The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck®...

34
The Price Advantage. Proven Stratos® technology now available in reasonably priced, basic version for pH, conductivity, or oxygen Isolation Amplifiers Transmitters Indicators Process Analytics Portable Meters Laboratory Meters Sensors Fittings Stratos® Eco

Transcript of The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck®...

Page 1: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

The Price Advantage.Proven Stratos® technology now available in reasonably

priced, basic version for pH, conductivity, or oxygen

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ingsStratos® Eco

Page 2: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

2

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings Stratos® Eco Series

For pH, ORP, Conductivity, or Dissolved Oxygen

Made in Germany

High-tech at an astonishingly good price – the new

Stratos® Eco series is intended for use in almost all

standard applications. The reliable 4-wire devices

for the measurement of pH and ORP values, conduc-

tivity, or dissolved oxygen are the result of efficient

product development. Superior Knick technology is

now also available in powerful basic versions.

4-Wire Process Analyzers of the Stratos® Eco Series Stratos® Eco 2405 pH for measuring pH and ORP values

Stratos® Eco 2405 Oxy for measuring dissolved oxygen

Stratos® Eco 2405 Cond for measuring conductivity using 2- or 4-electrode sensors

Stratos® Eco 2405 CondI for measuring conductivity using electrodeless sensors

Page 3: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

3

Uncompromising quality.

From the calibration timer to the automatic GainCheck® device self-test, from the red alarm LED to the SensoCheck® sensor monitoring feature: The Stratos® Eco models provide all the sophisticated functions of the proven Stratos® devices.

Globally ready.

The integrated VariPower® broad-range power supply unit allows trouble-free operation with any supply voltage (20 ... 253 V AC/DC) even with large fluctuations in the mains voltage.

Easy to operate.

The self-explaining user interface ensures comfortable and intui-tive handling. Measured values and temperature are displayed simultaneously on the large, high-contrast LC display. International icons provide operating messages and signal unusual operating states.

Mode indicators show the current operating mode. The flat and fully encap sulated electronics is integrated in the enclosure front. This has made ample room for cabling within the device and all parts are easy to access.

Proven Stratos® technology now at an astonishingly

good price!

Sensor measures reliably

Measured value still acceptable,

sensor should be checked and

replaced, if required

Unreliable measurement,

replace sensor

Preventive Maintenance Indication with SensoFace®

3-year warranty

Page 4: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

4

Stratos® Eco SeriesFor pH, ORP, Conductivity, or Dissolved Oxygen

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Cost-saving.

The circuit technology, unrivalled by other manufacturers, means fewer electronic components and reduced energy consumption. These advantages give rise to an MTBF value of 110 years – plus the unusually long warranty period of 3 years.Thus, Stratos® Eco represents outstanding availability and extraordinary reliability.

Simple to install.

Like other Stratos® products, the Eco series offers the option of wall, pole/post and panel installation. The installation is simplified by plug-in terminals and the pre-mountable plastic housing made from robust PBT.

Cable glands: M 20 x 1.5 and NPT 1/2" or rigid metallic conduit

Robust and safe.

• Robust PBT enclosure

• IP 65 protection

• Suitable for rigid metallic conduits

• Ex classification Class I Div. 2

Reliable Flexibility

Page 5: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

5

The Facts• Easy installation; large terminal compartment,

rear unit can be pre-assembled

• Wall, post/pipe, or panel mounting

• Application in hazardous locations (FM, CSA Class I Div. 2 / Zone 2)

• Large, high-contrast display

• Simple, intuitive operation

• Optical alarm signaling by bright, red LED

• Temperature indication either in ºC or ºF

• Easy calibration using process medium

• Second current output for temperature measurement

• Two floating relay contacts

• Sensocheck® continuous sensor monitoring

• SensoFace® preventive maintenance indication

• Automatic buffer recognition by Calimatic®

• VariPower® power supply for 20 to 253 V AC/DC

• Display of concentration (conductivity)

• Reset to factory settings possible

Process Analyzers of the Stratos® Eco Series

Page 6: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

6

Stratos® Eco SeriesFor pH, ORP, Conductivity, or Dissolved Oxygen

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Applications• Simple industrial applications

