EVALUATION OF THE CALIBRATION SYSTEM OF AN EBERT …
Transcript of EVALUATION OF THE CALIBRATION SYSTEM OF AN EBERT …
EVALUATION OF THE CALIBRATION SYSTEM
OF AN EBERT-FASTIE SPECTROGRAPH
BENJAMIN R. ARCHER, B.A.
A THESIS
IN
PHYSICS
Submitted to the Graduate Faculty of Texas Tech University in Partial Fulfillment of the Requirements for
the Degree of.
MASTER OF SCIENCE
Approved
Director
Accepted^
December, 1970
AtHoiS^
EOS rz 1970 NoJ57 top. 2
ACKNOWLEDGMENTS
I am deeply indebted to Professor Glen A. Mann for his direction of
this thesis and to William H. Almond for his assistance in the prepara
tion of the instrument.
"V-
11
TABLE OF CONTENTS
LIST OF TABLES iv
LIST OF FIGURES v
I. INTRODUCTION 1
II. INSTRUT'IENTATION 3
Spectrograph 3
Calibration System 6
III. THE 1-0 AND 2-0 INFRARED BANDS OF HCl l6 •
IV. EXPERTtffiNTAL OBSERVATIONS AND RESULTS 29
V. CONCLUSION 35
Sources of Error 35
Recommendations 38
LIST OF REFERENCES ^0
111
LIST OF TABLES
Table I Ca lcu la ted and Observed R Branch Frequencies in cm of
HCl35 Bands 28
Table I I Typical Data for Edser -But le r Band Spacing 32
Table I I I Edse r -But le r Band Reproduc ib i l i t y 33
Table IV Average Uncer ta in ty Per C a l i b r a t i o n Fringe 3^
i v
LIST OF FIGURES
Figure 1 Ebert- Fastie Infrared Spectograph 5
Figure 2 The Calibration System 7
Figure 3 Multiple Beam Interference in a Fabry-Perot Interferometer 12
Figure k Graphs of the Airy Function for Different Values of Reflectance 13
Figure 5 Energy Level Diagram Explaining the Fine Structure of a Rotation-Vibration Band 26
Figure 6 Infrared Absorption Spectrum of the 2-0 Band of HCl • . 27
CHAPTER I
INTRODUCTION
The successful operation of a high resolution infrared spectrograph
requires some suitable means of detemiining the spectral positions with a
high degree of precision. One such method, employed by the infrared spec
trometry group in the Physics Department at Texas Tech University, in
volves the use of a Fabry-Perot interferometer. "Calibration fringes"
(Edser-Butler bands or channeled spectra) that result serve to impress a
"wave number scale" on the infrared spectrum to be measured. Ideally,
these calibration fringes should be equally spaced in cm units.
In practice, however, the spacing is not found to be precisely con
stant. This leads to an uncertainty in the wavelength determination of
spectral lines. In order to determine the accuracy of future infrared
measurements made in this laboratory, the spacing repeatability of the
calibration fringes with particular interferometer plate separations has
been found. Also, the spacings of the fringes for various plate separa
tions and the associated uncertainties have been determined.
The 1-0 and 2-0 infrared absorption bands of Hydrogen Chloride gas
were chosen as the standards with which to analyze the calibration frin
ges. HCl was chosen for several reasons. First, both bands occur in the
region of maximum operating efficiency of the spectrograph. At grating
angles less than about 27 or greater than about 33 , adequate calibra
tion fringes cannot be found without compromising the infrared resolu
tion of the instrument. Both bands of HCl lie within this region of
1
maximum calibration signal intensity.
Secondly, there are at present two modes of operation in the system
for obtaining infrared data. In the first mode, a transmission filter is
inserted in front of the infrared detector to eliminate all radiation be
low 2.5 microns. The second mode occurs when the filter is removed. HCl
provides a means of checking fringe consistency for both infrared modes
since the 1-0 band is above and the 2-0 band is below 2.5 microns.
Finally, the infrared spectrum of HCl is easily interpreted and very
precise measurements have been made on both bands under consideration.
CHAPTER II
INSTRUMENTATION
Spectrograph
The data was obtained with an Ebert-Fastie vacuum spectrograph. This
2 instrument is explained in detail in Almond's thesis and is represented
in Figure 1. The infrared source is a carbon arc. A lead sulfide solid
state detector, cooled with liquid nitrogen, is used as the infrared de
tector. The grating contains 300 lines per millimeter and is blazed for
maximum diffraction of incident radiation of wavelength three microns.
The beam from the carbon arc, continuous infrared radiation, first
passes through the 2 meter path length absorption cell containing HCl.
The infrared spectrum is then scanned by rotation of the grating and is
recorded by one pen of a double-pen recorder. The other pen simultaneous
ly records the calibration fringes.
The theoretical resolution of the spectrograph at 3 microns is
-1 3
0.05 cm for a single pass of the grating. The resolution is a measure
of the minimum resolvable difference of wavelength at any given wave
length. In order that two lines be resolved, there must be a dip in the
intensity distribution. A generally accepted criterion for the resolu
tion of two lines is the Rayleigh criterion. According to this criterion,
two equal lines are considered resolved if the intensity at the saddle
2
poin t i s not g r e a t e r than S/TT t imes the i n t e n s i t y at the two maxima.
The reso lv ing power of a g r a t i ng i s defined by:
R.P. = A/5A ( I l - l )
where 6A is the smallest wavelength difference that produces resolved
images. If the Rayleigh criterion for the resolution is used, the images
must be separated by the angle
66 = A/NdCos0 (II-2)
where N is the number of rulings on the grating, d is the ruling separa
tion, and 6 is the angle of incident radiation.
