersoe, Denmark, 4 Department of Endocrinology, Aarhus University Hospital, Aalborg, Denmark, 5 Center of Biomedical Engineering, The Engineering College of Aarhus, Aarhus, Denmark Abbreviations:  (DAN) diabetic autonomic neuropathy, (DSP) digital signal processor, (E:I) inhalation/exhalation ratio, (ECG) electrocardiogram, (HRV) heart rate variability, (SD) standard deviation Keywords:  autonomic neuropathy, cardiac, complications, diabetes, neuropathy, risk stratication Corresponding Author: Niels Ejskjaer, M.D., Ph.D., Aarhus University Hospital, Department of Medicine M (Diabetes & Endocrinology) and Clinical Research Units, Aarhus University Hospital, Denmark, email address [email protected]  Journal of Diabetes Science and Technol ogy  Volume 2, Issue 4, July 2008 © Diabetes Technology Society Abstract Background: Diabetic autonomic neuropathy (DAN) is a very frequent complication in the diabetic population (type 1 and type 2 diabetes), and patients may suffer debilitating symptoms from various organ systems. In the less symptomatic and even in the asymptomatic condition it severely impacts health. Testing for DAN is currently time-consuming and costly due to the technical setups available today, therefore the examination may not be offered regularly. The purpose of this study was to evaluate the clinical performance of a pocket-size device for detecting DAN by measuring heart rate variability (HRV).  Method : Ten healthy young males and eight type 1 diabetes patients suffering symptomatic DAN were selected. The standardized spectral analysis equipment VariaPulse TF3 ®  (Sima Media, Olumouc, Czechoslovakia) was used as a reference method for evaluating a prototype of the pocket-size device according to a specied protocol. HRV, inhalation/exhalation ratio (E:I) (deep breathing test), and 30:15 ratio (response going from lying to standing) were measured using both methods. Statistical calculations were performed. Results: The correlation between the two devices was R 2  = 0.98 and R 2  = 0.81 when 30:15 ratio and E:I were measured, respectively. Bland-Altman plots showed suitable agreement between the two devices, substantiated by 95% limits of agreement of the differences of ±0.014 and ±0.033 when 30:15 ratio and E:I were measured, respectively. Conclusions: The pocket-size device was fully interchangeable with the hitherto-used, research-based setup. It proved highly suitable for ambulatory testing of autonomic nervous function and may facilitate screening for DAN according to Danish and international recommendations.  J Diabetes Sci Technol 20 08;2( 4):692-696 NEUROLOGICAL AND MICROVASCULAR FUNCTION

Transcript of rasio neuro pati

Page 1: rasio neuro pati

8/13/2019 rasio neuro pati

http://slidepdf.com/reader/full/rasio-neuro-pati 1/5

692

A Pocket-size Device to Detect Autonomic Neuropathy

Niels Ejskjaer, M.D., Ph.D.,1  Jesper Fleischer, M.Sc., BME,2 Jacob Fleischer, M.D.,3

Poul Erik Jacobsen, M.D.,4 Per Logstrup Poulsen, M.D., DMSci,1 Hans Nygaard, DMSci5

Author Afliations:  1Aarhus University Hospital, Department of Medicine M (Diabetes & Endocrinology) and Clinical Research Units, AarhusUniversity Hospital, Aarhus, Denmark, 2Medicus Engineering Ltd, The Engineering College of Aarhus, Aarhus, Denmark, 3Soendersoe GeneraPractice, Soendersoe, Denmark, 4Department of Endocrinology, Aarhus University Hospital, Aalborg, Denmark, 5Center of Biomedical EngineeringThe Engineering College of Aarhus, Aarhus, Denmark 

Abbreviations:  (DAN) diabetic autonomic neuropathy, (DSP) digital signal processor, (E:I) inhalation/exhalation ratio, (ECG) electrocardiogram,(HRV) heart rate variability, (SD) standard deviation

Keywords: autonomic neuropathy, cardiac, complications, diabetes, neuropathy, risk stratication

Corresponding Author: Niels Ejskjaer, M.D., Ph.D., Aarhus University Hospital, Department of Medicine M (Diabetes & Endocrinology) andClinical Research Units, Aarhus University Hospital, Denmark, email address [email protected]

 Journal of Diabetes Science and Technology

 Volume 2, Issue 4, July 2008© Diabetes Technology Society

Abstract

Background:

Diabetic autonomic neuropathy (DAN) is a very frequent complication in the diabetic population (type 1and type 2 diabetes), and patients may suffer debilitating symptoms from various organ systems. In the lesssymptomatic and even in the asymptomatic condition it severely impacts health. Testing for DAN is currently

time-consuming and costly due to the technical setups available today, therefore the examination may not beoffered regularly. The purpose of this study was to evaluate the clinical performance of a pocket-size devicefor detecting DAN by measuring heart rate variability (HRV).