• Downstream applications

• Water processing

• Aeration of natural waters

• Energy production

• Boiler water

• Cooling water

• Salt spring

• Seawater desalination

• Mineral water

• Regeneration of ion exchangers

• Drinking water / well water

• Reverse osmosis

• Fish farming

Sensors• pH

• ORP

• 2-electrode conductivity

• 4-electrode conductivity

• Electrodeless conductivity

• Dissolved oxygen

Complete Solutions for a Broad Range of Applications

Page 7: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

7

Stratos® Eco 2405 pH

Stratos® Eco 2405 Cond

Stratos® Eco 2405 CondI

Stratos® Eco 2405 Oxy

Page 8: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

8

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Stratos® Eco 2405 pH

Suitable Sensors:

• All customary pH sensors with nominal zero point at pH 7

• Combination electrodes or separate glass and reference electrodes

• A broad selection of temperature sensors (Pt 100, Pt 1000, NTC 30 kohms, NTC 8.55 kohms, or Balco 3 kohms)

• Common metallic ORP electrodes (gold, platinum)

pH Measurement:

• Bypass • Inline • Dip

Recommended Installation Sets

Page 9: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

9

ϑ

, Wiring

, Product Line

, Terminal Assignments

Order No.

Stratos®Eco 2405 pH Process analyzer for measuring pH values 2405 pH

Mounting accessories

Pipe-mount kit ZU 0274

Panel-mount kit ZU 0275

Protective hood ZU 0276

For further information, please refer to our »Sensors, Fittings, Accessories« catalog or to www.knick.de.

pH measurement Control room / station

Voltage supply

Glass electrode

Reference electrode

Pt 100 / Pt 1000 / NTC 30 kohms NTC 8.55 kohms Balco 3 kohms

pH, ORP, Sensocheck®

Temperature

Alarm contact

Limit value

20 ... 253 V AC/DC

mA

mA

0(4) ... 20 mA

0(4) ... 20 mA

pH/mV

Temp

Stratos® Eco

2405 pH

Voltage supply

Alarm

Limit value

Page 10: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

10

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

pH/mV input Input for pH or ORP sensors

Measuring range –1500 … +1500 mV

Display range pH value: –2.00 … 16.00 ORP: –1999 … +1999 mV

Glass electrode input!} Input resistance > 0.5 x 10!‡ ohms Input current < 2 x 10–!‡ A

Reference electrode input!} Input resistance > 1 x 10!= ohms Input current < 1 x 10–!= A

Measurement error1) 2) 3) pH value: < 0.02 TC: 0.002 pH/K (display) mV value: < 1 mV TC: 0.1 mV/K

pH sensor standardization*} pH calibration

Operating modes BUF Calibration with Calimatic automatic buffer recognition®: Buffer sets: -01- Knick/Mettler-Toledo 2.00 / 4.01 / 7.00 / 9.21 -02- Merck/Riedel de Haen 2.00 / 4.00 / 7.00 / 9.00 / 12.00 -03- Ciba (94) 2.06 / 4.00 / 7.00 / 10.00 -04- NIST technical 1.68 / 4.00 / 7.00 / 10.01 / 12.46 -05- NIST standard 1.680 / 4.008 / 6.865 / 9.184 -06- HACH 4.01 / 7.00 / 10.01 -07- WTW technical buffers 2.00 / 4.01 / 7.00 / 10.00 -08- Hamilton 4.01 / 7.00 / 10.01 MAN Manual calibration with input of individual buffer values DAT Data entry of pre-measured electrodes

Max. calibration range Asymmetry potential: ±60 mV; slope: 80 ... 103 % (47.5 ... 61 mV/pH)

ORP sensor standardization*} ORP calibration Max. calibration range ¥: –700 … +700 mV

Calibration timer 0000 ... 9999 h

Sensocheck® Automatic monitoring of glass electrode (can be disabled)

Sensoface® Provides information on the sensor condition, evaluation of zero/slope, response time, calibration interval, Sensocheck®

Temperature input*} Pt 100/Pt 1000/NTC 30 kohms/NTC 8.55 kohms/Balco 3 kohms; 2-wire connection, adjustable

Measuring range Pt 100/Pt 1000 –20.0 … +200.0 °C (–4 … +392 °F) NTC 30 kohms –20.0 … +150.0 °C (–4 … +302 °F) NTC 8.55 kohms –10.0 … +130.0 °C (+14 … +266 °F) Balco 3 kohms 0.0 … +100.0 °C (+32 … +212 °F)