The grating formula giving the relation between the wavelength and
angle of diffraction is
mA = dSine (II-3)
where the integer m is the order of diffraction. Differentiation of
(II-3) gives
66 = m6A/dCos0 (II-U)
Combining equations (II-2) and (II-U) the resolving power of the grating
spectrograph is obtained:
R.P. = A/6A = mN (II-5)
_A_ For t h i s instrument ^x ~ 62,U00. Thus, i f the s l i t width i s small,
the resolution at 3 microns, as s ta ted should be 0.05 cm for a single
pass of the grat ing. The actual resolu t ion , determined under convenient
operating condit ions, has been found to be approximately 0.06 cm . These
conditions were that a reasonable s ignal- to-noise r a t io was simultaneously
obtained for both the infrared and cal ibra t ion s ignals .
Source
Vacuum Chamber
Figure 1. Ebert-Fastie Infrared Spectrograph
Calibration System
For ca l ibra t ion of the spectrograph, two beams of radiation are sent
through the grat ing optics simultaneously. The beam from the carbon a r c ,
as mentioned above, i s detected by the infrared detector . The second
beam consists of radiat ion from a 300 watt zirconium arc that i s focused
on the entrance s l i t of the monochromator. This resu l t s in nearly mono
chromatic l igh t emerging frora'the exit s l i t . The emergent beam traverses
the Fabry-Perot interferometer af ter which i t passes through the Wadsworth
prism system and f a l l s on an RCA 1P21 photomultiplier tube. The Edser-
Butler bands tha t resu l t when the interferometer i s swept with white l igh t
that has been diffracted by the grating are detected by the phototube and
recorded by the second pen of the double pen recorder. These cal ibrat ion
fringes are recorded simultaneously with the infrared spectrum.
The ca l ibra t ion system must contain the means of overcoming any e r
rors introduced by non-uniform rotat ion of the grat ing, as well as supply
a uniform set of indicator marks which provide a means of determining the
frequency of the spect ra l l i n e s . Both of these functions are performed
effect ively by the use of Edser-Butler bands obtained with the instrumen
ta t ion shown in Figure 2. Since both the infrared aad cal ibrat ion beams
t r ave l through the same grating optics simultaneously, any s l ight non-uni
form rotat ion of the grat ing wi l l affect both beams in exactly the same
manner. For example, a minute hes i ta t ion in the rotat ion of the grating
wi l l r esu l t in a s l igh t ly expanded separation of two adjacent infrared
absorption l i n e s . However, t h i s separation is exactly compensated for
u Q)
•H H ft
•H -P I O
o PH
0)
:3
-p o ^ <u
PH 1
^ <u +3 a) S o ^ (U
>>«H ^ S cd f^
^ (U +J C!
o
o u
J=i o o o
o
o •H -P •H CO
o PH
(U ,Q ;3
-P O
o
(U
m >>
CO
O •H +J
•H
O
<u
EH
OJ
•H
P4
a w >>
CD
•H
8
by a similar increase in spacing in the calibration fringe system. Thus,
the calibration fringes do provide a convenient means of calibration as
well as of overcoming anom.alies in the rotation of the grating.
The reliability and reproducibility of the fringes depend primarily
on the characteristics of the Fabry-Perot interferometer. The interfero
meter used consisted of two partially reflecting circular plates, 1 1/8
inch in diameter. The surfaces of the plates were flat to 1/6 of the
wavelength of green light (56OO angstroms).
The plate spacing was mechanically varied and could be fixed at
1mm, 2mm, 3mm, or ilmm by inserting the proper magnesium spacer in order
to allow no further plate separation than that desired. Once this de
sirable separation was obtained, it was necessary to make the plates
accurately parallel. To accomplish this, a sodium arc was used to il
luminate the interferometer. This produced a series of concentric rings
which became readily visible to the eye when the plates were nearly paral
lel. When the plates become precisely parallel, and if they are exactly
plane, the rings remain the same size as the eye is moved to any point
of the field of view.
Once the plates had been aligned, the sodiimi arc was removed and
the monochromatic light from the exit slit was positioned to strike the
Interferometer at the center of the interference rings. The physical
size of the central ring depended upon the plate separation - the smaller
the separation, the larger the size of the ring.
Before calibration fringes can be obtained, it is necessary to sep
arate out the multiple diffraction orders that are obtained from the
grating. Each different frequency that passes through the interferome
ter will, if it is dispersed, produce its own series of equidistant maxi
ma. However, if the orders are not separated, the resulting output will
consist of the average intensity of all signals combined and will there
fore be highly periodic in amplitude and spacing. It is thus necessary
to incorporate the Wadsworth prism system in order to separate the orders.
This system consists of.-a 60° prism and a plane mirror introduced
into the beam path after it has traversed the interferometer. The beam
enters the prism at the angle of minimum deviation and is dispersed ac
cording to wavelength, thus elimj.nating the overlapping of the various
orders. The plane mirror is positioned to reflect the dispersed orders
back on a line parallel to the beam incident on the interferometer.- Each
of the separated images contains the central spot of the interference
pattern produced by the interferometer. This enables the phototube to
be positioned so as to observe the central ring of only one of the dis
persed orders. This arrangement is shown in Figure 2.
