COURIER 30XPHigh performance on-stream slurry
analyzer system
T E C H N O L O G Y T E C H N O L O G Y
Please contact us:
Australia Chile Mexico RussiaOutokumpu Technology Outokumpu Tcnica Outokumpu Mexicana ZAO MineralPty. Ltd. Chile Ltda. S.A. de C.V. Processing EngineersTel. (61) 8 9316 1966 Tel. (56) 2 274 4201 Tel. (52) 5 203 5968 Tel. (7) 812 324 8836Fax (61) 8 9316 1952 Fax (56) 2 223 4989 Fax (52) 5 203 1018 Fax (7) 812 327 8832
Brazil China Peru South AfricaOutokumpu Processos, Outokumpu Beijing Outokumpu Tcnica Outokumpu TechnologyEngenharia e Comrcio Ltda. Representative Office Per S.A.C. (Pty) Ltd.Tel. (55) 11 542 0099 Tel. (86) 10 6506 5951 Tel. (51) 1 221 0833 Tel. (27) 11 799 2410Fax (55) 11 542 9908 Fax (86) 10 6500 6512 Fax (51) 1 221 2633 Fax (27) 11 799 2449
Canada Japan Poland U.S.A.Outokumpu Technology Outokumpu Japan K.K. Outokumpu Poland Outokumpu Technology,Mississauga Tel. (81) 3 3661 6401 Sp. z o.o. Inc.Tel. (1) 905 671 3304 Fax (81) 3 3661 6421 Tel. (48) 22 825 8252 Tel. (1) 303 792 3110Fax (1) 905 671 8413 Fax (48) 22 825 1961 Fax (1) 303 799 6892
OUTOKUMPU MINTEC OYP.O. BOX 84, FIN-02201 ESPOO, FINLANDTEL. +358 9 4211, FAX +358 9 421 2530
HTTP://WWW.OUTOKUMPU.COM
0600E
Outokumpu has a long and broad experience in the recovery and refining of base and precious metals.The company has developed a comprehensive line of metallurgical processes and related advancedequipment. Outokumpu TankCell technology is used by major mining companies around the world.State of the art on-stream analyzers and automation systems have been used in Outokumpus concentratorsand refineries since late 1960s. The company is market and technology leader with more than 600 on-stream analyzer systems delivered to mineral processing and metallurgical applications.
C30XP_8_page _brochure_1099_3_pp.p65 15.6.2000, 15:522-3
keV 6 6,5 7 7,5 80
20
40
60
80
100
Mn
Fe
Co
keV 6 6,5 7 7,5 80
20
40
60
80
100
Mn
Fe
Co
S1
S3
S6
S4S2
S3
ROUGHERS SCAVENGERS
CLEANERS
S7
S1
PROCESS SAMPLING
MULTIPLEXING DIVISION
SCREENINGDE-AERATION
ANALYSIS
CALIBRATIONSAMPLE
FAST LOOP 100-300 L/MIN
FLOW CELL 20 L/MINMONITORING
ANDCONTROL
COMPOSITESAMPLESCOMPOSITESAMPLESCOMPOSITE
SAMPLES
CONTROLLER
OPERATOR STATION
COURIER 30 XP
SERIES 90-30
LOGIC POWER
OK CPU 731BUS
GE FanucMABLE
RUNBUS
PRO GRAM-
CONTROLLER LOCAL DISPLAY
24 AnalogOutputs
Optional Windows basedOperator Station collects,displays and reports assay anddiagnostic information from oneor several networked COURIER30XP analyzers. The optionalOutocal calibration softwaremakes the calibration procedurelogical and helps in keeping theanalyzer performanceoptimized.
PROSCON Lite controller withdistributed I/O and localdisplays can be connected withthe Operator Station to controlreagent feeders and other plantequipment.
PRIMARY SAMPLING
Well designed primary samplingarrangement is essential if theassays are to reflect the actual stateof a process. The primary samplerdirects a part of a process stream tothe multiplexer for secondarysampling. Gravity or processpressure transports the samplestream. Outokumpu has a varietyof primary sampler designs forpressurized flow from a pump,gravity flow in a pipe or launderand well-mixed vessel.
SECONDARY
SAMPLING
COURIER 30XP usesmultiplexers to remove air andtrash from the sample and toprovide a stable flow to themeasurement cell. Themultiplexers are organized tominimize the samplechangeover times in largesystems. While a sample of onemultiplexer is still beingmeasured, the next sample isalready waiting in the secondmultiplexer unit allowing fastmeasurement cycle for allstreams.
The swithing time between thesamples is used for the internalreference measurement. Themeasurement sequence is fullyprogrammable, so that criticalstreams can be measured morefrequently. Composite shiftsampling of all streams is astandard feature.
ANALYZER PROBE ANDCALIBRATION SAMPLER
The COURIER 30XP usesproven, simple and dependablewavelength dispersive, fixed-angle X-ray fluorescense(WDXRF) spectrometers. Thefactory positioned and tuned X-ray crystals maintain theabsolute stability. Fullyautomatic referencemeasurement compensates forany instrument drift.Temperature control of theinterior of a cabinet furthercontributes to the highperformance.
Automatic calibration sampler atthe outlet of the measurementcell ensures a reliablecalibration, since both theanalyzer and the laboratorymeasure exactly the samesample.
