HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput Yi Zhang 1 , Zhiqi Hao 1 , Markus Kellmann 2 and Andreas FR. Huhmer 1 1 Thermo Fisher Scientific, San Jose, CA, USA; 2 Thermo Fisher Scientific, Bremen, Germany

Transcript of HR/AM Targeted Peptide Quantification on a Q Exactive MS: A...

Page 1: HR/AM Targeted Peptide Quantification on a Q Exactive MS: A …apps.thermoscientific.com/media/SID/LSMS/PDF/ASMS/ASMS12/... · 2012-06-25 · FIGURE 7. HCD spectrum of peptide NGFILDGFPR*

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High ThroughputYi Zhang1, Zhiqi Hao1, Markus Kellmann2 and Andreas FR. Huhmer1

1Thermo Fisher Scientific, San Jose, CA, USA; 2Thermo Fisher Scientific, Bremen, Germany

Page 2: HR/AM Targeted Peptide Quantification on a Q Exactive MS: A …apps.thermoscientific.com/media/SID/LSMS/PDF/ASMS/ASMS12/... · 2012-06-25 · FIGURE 7. HCD spectrum of peptide NGFILDGFPR*

2 HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, andHigh Throughput Yi Zhang1, Zhiqi Hao1, Markus Kellmann2 and Andreas FR. Huhmer1

1Thermo Fisher Scientific, San Jose, CA, USA; 2Thermo Fisher Scientific, Bremen, Germany

OverviewPurpose: The performance of a novel hybrid quadrupole-OrbitrapTM mass spectrometer was evaluated in targeted peptide quantification.

Methods: Two high-resolution, accurate-mass (HR/AM) targeted quantification methods were used to analyze 14 peptide standards spiked into a complex E. colidigest background. The data were automatically processed using specialized software.

Results: A detection limit of 10 amol and a linear dynamic range of 4 orders were obtained with msx tSIM methods in a moderately complex background. A detection limit of 10–50 amol and a linear dynamic range of 3–4 orders were obtained with tHCD methods in a highly complex background.

IntroductionQuantitative proteomics enables the identification of a large number of protein candidates, which display biologically interesting dynamics on a global scale in the early discovery phase. A targeted MS approach, in particular selected reaction monitoring (SRM) on a triple quadrupole MS, has become the standard technique for quantitatively analyzing tens to hundreds of peptide candidates across large numbers of samples, either for understanding of signaling regulation or for verification and selection of potential biomarkers. However, the low resolution of triple quadrupole MS limits its ability to achieve high selectivity on the targets from complex backgrounds. It is also difficult to develop SRM method for peptides with high charge states or high mass or for modified peptides.

In this study, a true high-resolution, accurate-mass (HR/AM) mass spectrometer, was evaluated for targeted protein quantification. The dynamic range and LOD/LOQ of two HR/AM methods: spectrum multiplexing targeted selected ion monitoring (msx tSIM) and targeted HCD (tHCD), were investigated.

MethodsSample Preparation

Thermo Scientific heavy-isotope labeled peptide retention standards were spiked into either 5 ng/µL or 250 ng/µL E coli. whole cell tryptic digest to reach final concentrations of 0 amol/µL, 5 amol/µL, 25 amol/µL, 50 amol/µL, 500 amol/µL, 5 fmol/µL, and 50 fmol/µL.

Liquid Chromatography

Peptides from 2 µL of each sample were separated on a Michrom Magic™ C18 nanoLC column (75 μm x 15 cm, 3 μm particle) with a 1 hr gradient at a flow rate of 350 nL/min. LC solvents were 0.1% formic acid in H2O (Solvent A) and 0.1% formic acid in acetonitrile (Solvent B). The LC gradient was 0-2 min, 3% B; 2-47 min, 3%-30% B; 47-52 min, 30%-95% B; 52-54 min, 95% B; 54-55 min, 95%-5% B; 55-60 min, 5% B.

Mass Spectrometry

The samples with 5 ng/µL E coli digest background were analyzed in triplicate with a Full-msx tSIM method on a Thermo Scientific Q Exactive hybrid quadrupole-Orbitrap MS (Figure 1). The resolution was set at 140K for both scan types. The AGC target was 1E6 for full scan and 2E5 for SIM scan. The maximum ion injection time was 120 ms for full scan and 500 ms for SIM scan. The isolation width for SIM scan was 4 amu. The peptide retention standards were monitored over a 3 min window. The multiplexing level was set at 4, which allowed isolation and accumulation of up to four peptide targets in the C-trap before they were transferred to the Orbitrap mass analyzer for detection (Figure 2).

The samples with 250 ng/µL E coli digest background were analyzed in triplicate with a tHCD method. The resolution was set at 17.5K. The AGC target was 5E5. The maximum ion injection time was 500 ms. The quadrupole isolation width was 2.0 amu. The normalized collision energy (NCE) was 27%.The peptide retention standards were monitored over a 3 min window.

Data Analysis

The data were automatically processed using Thermo Scientific Pinpoint software. Extracted ion chromatograms for peptide retention standards were obtained with a 5 ppm mass tolerance. Peak areas were calculated with 7-point smoothed complete chromatograms. Linear fitting was performed with a weighing scheme of 1/x.

ResultsTargeted Quantification with msx tSIM

FIGURE 1. Schematic and key features of the Q Exactive MS.

Key Features of the Q Exactive MS

Incorporation of the S-lens at the source enhances sensitivity by up to 5 fold. Quadrupole mass filter enables precursor selection for data-dependent MS2 and

selected ion monitoring (SIM). Advanced signal processing increases resolution by about twofold at the same

transient length, which results in a maximum resolution of 140K at m/z 200 and a maximum scan speed of 12 Hz at a resolution of 17.5K.

Spectrum multiplexing (msx) and parallel ion injection/detection significantly improvesduty cycle.

FIGURE 2: Spectrum multiplexing and parallel ion injection/detection provide high-throughput analysis.

Orbitrap FTMS acquisition scan 2 AGC

Orbitrap FTMS acquisition scan 1 AGC

Collecting ions for scan 2 Collecting ions for scan 3

Up to 10 target ions can be isolated by the quadrupoles sequentially, stored in the C-trap, then transferred to the Orbitrap mass analyzer and detected with high resolution simultaneously. The isolation and trapping of the target ions is also concurrent with the Orbitrap mass analyzer detection of the previous ion packet.

Full or SIM spectra were acquired at three different resolutions: 35K, 70K and 140K. Baseline separation of two ions that are only 30 ppm apart was only achievable at the resolution of 140K. The high mass accuracy (< 5 ppm) and high resolution ensured accurate identification of target peptides from complex background with msx tSIM (Figure 3).

FIGURE 3: High resolution ensures accurate target selection.

FIGURE 4: High sensitivity with quadrupole-based SIM scan.

Target peptides in 10 ng of E coli digest were analyzed with alternating full scans and tSIM scans (Figure 4). The peptide GLILVGGYGTR* (558.3259, 2+) is barely visible in the full scan with a 600 amu isolation window (Figure 4A), but is accurately identified with a S/N of 65 and mass deviation of ~0.2 ppm in the SIM scan with a 4 amu isolation window (Figure 4B). The +1 isotope peak is also accurately identified with a S/N of 39 and mass deviation of ~1.4 ppm in the SIM scan.

FIGURE 5. Multiplexing improves throughput and results in accurate quantification.

17.0 17.2 17.4 17.6 17.8 18.0 18.2 18.4 18.6

Time (min)

020406080

1000

20406080

100

Rel

ativ

e A

bund

ance

020406080

100 18.12

18.11

18.10

GISNEGQNASIK*

5 ppm XIC

100 fmol

1 fmol

10 amol

Extracted ion chromatograms (XIC) of target peptides in SIM scans were obtained with 5 ppm mass tolerance. The XIC of GISNEGQNASIK* (613.3167, 2+) at 100 fmol, 1 fmol and 10 amol is displayed in Figure 5. Although GISNEGQNASIK* was monitored with two other peptides in the same time range, more than 10 SIM scans with a resolution of 140K were obtained across a 12 sec LC peak, even at a sample amount of 10 amol.

