2010 series warmers

37
nice NeÖtech Medical Systems P

Transcript of 2010 series warmers

nice NeOumltech Medical Systems Pvt Ltd

Presentation Flow

COMPANY OVERVIEW

CLINICAL ASPECTS OF nice 2010 SERIES WARMERS

PRODUCT OVERVIEW

TROUBLESHOOTING

Company INFO

Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive

for growth and delighted customers

Timeline-nice Neotech

2005bullFirst New Factory Unit 1 Opened

bullIncrease in production Line

2001bullRegistered as Pvt ltd

bullProduction started for different models

2013bullSecond New Factory Unit 2 Opened

1999bull Neotech Logo

Designed

bull Designed first incubator amp open

care prototype

2003bull Export to

Vietnam

2009-2011

bull CE certification for products

2015bull 125+skilled

employee strong firm

bull Multiple range of NICU products

1997bullStarted Neotech Medical Systems with 2 employees

bullFirst Infant Radiant Warmer

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

Why Infant Radiant Warmers are Used

Assist Thermoregulation

Body warming by providing Far IR radiant heat

Ensure baby is stabilized after birth

INFANT RADIANT WARMERS

MODES OF HEAT LOSS

Conduction Heat Loss Direct heat transfer from skin to object

Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

Electromagnetic Spectrum

Infrared Spectrum

Effects of Far IR in Infant body

Far Infrared Rays are invisible waves of energy

They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Presentation Flow

COMPANY OVERVIEW

CLINICAL ASPECTS OF nice 2010 SERIES WARMERS

PRODUCT OVERVIEW

TROUBLESHOOTING

Company INFO

Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive

for growth and delighted customers

Timeline-nice Neotech

2005bullFirst New Factory Unit 1 Opened

bullIncrease in production Line

2001bullRegistered as Pvt ltd

bullProduction started for different models

2013bullSecond New Factory Unit 2 Opened

1999bull Neotech Logo

Designed

bull Designed first incubator amp open

care prototype

2003bull Export to

Vietnam

2009-2011

bull CE certification for products

2015bull 125+skilled

employee strong firm

bull Multiple range of NICU products

1997bullStarted Neotech Medical Systems with 2 employees

bullFirst Infant Radiant Warmer

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

Why Infant Radiant Warmers are Used

Assist Thermoregulation

Body warming by providing Far IR radiant heat

Ensure baby is stabilized after birth

INFANT RADIANT WARMERS

MODES OF HEAT LOSS

Conduction Heat Loss Direct heat transfer from skin to object

Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

Electromagnetic Spectrum

Infrared Spectrum

Effects of Far IR in Infant body

Far Infrared Rays are invisible waves of energy

They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Company INFO

Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive

for growth and delighted customers

Timeline-nice Neotech

2005bullFirst New Factory Unit 1 Opened

bullIncrease in production Line

2001bullRegistered as Pvt ltd

bullProduction started for different models

2013bullSecond New Factory Unit 2 Opened

1999bull Neotech Logo

Designed

bull Designed first incubator amp open

care prototype

2003bull Export to

Vietnam

2009-2011

bull CE certification for products

2015bull 125+skilled

employee strong firm

bull Multiple range of NICU products

1997bullStarted Neotech Medical Systems with 2 employees

bullFirst Infant Radiant Warmer

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

Why Infant Radiant Warmers are Used

Assist Thermoregulation

Body warming by providing Far IR radiant heat

Ensure baby is stabilized after birth

INFANT RADIANT WARMERS

MODES OF HEAT LOSS

Conduction Heat Loss Direct heat transfer from skin to object

Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

Electromagnetic Spectrum

Infrared Spectrum

Effects of Far IR in Infant body

Far Infrared Rays are invisible waves of energy

They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Timeline-nice Neotech

2005bullFirst New Factory Unit 1 Opened

bullIncrease in production Line

2001bullRegistered as Pvt ltd

bullProduction started for different models

2013bullSecond New Factory Unit 2 Opened

1999bull Neotech Logo

Designed

bull Designed first incubator amp open

care prototype

2003bull Export to

Vietnam

2009-2011

bull CE certification for products

2015bull 125+skilled

employee strong firm

bull Multiple range of NICU products

1997bullStarted Neotech Medical Systems with 2 employees

bullFirst Infant Radiant Warmer

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

Why Infant Radiant Warmers are Used

Assist Thermoregulation

Body warming by providing Far IR radiant heat

Ensure baby is stabilized after birth

INFANT RADIANT WARMERS

MODES OF HEAT LOSS

Conduction Heat Loss Direct heat transfer from skin to object

Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

Electromagnetic Spectrum

Infrared Spectrum

