Manual
Microbiological Saftey Cabinet
BSC-11IIA2X
PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
MRC.VER.01-11.13
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
1
Table of Content
Introduction............................................................................................................. 2
1. Application Range ............................................................................................ 3
2. Technical Parameters ........................................................................................ 4
3. Performance Index ............................................................................................ 5
4. Product Characteristic....................................................................................... 6
5. Structure ............................................................................................................ 9
6. Installation and Instruction ............................................................................. 15
7. Maintainance and malfunction analysis.......................................................... 19
8. Notes ............................................................................................................... 23
9. Label Instruction ............................................................................................. 26
10. Guarantee ................................................................................................... 27
Appendix A Wiring Diagram................................................................................ 27
Appendix B Air Flow Draft and Protected Area................................................... 29
Appendix C BSC-11IIA2X Packing List........................................................... 30
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
2
Introduction
Dear User:
Welcome to choose MRC Cabinet. Thanks!
MRC cabinet can offer the protection to the operator, environment and
experimental material. It was designed to avoid the infectious aerosol and splash
when operators work with infectious material such as primary culture, bacterium
and diagnostic samples.
MRC Cabinet is the necessary equipment of laboratory used in microbiology,
biomedicine, gene recombine, animal experiment and biological goods, especially
used in the environment of non-germ and non-dust where the operators should be
protected to culture germ, such as health care, pharmacy and research.
Hope our product can bring you best help.
In order to make you understand more clearly our MRC cabinet, please read this
manual carefully before use it. It is very important for you to use our instrument
correctly and safely. Please put the manual in appropriate position in order to use it
conveniently.
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
3
1. Application Range
MRC cabinet is one kind of fume hood with negative pressure and filtered that can
protect the operator, environment and product. The front sash with inflow can
protect the operator, the vertical downward and filtered layer airflow can protect
product, and the pollutional airflow after passing the HEPA (ULPA) filter can
protect the environment. When the MRC cabinet was used in the experiment on
trace volatile toxic chemicals and trace radionuclide as the subsidiary additives
microbiology, it is necessary to contact the suitable exhaust hood.
Working Environment:
(1). only for the indoor.
(2).Temperature: 15OC~35OC;
(3).Relative humidity≤75%;
(4).Atmospheric pressure: 75kPa~106kPa;
No frost, Gel, seeping, raining or sunshine.
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
4
2. Technical Specification:
Our product completely fit in the MRC cabinet requirement of the pharmaceutical
industry standard
Basic Technical Specification as follow:
Model
Technical
Specification
BSC-11IIA2X
External Size 1100×750×2250 mm
Internal Size 940×600×660 mm
Power Supply AC 220V±10%/50 Hz±1 Hz, AC 110V±10%/60Hz±1 Hz,
Power ≤1350 W
Total Exhaust Volume 350 m3/h
Down Airflow 0.25~0.5m/s
Inflow Air ≥0.5m/s
UV lamp 18W
Fluorescent 21W
Filter Efficiency 99.99%(Diameter :0.3μm)
Noisy ≤65dB(A)
Note:
(1) Consumed power include the operating zone load power(the load power can’t
exceed 500W);
(2) The company reserves the right to change the design of products.
3. Performance Index
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
5
3.1 MRC Cabinet Performance
Personnel Protection: the microorganism germ number ≤5CFU;
Product Protection: the microorganism germ number ≤5CFU;
Cross Contamination Protection: the microorganism germ number≤2CFU。
3.2 Body Leakage-proof
When the MRC cabinet air pressure up to 500Pa, the air pressure is no less than
450 Pa after kept 30min
3.3 HEPA filter integrity
For the scanning and detecting filter, the rate of leakage is not more than 0.01% in
any point, For the no-scanning and no-detecting filter test points, the rate of
leakage is no more than 0.005%。
3.4 Vibration
The net Vibration amplitude is no more than 5 rms, when the frequency is from 10
Hz to 10 kHz.
3.5 Illumination
The average illumination is not less than 650 lx,The measured value of every
illumination is not less than 430 lx.
3.6 Mechanical performance.
The MRC cabinet design and structure can resist the turn-turtle and transformation
caused by the external force, the deflexibility caused by the working table load
and the tip-over caused by the workload.
