What went wrong in Gothenburg? or Why is it so hard to count platelets… Stefan Jacobsson överläkare Klinisk kemi, Sahlgrenska Universitetssjukhuset

Transcript of What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test...

Page 1: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

What went wrong in Gothenburg?or

Why is it so hard to count platelets…

Stefan Jacobsson

överläkareKlinisk kemi, Sahlgrenska Universitetssjukhuset

Page 2: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

…because they are so peculiar!

JCB9909028.V1.mov

Page 3: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Identification and quantification of intermediates in platelet release by immunofluorescence

microscopy.

Thon J N et al. J Cell Biol 2010;191:861-874

© 2010 Thon et al.

Page 4: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Direct visualization of proplatelet/platelet release in culture.

Thon J N et al. J Cell Biol 2010;191:861-874

© 2010 Thon et al.

Page 5: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Organization of the cytoskeleton in the mature platelet and proplatelet.

Hartwig J et al. J Thromb Haemost 2003;1:1580-1586© 2003 Hartwig et al.

Page 6: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Bending and branching of mouse proplatelets.

Hartwig J et al. J Thromb Haemost 2003;1:1580-1586© 2003 Hartwig et al.

Page 7: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

α-Granule delivery to the proplatelet tip.

Hartwig J et al. J Thromb Haemost 2003;1:1580-1586© 2003 Hartwig et al.

Page 8: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Even more strange in the disease state…

Page 9: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Platelets are so diverse…

• In the blood:

– Differ already as newborns

– Of different age

– Of different fate

• In the sample:

– Swell

– Partially activated

Heterogenous:

Size, form, content

Measuring characteristics

Page 10: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

So we should validate the platelet counting

performance with normal and pathological platelets

”Procedur för validering vid byte av metod, analys kit,

instrument etc” Id:AX0197 (Utför relevanta avsnitt).Datum

(utförande)

Bilaga nr Sign

Om resultaten inte finns som kopplade godkända pdf dokument i RMT+ så skall de skannas

och pdf kopplas till detta dokument

Selektivitet/specificitet

Arbetsområde (tekniskt intervall)

Interferenser

Carry over (smitta mellan prover/ mellan reagens)

Linearitet

Detektionsgräns/kvantiteringsgräns

Utbyte

Jämförelse patientprover

Inomserie variation *

Total variation *

Referensintervall (byte av) *

Backup (reserv) metod/instrument

2. Validering av metod

Page 11: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

OptSapph vs ADVIA, PLT, SU

Page 12: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

OptSapph vs ADVIA, PLT, ÖS

Page 13: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

OptSapph vs ADVIA, low PLT, SU

Page 14: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

OptSapph vs ADVIA, low PLT, SU

Page 15: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

OptSapph vs ADVIA, low PLT, SU

Page 16: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

OptSapph vs ADVIA, low PLT, ÖS

Page 17: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

OptSapph vs ADVIA, low PLT

Page 18: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

ImmCD61 vs ADVIA, low PLT

0

2

4

6

8

10

12

14

16

18

1 2 3 4

CD61

Optik S

Optik A

Page 19: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

ADVIA vs ADVIA, PLT

Analys diff i enheter mot ref. instr.

Välj referensinstrument eller mv från listan

-60,00

-40,00

-20,00

0,00

20,00

40,00

60,00

150,0 200,0 250,0 300,0 350,0 400,0 450,0

Analys och enhet

Anal

ys

diff

i en

hete

r

M1190 MIKROS S3609 S3611 S3612

Ö2701 Ö2702

Page 20: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

ADVIA vs ADVIA, PLT

TPK diff i enheter mot ref. instr.