The central ring, as well as all others in the concentric series,
is due to the multiple beam interference utilized in the Fabry-Perot in
terferometer. The primary beam is partially reflected and partially
transmitted at the first surface. The transmitted part is then reflected
back and forth, as shown in Figure 3, between the two surfaces. If r is
the coefficient of reflection and t the coefficient of transmission,
2 2 2 then the amplitudes of successive transmitted rays are E t , E t r , o o
2 k E t r , where E is the amplitude of the incident beam. The geometric o o
path difference between any two successive transmitted rays is 2dCos9,
10
where d is the separation between reflecting surfaces, and 6 is the angle
either ray makes with the surface normal. The corresponding phase dif
ference 6 between two successive rays is then given by
6 = 2kdCos0 = dCosG (II-6) A
where k = 27T/A. Taking t h i s phase d i f f e r e n c e i n t o account by i n c l u d i n g
t h e f a c t o r e and adding t h e ampl i tudes of a l l of t h e t r a n s m i t t e d r ays
y i e l d s :
E t ^ ^ ^2 ^ ^2 2 16 ^ ^2 U 2i6 ^ o ,^^ „x
E^ = E t + E t r e + E t r e + . . . = •;--rT ( I I - 7 ) • T o o o _ 2 10
1-r e The i n t e n s i t y of t h e t r a n s m i t t e d l i g h t i s given by
1-r e
I 12 where I = E is the intensity of the primary beam.
o ' o' Since a change of phase can occur on reflection, r is, in general a
k complex number and can be expressed as
r = Irle^'^^/^ (II-9)
where — i s t h e phase change fo r one r e f l e c t i o n . Fu r the rmore , i f t h e r e
f l e c t a n c e of one s u r f a c e i s denoted by R and t h e t r a n s m i t t a n c e by T, t h e n
R = | r | ^ , T = | t | ^ (11-10)
Now ( I I - 3 ) may be w r i t t e n as
T =: I _ T ^ ,-^ ( I l - l l ) T o , x2 ^R o- 2/A.
(1-R) 1 + o 2^^ ^9^ (1 -R)^
11
where
A = 6 + 6^ (11-12)
i s t he t o t a l phase d i f fe rence between two success ive t r ansmi t t ed beams.
The i n t e n s i t y , t h e r e f o r e , va r i e s with A according t o t h e funct ion:
ITR ' p A ~ o~A~ (11-13) 1 + - ^ Sin^(f) 1 4- F Sin^(l-)
(1-R)^ •' ^ ^
which is known as the "Airy function.' Here
F = — ^ (II-IM (1-R)^
is a measure of the sharpness of the interference fringes. This does
not refer to the "calibration fringes" recorded as wavelength standards
but only to the concentric ring series produced by a non-varying source .
The Airy function is shown in Figure k for various values of the
reflectance R. These curves show the intensity distribution of the frin
ges in multiple beam interference. If R is small, the interference frin
ges are broad and indistinct, whereas if R is close to one, the fringes
are very sharp.
The maximum and minimum values of I are
T^ I r ^ = 1 T(max) o ^_J^)2
= I —^-TT (11-15) T(min) - o ^^^^)2
«
Calibration fringes are produced by what is referred to as the
"scanning method." The rotation of the grating causes the center spot
12
E
Figure 3. Multiple Beam Interference in a Fabry-Perot Interferometer
13
o •H -P O
K
o CO
;=J
-p
c
• H
O
O
•H
a
;:<
•H <
O
m
ft ccJ
(D
w •H
H /q.Tsua:^ui
Ik
to alternate in intensity. This variation at the ring center, recorded
photometrically, is due to the fact that the monochromatic light, inci
dent on the interferometer at an angle 6, changes by a small amount AA
from A to A±AA as the grating rotates. In order to examine this varia
tion fully it is necessary to first develop the theory of the interfer
ence fringes produced by a single wavelength A. When a ray undergoes
two.reflections, it has a ph^se difference of
6 = -- n (2dCos0) (II-I6) A a
from a ray that travels through unreflected, where n is the index of a
refraction of air and is equal to one, 9 is the angle of incidence of the
incident ray, d is the plate separation, and m is the order number. For
constructive interference that results in a bright fringe occurring, 6
must be an integral multiple of 2IT. For an intensity maximum, therefore
mA = 2dSin6 (11-17)
Considering the order number m as constant, d i f ferent ia t ion of
(11-17) y i e lds :
+mAA = T2dSin0A0 (II-I8)
This equation shows that the result of increasing A is a shrinking of
the ring system and that a decrease in A causes an expansion of the rings
When white light is used, A is being continually varied. At those
wavelengths A = ^^^^^ for which m is an integer, maxima will occur. At m
such wavelengths, the intensity of light from the central ring (given in
equation (II-I5)) will be appreciable. This large signal will be regis-
15
tered by the photomultiplier tube and recorded as a calibration fringe
maximum on the chart recorder. As the wavelength continues to increase,
the interference fringe system will collapse, causing a gradual decrease
in intensity of the beam section observed by the phototube until I / . x
given in (11-15) is reached. The next maxima will occur when the wave
length change results in the order nmnber being changed by one. The
calibration fringes are thus'"a graphical record of the intensity pattern
recorded at the center of the ring system. The distribution is essen-
tially a plot of the Airy function (l + F Sin -)
The spacing of the fringes can now be determined by finding the
difference in cm" corresponding to two adjacent orders of interference.
