Flow Measurement SITRANS F US Clamp-on SITRANS FUP1010 (Portable) 4/318 Siemens FI 01 · 2012 4 Overview SITRANS FUP1010 clamp-on non-intrusive ultrasonic flow trans- mitter offers maximum versatility plus battery power for portable field use. It can operate in either WideBeam transit time or re- flexor (Doppler) mode, making it suitable for virtually any liquid, even those with high aeration or suspended solids. SITRANS FUP1010 is available in single and dual channel or dual path configurations, with IP67 weatherproof enclosure. Benefits Battery power facilitates field use; the meter is easily trans- ported from one installation to another – saving time for sur- veys, monitoring and temporary installations Weatherproof enclosure can be used outdoors and left in place without concern for rain damage Rugged plastic case enables it to withstand rough treatment that would destroy most other meters Versatility - there is no need to change meters when operating conditions change Easy installation; no need to cut pipe or stop flow Minimal maintenance; external sensors do not require peri- odic cleaning No moving parts to wear or foul No pressure drop or energy loss Wide turn-down ratio Choice of single or dual channel models minimizes total cost ZeroMatic Path automatically sets zero without stopping flow and reduces zero drift, even at low flow Note that the SITRANS FUP1010 flow transmitter is not avail- able with hazardous area approvals Application SITRANS FUP1010 is suitable for a wide variety of liquid appli- cations, including the following: Water industry - Raw water - Potable water - Chemicals Wastewater industry - Raw sewage - Effluent - Sludges - Mixed liquor - Chemicals HVAC industry - Chillers - Condensers - Hot and cold water systems - Thermal energy rate and total Power industry - Nuclear - Fossil - Hydroelectric Processing industry - Process control - Batching - Rate indication - Volumetric and mass measurement Design IP67 Weatherproof / Impact resistant enclosure constructed of mineral reinforced copolymer polypropylene - Single channel - Dual channel / dual path Function Integral 33 button keypad and large (128 x 240 pixel) graphic display visible up to 12 m (40 ft) away Current, voltage, frequency and RS232 outputs (see specification section for details) Optional current, voltage and temperature inputs (see specification section for details) ZeroMatic Path automatically sets zero Bi-directional flow operation 1 MByte data logger with both site and data logger storage English, Spanish, German, Italian and French language op- tions © Siemens AG 2011

Transcript of Flow Measurement SITRANS F US Clamp-on · Flow Measurement SITRANS F US Clamp-on SITRANS FUP1010...

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUP1010 (Portable)

4/318 Siemens FI 01 · 2012

4

■ Overview

SITRANS FUP1010 clamp-on non-intrusive ultrasonic flow trans-mitter offers maximum versatility plus battery power for portable field use. It can operate in either WideBeam transit time or re-flexor (Doppler) mode, making it suitable for virtually any liquid, even those with high aeration or suspended solids.

SITRANS FUP1010 is available in single and dual channel or dual path configurations, with IP67 weatherproof enclosure.

■ Benefits

• Battery power facilitates field use; the meter is easily trans-ported from one installation to another – saving time for sur-veys, monitoring and temporary installations

• Weatherproof enclosure can be used outdoors and left in place without concern for rain damage

• Rugged plastic case enables it to withstand rough treatment that would destroy most other meters

• Versatility - there is no need to change meters when operating conditions change

• Easy installation; no need to cut pipe or stop flow• Minimal maintenance; external sensors do not require peri-

odic cleaning• No moving parts to wear or foul• No pressure drop or energy loss• Wide turn-down ratio• Choice of single or dual channel models minimizes total cost• ZeroMatic Path automatically sets zero without stopping flow

and reduces zero drift, even at low flow• Note that the SITRANS FUP1010 flow transmitter is not avail-

able with hazardous area approvals

■ Application

SITRANS FUP1010 is suitable for a wide variety of liquid appli-cations, including the following:• Water industry

- Raw water- Potable water- Chemicals

• Wastewater industry - Raw sewage- Effluent- Sludges- Mixed liquor- Chemicals

• HVAC industry - Chillers- Condensers- Hot and cold water systems- Thermal energy rate and total

• Power industry - Nuclear- Fossil- Hydroelectric

• Processing industry - Process control- Batching- Rate indication- Volumetric and mass measurement

■ Design

• IP67 Weatherproof / Impact resistant enclosure constructed of mineral reinforced copolymer polypropylene- Single channel- Dual channel / dual path

■ Function

• Integral 33 button keypad and large (128 x 240 pixel) graphic display visible up to 12 m (40 ft) away

• Current, voltage, frequency and RS232 outputs(see specification section for details)

• Optional current, voltage and temperature inputs(see specification section for details)

• ZeroMatic Path automatically sets zero• Bi-directional flow operation• 1 MByte data logger with both site and data logger storage• English, Spanish, German, Italian and French language op-

tions

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUP1010 (Portable)

4/319Siemens FI 01 · 2012

4

■ Technical specifications

Input

Flow range ± 12 m/s (± 40 ft/s), bidirectional

Pipe size 6.4 mm … 9.14 m (0.25“ … 360“)

Inputs,single channel

• Current: 2 x 4 … 20 mA DC• Voltage: 2 x 0 … 10 V DC• Temperature: 2 x 4 wire 1 kΩ

RTD

Output

Outputs • Current: 2 x 4 … 20 mA DC(1 kΩ at 30 V DC)

• Voltage: 2 x 0 … 10 V DC(5 kΩ minimum)

• Status Alarm: 4 x SPDT Relays• Frequency: 2 x 0 … 5000 Hz• RS232

Accuracy

Accuracy ± 0.5 % … 2 % of flow,for velocities greater than 0.3 m/s (1 ft/s)± 0.0015 … 0.006 m/s (± 0.005 … 0.02 ft/s),for velocities less than 0.3 m/s (1 ft/s)

Batch repeatability ± 0.15 % of flow,for velocities greater than 0.3 m/s (1 ft/s)± 0.0005 m/s (± 0.0015 ft/s),for velocities less than 0.3 m/s (1 ft/s)

Rated operation conditions

Degree of protection

• Weatherproof/impact resistant • IP67

Liquid temperature

• Standard -40 … +120 °C (-40 … +250 °F)

• Optional -40 … +230 °C (-40 … +450 °F)

Ambient temperature -18 … +60 °C (0 … 140 °F)

Design

Dimensions see SITRANS F US Clamp-on "System info and selection guide"

Weight see diagrams

Power supply

Power Internal rechargeable battery

Battery operation 7 hours

Indication and operation

Data logger memory 1 MByte

Site storage memory 50 sites minimum

Display 128 x 240 pixel LCD with back-light

Keypad 33 keypad buttons with tactile feedback

Language options English, Spanish, German, Italian, French

Certificates and approvals

Unclassified locations ULULc

Classified locations

CE EMC Directive 2004/108/ECATEX Directive 94/9/EC

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUP1010 (Portable)

4/320 Siemens FI 01 · 2012

4

Standard MLFB for quick delivery on SITRANS FUP1010 Portable (excluding energy)

Selection and Ordering data Order No. Order Code

SITRANS FUP1010 (Portable)

