Spectrofuorometer lab. 1. Fluorescence spectroscopy or (spectrofluorometry) Is a type of...
-
Upload
claud-watts -
Category
Documents
-
view
221 -
download
0
Transcript of Spectrofuorometer lab. 1. Fluorescence spectroscopy or (spectrofluorometry) Is a type of...
Spectrofuorometerlab. 1
Fluorescence spectroscopy or (spectrofluorometry)
Is a type of electromagnetic spectroscopy which analyzes fluorescence from
a sample. It involves using a beam of light, that excites the electrons in
molecules of certain compounds and causes them to emit light.
fluorophore• A fluorophore (or fluorochrome, similarly to a chromophore) is a fluorescent chemical
compound that can re-emit light upon light excitation.
• Fluorophores typically contain several combined aromatic groups, or plane or cyclic
molecules with several π bonds.
• Fluorochromes absorb light of a particular wavelength, they are transiently excited and then
emit light of a longer wavelength as they return to their unexcited state.
• The emitted light is referred to as fluorescent light (fluorescence).
• The intensity of the emitted light is directly proportional to the amount of fluorochrome in
the sample.
• Not all the light energy that is absorbed by the fluorochrome is emitted as fluorescence
because some of the energy is quickly dissipated as heat or vibrational energy.
Instrument design
• The instrument design must provide some mechanism by which different
excitation and emission wavelengths of light are selected.
• This is achieved by use wavelength selectors which are
monochromators.
• Monochromators are device made of several components, one component
disperses the spectrum of visible light, while the other component is
adjustable, and is used to select the particular wavelength of light that
will pass through the monochromator .
Instrument design
• A spectrofluorometer must have an excitation monochromator and
emission monochromator because
• The absorption and emission spectrum of one compound differs from that of another compound.
• It should always be kept in mind that the wavelength selector must be adjusted for every compound tested.
Principle of spectrofuorometer function• The excitation light signals is visible light consisting of many wavelengths
radiating in all directions. Xenon arc lamp is often used as an alternative
source because
• A device known as excitation monochromators is used to focus the light onto a
prism or grating, which cause dispersion of the spectrum of light. Computer
driven exit slits determine the wavelength of the light that exits the
monochromators and enters the cuvette. it is crucial that all extraneous
wavelengths of the light be blocked before entering the cuvette because this
would contribute to the measured emission.
Principle of spectrofuorometer function• The excitation light penetrates the cuvette and is absorbed by the
fluorochrome, when the fluorochrome absorbs light energy it, it is excited
momentarily. As the fluorochrome reverts back to its more stable state,
light of longer wavelength is emitted. • Fluorescence is emitted in all directions. A slit in the emission
monochromator allow light to focused onto the wave selector. The
emission monochromator is set to the wavelength of light emitted as
the fluorochrome returned to its stable state.
Principle of spectrofuorometer function• The emission monochromator is spatially arranged at 90º to the
excitation monochromators to minimize the amount of excitation light
that reaches the detector .
• The detector is often a photomultiplier tube (electron tube) that absorbs
the emitted light and ejects electron in proportion to the amount of
light absorbed.
A schematic of spectrofluorometer
v
v
v
Determination of unknown quinine sulphate
concentration using calibration curve
• In five 50 ml volumetric flasks, prepare a serial dilutions of quinine sulphate by pipette 2,4,6,8,10 ml from the stock solution of Q.S (5 g/L).
• Complete to volume with 0.2N H2SO4 and mix well.• Set the λ excitation at 350 nm and λ emission at 450 nm.• Record the relative fluorescence intensity for each dilution, and for the
unknown sample.• Plot the calibration curve by plotting relative fluorescence intensity versus
concentration.• Determine the unknown conc. of QS from the graph. • 0
procedure
ml that you pipette from quinine sulphate stockCalculate conc. of Q.S for each diluted flask
Stock Q.S concentration (5 g / L)
Relative Fluorescence
intensity
2mlC X V = C; X V5) g / L (X 2 ml = C; X 50 ml
? from the instrument
4mlC X V = C; X V5) g / L (X 4 ml = C; X 50 ml
? from the instrument
6mlC X V = C; X V5) g / L (X 6 ml = C; X 50 ml
? from the instrument
8mlC X V = C; X V5) g / L (X 8 ml = C; X 50 ml
? from the instrument
10ml C X V = C; X V5) g / L (X 10 ml = C; X 50 ml
? from the instrument
Unknown sample conc.? from the instrument
From the calibration graph
calculation
Unknown concentration
How to determine unknown concentration from the calibration curve
Determination of unknown quinine sulphate
concentration using linear equation
XY
10.260
20.4120
30.6180
40.8244
51310
1 -press AC …... Mode ….. (choose 3: STAT)…... (Choose 2: A +B X)
2- Fill the schedule with the calculated concentration in X column, and with fluorescence intensity obtained in Y column. As shown.
After each number you add
press = to transfer to the next row
3- press AC….. Shift 1 ….. Choose 7: Reg ……. (Then choose 1 : A) …. Press = ……. You will have intercept value.
4- press AC….. Shift 1 ….. Choose 7: Reg ……. (Then choose 2 : B) …. Press = ……. You will have slope value.
Y = a + bx
Unknown Conc.
slope
Unknown fluorescence intensity
intercept
Several factors affect the amount of fluorescence that is measured and so not all
emitted light is measured. The most common factor is quenching .
Quenching refers to a decrease in fluorescent intensity. Quenching by the
solvent may occur when the fluorochrome interacts with molecules in the
solution. A variety of processes can result in quenching such as :
energy transfer, complex-formation and collisional quenching.
The most common chemical quenchers are:
•Oxygen , iodide ions, chloride ions and acrylamide.
Factors that affect fluorescence
measurements
processes can result in quenching
Determination of the quenching effect of KI on the fluorescence of
quinine sulphate
procedure
• In eight 100 ml volumetric flasks, add 16 ml of Q.S.
• Add 0,1,1.5,2.5,5,7,10,15 ml of KI ( stock conc. 0.1M) to each flask.
• Complete to volume with distilled water.
• Read and record relative fluorescence intensity for each flask then plot intensity versus conc. of KI
• Comment on the curve
calculationml of Q.S.ml of KI Calculate conc. of KI for each diluted flask
))Stock conc. 0.1M
Intensity
160C X V = C; X V0.1 M X 0 ml = C; X 100 ml
?from the instrument
161C X V = C; X V0.1 M X 1 ml = C; X 100 ml
?from the instrument
161.5C X V = C; X V0.1 M X 1.5ml = C; X 100 ml
?from the instrument
162.5C X V = C; X V0.1 M X 2.5ml = C; X 100 ml
?from the instrument
165C X V = C; X V0.1 M X 5 ml = C; X 100 ml
?from the instrument
167C X V = C; X V0.1 M X 7 ml = C; X 100 ml
?from the instrument
1610C X V = C; X V0.1 M X 10 ml = C; X 100 ml
?from the instrument
1615C X V = C; X V0.1 M X 15ml = C; X 100 ml
?from the instrument
Effect of of KI on the fluorescence of quinine
sulphate
What is the difference between Fluorescence spectroscopy and spectrophotometer?