As shown in Figure 6, a 4-order linear dynamic range was obtained for most peptides with the msx tSIM method. Table 1 lists the LOD and LOQ of target peptides for the method. The CV% from triplicate analyses is also included for the LOQ.

FIGURE 6. Four-order linear dynamic range with msx tSIM method.

Pea

k A

rea

Pea

k A

rea

(log 1

0)

Sample Amount (fmol)Sample Amount (log10)

R2 > 0.9998 R2 > 0.9997

IGDYAGIK*

SSAAPPPPPR*GISNEGQNASIK*

FIGURE 7. HCD spectrum of peptide NGFILDGFPR* (573.305, 2+) at 50 amol in 500 ng E coli tryptic digest background.

TABLE 1. LOD and LOQ of target peptides for the msx tSIM method.

LOD (amol) LOQ (amol) CV% (LOQ)IGDYAGIK* 10 50 3SSAAPPPPPR* 10 50 3HVLTSIGEK* 10 50 3LTILEELR* <10 10 4GLILVGGYGTR* 10 50 7NGFILDGFPR* <10 10 14SAAGAFGPELSR* <10 10 4GISNEGQNASIK* <10 10 10ELASGLSFPVGFK* <10 10 10TASEFDSAIAQDK* 10 50 1SFANQPLEVVYSK* <10 10 8ELGQSGVDTYLQTK* 10 50 7LSSEAPALFQFDLK* <50 50 3GILFVGSGVSGGEEGAR* 50 100 5

400 500 600 700 800 900 1000m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

714.3813y6

486.2699y4

601.2971y5

974.5300y8

With tHCD, high resolution and high mass accuracy allow simultaneous accurate identification of multiple fragment ions from complex mixed MS2 spectra (Figure 7). As shown in Figure 8, 3-4 orders of linear dynamic range were obtained for most peptides with the tHCD method. Table 2 lists the LOD and LOQ of target peptides for the tHCD method. The CV% from triplicate analyses is also included for the LOQ.

Targeted Quantification with tHCD

50 amol

ConclusionThe Q Exactive™ MS is well suited for targeted quantification providing two powerful HR/AM approaches: msx tSIM and tHCD. Both approaches demonstrate high selectivity, high sensitivity and high throughput due to several hardware and software innovations.

For the msx tSIM approach, high mass accuracy (< 5 ppm) and high resolution (140K) ensure accurate identification of target ions in the presence of a complex background, while spectrum multiplexing dramatically increases throughput without sacrificing resolution.

A LOD of 10 amol and a 4-order linear dynamic range was obtained for most peptides with the msx tSIM method in the presence of a medium complex background.

Different from triple-quadrupole-based SRM, all fragment ions are simultaneously detected with high mass accuracy and high resolution using the tHCD approach. The HR/AM of fragment ions provides unmatched selectivity. The simultaneous detection of all fragment ions allows more efficient sampling of target peptides and results in improved sensitivity. A scan speed of 12 Hz at a resolution of 17,500 ensures high throughput.

A LOD of 50 amol and 3–4 orders of linear dynamic range were obtained for most peptides with the tHCD method in the presence of a strong complex background.

FIGURE 8. Three to four orders of linear dynamic range with the tHCD method.

Pea

k A

rea

Pea

k A

rea

(log 1

0)

Sample Amount (fmol)Sample Amount (log10)

R2 > 0.999 R2 > 0.998

SFANQPLEVVYSK*LTILEELR*NGFILDGFPR*

TABLE 2. LOD and LOQ of target peptides for the tHCD method.

LOD (amol) LOQ (amol) CV% (LOQ)IGDYAGIK* >10 50 6SSAAPPPPPR* 10 50 11HVLTSIGEK* 50 100 11LTILEELR* 10 50 3GLILVGGYGTR* >10 50 8NGFILDGFPR* >10 50 6SAAGAFGPELSR* 10 50 4GISNEGQNASIK* 10 50 5ELASGLSFPVGFK* >10 50 2TASEFDSAIAQDK 10 50 1SFANQPLEVVYSK* <10 10 12ELGQSGVDTYLQTK* 10 50 6LSSEAPALFQFDLK* 10 50 6GILFVGSGVSGGEEGAR* >10 50 8

400 500 600 700 800 900 10000

20

40

60

80

100

Rel

ativ

e A

bund

ance

557.0 558.0 559.0 560.0m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

558.3 558.5 558.7

558.3302S/N: 7

Full Scan (400 -1000amu)Scan #: 3099

SIM Scan (556.3 - 560.3amu)Scan #: 3100

558.3 558.5 558.7

558.3258(S/N: 65)

558.8282(S/N: 39)

A.

B.

Capillary

S-Lens forImproved Sensitivity

BentFlatapole

Quadrupole Mass Filter forPrecursor Ion Selection

HCD Cell, C-Trap forSpectrum Multiplexing

Orbitrap withEnhanced Resolution

30 ppm

m/z

Magic is a trademark of Michrom BioResources, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

Page 3: HR/AM Targeted Peptide Quantification on a Q Exactive MS: A …apps.thermoscientific.com/media/SID/LSMS/PDF/ASMS/ASMS12/... · 2012-06-25 · FIGURE 7. HCD spectrum of peptide NGFILDGFPR*

3Thermo Scientific Poster Note • PN63586_E 06/12S

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, andHigh Throughput Yi Zhang1, Zhiqi Hao1, Markus Kellmann2 and Andreas FR. Huhmer1

1Thermo Fisher Scientific, San Jose, CA, USA; 2Thermo Fisher Scientific, Bremen, Germany

OverviewPurpose: The performance of a novel hybrid quadrupole-OrbitrapTM mass spectrometer was evaluated in targeted peptide quantification.

Methods: Two high-resolution, accurate-mass (HR/AM) targeted quantification methods were used to analyze 14 peptide standards spiked into a complex E. colidigest background. The data were automatically processed using specialized software.

Results: A detection limit of 10 amol and a linear dynamic range of 4 orders were obtained with msx tSIM methods in a moderately complex background. A detection limit of 10–50 amol and a linear dynamic range of 3–4 orders were obtained with tHCD methods in a highly complex background.

IntroductionQuantitative proteomics enables the identification of a large number of protein candidates, which display biologically interesting dynamics on a global scale in the early discovery phase. A targeted MS approach, in particular selected reaction monitoring (SRM) on a triple quadrupole MS, has become the standard technique for quantitatively analyzing tens to hundreds of peptide candidates across large numbers of samples, either for understanding of signaling regulation or for verification and selection of potential biomarkers. However, the low resolution of triple quadrupole MS limits its ability to achieve high selectivity on the targets from complex backgrounds. It is also difficult to develop SRM method for peptides with high charge states or high mass or for modified peptides.

In this study, a true high-resolution, accurate-mass (HR/AM) mass spectrometer, was evaluated for targeted protein quantification. The dynamic range and LOD/LOQ of two HR/AM methods: spectrum multiplexing targeted selected ion monitoring (msx tSIM) and targeted HCD (tHCD), were investigated.

MethodsSample Preparation

Thermo Scientific heavy-isotope labeled peptide retention standards were spiked into either 5 ng/µL or 250 ng/µL E coli. whole cell tryptic digest to reach final concentrations of 0 amol/µL, 5 amol/µL, 25 amol/µL, 50 amol/µL, 500 amol/µL, 5 fmol/µL, and 50 fmol/µL.

