Understanding Container and Closure System Integrity … 1_Booth_pres.pdf · Understanding...
Transcript of Understanding Container and Closure System Integrity … 1_Booth_pres.pdf · Understanding...
Understanding Container
and Closure System
Integrity Testing
Crystal Booth, M.M.
Masters of Microbiology from North Carolina State University
Former Associate Director of Microbiology at Novartis
Understanding Container and Closure System Integrity
Testing
Application of the Assay
What is the test used for?
Regulations governing the test
Review some regulatory observations regarding Container and Closure System
Integrity Testing (CCIT)
Examining some of the various testing methods
Developing and Validating the Method
Developing a method
Performing the validation
Establishing acceptance criteria
Performing the test
Stability Testing
Interactive Exercise
Using an interactive game, participants review information pertaining to
monitoring microbial metrics.
Application of the Assay
What is the Test Used For?
Container closure systems should maintain the sterility
and product quality of sterile final pharmaceutical,
biological, and vaccine products throughout their shelf-
life.4
Contaminants may include microorganisms, reactive gases, and
other substances.13
Container closure systems consist of primary packaging
components and secondary packaging components.13
Primary packaging components are those components that come
into direct contact with the product, such as a glass vial or
syringe.
Secondary packaging components, according to USP <1207> are
those components that are vital to ensure correct package
assembly, such as aluminum caps over stoppers.13
What is the Test Used For?
What is Container Closure Integrity Testing (CCIT)?
CCIT is an assay that evaluates the adequacy of
container closure systems to maintain a sterile barrier.
http://www.accuratuslabs.com/wordpress/wp-content/uploads/2015/05/container-closure-integrity-testing-300x300.jpg
Some Regulations Governing the Test 21CFR211.94 Drug Product Containers and Closures2
(a) Drug product containers and closures shall not be reactive, additive, or absorptive so as to alter the safety, identity, strength, quality, or purity of the drug beyond the official or established requirements.
(b) Container closure systems shall provide adequate protection against foreseeable external factors in storage and use that can cause deterioration or contamination of the drug product.
(c) Drug product containers and closures shall be clean and, where indicated by the nature of the drug, sterilized and processed to remove pyrogenic properties to assure that they are suitable for their intended use. Such depyrogenation processes shall be validated.
(d) Standards or specifications, methods of testing, and, where indicated, methods of cleaning, sterilizing, and processing to remove pyrogenic properties shall be written and followed for drug product containers and closures.
Some Regulations Governing the Test
European Commission EudraLex- The Rules Governing
Medicinal Products in the European Union
Annex 1- Manufacture of Sterile Medicinal Products
• 117. “Containers should be closed by appropriately validated
methods. Containers closed by fusion, e.g. glass or plastic
ampoules should be subject to 100% integrity testing. Samples of
other containers should be checked for integrity according to
appropriate procedures.”1
• *Repeated in PIC/S PE 009-12 (Pharmaceutical Inspection
Convention)12
Volume 4- Part II Basic Requirements for Active Substances
used as Starting Materials
• 9.20. “Containers should provide adequate protection against
deterioration or contamination of the intermediate or API that
may occur during transportation and recommended storage.”3
Some Regulations Governing the Test ICH Harmonized Triplicated Guideline, Quality of Biotechnological
Products: Stability Testing of Biotechnological/Biological Products
Q5C10
Sterility testing or alternatives (e.g. container/closure integrity testing)
should be performed at a minimum initially and at the end of the
proposed shelf-life. 10
Other Guidance Material
Guidance for the Industry “Container and Closure System Integrity Testing in
Lieu of Sterility Testing as a Component of the Stability Protocol for Sterile
Products.9
USP <1207> Sterile Product Packaging- Integrity Evaluation, and the new
updated drafts <1207>13, <1207.1>14, <1207.2>15 and <1207.3>16 Series
PDA Technical Report 27
PDA White Paper: Container Closure Integrity Control versus Integrity Testing
during Routine Manufacturing4
Some Regulations Governing the Test
FDA Compliance Program Guidance Manual, Chapter 56-
Drug Quality Assurance Program 7356.002A- 09/11/155
(5) Verification of Container and Closures. The physical and
chemical characteristics of containers and closures can be critical
to the sterility and stability of the finished product. Many
containers and closures look alike (color and dimensions), but are
made of different materials or have a different surface treatment
such as silicone on stoppers and ammonium sulfate on Type I glass.
Evaluate the firm’s procedures for assuring containers and closure
consistently meet appropriate specifications.
Determine what tests and examinations are done to verify the
containers and closures are made of the correct materials with the
correct dimensions (critical to ensuring continuing container-
closure integrity) and are free of critical defects.
Some Regulations Governing the Test
FDA Compliance Program Guidance Manual, Chapter 56-Drug Quality Assurance Program 7356.002A- 09/11/155
(6) Container / Closure Integrity. The integrity of the container / closure system is critical to assuring that all units of drug products remain sterile through shipment, storage and use. Leaking containers or closures lead to product contamination.
Reference: FDA’s 1994 Guidance for Industry for the Submission of Sterilization Process Validation in Applications for Human and Veterinary Drug Products.
