Intro to Flow Cytometry. James Marvin Director, Flow Cytometry Core Facility University of Utah Health Sciences Center Office 801-585-7382 Lab 801-581- 8641. Seventeen- colour flow cytometry: unravelling the immune system Nature Reviews Immunology, 2004. - PowerPoint PPT Presentation

Transcript of Intro to Flow Cytometry

Page 1: Intro to Flow Cytometry

Intro to Flow Cytometry

James MarvinDirector, Flow Cytometry Core FacilityUniversity of Utah Health Sciences CenterOffice 801-585-7382Lab 801-581-8641

Page 2: Intro to Flow Cytometry

Seventeen-colour flow cytometry: unravelling the immune systemNature Reviews Immunology, 2004

“This ain’t your grandma’s flow cytometer”

Page 3: Intro to Flow Cytometry

Uses of Flow Cytometry Immunophenotyping DNA cell cycle/tumor ploidy Membrane potential Ion flux Cell viability Intracellular protein staining pH changes Cell tracking and proliferation Sorting Redox state Chromatin structure Total protein Lipids Surface charge Membrane fusion/runover Enzyme activity Oxidative metabolism Sulfhydryl groups/glutathione DNA synthesis DNA degradation Gene expression

The uses of flow in research has boomed since the mid-1980s, and is now the gold standard for a variety of applications

Page 4: Intro to Flow Cytometry

Section I

Background Information on Flow Cytometry

Page 5: Intro to Flow Cytometry

Experimental Design Instrumentation“Flow Basics”

Analysis“Data Analysis”

Presentation“Data Analysis”

• Sample Procurement• Sample preparation• Fix/Perm• Which Fluorophore• Controls• Isotype?• Single color• FMO

• Appropriate Lasers• Appropriate Filters• Instrument Settings• Lin vs Log• Time• A, W, H

• Interpretation• Mean, Median• % +• CV• SD• Signal/Noise• Gating

• Histogram• Dot Plot• Density Plot• Overlay• Bar Graph

Many components to a successful assay

Page 6: Intro to Flow Cytometry

What Is Flow Cytometry?

Flow ~ motionCyto ~ cellMetry ~ measure

Measuring both intrinsic and extrinsic properties of cells while in a moving fluid stream

Page 7: Intro to Flow Cytometry
Page 8: Intro to Flow Cytometry

Cytometry vs. Flow CytometryCytometry/Microscopy

Localization of antigen is possible

Poor enumeration of cell subtypes

Limiting number of simultaneous measurements

Flow Cytometry. No ability to determine

localization (traditional flow cytometer)

Can analyze many cells in a short time frame. (30k/sec)

Can look at numerous parameters at once (>20 parameters)

Page 9: Intro to Flow Cytometry

Section II

The 4 Main Components of a Flow Cytometer

Page 10: Intro to Flow Cytometry

What Happens in a Flow Cytometer?

Cells in suspension flow single filethrough a focused laser where they scatter

light and emit fluorescence that is filtered, measured,

then converted to digitized values that are stored in a file

which can then be analyzed and interpreted within specialized software.

Interrogation

Fluidics

Electronics

Interpretation

Page 11: Intro to Flow Cytometry

The Fluidics System“Cells in suspension flow single file”

Cells must flow one-by-one into the cytometer to do single cell analysis

Accomplished through a pressurized laminar flow system.

The sample is injected into a sheath fluid as it passes through a small orifice (50um-300um)

Page 12: Intro to Flow Cytometry

Sheath and Core

Sheath

Core

Page 13: Intro to Flow Cytometry

Fluidics

V. Kachel, H. Fellner-Feldegg & E. Menke - MLM Chapt. 3

Notice how the ink is focused into a tight stream as it is drawn into the tube under laminar flow conditions.

PBS/SheathSample/cells

Laminar flowHydrodynamicFocusing

Laminar flow occurs when a fluid flows in parallel layers, with no disruption between the layers

Page 14: Intro to Flow Cytometry

Particle Orientation and Deformationa: Native human erythrocytes near the margin of the core stream of a short tube (orifice). The cells are uniformly oriented and elongated by the hydrodynamic forces of the inlet flow.

b: In the turbulent flow near the tube wall, the cells are deformed and disoriented in a very individual way. v>3 m/s.

V. Kachel, et al. - MLM Chapt. 3

Page 15: Intro to Flow Cytometry

What Happens in a Flow Cytometer (Simplified)•Cell flash.swf

Flow Cell- the place where hydrodynamically focused cells are delivered to the focused light source

Page 16: Intro to Flow Cytometry

Incoming Laser

Sample

Sheath Sheath

SampleCore

Stream

Low Differential High Differential or “turbulent flow”

Laser Focal Point

Gaussian- A “bell curved” normal distribution where the values and shape falls off quickly as you move away from central, most maximum point.