Liquid Chromatography

Peptides from 2 µL of each sample were separated on a Michrom Magic™ C18 nanoLC column (75 μm x 15 cm, 3 μm particle) with a 1 hr gradient at a flow rate of 350 nL/min. LC solvents were 0.1% formic acid in H2O (Solvent A) and 0.1% formic acid in acetonitrile (Solvent B). The LC gradient was 0-2 min, 3% B; 2-47 min, 3%-30% B; 47-52 min, 30%-95% B; 52-54 min, 95% B; 54-55 min, 95%-5% B; 55-60 min, 5% B.

Mass Spectrometry

The samples with 5 ng/µL E coli digest background were analyzed in triplicate with a Full-msx tSIM method on a Thermo Scientific Q Exactive hybrid quadrupole-Orbitrap MS (Figure 1). The resolution was set at 140K for both scan types. The AGC target was 1E6 for full scan and 2E5 for SIM scan. The maximum ion injection time was 120 ms for full scan and 500 ms for SIM scan. The isolation width for SIM scan was 4 amu. The peptide retention standards were monitored over a 3 min window. The multiplexing level was set at 4, which allowed isolation and accumulation of up to four peptide targets in the C-trap before they were transferred to the Orbitrap mass analyzer for detection (Figure 2).

The samples with 250 ng/µL E coli digest background were analyzed in triplicate with a tHCD method. The resolution was set at 17.5K. The AGC target was 5E5. The maximum ion injection time was 500 ms. The quadrupole isolation width was 2.0 amu. The normalized collision energy (NCE) was 27%.The peptide retention standards were monitored over a 3 min window.

Data Analysis

The data were automatically processed using Thermo Scientific Pinpoint software. Extracted ion chromatograms for peptide retention standards were obtained with a 5 ppm mass tolerance. Peak areas were calculated with 7-point smoothed complete chromatograms. Linear fitting was performed with a weighing scheme of 1/x.

ResultsTargeted Quantification with msx tSIM

FIGURE 1. Schematic and key features of the Q Exactive MS.

Key Features of the Q Exactive MS

Incorporation of the S-lens at the source enhances sensitivity by up to 5 fold. Quadrupole mass filter enables precursor selection for data-dependent MS2 and

selected ion monitoring (SIM). Advanced signal processing increases resolution by about twofold at the same

transient length, which results in a maximum resolution of 140K at m/z 200 and a maximum scan speed of 12 Hz at a resolution of 17.5K.

Spectrum multiplexing (msx) and parallel ion injection/detection significantly improvesduty cycle.

FIGURE 2: Spectrum multiplexing and parallel ion injection/detection provide high-throughput analysis.

Orbitrap FTMS acquisition scan 2 AGC

Orbitrap FTMS acquisition scan 1 AGC

Collecting ions for scan 2 Collecting ions for scan 3

Up to 10 target ions can be isolated by the quadrupoles sequentially, stored in the C-trap, then transferred to the Orbitrap mass analyzer and detected with high resolution simultaneously. The isolation and trapping of the target ions is also concurrent with the Orbitrap mass analyzer detection of the previous ion packet.

Full or SIM spectra were acquired at three different resolutions: 35K, 70K and 140K. Baseline separation of two ions that are only 30 ppm apart was only achievable at the resolution of 140K. The high mass accuracy (< 5 ppm) and high resolution ensured accurate identification of target peptides from complex background with msx tSIM (Figure 3).

FIGURE 3: High resolution ensures accurate target selection.

FIGURE 4: High sensitivity with quadrupole-based SIM scan.

Target peptides in 10 ng of E coli digest were analyzed with alternating full scans and tSIM scans (Figure 4). The peptide GLILVGGYGTR* (558.3259, 2+) is barely visible in the full scan with a 600 amu isolation window (Figure 4A), but is accurately identified with a S/N of 65 and mass deviation of ~0.2 ppm in the SIM scan with a 4 amu isolation window (Figure 4B). The +1 isotope peak is also accurately identified with a S/N of 39 and mass deviation of ~1.4 ppm in the SIM scan.

FIGURE 5. Multiplexing improves throughput and results in accurate quantification.

17.0 17.2 17.4 17.6 17.8 18.0 18.2 18.4 18.6

Time (min)

020406080

1000

20406080

100

Rel

ativ

e A

bund

ance

020406080

100 18.12

18.11

18.10

GISNEGQNASIK*

5 ppm XIC

100 fmol

1 fmol

10 amol

Extracted ion chromatograms (XIC) of target peptides in SIM scans were obtained with 5 ppm mass tolerance. The XIC of GISNEGQNASIK* (613.3167, 2+) at 100 fmol, 1 fmol and 10 amol is displayed in Figure 5. Although GISNEGQNASIK* was monitored with two other peptides in the same time range, more than 10 SIM scans with a resolution of 140K were obtained across a 12 sec LC peak, even at a sample amount of 10 amol.

As shown in Figure 6, a 4-order linear dynamic range was obtained for most peptides with the msx tSIM method. Table 1 lists the LOD and LOQ of target peptides for the method. The CV% from triplicate analyses is also included for the LOQ.

FIGURE 6. Four-order linear dynamic range with msx tSIM method.

Pea

k A

rea

Pea

k A

rea

(log 1

0)

Sample Amount (fmol)Sample Amount (log10)

R2 > 0.9998 R2 > 0.9997

IGDYAGIK*

SSAAPPPPPR*GISNEGQNASIK*

FIGURE 7. HCD spectrum of peptide NGFILDGFPR* (573.305, 2+) at 50 amol in 500 ng E coli tryptic digest background.

TABLE 1. LOD and LOQ of target peptides for the msx tSIM method.

LOD (amol) LOQ (amol) CV% (LOQ)IGDYAGIK* 10 50 3SSAAPPPPPR* 10 50 3HVLTSIGEK* 10 50 3LTILEELR* <10 10 4GLILVGGYGTR* 10 50 7NGFILDGFPR* <10 10 14SAAGAFGPELSR* <10 10 4GISNEGQNASIK* <10 10 10ELASGLSFPVGFK* <10 10 10TASEFDSAIAQDK* 10 50 1SFANQPLEVVYSK* <10 10 8ELGQSGVDTYLQTK* 10 50 7LSSEAPALFQFDLK* <50 50 3GILFVGSGVSGGEEGAR* 50 100 5

400 500 600 700 800 900 1000m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

714.3813y6

486.2699y4

601.2971y5

974.5300y8

With tHCD, high resolution and high mass accuracy allow simultaneous accurate identification of multiple fragment ions from complex mixed MS2 spectra (Figure 7). As shown in Figure 8, 3-4 orders of linear dynamic range were obtained for most peptides with the tHCD method. Table 2 lists the LOD and LOQ of target peptides for the tHCD method. The CV% from triplicate analyses is also included for the LOQ.

Targeted Quantification with tHCD

50 amol

ConclusionThe Q Exactive™ MS is well suited for targeted quantification providing two powerful HR/AM approaches: msx tSIM and tHCD. Both approaches demonstrate high selectivity, high sensitivity and high throughput due to several hardware and software innovations.

For the msx tSIM approach, high mass accuracy (< 5 ppm) and high resolution (140K) ensure accurate identification of target ions in the presence of a complex background, while spectrum multiplexing dramatically increases throughput without sacrificing resolution.

A LOD of 10 amol and a 4-order linear dynamic range was obtained for most peptides with the msx tSIM method in the presence of a medium complex background.

Different from triple-quadrupole-based SRM, all fragment ions are simultaneously detected with high mass accuracy and high resolution using the tHCD approach. The HR/AM of fragment ions provides unmatched selectivity. The simultaneous detection of all fragment ions allows more efficient sampling of target peptides and results in improved sensitivity. A scan speed of 12 Hz at a resolution of 17,500 ensures high throughput.

A LOD of 50 amol and 3–4 orders of linear dynamic range were obtained for most peptides with the tHCD method in the presence of a strong complex background.

FIGURE 8. Three to four orders of linear dynamic range with the tHCD method.

