ANSI-ASQ National Accreditation Board/ACLASS · 2015. 1. 9. · 3458A (Amp) and Fluke 8508A (Volts)...
Transcript of ANSI-ASQ National Accreditation Board/ACLASS · 2015. 1. 9. · 3458A (Amp) and Fluke 8508A (Volts)...
Version 009 Issued: 8/01/2014 Page 1 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
& CERTIFIED COMPLIANCE TO ANSI/NCSL Z540-1-1994
Departamento de Metrología, Avianca S.A. Calle 26 No. 103-08 Terminal Simón Bolívar Bogotá, Colombia
Ing. Marco A. Brugés www.avianca.com
[email protected] Phone: 5714578700 x 139, 263
CALIBRATION
Valid to: August 1, 2016 Certificate Number: AC-1327
I. Electromagnetic - DC/Low Frequency
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
DC Voltage - Source5
Up to 330 mV
330 mV to 3.3 V
(3.3 to 33) V
(33 to 330) V
330 V to 1 kV
20 µV/V + 1.8 µV
11 µV/V + 9.6 µV
12 µV/V + 0.11 mV
18 µV/V + 1.1 mV
18 µV/V + 3.9 mV
Fluke 5520A SC600
OEM*,CEM,IEC,
EURAMET,EA or
Internal Procedures
DC Voltage - Measure5
Up to 200 mV
200 mV to 2 V
(2 to 20) V
(20 to 200) V
200 V to 1 kV
Up to 6 kV
2.7 µV/V + 0.65 µV
2.7 µV/V + 1.8 µV
2.7 µV/V + 18 µV
4.0 µV/V + 0.18 mV
4.0 µV/V + 1.3 mV
1.6 mV/V
Fluke 8508A Option 01
with
VRS-100-5-1G-10kV
DC Current - Source5
with Current Clamp
Up to 330 µA
330 µA to 3.3 mA
(3.3 to 33) mA
(33 to 330) mA
330 mA to 1.1 A
(1.1 to 3) A
(3 to 11) A
(11 to 20.5) A
(20.5 to 1 025) A
0.15 mA/A + 20 nA
0.10 mA/A + 60 nA
0.10 mA/A + 0.50 µA
0.10 mA/A + 8 µA
0.20 mA/A + 50 µA
0.38 mA/A + 0.20 mA
0.50 mA/A + 0.80 mA
1 mA/A + 2 mA
3.4 mA/A
Fluke 5520A SC600
Fluke 5520A
with 5500A Coil
ANSI-ASQ National Accreditation Board/ACLASS
Version 009 Issued: 8/01/2014 Page 2 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
DC Current - Measure5
Up to 100 nA
100 nA to 1 µA
1 µA to 10 µA
10 µA to 100 µA
Up to 200 µA
200 µA to 2 mA
(2 to 20) mA
(20 to 200) mA
200 mA to 2 A
(2 to 20) A
(20 to 50) A
30 µA/A + 40 pA
20 µA/A + 50 pA
20 µA/A + 0.20 nA
20 µA/A + 0.90 nA
6.6 µA/A + 5 nA
6.6 µA/A + 20 nA
8 µA/A + 90 nA
34 µA/A + 3 µA
0.17 mA/A + 40 µA
0.38 mA/A + 2 mA
18 mA/A
Agilent 3488A
Fluke 8508A Opt 001
with Current Shunt
1651 Keithley
OEM*,CEM,IEC,
EURAMET,EA or
Internal Procedures
Resistance - Source5
Up to 11 Ω
(11 to 33) Ω
(33 to 110) Ω
(110 to 330) Ω
330 Ω to 1.1 kΩ
(1.1 to 3.3) kΩ
(3.3 to 11) kΩ
(11 to 33) kΩ
(33 to 110) kΩ
(110 to 330) kΩ
330 kΩ to 1.1 MΩ
(1.1 to 3.3) MΩ
(3.3 to 11) MΩ
(11 to 33) MΩ
(33 to 110) MΩ
(110 to 330) MΩ
330 MΩ to 1.1 GΩ
50 µΩ
100 µΩ
500 µΩ
(0.001 to 0.01) Ω
(0.01 to 1) Ω
(1 to 10) Ω
10 Ω to 10 kΩ
40 µΩ/Ω + 1.1 mΩ
30 µΩ/Ω + 1.6 mΩ
28 µΩ/Ω + 1.6 mΩ
28 µΩ/Ω + 2.7 mΩ
28 µΩ/Ω + 6 mΩ
28 µΩ/Ω + 33 mΩ
28 µΩ/Ω + 84 mΩ
28 µΩ/Ω + 0.33 Ω
28 µΩ/Ω + 0.90 Ω
32 µΩ/Ω + 3 Ω
32 µΩ/Ω + 6.9 Ω
60 µΩ/Ω + 47 Ω
0.13 mΩ/Ω + 0.21 kΩ
0.25 mΩ/Ω + 3.2 kΩ
0.50 mΩ/Ω + 9.6 kΩ
3 mΩ/Ω + 0.13 MΩ
15 mΩ/Ω + 1.4 MΩ
3 µΩ
3 µΩ
20 µΩ
0.10 mΩ/Ω + 0.2 mΩ
70 µΩ/Ω + 0.2 mΩ
70 µΩ/Ω
45 µΩ/Ω
Fluke 5520A SC600
Avtron
Shunts C2047
High Accuracy
Resistance Substituter
HARS-X-11-0.001
OEM*,CEM,
EURAMET,EA or
Internal Procedures
Version 009 Issued: 8/01/2014 Page 3 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Resistance - Source5
(cont.)