Adjustment range 10 K

Resolution 0.1 °C / 1 °F

Measurement error1) 2) 3) < 0.5 K (< 1K for Pt100; < 1K for NTC >100 °C)

Temperature compensation of process medium

Linear –19.99 … +19.9 %/K (reference temp 25 °C)

, Specifications

Stratos® Eco 2405 pH

Page 11: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

11

Output 1 0/4 … 20 mA, max. 10 V, floating (galvanically connected to output 2)

Measured variable*} pH or mV value

Overrange*} 22 mA in the case of error messages

Output filter*} Low-pass, filter time constant: 0 … 120 s

Measurement error!} < 0.3% current value + 0.05 mA

Start/end of scale*} Configurable within the measuring range for pH or mV

Permitted span pH 2.00 … 18.00 / 200 … 3000 mV

Output 2 0/4 … 20 mA, max. 10 V, floating (galvanically connected to output 1)

Measured variable Temperature

Overrange*} 22 mA in case of temp error messages

Output filter*} Low-pass, filter time constant: 0 … 120 s

Measurement error!} < 0.3% current value + 0.05 mA

Start/end of scale*} –20 … 200 °C / –4 … 392 °F

Permitted span 20 … 220 K (36 … 396 °F)

Alarm contact Relay contact, floating AC < 250 V / < 3 A / < 750 VA DC < 30 V / < 3 A / < 90 W

Contact response N/C (fail-safe type)

Alarm delay 10 s

Limit contact Relay contact, floating AC< 250 V / < 3 A / < 750 VA DC< 30 V / < 3 A / < 90 W

Contact response*} N/C or N/O

Delay*} 0000 ... 9999 s

Setpoints*} As desired within range

Hysteresis*} pH 0.00 … 5.00 / 0 … 500 mV

Specifications (continued)

Page 12: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

12

Display LC display, 7-segment with icons

Main display Character height 17 mm, unit symbols 10 mm

Secondary display Character height 10 mm, unit symbols 7 mm

Sensoface® 3 status indicators (friendly, neutral, sad face)

Mode indication 4 mode indicators "meas", "cal", "alarm", "config" Further icons for configuration and messages

Alarm indication Red LED in case of alarm

Keypad 5 keys: [cal] [conf] [?] [±] [enter]

Service functions

Current source Current specifiable for output 1 and 2 (00.00 ... 22.00 mA)

Device self-test Automatic memory test (RAM, FLASH, EEPROM)

Display test Display of all segments

Last Error Display of last error occurred

Sensor monitor Display of direct, uncorrected sensor signal

Data retention Parameters and calibration data > 10 years (EEPROM)

Protection against electric shock

Safe Isolation of all extra-low-voltage circuits against mains by double insulation to EN 61010-1

Power supply 24 (–15 %) ... 230 (+10 %) V AC/DC approx. 5 VA / 2.5 W AC: 45 … 65 Hz Overvoltage category II, Class II

Nominal operating conditions

Ambient temperature –20 … +55 °C

Transport/Storage temp –20 … +70 °C

Relative humidity 10 ... 95 % not condensing

EMC EN 61326-1, EN 61326-2-3 Emitted interference: Class B (residential area) Class A: for mains > 60 V DC Immunity to interference: Industry

Explosion protection FM NI, Class I Div. 2, Groups A,B,C & D T4, Ta = 55 °C; Type 2 NI, Class I Zone 2, Group IIC T4, Ta = 55 °C; Type 2 CSA Class I Div. 2, Groups A, B, C and D, T4 Ex nA IIC T4

Specifications (continued)

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Stratos® Eco 2405 pH

Page 13: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

13

Enclosure Molded enclosure made from PBT (polybutylene terephtalate)

Color Bluish gray RAL 7031

Assembly – Wall mounting – Pipe mounting ZU 0274 Ø 40 … 60 mm, _ 30 … 45 mm – Panel mounting ZU 0275

cutout to DIN 43 700, sealed against panel

Dimensions H 144 mm, W 144 mm, D 105 mm

Ingress protection IP 65 / NEMA 4X

Cable glands 3 knockouts for cable glands M20x1.5 2 knockouts for 1/2" NPT or rigid metallic conduit

Weight Approx. 1 kg

*) User-defined 1) To IEC 746 Part 1, at nominal operating conditions 2) ±1 count 3) Plus sensor error

Specifications (continued)

Page 14: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

14

Stratos® Eco 2405 Oxy

Suitable Sensors: • Customary amperometric sensors with

or without polarization voltage

• NTC 22 kohms, NTC 30 kohms temperature detectors

DO Measurement:

• Bypass • Inline • Dip

Recommended Installation Sets

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Page 15: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

15

, Wiring

, Terminal Assignments

Order No.

