Download - Instrument QC and Qualification. Overview Why QC is Important LSRII Optical Configuration LSRII QC LSRII Validation LSRII Optimization LSRII.

Transcript
Page 1: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Instrument QC and Qualification

Page 2: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Overview

Why QC is ImportantLSRII Optical ConfigurationLSRII QC

LSRII ValidationLSRII Optimization LSRII Calibration LSRII Standardization

Practice Analysis??

Page 3: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

3

Characterizing the Cytometer

Page 4: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Challenges…• Instrument - optical configuration,

optimization, standardization, and calibration

• Reagent - optimization and standardization

• Sample processing• Staining protocols• Data Analysis - compensation &

gating• Operators• Volume of data (death-by-excel!)

Duke University Medical Center

Page 5: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Two Methods of Instrument Charaterization

BD CS&T: Cytometer Setup and Tracking

Page 6: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Instrument Performance

• Parameters CS&T Perfetto• Laser (amps vs mW) yes yes• CV (resolution) yes (target) yes (range)• Signal-to-noise ratio NO yes (range)• Linearity Yes (range) yes (range)• Specific fluorescence NO yes (target)

• Automated YES NO

Page 7: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Instrument QC

Validation: optimal voltage range for each detector

Optimization: assay specific optimal voltage for each detector & assay specific target channels

Calibration: set detectors to assay specific target channel values for specimen acquisition

Standardization: Monitor trendlines using assay specific target channels over time

Initial Characterization & After Optical Service

Once Per Assay

Each Experiment &

Troubleshooting

Frequency Purpose

Page 8: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Key Performance Factors in High-Quality Flow Cytometry Data

• Resolution of subpopulations, including dim subpopulations• Sensitivity

• Relative measured values of fluorescence• Linearity and accuracy

• Reproducibility of results and cytometer performance• Tracking

• Comparison of results across time and amongst laboratories• Standardization

Page 9: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

LSRII qualification

Validation Linearity Resolution (CV) Signal-to-Noise Ratio (S:N) - LLOD & LLOQ

Optimization Select Optimal Voltages for Inst Performance with Assay-

Specific Reagents - reduce spillover (Specificity) Establish Assay-Specific Target Channels

Calibration (“Daily QC”) Set Daily Voltages Verify Voltages are within acceptable Range (P/F) Verify CVs are within acceptable Range (P/F) Set Target Channels for Specimen Acquisition (P/F)

Standardization - Reproducibility/Precision Record Daily Target Channel Values Record Daily Voltages Record Daily CVs Calculate Daily S:N Plot & Review Monthly Trendlines

Page 10: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Linearity

Definition: Proportionality of output to input

Method: Access linearity using the ratio of two pulses over voltage range of the PMT

Significance: Linearity is important for compensation

(Median Pk6 - Median Pk5)Median Pk5

= Constant

Page 11: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

• Defined as proportionality of output (MFI) to input (Fluorescence/ # of photons)

• Important for fluorescence compensation • Compensation of data in the last decade involves subtraction of

large numbers

• Small errors (non-linearity) in one or both large numbers can cause a large absolute error in the result

Linearity

0 1000 2000 3000

Ab = X Ab = 2XX Abs2X Abs =

1000 MFI2000 MFI

Page 12: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

73,000

365

D

1796

75

B

Linearity: Effect on Compensation

BD CompBeads stained with varying levels of FITC-Ab.Compensation was set using samples A and C.This cytometer had a 2% deviation from linearity above 50,000 units.

FITC

PE

A

68

80

Detector Median Fluorescence Intensity (MFI)

5921

79

C

• Compensation of data in the last decade involves subtraction of large numbers

• Errors (non-linearity) in one or both large numbers can cause a large absolute error in the result

Page 13: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

LSRII Qualification

Validation Linearity Resolution (CV) Signal-to-Noise Ratio (S:N) - LLOD & LLOQ

Optimization Select Optimal Voltages for Inst Performance with Assay-

Specific Reagents - reduce spillover (Specificity) Establish Assay-Specific Target Channels

Calibration (“Daily QC”) Set Daily Voltages Verify Voltages are within acceptable Range (P/F) Verify CVs are within acceptable Range (P/F) Set Target Channels for Specimen Acquisition (P/F)

Standardization - Reproducibility/Precision Record Daily Target Channel Values Record Daily Voltages Record Daily CVs Calculate Daily S:N Plot & Review Monthly Trend lines

Page 14: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

cvs

16

Page 15: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Instrument Sensitivity: Two Definitions

• Defining sensitivity 1. Threshold: Degree to which a flow cytometer can distinguish

particles dimly stained from a particle-free background. Usually used to distinguish populations on the basis of Molecules of Soluble Equivalent Fluorochrome (MESF).

2. Resolution: Degree to which a flow cytometer can distinguish unstained from dimly stained populations in a mixture.

• How to measure instrument-dependent sensitivity?• Resolution sensitivity is a function of three independent

instrument factors: • Br • Qr • Electronic noise (SDen)

Page 16: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Background contributions

18

Page 17: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

• High Br widens negative and dim populations. • High Br value = lower resolution • Low Br value = higher resolution

Why is Br important?