(10 to 100) kΩ
100 kΩ to 1 MΩ
(1 to 10) MΩ
(10 to 100) MΩ
100 MΩ to 1 GΩ
(1 to 10) GΩ
(10 to 301) GΩ
1 TΩ
10 TΩ
35 µΩ/Ω
0.13 mΩ/Ω
50 µΩ/Ω
225 µΩ/Ω
0.75 mΩ/Ω
0.75 mΩ/Ω
1 mΩ/Ω
24 GΩ
0.58 TΩ
High Accuracy
Resistance Substituter
HRRS-B-5-100K
HARS-X-11-0.001
High Resistance
VICTOREEN
VRS-100-5-1G-10kV
High-Resistance
High Voltage Decade
Substituter
OEM*,CEM,
EURAMET,EA or
Internal Procedures
Resistance - Measure5
Up to 2 Ω
(2 to 20) Ω
(2 to 200) Ω
200 Ω to 2 kΩ
(2 to 20) kΩ
(20 to 200) kΩ
200 kΩ to 2 MΩ
(2 to 20) MΩ
(20 to 200) MΩ
200 MΩ to 2 GΩ
(2 to 20) GΩ
(20 to 190) GΩ
10 µΩ/Ω + 19 µΩ
7 µΩ/Ω + 63 µΩ
7 µΩ/Ω + 5 mΩ
7 µΩ/Ω + 3 mΩ
7 µΩ/Ω + 31 mΩ
7 µΩ/Ω + 0.52 Ω
7 µΩ/Ω + 8.8 Ω
9 µΩ/Ω + 0.21 kΩ
30 µΩ/Ω + 14 kΩ
0.50 mΩ/Ω + 1.1 MΩ
0.50 mΩ/Ω + 13 MΩ
8.95 mΩ/Ω
Fluke 8508A Opt 01
Fluke 5520A and Agilent
3458A (Amp) and
Fluke 8508A (Volts)
Defibrillator5 -
Energy Source (10 to 360) J 1.4 J Fluke
Impulse 7000 DP
OEM*, CEM
EURAMET,EA
ABNT or Internal
Procedures
Inductance - Measure
at 1 kHz 0.00001 mH to 99 999 H 0.20 mH/H
RLC 1689 Digibridge
Quadtech OEM*, CEM,EA
ABNT or Internal
Procedures Inductance - Source
at 1 kHz 100 µH to 10 H 23 mH/H
Decade Inductor
1491
Version 009 Issued: 8/01/2014 Page 4 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
AC Voltage – Source5
(1 to 33) mV
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(33 to 330) mV
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
330 mV to 3.3 V
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 500) kHz
(3.3 to 33) V
(10 to 45) Hz
45 Hz to 10 kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(33 to 330) V
45 Hz to 1 kHz
(1 to 10) kHz
(10 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
330 V to 1.02 kV
45 Hz to 1 kHz
(1 to 5) kHz
(5 to 10) kHz
0.80 mV/V + 6.8 µV
0.15 mV/V + 6.8 µV
0.20 mV/V + 6.8 µV
1 mV/V + 10 µV
3.5 mV/V + 25 µV
8 mV/V + 55 µV
0.30 mV/V + 13 µV
0.15 mV/V + 13 µV
0.16 mV/V + 13 µV
0.35 mV/V + 17 µV
0.80 mV/V + 40 µV
2 mV/V + 1.1 mV
0.30 mV/V + 0.11 mV
0.15 mV/V + 0.11 mV
0.19 mV/V + 0.11 mV
0.30 mV/V + 0.13 mV
0.70 mV/V + 0.23 mV
2.4 mV/V + 1.2 mV
0.30 mV/V + 1.2 mV
0.15 mV/V + 1.1 mV
0.24 mV/V + 1.1 mV
0.35 mV/V + 1.5 mV
0.90 mV/V + 2.8 mV
0.19 mV/V + 13 mV
0.20 mV/V + 14 mV
0.25 mV/V + 14 mV
0.30 mV/V + 14 mV
2 mV/V + 59 mV
0.30 mV/V + 33 mV
0.25 mV/V + 36 mV
0.30 mV/V + 36 mV
Fluke 5520A SC600
OEM*, CEM,
EURAMET, EA or
Internal
Procedures
AC Voltage - Measure5
Up to 200 mV
(1 to 10) Hz
(10 to 40) Hz
(40 to 100) Hz
100 Hz to 2 kHz
(2 to 10) kHz
(10 to 30) kHz
(30 to 100) kHz
0.12 mV/V + 16 µV
0.12 mV/V + 7.4 µV
0.10 mV/V + 9.3 µV
0.10 mV/V + 7.4 µV
0.10 mV/V + 8.1 µV
0.30 mV/V + 8.0 µV
0.70 mV/V + 31 µV
Fluke 8508A Opt 001
Version 009 Issued: 8/01/2014 Page 5 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
AC Voltage – Measure5
(cont.)