Pea

k A

rea

Pea

k A

rea

(log 1

0)

Sample Amount (fmol)Sample Amount (log10)

R2 > 0.999 R2 > 0.998

SFANQPLEVVYSK*LTILEELR*NGFILDGFPR*

TABLE 2. LOD and LOQ of target peptides for the tHCD method.

LOD (amol) LOQ (amol) CV% (LOQ)IGDYAGIK* >10 50 6SSAAPPPPPR* 10 50 11HVLTSIGEK* 50 100 11LTILEELR* 10 50 3GLILVGGYGTR* >10 50 8NGFILDGFPR* >10 50 6SAAGAFGPELSR* 10 50 4GISNEGQNASIK* 10 50 5ELASGLSFPVGFK* >10 50 2TASEFDSAIAQDK 10 50 1SFANQPLEVVYSK* <10 10 12ELGQSGVDTYLQTK* 10 50 6LSSEAPALFQFDLK* 10 50 6GILFVGSGVSGGEEGAR* >10 50 8

400 500 600 700 800 900 10000

20

40

60

80

100

Rel

ativ

e A

bund

ance

557.0 558.0 559.0 560.0m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

558.3 558.5 558.7

558.3302S/N: 7

Full Scan (400 -1000amu)Scan #: 3099

SIM Scan (556.3 - 560.3amu)Scan #: 3100

558.3 558.5 558.7

558.3258(S/N: 65)

558.8282(S/N: 39)

A.

B.

Capillary

S-Lens forImproved Sensitivity

BentFlatapole

Quadrupole Mass Filter forPrecursor Ion Selection

HCD Cell, C-Trap forSpectrum Multiplexing

Orbitrap withEnhanced Resolution

30 ppm

m/z

Magic is a trademark of Michrom BioResources, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

Page 4: HR/AM Targeted Peptide Quantification on a Q Exactive MS: A …apps.thermoscientific.com/media/SID/LSMS/PDF/ASMS/ASMS12/... · 2012-06-25 · FIGURE 7. HCD spectrum of peptide NGFILDGFPR*

4 HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, andHigh Throughput Yi Zhang1, Zhiqi Hao1, Markus Kellmann2 and Andreas FR. Huhmer1

1Thermo Fisher Scientific, San Jose, CA, USA; 2Thermo Fisher Scientific, Bremen, Germany

OverviewPurpose: The performance of a novel hybrid quadrupole-OrbitrapTM mass spectrometer was evaluated in targeted peptide quantification.

Methods: Two high-resolution, accurate-mass (HR/AM) targeted quantification methods were used to analyze 14 peptide standards spiked into a complex E. colidigest background. The data were automatically processed using specialized software.

Results: A detection limit of 10 amol and a linear dynamic range of 4 orders were obtained with msx tSIM methods in a moderately complex background. A detection limit of 10–50 amol and a linear dynamic range of 3–4 orders were obtained with tHCD methods in a highly complex background.

IntroductionQuantitative proteomics enables the identification of a large number of protein candidates, which display biologically interesting dynamics on a global scale in the early discovery phase. A targeted MS approach, in particular selected reaction monitoring (SRM) on a triple quadrupole MS, has become the standard technique for quantitatively analyzing tens to hundreds of peptide candidates across large numbers of samples, either for understanding of signaling regulation or for verification and selection of potential biomarkers. However, the low resolution of triple quadrupole MS limits its ability to achieve high selectivity on the targets from complex backgrounds. It is also difficult to develop SRM method for peptides with high charge states or high mass or for modified peptides.

In this study, a true high-resolution, accurate-mass (HR/AM) mass spectrometer, was evaluated for targeted protein quantification. The dynamic range and LOD/LOQ of two HR/AM methods: spectrum multiplexing targeted selected ion monitoring (msx tSIM) and targeted HCD (tHCD), were investigated.

MethodsSample Preparation

Thermo Scientific heavy-isotope labeled peptide retention standards were spiked into either 5 ng/µL or 250 ng/µL E coli. whole cell tryptic digest to reach final concentrations of 0 amol/µL, 5 amol/µL, 25 amol/µL, 50 amol/µL, 500 amol/µL, 5 fmol/µL, and 50 fmol/µL.

Liquid Chromatography

Peptides from 2 µL of each sample were separated on a Michrom Magic™ C18 nanoLC column (75 μm x 15 cm, 3 μm particle) with a 1 hr gradient at a flow rate of 350 nL/min. LC solvents were 0.1% formic acid in H2O (Solvent A) and 0.1% formic acid in acetonitrile (Solvent B). The LC gradient was 0-2 min, 3% B; 2-47 min, 3%-30% B; 47-52 min, 30%-95% B; 52-54 min, 95% B; 54-55 min, 95%-5% B; 55-60 min, 5% B.

Mass Spectrometry

The samples with 5 ng/µL E coli digest background were analyzed in triplicate with a Full-msx tSIM method on a Thermo Scientific Q Exactive hybrid quadrupole-Orbitrap MS (Figure 1). The resolution was set at 140K for both scan types. The AGC target was 1E6 for full scan and 2E5 for SIM scan. The maximum ion injection time was 120 ms for full scan and 500 ms for SIM scan. The isolation width for SIM scan was 4 amu. The peptide retention standards were monitored over a 3 min window. The multiplexing level was set at 4, which allowed isolation and accumulation of up to four peptide targets in the C-trap before they were transferred to the Orbitrap mass analyzer for detection (Figure 2).

The samples with 250 ng/µL E coli digest background were analyzed in triplicate with a tHCD method. The resolution was set at 17.5K. The AGC target was 5E5. The maximum ion injection time was 500 ms. The quadrupole isolation width was 2.0 amu. The normalized collision energy (NCE) was 27%.The peptide retention standards were monitored over a 3 min window.

Data Analysis

The data were automatically processed using Thermo Scientific Pinpoint software. Extracted ion chromatograms for peptide retention standards were obtained with a 5 ppm mass tolerance. Peak areas were calculated with 7-point smoothed complete chromatograms. Linear fitting was performed with a weighing scheme of 1/x.

ResultsTargeted Quantification with msx tSIM

FIGURE 1. Schematic and key features of the Q Exactive MS.

Key Features of the Q Exactive MS

Incorporation of the S-lens at the source enhances sensitivity by up to 5 fold. Quadrupole mass filter enables precursor selection for data-dependent MS2 and

selected ion monitoring (SIM). Advanced signal processing increases resolution by about twofold at the same

transient length, which results in a maximum resolution of 140K at m/z 200 and a maximum scan speed of 12 Hz at a resolution of 17.5K.

Spectrum multiplexing (msx) and parallel ion injection/detection significantly improvesduty cycle.

FIGURE 2: Spectrum multiplexing and parallel ion injection/detection provide high-throughput analysis.

Orbitrap FTMS acquisition scan 2 AGC

Orbitrap FTMS acquisition scan 1 AGC

Collecting ions for scan 2 Collecting ions for scan 3

Up to 10 target ions can be isolated by the quadrupoles sequentially, stored in the C-trap, then transferred to the Orbitrap mass analyzer and detected with high resolution simultaneously. The isolation and trapping of the target ions is also concurrent with the Orbitrap mass analyzer detection of the previous ion packet.

Full or SIM spectra were acquired at three different resolutions: 35K, 70K and 140K. Baseline separation of two ions that are only 30 ppm apart was only achievable at the resolution of 140K. The high mass accuracy (< 5 ppm) and high resolution ensured accurate identification of target peptides from complex background with msx tSIM (Figure 3).

FIGURE 3: High resolution ensures accurate target selection.

FIGURE 4: High sensitivity with quadrupole-based SIM scan.