Evaluate the tests and studies performed to demonstrate the integrity of container / closure systems for all sterile drugs, including:
• Verify that all incoming container-closure components meet specifications, including all appropriate dimensions.
• Determine studies adequately simulate the stress conditions of the sterilization process, handling and storage.
• Verify that the units tested in validation are appropriate (e.g., for terminally sterilized drug product, the units selected should be exposed to the maximum sterilization cycles using the production process).
• Sensitivity of the test is specified.
• Container-closure integrity is demonstrated during validation and as part of the stability program (in lieu of sterility testing), over the shelf life of the product.
A Few Regulatory Observations
Regarding Container and Closure
System Integrity Testing
http://www.malacandra.me/assets/images/inspector_big_eye.png
A Few Regulatory Observations Regarding
Container and Closure System Integrity Testing
FDA Warning Letter Dated 24Jun09:
This is a repeat violation of the February 2007 inspection.
Your QCU failed to establish an adequate stability testing
program designed to evaluate the integrity of the container-
closure system. Specifically, SOP-QC-XXXX does not include
the storage orientation for liquid products. The stability
samples for liquid products should be stored in an upright or
inverted orientation in order to test and evaluate the integrity
of the bottle seal.
A Few Regulatory Observations Regarding
Container and Closure System Integrity Testing
FDA Warning Letter Dated 20Jan11:
Your firm failed to ensure your container closure system provided adequate protection against foreseeable external factors in storage and use that can cause deterioration or contamination of the drug product [21 C.F.R. § 211.94(b)].
• For example, your firm identified 542 incidences through consumer complaints of product defects such as, leaks, bursts, and premature activation during the period of January 2008 to August 2010. These are critical defects that can impact sterility and stability of your product. Your firm identified that the probable cause was the result of defective materials used in the manufacture of the container closure system.
• Your response is not adequate since the sampling plans described are not based on appropriate statistical criteria to sufficiently identify these known potential defects, especially given the history of the supplier for this container closure system.
• Furthermore, your final product inspection procedure and use of a (b)(4) does not appear to be effective in preventing shipments of product with critical defects to the marketplace. Additionally, our data indicates there may be other cases of foreign substances in products manufactured at your facility such as an insect found in the intravenous solution of XXXX and dirt reported inside of (b)(4) and the (b)(4) of XXXXX®. Please provide an evaluation of the suitability of any of these potentially affected lots.
A Few Regulatory Observations Regarding
Container and Closure System Integrity Testing
FDA Warning Letter Dated 31May13:
Your firm failed to ensure your container closure system
provided adequate protection against foreseeable external
factors in storage and use that can cause deterioration or
contamination of the drug product (21 CFR 211.94(b)).
For example, you received consumer complaints identifying at
least ten (10) membrane leaks and one hundred fifty-five
(155) inadequately-fitting blue caps during the period of
November 2011 to March 2013. These are critical defects that
can impact the sterility and stability of your products.
This is a repeat violation
A Few Regulatory Observations Regarding
Container and Closure System Integrity Testing
FDA Warning Letter Dated 02Nov15:
FDA investigators also noted CGMP violations at your facility,
causing your drug product(s) to be adulterated within the
meaning of section 501(a)(2)(B) of the FDCA. The violations
include, for example:
• Your firm failed to establish adequate written procedures designed to
assure batch uniformity and integrity of drug products that describe
the in-process controls, and tests, or examinations to be conducted on
appropriate samples of in-process materials of each batch (21 CFR
211.110(a)).
• Your firm failed to ensure container closure systems provide protection
against foreseeable external factors in storage and use that can cause
deterioration or contamination of the drug product (21 CFR 211.94(b)).
A Few Regulatory Observations Regarding
Container and Closure System Integrity Testing
FDA Warning Letter Dated 21APR16:
Puncturing a container compromises the integrity of the container closure
system, and each puncture increases the chances of contamination.
In response to our observation regarding (b)(4), you indicated that you
developed a container closure integrity test for the (b)(4) and plan to
demonstrate the sterility of the (b)(4) throughout its expiry. However, this
test only confirms the sampled portion is sterile and does not validate the
integrity of the container. Additionally, the tested portion may not have
detectable microbial contamination after the (b)(4) container loses
integrity, but the amount of bacterial endotoxin may still increase. Your
proposed corrective action is not adequate to demonstrate container
closure integrity. Therefore, your response does not indicate how you will
ensure your finished product is sterile and the amount of endotoxin is
within an acceptable limit after multiple punctures to the (b)(4)
containers.
Examining Some of the Various Testing Methods
Container closure integrity testing can be performed in
many different ways.
All methods have pros and cons.8
Some containers, such as ampoules, require 100% integrity
testing.4
New proposed changes to USP Chapter <1207> describes
several methods the user and experiment with.
The proposed update removes the requirement to
compare the results of other container closure integrity
(CCI) assays to the microbial ingress challenge.
The updated chapter series is expected to become official
in August 2016.
Examining Some of the Various Testing Methods
The chapter (proposed USP <1207>) divides the tests
into two major categories.