200 mV to 2 V
(1 to 10) Hz
(10 to 40) Hz
(40 to 100) Hz
100 Hz to 2 kHz
(2 to 10) kHz
(10 to 30) kHz
(30 to 100) kHz
(100 to 300) kHz
300 kHz to 1 MHz
(2 to 20) V
(1 to 10) Hz
(10 to 40) Hz
(40 to 100) Hz
100 Hz to 2 kHz
(2 to 10) kHz
(10 to 30) kHz
(30 to 100) kHz
(100 to 300) kHz
300 kHz to 1 MHz
(20 to 200) V
(1 to 10) Hz
(10 to 40) Hz
(40 to 100) Hz
100 Hz to 2 kHz
(2 to 10) kHz
(10 to 30) kHz
(30 to 100) kHz
(100 to 300) kHz
300 kHz to 1 MHz
200 V to 1.05 kV
(1 to 10) Hz
(10 to 40) Hz
40 Hz to 10 kHz
(10 to 30) kHz
(30 to 100) kHz
Up to 6 kV 60 Hz
0.10 mV/V + 0.15 mV
0.10 mV/V + 83 µV
80 µV/V + 46 µV
60 µV/V + 50 µV
80 µV/V + 57 µV
0.20 mV/V + 0.12 mV
0.50 mV/V + 0.34 mV
3 mV/V + 2.3 mV
10 mV/V + 21 mV
0.10 mV/V + 1.5 mV
0.10 mV/V + 5.8 mV
80 µV/V + 0.54 mV
60 µV/V + 0.54 mV
80 µV/V + 0.53 mV
0.20 mV/V + 0.58 mV
0.50 mV/V + 2.6 mV
3 mV/V + 22 mV
10 mV/V + 0.21 V
0.10 mV/V + 14 mV
0.10 mV/V + 6.6 mV
80 µV/V + 4.5 mV
60 µV/V + 4.5 mV
80 µV/V + 4.5 mV
0.20 mV/V + 6.4 mV
0.50 mV/V + 27 mV
3 mV/V + 0.20 V
10 mV/V + 2 V
0.10 mV/V + 80 mV
0.10 mV/V + 37 mV
80 µV/V + 0.12 V
0.20 mV/V + 29 mV
0.50 mV/V + 36 mV
15 mV/V
Fluke 8508A Opt 001
High Voltage Probe
80K-6 together with
Fluke 289
OEM*,CEM,
EURAMET,EA or
Internal
Procedures
Version 009 Issued: 8/01/2014 Page 6 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
AC Current - Source5
Clamp-On Current
(29 to 200) µA
(1 to 10) Hz
10 Hz to 10 kHz
(10 to 30) kHz
200 µA to 2 mA
(1 to 10) Hz
10 Hz to 10 kHz
(10 to 30) kHz
(2 to 20) mA
(1 to 10) Hz
10 Hz to 10 kHz
(10 to 30) kHz
(20 to 200) mA
(1 to 10) Hz
10 Hz to 10 kHz
(10 to 30) kHz
200 mA to 2 A
(1 to 10) Hz
10 Hz to 2 kHz
(2 to 10) kHz
(2 to 20) A
45 Hz to 2 kHz
(2 to 5) kHz
(20 to 1 025) A
(45 to 65) Hz
(65 to 400) Hz
0.25 mA/A + 24 nA
0.25 mA/A + 46 nA
0.60 mA/A + 46 nA
0.25 mA/A + 0.29 µA
0.25 mA/A + 0.27 µA
0.60 mA/A + 0.27 µA
0.25 mA/A + 2.9 µA
0.25 mA/A + 3.2 µA
0.60 mA/A + 3.2 µA
0.25 mA/A + 29 µA
0.25 mA/A + 29 µA
0.60 mA/A + 29 µA
0.60 mA/A + 0.29 mA
0.70 mA/A + 0.45 mA
3 mA/A + 1.1 mA
0.80 mA/A + 3.6 mA
2.5 mA/A + 11 mA
3.9 mA/A
8.9 mA/A
Fluke 5520A SC 600,
Fluke 8508A Opt 01
with 5500A Coil
OEM*, CEM
EURAMET or
Internal
Procedures
AC Current - Measure5
Up to 200 µA
(1 to 10) Hz
10 Hz to 10 kHz
(10 to 30) kHz
(30 to 100) kHz
200 µA to 2 mA
(1 to 10) Hz
10 Hz to 10 kHz
(10 to 30) kHz
(30 to 100) kHz
(2 to 20) mA
(1 to 10) Hz
10 Hz to 10 kHz
(10 to 30) kHz
(30 to 100) kHz
(20 to 200) mA
(1 to 10) Hz
10 Hz to 10 kHz
(10 to 30) kHz
0.25 mA/A + 29 nA
0.25 mA/A + 45 nA
0.60 mA/A + 45 nA
4 mA/A + 45 nA
0.25 mA/A + 0.29 µA
0.25 mA/A + 0.27 µA
0.60 mA/A + 0.27 µA
4 mA/A + 0.27 µA
0.25 mA/A + 2.9 µA
0.25 mA/A + 3.2 µA
0.60 mA/A + 3.2 µA
4 mA/A + 3.2 µA
0.25 mA/A + 29 µA
0.25 mA/A + 29 µA
0.60 mA/A + 29 µA
Fluke 8508A Opt 01
Version 009 Issued: 8/01/2014 Page 7 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
AC Current - Measure5
(cont.)