Target peptides in 10 ng of E coli digest were analyzed with alternating full scans and tSIM scans (Figure 4). The peptide GLILVGGYGTR* (558.3259, 2+) is barely visible in the full scan with a 600 amu isolation window (Figure 4A), but is accurately identified with a S/N of 65 and mass deviation of ~0.2 ppm in the SIM scan with a 4 amu isolation window (Figure 4B). The +1 isotope peak is also accurately identified with a S/N of 39 and mass deviation of ~1.4 ppm in the SIM scan.

FIGURE 5. Multiplexing improves throughput and results in accurate quantification.

17.0 17.2 17.4 17.6 17.8 18.0 18.2 18.4 18.6

Time (min)

020406080

1000

20406080

100

Rel

ativ

e A

bund

ance

020406080

100 18.12

18.11

18.10

GISNEGQNASIK*

5 ppm XIC

100 fmol

1 fmol

10 amol

Extracted ion chromatograms (XIC) of target peptides in SIM scans were obtained with 5 ppm mass tolerance. The XIC of GISNEGQNASIK* (613.3167, 2+) at 100 fmol, 1 fmol and 10 amol is displayed in Figure 5. Although GISNEGQNASIK* was monitored with two other peptides in the same time range, more than 10 SIM scans with a resolution of 140K were obtained across a 12 sec LC peak, even at a sample amount of 10 amol.

As shown in Figure 6, a 4-order linear dynamic range was obtained for most peptides with the msx tSIM method. Table 1 lists the LOD and LOQ of target peptides for the method. The CV% from triplicate analyses is also included for the LOQ.

FIGURE 6. Four-order linear dynamic range with msx tSIM method.

Pea

k A

rea

Pea

k A

rea

(log 1

0)

Sample Amount (fmol)Sample Amount (log10)

R2 > 0.9998 R2 > 0.9997

IGDYAGIK*

SSAAPPPPPR*GISNEGQNASIK*

FIGURE 7. HCD spectrum of peptide NGFILDGFPR* (573.305, 2+) at 50 amol in 500 ng E coli tryptic digest background.

TABLE 1. LOD and LOQ of target peptides for the msx tSIM method.

LOD (amol) LOQ (amol) CV% (LOQ)IGDYAGIK* 10 50 3SSAAPPPPPR* 10 50 3HVLTSIGEK* 10 50 3LTILEELR* <10 10 4GLILVGGYGTR* 10 50 7NGFILDGFPR* <10 10 14SAAGAFGPELSR* <10 10 4GISNEGQNASIK* <10 10 10ELASGLSFPVGFK* <10 10 10TASEFDSAIAQDK* 10 50 1SFANQPLEVVYSK* <10 10 8ELGQSGVDTYLQTK* 10 50 7LSSEAPALFQFDLK* <50 50 3GILFVGSGVSGGEEGAR* 50 100 5

400 500 600 700 800 900 1000m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

714.3813y6

486.2699y4

601.2971y5

974.5300y8

With tHCD, high resolution and high mass accuracy allow simultaneous accurate identification of multiple fragment ions from complex mixed MS2 spectra (Figure 7). As shown in Figure 8, 3-4 orders of linear dynamic range were obtained for most peptides with the tHCD method. Table 2 lists the LOD and LOQ of target peptides for the tHCD method. The CV% from triplicate analyses is also included for the LOQ.

Targeted Quantification with tHCD

50 amol

ConclusionThe Q Exactive™ MS is well suited for targeted quantification providing two powerful HR/AM approaches: msx tSIM and tHCD. Both approaches demonstrate high selectivity, high sensitivity and high throughput due to several hardware and software innovations.

For the msx tSIM approach, high mass accuracy (< 5 ppm) and high resolution (140K) ensure accurate identification of target ions in the presence of a complex background, while spectrum multiplexing dramatically increases throughput without sacrificing resolution.

A LOD of 10 amol and a 4-order linear dynamic range was obtained for most peptides with the msx tSIM method in the presence of a medium complex background.

Different from triple-quadrupole-based SRM, all fragment ions are simultaneously detected with high mass accuracy and high resolution using the tHCD approach. The HR/AM of fragment ions provides unmatched selectivity. The simultaneous detection of all fragment ions allows more efficient sampling of target peptides and results in improved sensitivity. A scan speed of 12 Hz at a resolution of 17,500 ensures high throughput.

A LOD of 50 amol and 3–4 orders of linear dynamic range were obtained for most peptides with the tHCD method in the presence of a strong complex background.

FIGURE 8. Three to four orders of linear dynamic range with the tHCD method.

Pea

k A

rea

Pea

k A

rea

(log 1

0)

Sample Amount (fmol)Sample Amount (log10)

R2 > 0.999 R2 > 0.998

SFANQPLEVVYSK*LTILEELR*NGFILDGFPR*

TABLE 2. LOD and LOQ of target peptides for the tHCD method.

LOD (amol) LOQ (amol) CV% (LOQ)IGDYAGIK* >10 50 6SSAAPPPPPR* 10 50 11HVLTSIGEK* 50 100 11LTILEELR* 10 50 3GLILVGGYGTR* >10 50 8NGFILDGFPR* >10 50 6SAAGAFGPELSR* 10 50 4GISNEGQNASIK* 10 50 5ELASGLSFPVGFK* >10 50 2TASEFDSAIAQDK 10 50 1SFANQPLEVVYSK* <10 10 12ELGQSGVDTYLQTK* 10 50 6LSSEAPALFQFDLK* 10 50 6GILFVGSGVSGGEEGAR* >10 50 8

400 500 600 700 800 900 10000

20

40

60

80

100

Rel

ativ

e A

bund

ance

557.0 558.0 559.0 560.0m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

558.3 558.5 558.7

558.3302S/N: 7

Full Scan (400 -1000amu)Scan #: 3099

SIM Scan (556.3 - 560.3amu)Scan #: 3100

558.3 558.5 558.7

558.3258(S/N: 65)

558.8282(S/N: 39)

A.

B.

Capillary

S-Lens forImproved Sensitivity

BentFlatapole

Quadrupole Mass Filter forPrecursor Ion Selection

HCD Cell, C-Trap forSpectrum Multiplexing

Orbitrap withEnhanced Resolution

30 ppm

m/z

Magic is a trademark of Michrom BioResources, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

Page 5: HR/AM Targeted Peptide Quantification on a Q Exactive MS: A …apps.thermoscientific.com/media/SID/LSMS/PDF/ASMS/ASMS12/... · 2012-06-25 · FIGURE 7. HCD spectrum of peptide NGFILDGFPR*

5Thermo Scientific Poster Note • PN63586_E 06/12S

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, andHigh Throughput Yi Zhang1, Zhiqi Hao1, Markus Kellmann2 and Andreas FR. Huhmer1

1Thermo Fisher Scientific, San Jose, CA, USA; 2Thermo Fisher Scientific, Bremen, Germany

OverviewPurpose: The performance of a novel hybrid quadrupole-OrbitrapTM mass spectrometer was evaluated in targeted peptide quantification.

Methods: Two high-resolution, accurate-mass (HR/AM) targeted quantification methods were used to analyze 14 peptide standards spiked into a complex E. colidigest background. The data were automatically processed using specialized software.

Results: A detection limit of 10 amol and a linear dynamic range of 4 orders were obtained with msx tSIM methods in a moderately complex background. A detection limit of 10–50 amol and a linear dynamic range of 3–4 orders were obtained with tHCD methods in a highly complex background.

IntroductionQuantitative proteomics enables the identification of a large number of protein candidates, which display biologically interesting dynamics on a global scale in the early discovery phase. A targeted MS approach, in particular selected reaction monitoring (SRM) on a triple quadrupole MS, has become the standard technique for quantitatively analyzing tens to hundreds of peptide candidates across large numbers of samples, either for understanding of signaling regulation or for verification and selection of potential biomarkers. However, the low resolution of triple quadrupole MS limits its ability to achieve high selectivity on the targets from complex backgrounds. It is also difficult to develop SRM method for peptides with high charge states or high mass or for modified peptides.