Deterministic- less subject to error and provide
quantitative results8
Probabilistic- more uncertainty in assay results,
includes traditional methods8
Testing methods included in the chapter were chosen
from peer reviewed articles and are supported by the
American Society for Testing and Materials (ASTM) 8
The user may choose from the methods listed or even
methods not listed as long as the final method is properly
validated and optimized for the closure system.8
Examining Some of the Various Testing Methods
Deterministic methods include:
Electrical Conductivity and Capacitance Test
(HVLD)
Laser-based Gas Headspace Analysis
Mass Extraction
Pressure Decay
Tracer Gas (vacuum mode)
Vacuum Decay
http://www.cincinnati-test.com/pressure-decay-test-systems.html
Examining Some of the Various Testing MethodsDeterministic
Leak Test
Technologies
Package Content
Requirements
Package
Requirements
Leak Size Class
(Detection Range)Measurement Outcome and Data Analysis
Effect of Method
on Package
Test Time
Order of
Magnitude
Electrical
conductivity and
capacitance
(high-voltage leak
detection)
Liquid (with no combustion
risk) must be more
electrically conductive
than package.
Product must be present
at leak site.
Less electrically
conductive than
liquid product.
3 to 6
Range varies with
product–package,
instrument, test
sample fixturing,
and method
parameters.
Quantitative measure of electrical current
passing through the test sample: provides an
indirect determination of leak presence as
shown by a drop in test sample electrical
resistivity, with a resultant increase in voltage
reading above a predetermined pass/fail limit.
Nondestructive,
although product
stability studies
are recommended
to verify effect of
exposure.
Seconds
Laser-based gas
headspace
analysis
Gas volume, path length,
and content must be
compatible with
instrument’s detection
capability.
Transparent or
translucent,
amber or
colorless.
1 to 6
Detection of
smallest leakage
rates is a function
of time span
between analyses.
Quantitative measure of gas headspace
content of the test sample by laser-based gas
analysis, for product requiring a headspace of
low oxygen concentration, low moisture vapor
concentration, and/or low absolute pressure.
To calculate the leak rate of the test sample,
one can use results compiled as a function of
time.
Nondestructive Seconds
Mass extraction Gas or liquid must be
present at leak site.
Presence of liquid at leak
site requires test
pressures below vapor
pressure.
Product must not clog leak
path.
Rigid, or flexible
with package
restraint
mechanism.
Porous package
requires masking
to limit gas flow
through porous
material.
3 to 6
Range varies with
product–package,
instrument, test
fixtures/chamber,
and method
parameters.
Quantitative measure of mass flow rate
resulting from test sample headspace escape
or liquid product volatilization within an
evacuated test chamber housing the test
sample.
Leak rate of the test sample may be estimated
by comparing the mass flow results of the test
sample to tests performed using leak rate
standards and positive controls.
Quantitative measure of pressure provides an
indication of larger leak presence.
Nondestructive Seconds to
minutes
Examining Some of the Various Testing MethodsDeterministic
Leak Test
Technologies
Package Content
Requirements
Package
Requirements
Leak Size Class
(Detection Range)Measurement Outcome and Data Analysis
Effect of Method
on Package
Test Time
Order of
Magnitude
Pressure
decay
Gas must be present
at leak site.
Product (especially
liquids or semi-
solids) must not
cover potential leak
sites.
Compatible with
pressure detection
mode.
Rigid, or flexible with
package restraint
mechanism.
3 to 6
Range varies with
product–package,
instrument, and
method parameters.
Quantitative measure of pressure drop within a
pressurized test sample; pressure drop results
from escape of gas through leak paths.
Results that are compiled as a function of time are
indicative of leakage rate of the whole test sample.
Nondestructive,
although means
used to access
test sample
interior may
compromise test
sample barrier.
Minutes to
days,
depending
on package
volume and
required leak
limit of
detection
Tracer gas
detection,
vacuum mode
Tracer gas must be
added to package.
Tracer gas must
have access to
package surfaces to
test package for
leaks.
Able to tolerate high-
vacuum test conditions
Rigid, or flexible with
package restraint
mechanism.
Limited tracer gas
permeability
1 to 4
Detection of larger
leaks is a function of
instrument capability
and test sample
fixturing.
Quantitative measure by spectroscopic analysis of
tracer gas leak rate while tracer gas is being
drawn out of a tracer-flooded test sample into a
vacuum chamber.
Leak rate of the total sample can be calculated by
normalizing the measured tracer leak rate by
tracer concentration in the test sample.
Nondestructive,
although access
to package interior
for tracer gas
introduction may
compromise test
sample barrier.
Seconds to
minutes
Vacuum decay Gas or liquid must
be present at leak
site.
Presence of liquid at
leak site requires
test pressures below
vapor pressure.
Product must not
clog leak path.
Rigid, or flexible with
package restraint
mechanism
Porous package
requires masking to
limit gas flow through
porous material.
3 to 6
Range varies with
product–package,
instrument, test
sample chamber,
and method
parameters.
Quantitative measure of pressure rise (vacuum
decay) within an evacuated test chamber housing
the test sample; vacuum decay results from
headspace escape from the test sample, or liquid
product volatilization.
Leak rate of the test sample may be estimated by
comparing vacuum decay results for the test
sample to results of tests performed using leak
rate standards and positive controls.