200 mA to 2 A
10 Hz to 2 kHz
(2 to 10) kHz
(10 to 30) kHz
(2 to 20) A
10 Hz to 2 kHz
(2 to 10) kHz
(20 to 50) A
60 Hz
0.60 mA/A + 0.29 mA
0.70 mA/A + 1.1 mA
3 mA/A + 1.1 mA
0.80 mA/A + 3.6 mA
2.5 mA/A + 3.6 mA
18 mA/A
Fluke 8508A Opt 01
with Current Shunt
1651 Keithley
OEM*, CEM
EURAMET,EA
ABNT or Internal
Procedures
Electrical Simulation of
Thermocouples5
Type B
Type C
Type E
Type J
Type K
(600 to 800) °C
(800 to 1 000) °C
(1 000 to 1 550) °C
(1 550 to 1 820) °C
(0 to 150) °C
(150 to 650) °C
(650 to 1 000) °C
(1 000 to 1 800) °C
(1 800 to 2 316) °C
(-250 to -100) °C
(-100 to -25) °C
(-25 to 350) °C
(350 to 650) °C
(650 to 1 000) °C
(-210 to -100) °C
(-100 to -30) °C
(-30 to 150) °C
(150 to 760) °C
(760 to 1 200) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 1 000) °C
(1 000 to 1 372) °C
0.44 °C
0.34 °C
0.3 °C
0.33 °C
0.3 °C
0.26 °C
0.31 °C
0.5 °C
0.84 °C
0.5 °C
0.16 °C
0.14 °C
0.16 °C
0.21 °C
0.27 °C
0.16 °C
0.14 °C
0.17 °C
0.23 °C
0.33 °C
0.18 °C
0.16 °C
0.26 °C
0.4 °C
Fluke 5520A SC600
Version 009 Issued: 8/01/2014 Page 8 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Electrical Simulation of
Thermocouples5 (cont.)
Type L
Type N
Type R
Type S
Type T
Type U
(-200 to -100) °C
(-100 to 800) °C
(800 to 900) °C
(-200 to -100) °C
(-100 to -25) °C
(-25 to 120) °C
(120 to 410) °C
(410 to 1 300) °C
(0 to 250) °C
(250 to 400) °C
(400 to 1 000) °C
(1 000 to 1 767) °C
(0 to 250) °C
(250 to 1 000) °C
(1 000 to 1 400) °C
(1 400 to 1 767) °C
(-250 to -150) °C
(-150 to 0) °C
(0 to 120) °C
(120 to 400) °C
(-200 to 0) °C
(0 to 600) °C
0.37 °C
0.26 °C
0.17 °C
0.4 °C
0.22 °C
0.19 °C
0.18 °C
0.27 °C
0.57 °C
0.35 °C
0.33 °C
0.4 °C
0.47 °C
0.36 °C
0.37 °C
0.46 °C
0.63 °C
0.24 °C
0.16 °C
0.14 °C
0.56 °C
0.27 °C
Fluke 5520A SC600
OEM*, CEM,EA
ABNT or Internal
Procedures
Electrical Simulation of
RTDs5
Pt 395, 100 Ω
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 300) °C
(300 to 400) °C
(400 to 630) °C
(630 to 800) °C
0.05 °C
0.05 °C
0.07 °C
0.09 °C
0.10 °C
0.12 °C
0.23 °C
Version 009 Issued: 8/01/2014 Page 9 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Electrical Simulation of
RTDs5 (cont.)
Pt 385, 200 Ω
Pt 385, 1000 Ω
Cu 427, 10 Ω
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
(-200 to -80) °C
(-80 to 0) °C
(0 to 100) °C
(100 to 260) °C
(260 to 300) °C
(300 to 400) °C
(400 to 600) °C
(600 to 630) °C
(-100 to 260) °C
0.04 °C
0.04 °C
0.04 °C
0.05 °C
0.12 °C
0.13 °C
0.14 °C
0.16 °C
0.03 °C
0.03 °C
0.04 °C
0.05 °C
0.06 °C
0.07 °C
0.07 °C
0.23 °C
0.3 °C
Fluke 5520A SC600
OEM*, CEM,EA
ABNT or
Internal
Procedures
Capacitance - Source
(1 kHz)
10 Hz to 10 kHz
10 Hz to 3 kHz
10 Hz to 1 kHz
10 Hz to 1 kHz
10 Hz to 1 kHz
(10 to 600) Hz
(10 to 300) Hz
(10 to 150) Hz
(10 to 120) Hz
(10 to 80) Hz
(0 to 50) Hz
(0 to 20) Hz
(0 to 6) Hz
(0 to 2) Hz
(0 to 0.6) Hz
(0 to 0.2) Hz
1 pF to 100 nF
(0.19 to 1.1) nF
(1.1 to 3.3) nF
(3.3 to 11) nF
(11 to 110) nF
(110 to 330) nF
330 nF to 1.1 µF
(1.1 to 3.3) µF
(3.3 to 11) µF
(11 to 33) µF
(33 to 110) µF
(110 to 330) µF
330 µF to 1.1 mF
(1.1 to 3.3) mF
(3.3 to 11) mF
(11 to 33) mF
(33 to 110) mF
0.50 F/F + 0.5 pF
5 mF/F + 10 pF
5 mF/F + 10 pF
2.5 mF/F + 10 pF
2.5 mF/F + 0.1 nF
2.5 mF/F + 0.3 nF
2.5 mF/F + 1 nF
2.5 mF/F + 3 nF
2.5 mF/F + 10 nF
4 mF/F + 30 nF
4.5 mF/F + 0.1 µF
4.5 mF/F + 0.3 µF
4.5 mF/F + 1 µF
4.5 mF/F + 3 µF
4.5 mF/F + 10 µF
7.5 mF/F + 30 µF
11 mF/F + 0.1 mF
Decade Capacitor 1413
Fluke 5520A SC600
Capacitance - Measure
at 1 kHz 0.00001 pF to 1 F 0.20 mF/F
RLC 1689 Digibridge
Quadtech
