2025 Partisol Sequential Sampler
description
Transcript of 2025 Partisol Sequential Sampler
![Page 1: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/1.jpg)
2025 Partisol Sequential Sampler
Deborah Bowe
Thermo Electron Corporation
![Page 2: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/2.jpg)
2
2025 Verification & Routine Maintenance
Verification• Ambient air temperature• Filter temperature• Ambient pressure• External leak check• Internal leak check• Flow rate
![Page 3: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/3.jpg)
3
2025 Verification & Routine Maintenance
Routine Maintenance• Filter cassettes: inspect damage or contamination
• Filter cassette magazines: Generally not necessary to clean, but if damaged certain parts can be replaced
• WINS impactor: In accordance with USEPA guidelines for PM2.5
• VSCC: In accordance with U.S. EPA guidelines
• External leak check: Every 4 weeks of use
• PM10 Inlet: Every 14 days of use
![Page 4: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/4.jpg)
4
2025 Verification & Routine Maintenance
Routine Maintenance• Particle trap filter: inspect 6 months, replace as
needed• V seals: Check and clean as needed• Rainhoods: Clean air screens 6 months• Batteries: Check every 6 months replace as
needed• Pump: Check every 6 months and rebuild; needed
average lifetime is 12 to 18 months
![Page 5: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/5.jpg)
5
2025 Plumbing Diagram
![Page 6: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/6.jpg)
6
2025 Backup Battery Check
Measure across 188_VBAT & PGND
Drops below 4.2VDC replace batteries
![Page 7: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/7.jpg)
7
2025 Pump CheckPressure Check: Quality Pump should give 40 psi on the exhaust
![Page 8: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/8.jpg)
8
Partisol 2000 Single Channel FRM Sampler
Deborah Bowe
Thermo Electron Corporation
![Page 9: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/9.jpg)
9
2000 FRM Audit Screen
Ambient temperature + 2 °C
Filter temperature + 2 °C
Ambient pressure + 10 mmHg
![Page 10: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/10.jpg)
10
2000 FRM External Leak CheckManual method
Record reading on gauge
30 seconds after pump
valve is closed
Should not drop more
than 8.5 inHg in 30
seconds. Corresponds to
80ml/min rate
![Page 11: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/11.jpg)
11
2000 FRM External Leak CheckSemi Automatic
Tolerance is 5” in 60 seconds
![Page 12: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/12.jpg)
12
2000 FRM Internal Leak Check
Manual: 8.5”Hg in 30 seconds
Semi-automatic: 9.3”Hg in 30 seconds
![Page 13: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/13.jpg)
13
Flow Schematic
![Page 14: 2025 Partisol Sequential Sampler](https://reader036.fdocuments.in/reader036/viewer/2022062314/568146c2550346895db3f7e6/html5/thumbnails/14.jpg)
14
2000 FRM Flow Audit
Streamline Pro, FTS or Bios flow meter
+ 5% tolerance