Nondestructive Seconds to
minutes
a The leak size class detection range cited for each technology allows for method-to-method comparison. The leak sizes actually detected by a fully developed and validated
deterministic method may deviate from these ranges, as noted.
Examining Some of the Various Testing Methods
Electrical Conductivity and Capacitance Test (also known as High Voltage Leak Detection HVLD)
Leak detection in walls of nonporous, rigid or flexible packaging containing liquid or semi-liquid product (e.g. ampoules) 15, 17
• High voltage/high frequency charge is applied across the container-closure system17
• A leak will cause an increase in current across the high voltage electrodes, triggering the reject mechanism for the leak detector17
Key factors include voltage level, probe positioning, container-closure system geometry and wall thickness, and product formulation17
Advantages Disadvantages
• Automation of the integrity test17
• Rapidity of the test procedure
(seconds) 17
• Can be used online17
• Nondestructive15
• Could potentially detect leak
locations15
• Quantitative Electrical Conductance
Measurements15
• Testing under normal atmosphere11
• High voltage/high voltage field could
potentially affect potency and stability
of some protein-based therapeutics. 17
• Requires conductive liquid fills11
Examining Some of the Various Testing Methods
Laser-based Gas Headspace Analysis Typically performed using non-contact methods, such as frequency
modulation spectroscopy17
• Near infrared diode laser light passes through the gas headspace region15
• Light is absorbed as a function of gas concentration and pressure15
• Absorption information is processed using phase-sensitive detection techniques15
• A microprocessor analyzes the data and yields the test results15
Can be used for lyophilized products or oxygen-sensitive liquid products17
Headspace gas analysis is rapid (seconds)—allows 100% inspection of oxygen sensitive products or products packaged under vacuum17
Nondestructive and provides quantitative results15
Key parameters: Headspace Volume, Package Temperature, Headspace Pressure and Vacuum, Sensitivity of the Headspace Analysis Instrumentation17
Examining Some of the Various Testing Methods
Mass Extraction Nondestructive and Quantitative15
Can be used for detecting leakage in nonporous, rigid or flexible packages15
• Packages with a porous component can be tested by masking the porous package component15
The test sample is placed inside a test chamber that is pneumatically connected to a mass extraction leak test system equipped with a vacuum generator package15
• The chamber is quickly evacuated for a predetermined time to reach a predetermined vacuum level. A series of evacuation cycles are performed, each intended to identify smaller leakage rates15
• After each cycle, the test system is isolated from the vacuum source and measurements of absolute pressure, pressure decay rate, and/or gas mass flow rate are captured15
• Readings greater than predetermined limits that were established using negative controls are indicative of container leakage, triggering test cycle abort15
• For those test samples passing all previous larger leak vacuum cycles, a final vacuum is drawn15
• With all flow from the test chamber directed through the mass flow sensor, the mass flow rate is measured. Mass flow above a predetermined limit established using negative controls is indicative of container leakage15
Advantages• Rapidity of the test
procedure 17
• Nondestructive15
• Quantitiative15
Disadvantages• Requires fixtures for
each container
closure system 17
• Proteinaceous
products may
interfere with defect
detection11
Examining Some of the Various Testing MethodsPressure Decay
Intended for integrity testing of the gas headspace region of the test sample (nonporous, rigid, or flexible packages) 17
• For this test, the container-closure system is placed in a test fixture that is either pressurized or evacuated17
• The test chamber is allowed to stabilize, and then the change in pressure or vacuum is measured over time17
• Pressure or vacuum can be measured directly or by differential pressure between the test chamber and a reference chamber17
Key test parameters include temperature, package geometry, test fixture geometry, volume of package headspace, water vapor pressure inside the package, stabilization time, and test time17
Advantages Disadvantages
• Rapidity of the test
procedure 17
• Can be used online17
• Nondestructive15
• Quantitiative15
• Requires fixtures for each
container closure system 17
• Proteinaceous products
may interfere with defect
detection11
Examining Some of the Various Testing Methods
Tracer Gas Detection (Vacuum Mode) Detects leakage from nonporous, rigid or flexible packages15
The test requires the presence of tracer gas inside the test sample package15
• Helium is the most commonly used tracer gas, and hydrogen is also used15
The leakage rate of tracer gas is quantitatively measured using a spectrometric analytical instrument specific for the tracer gas15
To perform the vacuum-mode test, test samples that have been fully or partially flooded with tracer gas are placed inside an evacuation chamber15
• The instrument's vacuum pump evacuates the test chamber or fixture, drawing any leaking tracer gas through the analyzer15
• The absolute leak rate of the test sample may be calculated by normalizing test results by the partial pressure of the tracer gas within the test sample at the time of test15
• Test sample leakage is judged acceptable if the absolute leak rate is below that which has been reported to put product quality at risk15
Advantages
• Rapidity of the test
procedure 17
• Nondestructive15
• Quantitiative15
• Sensitivity11
Disadvantages
• Requires helium-
containing
headspace (for
helium leak
detection) or
modified atmosphere
packaging (for
headspace oxygen
testing)11
Examining Some of the Various Testing Methods
Vacuum Decay
Nondestructive and quantitative15
Detects leaks in nonporous, rigid or flexible packages. Packages with a porous component can be tested by masking the porous package component. 15
• The test sample is placed in a closely fitting evacuation test chamber, which is equipped with an external vacuum source. 15
• The test chamber plus test system dead space are evacuated for a predetermined period of time. 15
• The targeted vacuum level chosen for the test is predetermined on the basis of the test sample type under evaluation. 15
• The rise in dead space pressure (i.e., vacuum decay) is monitored for a predetermined length of time using absolute and/or differential pressure transducers. 15
• A pressure increase that exceeds a predetermined pass/fail limit established using negative controls indicates container leakage. 15
Advantages• Rapidity of the test
procedure 17
• Nondestructive15
• Quantitiative15
Disadvantages• Requires fixtures for
each container
closure system 17
• Proteinaceous
products may
interfere with defect
detection11
Examining Some of the Various Testing Methods
Probabilistic Methods Include:
Microbial Challenge by Immersion
Tracer Liquid Tests (e.g. Dye Ingress)
Bubble Tests
Tracer Gas (Sniffer Mode)
Microbial Challenge by Immersion and Dye Ingress
are the most recognized leak test methods.