From the form of the Airy function it is evident that this is given by
setting
A , - A = 27r (11-19) n+1 n
then
A = 6 + 6 (11-20) n n r
where
6 = UTT/A dCose (11-21) n n
but V = 1/A gives the cm"" value of A. After substitution, it is n n '
found that the fringe spacing is given by:
V .. - V n+1 n 2dCose
(11-22)
CHAPTER III
THE 1-0 AND 2-0 INFRARED BANDS OF HCl
In order to attribute justifiable precision to the calibrated wave
number scale it is necessary to have accurate theoretical information on
the individual HCl lines used as standards. Thus, the theoretical infra
red spectrum must be calculated.
• _
For a diatomic molecule-'there are two modes of motion that are of
primary importance in determining its infrared spectrum. First, the
molecule can rotate as a whole about an axis passing through the center
of gravity and perpendicular to the internuclear axis, and second the
atoms can vibrate relative to each other along the internuclear axis.
To investigate what type of spectrum would be expected for such a
vibrating or rotating system, on the basis of quantum theory, one begins
with the so-called rigid rotator model. Consider the two atoms to be a
distance r apart, separated by a massless rigid rod.
In order to determine the possible energy states of such a rigid
rotator the Schrodinger equation must be solved.
2
1 ' (Sine -^) . - V — ^ - E^ = 0 (III-I) Sine 88 ' dB' „. 2. ..2 ,2 r
Sin 6 d4» h
where I is the moment of inertia of the molecule. The solution of this
7 equation, carried out in Pauling and Wilson, is found to be
^ = N P,(">(Cose)ei"* (III-2)
^r r J
where ip represents the rigid rotator eigenfunctions and N^ is a nor
malization constant. The energy eigenvalues are
16
17
y2 E = - ^ I • J(J+1) (III-3)
87T
where the rotational quantum number J can take the integral values
0,1,2,. . .
The absorption of a light quantum of the proper frequency produces
a transition from a lower to a higher rotational level. The wave number
of the absorbed radiation is
V = E'/hc - E"/hc (Ill-ii)
where E' and E" are rotational energies in the upper and lower states,
respectively. Define F(j) as the rotational energy such that
F(J) = E/hc = —~—J(J+1) = BJ(J+1) (III-5) 87T cl
where
B = ^ 87T^Cl
Thus,
V = F(J') - F(J") = BJ'(J'+1) - BJ"(J"+1) (III-6)
In order to calculate the frequencies that are actually absorbed,
it is necessary to know the selection rules for the quantum number J.
These rules are found by evaluating the m.atrix elements of the dipole
moment. If M is the constant dipole moment of the rotator, its com-o
ponents are
M = M SineCosd) X o ^
M = M SineSin* (III-7) y o ^ M^ = M^Cose .
18
The matrix elements to be evaluated are of the form
^ J'M'J"M" K. 1
*/'"' M^,/""" d. (III-8)
where i = x,y,z.
o
Evaluation of these elements show that they vanish except when
AJ = J'-J" = ±1.
Hence
V = F(J"+1) - F(J") = B(j"+l)(j"+2) - B(J"+1)
= 2B(J+1) J = J" = 0,1,2,. . . (III-9)
Thus, the spectrum of the simple r ig id ro ta tor consists of a ser ies
of equidistant l i n e s , the f i r s t at 2B (J=0). The separation of succes
sive l i ne s i s also 2B.
The simplest possible assumption about the form of the vibrat ions
in a diatomic molecule i s that each atom executes simple harmonic motion.
If the masses of the atoms are replaced by y = TiLm/m+m^, the reduced
mass, and the displacement of each atom from i t s equilibrium pos i t ion ,
TQ and rgp, by the d i s to r t ion of the distance between the two pa r t i c l e s
from the equilibrium bond length, then the vibration of the two pa r t i c l e
system may be t r ea t ed as a single vibrat ing p a r t i c l e . The Schrodinger
equation for the vibrat ions of a diatomic molecule with reduced mass y
i s
7^f^^I^\-^\ (iii-io) OlT y dX
19
the solut ion functions can be writ ten as
o —l/2ax
K " V H^(v^) V = 0 ,1 ,2 , . . . ( I l l - l l )
where v i s the v ibra t ional quantum number, N^ i s a normalizing constant ,
H^ i s the Hermite polynomial of degree v, and
a = 27T/yk/h
The allowed energies are given by the expression^
E^ = (v+l) hv^AT
o^ E^ = (v+l") ho) V = 0,1,2, . . .(III-12)
Since ^ = "2:^ /k/y i s the c lass ica l frequency of vibrat ion. In
order to express the resul t of (III-12) in cm" , the energy values are
transformed into "term values" (G(v)). Thus,
G(v) = E(v)/h = a)(v+|) ( III-13)
where cu i s measured in cm
The o s c i l l a t o r can be set in vibration by absorption of l ight of
the proper frequency. This frequency i s given by
V = G(v') - G(v") (III-IU)
In order to find the par t i cu la r t r ans i t ions that may occur, the
matrix elements of the dipole moment must be evaluated, t h i s time using
the vibrat ional wave fun ct ions. I t is found that the selection rule for
the v ibra t ional quantum number of the harmonic osc i l l a to r i s Av = +1.
20
Av = v' - v" = +1 for absorption
so
- V = G(v + 1) - G(v) =00 (III-15)
Thus, all of the allowed transitions give rise to the same frequency.
The frequency of radiated light is equal to the frequency of the oscilla
tor, regardless of what the Value of the initial state may be.
If the theoretical results obtained so far, are compared with the ob
served spectra of HCl, the following interpretation results. The spectrum
in the far infrared, since it consists of nearly equidistant lines, is a
rotation spectrum. The spectrum in the near infrared, since it consists
essentially of a single very intense line, is a vibration spectrum. On
this assumption the occurrence of bands with nearly two and three times
the frequency would be connected with deviations from the harmonic oscil
lator.