Version 009 Issued: 8/01/2014 Page 10 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Oscilloscopes
Amplitude - DC Signal
into 50 Ω Load
into 1 MΩ Load
Amplitude - Square wave
into 50 Ω Load
into 1 MΩ Load
Rise Time
Leveled Sine Wave
Relative to 50 kHz
Time Marker3
into 50 Ω Load
Wave Generator -
Amplitude
10 Hz to 10 kHz
Square, Sine, Triangle
into 1 MΩ Load
into 50 Ω Load
Frequency
0 V to ± 6.6 V
0 V to ± 130 V
±1 mV to ± 6.6 V p-p
(10 Hz to 10 kHz)
±1 mV to ± 130 V p-p
10 Hz to 1 kHz
(1 to 10) kHz
300 ps
5 mV to 5.5 V
50 kHz Reference
50 kHz to 100 MHz
(100 to 300) MHz
(300 to 600) MHz
5 s to 50 ms
20 ms to 100 ns
(50 to 20) ns
10 ns
(5 to 1) ns
1.8 mV to 55 V pp
1.8 mV to 2.5 V pp
10 Hz to 100 kHz
2,5 mV/V + 40 µV
0.50 mV/V + 40 µV
2.5 mV/V + 40 µV
1 mV/V + 40 µV
2.5 mV/V + 40 µV
+ 0 ps / -100 ps
20 mV/V + 0.30 mV
35 mV/V + 0.30 mV
40 mV/V + 0.30 mV
60 mV/V + 0.30 mV
(25 + 1 000t) µs/s
2.5 µs/s
2.5 µs/s
2.5 µs/s
2.5 µs/s
30 mV/V + 0.10 mV
30 mV/V + 0.10 mV
25 µHz/Hz + 15 mHz
Fluke 5520A SC600
OEM*, CEM,
EURAMET, IEE or
Internal
Procedures
DC Power - Source5
(33 mV to 1.02 kV)
(10.8 µW to 20.9 kW)
330 µA to 330 mA
330 mA to 3 A
(3 to 20.5) A
0.23 mW/W
0.22 mW/W
0.70 mW/W
AC Power - Source5
(10 to 65) Hz
1 mW to 99 W
100 W to 20 kW
(65 to 500) Hz
1 mW to 20 kW
500 Hz to 1 kHz
1 mW to 20 kW
(1 to 5) kHz
1 mW to 1.5 kW
0.15 mW/W
19 µW/W
19 µW/W
19 µW/W
8 mW/W
Version 009 Issued: 8/01/2014 Page 11 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
DC Power - Measure
Up to 7.5 W
(7.5 to 30) W
(30 to 120) W
(120 to 750) W
750 W to 3 kW
(3 to 12) kW
3 mW/W + 23 mW
3 mW/W + 9 mW
3 mW/W + 0.36 W
3 mW/W + 2.3 W
3 mW/W + 9 W
3 mW/W + 36 W Yokogawa WT110
Digital Power Meter
OEM* ,CEM, IEE
or Internal
Procedures
AC Power - Measure
10 Hz to 5 kHz
Up to 7.5 W
(7.5 to 30) W
(30 to 120) W
(120 to 750) W
750 W to 3 kW
(3 to 12) kW
5 mW/W + 23 mW
5 mW/W + 9 mW
5 mW/W + 0.36 W
5 mW/W + 2.3 W
5 mW/W + 9 W
5 mW/W + 36 W
II. Electromagnetic - RF / Microwave6
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
RF Power - Source
into 50 Ω load
100 kHz to 3.2 GHz
(-120 to 13) dBm
(3.2 to 20) GHz
(-60 to 20) dBm
9 kHz to 3.2 GHz
-55 dBm
(-50 to -40) dBm
(-39 to 20) dBm
0.18 dB
0.18 dB
1.7 dB
0.6 dB
0.3 dB
Rohde & Schwarz
SMB 100A
RF Signal Generator
E8257D and
Agilent E4418B with
Power Sensors,
E9300A, E4412A
and E9304A H18
OEM*,CEM, IEE or
Internal
Procedures
RF Power - Measure
into 50 Ω load
10 MHz to 18 GHz
-55 dBm
-50 dBm to -40 dBm
(-40 dBm to + 20 dBm)
9 kHz to 6 GHz
(-60 dBm to +20 dBm)
1.7 dB
0.5 dB
0.19 dB
0.18 dB
Agilent E4418B
w/E9300A/E4412A
Sensor
Agilent E4418B
with E9304A H18
Version 009 Issued: 8/01/2014 Page 12 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
RF Power - Source
into 50 Ω load
(0 to 100) W
(20 MHz to 1 GHz) 0.9 dB
RF Signal Generator
E8257D
RF Power Amplifier
Ophir 5300796 &
Ophir 5303031
Power Attenuator
PE7019-X, 40-3-43,
60-1-43
Power Meter E4418B with
Power Sensors, E9300A,
E4412A And E9304A H18
USB Power Sensor
NRP-Z91
OEM*,CEM, IEE or
Internal Procedures
RF Power - Measure
into 50 Ω load
(0 to 200) W
(20 MHz to 1 GHz) 0.9 dB
Power Attenuator
PE7019-X, 40-3-43,
60-1-43
Power Meter E4418B with
Power Sensors, E9300A,
E4412A And E9304A H18
USB Power Sensor
NRP-Z91
Version 009 Issued: 8/01/2014 Page 13 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
III. Time and Frequency
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Frequency - Source
Time Base @ 10 MHz
10 µHz to 1.64 GHz
(1.64 to 20) GHz
10 MHz
100 ns to 10 000 s
9 x 10-11
Hz/Hz
5 x 10-8
Hz/Hz
4 x 10-11
Hz/Hz
