Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

47
Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults

Transcript of Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Page 1: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Mass SpectrometryFault finding and trouble shooting

Or: Giving the bullet to those pesky faults

Page 2: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Disclaimer

3000 volts can kill

Protect your eyes

230 Volts can kill

Mass spectrometers can cause pain

Mental pain!

Please beware that any work on the instrumentis carried out at your own risk. Isomass Scientific cannot be held responsible for injury or damage caused by following these techniques.

Hot solderingirons can burn

Page 3: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Salesman’s view of solving instrument problems

New!

Page 4: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Reality

• Budget shortfalls– Not enough money to buy a new instrument– Not enough money to pay for service

Page 5: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

What to do• Learn how to look after your own instrument– Go on the manufactures training courses– Ask questions during the installation and any visit

made by an engineer– Regularly attend ASITA– Read the manual and other books– Take an electronics course

Page 6: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Fault Finding• Use the instrument supplier as a resource– The engineers are trained on the instrument– The engineers want to help– The engineers will have access to more

information• Take your time– Fault finding is not a race– More intervention usually leads to more damage– Take notes and pictures

Page 7: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Fault Finding• Beware of dangerous voltages that might be present

in the unit under investigation– The 230 volt supply into most units can be fatal if touched– Source supplies are usually at high voltage, > 3kv

• Be careful when connecting or disconnecting meter probes– A badly placed probe can cause a short and further

damage• Double check connections and meter setting before

turning on the unit under test

Page 8: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Fault Finding

• Check power supplies first– Fuses, check with a meter on resistance not just

visually– If a supply voltage is less than expected it could be

that the supply is damaged or the load is pulling the supply voltage down

– Regulated supplies should have very low a.c. ripple

– All electronics can be seen as either a supply or load

Page 9: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Fault FindingThink before acting #1

5 minutes thought can save hours of work

Page 10: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Fault FindingThink before acting #2

Think about the problem before acting

Page 11: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Art is a science• Learn to draw– Draw a block diagram of the fault area– Write down the inputs and outputs of each block– Make a list of possible causes

Page 12: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

My block diagram

Page 13: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Power Block Diagram

Page 14: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

A few simple techniques

• Swap– Exchange the suspected part

• Measure– Compared the measured parameter with the

expected value• Observe– Look to see what’s happening

• Write– Keep good notes

Page 15: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Swap

Pirani reads 1e-3mbar

Pirani reads atmosphere

Page 16: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Simplify

Pirani showingatmosphere after

10 minutes pumping

Use a blank at point 'A' totest the sub- system withinthe instrument as a whole

Page 17: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Voltage Measurement

Page 18: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Voltage Measurement

• Multimeter set to a.c. or d.c. voltage as required– Black lead in common socket– Red lead in V socket

• Voltage measured across one component measures across all components in parallel– Voltage is measured in parallel– Circuit is powered

• Connect red probe to one lead of the component and the black probe to the second lead– If meter is auto-ranging note the displayed multiplier

Page 19: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Current Measurement

Page 20: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Current Measurement

• Multimeter set to a.c. or d.c. current as required– Black lead in common socket– Red lead in mA or A socket as required, if in doubt start in A

socket• Multimeter must be in series with the component

– If necessary unsolder one lead from the board– Circuit is powered

• Connect red probe to the lead of the component and the black probe to the point the lead was previously connected to– If meter is auto-ranging note the displayed multiplier

• Often easier to measure voltage across a resister and use V=IR

Page 21: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Resistance Measurement

Page 22: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Resistance Measurement

• Multimeter set to resistance– Black lead in common socket– Red lead in Ω socket

• Component must be measured in isolation– Circuit is not powered– If necessary unsolder one lead from the board

• Connect red probe to one lead of the resistor and the black probe to the other lead– If meter is auto-ranging note the displayed multiplier

Page 23: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Start with the obvious

If a light does not work....

Check the light bulb before dismantling thelight switch, changing the fuse in the basement etc.

Page 24: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Start with the obvious

Two bulbs might have blown but it's more likely that there is a fault common to both e.g. the fuse in the basement

If two light bulbs do not work....

Page 25: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Fault FindingTake good, clear notes when the instrument is working correctly

and notes during any fault finding procedure

Don't forget, "RTFM" !

Page 26: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Power Control

230 Volts

Switched outletRelayControl

Page 27: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Vacuum control

Page 28: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Vacuum parts

Page 29: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Vacuum Start-up

Page 30: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Quick Rotary Pump Test

Page 31: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Penning Gauge

Page 32: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Turbo Pump Faults

Page 33: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Vacuum Service

• Rotary pump– Change the pump oil every 6 months

• Turbo pump– Change the oil wicks yearly

Page 34: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Ion Source Control

Page 35: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Ion Source Electronics

~4A~700µA~800µA

1.5mA

~4.2A

Page 36: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Ion source

Page 37: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Exploded Ion Source

Page 38: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Source Cleaning

Page 39: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Magnet Supply

Page 40: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Magnet Screwdriver Test

Page 41: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Inlets

Page 42: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

Dual Inlet Valve Control

Page 43: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

GC Combustion

Page 44: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

EA

Page 45: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

GasBench

Page 46: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.

GasBench

Page 47: Mass Spectrometry Fault finding and trouble shooting Or: Giving the bullet to those pesky faults.