In this study, a true high-resolution, accurate-mass (HR/AM) mass spectrometer, was evaluated for targeted protein quantification. The dynamic range and LOD/LOQ of two HR/AM methods: spectrum multiplexing targeted selected ion monitoring (msx tSIM) and targeted HCD (tHCD), were investigated.

MethodsSample Preparation

Thermo Scientific heavy-isotope labeled peptide retention standards were spiked into either 5 ng/µL or 250 ng/µL E coli. whole cell tryptic digest to reach final concentrations of 0 amol/µL, 5 amol/µL, 25 amol/µL, 50 amol/µL, 500 amol/µL, 5 fmol/µL, and 50 fmol/µL.

Liquid Chromatography

Peptides from 2 µL of each sample were separated on a Michrom Magic™ C18 nanoLC column (75 μm x 15 cm, 3 μm particle) with a 1 hr gradient at a flow rate of 350 nL/min. LC solvents were 0.1% formic acid in H2O (Solvent A) and 0.1% formic acid in acetonitrile (Solvent B). The LC gradient was 0-2 min, 3% B; 2-47 min, 3%-30% B; 47-52 min, 30%-95% B; 52-54 min, 95% B; 54-55 min, 95%-5% B; 55-60 min, 5% B.

Mass Spectrometry

The samples with 5 ng/µL E coli digest background were analyzed in triplicate with a Full-msx tSIM method on a Thermo Scientific Q Exactive hybrid quadrupole-Orbitrap MS (Figure 1). The resolution was set at 140K for both scan types. The AGC target was 1E6 for full scan and 2E5 for SIM scan. The maximum ion injection time was 120 ms for full scan and 500 ms for SIM scan. The isolation width for SIM scan was 4 amu. The peptide retention standards were monitored over a 3 min window. The multiplexing level was set at 4, which allowed isolation and accumulation of up to four peptide targets in the C-trap before they were transferred to the Orbitrap mass analyzer for detection (Figure 2).

The samples with 250 ng/µL E coli digest background were analyzed in triplicate with a tHCD method. The resolution was set at 17.5K. The AGC target was 5E5. The maximum ion injection time was 500 ms. The quadrupole isolation width was 2.0 amu. The normalized collision energy (NCE) was 27%.The peptide retention standards were monitored over a 3 min window.

Data Analysis

The data were automatically processed using Thermo Scientific Pinpoint software. Extracted ion chromatograms for peptide retention standards were obtained with a 5 ppm mass tolerance. Peak areas were calculated with 7-point smoothed complete chromatograms. Linear fitting was performed with a weighing scheme of 1/x.

ResultsTargeted Quantification with msx tSIM

FIGURE 1. Schematic and key features of the Q Exactive MS.

Key Features of the Q Exactive MS

Incorporation of the S-lens at the source enhances sensitivity by up to 5 fold. Quadrupole mass filter enables precursor selection for data-dependent MS2 and

selected ion monitoring (SIM). Advanced signal processing increases resolution by about twofold at the same

transient length, which results in a maximum resolution of 140K at m/z 200 and a maximum scan speed of 12 Hz at a resolution of 17.5K.

Spectrum multiplexing (msx) and parallel ion injection/detection significantly improvesduty cycle.

FIGURE 2: Spectrum multiplexing and parallel ion injection/detection provide high-throughput analysis.

Orbitrap FTMS acquisition scan 2 AGC

Orbitrap FTMS acquisition scan 1 AGC

Collecting ions for scan 2 Collecting ions for scan 3

Up to 10 target ions can be isolated by the quadrupoles sequentially, stored in the C-trap, then transferred to the Orbitrap mass analyzer and detected with high resolution simultaneously. The isolation and trapping of the target ions is also concurrent with the Orbitrap mass analyzer detection of the previous ion packet.

Full or SIM spectra were acquired at three different resolutions: 35K, 70K and 140K. Baseline separation of two ions that are only 30 ppm apart was only achievable at the resolution of 140K. The high mass accuracy (< 5 ppm) and high resolution ensured accurate identification of target peptides from complex background with msx tSIM (Figure 3).

FIGURE 3: High resolution ensures accurate target selection.

FIGURE 4: High sensitivity with quadrupole-based SIM scan.

Target peptides in 10 ng of E coli digest were analyzed with alternating full scans and tSIM scans (Figure 4). The peptide GLILVGGYGTR* (558.3259, 2+) is barely visible in the full scan with a 600 amu isolation window (Figure 4A), but is accurately identified with a S/N of 65 and mass deviation of ~0.2 ppm in the SIM scan with a 4 amu isolation window (Figure 4B). The +1 isotope peak is also accurately identified with a S/N of 39 and mass deviation of ~1.4 ppm in the SIM scan.

FIGURE 5. Multiplexing improves throughput and results in accurate quantification.

17.0 17.2 17.4 17.6 17.8 18.0 18.2 18.4 18.6

Time (min)

020406080

1000

20406080

100

Rel

ativ

e A

bund

ance

020406080

100 18.12

18.11

18.10

GISNEGQNASIK*

5 ppm XIC

100 fmol

1 fmol

10 amol

Extracted ion chromatograms (XIC) of target peptides in SIM scans were obtained with 5 ppm mass tolerance. The XIC of GISNEGQNASIK* (613.3167, 2+) at 100 fmol, 1 fmol and 10 amol is displayed in Figure 5. Although GISNEGQNASIK* was monitored with two other peptides in the same time range, more than 10 SIM scans with a resolution of 140K were obtained across a 12 sec LC peak, even at a sample amount of 10 amol.

As shown in Figure 6, a 4-order linear dynamic range was obtained for most peptides with the msx tSIM method. Table 1 lists the LOD and LOQ of target peptides for the method. The CV% from triplicate analyses is also included for the LOQ.

FIGURE 6. Four-order linear dynamic range with msx tSIM method.

Pea

k A

rea

Pea

k A

rea

(log 1

0)

Sample Amount (fmol)Sample Amount (log10)

R2 > 0.9998 R2 > 0.9997

IGDYAGIK*

SSAAPPPPPR*GISNEGQNASIK*

FIGURE 7. HCD spectrum of peptide NGFILDGFPR* (573.305, 2+) at 50 amol in 500 ng E coli tryptic digest background.

TABLE 1. LOD and LOQ of target peptides for the msx tSIM method.

LOD (amol) LOQ (amol) CV% (LOQ)IGDYAGIK* 10 50 3SSAAPPPPPR* 10 50 3HVLTSIGEK* 10 50 3LTILEELR* <10 10 4GLILVGGYGTR* 10 50 7NGFILDGFPR* <10 10 14SAAGAFGPELSR* <10 10 4GISNEGQNASIK* <10 10 10ELASGLSFPVGFK* <10 10 10TASEFDSAIAQDK* 10 50 1SFANQPLEVVYSK* <10 10 8ELGQSGVDTYLQTK* 10 50 7LSSEAPALFQFDLK* <50 50 3GILFVGSGVSGGEEGAR* 50 100 5

400 500 600 700 800 900 1000m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

714.3813y6

486.2699y4

601.2971y5

974.5300y8

With tHCD, high resolution and high mass accuracy allow simultaneous accurate identification of multiple fragment ions from complex mixed MS2 spectra (Figure 7). As shown in Figure 8, 3-4 orders of linear dynamic range were obtained for most peptides with the tHCD method. Table 2 lists the LOD and LOQ of target peptides for the tHCD method. The CV% from triplicate analyses is also included for the LOQ.

Targeted Quantification with tHCD

50 amol

ConclusionThe Q Exactive™ MS is well suited for targeted quantification providing two powerful HR/AM approaches: msx tSIM and tHCD. Both approaches demonstrate high selectivity, high sensitivity and high throughput due to several hardware and software innovations.

For the msx tSIM approach, high mass accuracy (< 5 ppm) and high resolution (140K) ensure accurate identification of target ions in the presence of a complex background, while spectrum multiplexing dramatically increases throughput without sacrificing resolution.