9 x 10-11
s/s
PTS-SG 1640
GPS Disciplined
Fluke 910 GPS Frequency
Standard
OEM* or
Internal Procedures
Frequency - Measure5
Time Base @ 10 MHz
1 Hz to 100 kHz
100 kHz to 225 MHz
225 MHz to 12.5 GHz
5.8 x 10-9
Hz/Hz
2.6 x 10-10
Hz/Hz
5.9 x 10-9
Hz/Hz
Agilent 53132A
enhanced with
Fluke 910A GPS (100 seconds gate time)
Agilent 53131A,
enhanced with
Fluke 910A GPS
OEM*CEM or
Internal Procedures
IV. Thermodynamic
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Temperature -
Measure5
(-200 to 650) ºC
(650 to 1 000) ºC
(1 000 to 1 450) ºC
0.02 ºC
0.36 ºC
0.81 ºC
Fluke 8508A
and Hart 5626 PRT
Fluke 8508A and
5650 Type S
Thermocouple
OEM*,CEM, NPL
ISBN, ASME,
ASTM or Internal
Procedures
Temperature - Source5
0 ºC
(-30 to 650) ºC
(650 to 1 000) ºC
(1 000 to 1 450) ºC
0.01° C
0.13 ºC
2.7 ºC
2.2 ºC
Fluke 8508A with 5626
PRT, Dry Block Venus,
Water Triple Point Cell
Isotech B8/30/130
Fluke 8508A with 5626
PRT, Dry Block Venus,
Dry Block 9144
Fluke 8508A with
5650 Type S
Thermocouple,
Fluke 9150-A-156
Furnace 47900
Version 009 Issued: 8/01/2014 Page 14 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Relative Humidity
Humidity Chamber -
Measure
(11 to 80) % RH 1.1 %RH Vaisala HMI41/46
Humidity Chamber OEM*, CEM, ASME
NPL ISBN, ASTM
or Internal
Procedures Relative Humidity
Saturated Salts -
Source5
LiCl salt (11.3 %RH)
MgCl2 salt (32.8 %RH)
NaCl salt (75.4 %RH)
K2SO4 (97.4 %RH)
1.3 %RH
1.3 %RH
1.5 %RH
2.0 % RH
Vaisala Humidity
Calibrator HMK15 with
Saturated Salt Solutions
V. Mechanical
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Balances/Scales5
(Resolution in
Parentheses)
Up to 300 g (0.0001 g)
300 g to 1 kg (0.001 g)
(1 to 6.5) kg (0.01 g)
(6.5 to 33) kg (0.1 g)
(33 to 50) kg (5 g)
(50 to 100) kg (10 g)
(100 to 200) kg (20 g)
(200 to 500) kg (50 g)
(500 to 1 000) kg (100 g)
(1 000 to 4 000) kg (200 g)
(4 000 to 5 000) kg (500 g)
0.24 mg
0.8 mg
9 mg
84 mg
4 g
8 g
20 g
40 g
80 g
0.2 kg
0.4 kg
Mass OIML Class E2,
F1 & M1
Mass ASTM E617
Class 1
ASTM E617-91 Class 6
& NIST 105-1
Class F Weights
OEM*, NIST
Handbook 44, OIML R
76-2, NTC 2031, SIM
Guidelines, CEM or
Internal Procedures
Mass - Measure
(1 to 500) mg
(1 to 20) g
50 g
100 g
200 g
300 g
500 g
1 kg
2 kg
5 kg
6 kg
10 kg
20 kg
50 lb
0.06 mg
0.07 mg
0.08 mg
0.1 mg
0.16 mg
0.24 mg
0.7 mg
1 mg
6 mg
14 mg
14 mg
63 mg
71 mg
71 mg
Electronic Comparators
Radwag XA310, Mettler
XS1003S,
A&D GX-6100 and
Vibra HJ-33KSE
with Mass ASTM
E617Class 1 and
OIML Class E2 & F1
OEM*,
NIST Handbook 44,
OIML-R111-2, NTC
1848, CEM, ASTM,
NTC or Internal
Procedures
Version 009 Issued: 8/01/2014 Page 15 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Air Flow (2.5 to 15) m/s
(492 to 2 953) FPM
0.12 m/s
24 FPM
Wind Tunnel
WTM-1000
Standard Anemometer
PAMB11
PRT with Readout 1522
OEM*,
IEC 61400-12-1 Annex
F, CEM, ASME,IEC or
Internal Procedures
Pressure5,
Absolute and Gauge
Mode
(0 to 6.9) kPa
(6.9 to 1 723) kPa
(1 723 to 6 894) kPa
(6 894 to 68 948) kPa
1 mPa/Pa + 3 Pa
172 Pa
80 µPa/Pa
0.2 mPa/Pa
Module 700P00 with
Fluke 754
ADT681-10-DP5-H2O
ADT681-05-DP20-H20
PAMB11
FLUKE RPM4
DHI RPM3
OEM*, OIML, ASME,
DKD R6-1,CEM,
OIML or Internal
Procedures
Vacuum5,
Absolute and Gauge
Mode
(-103 to 0) kPa
(-30 to 0) in·Hg
0,2 mPa/Pa
0,005 in·Hg
RPM4 A Mode
RPM3 A Mode,
Crystal IS33-
16/300PSI/SPSS
PAMB11
Force5 -
Source
Certified Mass
Up to 2 224 N 0.00026 N/N
Weight Classes
ASTM E617-91 Class 6
& NIST 105-1 Class F
OEM*,ASTM,
NIST Handbook 44,
NTC- ISO 7500-1,
CEM, OIML or Internal
Procedures
Force - Compression
and Weighing System
Source and Measure5