Page 17: Intro to Flow Cytometry

Low pressure

High pressure

FL30 1024 2048 3072 4096

Cou

nt

300280

260240

220200180

160140

120100

80

6040

200

68.70 19.16 9.56

G0/G1 CV= 2.42

G2/M

S phaseG0/G1

FL3

0 1024 2048 3072 4096

Cou

nt340320300280260240220200180160140120100

80604020

0

74.85 9.12 15.84

GO/G1 CV= 7.79

74.85 9.12 15.84

GO/G1 CV= 7.79

Page 18: Intro to Flow Cytometry

Fluidics Recap

Purpose is to have cells flow one-by-one past a light source.

Cells are “focused” due to hydrodynamic focusing and laminar flow.

Turbulent flow, caused by clogs or fluidic instability can cause imprecise data

Page 19: Intro to Flow Cytometry

What Happens in a Flow Cytometer?

Cells in suspension flow single filethrough a focused laser where they scatter

light and emit fluorescence that is filtered, measured,

and converted to digitized values that are stored in a file

Which can then be read by specialized software.

Interrogation

Fluidics

Electronics

Interpretation

Page 20: Intro to Flow Cytometry

InterrogationLight source needs to be focused on the

same point where cells are focused.

Light source 99%=Lasers Add optical bench

Page 21: Intro to Flow Cytometry

LasersLight amplification by stimulated emission of radiation

Lasers provide a single wavelength of light (monochromatic)

They can provide milliwatts to watts of power Low divergence Provide coherent light Gas, dye, or solid state

Coherent: all emmiting photons have same wavelength, phase and direction as stimulation photons

Page 22: Intro to Flow Cytometry

Light collection

Collected photons are the product of 488nm laser light scattering or bouncing off cells

488nm Information associated

with physical attributes of cells (size, granularity, refractive index)

Scatter FluorescenceVS Collected photons are

product of excitation with subsequent emission determined by fluorophore

350nm-800nm Readout of intrinsic

(autofluorescence) or extrinsic (intentional cell labeling) fluorescence

Page 23: Intro to Flow Cytometry

Forward Scatter

FSCDetector

Laser Beam

Original from Purdue University Cytometry Laboratories

.50-80

Page 24: Intro to Flow Cytometry

Forward Scatter

The intensity of forward scatter signal is often attributed to cell size, but is very complex and also reflects refractive index (membrane permeability), among other things

Forward Scatter=FSC=FALS=LALS

Page 25: Intro to Flow Cytometry

Side Scatter

FSCDetector

CollectionLens

SSCDetector

Laser Beam

Original from Purdue University Cytometry Laboratories

Page 26: Intro to Flow Cytometry

Side Scatter

Laser light that is scattered at 90 degrees to the axis of the laser path is detected in the Side Scatter Channel

The intensity of this signal is proportional to the amount of cytosolic structure in the cell (eg. granules, cell inclusions, drug delivery nanoparticles.)

Side Scatter=SSC=RALS=90 degree Scatter

Page 27: Intro to Flow Cytometry

Why Look at FSC v. SSC Since FSC ~ size and SSC ~ internal structure, a

correlated measurement between them can allow for differentiation of cell types in a heterogenous cell population

FSC

SSC

Lymphocytes

Monocytes

Granulocytes

RBCs, Debris,Dead Cells

LIVE

Dead

Page 28: Intro to Flow Cytometry

Multi-laser Instruments and Pinholes

Page 29: Intro to Flow Cytometry

Multi-laser Instruments and pinholes

Implications--Can separate completely overlapping emission profilesif originating off different lasers-Significantly reduces compensation

Page 30: Intro to Flow Cytometry

FluorescenceEn

ergy

Absorbed exciting light

Emitted fluorescence

S0Ground State

Excited higher energy states

S1

S2

S3

•As the laser interrogates the cell, fluorochromes on/in the cell (intrinsic or extrinsic) may absorb some of the light and become excited•As those fluorochromes leave their excited state, they release energy in the form of a photon with a specific wavelength, longer than the excitation wavelength

Stokes shift- the difference in wavelength between the absorption or excitation and the emission

Page 31: Intro to Flow Cytometry

Optical Filters Many wavelengths of light will be emitted from a cell,

we need a way to split the light into its specific wavelengths in order to detect them independently. This is done with filters

Optical filters are designed such that they absorb or reflect some wavelengths of light, while transmitting other.

3 types of filters Long Pass filter Short Pass filter Band Pass filter

Page 32: Intro to Flow Cytometry

Long Pass Filters

Transmit all wavelengths greater than specified wavelength Example: 500LP will transmit all wavelengths

greater than 500nm

400nm 500nm 600nm 700nm

Tran

smitt

ance

Page 33: Intro to Flow Cytometry

Short Pass Filter

Transmits all wavelengths less than specified wavelength Example: 600SP will transmit all wavelengths

less than 600nm.

400nm 500nm 600nm 700nm

Tran

smitt

ance

Original from Cytomation Training Manual, Modified by James Marvin

Page 34: Intro to Flow Cytometry

Band Pass Filter

Transmits a specific band of wavelengths Example: 550/20BP Filter will transmit

wavelengths of light between 540nm and 560nm (550/20 = 550+/-10, not 550+/-20)

400nm 500nm 600nm 700nm

Tran

smitt

ance

Page 35: Intro to Flow Cytometry

Dichroic Filters

Can be a long pass or short pass filter Depending on the specs of the filter, some of the

light is reflected and part of the light is transmitted and continues on.