The work table doesn’t cause the permanent deformation after it was loaded 23 kg
pressure.
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
6
3.7The voltage value will be up to 1390V AC in 5s and keep not to breakdown for
5s.
Leakage Current≤10mA。
Ground Resistance≤0.1Ω。
4. Product Characteristic
4.1LCD Display
Liquid Crystal Display: Place liquid crystal among two parallel glasses, there are
many vertical and horizontal tiny wires among the two glasses. The principle is
mainly through power on or not to control the crystal molecules to change
directions so as to reflect the light and create images. The LCD display is low
power consumption, no electromagnetic radiation.
Picture 1
The LCD display technology adopted by us could directly reflect the working
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
7
conditions of the equipment, such as the effective working status of the filter. This
makes more intuitive sense to the operator.
4.2REMOTE CONTROL
MRC Cabinet be installed a special controller to reduce the inconvenience of
close operation. It is small size and light weight, which makes it is convenient
to control the MRC cabinet all functions in the 6 m front of MRC cabinet. It
adopts Japanese NEC remote controller’s chip with many advantages, such as
anti-interference function, far control distance, high control accuracy etc.
(Draft 1)
Remote Control Function:
1、POWER 2、SUB(RESERVE)
3、INSTALL TIMER
4、CONFIRM 5、CANCEL
6、+ 7、-
8、BLOWER
9、UV(STERILIZATION)
10、LIGHT 11、SOCKET
12、E/D 13、UP 14、DOWN
4.3AUTOMATIC CONTROL OF FRONT PANEL
Front panel is electrically controlled. With a controller, the operator can
control the up and down of front panel without touching it. The motor is
imported from Japan with various speeds, and it bears many advantages, such
as accurate speed adjustment, powerful tripping force, and smooth rotating
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
8
and so on.
4.4RESERVED TIMER
MRC Cabinets shall be disinfected no less than 30 minutes before working
with it. Therefore, the operator has to wait for 30 minutes everyday before
working. Our cabinets were manufactured with the time reserve function that
we can fix the time to turn on and reserve the relative functions. The cabinets
have gotten ready for work before the operator’s arrival to save time. The
range of time reserve is 0-99 hours.
4.5STRUCTURE
4.51. The left, right sides and backside walls of Biological Safety Cabinets are
negative pressure enclosure design to prevent the leak of products.
4.52. Use of the technology of cold rolling steel with 1.5mm duster is to
increase the intensity of frame.
4.53. The working zone made of stainless steel, easy operation.
4.54. The support stand made of metals with a protection coatings.
4.55. The control panel is touched-model switch to make more convenient
operation and pretty appearance.
4.6. Caution
Digital differential pressure display;
Digital airflow velocity display;
Audible and visual alarm system.
4.6.1. Glass windows ultra-high alarm: when the current window height is more
than setting height, the equipment will issue voice alarm, and the LCD starts
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
9
to blink .Now you need to adjust the front window to the setting height.
4.6.2. Filter pressure differential warning: If the air filter of supply and exhaust
can’t reach to the stipulate requirement, the equipment will issue a voice warning
to inform user of changing the filter for user’s security.
5. Structure
5.1Front glass door driving system
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
10
Front glass door driving system consists of door-motor, front glass door, traction
structure, limit switch etc.
5.2Air filtering system
Air filtering system is the uppermost system to ensure equipment functions. It
consists of air supply fan, exhaust fan, air supply filter and exhaust air filter.
The main function of air filtering system is pushing clean air into workroom
continuously, the vertical clean air speed in work room is comply to the standard
to ensure that the clean level can meet the filed standard. The exhaust air is
purified, meanwhile, environmental protection.
5.3UV Light
UV Lamp is specially used to sterilize, lies inside the working zone fixed at
stainless steel top-inside, so that this could ensure the maximum lighting working
zone.
5.4 Fluorescent Light
Fluorescent Light ensures a certain degree of illumination. Light source consists a
pipe-type energy-saving fluorescent lamp, outside the area; the average lighting
degree inside the working zone meets the standard requirement.
5.5Foot switch
By controlling foot switch, the glass door can be controlled flexibly up and down
within the glass door active range.
