8/2/2012 1 Cranial SRS on a Siemens Artiste™ using the BrainLab iPlan® treatment planning system Olivier Gayou, PhD Allegheny General Hospital Pittsburgh, PA Outline SRS workflow Siemens and BrainLab equipment System interdependence Bumps on the road Experience at AGH Conclusion Objectives To assess the safety and usefulness of a multivendor solution for cranial SRS To understand the commissioning and QA steps of a Siemens Artiste/BrainLab iPlan hybrid system for cranial SRS To understand the limitations of the hybrid system

Transcript of Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™...

Page 1: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

1

Cranial SRS on a Siemens

Artiste™ using the BrainLab

iPlan® treatment planning system

Olivier Gayou, PhD

Allegheny General Hospital

Pittsburgh, PA

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

Objectives

� To assess the safety and usefulness of a multivendor solution for cranial

SRS

� To understand the commissioning

and QA steps of a Siemens Artiste/BrainLab iPlan hybrid system for cranial SRS

� To understand the limitations of the hybrid system

Page 2: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

2

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

SRS workflow at AGH

MRI

Frame

CT with localizer

FusionContouring

Planning

Evaluation and approval

Treatment

Datatransfer - OIS

Localization

Imaging Planning - Brainlab

Checking –Lantis, Diamond

Delivery – Siemens Artiste

Winston-Lutztest

2nd check

MV-CBCT

Page 3: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

3

SRS workflow at AGH

MRI Frame CT Fusion

Contour

-Rx

Plan EvalData

transfer

2nd

checkQA Treat

Patient

Neuro

surgeon

Radiation

Oncologist

Physicist

Radiology

Therapist

Team work

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

Page 4: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

4

Equipment: The BrainLab SRS system

iPlan®

Treatment planningsystem

Couch extension

ConesCone mount

Localizer

Head frame

Target positioner

Equipment: The BrainLab TPS

iPlan Image4.5

Localize

Fuse

Segment

iPlan Dose4.5.1

Conformal arc

Circular arc

Conformal beam

Dynamicarc

Static IMRT

CTMRIAngioPET

AutoManual

Pencil beamMonte-Carlo

Cones

Various MLCmMLC

iPlan®

Treatment planningsystem

Equipment: the Siemens Artiste™ linac

High-end Siemens linac

6X Flat beam300 MU/min

6ST FFF beam1000 MU/min

� 5 mm width at isocenter� Up to 40×40 cm2

� Tilted leaf, single-focus design� Leakage: 0.37%� 43 cm clearance to isocenter� Leaf speed 42.9±0.6 mm/s� Penumbra: 4.1±0.5 mm

Tacke et al, ‘6MV dosimetric characterization of the 160 MLC™, the new Siemens multileaf collimator,’ Med Phys 35(5), 1634-

1642 (2008)

With small field-restricting accessory only (SRS cones)

160 MLC™

Page 5: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

5

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

System interdependence: a simplistic

view

Deliverysystem

-

Siemens

Artiste™linear

accelerator

Planningsystem

-

BrainLab

iPlan®

TPS can model linac behavior

Linac can deliver TPS plan

System dedicated to SRSCan linac perform inSRS environment?

Dosimetric matching (calculated dose = delivered dose)

Geometric matching (plan isocenter = linac isocenter

Page 6: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

6

System interdependence: a detailed

view

Brainlabcouch

extension

Brainlab head frame

Brainlab localizer

Treatment setup

Siemens treatment couch

Treatment localizationSiemensMV-CBCT

Brainlab SRS cones

Siemens160-MLC

Treatment planning

Beam shaping devices

Treatment delivery

Brainlab iPlan®

SiemensArtiste™

attaches toattaches to

models

attaches to

System interdependence: a detailed

view

Brainlabcouch

extension

Brainlab head frame

Brainlab localizer

Treatment setup

Siemens treatment couch

Treatment localizationSiemensMV-CBCT

Brainlab SRS cones

Siemens160-MLC

Treatment planning

Beam shaping devices

Treatment delivery

Brainlab iPlan®

SiemensArtiste™

attaches toattaches to

Treatment couch

Siemens 550 TxT TT-S treatment couch

Max load 550lbs / 250kg

Isocentric displacement <1mm

Isocentric rotation ±115°

Absolute position accuracy±0.5mm (±1.0mm

vert)