Stratos® Eco 2405 Oxy Process analyzer for measuring dissolved oxygen 2405 Oxy

Mounting accessories

Pipe-mount kit ZU 0274

Panel-mount kit ZU 0275

Protective hood ZU 0276

, Product Line

For further information, please refer to our »Sensors, Fittings, Accessories« catalog or to www.knick.de.

O2 measurement Control room / station

Voltage supply

%, mg/l

Temperature

Alarm contact

Limit value

20 ... 253 V AC/DC

mA

mA

0(4) ... 20 mA

0(4) ... 20 mA

SAT %

Temp

Stratos® Eco

2405 Oxy

Voltage supply

Alarm

Limit value

O2 sensor

Page 16: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

16

, Specifications

Stratos® Eco 2405 Oxy

O2 input

Measuring current –2 ... +1800 nA

Resolution (at Vpol ≤ 800 mV and Vref ≤ 200 mV)

0.05 nA

Saturation (–10 ... 80 °C) 0 ... 200%

Measurement error1) 2) 3) 0.5 % meas.val. + 0.5 %

Concentration (–10 …80 °C)

0.00 ... 20.00 mg/l 0.00 ... 20.00 ppm

Measurement error1) 2) 3) 0.5 % meas.val. + 0.05 mg/l (0.05 ppm)

Permitted guard current ≤ 20 µA

Polarization voltage*} 0 … 1000 mV

Process pressure*} 0.000 … 9.999 bars (… 999.9 kPa / … 145.0 psi)

Salinity correction*} 00.00 ... 45.00 g/kg

Sensor standardization

Operating modes O2 saturation (automatic) O2 concentration (automatic) Product calibration Zero calibration

Calibration range: sensor type A Zero point ± 2 nA

Slope 25 ... 130 nA (at 25°C, 1013 mbars)

Calibration range: sensor type B Zero point ± 2 nA

Slope 200 ... 550 nA (at 25°C, 1013 mbars)

Calibration timer*} 0000 ... 9999 h

Pressure correction*} 0.000 ... 9.999 bars / 999.9 kPa / 145.0 psi

Sensocheck® Monitoring for short circuits / open circuits (can be disabled)

Sensoface® Provides information on the sensor condition (evaluation of zero/slope, response time, calibration interval, Sensocheck®)

Temperature input*} NTC 22 kohms / NTC 30 kohms, 2-wire connection, adjustable

Measuring range –20.0 ... +150.0 °C / –4 ... +302 °F

Adjustment range 10 K

Resolution 0.1 °C / 1 °F

Measurement error1) 2) 3) < 0.5 K (< 1 K at > 100°C)

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Page 17: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

17

Specifications (continued)

Output 1 0/4 ... 20 mA, max. 10 V, floating (galvanically connected to output 2)

Process variable*} O2 saturation/O2 concentration

Overrange*} 22 mA in the case of error messages

Output filter*} Low-pass, filter time constant 0 ... 120 s

Measurement error!} < 0.3 % current value + 0.05 mA

Start/end of scale As desired within range

Permitted span 5 ... 200 % / 0.5 ... 20 mg/l (ppm)

Output 2 0/4 ... 20 mA, max. 10 V, floating (galvanically connected to output 1)

Measured variable Temperature

Overrange*} 22 mA in case of temp error messages

Output filter*} Low-pass, filter time constant 0 ... 120 s

Measurement error!} < 0.3 % current value + 0.05 mA

Start/end of scale –20 ... +150 °C / –4 ... +302 °F

Permitted span 20 ... 170 K / 36 ... 306 °F

Alarm contact Relay contact, floating

Contact ratings AC < 250 V / < 3 A / < 750 VA DC < 30 V / < 3 A / < 90 W

Contact response N/C (fail-safe type)

Response delay 10 s

Limit contact Relay contact, floating AC < 250 V / < 3 A / < 750 VA DC < 30 V / < 3 A / < 90 W