Thus, the main characteristics of the infrared spectra are explained
by the models of the rigid rotator and the harmonic oscillator. However,
to obtain the finer details of the spectra, the models must be refined.
The rotator must be treated as nonrigid and the oscillator as anharmonic.
According to equation (III-IO), an harmonic oscillator is character
ized by a parabolic potential curve. In the actual molecule, however,
when the atoms are a great distance apart, the attractive force is zero,
and the potential energy is constant. This is not consistent with the
parabolic potential function in which the potential energy increases in
definitely with increasing distance from the equilibrium position.
21
As a f i r s t approximation to the actual potent ia l energy function
of the molecule a cubic term i s added to the quadratic function of the
harmonic o s c i l l a t o r . Thus,
U = f ( r - r ^ ) ^ - g ( r - r^ )^ ( I I I - I6 )
Here, the coefficient g is very much smaller than f and r is the equi
librium separation.
If (III-I6) is substituted into the Schrodinger equation as the po
tential function, the energy values of the anharmonic oscillator are
given by
E^ = hcw^Cv-t^) - hca)^x^(v+|-)^ + hca)^y^(v+|-) + . . .(III-I7)
and correspondingly the term values are
% = ^e^"""^^ " %\^^^^^ '' '^e^e^'^'i^^ + . . . ( I I I - I8 )
1 / ^ where the constant w x <<a) and oj y <<co x . Here oj = 7 and w x e e e e e e e 27TC y e e
and 0) y are anharmonicity constants. e' e "
Equation (III-I8) shows that the energy levels of the anharmonic
oscillator are not equidistant like those of the harmonic oscillator,
but that their separation decreases slowly with increasing v. However,
in order to consider the spectrum more fully, the selection rules must
be considered. They are
Av = 1,2,3
22
The t r ans i t i ons with Av=l are expected to give r i se to the most intense
t r a n s i t i o n s . I t can be seen that the t rans i t ions with Av=2,3,. . .
have approximately but not exact ly, two, three . . . times the frequency
of the t r ans i t i on Av=l. In spi te of the s l ight deviation from the c las
s i ca l r e s u l t , the 1-0, 2-0, 3-0, . . . are referred to as the fundamental,
f i r s t overtone, second overtone, and so on. The formula that gives the
posi t ions of the v ibra t ional - levels above the lowest vibrat ional level i s
V = G(v) - G(o) = w V - 0) X V + OJ X v- -(111-19) o o o o o
Thus far the models of the rigid rotator and the harmonic (or an
harmonic) oscillator have been used independently of each other. How
ever, it is obvious that the molecule cannot be a strictly rigid rotator
where it has the capability of vibrating. Therefore, a new model, in
which the atoms are joined by a massless spring instead of a rigid rod
will be considered. This model is called the nonrigid rotator.
In this system, an increase in angular speed will increase the cen
trifugal force which in turn will increase the internuclear distance and
consequently the moment of inertia. To a good approximation, the rota
tional terms of the nonrigid rotator are given by
F(J} = BJ(J+1) - DJ^(J+1)^ (III-20)
where D is the centrifugal distortion constant and D<<B. The selection
rule derived previously AJ = ±1 is valid whether or not the rotator is
rigid. Using (III-20) the spacing of lines is given by
23
V = F(J+1) - F(J) = 2B(J+1) - 4 D ( J + 1 ) ^ ( I I I - 2 1 )
The l i n e s are no longer e q u i d i s t a n t , as for a r i g i d r o t a t o r , but t h e i r
s epa ra t ion decreases s l i g h t l y with inc reas ing J .
To t h i s po in t t h e v i b r a t i o n and r o t a t i o n of t he molecule have been
regarded as be ing t o t a l l y independent. However, observed data show
s t rong i n d i c a t i o n s t h a t r o t a t i o n and v ib ra t ion take place simultaneously.
Hence, i t i s necessary t o consider t he v i b r a t i n g r o t a t o r .
During t h e v i b r a t i o n the i n t e r n u c l e a r dis tance and consequently the
moment of i n e r t i a and the rota . t ional constant B are changing. The r o t a
t i o n a l constant B i n t he v i b r a t i o n a l s t a t e v i s given by V
B = B - a (v-i^) + . . . ( I I I -22 ) V e e 2
where a i s a constant which i s small compared to B . Analogously, D , Q C V
representing the influence of centrifugal forces, is given by
D = D + 3 (v4-) + . . . (III-23) V e e ^ 2'
IfR 3 Here 3 <<D = , which r e f e r s t o t he completely v ib r a t i on l e s s s t a t e e e ^
We The r o t a t i o n a l terms in a given v i b r a t i o n a l l e v e l are
F ( J ) = B^J(J+1) - D^J^(J+1)^+ . . . (III-2I1)
The term values of a v i b r a t i n g r o t a t o r , obtained by t ak ing in to account
t he i n t e r a c t i o n of v i b r a t i o n and r o t a t i o n , are
2U
T = G(v) + F (J)
= 0) , ( v 4 ) - a)^xjv-f |)2 + . e e
+ B^J(J+1) - D^J^(J+1)^+. . .
( I I I - 2 5 )
Since the eigenfunct ions of the v i b r a t i n g r o t a t o r are e s s e n t i a l l y
the products of r o t a t o r and o s c i l l a t o r func t ions , the same se lec t ion
r u l e s hold as for these systems i n d i v i d u a l l y . Thus, v can change by any
i n t e g r a l amount although Av=+1 gives by far the most in tense t r a n s i t i o n .