• IP67 weatherproof battery powered L) 7ME3510-

77777 - 0 77 0 777

Number of channels/ultrasonic pathsSingle channel 1Dual channel/Dual path 2

Standard flowmeter types and I/O configurations

• Standard I/O- 2 x 4 ... 20 mA analog in- 2 x RTD input

C

Sensor cablesNo sensor cable A1 x PVC Jacket cable, length 6 m/20 ft B2 x PVC Jacket cable, length 6 m/20 ft C1 x PVC Jacket cable, length 15 m/50 ft D2 x PVC Jacket cable, length 15 m/50 ft E

RTD temperature Sensor(Mounting hardware & cable included)No RTDs 01 x standard clamp-on RTD, 6 m/20 ft cable 52 x standard clamp-on RTD with 6 m/20 ft cable 61 x standard clamp-on RTD with 15 m/50 ft cable 72 x standard clamp-on RTD with 15 m/50 ft cable 8

Battery charger optionsNo battery charger 0Charger Type A for Europe (CEE7/7) 1Charger Type G for U.S. (NEMA 5-15P) 5

Sensor for channel 1(includes pipe mounting kit and spacer bar for indicated max. outer diameter listed) See "Sensor selection charts" for specifica-tions.

no sensor AA2 universal Trackmount and straps

provided up to 75 mm (3”)

B

B3 universal Trackmount and straps provided up to125 mm (5”)

C

C3 universal Mounting frame and straps provided up to 300 mm (13”)

D

D3 universal Mounting frame and straps provided up to 600 mm (24”)

E

E2 universal Mounting frame and straps provided up to 600 mm (24”)

F

C1H (high precision) Mounting frame and straps provided up to 600 mm (24”)

M

C2H (high precision) Mounting frame and straps provided up to 600 mm (24”)

N

D1H (high precision) Mounting frame and straps provided up to 600 mm (24”)

P

D4H (high precision) Mounting frame and straps provided up to 600 mm (24”)

R

Doppler to 12" with strap kit (not for IP65 (NEMA 7))

S

D1H High temperature range 104 °C/ 220 °F HP2)

Z P 1 P

Sensor for channel 2(includes pipe mounting kit and spacer bar for indicated max. outer diameter listed)See "Sensor selection charts" for specifica-tions.

no sensor AA2 universal Trackmount and straps

provided up to 75 mm (3”)

B

B3 universal Trackmount and straps provided up to 125 mm (5”)

C

C3 universal Mounting frame and straps provided up to 300 mm (13”)

D

D3 universal Mounting frame and straps provided up to 600 mm (24”)

E

E2 universal Mounting frame and straps provided up to 600 mm (24”)

F

C1H (high precision) Mounting frame and straps provided up to 600 mm (24”)

M

C2H (high precision) Mounting frame and straps provided up to 600 mm (24”)

N

D1H (high precision) Mounting frame and straps provided up to 600 mm (24”)

P

D2H (high precision) Mounting frame and straps provided up to 600 mm (24”)

Q

Doppler to 12" with strap kit (not for IP65 (NEMA 7))

S

D1H High temperature range 104 °C / 220 °F HP2)

Z Q 1 P

Approvals: No options (UL, ULc, CE by default)

Standard MLFB product offering represents 4 to 6 weeks delivery time

L) Subject to export regulations AL: N, ECCN: 3A991X

Selection and Ordering data Order No. Order Code

SITRANS FUP1010 (Portable)

• IP67 weatherproof battery powered L) 7ME3510-

77777 - 0 77 0 777

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUP1010 (Portable)

4/321Siemens FI 01 · 2012

4

Selection and Ordering data Order No. Ord. code

SITRANS FUP1010 (Portable)

• IP67 weatherproof battery powered L) 7ME3510-

77777 - 0 77 0 777

Number of channels/ultrasonic pathsSingle channel 1Dual channel/Dual path 2

Standard flowmeter types and I/O configurations

• Standard I/O- Reflexor capable- Graphic display- 2 x 0 ... 10 V- 2 x 4 ... 20 mA- 2 x pulse outputs- 4 x status logic- 2 x 4 ... 20 mA analog in- 1 x RTD per channel

C

Sensor cables(select proper quantity of active channels)No sensor cable A

IP67 (weatherproof) only1 x PVC-jacket, length 6 m (20 ft)(for IP67 NEMA 6)2)

B

2 x PVC-jacket, length 6 m (20 ft)(for IP67 NEMA 6)2)

C

1 x PVC-jacket, length 15 m (50 ft)(for IP67 NEMA 6)2)

D

2 x PVC-jacket, length 15 m (50 ft)(for IP67 NEMA 6)2)

E

RTD temperature sensor(for type 3 meter only, mounting hardware and cable included)No RTDs 0IP67 (weatherproof) only1 x standard clamp-on RTD (NEMA 6) with 6 m (20 ft) cable1) 5

2 x standard clamp-on RTD (NEMA 6) with 6 m (20 ft) cable1)

6

1 x standard clamp-on RTD (NEMA 6) with 15 m (50 ft) cable1)

7

2 x standard clamp-on RTD (NEMA 6) with 15 m (50 ft) cable1)

8

Battery charger options

no battery charger 0Charger Type A for Europe (CEE7/7) 1Charger Type C for Australia (AS3112) 2Charger Type D for U.K. (BS1363) 3Charger Type J for Japan (JIS8303) 4Charger Type G for U.S. (NEMA 5-15P) 5Charger Type L for Switzerland (SEV1011) 6

Sensor for channel 1Including pipe mounting tracks for sizes A & B sensors indented for pipe with a OD less than 125 mm (5") and mounting frame/spacer bars for sizes C, D & E sensors. Straps provided are for the indicated maximum OD listed below. Strap kits are available to accommodate larger pipes (refer to spare part list). Refer to "Sensor Selection Charts" for the sensor suitability of pipe size and wall thickness.

no sensor AA2 universal Trackmount and straps

provided up to75 mm (3”)

B

B3 universal Trackmount and straps provided up to 125 mm (5”)

C

C3 universal Mounting frame and straps provided up to 300 mm (13”)

D

D3 universal Mounting frame and straps provided up to 600 mm (24”)

E

E2 universal Mounting frame and straps provided up to 600 mm (24”)

F

For the following A1H to D4H sensors, tem-perature range is -40 °C to 65 °C(-41 °F to 150 °F), nominal 21 °C (70 °F):A2H (high precision) Trackmount and straps

provided up to 75 mm (3”)

H

A3H (high precision) Trackmount and straps provided up to 75 mm (3”)

J

B1H (high precision) Trackmount and straps provided up to 125 mm (5”)

K

B2H (high precision) Trackmount and straps provided up to 125 mm (5)

L

C1H (high precision) Mounting frame and straps provided up to 600 mm (24”)

M

C2H (high precision) Mounting frame and straps provided up to 600 mm (24”)

N

D1H (high precision) Mounting frame and straps provided up to 600 mm (24”)

P

D2H (high precision) Mounting frame and straps provided up to 600 mm (24”)

Q

D4H (high precision) Mounting frame and straps provided up to 600 mm (24”)

R

Doppler to 12“ with chain kit SHigh temperature sensor size 2 for up to 230 °C (446 °F) (30 to 200 mm diam. (1.18 to 7.67 inch diam.))

Z P 1 A

High temperature sensor size 3 for up to 230 °C (446 °F) (150 to 610 mm diam. (5.90 to 24 inch diam.))