COURIER 30XP system
ANALYZER CONTROL
The analyzer allows a truly stand-alone operation: it does samplingcontrol, XRF measurement,multimodel assay calculation andbuffering of calibration X-ray data.The Probe Control Set is used foranalyzer operation control and todisplay assay results. The unit pro-vides an optional connection toOperator Station and/or an auto-mation system.The plant automa-tion system can be used to con-trol remotely sample bypass or animmediate measurement requestfunctions.
OPERATOR STATION
NLA launder sampler withmechanical cutter cleaner
PSA pressure pipe samplerfor pump outlet
SUPPORTOutokumpu providesconsulting, sampling systemdesign, installation,commissioning and trainingservices. The long term efficient
operation of the system isensured by a world widecustomer support networkproviding on-site serviceprograms, spare parts supply,training and technical assistance.
C30XP_8_page _brochure_1099_3_pp.p65 15.6.2000, 15:536-7
AnalyzerperformancePerformance depends on actualsample characteristics and howthe analyzer configuration isoptimized for a particularapplication
Measurement timeSelectable for each samplestream. Typically effectivemeasurement time is 15 - 30seconds per stream. Total cycletime for 24 streams is 12 - 15minutes.
Number ofsimultaneous assaysFive elements and solids contentfrom each sample.
Multimodel assaycalculationFor varying ore types theanalyzer assay calculation canbe programmed to automaticallyswitch to the best model.Criteria for switching thecalculation model is set by theuser. Outocal calibrationprogram supports multimodeldevelopment.
Measurableconcentration range0 ... 100%
Minimum detectionlimitFor most elements 3 ... 30 ppmin slurries, 1 .... 10 ppm insolutions
Repeatability
The standard deviation ofrepeated measurements ofconstant sample at 1% level isless than 0.01% absolute.
RepeatabilityShort term stability is 0.1%relative under the specifiedoperating conditions. Thereference measurementsautomatically compensate for along term drift.
ResolutionSuperior resolution of crystalspectrometers (down to 30 eV )allows measurement ofneighboring elements withminimal line overlap
Accuracy for slurriesFor single measurements withtypical measurement time, onesigma relative error is typically3 ... 6% for low concentrationsand 1 ... 4% for highconcentrations. The accuracy oflaboratory assays used forcomparison may significantlyaffect these figures.
Accuracy figures obtained bycomparing filtered measurementsor long term averages are betterin most cases.
Communicationconnection options
Operator StationFor transferring information toand from analyzer.
Third party DCSsystemSerial line connection, MODBUSRTU protocol, for readingprocess analysis results andanalyzer status.
PROSCON 2100automation systemStandard module for transferringinformation to and fromPROSCON 2100 system.
Other availableoptionsControlled on-demandsampling option for remotecontrol of sampler and sampleline washing.
Slurry flow speed reducers forprocess sample entering tosecondary sampling device.
Flexible installation for slurryapplications (secondarysampling system at the samelevel or lower than the analyzerprobe).
Briquette carousel for easymeasurement of ten briquetteset.
Outoclear sample handlingsystem for solution samples.
Closed loop water coolingsystem for sites where
continuous supply of coolingwater is not available.
Remote diagnostics
Installationrequirements
Dimensions andweightAnalyzer probe:
767(d) * 654(w) * 1534(h) mm
30,2"(d) * 25,7"(w) * 60,4"(h)
Weight 160 kg 352 lbs
Probe Control Set
385(d) * 600(w) * 820(h) mm
15,2"(d) * 23,6"(w) * 32.3" (h)
weight 50 kg, 110 lbs
EnvironmentalclassificationAnalyzer probe and probecontrol set IP 56 (NEMA 4)
Power supplySingle phase AC, 220V 10% at5A, 50/60 Hz or 110 V 10% at10 A, 50/60 Hz
No-break power system isrecommended for sites withfrequent power shortages
Cooling waterSand filtered raw water foranalyzer probe cooling.
Pressure 1 ... 6 bar (14.5 ... 87psi), temperature +5 ... +35 C(41 ... 95 F)
Consumption max. 8 l/min (2.3gal/min, depending ontemperature)
Flushing waterSand filtered raw water forflushing multiplexers
Pressure 2 ... 6 bar (29 ... 87 psi)
Consumption 50 l/min (14.2 gal/min) momentary, 30 l/min (8.5gal/min) average for twomultiplexers
Compressed airCompressed air for multiplexers,calibration sampler anddemultiplexer actuators
Pressure 5 ... 10 bar (72.5 ... 145psi)
Consumption 600 l/minmomentary, 16 l/min averagefor two multiplexers
Operating ambienttemperature+5 ... +40 C (+41...104 F)
Storage temperature-25 ... +60 C (-13 ... +140 F)
ISO 9001CERTIFIED
COURIER 30XPSPECIFICATIONS
Analysis method
Wavelength dispersive X-rayfluorescence spectrometry, X-raytube excitation, fixed crystalJohansson type spectrometers.
Element rangeFrom Titanium to Uranium(atomic numbers 22 to 92)
Measurement channelsUp to six fixed spectrometers
Number of samplestreamsUp to twenty-four samplesstreams per analyzer probe, 6per multiplexer
Slurry sample flowratesPrimary sample:
Recommended 80 ... 200 l/min(20 ... 57 gal/min)
Maximum 300l/min (85 gal/min)
Minimum 40 l/min (11gal/min)
Slurry measurementcell flow20 ... 30 l/min (6 ... 8 gal/min)
Slurry solids contentTypically 5 ... 45%, samplehandling sets high limit,analyzer performance set thelow limit