The value of J can only change by un i ty .
I f a t r a n s i t i o n from v' t o v" i s considered, the loca t ions of the
r e s u l t i n g l i n e s are given by
V = v^ + B , J ' ( J ' + 1 ) - D , J ' 2 ( J ' + 1 ) 2 _ B „J"(J"+1)
+ D „J"^(J"+1)2 ( I I I -26 )
where v = G ( V ' ) - G(v") , the frequency of the pure v ib r a t i ona l t r a n s i
t i o n . I f t he AJ=+1 t r a n s i t i o n i s defined as the R branch and the AJ=-1
as t he P branch the r e s u l t i n g l i n e s are
>R = ^^ + (2B ,-kl) , ) + (3B ,-B „-12D , ) J + (B ,-B „-13D ,+D „ ) j ' n O V V V V V V V V V
- ( 6 D , + 2 D „ )J -^ - (D ,+D „)J V V V V
( I I I -27 )
vp = v^ + (B ,-B „)J + (B ,-B ,,-D ,-D „)J^ + (-2D ,+2D „)J^ r O ^ V V V V V V V V
- (D ,+D „ ) J V V
(III-28)
25
A single formula can be used to represent the two branches - namely
V = v^ + (B^,+B^.,)m + (B^,-B^„-D^,+D^.,)m2
-2(D^,+D^„)m2 - (D^,-D^„)m^ (III-29)
where m is an integral running number that takes the values 1,2, . . .
for the R branch (that is, m=J+l) and the values -1,-2, . . . for the P
branch (m=-J). Thus, there is a single series of lines for which a line
is missing at m=0. The missing line at v=v is called the band origin.
o ^
F i g u r e 5 i s an energy l e v e l diagram t h a t e x p l a i n s t h e f i n e s t r u c t u r e
of a. r o t a t i o n - v i b r a t i o n band. In g e n e r a l , t h e s e p a r a t i o n of two v i b r a
t i o n a l l e v e l s i s much g r e a t e r t h a n t h e spac ing of t h e r o t a t i o n a l l e v e l s
shown i n t h e f i g u r e . The schemat ic spec t rogram below t h e f i g u r e r e p r e
s e n t s t h e e x p e c t e d v i b r a t i o n - r o t a t i o n s p e c t r a . F igure 6 shows t h e a c t u a l
i n f r a r e d s p e c t r a of HCl fo r t h e 2-0 band. The s h o r t e r peaks r e p r e s e n t 37 35 .
t h e i s o t o p e HCl . The e v a l u a t i o n , however , was done only wi th HCl
The c a l c u l a t i o n of t h e s e l i n e p o s i t i o n s as w e l l as a c c u r a t e e x p e r i
men ta l measurements , have been c a r r i e d out and p u b l i s h e d by D. H. Rank.
His r e s u l t s a re d i s p l a y e d i n Table 1 fo r v a l u e s of J up t o and i n c l u d i n g
J=10 .
26
v=2
v=0
J'
5 -
k -
1 -0 -
J"
5
1 0
r 'T
CM O H
PH
OJ
PH
oo P
. ^
PH
LTN
P-.
h
..A..—-—a.
I 1 ji • »-jfc-y
V.
Figure 5 Energy Level Diagram Explaining the Fine Structure of a Rotation-Vibration Band
o
o
pq
o I
CM
(U
- P
O
- P o (U
ft CO
o •H -p P^ U O CO
(U
u
M
27
>
5 rz:> n _ z>
13-___ i_
L_
U
•H
-i-
28
TABLE I
CALCULATED AND OBSERVED R BRANCH FREQUENCIES IN CM " OF HCl" ^ BANDS
1-0 Band
J (Ca lc )
R(J) (Obs)
Calc - Obs
X 10"^
0
1
2
3
h
5
6
7
8
9
10
2-0 Band
0
1
2
3
k
5
6
7
8
9
10
2906.2I179
2925.8977
29i|I | .9ili6
2963.2866
2981.0015
2998.0)476
30lU.U83^
3030.0876
30U5.0592
3059.3171
3072.8509
5687 .651^
5706.0951
5723.3033
5739.26^3
5753.9665
5767.3989
5779.5510
5790.U121
5799.9725
5806.2225
5815.1527
.2521
.8950
.915^
.2865
.0013
.0I438
.lilli^
.0862
.0569
.3179
.8U90
.61+98
.0932
.302U
.2651+
.9673
.3980
.51^98
.U1U7
.9688
.2205
. l i^8l
-U.2
+2 .7
- 0 . 8
+ 0 . 1
+0.2
+ 3 .8
+2.0
+l.k
+ 2 . 3
- 0 . 8
+1.9
+1.6
+1.9
+0.9
- 1 . 1
- 0 . 3
+0.9
+1.2
- 2 . 6
+3.7
+2.0
+I1.6
CHAPTER IV
EXPERIMENTAL OBSERVATIONS AND RESULTS
In analyzing the data, great care was taken to justify the degree
of accuracy that is presented. At least four independent spectra were
made with each spacer. The results for each spacer were determined us
ing both the 1-0 and 2-0 bands of HCl as standards. A comprehensive
error analysis, as described in Errors of Observation and Their Treat
ment , was then made. A brief indication of this analysis follows.
If X , Xp, . . • J x^ represent the averages of fringe spacings be
tween adjacent HCl absorption peaks in a single spectrum, the most pro
bable value is the arithmetic mean.