Z P 1 B

High temperature sensor size 4 for up to 230 °C (446 °F) (400 to 1200 mm diam. (15.75 to 47.25 inch diam.))

Z P 1 C

L) Subject to export regulations AL: N, ECCN: 3A991X

Selection and Ordering data Order No. Ord. code

SITRANS FUP1010 (Portable)

• IP67 weatherproof battery powered L) 7ME3510-

77777 - 0 77 0 777

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUP1010 (Portable)

4/322 Siemens FI 01 · 2012

4

L) Subject to export regulations AL: N, ECCN: 3A991X.

D) Subject to export regulations AL: N, ECCN: EAR99H.

Sensor for channel 1 (continued)

For the following B1H to D4H sensors, tem-perature range is -1 °C up to 104 °C (30 °F up to 220 °F), nominal 65 °C (150 °F):B1H (high temperature range HP) Z P 1 KB2H (high temperature range HP) Z P 1 LC1H (high temperature range HP) Z P 1 MC2H (high temperature range HP) Z P 1 ND1H (high temperature range HP) Z P 1 PD2H (high temperature range HP) Z P 1 QD4H (high temperature range HP) Z P 1 R

Sensor for channel 2(includes pipe mounting kit and spacer bar for indicated max. outer diameter listed)See "Sensor selection charts" for specifica-tions.

no sensor AA2 universal Trackmount and straps

provided up to 75 mm (3”)

B

B3 universal Trackmount and straps provided up to 125 mm (5”)

C

C3 universal Mounting frame and straps provided up to 300 mm (13”)

D

D3 universal Mounting frame and straps provided up to 600 mm (24”)

E

E2 universal Mounting frame and straps provided up to 600 mm (24”)

F

For the following A1H to D4H sensors, tem-perature range is -40 °C to 65 °C(-41 °F to 150 °F), nominal 21 °C (70 °F):A2H (high precision) Trackmount and straps

provided up to 75 mm (3”)

H

A3H (high precision) Trackmount and straps provided up to75 mm (3”)

J

B1H (high precision) Trackmount and straps provided up to 125 mm (5”)

K

B2H (high precision) Trackmount and straps provided up to 125 mm (5")

L

C1H (high precision) Mounting frame and straps provided up to 600 mm (24”)

M

C2H (high precision) Mounting frame and straps provided up to 600 mm (24”)

N

D1H (high precision) Mounting frame and straps provided up to 600 mm (24”)

P

D2H (high precision) Mounting frame and straps provided up to 600 mm (24”)

Q

D4H (high precision) Mounting frame and straps provided up to 600 mm (24”)

R

Doppler to 12“ with chain kit S

Selection and Ordering data Order No. Ord. code

SITRANS FUP1010 (Portable)

• IP67 weatherproof battery powered L) 7ME3510-

77777 - 0 77 0 777

Sensor for channel 2 (continued)

High temperature sensor size 2 for up to 230 °C (446 °F) (30 to 200 mm diam. (1.18 to 7.67 inch diam.))

Z Q 1 A

High temperature sensor size 3 for up to 230 °C (446 °F) (150 to 610 mm diam. (5.90 to 24 inch diam.))

Z Q 1 B

High temperature sensor size 4 for up to 230 °C (446 °F) (400 to 1200 mm diam. (15.75 to 47.25 inch diam.))

Z Q 1 C

For the following B1H to D4H sensors, tem-perature range is -1 °C up to 104 °C (30 °F up to 220 °F), nominal 65 °C (150 °F):B1H (high temperature range HP) Z Q 1 KB2H (high temperature range HP) Z Q 1 LC1H (high temperature range HP) Z Q 1 MC2H (high temperature range HP) Z Q 1 ND1H (high temperature range HP) Z Q 1 PD2H (high temperature range HP) Z Q 1 QD4H (high temperature range HP) Z Q 1 R

Approvals: No options (UL, ULc, CE by default)1) -40 ... +200 °C (-40 ... +392 °F)2) -40 ... +80 °C (-40 ... +176 °F)

Selection and Ordering data Order No.

Operating Instructions for SITRANS FUP1010

English IP67 Weatherproof D) A5E02951522

German IP67 Weatherproof D) A5E02951534

This device is shipped with a Quick Start Guide and a CD containing further SITRANS F literature.All literature is also available for free at: http://www.siemens.com/flowdocumentation

Selection and Ordering data Order No. Ord. code

SITRANS FUP1010 (Portable)

• IP67 weatherproof battery powered L) 7ME3510-

77777 - 0 77 0 777

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUP1010 (Portable)

4/323Siemens FI 01 · 2012

4

■ MLFB example

Application example

A general survey portable flowmeter is required for pipes sizes ranging from 76 ... 500 mm (3" ... 20") with both cast iron and steel material. Doppler may be required as liquid may be mod-erately aerated.

Requires language support for German.

MLFB Order No.: 7ME3511-2AB01-0DE0-ZB10

Universal sensor selection chart

High precision sensor selection chart

Selection and Ordering data Order code

Further designsPlease add "-Z" to Order No. and specify Order code(s).

Languages (Meter and Documentation), English (default). All languages now come standard in all flow-meters

Wet flow transfer calibration (priced on request)

6 point calibration 2/water (Price per channel)

• 2SS40 pipe D01• 3CS40 pipe D02• 4CS40 pipe D03• 4SS40 pipe D04

• 6CS40 pipe D05• 6SS40 pipe D06• 6CS120 pipe D07• 8CS40 pipe D08

• 8SS40 pipe D09• 8CS120 pipe D10• 10CS Standard pipe D11• 10CS40 pipe D12

• 10SS40 pipe D13• 12CS Standard pipe D14• 12CS40 pipe D15• 14CS30 pipe D16

• 14CS40 pipe D17• 16CS Standard pipe D18• 16CS40 pipe D19• 18CS Standard pipe D20

• 20CS20 pipe D21• 20CS30 pipe D22• 24CS Standard pipe D23• 24CS20 pipe D24

• 24CS30 pipe D25• 30CS Standard pipe D26• 36CS Standard pipe D27• Other pipe, other liquid, additional points, witness Y28

1x Insertion RTD with thermowell and lagging M1A2x Insertion RTD with thermowell and lagging M1B

Tag name plate

• Stainless steel tag with 3.2 mm (0.13 inch) character size (68 characters max.)

Y19

1) Supplied spacer bar supports pipes up to 1050 mm (42 inches). For pipes larger than 1050 mm (42 inches) purchase also, spare part 7ME3960-0MS40 (1012BN-4).

2) Supplied spacer bar supports pipes up to 750 mm (30 inches). For pipes larger than 750 mm (30 inches) purchase also, spare part 7ME3960-0MS40 (1012BN-4).

Selection and Ordering data Order No. Ord. code

SITRANS FUP1010 meter family

7 M E 3 5 1 7 - 77777 - 0 77 0 777

IP67 weatherproof 1

Dual channel 2

Portable Type 3 I/O with Doppler capable, temperature

C

1 x PVC-Jacket, length 6 m (20 ft) B

No RTDs required 0

Charger Type A for Europe (CEE7/7)

1

Sensor for DN 50 ... DN 300(2“ ... 12“) pipes

D

Sensor for DN 200 ... DN 600 (8“ ... 24“) pipes

E

German language support B 1 0

Based on pipe size (pipes other than steel)

Pipe size OrderCode

Outer diameter range (mm)

Outer diameter range (inches)

min. max. min. max.