5.6 Control Panel
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
11
1. UV key、2. LIGHT key、3 Fan control key
4.Socket(SOCKET)、5.Glass door up moving key (UP)
6.Glass door down moving key(DOWN)、
7.The key to remove the alarm、
8.Power key(POWER)、9. Alarm status display、
10. Mute status display、11. Please replace filter、12. Temperature Display、13.
Working time display、
14. Humidity display、15. Glass windows down status display、16. Glass windows
up status display、17. Foot switch working status display、18. Uv lamp working
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
12
status display、19. Socket working status display20. Light working status display
21. Fan working status display 22. Exhaust filter differential pressure23. Inflow
velocity 24 Down flow filter differential pressure 25. Down flow velocity 26
Filters working time display 27 UV lamp working time display
5.6.1 LCD display window
The output part of Human-equipment exchange, which we can understand the
state of equipment and performance.
(2) Remote-control receive window
This window receive the order from remote control
(3) Press button lightly
The key point of Human-equipment exchange is to operate the equipment through
light-touch. When Remote control is out of work, this equipment can achieve the
basically functions by the button.
Power Button: POWER, in addition to reservation, in control of the other switch;
Light button: LIGHT, is the key to control the lamp lighting. Press every time, the
state of LCD and the lamp correspondingly change. That is, the light turn on to off
or turns off to on.
Sterilization Button: UV, the ultraviolet lamp controls switch. The state of LCD
and the lamp correspondingly change. That is, the light turn on to off or turns off
to on. The key acts only when the glass window closed.
Fan Control Button: in control of the fan working condition. Press every time, the
state of LCD and the lamp correspondingly change. The button cannot act when
the glass window closed.
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
13
Socket: SOCKET, is the control button of socket working situation. Every time you
press it, on / off situation and the relevant indication situation on LCD display will
change.
Glass Window up Button: UP, press up button sustaining, the glass window rises .To
height of 200mm or releasing the button, the glass window immediately stops.
Press again and the glass window rises again.
Glass Window down Button: DOWN. Keep pressing the down button, the glass
window will go down until 200mm to the bottom or you release the button. The
glass window will continue going down when you press the button again.
Mute button: MUTE. The alarm will stop when you press this button, and will
continue if you press again.
There are 8 usually-used buttons on the operational board. Reservation and timing
function should be operated by the remote control. Button existed on both
operational board and the remote control has the same function.
5.7. Airway tube
Ventilation vent of pressure difference sensor.
User must not block this ventilation vent, and nothing should be hanged on this
device.
5.8. Power lock
When power cord is connected with electricity, unlock power lock with key, and
device starts to work.
5.9. Remote control
5.91. Reservation(SUB)
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
14
a、 Plug in, open power locker, press subscribe key. (SUB);
b、 Through +、 - key to adjust time division, then press confirm key to
confirm.(CONFIRM);Then adjust time 10 division, hour division and 10
hour division using same method.
c、 After confirm time, select relative function key (for example UV), relative
display will be light.
d、 Press power key again(POWER), subscribe function will be started.
Subscribed time period begins count down, when timing shows 0, relative
function will be started.
5.92.Timer(INSTALL TIMER)
a、 Plug on power, open electrical socket, press down power key(POWER),
through select relative function key(eg: UV),relative function will be started.
b、 Press down timer key(INSTALL TIMER), through +、- key to adjust time
division, and then press confirm key to confirm(CONFIRM); Then adjust
time 10 division, hour division and 10 hour division using same method.
c、 After confirm time, timer function will be started. When count down to 0,
turnoff all the functions, cabinet will be on standby condition.
5.93. Other key’s function is the same as light touch key on control panel. Pls refer
to 5.6.3.
5.10. Water proof socket
Water proof socket is placed at right side of operating area, which can supply
power to inside equipments. User controls this supplying power or not through
this socket.(SOCKET)
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
15
(1)Total equipments power of socket should not exceed 500W;
(2)Water proof socket can only proof water when its front cover put
down. Once front cover is opened, this socket can not be called water proof
socket.
5.11.Fuse protection(Protective tube)
This cabinet equips includes general power fuse (Protective tube) and water proof
socket fuse (Protective tube), which locates near the electrical wire side at back
side of cabinet. Fuse (Protective tube)’s relative specification and model, pls refer
to 9.3.