Low Br High Br

Page 18: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Br: Optical Background from Propidium Iodide

• Example: It is common to use propidium iodide (PI) to distinguish live from dead cells. Propidium iodide was added in increasing amounts to the buffer containing beads, and Qr and Br were estimated.

• Residual PI in your sample tube will increase Br, which will reduce sensitivity.

PerCP

0100020003000400050006000700080009000

0 1 2 3 4 5 6PI-free dye (µg)

Br

0

0.01

0.02

0.03

0.04

0.05

Qr

BrQr

Page 19: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

23

Page 20: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

PMT 1

PMT 2

Qr =

Qr =

Qr# fluorescence molecules

# photoelectrons=

= 0.252 photoelectrons

8 fluorescence molecules

= 0.1251 photoelectron

8 fluorescence molecules

Relative Q: Qr

• Qr is photoelectrons per fluorescence unit and indicates how bright a reagent will appear on the sample when measured in a specific detector.

Page 21: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

A system with a higher Qr has a better resolution than a system with a lower Qr.

Low Qr value = lower resolution

High Qr value = higher resolution

High Qr Low Qr

Why is Qr important?

Page 22: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

•Laser power

•Optical efficiency

•PMT sensitivity (red spectrum)•Poor PMT performance •Dirty flow cell•Dirty or degraded filter

What Factors Affect Qr?

Page 23: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Population resolution for a given fluorescence parameter is decreased by increased spread due to spillover from other fluorochromes.

Spillover Decreases Resolution Sensitivity

Spread from APC Cy-7 background

Page 24: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

29

Page 25: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

30

Page 26: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

31

Page 27: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

32

Page 28: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Methods Used for Validation

Which Beads to UseWhat Values to PlotSelection Criteria

Page 29: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Duke LSRII Optical Configuration

34

Green Laser

(532nm)

Duke LSRII

PE

Cy7

PE

Cy5

PE

TR

PE

Cy5.5

PE-

Green780/40

610/20

710/50

660/40

575/25

740LP640LP

600LP

690LP

Violet Laser

(407nm)

Duke LSR II

QDot

705

QDot

605

QDot

545

QDot

585

QDot

655

QDot

565705/70

585/42

660/40

605/40

560/40

670LP595LP

570LP

630LP

557LP

560/40

535LP

Am

Cyan

515/20

505LP

Cas

Blue45

0/50

Blue Laser

(488nm)

Duke LSRII

SSC

Blue

FITC

515/20

505LP

575/25

550LP

PE ミBlue

488LP

Red Laser

(635nm)

Duke LSRII

APC

Alexa

680

710/50

685LP

780/60

740LP

APC

Cy7

660/20

Duke University Medical Center

Page 30: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Beads Used for LSRII Validation

• 8 Peak Rainbow: 300-900v in 50v increments (all PMTs)• Non-fluorescent to Very Bright; Broad Spectrum Ex & Em• 8pks• Run: Once, then after optical service - “High” flow rate &

LOG scale• Uses: Validation

• Check Linearity: (MedianPk6-MedianPk5)/MedianPk5• Check CV: CV Pk5• Check S:N (LLOD & LLOQ): MedianPk5/MedianBlank

• Unstained Comp Beads: 300-900v in 50v increments (all PMTs)• Non-fluorescent• 1pk• Run: Once, then after optical service- “High” flow rate &

LOG scale• Uses: Validation

• Check S:N (LLOD & LLOQ): MedianPk5/MedianBlank

Page 31: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Example of Optimal PMT Performance

0 102 103 104 105

Green B-A

0

1000

2000

3000

4000

SS

C-A

0 102 103 104 105

Green B-A

0

1000

2000

3000

4000

SS

C-A

0 102 103 104 105

Green B-A

0

1000

2000

3000

4000

SS

C-A

0 102 103 104 105

Green B-A

0

1000

2000

3000

4000

SS

C-A

0 102 103 104 105

Green B-A

0

1000

2000

3000

4000

SS

C-A

0 102 103 104 105

Green B-A

0

1000

2000

3000

4000

SS

C-A

0 102 103 104 105

Green B-A

0

1000

2000

3000

4000

SS

C-A

0 102 103 104 105

Green B-A

0

1000

2000

3000

4000

SS

C-A

300 Volts 350 Volts 400 Volts 450 Volts 500 Volts

550 Volts 600 Volts 650 Volts 700 Volts 750 Volts

0 102 103 104 105

Green B-A

0

1000

2000

3000

4000

SS

C-A

0 102 103 104 105

Green B-A

0

1000

2000

3000

4000

SS

C-A

Page 32: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Criteria for the Selection of Voltage Ranges

Lowest CV possibleLowest Voltage possibleHighest MFI possibleLowest background possible

Page 33: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Example of Optimal PMT Performance & Voltage Selection Criteria (Green B)

GreenB Voltage vs. CV, LIN and Ratio

0

10

20

30

40

50

60

70

300 350 400 450 500 550 600 650 700 750

Voltage

CV

or

LIN

0

50

100

150

200

250

300

350

400

450

500

Rat

io

CV Lin Ratio

Optimal Voltages: 450-650

Selection Criteria:LinearLowest CVHighest S:NLowest Voltage

Linearity = (M2-M1)/M1Ratio = M1/B

Page 34: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Red A (APC-Cy7):Before (30Mar06) & After (10Apr06) Replacing PMT