Corrective position accuracy±0.2mm (±1.0mm

vert)

BrainLab couch extension

Locking mechanism

X,Y,Z and tilt microadjustments

Green locking indicator

Page 7: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

7

BrainLab head frame

Immobilization of head in frameImmobilization of frame on couch

extension (linac isocenter)

Head frame = frame of referenceIsocenter calculated in frame of reference

System interdependence: a detailed

view

Brainlabcouch

extension

Brainlab head frame

Brainlab localizer

Treatment setup

Siemens treatment couch

Treatment localizationSiemensMV-CBCT

Brainlab SRS cones

Siemens160-MLC

Treatment planning

Beam shaping devices

Treatment delivery

Brainlab iPlan®

SiemensArtiste™

attaches toattaches to

Data acquisition for the 160 MLC

• Output• Leakage• PDD/TMR• Scatter factors• Diagonal radial profiles

• Transversal profiles

BrainLab-provided Excel template

(1) Read instructions

(2) Perform measurements

(3) Send to BrainLab for verification and consistency(4) Enter in Beam Profile Editor(5) Take verification measurements

Page 8: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

8

Data acquisition for 160 MLC: absolute

output and leakage

SSD=100cmd=10cm10×10 cm2

0.6 cc Farmer chamber

NLout=0.657 cGy/MU

Leakage measurements:

Open jaws/closed MLC: 2.1×10-3

Closed jaws/closed MLC: 1.1×10-4

Data acquisition for 160 MLC: PDD

� PDD at SSD=100cm� 10 square field sizes� Wellhofer cc01 chamber

Data acquisition for 160 MLC: scatter

factors

� MLC-definedsquare fields� Different jaw overlap� Wellhofer cc01 chamber� PTW photon diode

Page 9: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

9

Data acquisition for 160 MLC: diagonal

profile

� SSD=100cm� 8 depths (0.5, 1.4, 2.5, 5, 10, 20 and 35 cm)� Wellhofer cc01 chamber

35 cm

20 cm

10 cm

1.4 cm

2.5 cm

5 cm

0.5 cm

Data acquisition for 160 MLC:

transversal profile

x

y

� Isocentric� 3 different depths (dmax, 10 and 20cm)� PTW diode� Purpose: adjust the source function correction and radiological field correction (simulate effects of

extended source and other beam smearing effects)

System interdependence: a detailed

view

Brainlabcouch

extension

Brainlab head frame

Brainlab localizer

Treatment setup

Siemens treatment couch

Treatment localizationSiemensMV-CBCT

Brainlab SRS cones

Siemens160-MLC

Treatment planning

Beam shaping devices

Treatment delivery

Brainlab iPlan®

SiemensArtiste™

attaches toattaches to

models

Page 10: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

10

BrainLab cones on the Siemens Artiste

+ + =

� Accessory interlock (code STEREO in slot 3)� detects presence of cone mount, but NOT cone size