6. Installation and instructions
6.1. Installation
6.11Remove all the package materials;
6.12Check if there is scratch, out of shape or foreign matter on cabinet outside
surface;
6.13Carefully check with accessories and materials, referring to packing list in this
instruction book;
6.14Move the whole cabinet to final placing position, where is nearer and
convenient to install;
Supporting frame is placed on cabinet top when transportation, take it
out before installation. Strictly prohibit putting bottom upside when moving,
strictly prohibiting putting id down or disassembling.
6.15Base stand assembling
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
16
Refer to Draft 5 to assemble base stand.
Picture 5
Take out M10×50 inner hexagonal cylinder bolt, flat gasket 10, and spring
gasket 10 from accessories. Pls refer to picture 5 to assemble base stand. It must
be assembled fasten and firmly.
6.16Connect main machine and base stand
Refer to Picture 6 to connect main machine and base stand.
Draft 6
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
17
1. Base stand connecting parts right、2. Base stand connecting parts left、3.
M10cover type nut、4. M10×55 hexagon-headed bolt+ flat gasket10+spring
gasket 10
Firm and fasten it according to Draft 6 is ok.
6.17Complete machine placing
Biological safety cabinet should be placed in air flow protected well area, avoiding
influence of air flow from ventilating system, air conditioning, door, window or
person moving. Test shows if other influence air flow exceeds suction air flow
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
18
speed at cabinet air intake side, room polluted air will enter into working area of
the cabinet. So place cabinet correctly, and make it at correct position is very
necessary. Treat well the relations between cabinet air exhaust and room inside air
exhaust or exhaust duct. Cabinet’s air exhausted from upper side of cabinet. When
place cabinet, air exhaust system should not be restricted or effected. Cabinet
should be placed at the lower reaches of the air flow. Room sides should be at least
300mm on each side of the cabinet, for convenient checking.
6.2. Instructions for use
6.21 Connect with 220V、50HZ alternating current power supply;
6.22 Open power locker, power on the cabinet, VFD twinkle 3 times, buzz 3 times
at the same time. And then cabinet will be on standby condition, waiting for
operator’s press key or remote control to input execute.
6.23 Press power key, pressure difference shows sending air pressure difference,
number value is 0. After press power key, following functions can be then realized:
light, UV sterilization, motor, send/exhaust wind display transform, socket,
electrical door up and down, timer; While subscribe must be down before pressing
power key(relative press key, function and operation, pls refer to 5.6 and 5.9)
UV sterilization function will not be selected when glass door open or
any other key pressed down
6.24Before using, down electrical door rail to the bottom, open UV light,
sterilizing above 0.5hour.
(1)When sterilizing, person should leave room, for protect eyes and skin,
avoiding damage from expose carelessly.
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
19
(2)UV light strength should be tested at regular intervals according to
supplier’s description. Test quarterly is suggested. Change new one when
disqualification.
6.25Move electrical door bottom rail at the 200mm height of operating surface,
open motor, after running 0.5hour, normal inside test operating can be done.
For guarantee operating safety, pls put in testing materials ahead of
time inside the cabinet, and keep electrical door rail at the 200mm height
of operating surface when testing.
6.26After usage, down electrical door rail to bottom line, opens UV light,
sterilizing 0.5hour above, and then shut down the cabinets.
7. Maintenance and Analysis of Common Failures
Because the statistics of operating time will directly affect the judgment of
maintenance needs, we recommend preparing a detailed record of operating time
for reference and inquiry when using!
7.1. Clean the cabinet surface
7.1.1. Clean the operating area surface
Rub the entire surface with a soft cotton cloth soaked in concentrated liquid soap,
wipe up soap with the cotton cloth or towel soaked in clean hot water or warm
water, and then dry the surface with another cotton cloth or towel rapidly.
For the contaminated or traces of the work surface, fluid tank, etc., use medical
alcohol or other disinfectant to wipe.
Disinfectants can not damage for 304 stainless steel.
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
20
7.1.2. Clean the the external surface and glass door.
As to any of the non-abrasive household cleanser, use a soft cotton cloth or towel
to wipe.