RedA Voltage vs. CV, LIN and Ratio

0

20

40

60

80

100

120

140

160

180

200

300 350 400 450 500 550 600 650 700 750

VoltageC

V o

r L

IN

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rat

io

CV Lin Ratio

RedA Voltage vs. CV, LIN and Ratio

0

20

40

60

80

100

120

140

160

180

200

300 350 400 450 500 550 600 650 700 750

Voltage

CV

or

LIN

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rat

io

CV Lin Ratio

Optimal Voltages: 625-700Optimal Voltages: IndeterminateHigh CV’s; poor S:N

30 March 2006 10 April 2006

Page 35: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Faulty PMT on install…

0 102 103 104 105

Red A-A

0

1000

2000

3000

4000

SS

C-A

650 Volts

0 102 103 104 105

Red C-A

0

1000

2000

3000

4000

SS

C-A

650 Volts

Before After

Page 36: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

LSRII Qualification

Validation Linearity Resolution Signal-to-Noise Ratio (S:N) - LLOD & LLOQ

Optimization Select Optimal Voltages for Inst Performance with Assay-

Specific Reagents - reduce spillover (Specificity) Establish Assay-Specific Target Channels

Calibration (“Daily QC”) Set Daily Voltages Verify Voltages are within acceptable Range (P/F) Verify CVs are within acceptable Range (P/F) Set Target Channels for Specimen Acquisition (P/F)

Standardization - Reproducibility/Precision Record Daily Target Channel Values Record Daily Voltages Record Daily CVs Calculate Daily S:N Plot & Review Monthly Trend lines

Page 37: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Beads Used for LSRII Optimization

• STAINED Comp Beads: validated range in 50v increments (each PMT)

• Assay Specific fluorescence• 1pk• Run: Once, then after optical service- “High” flow rate & LOG scale• Uses: Optimization

• Optimize PRIMARY Fluorescence (Primary>Secondary)

• Unstained Comp Beads: optimized voltages (all PMTs)• Non-fluorescent• 1pk• Run: Once, then after optical service- “High” flow rate & LOG scale• Uses: Validation

• Check S:N (LLOD & LLOQ)

• 1x or Midrange Rainbow: optimized voltages (all PMTs)• Moderate fluorescence, near celllular expression; Broad Spectrum

Ex & Em• 1pk• Run: Once (after optimization) - “High” flow rate & LOG scale• Uses: Validation

• Establish Target Channels Linearity: medians = assay specific target channels

Page 38: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Criteria for the Selection Optimal PMT Voltages

The primary fluorescence should be the highest in the respective detector relative to all secondary detectors.

Page 39: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Green A: TNF PE-Cy7460v Baseline & Final

