Mass Spectrometry Fault finding and trouble shooting
description
Transcript of Mass Spectrometry Fault finding and trouble shooting
![Page 1: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/1.jpg)
Mass SpectrometryFault finding and trouble shooting
Or: Giving the bullet to those pesky faults
![Page 2: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/2.jpg)
Disclaimer
3000 volts can kill
Protect your eyes
230 Volts can kill
Mass spectrometers can cause pain
Mental pain!
Please beware that any work on the instrumentis carried out at your own risk. Isomass Scientific cannot be held responsible for injury or damage caused by following these techniques.
Hot solderingirons can burn
![Page 3: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/3.jpg)
Salesman’s view of solving instrument problems
New!
![Page 4: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/4.jpg)
Reality
• Budget shortfalls– Not enough money to buy a new instrument– Not enough money to pay for service
![Page 5: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/5.jpg)
What to do• Learn how to look after your own instrument– Go on the manufactures training courses– Ask questions during the installation and any visit
made by an engineer– Regularly attend ASITA– Read the manual and other books– Take an electronics course
![Page 6: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/6.jpg)
Fault Finding• Use the instrument supplier as a resource– The engineers are trained on the instrument– The engineers want to help– The engineers will have access to more
information• Take your time– Fault finding is not a race– More intervention usually leads to more damage– Take notes and pictures
![Page 7: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/7.jpg)
Fault Finding• Beware of dangerous voltages that might be present
in the unit under investigation– The 230 volt supply into most units can be fatal if touched– Source supplies are usually at high voltage, > 3kv
• Be careful when connecting or disconnecting meter probes– A badly placed probe can cause a short and further
damage• Double check connections and meter setting before
turning on the unit under test
![Page 8: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/8.jpg)
Fault Finding
• Check power supplies first– Fuses, check with a meter on resistance not just
visually– If a supply voltage is less than expected it could be
that the supply is damaged or the load is pulling the supply voltage down
– Regulated supplies should have very low a.c. ripple– All electronics can be seen as either a supply or
load
![Page 9: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/9.jpg)
Fault FindingThink before acting #1
5 minutes thought can save hours of work
![Page 10: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/10.jpg)
Fault FindingThink before acting #2
Think about the problem before acting
![Page 11: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/11.jpg)
Art is a science• Learn to draw– Draw a block diagram of the fault area– Write down the inputs and outputs of each block– Make a list of possible causes
![Page 12: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/12.jpg)
My block diagram
![Page 13: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/13.jpg)
Power Block Diagram
![Page 14: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/14.jpg)
A few simple techniques
• Swap– Exchange the suspected part
• Measure– Compared the measured parameter with the
expected value• Observe– Look to see what’s happening
• Write– Keep good notes
![Page 15: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/15.jpg)
Swap
Pirani reads 1e-3mbar
Pirani reads atmosphere
![Page 16: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/16.jpg)
Simplify
Pirani showingatmosphere after
10 minutes pumping
Use a blank at point 'A' totest the sub- system withinthe instrument as a whole
![Page 17: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/17.jpg)
Voltage Measurement
![Page 18: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/18.jpg)
Voltage Measurement
• Multimeter set to a.c. or d.c. voltage as required– Black lead in common socket– Red lead in V socket
• Voltage measured across one component measures across all components in parallel– Voltage is measured in parallel– Circuit is powered
• Connect red probe to one lead of the component and the black probe to the second lead– If meter is auto-ranging note the displayed multiplier
![Page 19: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/19.jpg)
Current Measurement
![Page 20: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/20.jpg)
Current Measurement• Multimeter set to a.c. or d.c. current as required
– Black lead in common socket– Red lead in mA or A socket as required, if in doubt start in A
socket• Multimeter must be in series with the component
– If necessary unsolder one lead from the board– Circuit is powered
• Connect red probe to the lead of the component and the black probe to the point the lead was previously connected to– If meter is auto-ranging note the displayed multiplier
• Often easier to measure voltage across a resister and use V=IR
![Page 21: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/21.jpg)
Resistance Measurement
![Page 22: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/22.jpg)
Resistance Measurement
• Multimeter set to resistance– Black lead in common socket– Red lead in Ω socket
• Component must be measured in isolation– Circuit is not powered– If necessary unsolder one lead from the board
• Connect red probe to one lead of the resistor and the black probe to the other lead– If meter is auto-ranging note the displayed multiplier
![Page 23: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/23.jpg)
Start with the obvious
If a light does not work....
Check the light bulb before dismantling thelight switch, changing the fuse in the basement etc.
![Page 24: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/24.jpg)
Start with the obvious
Two bulbs might have blown but it's more likely that there is a fault common to both e.g. the fuse in the basement
If two light bulbs do not work....
![Page 25: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/25.jpg)
Fault FindingTake good, clear notes when the instrument is working correctly
and notes during any fault finding procedure
Don't forget, "RTFM" !
![Page 26: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/26.jpg)
Power Control
230 Volts
Switched outletRelayControl
![Page 27: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/27.jpg)
Vacuum control
![Page 28: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/28.jpg)
Vacuum parts
![Page 29: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/29.jpg)
Vacuum Start-up
![Page 30: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/30.jpg)
Quick Rotary Pump Test
![Page 31: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/31.jpg)
Penning Gauge
![Page 32: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/32.jpg)
Turbo Pump Faults
![Page 33: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/33.jpg)
Vacuum Service
• Rotary pump– Change the pump oil every 6 months
• Turbo pump– Change the oil wicks yearly
![Page 34: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/34.jpg)
Ion Source Control
![Page 35: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/35.jpg)
Ion Source Electronics
~4A ~700µA~800µA
1.5mA
~4.2A
![Page 36: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/36.jpg)
Ion source
![Page 37: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/37.jpg)
Exploded Ion Source
![Page 38: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/38.jpg)
Source Cleaning
![Page 39: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/39.jpg)
Magnet Supply
![Page 40: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/40.jpg)
Magnet Screwdriver Test
![Page 41: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/41.jpg)
Inlets
![Page 42: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/42.jpg)
Dual Inlet Valve Control
![Page 43: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/43.jpg)
GC Combustion
![Page 44: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/44.jpg)
EA
![Page 45: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/45.jpg)
GasBench
![Page 46: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/46.jpg)
GasBench
![Page 47: Mass Spectrometry Fault finding and trouble shooting](https://reader035.fdocuments.in/reader035/viewer/2022081502/56816462550346895dd64031/html5/thumbnails/47.jpg)