Step 2. Pressure test
Schritt 2. Dichtigkeitsprüfung
Main menu (Hauptmenu (1)) select Data set (Messreihe angeben)
� (5) Loading
� PaulBMBF.P64 [sidenote-- include a (site measured at).P64]
� (3) Display table (Tabelle anzeigen)
-Enter in missing port number: with row F3 paste in, type in corresponding port
numbers, and switch it with the next row so the changes are assumed.
- Correctly choose calibration gas in the data set! E0, E1, and E2 for the small/short
measurement range, and E1, E2, E3 for the longer measurement range.
�In the correct numerical order according to the protocol, plug the glass bottles in the ports,
but DON’T open them.
� In main menu (haupmenü), select (2), Pressure test / dichtigkeitsprüfung.
� Press “enter” and affirm it with j.
� If it is sealed (it will show Dichtigkeit i.O.), continue to step three.
� If the pressure test is not i.O. check the marked bottles on the right side of port, if
necessary separately check the pressure on the vacuum pump of the o.a. valves.
� In main menu (haupmenü 2) start a new pressure test.