 Method:

Ten healthy young males and eight type 1 diabetes patients suffering symptomatic DAN were selected. Thestandardized spectral analysis equipment VariaPulse TF3® (Sima Media, Olumouc, Czechoslovakia) was used asa reference method for evaluating a prototype of the pocket-size device according to a specied protocol. HRV,inhalation/exhalation ratio (E:I) (deep breathing test), and 30:15 ratio (response going from lying to standing)were measured using both methods. Statistical calculations were performed.

Results:

The correlation between the two devices was R2 = 0.98 and R2 = 0.81 when 30:15 ratio and E:I were measured,respectively. Bland-Altman plots showed suitable agreement between the two devices, substantiated by95% limits of agreement of the differences of ±0.014 and ±0.033 when 30:15 ratio and E:I were measured,respectively.

Conclusions:

The pocket-size device was fully interchangeable with the hitherto-used, research-based setup. It proved highlysuitable for ambulatory testing of autonomic nervous function and may facilitate screening for DAN accordingto Danish and international recommendations.

 J Diabetes Sci Technol 2008;2(4):692-696

NEUROLOGICAL ANDMICROVASCULAR FUNCTION

Page 2: rasio neuro pati

8/13/2019 rasio neuro pati

http://slidepdf.com/reader/full/rasio-neuro-pati 2/5

693

 A Pocket-size Device to Detect Autonomic Neuropathy Ejskjae

www.journalofdst.orgJ Diabetes Sci Technol Vol 2, Issue 4, July 2008

Heart rate variability (HRV) measures are animportant means of determining cardiac autonomicfunction.1  Detecting and measuring autonomicdysfunction in the diabetic population is of particularinterest and importance. First of all, it may identify

patients at risk of developing late complications ofdiabetes, some of which may be threatening to quality oflife and general health. This may in turn enable earlierintervention and treatment for the individual patient.2 Diminished HRV is associated with an increased risk ofcoronary heart disease and death, and it is hypothesizedthat low HRV is indicative of poor general health. 3  Ofparticular clinic interest within the diabetic populationhas been “the sudden death in bed syndrome” whichhas been connected to diabetic autonomic neuropathy(DAN).4  It has been shown that low heart rate variabilityis a risk factor for sudden cardiac death in type 2

diabetes.5  HRV measures are important diagnostic toolsfor the physician in reaching a correct diagnosis anddetermining whether autonomic dysfunction plays a rolein the patient’s symptoms.

In spite of the importance of determining DAN, testinghas rarely been available in hospitals and outpatientsettings, and then primarily for research use. Possibleexplanations for this may be that the apparatus availablehas been both very time- and cost-consuming to useand of dimensions not allowing bedside testing. Theseare major obstacles to general testing of the diabetic

populations, as is recommended,2,3  and the introductionof the pocket-size device described in this article mayoffer solutions to overcome these barriers.

Therefore, the purpose of this study was to evaluate theclinical performance of a pocket-size prototype devicefor detecting DAN by measuring HRV.

Materials and Methods

The clinical evaluation of the new device was performedon 10 healthy volunteers and 8 type 1 diabetes patients

suffering symptomatic DAN, using recordings fromVariaPulse TF3®  (Sima Media, Olumouc, Czechoslovakia)as reference (Table 1). All type 1 diabetes patients wererecruited from the neuropathy clinic in the department ofmedicine. All demonstrated two or more of the followingsymptoms of autonomic dysfunction: gustatory sweating,postural hypotension, diabetic gastroparesis, diabeticdiarrhea, or neurogenic bladder.

The pocket-size prototype device and clinical setup isshown in Figure 1a. The volunteer places his left and

right hands on the two separate metal electrodes, andthe device records an electrocardiogram (ECG) signafrom which the heart rate and heart rate variabilityare calculated and shown in the display. An onboardgraphical user interface guides the user through the

standard tests to assess autonomic function. The patientswere instructed to follow the graphical elements on thedisplay. In the deep breathing [inhalation/exhalationratio (E:I)] test, the graphical user interface visualizes the breathing pattern with a continuously moving bar anda short text message. When the bar moves up, the text

“Breathe in” appears, and when the bar moves down thedisplay shows the text “Breathe out.”