7.2. Comprehensive maintenance period
We suggest comprehensive maintenance period is one year or 1000 hours of work.
7.3. Maintenance methods
7.3.1. Daily or weekly maintenance
(1) Disinfect and clean operating area (see 7.1.1);
(2)Clean the external surface and glass door around the operating area (see 7.1.2);
(3) Check the various functions of equipment;
(4) Record this maintenance result;
7.3.2Monthly maintenance
(1) Clean the the external surface and glass door. (See 7.1.2);
(2)Wipe the worktop surfaces, wall surface of operating area (excluding the well
air net of the top of operating area), and the inner surface of glass door with 70﹪
medical alcohol or household bleach diluted 1:100 (ie, 0.05% sodium
hypochlorite); Then use the sterile water to wipe again in order to eliminate the
rest chlorine.
(3) Check the various functions of equipment;
(4) Record this maintenance result;
7.3.3Annual maintenance
(1)Check the two conveyor belts of front glass door drive unit, and ensure that
their tightness is coincident.
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
21
(2)Check the UV lamp and fluorescent lamps.
(3)Apply for testing the overall performance of cabinet on an annual basis to
ensure the performance safety. User is responsible for testing costs.
(4)Record this maintenance result.
7.4 Analysis of Common Failures
Please confirm whether the power is connected or not, whether the power cord is
obvious damaged or not, whether the fuse is good or not, and whether the power
locks are in the open state or not before fault diagnosis.
Symptom Fault analysis
UV lamp not bright or abnormal.
• Whether lamp foot connect is good or not.
• Whether tube is good or not.
• Whether glass doors are all down or not.
Light not bright or abnormal. • Whether light foot connect is good or not.
• Whether tube is good or not.
Buttons not working or abnormal. • Whether there is obvious damage to
buttons.
Remote control not working.
• Whether the remote control batteries are
charged.
• Whether there is obvious damage to remote
control.
Equipment not electricity
• Whether the power connect is good or not.
• Whether there is obvious damage to power
code.
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
22
• Whether fuse is good or not.
• Whether the power locks are in the open
state.
The socket of operating area
without electricity
• Whether socket button starts up or not.
• Whether socket is good or not.
(1)The above electrical parts must be operated by a qualified electrician in safey
conditions (cutting off power supply). The other parts are not allowed to remove;
otherwise the user should take responsibility by them;
(2)When failures not in the above occur, and the operator can’t solve, please
notify our maintenance department immediately. For your safety, please do not
maintenance equipment by yourself;
(3)The maintenance of this equipment is undertaken by trained and recognized
technicians;
(4)If you need to order parts, contact our technical service department, and please
indicate the model and serial number of the cabinet purchased.
7.5 Storage conditions
Safety cabinet is stored in warehouse in which relative humidity is no more than
75%, temperature stays below 40OC, ventilation is good with no acid, alkali and
other corrosive gases, and its storage period should be not exceeding one year. If
the period is more than one year, check the cabinet out of the box, and let the
qualified cabinet enter circulation.
8. Notice
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
23
8.1Before connecting the AC power supply, ensure inputted voltage of safety
cabinet is the same as the supply voltage and both with stability, and ensure that
the rated load of power socket is not less than requirements; a third leg
grounded plug, which can only be used with a grounded power socket, is a
safety device. If you can’t plug in socket, turn to electrician for assistance.
8.2Slow-moving: In order to avoid affecting the normal air flow state, the operator
who moves his arms in and out of the cabinet is required to be careful to
maintain the integrity of the front opening air. Arms should be slowly and
perpendicular while coming in and out of the front opening. Hands and arms
stretching to the cabinet wait about one minute in which cabinet adjusts itself
and makes the air inside sweep the surface of the hands and arms, then hands
and arms can process the items. In order to minimize the times of arms into and
out of the opening, perpare all the necessary items in the cabinet before starting
the experiment;
8.3 Moving principles of different samples in cabinet: When two or more samples
need to be moved, be sure that low-polluting samples move to high-polluting
samples. Movement of items should also follow the principles of slow-moving.