44

0 10 2 10 3 10 4 10 5

Blue B-A0 10 2 10 3 10 4 10 5

Green A-A

0 10 2 10 3 10 4 10 5

Green B-A

0 10 2 10 3 10 4 10 5

Green C-A

0 10 2 10 3 10 4 10 5

Red A-A

0 10 2 10 3 10 4 10 5

Green D-A

0 10 2 10 3 10 4 10 5

Green E-A

0 10 2 10 3 10 4 10 5

Red B-A

0 10 2 10 3 10 4 10 5

Red C-A

0 10 2 10 3 10 4 10 5

Violet A-A

0 10 2 10 3 10 4 10 5

Violet B-A

0 10 2 10 3 10 4 10 5

Violet C-A

0 10 2 10 3 10 4 10 5

Violet D-A

0 10 2 10 3 10 4 10 5

Violet E-A

0 10 2 10 3 10 4 10 5

Violet F-A

0 10 2 10 3 10 4 10 5

Violet G-A

0 10 2 10 3 10 4 10 5

Violet H-A

(Blue B-A): GeoMean = 19 (Green A-A): GeoMean = 5462

(Green B-A): GeoMean = 44.7

(Green C-A): GeoMean = 26.3

(Green D-A): GeoMean = 99.2

(Green E-A): GeoMean = 90.2

(Red A-A): GeoMean = 129

(Red B-A): GeoMean = 5.69

(Red C-A): GeoMean = 15

(Violet A-A): GeoMean = 8.05

(Violet B-A): GeoMean = 47

(Violet C-A): GeoMean = 12.3

(Violet D-A): GeoMean = 21.4

(Violet E-A): GeoMean = 7.82

(Violet F-A): GeoMean = 32.2

(Violet G-A): GeoMean = 9.43

(Violet H-A): GeoMean = 18.8TNFa PE-Cy7 DukeJY_460v baseline.fcs

Page 40: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

45

0 10 2 10 3 10 4 10 5

Blue B-A0 10 2 10 3 10 4 10 5

Green A-A

0 10 2 10 3 10 4 10 5

Green B-A

0 10 2 10 3 10 4 10 5

Green C-A

0 10 2 10 3 10 4 10 5

Red A-A

0 10 2 10 3 10 4 10 5

Green D-A

0 10 2 10 3 10 4 10 5

Green E-A

0 10 2 10 3 10 4 10 5

Red B-A

0 10 2 10 3 10 4 10 5

Red C-A

0 10 2 10 3 10 4 10 5

Violet A-A

0 10 2 10 3 10 4 10 5

Violet B-A

0 10 2 10 3 10 4 10 5

Violet C-A

0 10 2 10 3 10 4 10 5

Violet D-A

0 10 2 10 3 10 4 10 5

Violet E-A

0 10 2 10 3 10 4 10 5

Violet F-A

0 10 2 10 3 10 4 10 5

Violet G-A

0 10 2 10 3 10 4 10 5

Violet H-A

(Blue B-A): GeoMean = 14.1 (Green A-A): GeoMean = 617

(Green B-A): GeoMean = 986

(Green C-A): GeoMean = 165

(Green D-A): GeoMean = 3.39

(Green E-A): GeoMean = 0.98

(Red A-A): GeoMean = 295

(Red B-A): GeoMean = 870

(Red C-A): GeoMean = 97.4

(Violet A-A): GeoMean = 1890

(Violet B-A): GeoMean = 206

(Violet C-A): GeoMean = 6.55

(Violet D-A): GeoMean = 8.57

(Violet E-A): GeoMean = 3.85

(Violet F-A): GeoMean = 32.2

(Violet G-A): GeoMean = 9.64

(Violet H-A): GeoMean = 18.7CD8 PerCPCy55 DukeJY_450vbasline.fcsGreen B: CD8 PerCP-Cy5.5

450v Baseline

Page 41: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Green B: CD8 PerCP-Cy5.5550v Final46

0 10 2 10 3 10 4 10 5

Blue B-A0 10 2 10 3 10 4 10 5

Green A-A

0 10 2 10 3 10 4 10 5

Green B-A

0 10 2 10 3 10 4 10 5

Green C-A

0 10 2 10 3 10 4 10 5

Red A-A

0 10 2 10 3 10 4 10 5

Green D-A

0 10 2 10 3 10 4 10 5

Green E-A

0 10 2 10 3 10 4 10 5

Red B-A

0 10 2 10 3 10 4 10 5

Red C-A

0 10 2 10 3 10 4 10 5

Violet A-A

0 10 2 10 3 10 4 10 5

Violet B-A

0 10 2 10 3 10 4 10 5

Violet C-A

0 10 2 10 3 10 4 10 5

Violet D-A

0 10 2 10 3 10 4 10 5

Violet E-A

0 10 2 10 3 10 4 10 5

Violet F-A

0 10 2 10 3 10 4 10 5

Violet G-A

0 10 2 10 3 10 4 10 5

Violet H-A

(Blue B-A): GeoMean = 15.1 (Green A-A): GeoMean = 609

(Green B-A): GeoMean = 4343

(Green C-A): GeoMean = 442

(Green D-A): GeoMean = 18

(Green E-A): GeoMean = 10.7

(Red A-A): GeoMean = 549

(Red B-A): GeoMean = 1181

(Red C-A): GeoMean = 114

(Violet A-A): GeoMean = 3418

(Violet B-A): GeoMean = 206

(Violet C-A): GeoMean = 37.2

(Violet D-A): GeoMean = 26.3

(Violet E-A): GeoMean = 28.2

(Violet F-A): GeoMean = 274

(Violet G-A): GeoMean = 192

(Violet H-A): GeoMean = 91.4CD8 PerCPCy55 DukeJY_550v final.fcs

Page 42: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Green C: CD27 PE-Cy5350v Baseline47

0 10 2 10 3 10 4 10 5

Blue B-A0 10 2 10 3 10 4 10 5

Green A-A

0 10 2 10 3 10 4 10 5

Green B-A

0 10 2 10 3 10 4 10 5

Green C-A

0 10 2 10 3 10 4 10 5

Red A-A

0 10 2 10 3 10 4 10 5

Green D-A

0 10 2 10 3 10 4 10 5

Green E-A

0 10 2 10 3 10 4 10 5

Red B-A

0 10 2 10 3 10 4 10 5

Red C-A

0 10 2 10 3 10 4 10 5

Violet A-A

0 10 2 10 3 10 4 10 5

Violet B-A

0 10 2 10 3 10 4 10 5

Violet C-A

0 10 2 10 3 10 4 10 5

Violet D-A

0 10 2 10 3 10 4 10 5

Violet E-A

0 10 2 10 3 10 4 10 5

Violet F-A

0 10 2 10 3 10 4 10 5

Violet G-A

0 10 2 10 3 10 4 10 5

Violet H-A

(Blue B-A): GeoMean = 23.7 (Green A-A): GeoMean = 1497

(Green B-A): GeoMean = 6724

(Green C-A): GeoMean = 2747

(Green D-A): GeoMean = 305

(Green E-A): GeoMean = 284

(Red A-A): GeoMean = 368

(Red B-A): GeoMean = 1324

(Red C-A): GeoMean = 1663

(Violet A-A): GeoMean = 321

(Violet B-A): GeoMean = 267

(Violet C-A): GeoMean = 25.6

(Violet D-A): GeoMean = 53.2

(Violet E-A): GeoMean = 16.8

(Violet F-A): GeoMean = 39.1

(Violet G-A): GeoMean = 11.1

(Violet H-A): GeoMean = 19.4CD27 PE-Cy5 VRC_350v baseline.fcs

Page 43: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Green C: CD27 PE-Cy5450v Final48