Table 1.Subject Characteristics

Parameters Control DAN patients

Number of subjects 10 8

 Age (years)* 34 ± 4 43 ± 12

Systolic/diastolic (mm Hg)* 129/79 ± 12/10 141/91 ± 20/17

Heart Rate (beats/min)* 56 ± 9 82 ± 11

* Mean ± standard deviation

A bedside screening device warrants a user-friendlyinstrument that does not require substantial time forpreparation and data analysis. The present prototypedevice was optimized for easy ad-hoc bedside

examination requiring no preparation before use. Forexample, the integrated metal electrodes were speciallydesigned for dry operation, and therefore need no gelThe device is based on two-electrode ECG recordingsmeasured by two hand electrodes, and is powered bytwo 1.5-v AAA batteries to secure mobility and patientsafety.

The signal processing was performed automatically by a digital signal processor (DSP), where the discreteheartbeats were detected using a correlation algorithmadapting to individual R  wave shapes to account for biological variation. Ectopic heart beats were removedautomatically by digital ltering. The correlationalgorithm is an optimized version of an earlier clinicallytested device showing no signicant systematic error ofthe detecting software compared with standard ECGdevices. Correlation between the methods was R2  = 1.00and R2  = 0.99 when measuring heart rate and the timedomain HRV parameter, root mean square successivedifference, respectively.6  A Preliminary test on the MITBIH Database indicates a high precision.

Page 3: rasio neuro pati

8/13/2019 rasio neuro pati

http://slidepdf.com/reader/full/rasio-neuro-pati 3/5

694

 A Pocket-size Device to Detect Autonomic Neuropathy Ejskjae

www.journalofdst.orgJ Diabetes Sci Technol Vol 2, Issue 4, July 2008

circumstances by the same personnel, with onemeasurement followed immediately by the next.

To show that the recorded differences were not asystematic error of the prototype device, we madea control measurement with the VariaPulse TF3immediately after each recording with the prototype

device.

Results

The correlations were rst visualized in a scatter plot(Figure 2) and then calculated using the Pearson R2  testand the mean difference between the methods. A highdegree of correlation between the two methods is evidentwith Pearson R2  and mean difference values shown inTable 2.

Figure 1.  The prototype device. (a)  The user places her hands on thetwo electrodes, and heart rate and heart rate variability are measuredand shown in the display. The onboard user interface guides the userthrough the tests, in this case, the deep breathing test. (b)  Showsambulatory testing.

For logistical and practical reasons, measurements withthe two setups were not performed simultaneously. Itwas not possible to synchronize the two devices, whichof course may lead to differences in the recorded datacaused by the time difference between measurements.However, testing was performed under identical

Figure 2.  Prototype device and stationary equipment, with line oequality.

Page 4: rasio neuro pati

8/13/2019 rasio neuro pati

http://slidepdf.com/reader/full/rasio-neuro-pati 4/5

695

 A Pocket-size Device to Detect Autonomic Neuropathy Ejskjae

www.journalofdst.orgJ Diabetes Sci Technol Vol 2, Issue 4, July 2008

Note that Figures 2–4 have different scales for the A plotcompared to the B  plot because of the great differencein the recorded data between the control group and thediabetic group.

The high correlation between the two devices indicatesthat HRV measurements by the pocket-size device and

the stationary VariaPulse TF3 device are highly correlated, but not necessarily in agreement.7,8  Bland-Altman plots

Table 2.Pearson R2  and Mean Difference

Control Group Patients With DAN

R2 Mean

difference  R2 Mean

difference

30:15 ratio 0.92 -0.031 0.98 0.003

E:I 0.81 -0.033 0.81 0.001

Table 3.Limit of agreement

Control Group Patients With DAN

±1.96·SD ±1.96·SD

30:15 ratio ±0.17 ±0.014

E:I ±0.14 ±0.033

Figure 3.  E:I test: difference of average of prototype device andstationary equipment, with mean and 95% limits of agreement.

Figure 4.  30:15 test: difference of average of prototype device andstationary equipment, with mean and 95% limits of agreement (6patients only; 2 failed due to amputation/poor health).

were applied to examine the agreement between the twomethods. Figure 3  and Figure 4  depict a high level ofagreement between the novel device and VariaPulse TF3This is substantiated by 95% limits of agreement [±1.96standard deviation (SD)] of the difference showed inTable 3.

