Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

14
utorial on the Ortec 935 Quad Constant Fraction Discriminato Kevin Carnes Fall, 2004

description

Tutorial on the Ortec 935 Quad Constant Fraction Discriminator. Kevin Carnes Fall, 2004. Threshold Monitor, -200mV to -10V. Threshold Adjust, -20mV to -1000mV. This is the input pulse used for simulating the operation of the Ortec 935. This shows how the input signal is split, - PowerPoint PPT Presentation

Transcript of Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Page 1: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Kevin CarnesFall, 2004

Page 2: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Threshold Adjust, -20mV to -1000mV

Threshold Monitor,-200mV to -10V

Page 3: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

This is the inputpulse used for simulating the operation of the Ortec 935.

Page 4: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

This shows how theinput signal is split, one half attenuatedto 20% of its originalheight, and the otherhalf delayed. The two are then combined to producethe signal thatcan be seen at theCF Monitor.

Page 5: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Three CF Monitorsignals producedwith different delays.

Page 6: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

tdext = T + 1.0 ns, W1 outtdext = T – 0.7 ns, W1 in

Constant FractionMonitor

The equations below arefrom the 935 manual. Theygive the amount of delay produced from the externalcable to achieve a specificinternal delay T. Note thatthe value depends on whether or not the internaljumper W1 is in or out.

Page 7: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

This is an actual scopetrace of the CF Monitoroutput from the 935with a reasonably longdelay cable.

Page 8: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

CF Monitor scope trace with much shorter delay cable.

Page 9: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Walk Adjust,+/- 15 mV

Walk Adjust TestPoint, reads actual voltage

Page 10: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Zero crossing pointof simulated CFMonitor, with longcable (lg) and shortcable. Walk adjustis set to + 10 mV and – 10 mV.

Page 11: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Width Adjust, <4ns to>400ns, depending oninternal jumpers, updateor block

Page 12: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Scope trace of 935 output into 50 ohms, other two outputs in this channelunterminated.

Page 13: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Scope trace of 935output into 50 ohms, second outputterminated in 50 ohms, third output unterminated.

Page 14: Tutorial on the Ortec 935 Quad Constant Fraction Discriminator

Scope trace of 935output into 50 ohms,both of the other outputs terminated in50 ohms.