Effects of Far IR in Infant body

Far Infrared Rays are invisible waves of energy

They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Why nice Neotech

User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction

Why Infant Radiant Warmers are Used

Assist Thermoregulation

Body warming by providing Far IR radiant heat

Ensure baby is stabilized after birth

INFANT RADIANT WARMERS

MODES OF HEAT LOSS

Conduction Heat Loss Direct heat transfer from skin to object

Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

Electromagnetic Spectrum

Infrared Spectrum

Effects of Far IR in Infant body

Far Infrared Rays are invisible waves of energy

They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Why Infant Radiant Warmers are Used

Assist Thermoregulation

Body warming by providing Far IR radiant heat

Ensure baby is stabilized after birth

INFANT RADIANT WARMERS

MODES OF HEAT LOSS

Conduction Heat Loss Direct heat transfer from skin to object

Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

Electromagnetic Spectrum

Infrared Spectrum

Effects of Far IR in Infant body

Far Infrared Rays are invisible waves of energy

They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

MODES OF HEAT LOSS

Conduction Heat Loss Direct heat transfer from skin to object

Convection Heat Loss (i) Heat loss through air flow (ii) Also depends on Air Temperature

Radiant Heat Loss Direct transfer by electromagnetic radiation in infrared spectrum

(i) Heat gained by radiation from external radiant energy source (ii) Heat lost by radiation to cooler walls of incubator

Evaporation Heat Loss Heat loss when water evaporates from skin and respiratory track Depends on maximum relative humidity of surrounding Less Humidity = More Evaporation

Heat Loss(wattsmeter square) Evaporation 60-80 wattsmeter square Radiation 50 wattsmeter square Convection 25 wattsmeter square Evaporation Negligible Total Heat Loss 135-155 wattsmeter square

Electromagnetic Spectrum

Infrared Spectrum

Effects of Far IR in Infant body

Far Infrared Rays are invisible waves of energy

They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Electromagnetic Spectrum

Infrared Spectrum

Effects of Far IR in Infant body

Far Infrared Rays are invisible waves of energy

They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Effects of Far IR in Infant body

Far Infrared Rays are invisible waves of energy

They have the ability to penetrate all layers of the infant body penetrating into the inner-most regions of the tissues muscles and boneFar Infrared Rays very gently increases blood flow by expanding the capillaries that carry blood It also increases oxygenation and regeneration of the bloodFar Infrared boosts the immune system increasing the amount of white blood cells and killer T-cellsFar Infrared can also increase the processing enzyme activity in digestive tract as well as boost metabolism

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Types of heaters Quartz

A quartz heater is a type of infrared heater The heating comes from a heating element encased inside a quartz tube The element produces a wavelength of infrared radiation The quartz protects the heating element and stops convection heat from escaping

Far IR Infrared

Infrared radiation is a type of electromagnetic radiation just below the frequency of visible light Pretty much everything emits infrared radiation The hotter something is the more infrared it emits Every molecule also absorbs infrared at a certain wavelength When a molecule absorbs infrared energy it heats up The IR rays by deep penetration in skin ensures the growth of cells by stimulating it

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Ceramic Heaters

Ceramic heaters are heaters that generate heat by passing electricity through heating wires embedded in ceramic plates The ceramic coating tend to crack over a period of time and can fall over the baby causing havoc

Tubular Heater

A tubular heater is designed from metal tubing There are chances of heat loss and heat conduction in relative heating methods compared with other heating methods like resistance heating infrared heating The heating pipe has higher cost due to metal material

Types of heaters(Contrsquodhelliphellip)

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Benefits of Heaterhellip

QUARTZ

bullInstant Heating bullCost efficient bullFar IR heating stimulates skin cellsbullLong lifebullDoes not crack

CERAM

IC

bullTakes time to heat bullNot cost effective bullNo Far IR heatingbullCracks in a period of time (ceramic coating)

TUBULAR

bullRisk of firebullHigh operating costbullDelay in heatingbullNo Far IR heating

Far IR Heating stimulates the cells of the baby by skin penetration and in turn helps growth of baby

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

A range of Baby friendly equipments 6 Layer security Introducing nice NeOumltech 2010 series Infant Radiant Warmers