The limits of agreement between the two devices are±0.14, indicating that 95% of the difference in E:I was

Page 5: rasio neuro pati

8/13/2019 rasio neuro pati

http://slidepdf.com/reader/full/rasio-neuro-pati 5/5

696

 A Pocket-size Device to Detect Autonomic Neuropathy Ejskjae

www.journalofdst.orgJ Diabetes Sci Technol Vol 2, Issue 4, July 2008

 between 0.107 and -0.173 for the healthy control groupwith a mean and standard deviation of 1.37 ± 0.15. Whenmeasuring E:I two times in a row with the VariaPulseTF3 (with a 1-minute break between the measurements),the limits of agreement were ±0.2, indicating a differencein E:I between -0.14 and 0.27 for the same healthy controlgroup.

The limits of agreement between the two devices in thegroup of DAN patients with a response going from lyingto standing (30:15 ratio) measured in the range of 1.05 ±0.04 were ±0.014, indicating that 95% of the difference inthe 30:15 ratio was between 0.01 and -0.017 for this group.These results indicate a good agreement between the twomethods.

Discussion

Traditionally, DAN screening devices require manual

data analysis and attachment of electrodes or chest beltto the patient beforehand and removal afterward, whichprolongs the examination unnecessarily, resulting inincreased time consumption and cost. To overcomethese obstacles to regular testing, we have examined anew pocket-size device. With only a short introductionto the device, and especially the graphical and textualinformation on the display, the patients were able toperform the tests correctly.

We admit that only two measures of HRV (E:I and 30:15ratio) have been tested and not the Valsalva maneuver.Ideally, all three tests should be performed in a much broader investigation in the diabetic population. Thenew version of the device is able to measure the Valsalvamaneuver, and results will be published accordingly.

In statistical terms, we have shown that the minor timelapse between testing one apparatus and the other is ofno signicance. Furthermore, we have shown that thetested pocket-size device is fully interchangeable withthe well-known stationary setups when used for clinicalexaminations, both in practical and statistical terms.

Conclusion

A pocket-size prototype device for estimating cardiacautonomic function by measuring HRV was comparedto stationary equipment used for diabetic autonomicfunction testing in a pilot investigation with a limitednumber of patients in our department. A statisticallysignicant degree of correlation and agreement

 between measurements using the two setups was found,suggesting that this new pocket-size device is suitable

for point-of-care screening for diabetic autonomicdysfunction and may facilitate screening according toDanish and international recommendations.

Acknowledgements:

The authors would like to thank Medicus Engineering Ltd, AarhusDenmark, for funding the development of the present prototypeBiotechnician Merete Moeller at the Clinical Research Department of

Department of Medicine M, Aarhus University Hospital, is thankedfor valuable assistance. Volunteers and patients are likewise warmlythanked for their time and cooperation.

Funding:

Equipment was made available by Medicus Engineering Ltd, AarhusDenmark.

Disclosure:

 Jesper Fleischer is co-founder of Medicus Engineering Ltd, AarhusDenmark.

References:

Ewing DJ, Martyn CN, Young RJ, Clarke BF. The value of

cardiovascular autonomic function tests: 10 years experience indiabetes. Diabetes Care 1985;8(5):491-8.

Vinik AI, Erbas T. Recognizing and treating diabetic autonomicneuropathy. Cleve Clin J Med. 2001;68(11):928-30, 932, 934-44.

Dekker JM, Crow RS, Folsom AR, Hannan PJ, Liao D, Swenne CASchouten EG. Low heart rate variability in a 2-minute rhythm strippredicts risk of coronary heart disease and mortality from severacauses: the ARIC Study. Atherosclerosis Risk In CommunitiesCirculation. 2000;102(11):1239-44.

Vinik AI, Maser RE, Mitchell BD, Freeman R. Diabetic autonomicneuropathy. Diabetes Care. 2003;26(5):1553-79.

Kataoka M, Ito C, Sasaki H, Yamane K, Kohno N. Low hearrate variability is a risk factor for sudden cardiac death in type 2diabetes. Diabetes Res Clin Pract. 2004;64(1):51-8.

Kristiansen NK, Fleischer J, Jensen MS, Andersen KS, Nygaard HDesign and evaluation of a handheld impedance plethysmographfor measuring heart rate variability. Med Biol Eng Comput2005;43(4):516-21.

Bland J.M, Altman, DG. Statistical methods for assessingagreement between two methods of clinical measurement. Lance1986;1(8476):307-10.

Bland JM, Altman DG. Comparing methods of measurement: whyplotting difference against standard method is misleading. Lance1995;346(8982):1085-7.

1.

2.

3.

4.

5.

6.

7.

8.