DichroicFilter

Detector 1

Detector 2

Page 36: Intro to Flow Cytometry

BD optical layout

Page 37: Intro to Flow Cytometry

Spectra of Common Fluorochromes with Typical Filters

Page 38: Intro to Flow Cytometry

Spatial separation

Page 39: Intro to Flow Cytometry

CompensationFluorochromes typically fluoresce over a

large part of the spectrum (100nm or more)Depending on filter arrangement, a detector

may see some fluorescence from more than 1 fluorochrome. (referred to as bleed over)

You need to “compensate” for this bleed over so that 1 detector reports signal from only 1 fluorochrome

Page 40: Intro to Flow Cytometry
Page 41: Intro to Flow Cytometry

Compensation-Practical Eg.

compensationexample.swf

Page 42: Intro to Flow Cytometry

Interrogation Recap A focused light source (laser) interrogates a cell and

scatters light That scattered light travels down a channel to a detector FSC ~ size and cell membrane integrity SSC ~ internal cytosolic structure Fluorochromes on/in the cell will become excited by the

laser and emit photons These photons travel down channels and are steered and

split by dichroic (LP/SP) filters

Page 43: Intro to Flow Cytometry

What Happens in a Flow Cytometer?

Cells in suspension flow single file Through a focused laser where they scatter

light and emit fluorescence that is filtered, measured

and converted to digitized values that are stored in a file

Which can then be read by specialized software.

Interrogation

Fluidics

Electronics

Interpretation

Page 44: Intro to Flow Cytometry

Electronics

Detectors basically collect photons of light and convert them to an electrical current

The electronics must process that light signal and convert the current to a digitized value/# that the computer can graph

Page 45: Intro to Flow Cytometry

DetectorsThere are two main types of photo detectors

used in flow cytometry Photodiodes

Used for strong signals, when saturation is a potential problem (eg. FSC detector)

Photomultiplier tubes (PMT) Used for detecting small amounts of fluorescence

emitted from fluorochromes. Incredible Gain (amplification-up to 10million

times) Low noise

Page 46: Intro to Flow Cytometry

Photodiodes and PMTs

Photo Detectors usually have a band pass filter in front of them to only allow a specific band width of light to reach it

Therefore, each detector has a range of light it can detect, once a filter has been placed in front of it.

Page 47: Intro to Flow Cytometry

Photoelectric Effect

Einstein- Nobel Prize 1921

pmt.swf

Photons -> Photoelectrons -> Electrons

cellflashelectronics.swf

Electric pulse generation

Page 48: Intro to Flow Cytometry

Detector names

Page 49: Intro to Flow Cytometry

Measurements of the Pulse

Pulse HeightPulse Width

Pulse Area

Time

Mea

sure

d C

urre

nt a

t det

ecto

r

Page 50: Intro to Flow Cytometry

0

10

(Vol

ts)

Rel

ativ

e B

right

ness

6.21 volts

1.23 volts

3.54 volts

101

102

103

104

1

ADCAnalog to Digital Conversion

Count

Page 51: Intro to Flow Cytometry

Does voltage setting matter?

Voltage=362 292 272 252 522

Page 52: Intro to Flow Cytometry

FSC SSC FITC PE APC APC-Cy7FCS Fileor

List Mode File

Page 53: Intro to Flow Cytometry

Electronics Recap Photons ElectronsVoltage pulseDigital # The varying number of photons reaching the

detector are converted to a proportional number of electrons

The number of electrons exiting a PMT can be multiplied by making more electrons available to the detector (increase Voltage input)

The current generated goes to a log or linear amplifier where it is amplified (if desired) and is converted to a voltage pulse

The voltage pulse goes to the ADC to be digitized The values are placed into a List Mode File

Page 54: Intro to Flow Cytometry

What Happens in a Flow Cytometer?

Cells in suspension flow single file pasta focused laser where they scatter light and

emit fluorescence that is collected, filtered and converted to digitized values that are

stored in a file Which can then be read by specialized

software.

Interrogation

Fluidics

Electronics

Interpretation

Page 55: Intro to Flow Cytometry

See you at Data Analysis

Page 56: Intro to Flow Cytometry

AntibodyAntigen binding site

Page 57: Intro to Flow Cytometry

Immunophenotyping

Page 58: Intro to Flow Cytometry

roGFPRedox senstitive biosensor

Page 59: Intro to Flow Cytometry

Ca2+ Flux

Indo-1 Ca2+ freeEmission=500nm

Indo-1 Ca2+ boundEmission=395nm

Page 60: Intro to Flow Cytometry

Apoptosis

Live

Apoptotic

Annexin V

MTR

PI

FLIC

A

Page 61: Intro to Flow Cytometry

Cell cycle

Page 62: Intro to Flow Cytometry

Sorting

Last attached droplet