Lumax 540Thoracic Impedance – Possible Parameter Settings

Option 1

Option 2

Option 3

Please note: As long as HM is OFF, there is no access to TI

Option 4

If HM is ON, TI can be enabled ordisabled individually

TI (HM) = ON and TI (statistics) = OFF means that the TI-values will only be transmitted via HM and not be storedInside the ICD.

TI (HM) = ON and TI (statistics) = ON means that the TI-values will be transmitted via HM and storedinside the ICD.(The interrogation time during next FU is significantly longer if TI (statistics) is ON)

Page 3: 0 © Dr. A. Kucher Technical Service Lumax 540 (VR-T, DR-T, HF-T, DX) Thoracic Impedance CRT Global Technical Services 08.12.2011.

3© Dr. A. KucherTechnical Service

Lumax 540Thoracic Impedance – Trend Curve via HM

Page 4: 0 © Dr. A. Kucher Technical Service Lumax 540 (VR-T, DR-T, HF-T, DX) Thoracic Impedance CRT Global Technical Services 08.12.2011.

4© Dr. A. KucherTechnical Service

Lumax 540Thoracic Impedance – Trend Curve stored in the ICD