Contact response*} N/C or N/O

Delay*} 0000 ... 9999 s

Setpoints*} Within selected range

Hysteresis*} 0.0 ... 50.0 % / 0.00 ... 5.00 mg/l (ppm)

Page 18: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

18

Specifications (continued)

Stratos® Eco 2405 Oxy

Display LC display, 7-segment with icons

Main display Character height 17 mm, unit symbols 10 mm

Secondary display Character height 10 mm, unit symbols 7 mm

Sensoface® 3 status indicators (friendly, neutral, sad face)

Mode indication 4 mode indicators "meas", "cal", "alarm", "config" Further icons for configuration and messages

Alarm indication Red LED in case of alarm

Keypad 5 keys: [cal] [conf] [?] [±] [enter]

Service functions

Current source Current specifiable for output 1 and 2 (00.00 ... 22.00 mA)

Device self-test Automatic memory test (RAM, FLASH, EEPROM)

Display test Display of all segments

Last Error Display of last error occurred

Sensor monitor Display of direct, uncorrected sensor signal

Data retention Parameters and calibration data > 10 years (EEPROM)

Protection against electric shock Safe Isolation of all extra-low-voltage circuits against mains by double insulation to EN 61010-1

Power supply 24 (–15 %) ... 230 (+10 %) V AC/DC approx. 5 VA / 2.5 W AC: 45 … 65 Hz Overvoltage category II, Class II

Nominal operating conditions

Ambient temperature –20 … +55 °C

Transport/Storage temp –20 … +70 °C

Relative humidity 10 ... 95 % not condensing

EMC EN 61326-1, EN 61326-2-3 Emitted interference: Class B (residential area) Class A: for mains > 60 V DC Immunity to interference: Industry

Explosion protection FM NI, Class I Div. 2, Groups A,B,C & D T4, Ta = 55 °C; Type 2 NI, Class I Zone 2, Group IIC T4, Ta = 55 °C; Type 2 CSA Class I Div. 2, Groups A, B, C and D, T4 Ex nA IIC T4

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Page 19: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

19

Specifications (continued)

Enclosure Molded enclosure made from PBT (polybutylene terephtalate)

Color Bluish gray RAL 7031

Assembly – Wall mounting – Pipe mounting ZU 0274 Ø 40 … 60 mm, _ 30 … 45 mm – Panel mounting ZU 0275

cutout to DIN 43 700, sealed against panel

Dimensions H 144 mm, W 144 mm, D 105 mm

Ingress protection IP 65 / NEMA 4X

Cable glands 3 knockouts for cable glands M20x1.5 2 knockouts for 1/2" NPT or rigid metallic conduit

Weight Approx. 1 kg

*) User-defined 1) To IEC 746 Part 1, at nominal operating conditions 2) ±1 count 3) Plus sensor error

Page 20: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

20

• Bypass • Inline

Recommended Installation Sets

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Stratos® Eco 2405 Cond

Suitable Sensors:

• All known types of 2-electrode sensors

• All common types of 4-electrode sensors

• Cell constant selectable from 0.005 to 20 cm–!

• Selectable temperature detectors Pt 100, Pt 1000, NTC 30 kohms, NTC 8.55 kohms

Conductive Conductivity Measurement:

Page 21: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

21

ϑ

, Wiring

, Terminal Assignments

, Product LineOrder No.

Stratos® Eco 2405 Cond Process analyzer for measuring conductivity using 2- and 4-electrode sensors

2405 Cond

Mounting accessories

Pipe-mount kit ZU 0274

Panel-mount kit ZU 0275

Protective hood ZU 0276

For further information, please refer to our »Sensors, Fittings, Accessories« catalog or to www.knick.de.

Conductivity measurement

Control room / station

Voltage supply

Pt 100 / Pt 1000 / NTC 30 kohms NTC 8.5 kohms

mS, %, SAL Sensocheck®

Temperature

Alarm contact

Limit value

20 ... 253 V AC/DC

mA

mA

0(4) ... 20 mA

0(4) ... 20 mA

mS/cm

Temp

Stratos® Eco

2405 Cond

Voltage supply

Alarm

Limit value

Page 22: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

22

Conductivity input Input for 2-electrode or 4-electrode sensors

Effective ranges 0.2 µS · c … 1000 mS · c

Ranges*} Conductivity 0.000 … 9.999 µS/cm 0.00 … 99.99 µS/cm 0.0 … 999.9 µS/cm 0 … 9999 µS/cm 0.000 … 9.999 mS/cm 0.00 … 99.99 mS/cm 0.0 … 999.9 mS/cm 0.000 … 9.999 S/m 0.00 … 99.99 S/m Resistivity 0.00 … 99.99 Mohms · cm Concentration 0.00 … 9.99 % Salinity 0.0 … 45.0 ‰ (0 … 35 °C)