A LOD of 10 amol and a 4-order linear dynamic range was obtained for most peptides with the msx tSIM method in the presence of a medium complex background.

Different from triple-quadrupole-based SRM, all fragment ions are simultaneously detected with high mass accuracy and high resolution using the tHCD approach. The HR/AM of fragment ions provides unmatched selectivity. The simultaneous detection of all fragment ions allows more efficient sampling of target peptides and results in improved sensitivity. A scan speed of 12 Hz at a resolution of 17,500 ensures high throughput.

A LOD of 50 amol and 3–4 orders of linear dynamic range were obtained for most peptides with the tHCD method in the presence of a strong complex background.

FIGURE 8. Three to four orders of linear dynamic range with the tHCD method.

Pea

k A

rea

Pea

k A

rea

(log 1

0)

Sample Amount (fmol)Sample Amount (log10)

R2 > 0.999 R2 > 0.998

SFANQPLEVVYSK*LTILEELR*NGFILDGFPR*

TABLE 2. LOD and LOQ of target peptides for the tHCD method.

LOD (amol) LOQ (amol) CV% (LOQ)IGDYAGIK* >10 50 6SSAAPPPPPR* 10 50 11HVLTSIGEK* 50 100 11LTILEELR* 10 50 3GLILVGGYGTR* >10 50 8NGFILDGFPR* >10 50 6SAAGAFGPELSR* 10 50 4GISNEGQNASIK* 10 50 5ELASGLSFPVGFK* >10 50 2TASEFDSAIAQDK 10 50 1SFANQPLEVVYSK* <10 10 12ELGQSGVDTYLQTK* 10 50 6LSSEAPALFQFDLK* 10 50 6GILFVGSGVSGGEEGAR* >10 50 8

400 500 600 700 800 900 10000

20

40

60

80

100

Rel

ativ

e A

bund

ance

557.0 558.0 559.0 560.0m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

558.3 558.5 558.7

558.3302S/N: 7

Full Scan (400 -1000amu)Scan #: 3099

SIM Scan (556.3 - 560.3amu)Scan #: 3100

558.3 558.5 558.7

558.3258(S/N: 65)

558.8282(S/N: 39)

A.

B.

Capillary

S-Lens forImproved Sensitivity

BentFlatapole

Quadrupole Mass Filter forPrecursor Ion Selection

HCD Cell, C-Trap forSpectrum Multiplexing

Orbitrap withEnhanced Resolution

30 ppm

m/z

Magic is a trademark of Michrom BioResources, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

Page 6: HR/AM Targeted Peptide Quantification on a Q Exactive MS: A …apps.thermoscientific.com/media/SID/LSMS/PDF/ASMS/ASMS12/... · 2012-06-25 · FIGURE 7. HCD spectrum of peptide NGFILDGFPR*

6 HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, andHigh Throughput Yi Zhang1, Zhiqi Hao1, Markus Kellmann2 and Andreas FR. Huhmer1

1Thermo Fisher Scientific, San Jose, CA, USA; 2Thermo Fisher Scientific, Bremen, Germany

OverviewPurpose: The performance of a novel hybrid quadrupole-OrbitrapTM mass spectrometer was evaluated in targeted peptide quantification.

Methods: Two high-resolution, accurate-mass (HR/AM) targeted quantification methods were used to analyze 14 peptide standards spiked into a complex E. colidigest background. The data were automatically processed using specialized software.

Results: A detection limit of 10 amol and a linear dynamic range of 4 orders were obtained with msx tSIM methods in a moderately complex background. A detection limit of 10–50 amol and a linear dynamic range of 3–4 orders were obtained with tHCD methods in a highly complex background.

IntroductionQuantitative proteomics enables the identification of a large number of protein candidates, which display biologically interesting dynamics on a global scale in the early discovery phase. A targeted MS approach, in particular selected reaction monitoring (SRM) on a triple quadrupole MS, has become the standard technique for quantitatively analyzing tens to hundreds of peptide candidates across large numbers of samples, either for understanding of signaling regulation or for verification and selection of potential biomarkers. However, the low resolution of triple quadrupole MS limits its ability to achieve high selectivity on the targets from complex backgrounds. It is also difficult to develop SRM method for peptides with high charge states or high mass or for modified peptides.

In this study, a true high-resolution, accurate-mass (HR/AM) mass spectrometer, was evaluated for targeted protein quantification. The dynamic range and LOD/LOQ of two HR/AM methods: spectrum multiplexing targeted selected ion monitoring (msx tSIM) and targeted HCD (tHCD), were investigated.

MethodsSample Preparation

Thermo Scientific heavy-isotope labeled peptide retention standards were spiked into either 5 ng/µL or 250 ng/µL E coli. whole cell tryptic digest to reach final concentrations of 0 amol/µL, 5 amol/µL, 25 amol/µL, 50 amol/µL, 500 amol/µL, 5 fmol/µL, and 50 fmol/µL.

Liquid Chromatography

Peptides from 2 µL of each sample were separated on a Michrom Magic™ C18 nanoLC column (75 μm x 15 cm, 3 μm particle) with a 1 hr gradient at a flow rate of 350 nL/min. LC solvents were 0.1% formic acid in H2O (Solvent A) and 0.1% formic acid in acetonitrile (Solvent B). The LC gradient was 0-2 min, 3% B; 2-47 min, 3%-30% B; 47-52 min, 30%-95% B; 52-54 min, 95% B; 54-55 min, 95%-5% B; 55-60 min, 5% B.

Mass Spectrometry

The samples with 5 ng/µL E coli digest background were analyzed in triplicate with a Full-msx tSIM method on a Thermo Scientific Q Exactive hybrid quadrupole-Orbitrap MS (Figure 1). The resolution was set at 140K for both scan types. The AGC target was 1E6 for full scan and 2E5 for SIM scan. The maximum ion injection time was 120 ms for full scan and 500 ms for SIM scan. The isolation width for SIM scan was 4 amu. The peptide retention standards were monitored over a 3 min window. The multiplexing level was set at 4, which allowed isolation and accumulation of up to four peptide targets in the C-trap before they were transferred to the Orbitrap mass analyzer for detection (Figure 2).

The samples with 250 ng/µL E coli digest background were analyzed in triplicate with a tHCD method. The resolution was set at 17.5K. The AGC target was 5E5. The maximum ion injection time was 500 ms. The quadrupole isolation width was 2.0 amu. The normalized collision energy (NCE) was 27%.The peptide retention standards were monitored over a 3 min window.

Data Analysis

The data were automatically processed using Thermo Scientific Pinpoint software. Extracted ion chromatograms for peptide retention standards were obtained with a 5 ppm mass tolerance. Peak areas were calculated with 7-point smoothed complete chromatograms. Linear fitting was performed with a weighing scheme of 1/x.

ResultsTargeted Quantification with msx tSIM

FIGURE 1. Schematic and key features of the Q Exactive MS.

Key Features of the Q Exactive MS

Incorporation of the S-lens at the source enhances sensitivity by up to 5 fold. Quadrupole mass filter enables precursor selection for data-dependent MS2 and

selected ion monitoring (SIM). Advanced signal processing increases resolution by about twofold at the same

transient length, which results in a maximum resolution of 140K at m/z 200 and a maximum scan speed of 12 Hz at a resolution of 17.5K.

Spectrum multiplexing (msx) and parallel ion injection/detection significantly improvesduty cycle.

FIGURE 2: Spectrum multiplexing and parallel ion injection/detection provide high-throughput analysis.

Orbitrap FTMS acquisition scan 2 AGC

Orbitrap FTMS acquisition scan 1 AGC

Collecting ions for scan 2 Collecting ions for scan 3

Up to 10 target ions can be isolated by the quadrupoles sequentially, stored in the C-trap, then transferred to the Orbitrap mass analyzer and detected with high resolution simultaneously. The isolation and trapping of the target ions is also concurrent with the Orbitrap mass analyzer detection of the previous ion packet.

