Step 2. Pressure test Schritt 2. Dichtigkeitsprüfung Main menu (Hauptmenu (1)) select Data set (Messreihe angeben) (5) Loading PaulBMBF.P64 [sidenote-- include a (site measured at).P64] (3) Display table (Tabelle anzeigen) -Enter in missing port number: with row F3 paste in, type in corresponding port numbers, and switch it with the next row so the changes are assumed. - Correctly choose calibration gas in the data set! E0, E1, and E2 for the small/short measurement range, and E1, E2, E3 for the longer measurement range. In the correct numerical order according to the protocol, plug the glass bottles in the ports, but DON’T open them. In main menu (haupmenü), select (2), Pressure test / dichtigkeitsprüfung. Press “enter” and affirm it with j. If it is sealed (it will show Dichtigkeit i.O.), continue to step three. If the pressure test is not i.O. check the marked bottles on the right side of port, if necessary separately check the pressure on the vacuum pump of the o.a. valves. In main menu (haupmenü 2) start a new pressure test.

Transcript of Gc step 2

Page 1: Gc step 2

Step 2. Pressure test

Schritt 2. Dichtigkeitsprüfung

Main menu (Hauptmenu (1)) select Data set (Messreihe angeben)

� (5) Loading

� PaulBMBF.P64 [sidenote-- include a (site measured at).P64]

� (3) Display table (Tabelle anzeigen)

-Enter in missing port number: with row F3 paste in, type in corresponding port

numbers, and switch it with the next row so the changes are assumed.

- Correctly choose calibration gas in the data set! E0, E1, and E2 for the small/short

measurement range, and E1, E2, E3 for the longer measurement range.

�In the correct numerical order according to the protocol, plug the glass bottles in the ports,

but DON’T open them.

� In main menu (haupmenü), select (2), Pressure test / dichtigkeitsprüfung.

� Press “enter” and affirm it with j.

� If it is sealed (it will show Dichtigkeit i.O.), continue to step three.

� If the pressure test is not i.O. check the marked bottles on the right side of port, if

necessary separately check the pressure on the vacuum pump of the o.a. valves.

� In main menu (haupmenü 2) start a new pressure test.