Response time (T90) < 1 s (Sensocheck off) < 3 s (Sensocheck on)

Meas. error1,2,3) < 1 % meas. val. + 0.4 µS · c

Concentration determination -01- NaCl 0.00 ... 9.99% by wt (0 … 60 °C) -02- HCl 0.00 ... 9.99% by wt (–20 … 50 °C) -03- NaOH 0.00 ... 9.99% by wt (0 … 100 °C) -04- H2SO4 0.00 ... 9.99% by wt (–17 … 110 °C) -05- HNO3 0.00 ... 9.99% by wt (-20 … 50 °C)

Sensor standardization Input of cell constant with simultaneous display of conductivity and –temperatureInput of conductivity of calibration solution with simultaneous display of –cell constant and temperatureProduct calibration –Temperature probe adjustment –

Permitted cell constant 0.0050 … 19.9999 cm–!

Sensocheck® Polarization detection and monitoring of cable capacitance

Sensoface® Provides information on the sensor condition (Sensocheck®)

Sensor monitor Direct display of measured values from sensor for validation (resistance/temperature)

Temperature input*} Pt100/Pt1000/NTC 30 kΩ/NTC 8.55 kΩ (Betatherm) 2-wire connection, adjustable

Measuring range Pt 100/Pt 1000 –20.0 … +200.0 °C (–4 … +392 °F) NTC 30 kohms –20.0 … +150.0 °C (–4 … +302 °F) NTC 8.55 kohms –10.0 … +130.0 °C (+14 … +266 °F)

Resolution 0.1 °C / 0,1 °F

Meas. error1,2,3) < 0.5 K (< 1K for Pt100; < 1K for NTC >100°C)

Temperature compensation*} (ref. temperature 25°C)

(OFF) without (Lin) linear characteristic 00.00 … 19.99 %/K (NLF) natural waters to EN 27888 (NaCl) ultrapure water with NaCl traces (0...120°C) (HCl) ultrapure water with HCl traces (0...120°C) (NH3) ultrapure water with NH3 traces (0...120°C)

Stratos® Eco 2405 Cond

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

, Specifications

Page 23: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

23

Output 1 0/4 ... 20 mA, max. 10 V, floating (galvanically connected to output 2)

Measured variable*} Conductivity, resistivity, concentration, salinity

Overrange*} 22 mA in the case of error messages

Output filter *} Low-pass, filter time constant 0 ... 120 s

Measurement error!} < 0.3% current value + 0.05 mA

Start/end of scale*} As desired within range

Minimum span 5 % of selected range

Output 2 0/4 ... 20 mA, max. 10 V, floating (galvanically connected to output 1)

Measured variable Temperature

Overrange*} 22 mA in case of temp error messages

Output filter *} Low-pass, filter time constant 0 ... 120 s

Measurement error!} < 0.3 % current value + 0.05 mA

Start/end of scale*} –20 ... 300 °C / –4 ... 572 °F

Permitted span 20 ... 320 K / 36 ... 576 °F

Alarm contact Relay contact, floating AC < 250 V / < 3 A / < 750 VA DC < 30 V / < 3 A / < 90 W

Contact response N/C (fail-safe type)

Alarm delay*} 10 s

Limit contact Relay contact, floating AC < 250 V / < 3 A / < 750 VA DC < 30 V / < 3 A / < 90 W

Contact response*} N/O or N/C

Delay*} 0000 ... 9999 s

Setpoints*} As desired within range

Hysteresis*} 0 ... 50 % full scale

Specifications (continued)

Page 24: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

24

Display LC display, 7-segment with icons

Main display Character height 17 mm, unit symbols 10 mm

Secondary display Character height 10 mm, unit symbols 7 mm

Sensoface® 3 status indicators (friendly, neutral, sad face)

Mode indication 4 mode indicators "meas", "cal", "alarm", "config" Further icons for configuration and messages