0 10 2 10 3 10 4 10 5

Blue B-A0 10 2 10 3 10 4 10 5

Green A-A

0 10 2 10 3 10 4 10 5

Green B-A

0 10 2 10 3 10 4 10 5

Green C-A

0 10 2 10 3 10 4 10 5

Red A-A

0 10 2 10 3 10 4 10 5

Green D-A

0 10 2 10 3 10 4 10 5

Green E-A

0 10 2 10 3 10 4 10 5

Red B-A

0 10 2 10 3 10 4 10 5

Red C-A

0 10 2 10 3 10 4 10 5

Violet A-A

0 10 2 10 3 10 4 10 5

Violet B-A

0 10 2 10 3 10 4 10 5

Violet C-A

0 10 2 10 3 10 4 10 5

Violet D-A

0 10 2 10 3 10 4 10 5

Violet E-A

0 10 2 10 3 10 4 10 5

Violet F-A

0 10 2 10 3 10 4 10 5

Violet G-A

0 10 2 10 3 10 4 10 5

Violet H-A

(Blue B-A): GeoMean = 23.4 (Green A-A): GeoMean = 1507

(Green B-A): GeoMean = 13446

(Green C-A): GeoMean = 17138

(Green D-A): GeoMean = 1379

(Green E-A): GeoMean = 1689

(Red A-A): GeoMean = 709

(Red B-A): GeoMean = 1805

(Red C-A): GeoMean = 1904

(Violet A-A): GeoMean = 563

(Violet B-A): GeoMean = 252

(Violet C-A): GeoMean = 114

(Violet D-A): GeoMean = 129

(Violet E-A): GeoMean = 82.4

(Violet F-A): GeoMean = 306

(Violet G-A): GeoMean = 196

(Violet H-A): GeoMean = 91.5CD27 PE-Cy5 VRC_450v final.fcs

(Blue B-A): GeoMean = 33.7

Page 44: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Green D: CD45RO PE-TR430v Baseline & Final

49

0 10 2 10 3 10 4 10 5

Blue B-A0 10 2 10 3 10 4 10 5

Green A-A

0 10 2 10 3 10 4 10 5

Green B-A

0 10 2 10 3 10 4 10 5

Green C-A

0 10 2 10 3 10 4 10 5

Red A-A

0 10 2 10 3 10 4 10 5

Green D-A

0 10 2 10 3 10 4 10 5

Green E-A

0 10 2 10 3 10 4 10 5

Red B-A

0 10 2 10 3 10 4 10 5

Red C-A

0 10 2 10 3 10 4 10 5

Violet A-A

0 10 2 10 3 10 4 10 5

Violet B-A

0 10 2 10 3 10 4 10 5

Violet C-A

0 10 2 10 3 10 4 10 5

Violet D-A

0 10 2 10 3 10 4 10 5

Violet E-A

0 10 2 10 3 10 4 10 5

Violet F-A

0 10 2 10 3 10 4 10 5

Violet G-A

0 10 2 10 3 10 4 10 5

Violet H-A

(Blue B-A): GeoMean = 33.7 (Green A-A): GeoMean = 614

(Green B-A): GeoMean = 2966

(Green C-A): GeoMean = 4539

(Green D-A): GeoMean = 22260

(Green E-A): GeoMean = 252

(Red A-A): GeoMean = 6.68

(Red B-A): GeoMean = 24.6

(Red C-A): GeoMean = 42.5

(Violet A-A): GeoMean = 139

(Violet B-A): GeoMean = 229

(Violet C-A): GeoMean = 251

(Violet D-A): GeoMean = 50.7

(Violet E-A): GeoMean = 10.9

(Violet F-A): GeoMean = 34.7

(Violet G-A): GeoMean = 9.53

(Violet H-A): GeoMean = 19.1CD45RO PE-TR VRC_430v baeline.fcs

Page 45: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Green E: MIP1b PE350v Baseline & Final50

0 10 2 10 3 10 4 10 5

Blue B-A0 10 2 10 3 10 4 10 5

Green A-A

0 10 2 10 3 10 4 10 5

Green B-A

0 10 2 10 3 10 4 10 5

Green C-A

0 10 2 10 3 10 4 10 5

Red A-A

0 10 2 10 3 10 4 10 5

Green D-A

0 10 2 10 3 10 4 10 5

Green E-A

0 10 2 10 3 10 4 10 5

Red B-A

0 10 2 10 3 10 4 10 5

Red C-A

0 10 2 10 3 10 4 10 5

Violet A-A

0 10 2 10 3 10 4 10 5

Violet B-A

0 10 2 10 3 10 4 10 5

Violet C-A

0 10 2 10 3 10 4 10 5

Violet D-A

0 10 2 10 3 10 4 10 5

Violet E-A

0 10 2 10 3 10 4 10 5

Violet F-A

0 10 2 10 3 10 4 10 5

Violet G-A

0 10 2 10 3 10 4 10 5

Violet H-A

(Blue B-A): GeoMean = 44.3 (Green A-A): GeoMean = 269

(Green B-A): GeoMean = 1604

(Green C-A): GeoMean = 3200

(Green D-A): GeoMean = 22352

(Green E-A): GeoMean = 28006

(Red A-A): GeoMean = 3.41

(Red B-A): GeoMean = 8.53

(Red C-A): GeoMean = 24.4

(Violet A-A): GeoMean = 95.4

(Violet B-A): GeoMean = 222

(Violet C-A): GeoMean = 395

(Violet D-A): GeoMean = 925

(Violet E-A): GeoMean = 244

(Violet F-A): GeoMean = 90.3

(Violet G-A): GeoMean = 10

(Violet H-A): GeoMean = 18.6MIP1b PE DukeJY_450v baseline.fcs

(Blue B-A): GeoMean = 47.2 (Green A-A): GeoMean = 267 (Red A-A): GeoMean = 4.12 (Violet A-A): GeoMean = 175