100 N to 100 kN
(100 to 445) kN
0.00034 N/N
0.00066 N/N
MGC plus ML38/AP03i
with Load Cells U9B
HBM, Z30A HBM, Z4A
HBM and C2P1 BLH OEM*, ASTM E74,
NTC/ISO 7500-1,
CEM, OIML or Internal
Procedures Force - Tension
Source and Measure5
100 N to100 kN
0.00034 N/N
MGC plus ML38/AP03i
with Load Cells U9B
HBM, Z30A HBM, Z4A
HBM
Version 009 Issued: 8/01/2014 Page 16 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Torque Transducer (0,014 to 2 712) N.m
2 oz·in to 2 000 lbf·ft 0.11 % of reading
NIST Class F Weights
with
2.5 in Radius Wheel,
5 in Radius Wheel,
10 in Radius Butterfly,
and 40 in Radius Arm OEM*, ASME
B107.300, DIN EN, ISO 6789, EURAMET/cg-14
CEM, OIML or Internal
Procedures
Torque Wrenches5
2 oz·in to 2 000 lbf·ft
(0.014 to 2 712) N.m
0.25 % of reading TTC2000 Torque
Calibrator Snap-On
Rotary Torque Tools5 (15 to 300) lbf·ft
(20 to 407) N.m 0.13 % of reading
Torque Tool Tester and
Impulse Rotary
Transducer
Norbar 43228
VI. Dimensional
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Length (Source) 2,5
Calipers Outside.
Up to 24 in
(24 to 60) in
(Up to 609) mm
(320 + 7.0L) µin
(330 + 8.5L) µin
(8.1 + 0.0072L) µm
Gage Blocks Grade 0
OEM*, NTC, JIS,BS,
CEM, OIML,ASME
or Internal
Procedures
Length (Source) 2,5
Calipers Depth
Up to 12 in
(Up to 304) mm
(320 + 8.3L) µin
(8.1 + 0.0083L) µm
Gage Blocks Grade 0
Flatness Surface
Length (Source) 2,5
Calipers Inside
Up to 16 in
(Up to 609) mm
(320 + 7.0L) µin
(8.1 + 0,008L) µm
Gage Blocks Grade 0
Ceramic Micro Checker
515-586
Version 009 Issued: 8/01/2014 Page 17 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Length (Source) 2,5
Micrometers
Outside
Up to 1 in
(1 to 24) in
(24 to 40) in
Up to 25 mm
(25 to 609) mm
(5.7 + 7.0L) µin
(35 + 7.0L) µin
(190 + 7.0L) µin
(0.18 + 0.0071L) µm
(0.89 + 0.0071L) µm
Gage Blocks Grade 0
Optical Flat
OEM*, JIS,
BS, ASME B89, CEM,
OIML, BS or Internal
Procedures
Length (Source) 2,5
Micrometers
Inside.
Up to 16 in
Up to 600 mm
(42 + 7.0 L) µin
(1.4 + 0.007L) µm
Gage Blocks Grade 0
Ceramic Micro Checker
515-586
Hite Icator
OEM*, JIS,
BS, ASME B89, CEM,
OIML or Internal
Procedures
Length (Source) 2,5
Micrometers Depth,
Height, Dial Indicators
Up to 12 in
Up to 750 mm
(35 + 9.9 L) µin
(6.4 + 0.012 L) µm
Gage Blocks Grade 0
Depth Micro Checker
515-575
Granite Surface Plate
517-164
OEM*,
BS, ASME, CEM,
OIML or Internal
Procedures
Length (Source) 2,5
Dial Gage Tester
Up to 1 in
Up to 25 mm
(19 + 7.0 L) µin
(0.45 + 0.007 L) µm Gage Blocks Grade 0
OEM* ASME,
CEM, OIML or
Internal Procedures
Length (Source)5
Dial Indicators
Up to 1 in
Up to 25 mm
120 µin
3.0 µm
Dial Gage Tester
170-102M-2 and
170-101E
OEM*, NTC,ASME
BS , CEM, OIML or
Internal Procedures
Length - Measure5
Up to 1 in
(1 to 12) in
(12 to 48) in
Up to 25 mm
(25 to 101) mm
26 µin
1 200 µin
1 800 µin
1.2 µm
37 µm
Digimatic Micrometer
293-130
Digimatic Caliper
500 Series
OEM*, CEM,
OIML,ASME or
Internal Procedures
Length - Ruler Up to 36 in
Up to 1 000 mm
2 700 µin
160 µm
Standard Steel Ruler
Magnifying Glass
183-304
Granite Surface Plate
517-164
OEM*, CEM, OIML
or Internal Procedures
Length -
Flexometer
Up to 4 000 mm
(4 000 to 8 000) mm
(8 000 to 10 000) mm
0.95 mm
1.9 mm
2.9 mm
Standard Flexometer
CM182
Magnifying Glass
183-304
OEM*,CEM, ASME,
OIML, NOM-046
SCFI or Internal
Procedures
Version 009 Issued: 8/01/2014 Page 18 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Angle - Source
Clinometer, Level,
Protractor
(0 to 180) °
0.036 °
Mitutoyo Angle Blocks
981-102
Granite Surface Plate
517-164
Precision level