USP <1207> Series is encouraging a move toward
the deterministic methods.
Examining Some of the Various Testing Methods
Probabilistic
Leak Test
Technologies
Package Content
Requirements
Package
Requirements
Leak Size Class
(Detection Range)
Measurement Outcome and Data
Analysis
Effect of Method
on Package
Test Time
Order of
Magnitude
Bubble emission Gas must be present at
leak site.
Product (especially liquids
or semi-solids) must not
cover package surfaces to
be leak tested.
Rigid, or flexible
with package
restraint
mechanism.
4 to 6
Range varies with
product–package,
test sample fixturing
and positioning,
method parameters,
and analyst technique
and skill.
Qualitative measure by visual inspection of
bubble emission caused by escape of test
sample headspace while sample is
submerged and exposed to differential
pressure conditions. Alternatively, sample
surfaces may be exposed to surfactant.
Continuous bubble emission indicates
presence of a leak, its location, and its
relative size.
Destructive Minutes
Microbial
challenge,
immersion
exposure
Growth-supportive media
or product.
Presence of liquid at the
leak site required for
method reliability.
Able to tolerate
pressure and
immersion
challenge.
Rigid, or flexible
with package
restraint
mechanism.
4 to 6
Range varies with
container-closure,
test sample fixturing
and positioning,
challenge condition
severity, and inherent
biological variability.
Qualitative measure by visual inspection of
microorganism growth inside test samples
filled with growth-supportive media or
product, post immersion in heavily
contaminated challenge media while
exposed to differential pressure conditions,
followed by incubation to encourage
microbial growth.
Growth in the test sample indicates the
presence of test sample leak site(s)
capable of allowing passive or active entry
of microbes.
Destructive Weeks
Examining Some of the Various Testing MethodsProbabilistic
Leak Test
Technologies
Package Content
Requirements
Package
Requirements
Leak Size Class
(Detection Range)
Measurement Outcome and Data
Analysis
Effect of Method
on Package
Test Time
Order of
Magnitude
Tracer gas
detection, sniffer
mode
Tracer gas must be added
to package.
Tracer gas must have
access to package surfaces
to be tested for leaks.
Leak site
accessible to
probe.
Limited tracer gas
permeability.
3 to 6
Range varies with test
sample, method
parameters, test sample
fixturing, and analyst
technique and skill.
Smaller leak detection
may be possible under
optimum test conditions.
Quantitative measure by spectroscopic
analysis of tracer gas near the outer
surfaces of the tracer-flooded test
sample, sampled using a sniffer probe.
Tracer presence above a pass/fail limit
indicates presence and location of a leak
site.
Nondestructive,
although access to
package interior for
tracer gas
introduction may
compromise test
sample barrier.
Seconds to
minutes
Tracer liquid Contents must be
compatible with liquid tracer.
Product must not clog leak
path.
Rigid, or flexible
with package
restraint
mechanism.
Compatible with
liquid tracer
detection mode.
4 to 6
Able to tolerate liquid
immersion.
Smaller leak detection
may be possible under
optimal test conditions
employing chemical
analysis tracer detection.
Range varies with
container-closure, test
sample fixturing and
positioning, challenge
condition severity, and
tracer liquid content.
Measure of tracer in test sample
previously submerged in tracer-charged
liquid while exposed to differential
pressure conditions. Alternatively, tracer-
charged test samples may be submerged
in tracer-free collection fluid.
Tracer presence indicates leak site(s)
capable of allowing tracer passage, and
tracer magnitude may indicate relative
leak size (assuming a single-leak
pathway).
Tracer migration measurement may be
quantitative (by chemical analysis;
preferred approach for small leak
detection) or qualitative (by visual
inspection).
Destructive Minutes to
hours
a The leak size class detection range cited for each technology allows for method-to-method comparison. The leak sizes actually detected by a fully developed and validated deterministic method
may deviate from these ranges, as noted.