A2 B 12,7 50,8 0.5 2

B3 C 19 127 0.75 5

C3 D 51 305 2 12

D3 E 203 610 8 24

E2 F 254 6096 10 249

Based on pipe wall thickness (steel pipes only)

Pipe Wall OrderCode

Pipe Wall [mm] Pipe Wall [inches]

min. max. min. max.

A1H G 0,64 1,02 0.025 0.04

A2H H 1,02 1,52 0.04 0.06

A3H J 1,52 2,03 0.06 0.08

B1H K 2,03 3,05 0.08 0.12

B2H L 3,05 4,06 0.12 0.16

C1H M 4,06 5,84 0.16 0.23

C2H N 5,84 8,13 0.23 0.32

D1H P 8,13 11,18 0.32 0.44

D2H Q 11,18 15,75 0.44 0.62

D4H R 15,75 31,75 0.62 1.25

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-onSITRANS FUP1010Water and Liquid check metering kits

4/324 Siemens FI 01 · 2012

4

■ Overview

There are two check kits available: SITRANS FUP1010 Water check metering kit for water and wastewater applications, and SITRANS FUP1010 Liquid check metering kit for liquid applica-tions other than water. The kits have been developed especially for verifying the accuracy and performance of any brand or type of flowmeter. They can be used to verify the performance of me-ters based on any existing flow measurement principle: orifice, electromagnetic, ultrasonic, rotary piston, coriolis, etc. In addi-tion, they measure practically all conductive or non-conductive clean or moderately aerated liquids or liquids with suspended solids.

■ Benefits

• Performance check or verification of any type or brand of flow-meter

• Field use is facilitated by meter portability and 7 hours or nor-mal battery operation.

• Weatherproof enclosure withstands even severe weather con-ditions

• 1 MByte datalogger capability downloadable to PC via in-cluded RS232 cable

• Fast, easy and cost-efficient on-site measurement of any con-voluted pipe from 20 to 1200 mm (0.75 to 48“)

• Delivered as an all inclusive kit in a sturdy rolling case that holds all the equipment needed to conduct performance and verification tests (cables, multiple sensors, flow transmitter etc.)

■ Application

The SITRANS FUP1010 Water and Liquid Check Metering Kits measure practically all conductive or non-conductive clean or moderately aerated liquids or liquids with suspended solids.This basic feature enables the performance check and verification of existing meters used in various water and wastewater applica-tions such as:

Raw Water and sewage• Potable water• Chemicals• Effluent and sludges• Process control• Batching• Rate indication• Hot and cold water systems

■ Design

• IP67 weatherproof/impact resistant enclosure, constructed of mineral reinforced copolymer polypropylene

• Single channel

■ Function

• Integral 33 button keypad and large (128 x 240 pixel) graphic display visible up to 12 m (40 ft) away

• Current, voltage, frequency and RS232 outputs (see Technical specification section for details)

• Optional current, voltage and temperature inputs (see specifi-cation section for details)

• ZeroMatic Path automatically sets zero• Bi-directional flow operation• 1 MByte data logger with both site and data logger storage• English, Spanish, German, Italian and French language op-

tions

■ Technical specifications

Pipe sizes

• Water Check Metering Kit 50 ... 1050 mm (2 ... 42“)

• Liquid Check Metering Kit 20 ... 1200 mm (0.75 ... 48“)

Accuracy ±0.5 % ... ±2.0 % of flow rate

Flow range 12 m/s (40 ft/s) bidirectional

Media temperature -40 ... +104 °C (-40 ... 220 °F)

Enclosure ratings IP67 (Weatherproof)

See page 4/319 for complete technical specifications

Cerificates and approvals

Unclassified locations ULULc

Classified locations

CE EMV Directive 2004/108/ECATEX Directive 94/9/EC

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUP1010Water and Liquid check metering kits

4/325Siemens FI 01 · 2012

4

D) Subject to export regulations AL: N, ECCN: EAR99H.

Selection and Ordering data Order No.

SITRANS FUP1010 Water Check Metering Kit CQO:FUPW-WWKITD)

Content of delivery1 Single channel portable submersible

flow transmitter1 pair Universal sensor C31 pair Universal sensor E21 pair Doppler sensors1 pair Mounting Ezclamp (2 mounting

Ezclamp chains)1 Ladder chain1 Battery charger1 pair 20 ft sensor cable1 Cable - 1010WP/WDP to PC1 PinStop spacer bar (universal)1 Flow case1 Flowmeter manual1 Laminated card set1 Certificate of intrinsic calibration

Selection and Ordering data Order No.

SITRANS FUP1010 Liquid Check Metering Kit CQO:FUS-LIQKITD)

Content of delivery1 Single channel portable submersible

transmitter1 pair Universal sensor B31 pair Universal sensor C31 pair Universal sensor D31 pair Universal sensor E21 pair Doppler sensors1 pair Sensor cables 6m (20 ft)1 pair Mounting track1 pair Mounting Ezclamp1 Spacer bar (portable)1 Ladder chain1 Battery charger1 RS232 cable for PC connection1 Flow case1 Clamp-on flowmeter manual CD1 Flowmeter manual1 Laminated card set1 Certificate of intrinsic calibration

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUE1010 (Energy)

4/326 Siemens FI 01 · 2012

4

■ Overview

SITRANS FUE1010 is a highly accurate clamp-on non-intrusive ultrasonic flow transmitter for revenue grade thermal energy sub-metering and energy efficiency distribution monitoring, with a real time coefficient of performance (COP) for HVAC systems.

SITRANS FUE1010 is available in single and dual channel or dual path configurations, with your choice of IP65 (NEMA 4X) dedicated wall mount or IP40 (NEMA 1) portable enclosures.

■ Benefits

• Measures energy rate and total consumption with highest accuracy available

• Accurately measures at both low flow rates and low differential temperatures

• Easy installation; no need to cut pipe or stop flow• Minimal maintenance; external sensors do not require

periodic cleaning• No moving parts to foul or wear • No pressure drop or energy loss• Wide turn-down ratio• Choice of single or dual channel / dual path or dual mode

operation: - Dual channel operation reduces the cost for the system on a

per channel measurement basis and permits measuring hot and chilled water lines at the same time

- Dual path capability insures high flow measurement accu-racy on installations with less than desirable piping runs

• Ability to operate in either Wide-Beam Transit-time or reflexor (Doppler) mode for applications with high aeration

• ZeroMatic Path automatically sets zero without stopping flow and reduces zero drift, even at low flow

■ Application

SITRANS FUE1010 is ideally suited to thermal energy / power in-dustry applications, including:• Chilled water sub-metering• Hot water sub-metering• Condenser water• Glycol• Thermal storage• Lake source cooling

■ Design

SITRANS FUE1010 is available in three configurations: • IP65 (NEMA 4X) wall mount enclosure constructed of fiber-

glass reinforced polyester with stainless steel hardware and polyester keypad- Single channel- Dual channel / dual path

• IP40 (NEMA 1) Portable impact resistant enclosure con-structed of mineral reinforced copolymer polypropylene- Dual channel / dual path

■ Function

• Flow transmitter has an integral 33 button keypad and large (128 x 240 pixel) graphic display visible up to 12 m (40 ft) away

• 4-wire 1000 Ω platinum RTD’s for supply and return tempera-ture measurements are precision matched to within0.01 °C (0.02 °F)