Alarm indication Red LED in case of alarm

Keypad 5 keys: [cal] [conf] [?] [±] [enter]

Service functions

Current source Current specifiable for output 1 and 2 (0.00 ... 22.00 mA)

Device self-test Automatic memory test (RAM, FLASH, EEPROM)

Display test Display of all segments

Last Error Display of last error occurred

Sensor monitor Display of direct, uncorrected sensor signal (resistance / temperature)

Data retention Parameters and calibration data > 10 years (EEPROM)

Protection against electric shock Safe Isolation of all extra-low-voltage circuits against mains by double insulation to EN 61010-1

Power supply 24 (–15 %) ... 230 (+10 %) V AC/DC approx. 5 VA / 2.5 W AC: 45 … 65 Hz Overvoltage category II, Class II

Nominal operating conditions

Ambient temperature –20 … +55 °C

Transport/Storage temp –20 … +70 °C

Relative humidity 10 ... 95 % not condensing

EMC EN 61326-1, EN 61326-2-3 Emitted interference: Class B (residential area) Class A: for mains > 60 V DC Immunity to interference: Industry

Explosion protection FM NI, Class I Div. 2, Groups A,B,C & D T4, Ta = 55 °C; Type 2 NI, Class I Zone 2, Group IIC T4, Ta = 55 °C; Type 2 CSA Class I Div. 2, Groups A, B, C and D, T4 Ex nA IIC T4

Stratos® Eco 2405 Cond

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Specifications (continued)

Page 25: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

25

Enclosure Molded enclosure made from PBT (polybutylene terephtalate)

Color Bluish gray RAL 7031

Assembly – Wall mounting – Pipe mounting ZU 0274 Ø 40 … 60 mm, _ 30 … 45 mm – Panel mounting ZU 0275

cutout to DIN 43 700, sealed against panel

Dimensions H 144 mm, W 144 mm, D 105 mm

Ingress protection IP 65 / NEMA 4X

Cable glands 3 knockouts for cable glands M20x1.5 2 knockouts for 1/2" NPT or rigid metallic conduit

Weight Approx. 1 kg

*) User-defined 1) To IEC 746 Part 1, at nominal operating conditions 2) ±1 count 3) Plus sensor error

Specifications (continued)

Page 26: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

26

• Bypass • Dip

Recommended Installation Sets

Stratos® Eco 2405 CondI

Suitable Sensors:

• A wide variety of European or US electrodeless sensors

• Temperature detectors Pt 100, Pt 1000, NTC 100 kohms

Inductive Conductivity Measurement:

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Page 27: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

27

, Wiring

, Terminal Assignments

, Product LineOrder No.

Stratos® Eco 2405 CondI Process analyzer for measuring conductivity using electrodeless sensors

2405 CondI

Mounting accessories

Pipe-mount kit ZU 0274

Panel-mount kit ZU 0275

Protective hood ZU 0276

For further information, please refer to our »Sensors, Fittings, Accessories« catalog or to www.knick.de.

Conductivity measurement

Control room / station

Voltage supply

Pt 100 / Pt 1000 / NTC 100 kohms

mS, %, SAL Sensocheck®

Temperature

Alarm contact

Limit value

20 ... 253 V AC/DC

mA

mA

0(4) ... 20 mA

0(4) ... 20 mA

mS/cm

Temp

Stratos® Eco

2405 CondI

Voltage supply

Alarm

Limit value

Page 28: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

28

Stratos® Eco 2405 CondI

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

, Specifications

Cond input Input for electrodeless conductivity sensors: SE 655, SE 656, SE 660 and others

Display range Conductivity: 0.000 … 1999 mS/cm Concentration: 0.00 ... 9.99 % by wt Salinity: 0.0 ... 45 ‰ (0 ... 35 °C)

Ranges Conductivity 0.000 ... 9.999 mS/cm 0.00 … 99.99 mS/cm 0.0 … 999.9 mS/cm 0 … 1999 mS/cm 0.000 … 9.999 S/m 0.00 … 99.99 S/m Concentration 0.00 … 9.99 % 10.0 … 100.0 % Salinity 0.0 … 45.0 ‰ (0 … 35 °C)

Response time (T90) Approx. 2 s

Meas. error1,2,3) 1% meas. val. + 0.005 mS

Temperature compensation*} (ref. temperature 25°C)