Page 46: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Duke LSRII Optimization of

Specific Fluorescence Violet

Red

Green

Blue

Primary Fluorescence

SecondaryFluorescence

Page 47: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Neg Comp Beads52

0 10 2 10 3 10 4 10 5

Blue B-A0 10 2 10 3 10 4 10 5

Green A-A

0 10 2 10 3 10 4 10 5

Green B-A

0 10 2 10 3 10 4 10 5

Green C-A

0 10 2 10 3 10 4 10 5

Red A-A

0 10 2 10 3 10 4 10 5

Green D-A

0 10 2 10 3 10 4 10 5

Green E-A

0 10 2 10 3 10 4 10 5

Red B-A

0 10 2 10 3 10 4 10 5

Red C-A

0 10 2 10 3 10 4 10 5

Violet A-A

0 10 2 10 3 10 4 10 5

Violet B-A

0 10 2 10 3 10 4 10 5

Violet C-A

0 10 2 10 3 10 4 10 5

Violet D-A

0 10 2 10 3 10 4 10 5

Violet E-A

0 10 2 10 3 10 4 10 5

Violet F-A

0 10 2 10 3 10 4 10 5

Violet G-A

0 10 2 10 3 10 4 10 5

Violet H-A

(Blue B-A): GeoMean = 15.3 (Green A-A): GeoMean = 9.54

(Green B-A): GeoMean = 51.2

(Green C-A): GeoMean = 72.1

(Green D-A): GeoMean = 250

(Green E-A): GeoMean = 214

(Red A-A): GeoMean = 28.4

(Red B-A): GeoMean = 42.1

(Red C-A): GeoMean = 41.2

(Violet A-A): GeoMean = 10.6

(Violet B-A): GeoMean = 85.3

(Violet C-A): GeoMean = 38.8

(Violet D-A): GeoMean = 30.1

(Violet E-A): GeoMean = 29.5

(Violet F-A): GeoMean = 235

(Violet G-A): GeoMean = 178

(Violet H-A): GeoMean = 90.9Controls_Neg Comp.fcs

Page 48: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

LSRII Qualification

Validation Linearity Resolution Signal-to-Noise Ratio (S:N) - LLOD & LLOQ

Optimization Select Optimal Voltages for Inst Performance with Assay-

Specific Reagents - reduce spillover (Specificity) Establish Assay-Specific Target Channels

Calibration (“Daily QC”) Set Daily Voltages Verify Voltages are within acceptable Range (P/F) Verify CVs are within acceptable Range (P/F) Set Target Channels for Specimen Acquisition (P/F)

Standardization - Reproducibility/Precision Record Daily Target Channel Values Record Daily Voltages Record Daily CVs Calculate Daily S:N Plot & Review Monthly Trend lines

Page 49: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

1x(Target Channel Values) 54

0 10 2 10 3 10 4 10 5

Blue B-A0 10 2 10 3 10 4 10 5

Green A-A

0 10 2 10 3 10 4 10 5

Green B-A

0 10 2 10 3 10 4 10 5

Green C-A

0 10 2 10 3 10 4 10 5

Red A-A

0 10 2 10 3 10 4 10 5

Green D-A

0 10 2 10 3 10 4 10 5

Green E-A

0 10 2 10 3 10 4 10 5

Red B-A

0 10 2 10 3 10 4 10 5

Red C-A

0 10 2 10 3 10 4 10 5

Violet A-A

0 10 2 10 3 10 4 10 5

Violet B-A

0 10 2 10 3 10 4 10 5

Violet C-A

0 10 2 10 3 10 4 10 5

Violet D-A

0 10 2 10 3 10 4 10 5

Violet E-A

0 10 2 10 3 10 4 10 5

Violet F-A

0 10 2 10 3 10 4 10 5

Violet G-A

0 10 2 10 3 10 4 10 5

Violet H-A

(Blue B-A): GeoMean = 4587 (Green A-A): GeoMean = 1505

(Green B-A): GeoMean = 11590

(Green C-A): GeoMean = 22057

(Green D-A): GeoMean = 65203

(Green E-A): GeoMean = 30710

(Red A-A): GeoMean = 2172

(Red B-A): GeoMean = 3927

(Red C-A): GeoMean = 3934

(Violet A-A): GeoMean = 712

(Violet B-A): GeoMean = 970

(Violet C-A): GeoMean = 3283

(Violet D-A): GeoMean = 2539

(Violet E-A): GeoMean = 3052

(Violet F-A): GeoMean = 12196

(Violet G-A): GeoMean = 27119

(Violet H-A): GeoMean = 4157Controls_1x.fcs

Page 50: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Target Channels & Ranges