Välj referensinstrument eller mv från listan

-40,00

-30,00

-20,00

-10,00

0,00

10,00

20,00

30,00

40,00

50 100 150 200 250 300 350 400

TPK 10E9/L

TP

K

dif

f i en

het

er

M1190

MIKROS

S3611

S3612

Ö2701

Page 21: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Carry over, PLT

Sysmex test XT-200i Operator's Guide

(L1 - L3) / (H3 - L3) (L1 - L3) / H1

H1 H2 H3 L1 L2 L3 % Carry over % Carry over

1045 1008 1039 12 13 14 -0,195% -0,191%

971 959 950 14 13 9 0,531% 0,515%

1063 1062 1059 16 7 8 0,761% 0,753%

876 877 881 21 18 18 0,348% 0,342%

697 702 685 21 19 15 0,896% 0,861%

542 547 531 11 10 10 0,192% 0,185%

Page 22: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

What we found, in short…

• Positive bias

• Significant carry over

• Difficulties in harmonizing instruments

Page 23: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

How did Siemens explain all this

• Manufacturing flaws

• Installation adjustments suboptimal

• Calibration suboptimal

• Calibrator bias

– Reference method deviations

• Optical alignment material malfunctioning

• Method adjustments necessary

Page 24: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Platelet count – yearly mean values

Preliminary results for 2010

-6

-4

-2

0

2

4

6

8

10

12

2005 2006 2007 2008 2009 2010

Me

an y

ear

ly d

evi

atio

n (%

) fro

m s

urv

ey

me

an

ADVIA 120 ADVIA 2120 ADVIA 2120i

Page 25: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Kalibrering, specifikation

Alla tilldelade värden för ADVIA SETpoint Hematology-kalibrator och ADVIA TESTpoint

Hematology-kontroller är baserade på NISTs, ICSHs eller NCCLSs rekommenderade

procedurer eller godkända referensprocedurer och material.

...

Referensräkningen för PLT utförs med hemocytometer och faskontrastmikroskop. Makro-

spädningen för PLT är 1:125 med 1 % ammoniumoxalat som spädningsvätska.

De systemspecifika värdena (SSV) har tilldelats för ADVIA 120 SETpoint Hematology-

kalibrator- och ADVIA 120 TESTpoint Hematologykontroll-produkterna genom att

analysera varje lotnummer i ett ADVIA 2120/2120i Hematology-system som kalibrerades

med färska, normala helblodsprov som tilldelats värdena RBC, WBC, PLT, HCT och

HGB med hjälp av den metod som beskrivs ovan.

ADVIA 120 OPTIpoint-värdena tilldelas med ett ADVIA 2120/2120i Hematology-system

efter att systemet har standardiserats med tre referensoljor.

Page 26: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Carry over, specifikation

Prestandaegenskaper: Carryover

Prov-carryover för ADVIA 120 Hematology-systemet uppmättes för WBC, RBC, HGB och

PLT. Prestandaspecifikationen för carryover är ≤1% för samtliga parametrar.

Carryover utvärderades i ADVIA 120 Hematology-systemet genom aspirering av en hög

pool förberedd för linjär testning, följt av 3 aspirationer av cellfri plasma.

Procentuell carryover beräknades som:

% carryover = [(L1 - L3) / H] x 100

L1 = räkningen för den första aspirationen av cellfri plasma

L3 = räkningen för den tredje aspirationen av cellfri plasma

H = räkningen från den höga poolen

Följande resultat erhölls:

Parameter Observerad carryover

WBC 0,4 %

RBC 0,2 %

HGB 0,2 %

PLT 0,1 %

Page 27: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Ongoing workMPV, PDV, pathological samples…

-2

-1,5

-1

-0,5

0

0,5

1

6,00 6,50 7,00 7,50 8,00 8,50 9,00 9,50 10,00

MPV (mv)

Dif

fere

ns C

elld

yn

- A

DV

IA

Page 28: What went wrong in Gothenburg? - Start | Equalis went wrong in gothenburg_s... · Sysmex test XT-200i Operator's Guide (L1 - L3) / (H3 ... har tilldelats för ADVIA 120 SETpoint Hematology-kalibrator-

Thank You for listening!

And thanks to all my coworkers at:

Sahlgrenska Universitetssjukhuset

SÄS, NÄL, Kungälvs sjukhus

Stefan Jacobsson