Full or SIM spectra were acquired at three different resolutions: 35K, 70K and 140K. Baseline separation of two ions that are only 30 ppm apart was only achievable at the resolution of 140K. The high mass accuracy (< 5 ppm) and high resolution ensured accurate identification of target peptides from complex background with msx tSIM (Figure 3).

FIGURE 3: High resolution ensures accurate target selection.

FIGURE 4: High sensitivity with quadrupole-based SIM scan.

Target peptides in 10 ng of E coli digest were analyzed with alternating full scans and tSIM scans (Figure 4). The peptide GLILVGGYGTR* (558.3259, 2+) is barely visible in the full scan with a 600 amu isolation window (Figure 4A), but is accurately identified with a S/N of 65 and mass deviation of ~0.2 ppm in the SIM scan with a 4 amu isolation window (Figure 4B). The +1 isotope peak is also accurately identified with a S/N of 39 and mass deviation of ~1.4 ppm in the SIM scan.

FIGURE 5. Multiplexing improves throughput and results in accurate quantification.

17.0 17.2 17.4 17.6 17.8 18.0 18.2 18.4 18.6

Time (min)

020406080

1000

20406080

100

Rel

ativ

e A

bund

ance

020406080

100 18.12

18.11

18.10

GISNEGQNASIK*

5 ppm XIC

100 fmol

1 fmol

10 amol

Extracted ion chromatograms (XIC) of target peptides in SIM scans were obtained with 5 ppm mass tolerance. The XIC of GISNEGQNASIK* (613.3167, 2+) at 100 fmol, 1 fmol and 10 amol is displayed in Figure 5. Although GISNEGQNASIK* was monitored with two other peptides in the same time range, more than 10 SIM scans with a resolution of 140K were obtained across a 12 sec LC peak, even at a sample amount of 10 amol.

As shown in Figure 6, a 4-order linear dynamic range was obtained for most peptides with the msx tSIM method. Table 1 lists the LOD and LOQ of target peptides for the method. The CV% from triplicate analyses is also included for the LOQ.

FIGURE 6. Four-order linear dynamic range with msx tSIM method.

Pea

k A

rea

Pea

k A

rea

(log 1

0)

Sample Amount (fmol)Sample Amount (log10)

R2 > 0.9998 R2 > 0.9997

IGDYAGIK*

SSAAPPPPPR*GISNEGQNASIK*

FIGURE 7. HCD spectrum of peptide NGFILDGFPR* (573.305, 2+) at 50 amol in 500 ng E coli tryptic digest background.

TABLE 1. LOD and LOQ of target peptides for the msx tSIM method.

LOD (amol) LOQ (amol) CV% (LOQ)IGDYAGIK* 10 50 3SSAAPPPPPR* 10 50 3HVLTSIGEK* 10 50 3LTILEELR* <10 10 4GLILVGGYGTR* 10 50 7NGFILDGFPR* <10 10 14SAAGAFGPELSR* <10 10 4GISNEGQNASIK* <10 10 10ELASGLSFPVGFK* <10 10 10TASEFDSAIAQDK* 10 50 1SFANQPLEVVYSK* <10 10 8ELGQSGVDTYLQTK* 10 50 7LSSEAPALFQFDLK* <50 50 3GILFVGSGVSGGEEGAR* 50 100 5

400 500 600 700 800 900 1000m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

714.3813y6

486.2699y4

601.2971y5

974.5300y8

With tHCD, high resolution and high mass accuracy allow simultaneous accurate identification of multiple fragment ions from complex mixed MS2 spectra (Figure 7). As shown in Figure 8, 3-4 orders of linear dynamic range were obtained for most peptides with the tHCD method. Table 2 lists the LOD and LOQ of target peptides for the tHCD method. The CV% from triplicate analyses is also included for the LOQ.

Targeted Quantification with tHCD

50 amol

ConclusionThe Q Exactive™ MS is well suited for targeted quantification providing two powerful HR/AM approaches: msx tSIM and tHCD. Both approaches demonstrate high selectivity, high sensitivity and high throughput due to several hardware and software innovations.

For the msx tSIM approach, high mass accuracy (< 5 ppm) and high resolution (140K) ensure accurate identification of target ions in the presence of a complex background, while spectrum multiplexing dramatically increases throughput without sacrificing resolution.

A LOD of 10 amol and a 4-order linear dynamic range was obtained for most peptides with the msx tSIM method in the presence of a medium complex background.

Different from triple-quadrupole-based SRM, all fragment ions are simultaneously detected with high mass accuracy and high resolution using the tHCD approach. The HR/AM of fragment ions provides unmatched selectivity. The simultaneous detection of all fragment ions allows more efficient sampling of target peptides and results in improved sensitivity. A scan speed of 12 Hz at a resolution of 17,500 ensures high throughput.

A LOD of 50 amol and 3–4 orders of linear dynamic range were obtained for most peptides with the tHCD method in the presence of a strong complex background.

FIGURE 8. Three to four orders of linear dynamic range with the tHCD method.

Pea

k A

rea

Pea

k A

rea

(log 1

0)

Sample Amount (fmol)Sample Amount (log10)

R2 > 0.999 R2 > 0.998

SFANQPLEVVYSK*LTILEELR*NGFILDGFPR*

TABLE 2. LOD and LOQ of target peptides for the tHCD method.

LOD (amol) LOQ (amol) CV% (LOQ)IGDYAGIK* >10 50 6SSAAPPPPPR* 10 50 11HVLTSIGEK* 50 100 11LTILEELR* 10 50 3GLILVGGYGTR* >10 50 8NGFILDGFPR* >10 50 6SAAGAFGPELSR* 10 50 4GISNEGQNASIK* 10 50 5ELASGLSFPVGFK* >10 50 2TASEFDSAIAQDK 10 50 1SFANQPLEVVYSK* <10 10 12ELGQSGVDTYLQTK* 10 50 6LSSEAPALFQFDLK* 10 50 6GILFVGSGVSGGEEGAR* >10 50 8

400 500 600 700 800 900 10000

20

40

60

80

100

Rel

ativ

e A

bund

ance

557.0 558.0 559.0 560.0m/z

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

558.3 558.5 558.7

558.3302S/N: 7

Full Scan (400 -1000amu)Scan #: 3099

SIM Scan (556.3 - 560.3amu)Scan #: 3100

558.3 558.5 558.7

558.3258(S/N: 65)

558.8282(S/N: 39)

A.

B.

Capillary

S-Lens forImproved Sensitivity

BentFlatapole

Quadrupole Mass Filter forPrecursor Ion Selection

HCD Cell, C-Trap forSpectrum Multiplexing

Orbitrap withEnhanced Resolution

30 ppm

m/z

Magic is a trademark of Michrom BioResources, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

Page 7: HR/AM Targeted Peptide Quantification on a Q Exactive MS: A …apps.thermoscientific.com/media/SID/LSMS/PDF/ASMS/ASMS12/... · 2012-06-25 · FIGURE 7. HCD spectrum of peptide NGFILDGFPR*

Thermo Fisher Scientific, San Jose, CA USA is ISO Certified.

PN63586_E 06/12S

Africa-Other +27 11 570 1840Australia +61 3 9757 4300Austria +43 1 333 50 34 0Belgium +32 53 73 42 41Canada +1 800 530 8447China +86 10 8419 3588Denmark +45 70 23 62 60

Europe-Other +43 1 333 50 34 0Finland/Norway/Sweden +46 8 556 468 00France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9434Italy +39 02 950 591

Japan +81 45 453 9100Latin America +1 561 688 8700Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Russia/CIS +43 1 333 50 34 0South Africa +27 11 570 1840

Spain +34 914 845 965Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752

www.thermoscientific.com©2012 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.