X = Cx^ + x^ + . . • x^)
N (lV-1)
The s tandard e r r o r of the mean i s — , where a i s the s tandard devia-
t i o n for the se t of measurements. The bes t es t imate of a i s found t o be
- (x^-x) + (x^-x) + . . . (x^-x) a2 = _ 1 ^ 1 (lV-2)
N-1
Hence, the measured spacing is reported as " x + ".
Table 2 shows typical data obtained for the 1-0 and 2-0 bands.
The data presented was obtained with the Umm spacer. The first column
indicates the absorption peak number of the HCl. The second gives the
separation of adjacent peaks in fringe numbers. The separation of
these peaks in cm""'' is listed in the next column. In the final column,
the fringe spacings in cm" are tabulated. The most probable value of
29
30
fringe spacing and the standard error of the mean for each band is also
reported.
Table 3 shows the reproducibility of fringe separation. The follow
ing procedure was used to determine this reproducibility; (l) the spac
er \mder investigation, was inserted and the interferometer plates made
parallel, (2) the spectrum was run, (3) the spacer was removed, (U) the
interferometer plates were closed down and then reopened to accomodate
the spacer again, (5) the plates were realigned and the next spectrum
was made. It was found that after replacing the spacer the interfero
meter plates did not assume the original parallel position. In some
cases, when using the larger plate separations, the interference pattern
as viewed with the sodium arc had entirely disappeared. The adjustment
necessary to reacquire the pattern was small, however, as was the inher
ent error in the reproducibility. It was also found that the interfero
meter plates could not be adjusted to precise parallelness. The probable
reasons for and the effect of this will be discussed in the following
chapter.
The ratio of the average spacings of the fringes in the 1-0 band
with respect to the 2-0 band are listed in column three of Table 3.
These ratios are approximately constant. A value in the neighborhood of
.58 will encompass all the ratios if the inherent uncertainty in each is
considered. This indicates that the separation in cm of a pair of
fringes used with the infrared region above 2.5 microns will represent
about .58 of the separation in cm" of the same pair of fringes used
with the region below 2.5 microns. The 2.5 micron division is due, as
31
previously explained, to the fact that the 1-0 band was examined using
the 2.5 micron infrared filter, while the 2-0 band was examined without
it.
Table k lists the average uncertainty per fringe for the various
spacings. These were compiled by computing the fringe spacing uncerta.in-
ty in each spectrum and then finding the average of these uncertainties
for*each spacer.
32
TABLE I I
TYPICAL DATA FOR EDSER-BUTLER BAND SPACING
{hzam Spacer )
1-0 Band
Pe'ak
No.
R(0)
R ( l )
R(2)
R(3)
R(l+)
R(5)
R(6)
R(7)
R(8)
HCl
F r i r
S e p a r a t i o n s in
ige Numbers
97 .52
9^^.66
9 1 . 3 1
88.22
Qk.86
8 1 . lU
77 .98
7^^.32
HCl S e p a r a t i o n s in
cim*~-'-
19.6ii98
19.0169
18.3720
17.71I+9
17.0U61
16.3658
15.67I42
l i [ . 9 7 l 6
F r inge Spacing i n
cm"-'-
.2015
.2009
.2012
.2008
.2009
.2017
.2010
.20lii
Most P robab le Value: .2012 + .0002
2-0
R(o;
R( l -
R(2;
R(3 .
Bik,
R(5;
R(6;
R(7;
R(8;
Band
) 53.15
) 1+9.75
) i+6.12
) 1+2.73
) 38 .73
) • 35.02
I 31 .38
) 27.1+2
18.1+1+37
17.2082
15.9610
IU.7022
I3.I+32I+
12 .1521
10 .8611
9.560I+
t P robab l e Value :
.31+71
.31+59
.31+61
.31+1+1
.31+68
.31+70
.3961
.3U87
.31+65 ± .0006
33
TABLE III
EDSER-BUTLER BAND REPRODUCIBILITY
Spacer
1 mm
2 mm
3 mm
k mm
Band . *
1-0
.582 ± .051+
.367 ± .007
.2588 ± .0008
.2006 ± .0001+
2-0
1.050 ± .Ol+l
.650 ± .006
.1+1+91 ± .0006
.31+62 ± .0002
Ratio ^ ^
552 ± .069
56I+ ± .016
5762 ± .0026
579^ ± .0015
TABLE IV
AVERAGE UNCERTAINTY PER CALIBRATION FRINGE
Band Spacer
1-0 2-0
31+
1mm .0016 .0031
2mm .0006 .0011
3mm .0005 .0009
i+mm .0003 .0008
CHAPTER V
CONCLUSION
Sources of Error
The data displayed in Tables III and IV indicates that the best
possible accuracy that may be obtained with the present instrumentation
results when the 1+ mm spacer is used. The accuracy drops off so consi
derably for separation of 1 mm and 2 mm that these spacers do not ap
pear to be suitable for infrared calibration. This increase in uncer
tainty occurs because the large widely-spaced calibration fringes that
result when the 1 and 2 mm spacers are used, are not suited to the pre
cise evaluation that high resolution analysis requires. The exact cen
ter line, for example, of the large, rounded peaks produced is difficult
to find. Whereas, the centers of the sharper, pointed peaks obtained
with k mm separation are readily ascertained. Furthermore, a small par
titioning error in a fringe produced with small separation leads to much
greater uncertainty than the same error produced with large separations.
This is a result of the fact that there are many more calibration fringes
between the same HCl peaks, when larger separations are used.
There are other factors that contribute to the uncertainties also.