Page 51: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

LSRII Qualification

Validation Linearity Resolution Signal-to-Noise Ratio (S:N) - LLOD & LLOQ

Optimization Select Optimal Voltages for Inst Performance with Assay-

Specific Reagents - reduce spillover (Specificity) Establish Assay-Specific Target Channels

Calibration (“Daily QC”) Set Daily Voltages Verify Voltages are within acceptable Range (P/F) Verify CVs are within acceptable Range (P/F) Set Target Channels for Specimen Acquisition (P/F)

Standardization - Reproducibility/Precision Record Daily Target Channel Values Record Daily Voltages Record Daily CVs Calculate Daily S:N Plot & Review Monthly Trend lines

Duke University Medical Center

Page 52: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Daily Calibration• Mid-Range or “1x” Rainbow Beads: TC settings

• Moderate fluorescence, similar to cells; Broad Spectrum Ex & Em• 1pk• Run: Morning & Before each Exp - “High” flow rate & LOG scale• Uses: Calibration & Standardization

• Set Assay-Specific Target Channels (TCs)• Determine Daily Voltage Settings• Daily QC: P/F (CVs, voltages, trends)• LSRII Performance Verification for Exp Run: P/F (CVs & TCs)

• 8 Peak Rainbow & Unstained Comp Beads: TC Settings• Non-fluorescent to Very Bright; Broad Spectrum Ex & Em• 8pk and 1pk• Run: Once Daily - “High” flow rate & LOG scale• Uses: Standardization

• Check Linearity• Check S:N (LLOD & LLOQ)• Check Specificity• Check Resolution

• Ultra Rainbow Beads: Mean ch ~150,000• VERY Bright; Broad Spectrum Ex & Em• 1pk• Run: Once Daily - “Low” flow rate & linear scale• Uses: Manufacturer Recommended

• Daily QC: P/F (CV & voltages)• Check/Adjust Area Scaling• Check/Adjust Time Delay• Check/Adjust Window Extension

Duke University Medical Center

Page 53: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Daily Standardization

• Mid-Range or “1x” Rainbow Beads:• Record Daily For each PMT & Scatter

• CV• HV• Median

• Plot Monthly For each PMT & Scatter• CV• HV vs Median

• 8 Peak Rainbow & Unstained Comp Beads:• Record Daily For each PMT & Scatter

• Median• Neg Comp Beads• Peak 5• Peak 6

• Frequency• Peak 5• Peak 6

• CV: peak 5• HV

• Plot Monthly For each PMT & Scatter• Linearity: Medianpk6-Medianpk5/Medianpk5• Signal:Noise: Medianpk5/MedianNeg• Q & B???

• Ultra Rainbow Beads:• Record Daily For each PMT & Scatter

• CV• HV• Median

• Plot Monthly For each PMT & Scatter• CV• HV vs Median

Duke University Medical Center

Page 54: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

1x Blue B Trend line

59

1X Blue B (+/- 5% of Mean)

425

435

445

455

465

475

8/7/

2006

8/14

/200

6

8/21

/200

6

8/28

/200

6

9/4/

2006

9/11

/200

6

9/18

/200

6

9/25

/200

6

10/2

/200

6

10/9

/200

6

10/1

6/20

06

10/2

3/20

06

10/3

0/20

06

11/6

/200

6

11/1

3/20

06

Vo

ltag

e

0

1,000

2,000

3,000

4,000

5,000

6,000

7,000

8,000

9,000

10,000

Med

ian

1X Voltages

1X Medians

Duke University Medical Center

Page 55: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Ratio and Voltage Variation

0

100

300

500

700

1 2 3 4 5 6 7 8 9 10 11

B515 G610 R660 V585

0

500

1000

1500

2000

2500

3000

1 2 3 4 5 6 7 8 9 1 11

Days

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

Ratio CV

Duke University Medical Center

Page 56: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

LASERs have a limited lifespan…

Laser Performance

5.68

5.70

5.72

5.74

5.76

5.78

5.80

5.82

5.84

Date

Las

er C

urr

ent

(Am

ps)

14.80

15.00

15.20

15.40

15.60

15.80

16.00

La

se

r P

ow

er

(mW

att

s)

Laser CurrentLaser Power

Page 57: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Overview

Why QC is ImportantLSRII Optical ConfigurationLSRII QC

LSRII Validation LSRII Optimization LSRII Calibration LSRII Standardization

Practice Analysis??

Duke University Medical Center

Page 58: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

8 peak beads

0 10 2 10 3 10 4 10 5

0

200

400

600

800

1000

0 10 2 10 3 10 4 10 5

0

200

400

600

800

1000

0 10 2 10 3 10 4 10 5

0

300

600

900

1200

0 10 2 10 3 10 4 10 5

0

200

400

600

800

1000

0 10 2 10 3 10 4 10 5

0

200

400

600

800

0 10 2 10 3 10 4 10 50

200

400

600

# C

ells

Example: Stain Index, Qr, and Br Across LaboratoriesExample : FITC Channel

Lab SI Qr Br

1 77.4 0.015 233

2 134.5 0.028 216

3 55.6 0.015 976

4 80.5 0.01 298

5 66 0.01 613

6 13.7 0.007 2322

7 16.2 0.018 2768

Low SI = Bad Resolution; correlates with low Qr and high Br

High SI = Good Resolution; correlates with high Qr and low Br

Low SI = Bad Resolution; correlates with high Br

0 10 2 10 3 10 4 10 50

300

600

900

1200

Impact of Instrument Performance in Quality Assurance of Multicolor Flow Cytometry Assays.Jaimes MC,1 Stall A,1 Inokuma M,1 Hanley MB,1 Maino S,1 D’Souza MP,2 and Yan M.1 ISAC 20101BD Biosciences, San Jose, CA 95131; 2Division of AIDS, National Institute of Allergy and Infectious Diseases, National Institutes of Health, Bethesda, MD 20892

