12/3/2018 1 © MFMER | slide-1 Development of a Standardized Visual Inspection Aid for Plasma © MFMER | slide-2 Manish J. Gandhi, M.D. Associate Professor of Laboratory Medicine and Pathology Division of Transfusion Medicine James R. Stubbs, M.D. Associate Professor of Laboratory Medicine and Pathology Chair, Division of Transfusion Medicine Erik J. Levorson Clinical Laboratory Technologist Component Laboratory Jessie A. Swanson, M.L.S Clinical Laboratory Technologist Component Laboratory Kassandra L. Poffenberger Education Specialist I Component Laboratory Katelyn K. Heaser Clinical Laboratory Technologist Component Laboratory Lorraine C. Steele* Clinical Laboratory Technologist Component Laboratory * Denotes Speaker Transfusion Medicine Department of Laboratory Medicine and Pathology Mayo Clinic, Rochester, Minnesota

Transcript of Development of a Standardized Visual Inspection Aid for Plasma · 12/3/2018  · • Used plasma...

12/3/2018

1

© MFMER | slide-1

Development of a Standardized Visual Inspection Aid for Plasma

© MFMER | slide-2

Manish J. Gandhi, M.D.Associate Professor of Laboratory Medicine and PathologyDivision of Transfusion Medicine

James R. Stubbs, M.D.Associate Professor of Laboratory Medicine and PathologyChair, Division of Transfusion Medicine

Erik J. LevorsonClinical Laboratory TechnologistComponent Laboratory

Jessie A. Swanson, M.L.SClinical Laboratory TechnologistComponent Laboratory

Kassandra L. PoffenbergerEducation Specialist I Component Laboratory

Katelyn K. HeaserClinical Laboratory TechnologistComponent Laboratory

Lorraine C. Steele*Clinical Laboratory TechnologistComponent Laboratory

* Denotes Speaker

Transfusion Medicine

Department of Laboratory Medicine and Pathology

Mayo Clinic, Rochester, Minnesota

12/3/2018

2

© MFMER | slide-3

Disclosure

• None

© MFMER | slide-4

Objectives

• Explain Whole Blood Processing

• Define Plasma Inspection Criteria

• Differentiate Plasma Quality Acceptability

• Describe Standardization of Plasma Quality Interpretation

• Evaluate Success of Plasma Inspection Aid

12/3/2018

3

© MFMER | slide-5

Whole Blood Processing - 1

• Whole blood (WB) is delivered by Donor Services

• Product code labels are applied and units are hung to filter

© MFMER | slide-6

Whole Blood Processing - 2

• Filtered WB units are centrifuged.

• Then the plasma is manually expressed into a separate bag.

12/3/2018

4

© MFMER | slide-7

Whole Blood Processing - 3

• Red Blood Cell (RBC) and plasma bags are separated

• Visual inspection is performed

• RBC has segments made, is documented in the computer, and refrigerated

• The plasma is weighed, documented and frozen

© MFMER | slide-8

Inspection Criteria for Plasma

• Color

• Lipemia

• Clots

• Contaminants

• Bag issues

12/3/2018

5

© MFMER | slide-9

• Is there a standard for RBCs in plasma?

• What volume of RBCs is acceptable in plasma?

• What does plasma look like at different RBC concentrations?

© MFMER | slide-10

Is there a standard for RBCs in plasma?

• No regulatory standard

• The American National Red Cross “Visual Inspection Reference Guide”

• Photos of normal plasma• Photos of normal plasma with red cells• No color or RBC concentration cut-off

12/3/2018

6

© MFMER | slide-11

What volume of RBCs is acceptable in plasma?

• Pathogen Reduction Technology we use for platelets has a specification of <4 x106 RBCs/mL.

• This RBC concentration is an instrument limitation, not a patient care recommendation.

• The guideline was used to help set a cut-off for acceptable vs unacceptable plasma.

© MFMER | slide-12

Plasma: Variation in RBC Concentration0.0 mL – 0.50 mL

0.0mL 0.15mL0.10mL0.05mL 0.30mL0.20mL 0.50mL0.40mL

12/3/2018

7

© MFMER | slide-13

Determination of Visual Cut-off

• Color differences were observed by multiple techs to establish a visual cut-off

• Visual cut off was determined to have an RBC concentration of approx. 3.3 x106

RBCs/mL

© MFMER | slide-14

Standardized Visual Inspection Aid

• No regulatory standards for red blood cells (RBCs) allowed in plasma.

• No visual inspection guide available that shows unacceptable limits for plasma.

• A standardized plasma inspection aid could be used to provide consistency across multiple work units.

12/3/2018

8

© MFMER | slide-15

Creation of Plasma Inspection Aid

• Used plasma units that were created when determining the variation in RBC concentration and its effect on the plasma color

• Cut off for acceptable vs unacceptable was determined visually as well as mathematically

• Professional photos were taken of sample units to provide an easy to use aid

© MFMER | slide-16

12/3/2018

9

© MFMER | slide-17

Implementation and Outcomes

• Standard visual cut-off easier to determine plasma outcome.

• Discard rate: implementation 2%, improved to 1%

• Easy to use, convenient

• Revised cut-off concentration is 2.2 x106

RBCs/mL

© MFMER | slide-18

Survey Results

• 84% use the aid

• 87% aid is helpful

• 92% commented on how aid improved visual inspection

• Consistency, Standardization and Scientific Support

0

2

4

6

8

10

12

14

16

18

20

Strongly Agree Agree Disagree

Utilize Aid Aid is Helpful

12/3/2018

10

© MFMER | slide-19

© MFMER | slide-20

Review of Objectives

• Explain Whole Blood Processing

• Define Plasma Inspection Criteria

• Differentiate Plasma Quality Acceptability

• Describe Standardization of Plasma Quality Interpretation

• Evaluate Success of Plasma Inspection Aid