1004[1]
Transcript of 1004[1]
![Page 1: 1004[1]](https://reader031.fdocuments.in/reader031/viewer/2022021316/577d367b1a28ab3a6b933486/html5/thumbnails/1.jpg)
8/8/2019 1004[1]
http://slidepdf.com/reader/full/10041 1/7
100-1004 Application Note 1004
Application Note 1004
Understanding Noise Equivalent Power in RadiometricDetectors and Instruments
Sid Levingston
03/20/2007
![Page 2: 1004[1]](https://reader031.fdocuments.in/reader031/viewer/2022021316/577d367b1a28ab3a6b933486/html5/thumbnails/2.jpg)
8/8/2019 1004[1]
http://slidepdf.com/reader/full/10041 2/7
100-1004 Application Note 1004
Noise Equivalent Power, or NEP, is a basic indicator of detector performance. NEP is thenoise floor of a detector, normalized to a 1Hz bandwidth. NEP is expressed in Watts per
square root bandwidth. To derive NEP, two parameters must be measured: the detector responsivity at a specified frequency, and the detector voltage noise at the same
frequency. NEP can then be calculated as:
o f freqeuncy someat Rv
Hz Voltage Noise NEP ,=
If the electrical pole of the detector circuit shown in figure 1 is much less than the unity
gain bandwidth of the amplifier, the responsivity for the detector is given by:
( )
TauThermal ic Pyroelectr f C R
f
f
f
f
f
f
f R R
f Rv
therm
f f
pole
poletherm
therm
f i
=⋅⋅⋅
=
+⋅
+
⋅⋅
=
π 2
1
11
22
If we specify that the frequency of measurement is higher then the thermal frequency and
lower than the electrical pole, then the responsivity can be simplified to:
( )
( )2
21 f f
f i
C R f
R R f Rv
⋅⋅⋅⋅+
⋅
=π
RF
VoCDRD
en
Vjin
CF
Figure 1, Current Mode Radiometer with Noise Sources
![Page 3: 1004[1]](https://reader031.fdocuments.in/reader031/viewer/2022021316/577d367b1a28ab3a6b933486/html5/thumbnails/3.jpg)
8/8/2019 1004[1]
http://slidepdf.com/reader/full/10041 3/7
100-1004 Application Note 1004
The noise in the circuit has three sources we need to consider. There is the noise voltage
of the amplifier, the Johnson noise voltage of the feedback resistor, and the noiseresulting from the noise current of the amplifier flowing through the feedback resistor. A
fourth noise source is the loss tangent noise of the detector, but its contribution is much
smaller than the main three and can be ignored. Note that the noise voltage of theamplifier will be amplified by the noise gain of the amplifier. For a Pyroelectric detector,the detector shunt resistance is much larger than the feedback resistor so that the noise
gain is dominated by the ratio of the detector capacitance to the feedback capacitance.This is not always true for photodiodes, so care must be taken when deriving NEP for
those cases. The dominant noise sources for a Pyroelectric sensor, normalized to a 1Hz bandwidth are given by:
( )( )( )
( )Noise Amp
C R f
C C R f e f V
f f
D f f n
e2
2
21
21
⋅⋅⋅⋅+
+⋅⋅⋅⋅+⋅
=π
π
( )( )
Noise JohnsonC R f
RT K f V
f f
f
J 2
21
4
⋅⋅⋅⋅+
⋅⋅⋅
=π
( )( )
NoiseCurrent C R f
I q R f V
f f
bias f
i2
21
2
⋅⋅⋅⋅+
⋅⋅⋅
=π
The total noise will be the rms sum of these three sources:
( )( )( )( )
( )2222
21
2421
f f
bias f f D f f n
noise
C R f
I q R RT K C C R f e f V
⋅⋅⋅⋅+
⋅⋅⋅+⋅⋅⋅++⋅⋅⋅⋅+⋅
=π
π
If we divide this by the responsivity, the equation for NEP becomes:
( )( )( )( )
i f
bias f f D f f n
R R
I q R RT K C C R f e f NEP
⋅
⋅⋅⋅+⋅⋅⋅++⋅⋅⋅⋅+⋅
=2421
222π
Assuming a large feedback resistor and that the detector is in the flat bandwidth region,this will simplify further to:
![Page 4: 1004[1]](https://reader031.fdocuments.in/reader031/viewer/2022021316/577d367b1a28ab3a6b933486/html5/thumbnails/4.jpg)
8/8/2019 1004[1]
http://slidepdf.com/reader/full/10041 4/7
100-1004 Application Note 1004
( )( )
( )( )( )( )
( )
( )( )( )
i
bias
f
D f
f
n
i f
bias f f D f f n
D f f
R
I q R
T K C C f
R
e
f NEP
R R
I q R RT K C C R f e f NEP
C C R f
⋅⋅+⋅⋅
++⋅⋅⋅⋅
≈
⋅
⋅⋅⋅+⋅⋅⋅++⋅⋅⋅⋅⋅
≈
+⋅⋅⋅⋅<<
24
2
242
21
2
2
222
2
π
π
π
It can now be seen that for a given detector with a characteristic responsivity andcapacitance, reducing NEP can be accomplished by: increasing R f , decreasing I bias,
decreasing the amplifier noise, or decreasing Cf . Unfortunately, not all of these
parameters can be independently changed. Since the frequency response and amplifier stability of the circuit are controlled by the feedback capacitor and feedback resistor,neither can be changed without verifying that performance hasn’t been compromised. As
the feedback resistor is increased, the feedback capacitor must be decreased to maintain bandwidth. At some point the feedback capacitor can not be reduced due to circuit
parasitic capacitance. NEP improvement will stop as the I bias contribution starts todominate. The amplifier noise is also of no consequence for feedback resistors larger than
100K_, so reducing it has no benefit. The NEP for a typical detector is plotted in figure 2for increasing R feedback. NEP improvement is negligible for values bigger than 100G_.