Page 59: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Plots gated on lymphocytes

Plots gated on CD3+

cells

Plots gated on CD3+CD8+

cells

Correlation of SI, Qr, and Br with Assay Performance

Lab 2 Lab 4 Lab 6 Lab 7

Impact of Instrument Performance in Quality Assurance of Multicolor Flow Cytometry Assays.Jaimes MC, Stall A, Inokuma M, et al. ISAC 2010

Page 60: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

65

BD FACSCanto II

1 Assay – 4 Platforms

BD LSR II

BD LSRFortessaBD FACSAria III

• CD4 V450 • CD8 PE• CD3 FITC• CD20 APC

Page 61: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Thank You!!

66

Page 62: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Defining Instrument Performance and Sensitivity: Qr, Br, and SDen

Page 63: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Qr: Anti-CD10 PE Example

Trans-mission

100% 83

35.5% 55

17.8% 35

7.1% 20

3.5% 14

SICorrected

The laser and detectors were attenuated by ND filters over a 30-fold range to illustrate the effects of decreasing detector sensitivity on population resolution.

Qr=0.087

Qr=0.227

Qr=0.038

Qr=0.014

Qr=0.007

Dim Population

Page 64: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

• Br is a measure of true optical background in the fluorescence detector, which helps indicate how easily (dim) signals may be resolved from unstained cells in that detector.

Unbound antibody or fluorochrome

Spectral overlap on a cell

Cell autofluorescence

Raman scatter

Scatter from the flow cell and ambient light.

Relative Background: Br

• Factors affecting Br: dirty flow cell, damaged optical component

Page 65: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

• High Br widens negative and dim populations. • High Qr value = lower resolution • Low Qr value = higher resolution

Why is Br important?

Low Br High Br

Page 66: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Br: Optical Background from Propidium Iodide

• Example: It is common to use propidium iodide (PI) to distinguish live from dead cells. Propidium iodide was added in increasing amounts to the buffer containing beads, and Qr and Br were estimated.

• Residual PI in your sample tube will increase Br, which will reduce sensitivity.

PerCP

0100020003000400050006000700080009000

0 1 2 3 4 5 6PI-free dye (µg)

Br

0

0.01

0.02

0.03

0.04

0.05

Qr

BrQr

Page 67: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Electronic Noise (SDen)

• SDen is the background signal due to electronics:• Contributed by:

• PMT connections / PMT noise• Cables too near power sources• Digital error

• Broadens the distribution of unstained or dim particles• Increases in electronic noise results in decreased

resolution sensitivity

Most important for channels with low cellular autofluorescence• APC-Cy™7, PE-Cy™7, PerCP-Cy™5.5

Cy™ is a trademark of Amersham Biosciences Corp. Cy™ dyes are subject to proprietary rights of Amersham Biosciences Corp and Carnegie Mellon University and are made and sold under license from Amersham Biosciences Corp only for research and in vitro diagnostic use. Any other use requires a commercial sublicense from Amersham Biosciences Corp, 800 Centennial Avenue, Piscataway, NJ 08855-1327, USA.

Page 68: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Summary: Instrument Performance and Sensitivity

• Qr and Br are independent variables, but both affect sensitivity.

• Increases in Br or decreases in Qr can reduce sensitivity and the ability to resolve dim populations.

• On digital instruments, BD FACSDiva™ software v6 and CS&T provides the capability to track performance data for all of these metrics, also allowing users to compare performance between instruments.

BrQr¥ ySensitivit relative

• Instrument performance can have a significant impact on the performance of an assay.

Page 69: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Resolution vs. Background

NegativePopulation

PositivePopulation

Negative population has low backgroundPopulations well resolvedNegative population has high backgroundPopulations not resolved

Negative population has low background high CV (spread)Populations not resolved

The ability to resolve populations is a function of both background and spread of the negative population.

Page 70: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

Measuring Sensitivity: The Stain Index

• The Stain Index is a measure of reagent performance on a specific cytometer, a normalized signal over background metric.

negative

negativepositive

rSD2

medianmedian

Negative ofWidth

Brightness Index Stain

Brightness

Width of negative

• The brightness is a function of the assay (antigen density, fluorochrome used).• The width of the negative is a function of:

– Instrument performance (Qr, Br, and SDen) [single-color]– The assay

(Fluorescence spillover / compensation) [multicolor] The cell population

Page 71: Instrument QC and Qualification. Overview  Why QC is Important  LSRII Optical Configuration  LSRII QC  LSRII Validation  LSRII Optimization  LSRII.

76

Gel (from coupling) found all over the flow cell????