Examining Some of the Various Testing Methods
Microbial Challenge by Immersion
Suitable for any container-closure system that can withstand immersion and pressure changes.17
Test article is immersed in a broth containing the test organism. 17
• Brevundimonas diminuta, Serratia marcescens, Escherichia coli and other organisms have been used for this test. 17
This test may be static (no pressure or vacuum) or dynamic (pressure and vacuum are applied, typically to simulate air transport) 17
Key test factors include: bacterial size and motility, differential pressure, challenge media, exposure time, and viable count in the challenge media. 17
Advantages• Widely used for
decades11
• Industry and
regulatory
familiarity11
• Readily incorporated
into media fills11
Disadvantages• Less sensitive11
• Detection
probabilistic for
small-size defects
(i.e. <20 µm) 11
• Destructive11
Examining Some of the Various Testing Methods
Tracer Liquid Tests (e.g. Dye Ingress)
The dye leak test is the most common liquid tracer test.15
• The container is immersed in a methylene blue solution and pressure and vacuum are applied to the container. 15
• The container inspected visually or via spectrophotometry (preferred method). 15
Key factors include differential pressure, compatibility of the dye with the product, liquid viscosity and surface tension, inspector training and experience (for visual inspection) or assay sensitivity (for spectrophotometry). 15
Qualitative and destructive17
Advantages• Widely used for
decades11
• Industry and
regulatory
familiarity11
Disadvantages• Less sensitive11
• Detection
probabilistic for
small-size defects
(i.e. <20 µm) 11
• Destructive11
Examining Some of the Various Testing Methods
Bubble Tests
The item under test is pressurized to about 3 psig and
immersed in a bath containing water or water and
surfactant (e.g., polysorbate 80).15
• This test can detect leaks as small as 10-5 mbar-L/sec.15
Key test factors include differential pressure, test
time, immersion fluid surface tension, visible
inspection conditions (e.g., light intensity,
magnification, and background), and visual
inspector training and experience.15
Qualitative and destructive17
http://flexpakinc.com/gallery/
Examining Some of the Various Testing Methods
Tracer Gas (Sniffer Mode)
Test samples are flooded completely or partially with the tracer gas via one of several options. 17
• Piercing a closed test sample to introduce pressurized tracer gas (sealant is applied to close the puncture site) 17
• Flooding the test sample before package closure17
• “Soaking” a closed test sample by pressurizing with tracer gas (most applicable to larger leak detection) 17
Test samples are checked for leakage by scanning the outer package surfaces using a vacuum wand17
The use of negative and positive controls along with the test samples provides evidence of test method limit of detection.17
The sniffer mode is generally chosen when the leak location must be identified. 17
Developing and
Validating the Method
Developing a Method
Package integrity verification occurs during three product
life cycle phases:14
1. The development and validation of the product-
package system
2. Product manufacturing
3. Commercial product shelf-life stability assessments
Testing needs to show the complete picture of container
CCI over the lifecycle of the product.
Any leak test requires optimization for each product-
package application.13
Developing a Method
All methods have limitations, but the following aspects should be considered.11
Methods must be suitable for its intended use
Methods must be applicable to the specific drug product-package (e.g. drug products can interact with CCI defects)
Methods must detect leaks effectively
Non-destructive CCI testing
Choose a preliminary method following vendor’s recommendations or literature research.
Choice of method depends on specific desired outcomes14
1. Detecting the presence of leak paths
2. Determining the location of leak paths
3. Measuring leak rate for the whole package
4. Evaluating the potential for microbial ingress
Other Criteria• Liquid or solid product?
• Headspace gas?
• Headspace vacuum?
• No Headspace?
• Package design?
• Rigid or flexible package?
• Moveable components?
• Polymeric materials?
• Metallic materials?
• Transparent or Opaque
materials?
• Porous materials?
• Multi-dose systems?
Developing a Method
Positive and negative controls need to be created for the assay.
Positive controls are needed to simulate defects.
• Leaks that occur naturally are rarely uniform holes or channels. They are generally complex tortuous paths.13
The controls will be tested along with intact samples under different testing parameters. 11
Negative and positive controls are designed and assembled appropriately with consideration of container–closure design, materials of construction, characteristics of anticipated package leaks, and impact of product contents on test results.14
Developing a Method
Many methods are available to create positive controls
Positive Control Type Advantages Disadvantages
Micro-pipettes11
Easy sample
preparation
Fragile
Broken tips may not be
easily detected
Laser drilled
holes11
Sample geometry can
remain unchanged
Better resemble natural
defects in glass (cracks)
and polymer (pinholes)
Cost
“Hole” size of laser-
drilled effects needs to
be calibrated
Microtubes11
Easy sample
preparation
Robust
Easy to use
The length of the
microtube defects
usually longer than that
of the typical “real-world”
defects
Wire
Easy to prepare
Easy to use
Can create a tenting
effect where the size of
the control is not
accurate
Developing a Method
In addition to choosing a preliminary method, creating
positive and negative control samples, acceptance
criteria must be predetermined.
