Total X-ray Fluorescence Spectroscopy (TXRF) · Total X-ray Fluorescence Spectroscopy (TXRF)...
Transcript of Total X-ray Fluorescence Spectroscopy (TXRF) · Total X-ray Fluorescence Spectroscopy (TXRF)...
Total X-ray Fluorescence Spectroscopy (TXRF)
Contact: World Agroforestry Centre (ICRAF), P.O. Box 30677-00100 Nairobi, Kenya. Tel: +254 020 722 4000. www.worldagroforestry.org
• TXRF is a versatile analytical technique for determining elemental content in liquids, solids and loose powders.
• The main principle is that atoms, when irradiated with X-rays, radiate
secondary X-rays – the fluorescence radiation.
• Each element is associated with a specific wavelength and energy of
the fluorescence radiation.
• The concentration of each element is calculated using the fluorescence
intensity.
• TXRF combines high accuracy and precision with simple and fast
sample preparation for the analysis of elements from Sodium (Na; 11) to Uranium (U; 92) in concentration ranges from 100% to sub-ppm level.
• TXRF is a powerful diagnostic tool for agriculture and environment.
• TXRF analysis is suitable for all types of samples
• Sample digestion is not normally required.
• Samples are prepared on a sample tray
that reflects X-ray radiation.
• Trays have a diameter of 30 mm,
made of acrylic or quartz glass.
• Liquids are prepared directly on the sample tray.
• Several μl are transferred to the glass disc
using a pipette and evaporated (Figure 2).
• Powdered samples are finely ground (<50 μm),
prepared in suspensions and transferred to the
sample tray.
• Powder samples (a few μg of sample) can also
be transferred directly, using a Q-tip,
for semi-quantitative analysis.
• Typical Lower Limits of Detection (LLD) are presented in Figure 5.
• The LLD for many elements is 1-10 μg/l.
• Elements from Na (11) to U (92) are detected
• Applications include:
• Plant trace element diagnostics
• Soil element fingerprinting
• Soil heavy metal screening
• Soil solution/extract analysis
• Water quality analysis
• All elements from Sodium to Uranium (excl. Niobium, Molybdenum and Technetium) can be analyzed (Figure 1).
• TXRF analysis is based on internal standardization; an element, which is not present in the sample, must be added for quantification purposes (Figure 4). • Measurement of the complete spectrum; all detectable elements are measured simultaneously.
• The working principle of TXRF spectroscopy as realized in the S2 PICOFOX spectrometer is shown in Figure 3.
• A primary beam generated by a X-ray tube is monochromatized by
Bragg-reflection on a Ni/C multilayer.
• Monochromatized X-rays hit a polished sample carrier (quartz glass or
acrylic glass) at a very shallow angle of incidence.
• X-rays are totally reflected by the surface of the sample carrier at a
very small angle (0.3 – 0.6 °).
• Fluorescence radiation is emitted only by the sample deposited on the
carrier surface.
• The characteristic fluorescence radiation emitted by the sample is
detected by an energy-dispersive detector.
• Intensity is measured by means of an amplifier coupled to a
multichannel analyzer.
Fig. 1: Elements analyzed; Na (11) to U (92)
Fig. 4: Typical spectrum showing peaks of several elements
Fig. 3: Schematic working principle of the S2 PICOFOX spectrometer
Fig.2: Sample preparation steps for the TXRF analysis of liquids
S2 PICOFOX TXRF spectrometer
Fig. 5. Lower Limits of Detection
• Fast and simple sample preparation; no need for pressed pellets • Simultaneous multi-element trace analysis without external calibration • Analysis of small sample amounts in nanogram or microgram range • Suitable for various sample types: solids, powders, liquids, suspensions • Portable system for fast in-field analyses; low power rating • Low maintenance and operating cost • Highly reproducible • Low limits of detection
Sample trays