Designed to perfection for your baby

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

nice 2010 series Infant Radiant Warmers

nice 2010 BC With bed and storage Cabinet Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display Cabinet for easy Storage

nice 2010 B With bed and storage tray Servo skin manual mode pre-warm and safe mode APGAR timer Large LED display

nice 2010 Servo skin manual mode pre-warm and safe mode APGAR Timer Standalone Height Adjustable Large LED display

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

bullMicro controller based electronicsbull6 Layer securitybullServo Skin Manual Pre-warm and Safe mode controlsbullCalibration Free temperature probebullHigh Intensity Quartz Heating element for effective radiationbullHalogen Observational LampbullParabolic reflector for uniform heatingbullMemory Backup restores previously entered data in case of power failurebullComprehensive alarms and safety cut offbullInstant warming and uniform heatingbullFar IR heater ensures instant warmingbullAutomatic safety heater cutoff at 39 degC at any mode bullApgar Timer

Features

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

How to operate

Control Panel Operation Operating Modes Alarms Heater Module Bed Platform Operation Side Panel Operation X-ray procedure

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Control Panel Operation

1 Skin Temperature Display 15

Manual ModeIndicator

2 Sensor Failure Indicator 16

Skin Mode Indicator

3 Heater Failure Indicator 17

Pre Warm Mode Indicator

4 System Failure Indicator 18

Increase Key

5 PowerFailure Indicator 19

Mute Key

6 Skin Temperature gt 39degC Indicator

20

Heater OutputIndicator

7 Lamp Indicator 21

Set TemperatureManualAPGAR Timer

8 Lock Key 22

APGAR Timer Indicator

9 Lock Indicator 23

Mute ON Indicator

10

Lamp Key 24

Manual Timer Indicator

11

APGAR Kay 25

Skin Temperature Low Indicator

12

Decrease Key 26

Skin Temperature High Indicator

13

Mode Key 27

Alarm Indicator

14

Safe Mode Indicator

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Operating Modes

Manual Mode Operation

1048729 The environment temperature in bed platform area is manually control by varying the percentage of heater output

1048729 Manual Timer Setting ndash Press Mode key and change into Manual Mode then Press Combination of Mute Key and increase ()Decrease () Key Set the Manual Timer (1 min ndash 15 min)

1048729 of Heater output Setting ndash Press increase ()Decrease () touch switch to adjust the of Heater Output in 10 increments

1048729 Manual timer alarm for 15 min will be on as an indication to monitor the patient temperature at a time the heater output is reduced

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Skin Mode Operation

Note The skin temperature probe must be properly attached before choosing Skin mode operation

1048729 The user must select a Skin control temperature (Set required Temperature) when the Infant Radiant Warmer is used in the Skin mode

1048729 How to set temperature in skin mode The Skin control temperature is adjusted by pressing the increase () and decrease () switches The temperature can be adjusted from 32degC to 38degC

APGAR Timer The APGAR display indicates the APGAR timer select in the APGAR Mode

Operating Modes(contrsquodhellip)

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Operating Modes(contrsquodhellip)

Pre-Warm Mode

bullUsed to heat the mattress surface before placing the baby thus eliminating heat loss by conductionbullHeater output at 100 for a period of 5 mins bullUniform heating of the mattress surface thereafter reducing the heater output to 30 (AFTER 5 MINS)

Safe Mode

bullServo controlled Heater output at 40 maximumbull Audio Alarm free modebullMaintains a safe ambient temperature when the baby is unattended

Note All the above mentioned modes can only be accessible when the keypad lock is off condition

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

AlarmsSkin Probe Failure Alarm Active in skin mode When probe is disconnected from console

Skin Temperature High Active in the skin mode of operation When infant temp greater than set temp(+1degC)

Skin Temperature LowActive in the skin mode of operation When infant temp less than set temp(-1degC)

Manual Timer Alarm The manual timer alarm activates in the manual mode of operation when the therapy time set has elapsed(15 mins by default)

Heater Fail Alarm The Heater failure alarm activates if the heater fails in any mode of operationSystem Failure Alarm The system failure alarm activates in any mode of operation if any components are defect

Power Failure Alarm The power failure alarm activates in any mode of operation if the power cable gets disconnected or when there is no AC main power

Thermal Cut-off Alarm When temp exceeds 39degC in any mode The heater deactivates and the patient temperature displays ldquootrdquo

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Heater Module

Parabolic design

Even Distribution of heat over bed surface

Extensive safety

measures(6 Layer safety for thermal

cut- off)

