Step 2. Pressure test

Schritt 2. Dichtigkeitsprüfung

Main menu (Hauptmenu (1)) select Data set (Messreihe angeben)

� (5) Loading

� PaulBMBF.P64 [sidenote-- include a (site measured at).P64]

� (3) Display table (Tabelle anzeigen)

-Enter in missing port number: with row F3 paste in, type in corresponding port

numbers, and switch it with the next row so the changes are assumed.

- Correctly choose calibration gas in the data set! E0, E1, and E2 for the small/short

measurement range, and E1, E2, E3 for the longer measurement range.

�In the correct numerical order according to the protocol, plug the glass bottles in the ports,

but DON’T open them.

� In main menu (haupmenü), select (2), Pressure test / dichtigkeitsprüfung.

� Press “enter” and affirm it with j.

� If it is sealed (it will show Dichtigkeit i.O.), continue to step three.

� If the pressure test is not i.O. check the marked bottles on the right side of port, if

necessary separately check the pressure on the vacuum pump of the o.a. valves.

� In main menu (haupmenü 2) start a new pressure test.