There are variations between Rank's calculated and observed values for
the HCl absorption lines. Table I shows that the average difference is
about .002 cm" . This indicates that the uncertainty is not entirely due
to fringe irregularities but is partly a result of instrumental inaccura
cies that are to be expected when observing the positions of the infrared
of HCl. 35
36
Another source of error mentioned in the preceding chapter is that
the plates could not be adjusted to exact parallelness. This problem
appears to be due to the surface defects of the Fabry-Perot plates,
which according to Chabbal,^^ can significantly change the properties of
the interferometer. The ordinary optical flat of 1/1+ wavelength flat
ness is not good enough for precise work. A flatness of the order of
1/20 to 1/100 wavelength is strongly recommended. As stated previously,
the plates used in obtaining the data presented had a listed flatness of
1/6 wavelength - clearly a minimal value for interferometer plates.
A final, and probably the smallest source of error is due to the
fact that the temperature of the interferometer plates and spacer was
not regulated. During periods in which the data was taken, both were ex
posed to the temperature of the room. The magnesium spacers used in the
-6
interferometer had a coefficient of linear expansion of about 26 x lO" 1/C
The maximum possible change in temperature that was likely to occur dur
ing a data run was about 5°. This amounts to a maximum change in length
of the spacer of about .005 mm. On the basis of many observations of the
behavior of the interference pattern with small displacements of the
plates, it seems reasonable to conclude that distortion due to tempera
ture expansion is small compared to the other errors mentioned.
The ratio of fringe spacing (from Table III) , as discussed in the
prior section, was found to be about O.58. This can be attributed to a
change in the angular dispersion of the grating for different orders of
diffraction. The theoretical angular separation between two spectral
37
lines differing in wavelength by AX is
^^ " dCose
where n is the order number and d is the grating spacing. Obviously,
therefore, the dispersion of the 2-0 band should be about twice that of
the 1-0 band. The fact that the ratio is not exactly 0.50 indicates
that the dispersion in the second order falls off from the expected
value.
38
Recommendations
Despite the above listed sources of error in the present calibra
tion system, its accuracy is nearly equivalent to or surpasses many of
the present techniques employed for wavelength calibration. It is, how
ever, very far from the accuracy claimed by Professor T. H. Edwards at
13 Michigan State University. Using fused quartz plates aluminized to
ahout 80^ reflectivity and a fused quartz spacer, fringe spacing was de
termined to ±0.000001 cm""''.
It is interesting to note some of the characteristics of this in
terferometer. Fused quartz is used because it has one of the lowest
-6 n coefficients of linear expansion, .1+2 x 10 l/C . The importance of
the high reflectivity was shown in Figure 1+, the graph of the Airy func
tion. As can be seen, the fringes become sharper for large values of R.
It is doubtful that the cost involved in this kind of accuracy is
necessary in the Ebert-Fastie system discussed in this thesis. However,
if the resolution is improved, which can be accomplished by the multiple
II+
pass system described in Almond's thesis, an improvement in the cali
bration system would seem desirable. This could be accomplished by con
struction of a Fabry-Perot interferometer of high reflectance using high
quality plates and fused-quartz spacers as described in Born and Wolf.
For general infrared calibration, a spacer of about 3-5 mm would be de
sirable. This would allow good accuracy as well as permit excellent frin
ges to be readily obtained — a combination that is difficult to achieve
with 1+ mm separation.
39
Since t h i s study was completed, high, opt ical quali ty plates have
been obtained and the construction of a Fabry-Perot e talon, with the
spacing recommended in t h i s work, wi l l soon be undertaken.
LIST OF REFERENCES
1. D. H. Rank, B. S. Rao, and T. A. Wiggins, "Molecular Constants of HCl-^^ J. Mol. Spectry., IJ (1965) , pp. 122-130.
2. W. H. Almond, "The Construction and Calibration of an Ebert-Fastie Vacuum Spectrograph," (Texas Technological College, 1966, Unpublished M.S. Thesis).
3. Ibid. , p. 22.
h. G. R. Fowles , I n t r o d u c t i o n t o Modern Op t i c s ( H o l t , R i n e h a r t , and Wins ton , I n c . , New York, I 9 6 8 ) , pp . 8 6 - 9 0 .
5 . M. Born and E. Wolf, P r i n c i p l e s of Op t i c s (The Macmillan Company, New York, I96I+), p . 326.
6 . K. W. M e i s s n e r , " I n t e r f e r e n c e S p e c t r o s c o p y , " J . Opt. Soc. Am., 13 (19I+I) , pp . I+O5-I+O9.
7 . L. P a u l i n g and E. B. Wi l son , I n t r o d u c t i o n t o Quantum Mechanics (McGraw-Hill Book Company, I n c . , New York, 1 9 3 5 ) , pp . 263-27I+.
8 . G. H e r z b e r g , S p e c t r a of Diatomic Molecules (D. Van Nost rand Company I n c . , T o r o n t o , 1 9 5 0 ) , p . 7 2 .
9 . I b i d . , p . 7 5 .
10. Rank, Rao, and Wiggins, op. cit.
11. J. Topping, Errors of Observation and Their Treatment (Chapman and Hall, London, I961).
12. R. Chabbal, "Finesse Limite D'un Fabray-Perot Forme de Lames Impar-fartes," J. Phys. Rad. , 1£ (1958), pp. 295-300.
13. D. H. Rank, K. N. Rao, C. J. Humphreys •> Wavelength Standards in the Infrared (Academic Press, New York, I966) , p. 16O.
II+. Almond, op. cit. , pp. 22-21+.
15. Born and Wolf, op. cit. , p. 330.