960-616
OEM* CEM, OIML,
ASME, DIN, NTC or
Internal Procedures
Angle5 - Measure
(0 to 180) ° 0.063°
Rank Taylor Hobson
Clinometer 142/42
Flatness5 (9.3 to 120) in
236 mm to 3 m
140 µin
3.5 µm
Flatness Gage LAP5
Flatness Reference
G-81703
Precision Level
960-616
OEM*, OIML, ASME,
BS,
CEM or Internal
Procedures
Holes 5
(0.011 to 0.5) in 130 µin C10P Meyer Gage
Makers Set Class Z
OEM* CEM, OIML,
DIN or Internal
Procedures
VII. ANSI/NCSL Z540-1-1994 Only (ISO/IEC 17025 Traceability Requirements Not Verified)
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
AC Phase Angle
- Measure
(0 to 180)° 10 Hz
50 Hz to 5 kHz
20 kHz
32 kHz
54 kHz
100 kHz
0.1 °
0.05 °
0.2 °
0.32 °
0.54 °
1 °
North Atlantic 2250
OEM* ,CEM, IEE or
Internal
Procedures Synchro Resolver
- Angle Source
Phase Shift (0 to 180) °
(2 to 115) V rms 47 Hz to 2 kHz (2 to 20) kHz
0.5 °
5 °
North Atlantic 5300
Phase Meter
Radial Bearing (0 to 340) ° 0.03 °
478A-3 Zifor III
Collins
Version 009 Issued: 8/01/2014 Page 19 of 19
500 Montgomery St. Suite 625│ Alexandria, VA 22314│703-836-0025 │ www.aclasscorp.com
VII (cont.). ANSI/NCSL Z540-1-1994 Only (ISO/IEC 17025 Traceability Requirements Not Verified)
PARAMETER/
EQUIPMENT RANGE
CALIBRATION &
MEASUREMENT
CAPABILITY
[EXPRESSED AS
UNCERTAINTY(+)]
REFERENCE
STANDARD OR
EQUIPMENT
METHOD(S)
Distortion - Measure
100 mV to 130 V
(0 to 0.1) % THD
10 Hz to 500 kHz
500 kHz to 1 MHz
(0.1 to 2 ) % THD
10 Hz to 500 kHz
500 kHz to 1 MHz
(2 to 20) % THD
10 Hz to 500 kHz
500 kHz to 1 MHz
0.02 %
0.03 %
0.08 %
0.01 %
0.08 %
0.13 %
Krohn - Hite 6900B
OEM* ,CEM, IEE or
Internal
Procedures
Temperature - Infrared
Non-Contact5
(50 to 500) ºC 0.88 ºC Hart 9132
Blackbody Source
OEM*,CEM, NPL
ISBN, ASME,
ASTM or Internal
Procedures
Altitude/
Static Pressure
(RVSM Compliant)
(Ps) Altitude
(Pt) Airspeed
(-1 500 to 80 000) ft
(31.58 to 0.815) in·Hg
(25 to 1 000) KTS
(29 951 to 103 467)
in·Hg
0,003 in·Hg @ 0 ft
0,003 in·Hg @ 0 Knots
PPCM
18905810002
TVCS-9100
Kollsman
Pressure Indicator Dual
Channel PAMB11
OEM* or Internal
Procedures
Sound Pressure Level
Source
1 kHz
@ 0 dBa = 20 µPa
114 dBa or 94 dBa
0.5 dB Simpson 887-2
Sound Level Calibrator
OEM* CEM, OIML,
IEC or Internal
Procedures
Crimping Tools5
(0.025 to 0.072) in 44 µin
Digital Micrometer, C10P
Meyer Gage Makers Set
Class Z and
Go-No Go Gages
OEM* CEM, OIML,
ASME, DIN, NTC or
Internal Procedures
Notes: 1. Calibration and Measurement Capabilities (CMC) (Expanded Uncertainties) are based on approximately a 95% confidence interval, using a
coverage of k=2.
2. The use of (L) signifies an expression of Length in inches or mm depending on the unit system. 3. The use of (t) refers to Time in seconds.
4. OEM* (Original Equipment Manufacturer Procedure), CEM (Spanish metrology center procedure), OIML (Organization International of
Legal Metrology, ASME(American Society of Mechanical Engineer procedure), BS ( British standard), NTC (Colombian technical standard),EURAMET, EA(European standard), other XXX (standard from national or international metrological institutes)
5. This laboratory also offers this calibration service on-site at customer-designated locations. Since on-site conditions are typically more
variable than those in the laboratory, larger measurement uncertainties may be anticipated. 6. Uncertainties listed for Electromagnetic - RF/Microwave do not include possible contributions from mismatch error.
7. This scope is part of and must be included with the Certificate of Accreditation No AC-1327.
_________________________ Vice President