• All negative controls must pass
• All positive controls with leaks at or above the claimed
limit of detection must fail
• A lower limit of detection must be established
• An upper limit of detection should be established
http://www.rxpax.com/training-services/
Developing a Method
USP <1207.1> and ICH Q2 (R1) describes how to determine and establish a limit of detection (LOD) for the method chosen. 14
The LOD is the smallest leakage rate or leak size that the method can reliably detect, given the product-package of interest.14
During method development trials all negative controls must pass and all positive controls with leaks at or above the claimed limit of detection must fail. 14
The LOD for a given method is defined as the smallest-leak positive control subset that consistently demonstrates leakage in 100% of the positive-control subset units at that defect size and larger. 14
• The percentage of positive controls to be detected must be specified in the validation study protocol by the end user. 14
Multiple methods are available to determine limit of detection criteria.
Developing a Method
USP <1207.1> also describes the need to verify the
largest leak detection capability or upper limit of
detection.
All analytical methods have optimum measurement ranges. 14
All leak test method selection should also take into account the largest
leak sizes likely to occur in the sample population. 14
Once an optimized method is created, multiple lots
should be tested representing the package integrity at
the extremes of finished product-package profiles. 14
e.g. relevant product variations, including various packaging component
sources/lots, drug products batches, as well as packaging sites and
lines.11
The quantities to be tested should be sufficient to provide adequate
assurance of package integrity. 14
Developing a Method
At this point
A method has been chosen
Target acceptance criteria has been established
Positive and negative controls have been created
Leak test method parameters have been developed
and optimized through experimentation to ensure
sensitive, accurate, robust, and reproducible leak
detection for specific product–package systems.14
And successful qualification(s) have been performed
A validation protocol should be written outlining
the successful trials and parameters created
during the method develop phase.
Performing the Validation
Container closure integrity methods need to be validated for
specific drug-product packages
Various components, such as the drug product, can affect the
testing outcome.
Validation of the final leak test method is required to
demonstrate test method detection limit, accuracy, range,
robustness, and precision.11, 13
Method accuracy is the ability of a leak test to correctly identify or size
leaks (as per the intended leak test outcome).14
Method precision is a measure of test result reproducibility and is proven
during method validation by testing a randomly mixed population of
negative and positive controls over multiple days by multiple operators
and, when possible, using multiple test instruments.14
USP <1225> Validation of Compendial Procedures and ICH Q2(R1)
provide good guidance on method validations
Performing the Validation
The established validation protocol should be followed when
performing the validation.
Validations are typically performed in triplicate by performing
the method created in the method development phase.
USP <1207.1> recommends randomly mixing negative and positive
control populations over multiple days by multiple operators and, when
possible, using multiple test instruments.14
Quantities of samples to be tested must be sufficient to
provide adequate assurance of package integrity and will likely
vary on the basis of:
1. The complexity of the product–package 14
2. The specifics of the user specification requirements14
3. The prior experience of the producer. 14
Performing the Validation
The method validation protocol should be written to describe introducing a number of defects of known size or leak rate as controls
Acceptance criteria for leak test method validation should include the following:
1. All negative controls pass (no leaks are identified) 14
2. All positive controls with leaks at or above the designated limit of detection fail (leaks are detected). 14
3. An integral package is one that conforms to specific product-package maximum allowable leakage limits.13
For some test methods (e.g. liquid tracer leak detection by mass spectrophotometric analysis), test blanks are also included as part of method validation and routine testing. 14
Blanks are not equivalent to, and should not substitute for, negative controls. 14
New USP <1207>
series removes the
requirement to
compare to microbial
ingress
Performing the Validation
After the successful validation, a report should be written
describing the acceptable parameters, acceptance criteria,
validation results, and the method to be used for routine
testing and stability testing.
Establishing Acceptance Criteria Acceptance criteria was established in the method development
phase.
A lower limit of detection was established and demonstrated
An upper limit of detection was established and
demonstrated
Positive and negative controls were created
• All positive controls must fail
• All negative controls must pass
Leak test method parameters were developed and optimized
through experimentation (e.g. amount of vacuum to pull)
Units that test to be non-integral need to be investigated
Performing the Test After the method is develop and validated, a routine testing SOP
(standard operating procedure or test method) should be
established.
The routine SOP should lock down all the parameters and
acceptance criteria for the specific product/container-closure
system so that the test is performed the same way every time.
“The manufacturer should be able to justify the amount of testing
required on the basis of statistical process control results generated
during the validation phase, and later, on the basis of routine
manufacturing product-quality trending analyses.”14
Quantities of samples to be tested must be sufficient to provide adequate
assurance of package integrity.14
Package integrity should be re-evaluated when changes are
required in package design, package materials, or
manufacturing/processing conditions.14
Stability Testing
The routine testing SOP should be utilized when performing
CCIT for stability testing.
Container-closure integrity should be demonstrated as part of
the stability program over the shelf life of the product for
new and existing products.5
Sterility testing or alternatives (e.g. container/closure
integrity testing) should be performed at a minimum
initially and at the end of the proposed shelf-life. 10
Container-closure integrity testing can be used to replace
sterility testing in stability protocols.4, 9
CCIT may not replace the sterility test for release
testing.4, 9
Stability Testing
If a non-destructive test has been validated for the specific
container-closure system, it is useful during stability studies.