Far IR heating ensures

stimulation of cells for growth

The parabolic design of the heater module ensures that the Far IR radiant energy is distributed very evenly over the bed surface The IR rays by deep penetration in skin ensures the growth of cells by stimulating itWith 6 layer safety measures the chances of hypo and hyperthermia is ruled out The heater also comes with a thermostat temperature control for enhanced protection

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Bed Platform Operation

bullThe bed platform tilts for Trendelenburg and Fowler positioning capabilities Rotate the knob to adjust tilts

bull Caution Make sure that an addition support is provided to hold baby firmly when the bed is in tilted position

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Side Panel Operation

bullTo open and lower the side panel of the bed lift latch and swing the acrylic panel downbull To raise a side panel swing it to the upright Position and engage the latch tightly

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

X-ray Procedure

bullThe X-ray cassette may be placed in the slot under the bed platform or in the X-raycassette tray if installed

bullThe heater housing rotates to the left or right for X-ray procedures

bull The X-ray cassette tray facilitates X-ray procedures while patients occupy the Infant Radiant Warmer bed

bullAn X-ray cassette can be placed on the tray and slid into the cavity below the bed without disturbing the patient

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Comparison of nice 2010 Infant Radiant Warmers

nice 2010 BC nice 2010 B nice 2010

Skin Mode Y Y Y

Manual Mode Y Y Y

Safe Mode Y Y Y

Pre-warm Mode Y Y Y

X Ray Facility Y Y N

Bed Y Y N

Pivoting Warmer ModuleY Y Y

Apgar Timer Y Y Y

Height Adjustment N N Y

Optional Accessories Shelf O2 Flow meter O2 Driven slow suction Ambu bag Ceramic Heater O2 Cylinder Mounting Provision

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

How nice Infant Radiant Warmers fare over competitors

nice 2010 series warmers

User-friendly easy to access all parameters

Long life quartz heater has a long life compared to other heaters

6 Layer safety for the protection of infant from burning

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

6 Layer Safety Features

Alarms if infant temperature greater or less than Set Temperature

Thermal cut off if infant temperature exceeds 39 degC

Heater output reduces to 30 after 15 mins in Manual mode

Heater output reduces to 30 if electronics senses no variable feedback

Heater output reduces to 30 if there is a sudden decrease in temperature when the probe comes off from the baby

Thermostat controlled in an event of breakdown if temp exceeds permissible limits- component relay micro-controller failure

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

TROUBLESHOOTING

AND MAINTENANCE

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Cleaning Procedure

1 Clean the Skin Temperature Probe by gently wiping with a soft damp cloth containing detergent of disinfecting solutions

2 Clean the exterior of the Infant Radiant Warmer using mild detergent solution applied with a damp cloth or sponge Do not allow liquids to seep into the electrical housing

3 The mattress x-ray tray bed amp side panels may be cleaned by using disinfecting agent safe for use on the material

Cleaning Solutions for Mattress

Generic Formulations Maximum concentration level

Hydrogen Peroxide 6

Quaternary Compound 05 aqueous solution

Quaternary Ammonium 02

Lodophors 50mglitre

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Preventive Maintenance

1048705 Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repair1048729 Always disconnect power supply before any maintenance1048729 Use only nice Neotechrsquos original parts for maintenance

Warning Periodically check the insulation and the connection of the cable it may cause fire because of insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment

1048729 Side Panels visually inspect the panels every day for any sign of breakages or loose fit 1048729 Temperature measuring system Measuring system should be checked by trained service personnel with calibrated test probe every yearNote No need of calibration if a new temperature probe is replaced1048729 Bed UpDown Tilting To be inspected once in every three months by the trained technical personnel1048729 Inspection amp Maintenance Yearly by trained Service Personnel

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

TroubleshootingGeneral System FailureSystem Fault Infant Radiant Warmer

Sl No Problem Cause Remedy

1 Power failure LED is ON with audio alarm

Power FailureCheck the unit is Plugged in to Main supply

Check the mains are switched ON

Else Contact nice Neotech

2System Failure LED is ON with audio alarm

System Failure Contact nice Neotech

General System Failure

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

System Fault Infant Radiant WarmerSl No Problems Possible Causes and Remedy

1 No Display on the Control PanelCheck AC Supply Voltage

Check if Power Cord is disconnected

Check Fuse in the Rear side of Pillar

Contact nice Neotech

2 No Heater OutputCheck Link cord connected between Control Module toHeater Module

Set Temperature may be less than Actual Temperature

Check Heater Connection

Contact nice Neotech

3 Equipment Working But Observation Lamp amp Heater output not working

Check Link cord connected between Control Module toHeater Module

Check Fuse in the Rear side of Heater Module

Check Transformer Voltage - 12V AC

Replace Lamp

4 Equipment not give the Audio AlarmCheck the Temperature it may near to the Settemperature

Check Mute ON

Contact nice Neotech

5 Patient Temperature Display shows ldquoS-Ordquo

with Continuous Alarm(S-O Sensor Open)