� Automatically sets MLC and y-jaws to 5×5 field size

Data acquisition for BrainLab cones

Scatter factors

Cone profiles

PDD

System interdependence: a detailed

view

Brainlabcouch

extension

Brainlab head frame

Brainlab localizer

Treatment setup

Siemens treatment couch

Treatment localizationSiemensMV-CBCT

Brainlab SRS cones

Siemens160-MLC

Treatment planning

Beam shaping devices

Treatment delivery

Brainlab iPlan®

SiemensArtiste™

attaches toattaches to

models

Page 11: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

11

Isocenter verification

Plan isocenter defined with respect to head frame using BrainLab localizer

Room lasers point to linac isocenter –verified by Winston-Lutz test

Target positioning sheets point to physical location of plan isocenter in frame of

reference

Align to lasers

End-to-end accuracy: 1.2mm

System interdependence: a detailed

view

Brainlabcouch

extension

Brainlab head frame

Brainlab localizer

Treatment setup

Siemens treatment couch

Treatment localizationSiemensMV-CBCT

Brainlab SRS cones

Siemens160-MLC

Treatment planning

Beam shaping devices

Treatment delivery

Brainlab iPlan®

SiemensArtiste™

attaches toattaches to

models

Frame/anatomy integrity verification

� Register MV-CBCT with planning CT• Skull contour from TPS• Align to MV-CBCT

� Verifies head position in head frame

No shift = no displacement

Page 12: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

12

Frame/anatomy integrity verification

Full export DICOM license

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

Page 13: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

13

Bumps on the road…

(1) Original design of accessory mount on Artiste™: the cone mount could not be inserted

� We filed our mount� Siemens redesigned mount for Artiste

Bumps on the road…

(2) Small field size scatter factor measurement inconsistent with other sites

~2%

~4.6%

� Repeat measurement with careful alignment of stereotactic diode

10×10 MLC field size, vary y-jaw opening

Cone size safety interlock

� Linac has no knowledge of cone size� Potential major radiation error� 2 person check of cone size� User defined pause

� Prompt team to switch cone size

Page 14: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

14

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

Outline

� SRS workflow

� Siemens and BrainLab equipment

� System interdependence

� Bumps on the road

� Experience at AGH

� Conclusion

Artiste + BrainLab SRS experience at

AGH

March 2008 May 2009

Varian 2100C + Radionics

BrainLab BrainScan® +

Siemens Primus™

BrainLab iPlan® + Siemens Artiste™

(up to July 2011)� 232 SRS patients

Metastatic disease 187Meningioma 14AVM 14Acoustic neuroma 10Other 7

Page 15: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

15

Case example

� 75 year old Caucasian male patient� Stage IV, T3N3M1b adenocarcinoma of the left lower lobe of the lung with associated bilateral mediastinal and hilar disease� 3 brain metastases� Whole brain radiation therapy 4/12 followed by palliative chemoradiation to the thorax (5/12)� Treated with SRS boost to brain mets (July 2012)

7am 8am 9am 12pm10am 11am 2pm1pm 4pm3pm 6pm5pm 7pm

MRI

Frame placement

Plan designed on MRI

Frame adjustment

CT

Plan finalized on CT and approved

Data transfer and 2nd check

Treatment

Frame issue

10 hours

Cross system check

Case example

� Right temporal� Right superior parietal� Right inferior parietal� 1600 cGy, 80% isodose line� 1 mm PTV margin

Conclusion

SRS-dedicated treatmentplanning system with

accessories -

BrainLab iPlan®

SRS-capable linearaccelerator

-Siemens Artiste™

Successful SRS program

� Take advantage of industry standards (DICOM, etc…)� Requires careful cross checking of information between systems� Centrality of physicist role in proper use of multiplatform system� “Rigidity” of cranial SRS treatment well suited for multivendor system

→ what you see is what you get

→ no need for adaptive→ no need for real time adjustment

Page 16: Thu.218.1030.02.Gayou.pptamos3.aapm.org/abstracts/pdf/68-19924-235349-85578.pdf · Artiste™ attaches to attaches to Data acquisition for the 160 MLC •Output •Leakage •PDD/TMR

8/2/2012

16

Acknowledgements

� Physicists, especially

� Mark Johnson, MS

� James MacKenzie, PhD

� Yongbok Kim, PhD

� Physicians, especially

� Russell Fuhrer, MD

� Stephen Karlovits, MD

� Athanasios Colonias, MD

� Therapists

� The whole department…