NEP vs. R Feedback
0.1
1
10
100
1000
1.E+06 1.E+08 1.E+10 1.E+12 1.E+14
R Feedback
N E P
( n W / r o o t H z )
NEP
Figure 2, NEP vs. R Feedback
![Page 5: 1004[1]](https://reader031.fdocuments.in/reader031/viewer/2022021316/577d367b1a28ab3a6b933486/html5/thumbnails/5.jpg)
8/8/2019 1004[1]
http://slidepdf.com/reader/full/10041 5/7
100-1004 Application Note 1004
At Spectrum Detector our low noise detectors use a 100G_ feedback resistor and we takegreat care to ensure that we use the lowest input bias current amplifiers available. We test
each device to ensure it meets our standards. Spectrum Detector sells the SPH-42, anultra low NEP detector. The parameters for this detector are:
( )
K T and Hz f at
Hz
nW f NEP
W A R R pF C pF C fA I
Hz nV e i f f Dbiasn
2985
9.01055.
1081064.1101.8
55.1012.02225013
6
323138
11
==
=⋅
⋅+⋅+⋅
≈
======
−
−−−
µ
The measured NEP of a typical SPH-42 detector is 0.82nW/root (Hz). Note that the
amplifier noise (the first term) is several orders of magnitude less than the other sourcesand does not contribute to NEP. The theoretical lowest NEP for the detector is limited bythe Johnson noise of the feedback resistor and would be 0.74nW/root (Hz).
When deriving NEP for a photodiode the shunt resistance of the device comes into play.
This is because for a Germanium device, the shunt resistance can be on the order of 10K_. Silicon diodes have shunt resistances on the order of 100M_. Since the current
responsivity for these devices is one million times higher than a Pyroelectric, thefeedback resistors are proportionally smaller. This means the feedback capacitor is larger
to maintain a similar bandwidth. The resistor ratios now dominate the capacitor ratios.The dominate noise sources will be:
( )( )
Noise AmpC R f
e f V
f f
n
e2
21 ⋅⋅⋅⋅+=
π
( )( )
Noise JohnsonC R f
RT K f V
f f
f
J 2
21
4
⋅⋅⋅⋅+
⋅⋅⋅
=π
( ) ( ) NoiseCurrent C R f
I q R
f V f f
bias f
i2
21
2
⋅⋅⋅⋅+
⋅⋅⋅
=π
The total noise will be the rms sum of these three sources times the voltage gain provided by the feedback resistor and shunt resistor:
![Page 6: 1004[1]](https://reader031.fdocuments.in/reader031/viewer/2022021316/577d367b1a28ab3a6b933486/html5/thumbnails/6.jpg)
8/8/2019 1004[1]
http://slidepdf.com/reader/full/10041 6/7
100-1004 Application Note 1004
( )( )
+⋅
⋅⋅⋅⋅+
⋅⋅⋅+⋅⋅⋅+=
D
f
f f
bias f f n
noise R
R
C R f
I q R RT K e f V 1
21
24
2
22
π
For a Silicon device, the feedback to shunt resistor ratio is much less than 1, and thissimplifies to:
( )( )2
22
21
24
f f
bias f f n
noise
C R f
I q R RT K e f V
⋅⋅⋅⋅+
⋅⋅⋅+⋅⋅⋅+=
π
Rv will be the same as with the simplified Pyroelectric equation, so NEP for the twocases are:
Germanium for R
R
R
I q R
T K
R
e
NEP
Silicon for R
I q R
T K
R
e
NEP
D
f
i
bias
f f
n
i
bias
f f
n
+⋅
⋅⋅+
⋅⋅+
≈
⋅⋅+
⋅⋅+
≈
1
24
24
2
2
2
2
While it appears at first glance that we can increase the feedback resistor to improve the
Silicon device, there is a practical limit set by the detector maximum voltage output and bandwidth requirements. With the Germanium device, the shunt resistance is a device
parameter we cannot control, so it cannot be arbitrarily increased. Spectrum Detector sells the SSI-A-45 Silicon Detector Instrument. The parameters for this detector are:
K T and Hz f at
Hz
pW NEP
W
A R R R fA I
Hz
nV e i f Dbiasn
2985
26.05.0
1081064.11069.1
50.101025013
322628
68
==
=
⋅+⋅+⋅
≈
=====
−−−
![Page 7: 1004[1]](https://reader031.fdocuments.in/reader031/viewer/2022021316/577d367b1a28ab3a6b933486/html5/thumbnails/7.jpg)
8/8/2019 1004[1]
http://slidepdf.com/reader/full/10041 7/7
100-1004 Application Note 1004
The measured NEP of a typical SSI-A-45 Silicon Detector Instrument is 0.27pW/root(Hz).
Spectrum Detector sells the SGI-A-45 Germanium Detector Instrument. The parameters
for this detector are:
K T and Hz f at
Hz
pW NEP
W
A R R R fA I
Hz
nV e i f Dbiasn
2985
0.3105
101
9.0
1081064.11069.1
90.1010525013
4
6322628
64
==
=
⋅+⋅
⋅+⋅+⋅≈
==⋅===
−−−
The measured NEP of a typical SGI-A-45 Germanium detector is 2.6pW/root (Hz).
Understanding the sources of noise in Radiometric Detectors and Instruments provides a
standard of performance based on solid theoretical limits. Spectrum Detector, Inc. buildsand sells Radiometers that perform to those theoretical limits. This ensures that our
customers are getting the best device available.
updated 11/28/2007