• Temperature is factory calibrated with built-in field calibrator.• Built-in energy/BTU mode• Detection of aeration and cavitation caused by worn or dam-

aged impellers, misaligned shafts, etc.• Reverse flow and empty pipe detection• Chiller efficiency analysis: accepts an independent analog in-

put representing kW usage for calculation of the following functions which can be selected for data logging or output purposes: - Cooling load (kW/ton)- Coefficient of performance (COP)- Energy efficiency ratio (EER)

• Optional current inputs• Digital communication options:

- MODBUS / Metasys N2 (IP65 (NEMA 4X) only)- RS232 Serial digital port (standard)

• ZeroMatic Path automatically sets zero• Bi-directional flow operation• 1 MByte data logger with both site and data logger storage• English, Spanish, German, Italian and French language op-

tions

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUE1010 (Energy)

4/327Siemens FI 01 · 2012

4

■ Technical specifications

Input

Flow range 0 … 12 m/s (0 … 40 ft/s), bi-directional

Flow sensitivity 0.0003 m/s (0.001 ft/s)

Pipe size 6.4 mm … 9.14 m (0.25“ … 360“)

Inputs per channel • Current: 2 x 4 … 20 mA• Voltage: 2 x 0 … 10 V DC• Temperature: 2 x 4 wire 1 kΩ

RTD• Totalizer commands (clear/hold)

Output

Standard outputs • Current: 2 x 4 … 20 mA DC (1 kΩ at 30 V DC)

• Voltage: 2 x 0 … 10 V DC (5 kΩ minimum)

• Status Alarm: 4 x SPDT Relays• Mercury wetted relays• Frequency: 2 x 0 … 5000 Hz• RS232

Optional outputs • Expanded I/Os (4 additional 4 ... 20 mA outputs) with form c relays

• Expanded I/Os with Mercury wetted relays

Accuracy

Accuracy ± 0.5 % ... 1.0 % of flow,for velocities greater than 0.3 m/s (1 ft/s)± 0.0015 ... 0.003 m/s (± 0.005 ... 0.01 ft/s),for velocities less than 0.3 m/s (1 ft/s)

Batch repeatability ± 0.15 % of flow,for velocities greater than 0.3 m/s (1 ft/s)± 0.0005 m/s (± 0.0015 ft/s),for velocities less than 0.3 m/s (1 ft/s)

Rated operation conditions

Degree of protection Wall mount enclosure:IP65 (NEMA 4X)Portable enclosure:IP40 (NEMA 1)

Liquid temperature

• Standard -40 … +120 °C (-40 … +250 °F)

• Optional -40 … +230 °C (-40 … +450 °F)

Sensor temperature

• Standard -40 … +120 °C (-40 … +250 °F)

• Optional -62 … +232 °C (-80 … +450 °F)

Ambient temperature -18 ... +60 °C (0 ... 140 °F)

Design

Dimensions see SITRANS F US Clamp-on "System info and selection guide"

Weight see diagrams

Power supply

Dedicated 90 … 240 V AC, 50 ... 60 Hz, 30 VA or14.0 … 18.5 V DC

Portable enclosure Rechargeable battery

Indication and operation

Data logger memory 1 Mbyte of storage

Display 128 x 240 pixel LCD with back-light

Keypad 33 keypad buttons with tactile feedback

Language options English, Spanish, German, Italian, French

Certificates and approvals

Dedicated enclosures

FM and CSA ratings • TransmitterNI Class I, Div 2S Class II, Div 2• SensorI.S. Class I, II, Div 1

CE EMC Directive 2004/108/ECATEX Directive 94/9/EC

C-TICK

ATEX ratings • Transmitter:Ex II (1) G [Ex ia] IICEx II 3 (1) G Ex nC [ia] IIC T5

• Sensors:Ex II 1 G Ex ia IIC T5

INMETRO ratings • Transmitter:[BR-Ex ia] IICBR-Ex nc [ia] IIC T5

• Sensors:BR-Ex ia IIC T5

Portable enclosures ULULc

CE EMC Directive 2004/108/ECATEX Directive 94/9/EC

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUE1010 (Energy)

4/328 Siemens FI 01 · 2012

4

Standard MLFB for quick delivery on SITRANS FUE1010 (Energy system)

Selection and Ordering data Order No. Order code

SITRANS FUE1010 (Energy) L) 7 M E 3 5 0 7 - 77777 - 0 777 777 777 + 777 + 777

DesignDedicatedIP65 (NEMA 4X) wall mount 0 K 0 2 + K 0 2 + R 0 2PortableIP40 (NEMA 1) Battery powered 2 K 0 1 + K 0 1 + R 0 1

Number of channels/ultrasonic pathsDedicated metersSingle channel 1Portable metersDual channel/Dual path 4

Flowmeter functions and I/O configurations• Portable Standard I/O C

- Energy efficiency COP/EER output- 2x 4-20mA analog input

• Dedicated Standard I/O F- Reflexor Capability- Energy efficiency COP/EER output- 2x 4-20mA analog input

Meter power options90 ... 240 V AC (Dedicated only) ACharger Type A for Europe (CEE7/7) CCharger Type K for U.S. (NEMA 5-15P) GNo charger J

Communication optionsRS232 (standard) 0

RTD temperature sensor pairNo RTDs (Note: Temperature input is required for Energy systems) 01x Pair Std clamp-on RTD (NEMA 4X only)3) 12x Pair Std clamp-on RTD (For Dual Channel NEMA 4X only)3) 21x Pair Std clamp-on RTD (For NEMA 12 Portable)3) 32x Pair Std clamp-on RTD (For Dual Channel NEMA 1 Portable)3) 41x Insertion RTD with Thermowell and Lagging M 1 A2x Insertion RTD with Thermowell and Lagging M 1 B

Sensor for channel 1(includes pipe mounting kit and spacer bar for indicated max. OD listed) See "Sensor selection charts" for specifications.no sensor AA2 universal Trackmount and straps provided up to 75 mm (3") BB3 universal Trackmount and straps provided up to 125 mm (5") CC3 universal Mounting frame and straps provided up to 300 mm (13") DD3 universal Mounting frame and straps provided up to 600 mm (24") EE2 universal Mounting frame and straps provided up to 1200 mm (48")1)4) FC1H (high precision)Mounting frame and straps provided up to 1200 mm (48")4) MC2H (high precision)Mounting frame and straps provided up to 1200 mm (48")4) ND1H (high precision)Mounting frame and straps provided up to 1200 mm (48")4) PD2H (high precision)Mounting frame and straps provided up to 1200 mm (48")4) QDoppler to 12" with strap kit (not for IP65 (NEMA7)) SD1H High temperature range 104 °C/220 °F HP2) Z P 1 P

L) Subject to export regulations AL: N, ECCN: 3A991X.

