12/3/2018
1
© MFMER | slide-1
Development of a Standardized Visual Inspection Aid for Plasma
© MFMER | slide-2
Manish J. Gandhi, M.D.Associate Professor of Laboratory Medicine and PathologyDivision of Transfusion Medicine
James R. Stubbs, M.D.Associate Professor of Laboratory Medicine and PathologyChair, Division of Transfusion Medicine
Erik J. LevorsonClinical Laboratory TechnologistComponent Laboratory
Jessie A. Swanson, M.L.SClinical Laboratory TechnologistComponent Laboratory
Kassandra L. PoffenbergerEducation Specialist I Component Laboratory
Katelyn K. HeaserClinical Laboratory TechnologistComponent Laboratory
Lorraine C. Steele*Clinical Laboratory TechnologistComponent Laboratory
* Denotes Speaker
Transfusion Medicine
Department of Laboratory Medicine and Pathology
Mayo Clinic, Rochester, Minnesota
12/3/2018
2
© MFMER | slide-3
Disclosure
• None
© MFMER | slide-4
Objectives
• Explain Whole Blood Processing
• Define Plasma Inspection Criteria
• Differentiate Plasma Quality Acceptability
• Describe Standardization of Plasma Quality Interpretation
• Evaluate Success of Plasma Inspection Aid
12/3/2018
3
© MFMER | slide-5
Whole Blood Processing - 1
• Whole blood (WB) is delivered by Donor Services
• Product code labels are applied and units are hung to filter
© MFMER | slide-6
Whole Blood Processing - 2
• Filtered WB units are centrifuged.
• Then the plasma is manually expressed into a separate bag.
12/3/2018
4
© MFMER | slide-7
Whole Blood Processing - 3
• Red Blood Cell (RBC) and plasma bags are separated
• Visual inspection is performed
• RBC has segments made, is documented in the computer, and refrigerated
• The plasma is weighed, documented and frozen
© MFMER | slide-8
Inspection Criteria for Plasma
• Color
• Lipemia
• Clots
• Contaminants
• Bag issues
12/3/2018
5
© MFMER | slide-9
• Is there a standard for RBCs in plasma?
• What volume of RBCs is acceptable in plasma?
• What does plasma look like at different RBC concentrations?
© MFMER | slide-10
Is there a standard for RBCs in plasma?
• No regulatory standard
• The American National Red Cross “Visual Inspection Reference Guide”
• Photos of normal plasma• Photos of normal plasma with red cells• No color or RBC concentration cut-off
12/3/2018
6
© MFMER | slide-11
What volume of RBCs is acceptable in plasma?
• Pathogen Reduction Technology we use for platelets has a specification of <4 x106 RBCs/mL.
• This RBC concentration is an instrument limitation, not a patient care recommendation.
• The guideline was used to help set a cut-off for acceptable vs unacceptable plasma.
© MFMER | slide-12
Plasma: Variation in RBC Concentration0.0 mL – 0.50 mL
0.0mL 0.15mL0.10mL0.05mL 0.30mL0.20mL 0.50mL0.40mL
12/3/2018
7
© MFMER | slide-13
Determination of Visual Cut-off
• Color differences were observed by multiple techs to establish a visual cut-off
• Visual cut off was determined to have an RBC concentration of approx. 3.3 x106
RBCs/mL
© MFMER | slide-14
Standardized Visual Inspection Aid
• No regulatory standards for red blood cells (RBCs) allowed in plasma.
• No visual inspection guide available that shows unacceptable limits for plasma.
• A standardized plasma inspection aid could be used to provide consistency across multiple work units.
12/3/2018
8
© MFMER | slide-15
Creation of Plasma Inspection Aid
• Used plasma units that were created when determining the variation in RBC concentration and its effect on the plasma color
• Cut off for acceptable vs unacceptable was determined visually as well as mathematically
• Professional photos were taken of sample units to provide an easy to use aid
© MFMER | slide-16
12/3/2018
9
© MFMER | slide-17
Implementation and Outcomes
• Standard visual cut-off easier to determine plasma outcome.
• Discard rate: implementation 2%, improved to 1%
• Easy to use, convenient
• Revised cut-off concentration is 2.2 x106
RBCs/mL
© MFMER | slide-18
Survey Results
• 84% use the aid
• 87% aid is helpful
• 92% commented on how aid improved visual inspection
• Consistency, Standardization and Scientific Support
0
2
4
6
8
10
12
14
16
18
20
Strongly Agree Agree Disagree
Utilize Aid Aid is Helpful
12/3/2018
10
© MFMER | slide-19
© MFMER | slide-20
Review of Objectives
• Explain Whole Blood Processing
• Define Plasma Inspection Criteria
• Differentiate Plasma Quality Acceptability
• Describe Standardization of Plasma Quality Interpretation
• Evaluate Success of Plasma Inspection Aid