8.4Items placed in parallel: items should be placed in the cabinet as the word
showing horizontally, avoiding cross-contamination between items caused by
return air, and avoiding blocking the return air grille in the back;
8.5 In order to avoid items being sucked into the vacuum suction duct or the
blower, do not place soft, slight items (for example: soft tissue) on the surface
during the using process;
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
24
8.6The maximum weight of items placed in the cabinet is no more than
23Kg/25×25cm2;
8.7Avoid vibration: avoid using vibration equipment (eg centrifuges, vortex
oscillator, etc.) in the cabinet. Vibration would make granular materials
accumulating in the membrane shake off, leading to lower cleanliness of
operating area. The failure of Balance in the front operating surface can also
cause the pollution on the operator;
8.8Open fire disabled: prohibit the use of open fire in cabinet! Using of fire will
lead to airflow disorder, and filter damage. High-temperature sterilization are
required during the experiment, and the use of infrared sterilizer are highly
recommend;
8.9High efficiency filter has its own life. With the use of prolonged, dust and
bacteria accumulate in the filter resulting in high efficiency filter pressure loss.
When the loss increases to sound an alarm, contact our service department to
replace high efficiency filter, otherwise it will affect the safety performance of
the equipment. And the used filter should be processed as medical waste;
8.10The covers of blower and its downside steel plate are static, and after the side
door is a negative pressure air duct baffle. The air duct was sealed strictly in the
factory and operator is required to maintain its airtightness. The operator must
not remove or loose screws of those parts. if necessary, it must be handled by
the service personal;
8.11 Grille on the surface near the front window is used for both air intake and
sewage (discharge interface under surface); do not block up grille near the
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
25
front window, otherwise it will affect the cabinet's airflow. Recommend the
use of armrest, solving this problem and reducing the operator's wrist
fatigue;
8.12 Long-term use of biological safety cabinets will inevitably cause pollution
(ie HEPA filters, corner cabinets, etc.). In order to remove the pollution
better, recommend to sterilize using formalin (formaldehyde) fumigation
sterilizer one time per 500 hours, and then clear formaldehyde gas using
ammonium hydrogen carbonate neutralizer. Make sure that there will be no
escape of sterilization gas during the process.
8.13 Its storage period should be not exceeding one year. If the period is more
than one year, check the cabinet out of the box, and let the qualified cabinet
enter circulation.
Seriously declare: we will take no responsibility for risks caused by the
operation not according to regulations!
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
26
9. Label description
9.1 Corporate logo (Draft 7)
(Draft 7)
9.2Biological hazard label (Draft 8)
(Draft 8)
9.3 Fuse label (Draft 9)
a b c
(Draft 9)
a、10A power fuse label; b、Neutral wire 10A fuse label ; c、Operating area 5A
fuse socket labels
9.4 Ground label (Draft 10)
(Draft 10)
9.5 Glass door superelevation warning label (Draft 11)
(Draft 11)
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
27
10. Warranty
10.1 Warranty is 12 months from purchasing day (excluding UV lamp and
Fluorescent tube, filter).
10.2 If failure or damage to instruments and equipment during the warranty period
is caused by improper use, we will take no responsibility.
10.3 We charge proper costs for maintenance out of warranty period.
10.4 Life time of this equipment is 10 years.
10.5 We will support trained and recognized technicians or units drawings and part
of the necessary technical data.
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
28
Appendix A Circuit diagram
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
29
Appendix B Air flow pattern and protected area
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
30
Appendix C BSC-11IIA2-X Packing list
Name Quantity
BSC-11IIA2-X Body 1 unit
BSC-11IIA2-X Base 1 pc
RVV Power cord 1 pc
10A Fuse 1 pc
5A Fuse 1 pc
Fluorescent tube 1 pc
UV tube 2 pcs
Romote control(including battery) 1 pc
Key 2 pc
Foot swith 1 pc
BSC-11IIA2-X manual 1 pc
Survey report 1 pc
Certification 1 pc
Warranty card 1 pc
Drain valve 1group
Hexagon socket cap head screws M10×50 8 pc
Hexagonal head screw M10×55+ Flap
gasket10+ Grower washer 10+ M10 nut 3groups
Flap gasket10 8 pc
Grower washer 10 8 pc
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 [email protected]
31
M10 Nut 6 pc
Inner hexagon spanner 1 pc
Fastenings 1 pc the left and right