© Siemens AG 2012

Modification 3/2012

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUE1010 (Energy)

4/329Siemens FI 01 · 2012

4

Sensor for channel 2(includes pipe mounting kit and spacer bar for indicated max. OD listed)See "Sensor selection charts" for specifications.no sensor AA2 universal Trackmount and straps provided up to 75 mm (3") BB3 universal Trackmount and straps provided up to 125 mm (5") CC3 universal Mounting frame and straps provided up to 300 mm (13") DD3 universal Mounting frame and straps provided up to 600 mm (24") EE2 universal Mounting frame and straps provided up to 1200 mm (48")1)4) FC1H (high precision)Mounting frame and straps provided up to 1200 mm (48")4) MC2H (high precision)Mounting frame and straps provided up to 1200 mm (48")4) ND1H (high precision)Mounting frame and straps provided up to 1200 mm (48")4) PD4H (high precision)Mounting frame and straps provided up to 1200 mm (48")4) RDoppler to 12" with strap kit (not for IP65 (NEMA7)) SD1H High temperature range 104 °C/220 °F HP2) Z Q 1 P

Approvals UL/Portable 0FM, CSA, CE, Dedicated 1

1) Supplied spacer bar supports pipes up to 1050 mm (42 inches). For pipes larger than 1050 mm (42 inches) purchase also, spare part 7ME3960-0MS40 (1012BN-4)

2) Supplied spacer bar supports pipes up to 750 mm (30 inches). For pipes larger than 750 mm (30 inches) purchase also, spare part 7ME3960-0MS40 (1012BN-4)

3) Requires two R** cables per one RTD pair4) 600 mm (24") for portable systems only

L) Subject to export regulations AL: N, ECCN: 3A991X.

Standard MLFB product offering represents 4 to 6 weeks delivery timeFor sensor and RTD cables for quick delivery see tables at end of section

Selection and Ordering data Order No. Order code

SITRANS FUE1010 (Energy) L) 7 M E 3 5 0 7 - 7777 0 - 7777 777 777 + 777 + 777

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUE1010 (Energy)

4/330 Siemens FI 01 · 2012

4

Selection and Ordering data Order No. Ord. code

SITRANS FUE1010 (Energy)

• DedicatedIP65 (NEMA 4X) wall mount L) 7ME3500-

• PortableIP40 (NEMA 1) Battery powered L) 7ME3502-

77777 - 0 777 777

Number of channels/ultrasonic pathsDedicated meter

Dedicated meterSingle channel 1Dual channel/Dual path 2

PortablesDual channel/Dual path 4

Flowmeter functions and I/O configura-tions

• Portable Standard I/O- Reflexor capability- Graphic display- 2 x 0 ... 10 V- 2 x 4 ... 20 mA- 2 x pulse output- 4 x status logic- Energy efficiency COP/EER output- 2 x 4 ... 20 mA analog input

C

• Dedicated Standard I/O- Reflexor capability- Graphic display- 2 x 0 ... 10 V- 2 x 4 ... 20 mA- 2 x pulse output- 4 x relay C type- Energy efficiency COP/EER output- 2 x 4 ... 20 mA analog input

F

• Standard I/O with Mercury wetted relays Z J 1 A

• Extended output adder plus standard inputs (4 additional 4 ... 20 mA outputs) and form C relay

Z J 1 B

• Extended output adder plus standard inputs (4 additional 4 ... 20 mA outputs) and Mercu-ry wetted relays

Z J 1 C

Meter power options

90 ... 240 V AC (Dedicated only) A9 ... 36 V DC (Dedicated only) BCharger Type A for Europe (CEE7/7) CCharger Type C for Australia (AS3112) DCharger Type D for U.K. (BS1363) ECharger Type J for Japan (JIS8303) FCharger Type K for U.S. (NEMA 5-15P) GCharger Type L for Switzerland (SEV1011) HNo Charger J

External 4 hours battery with US plug for Por-table

Z K 1 A

External 4 hours battery with European plug for Portable

Z K 1 B

Communication options

RS232 (standard) 0MODBUS (dedicated only) 1

RTD temperature sensor(includes mounting hardware for pipes above 1.5“ outer diameter)

No RTDs (Note: temperature input is required for energy system)

0

1 x pair standard clamp-on RTD (NEMA 4X only)3)

1

2 x pair standard clamp-on RTD (for dual channel NEMA 4X only)3)

2

1 x pair standard clamp-on RTD (NEMA 1 Portable)3)

3

2 x pair standard clamp-on RTD (for dual channel NEMA 1 Portable)3)

4

1 x Insertion style RTD with thermowell and lagging3)

9 M 1 A

2 x Insertion style RTD with thermowell and lagging3)

9 M 1 B

Sensor for channel 1Including pipe mounting tracks for sizes A & B sensors indented for pipe with a OD less than 125 mm (5") and mounting frame/spacer bars for sizes C, D & E sensors. Straps provided are for the indicated maxi-mum OD listed below. Strap kits are available to accommodate larger pipes (refer to spare part list). Refer to "Sensor Selection Charts" for the sensor suitability of pipe size and wall thickness.

No sensor AA2 universal Trackmount and straps

provided up to 75 mm (3”)

B

B3 universal Trackmount and straps provided up to 125 mm (5”)

C

C3 universal Mounting frame and straps provided up to 300 mm (13”)

D

D3 universal Mounting frame and straps provided up to 600 mm (24”)

E

E2 universal Mounting frame and straps provided up to 1200 mm (48”)1)4)

F

For the following A1H to D4H sensors, temperature range is -40 °C ... 65 °C (-41 °F ... 150 °F), nominal 21 °C (70 °F):A2H (high precision) Trackmount and straps

provided up to 75 mm (3”)

H

A3H (high precision) Trackmount and straps provided up to 75 mm (3”)

J

B1H (high precision) Trackmount and straps provided up to 125 mm (5”)

K

1) Supplied spacer bar supports pipes up to 1050 mm (42 inches). For pipes larger than 1050 mm (42 inches) purchase also, spare part 7ME3960-0MS40 (1012BN-4).

2) Supplied spacer bar supports pipes up to 750 mm (30 inches). For pipes larger than 750 mm (30 inches) purchase also, spare part 7ME3960-0MS40 (1012BN-4).

3) Requires two R** cables per one RTD pair4) 600 mm (24") for portable systems only

L) Subject to export regulations AL: N, ECCN: 3A991X.

Selection and Ordering data Order No. Ord. code

SITRANS FUE1010 (Energy)

• DedicatedIP65 (NEMA 4X) wall mount L) 7ME3500-

• PortableIP40 (NEMA 1) Battery powered L) 7ME3502-

77777 - 0 777 777

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUE1010 (Energy)

4/331Siemens FI 01 · 2012

4

Sensor for channel 1 (continued)B2H (high precision) Trackmount and straps

provided up to 125 mm (5")

L

C1H (high precision) Mounting frame and straps provided up to 1200 mm (48”)3)

M

C2H (high precision) Mounting frame and straps provided up to 1200 mm (48”)3)

N

D1H (high precision) Mounting frame and straps provided up to 1200 mm (48”)2)3)

P

D2H (high precision) Trackmount and straps provided up to1200 mm (48”)2)3)

Q

D4H (high precision) Trackmount and straps provided up to1200 mm (48”)2)3)

R

Doppler to 12“ with strap kit S

High temperature sensor size 2 for up to 230 °C (446 °F) (30 to 200 mm diam. (1.18 to 7.67 inch diam.))

Z P 1 A

High temperature sensor size 3 for up to 230 °C (446 °F) (150 to 610 mm diam.(5.90 to 24 inch diam.))

Z P 1 B

High temperature sensor size 4 for up to 230 °C (446 °F) (400 to 1200 mm diam. (15.75 to 47.25 inch diam.))