Check if skin Probe is Disconnected from theEquipment

If skin probe is defective replace new probe

Contact nice Neotech

7 Display shows H - F (H-F Heater Failure)Check Link cord connected between Control Module toHeater Module

Check Heater Connection

Contact nice Neotech

8Display Show O ndash t(O-t Over Temperature ie Temp gt 39 deg C)

Check Environment Temperature above 39deg C

Check Infant Fever

9 Continuous Skin Temperature High AlarmCheck Patient Temperature is 1degC above Set

Temperature

Check Environment Temperature is 1degC above SetTemperature

Contact nice Neotech

10 Continuous Skin Temperature Low Alarm Check Patient Temperature is 1degC below Set

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Temperature

Probe may be wet

Contact nice Neotech

13 Temperature over Mattress does not riseCheck Low ambient Temperature it should be approximately (22-30degC)

Check draft around air conditioning outletCheck Percentage Heater Output is Set at low in ManualMode

Low Supply Voltage

14 Temperature over Mattress rises excessivelyCheck direct Sun light Heating Device or other HeatSource

Percentage Heater Output is Set at High in ManualMode

15 Temperature over Mattress dose not stabilizeCheck unstable ambient temperature (should be 22-30degC)

Check for Strong Air conditioning draft

16Temperature Display does not display InfantsTemperature accurately

Probe Thermal Sensitive Portion attached improperlyCheck Thermal Sensitive Portion of Probe covered improperly

17Baby skin Temperature not correlate with display Temperature

Check sensitive portion of the Probe is fixed with babyskin properly

Check Probe is fixed to the lower abdomen of baby

18 Heater output is off after some time in manual modeCheck Manual Timer Time Can be set 2 to 30 minutesCheck the Heater output is off after set timeIf Heater output is off before set time contact niceNeotech

19 In Skin Mode Baby getting over heat but equipment gives Low Temperature alarm

Check the probe is intact with lower abdomen of babyCheck the probe is removed and kept awayCheck the Set Temperature is highCheck the Set Heater output is high ( Manual Mode)

System Fault Infant Radiant Warmer(Contrsquodhellip)

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant

Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal

components Do not autoclave or gas sterilize the skin temperature probe Do not

immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe

Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment

should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech

Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair

and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the

equipment performance

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Certifications and Approvals

Electrical Safety IEC 60601-1 IEC 60601-1-4

Particular product IEC 60601-2-21

EMC Safety IEC 60601-1-2

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us

Contact Us

nice Neotech Medical Systems Private Limited

No 85 Krishna Industrial Estate

Mettukuppam Vannagaram

Chennai ndash 600095 Tamil Nadu

INDIA

Tel +91-44-2476259424764608

Fax +91-44-24766920

Web wwwniceneotechcom

E-mail infoniceneotechcom marketingniceneotechcom

Overseas Enquiry overseasniceneotechcom

Toll Free +91-1800-425-2594

  • nice NeOumltech Medical Systems Pvt Ltd
  • Presentation Flow
  • Company INFO
  • Timeline-nice Neotech
  • Why nice Neotech
  • Why Infant Radiant Warmers are Used
  • MODES OF HEAT LOSS
  • Slide 8
  • Effects of Far IR in Infant body
  • Types of heaters
  • Slide 11
  • Benefits of Heaterhellip
  • Slide 13
  • Slide 14
  • Slide 15
  • How to operate
  • Control Panel Operation
  • Operating Modes
  • Slide 19
  • Slide 20
  • Alarms
  • Heater Module
  • Bed Platform Operation
  • Side Panel Operation
  • X-ray Procedure
  • Comparison of nice 2010 Infant Radiant Warmers
  • How nice Infant Radiant Warmers fare over competitors
  • Slide 28
  • Slide 29
  • Cleaning Procedure
  • Preventive Maintenance
  • Troubleshooting
  • System Fault Infant Radiant Warmer
  • Slide 34
  • Slide 35
  • Certifications and Approvals
  • Contact Us