The same container can be used throughout the stability
period.
This saves money and allows for more meaningful profiles
of container-closure integrity.17
Where the same strength and exact container/closure system
is used for 3 or more fill contents, the manufacturer may
elect to place only the smallest and largest container size into
the stability program, i.e., bracketing.10
Summary Container Closure Integrity Testing evaluates the adequacy of
container closure barrier systems to maintain a sterile barrier.
Some guidance documents and regulations that discuss CCIT include
FDA 21CFR211.94
European Commission EudraLex Annex 1 and Volume 4- Part II
ICH Q5CC
Guidance for the Industry “Container and Closure System Integrity Testing in
Lieu of Sterility Testing as a Component of the Stability Protocol for Sterile
Products”
USP <1207>
PDA Technical Report 27
PDA White Paper- “Container Closure Integrity Control versus Integrity Testing
during Routing Manufacturing”
FDA Compliance Program Guidance Manual, Chapter 56- Drug Quality
Assurance Program 7356.002A
Summary Container Closure Integrity Testing is currently reviewed by
regulators
Many different methods are available to perform CCIT.
The proposed USP <1207> update divides the methods into deterministic
and probabilistic.
There is a move to move toward more deterministic methods.
The USP <1207> revised guidance series is expected to become
effective in August 2016 and removes the recommendation to
compare CCIT methods to microbial ingress testing
Methods described in the proposed USP <1207> update include:
Electrical Conductivity and Capacity Test (HVLD)
Mass Extraction
Pressure Decay
Tracer Gas (vacuum mode)
Vacuum Decay
Microbial Challenge by Immersion
Tracer Liquid Tests (e.g. Dye Ingress)
Bubble Tests Tracer Gas (Sniffer
Mode)
Summary
Package integrity verification occurs during
three product life cycle phases:14
1. The development and validation of the product-package
system
2. Product manufacturing
3. Commercial product shelf-life stability assessments
CCIT needs to show container integrity over the
life cycle of the product.
CCIT methods require optimization for each
product-container/closure application.
All methods have limitations
Summary
During method development, a preliminary method is
chosen, optimized, and qualified.
Target acceptance criteria will be established
Positive and negative controls will be created
Leak test method parameters will be developed and
optimized through experimentation.14
Qualification(s) trials will be performed
Quantities for testing should be sufficient to provide
adequate assurance of package integrity. 14
Summary Container closure integrity needs to be validated for
each specific drug-product package.
Validations are typically performed in triplicate.
USP <1207.1> recommends to randomly mix negative and positive control populations over multiple days, with multiple operators, and if possible, using multiple test instruments.
Acceptance criteria should include the following:
1. All negative controls pass (no leaks are identified) 14
2. All positive controls with leaks at or above the designated limit of detection fail (leaks are detected). 14
3. An integral package is one that conforms to specific product-package maximum allowable leakage limits.13
4. Upper and lower limits of detection should be established as well as any key testing parameters.
Summary
Routine and stability testing should be performed per
an approved standard operating procedure or
method.
Package integrity should be re-evaluated when
changes are required in package design, package
materials, or manufacturing/processing conditions.14
CCIT can be used in place of sterility testing during
stability studies.
Non-destructive CCIT methods save product and
money.
Bracketing strategies can be utilized during stability
studies when appropriate.10
Using an interactive game, participants review information pertaining to
monitoring microbial metrics.
BINGO1. Container closure systems should maintain the ___________and
product quality of sterile finished products.
Sterility
2. All methods have pros and __________.
Cons
3. Some container, such as ________, require 100% integrity testing.
Ampoules
4. The proposed USP <1207> revisions remove the requirement to
compare CCIT results to the ___________ingress challenge.
Microbial
5. The proposed USP <1207> divides the test into _______ major
categories.
Two
BINGO6. _____________ assays are less prone to error.
Deterministic
7. The microbial ingress challenge is a ______________ assay.
Probabilistic
8. The _______ leak test is the most common liquid tracer
test.
Dye
9. Testing should show the complete CCI picture over the
___________ of the product
Lifecycle
10. All __________ controls with leaks at or above the claimed
limit of detection must fail.
Positive
BINGO11. A lower limit of _______________ must be established.
Detection
12. Quantities of samples to be tested must be ____________ to provide
adequate assurance of package integrity.
Adequate
13. Package integrity should be ______________ when changes are
required in package design.
Re-evaluated
14. All __________ controls must pass.
Negative
15. CCIT can be used to replace sterility testing in ___________
protocols
Stability
BINGO16. Container __________ integrity methods needs to be validated for
specific drug product packages.
Closure
17. Revisions to USP <1207> series are expected to become official in
__________ 2016.
August
18. Package integrity verification occurs during ________ product life
cycle phases.
Three
19. ____________ drilled holes can be used to create positive controls.
Laser
20. Choose a preliminary method following vendor’s recommendations or
____________ research.
Literature
BINGO21. The vacuum decay method is _____________ and quantitative.
Nondestructive
22. The dye leak test is destructive and _______________.
Qualitative
23. CCIT can be performed in ___________ different ways.
Many
24. _____________ may not replace the sterility test for release testing.
CCIT