Z P 1 C

For the following B1H to D4H sensors, tem-perature range is -1 °C up to 104 °C (30 °F up to 220 °F), nominal 65 °C (150 °F):B1H (high temperature range HP) Z P 1 KB2H (high temperature range HP) Z P 1 LC1H (high temperature range HP) Z P 1 MC2H (high temperature range HP) Z P 1 ND1H (high temperature range HP)2) Z P 1 PD2H (high temperature range HP)2) Z P 1 QD4H (high temperature range HP)2) Z P 1 R

Sensor for channel 2(includes pipe mounting kit for indicated max. outer diameter listed) See "Sensor selection charts" for specifications.

no sensor AA2 universal Trackmount and straps

provided up to75 mm (3”)

B

B3 universal Trackmount and straps provided up to125 mm (5”)

C

C3 universal Mounting frame and straps provided up to 300 mm (13”)

D

D3 universal Mounting frame and straps provided up to 600 mm (24”)

E

E2 universal Mounting frame and straps provided up to 1200 mm (48”)1)3)

F

Selection and Ordering data Order No. Ord. code

SITRANS FUE1010 (Energy)

• DedicatedIP65 (NEMA 4X) wall mount L) 7ME3500-

• PortableIP40 (NEMA 1) Battery powered L) 7ME3502-

77777 - 0 777 777

Sensor for channel 2 (continued)

For the following A1H to D4H sensors, tem-perature range is -40 °C to 65 °C (-41 °F to 150 °F), nominal 21 °C (70 °F):A2H (high precision) Trackmount and straps

provided up to 75 mm (3”)H

A3H (high precision) Trackmount and straps provided up to 75 mm (3”)

J

B1H (high precision) Trackmount and straps provided up to 125 mm (5”)

K

B2H (high precision) Trackmount and straps provided up to 125 mm (5")

L

C1H (high precision) Mounting frame and straps provided up to 1200 mm (48”)3)

M

C2H (high precision) Mounting frame and straps provided up to 1200 mm (48”)3)

N

D1H (high precision) Mounting frame and straps provided up to 1200 mm (48”)2)3)

P

D2H (high precision) Mounting frame and straps provided up to 1200 mm (48”)2)3)

Q

D4H (high precision) Mounting frame and straps provided up to 1200 mm (48”)2)3)

R

Doppler to 12“ with strap kit S

High temperature sensor size 2 for up to 230 °C (446 °F) (30 to 200 mm diam. (1.18 to 7.67 inch diam.))

Z Q 1 A

High temperature sensor size 3 for up to 230 °C (446 °F) (150 to 610 mm diam.(5.90 to 24 inch diam.))

Z Q 1 B

High temperature sensor size 4 for up to 230 °C (446 °F) (400 to 1200 mm diam. (15.75 to 47.25 inch diam.))

Z Q 1 C

For the following B1H to D4H sensors, tem-perature range is -1 °C up to 104 °C (30 °F up to 220 °F), nominal 65 °C (150 °F):B1H (high temperature range HP) Z Q 1 KB2H (high temperature range HP) Z Q 1 LC1H (high temperature range HP) Z Q 1 MC2H (high temperature range HP) Z Q 1 ND1H (high temperature range HP)2) Z Q 1 PD2H (high temperature range HP)2) Z Q 1 QD4H (high temperature range HP)2) Z Q 1 R

ApprovalsFM/CSA/CE Dedicated 1UL/ULc/CE Portable 01) Supplied spacer bar supports pipes up to 1050 mm (42 inches). For pipes

larger than 1050 mm (42 inches) purchase also, spare part 7ME3960-0MS40 (1012BN-4).

2) Supplied spacer bar supports pipes up to 750 mm (30 inches). For pipes larger than 750 mm (30 inches) purchase also, spare part 7ME3960-0MS40 (1012BN-4).

3) 600 mm (24") for portable systems only

L) Subject to export regulations AL: N, ECCN: 3A991X.

Selection and Ordering data Order No. Ord. code

SITRANS FUE1010 (Energy)

• DedicatedIP65 (NEMA 4X) wall mount L) 7ME3500-

• PortableIP40 (NEMA 1) Battery powered L) 7ME3502-

77777 - 0 777 777

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUE1010 (Energy)

4/332 Siemens FI 01 · 2012

4

■ MLFB example

Application example

A dedicated clamp-on energy meter is required for two separate return lines. Both will use clamp-on RTDs for the supply and re-turn lines. AC power is available and data access will be via MODBUS communication.

Pipe 1 is a DN150 (6") schedule 40 carbon steel linePipe 2 is a DN 300 (12") ductile iron line

MLFB Order No.: 7ME3500-2DA10-2NE0-ZK03 + K05 + R03 + R05 + R02 + R03

D) Subject to export regulations AL: N, ECCN: EAR99H.

Selection and Ordering data Order code

Further designsPlease add "-Z" to Order No. and specify Order code(s).

Cable assembly for sensors (add for # of channels)See "Sensor cable selection chart" K..

Cable assembly for RTDs (add for # of RTDs)See "RTD cable selection chart" R..

Cable termination kit (for one cable pair) dedicated only

• Termination for standard, plenum and armored sensor cable

T01

• Termination for submersible sensor cable T11

• RTD cable termination kit for standard RTD T21• RTD cable termination kit for submersible RTD T31• Insert RTD cable termination kit T41

Wet flow transfer calibration (priced on request)

6 point calibration 2/water (Price per channel)

• 2SS40 pipe D01• 3CS40 pipe D02• 4CS40 pipe D03• 4SS40 pipe D04

• 6CS40 pipe D05• 6SS40 pipe D06• 6CS120 pipe D07• 8CS40 pipe D08

• 8SS40 pipe D09• 8CS120 pipe D10• 10CS Standard pipe D11• 10CS40 pipe D12

• 10SS40 pipe D13• 12CS Standard pipe D14• 12CS40 pipe D15• 14CS30 pipe D16

• 14CS40 pipe D17• 16CS Standard pipe D18• 16CS40 pipe D19• 18CS Standard pipe D20

• 20CS20 pipe D21• 20CS30 pipe D22• 24CS Standard pipe D23• 24CS20 pipe D24

• 24CS30 pipe D25• 30CS Standard pipe D26• 36CS Standard pipe D27• Other pipe, other liquid, additional points, witness Y28

Tag name plate

• Stainless steel tag with 3.2 mm (0.13 inch) character size (68 characters max.)

Y19

Selection and Ordering data Order No. Ord. code

SITRANS FUE1010 meter family

7 M E 3 5 0 7 - 7777 0 - 7777 777

IP65 (NEMA 4X) enclosure 0

Dual channel 2

Dedicated Type 1 I/O option D

90 ... 230 V AC power option A

MODBUS option 1

2 pairs of clamp-on RTDs 2

Sensor code for 6" pipe N

Sensor code for 12" pipe E

No approval required 0

30 m (100 ft) sensor cable for channel 1

K 0 3

61 m (200 ft) sensor cable for channel 1

K 0 5

30 m (100 ft) cable for RTD 1 R 0 3

61 m (200 ft) cable for RTD 2 R 0 5

15 m (50 ft) cable for RTD 3 R 0 2

30 m (100 ft) cable for RTD 4 R 0 3

Selection and Ordering data Order code

Operating Instructions for SITRANS FUE1010

English NEMA 4X Wall mount D) A5E03086491

German NEMA 4X Wall mount D) A5E03086492

English IP40 NEMA 1 Battery powered A5E02951524

German IP40 NEMA 1 Battery powered A5E02951536

This device is shipped with a Quick Start Guide and a CD containingfurther SITRANS F literature.All literature is also available for free at: http://www.siemens.com/flowdocumentation

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUE1010 (Energy)

4/333Siemens FI 01 · 2012

4

Universal sensor selection chart

High precision sensor selection chart

Sensor cable selection chart

RTD cable selection chart

Based on pipe size (pipes other than steel)

Pipe size OrderCode

Outer diameter range (mm)

Outer diameter range (inches)

min. max. min. max.

A2 B 12,7 50,8 0.5 2

B3 C 19 127 0.75 5

C3 D 51 305 2 12

D3 E 203 610 8 24

E2 F 254 6096 10 249

Based on pipe wall thickness (steel pipes only)

Pipe Wall OrderCode

Pipe Wall [mm] Pipe Wall [inches]

min. max. min. max.

A1H G 0,64 1,02 0.025 0.04

A2H H 1,02 1,52 0.04 0.06

A3H J 1,52 2,03 0.06 0.08

B1H K 2,03 3,05 0.08 0.12

B2H L 3,05 4,06 0.12 0.16

C1H M 4,06 5,84 0.16 0.23

C2H N 5,84 8,13 0.23 0.32

D1H P 8,13 11,18 0.32 0.44

D2H Q 11,18 15,75 0.44 0.62

D4H R 15,75 31,75 0.62 1.25

Sensor cable codes for length and type options

Cable length m (ft)

Standard (PVCjacket)

-40…+80 °C(-40…+176 °F)

Submersible1)

(polyethylene jacket)-40…+80 °C(-40…+176 °F)

1) Submersible and armored sensor cable is not available for portable ver-sions.

2) Standard MLFB for quick delivery

Plenum Rated(teflon jacket)

-40…+200 °C(-40…+392 °F)

Armored1)

-40…+80 °C(-40…+176 °F)

Order code

6 (20) K012) K11 K21 K31

15 (50) K02 K122) K22 K322)

30 (100) K032) K132) K23 K33

46 (150) K042) K14 K24 K34

61 (200) K05 K15 K25 K35

91 (300) K062) K16 K26 K36

RTD cable codes for length and type

Cable length m (ft) Standard (teflon wrapped)-40 ... +200 °C(-40 … +392 °F)

Insert1)

-40 ... +200 °C(-40 … +392 °F)

1) Submersible RTD cable is not available for portable versions.2) Standard MLFB for quick delivery

Order code

6 (20) R012) R21

15 (50) R022) R22

30 (100) R032) R23

46 (150) R04 R24

61 (200) R05 R25

91 (300) R06 R26

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-onSITRANS FUE1010 (Energy)HVAC Check metering kit

4/334 Siemens FI 01 · 2012

4

■ Overview

The SITRANS FUE1010 dual channel clamp-on check metering kit is an all inclusive HVAC chilled water kit developed especially for verifying the accuracy and performance of any brand or type of flowmeter. The meter's portability makes it capable of verifying the performance of meters based on any existing flow measure-ment principle: electromagnetic, vortex, insertion turbine, or ul-trasonic. Perfect for areas where no metering exists. Ideal for balancing building performance. It accurately computes flow over an extremely wide range and measures practically all con-ductive or non-conductive clean or moderately aerated liquids or liquids with suspended solids. Dual channel models can mea-sure two separate applications at the same time.

■ Benefits• Performance check or verification of any type or brand of flow-

meter• Measures energy rate and total consumption with highest

accuracy available• Accurately measures at both low flow rates and low differential

temperatures• Field use is facilitated by meter portability charge for 4 hours

of normal operation • 1 MByte datalogger capability downloadable to PC via in-

cluded RS232 cable• Performs fast, easy and cost-efficient on-site measurement of

any convoluted pipe from 25.4 mm to 9.14 m (1.0" to 360")• Delivered as an all inclusive kit with all the equipment needed

to conduct performance and verification tests (cables, multiple sensors, flow transmitter etc.)

• Comes in a sturdy rolling case with a telescope handle that holds all the equipment needed to conduct performance and verification tests.

■ Application

The SITRANS FUE1010 Check Metering Kit is a highly accurate clamp-on non-intrusive ultrasonic flow display transmitter or rev-enue grade thermal energy sub-metering and energy efficiency distribution monitoring, with a real time coefficient of perfor-mance (COP) for HVAC systems. This kit is ideal for applications which include:• Chilled water sub-metering• Condenser water• Potable water• Ammonia and glycol• River and lake water• Lake source cooling

■ Design• IP40 (NEMA 1) Impact resistant enclosure, constructed of flame

retardant ABS with polycarbonate display and polyester keypad• Dual channel/dual path

■ Function• Integral 33 button keypad and large (128 x 240 pixel) graphic

display visible up to 12 m (40 ft) away• 4-wire 1000 Ω platinum RTD's for supply and return tempera-

ture measurements are precision matched to within 0.01 °C (0.02 °F)

• Chiller efficiency analysis: accepts an independent analoginput representing kW usage for calculation of the following functions which can be selected for data logging or output purposes:- Cooling load (kW/ton)- Coefficient of performance (COP)- Energy efficiency ratio (EER)

• Temperature is factory calibrated with built-in field calibrator• Built-in energy/BTU mode• Detection of aeration and cavitation caused by worn or

damaged impellers, misaligned shafts, etc.• Current, voltage, frequency and RS232 outputs

(see specification section for details)• Optional current, voltage and temperature inputs

(see specification section for details)• ZeroMatic Path automatically sets zero• Bi-directional flow operation• 1 MByte data logger with both site and data logger storage• English, Spanish, German, Italian and French language op-

tions

© Siemens AG 2011

Flow MeasurementSITRANS F US Clamp-on

SITRANS FUE1010 (Energy)Check metering kit

4/335Siemens FI 01 · 2012

4

■ Technical specifications

D) Subject to export regulations AL: N, ECCN: EAR99H.

Pipe sizes 25.4 mm to 9.14 m (1 ... 360“)Accuracy ± 0.5 % ... ± 2.0 % of flow rateFlow range 12 m/s (40 ft/s) bidirectionalMedia temperature -40 ... +104 °C (-40 ... 220 °F)Enclosure ratings IP40 (NEMA 1) impact resistant

See page 4/327 for complete technical specifications

Cerificates and approvalsPortable enclosuresUnclassified locations UL

ULc

Classified locations

CE EMC Directive 2004/108/ECATEX Directive 94/9/EC

Selection and Ordering data Order No.

Energy check metering kit D) CQO:FUEHVACKIT

Content of delivery1 Dual channel portable submersible flow

transmitter1 pair Universal sensors C31 pair Doppler sensors1 pair High precision sensors C21 pair High precision sensors D12 pairs RTDs2 pairs Mounting Ezclamp (4 mounting Ezclamp

chains)1 Battery charger2 pairs 6 m (20 ft) sensor cables1 RS232 cable4 RTD cable 6 m (20 ft)4 Mountings for RTDs1 Spacer bar (Portable)2 F connector to BNC1 Flow case1 Flow meter manual1 Laminated card set1 Certificate of intrinsic calibration