(OFF) without (LIN) linear characteristic 00.00 … 19.99 %/K (NLF) natural waters to EN 27888 (0 ... 35°C)

Concentration determination -01- NaCl 0.00 ... 9.99 % by wt (0 ... 60 °C) -02- HCl 0.00 ... 9.99 % by wt (–20 ... 50 °C) -03- NaOH 0.00 ... 9.99 % by wt (0 ... 100 °C) -04- H2SO4 0.00 ... 9.99 % by wt (–17 ... 110 °C) -05- HNO3 0.00 ... 9.99 % by wt (–20 ... 50 °C)

Sensor standardization Entry of cell factor with simultaneous display of conductivity and –temperatureEntry of conductivity of calibration solution with simultaneous display of –cell factor and temperatureProduct calibration –Zero adjustment –Temperature probe adjustment –

Permitted cell factor 0.100 ... 19.999

Permitted transfer ratio 1.00 ... 199.99

Permitted zero offset ± 0.5 mS/cm

Sensocheck® – Primary coil and lines monitored for short circuit – Secondary coil and lines monitored for open circuit

Sensoface® Provides information on the sensor condition (evaluation of zero point, Sensocheck®)

Sensor monitor Sensor monitor for validation of sensor and complete measured-value processing (display: resistance / temperature)

Page 29: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

29

Specifications (continued)

Temperature input*} Pt 100/Pt 1000/NTC 100 kohms, 2-wire connection, adjustable

Measuring range Pt 100/Pt 1000 –20.0 … +200.0 °C (–4 … +392 °F) NTC 100 kohms –20.0 … +130.0 °C (–4 … +266 °F)

Resolution 0.1 °C / 0,1 °F

Measurement error1) 2) 3) < 0.5 K (< 1K for Pt100; < 1K for NTC >100°C)

Output 1 0/4 ... 20 mA, max. 10 V, floating (galvanically connected to output 2)

Process variable *} Conductivity, concentration, or salinity

Overrange*} 22 mA in the case of error messages

Output filter*} (attenuation) Low-pass, filter time constant 0 ... 120 s

Measurement error!} < 0.3 % current value + 0.05 mA

Start/end of scale As desired within range

Minimum span 5 % of selected range

Output 2 0/4 ... 20 mA, max. 10 V, floating (galvanically connected to output 1)

Measured variable Temperature

Overrange*} 22 mA in the case of error messages

Output filter*} (attenuation) Low-pass, filter time constant 0 ... 120 s

Measurement error!} < 0.3% current value + 0.05 mA

Start/end of scale*} –20 ... 300 °C / –4 ... 572 °F

Permitted span 20 ... 320 K / 36 ... 576 °F

Alarm contact Relay contact, floating AC < 250 V / < 3 A / < 750 VA DC < 30 V / < 3 A / < 90 W

Contact response N/C (fail-safe type)

Alarm delay 10 s

Limit contact Relay contact, floating AC < 250 V / < 3 A / < 750 VA DC < 30 V / < 3 A / < 90 W

Contact response*} N/C or N/O

Delay*} 0000 ... 9999 s

Setpoints*} As desired within range

Hysteresis*} 0 ... 50 % full scale

Page 30: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

30

Display LC display, 7-segment with icons

Main display Character height 17 mm, unit symbols 10 mm

Secondary display Character height 10 mm, unit symbols 7 mm

Sensoface® 3 status indicators (friendly, neutral, sad face)

Mode indication 4 mode indicators "meas", "cal", "alarm", "config" Further icons for configuration and messages

Alarm indication Red LED in case of alarm

Keypad 5 keys: [cal] [conf] [?] [±] [enter]

Service functions

Current source Current specifiable for output 1 and 2 (00.00 ... 22.00 mA)

Device self-test Automatic memory test (RAM, FLASH, EEPROM)

Display test Display of all segments

Last Error Display of last error occurred

Sensor monitor Display of direct, uncorrected sensor signal

Data retention Parameters and calibration data > 10 years (EEPROM)

Protection against electric shock Safe Isolation of all extra-low-voltage circuits against mains by double insulation to EN 61010-1

Power supply 24 (–15 %) ... 230 (+10 %) V AC/DC approx. 5 VA / 2.5 W AC: 45 … 65 Hz Overvoltage category II, Class II

Nominal operating conditions

Ambient temperature –20 … +55 °C

Transport/Storage temp –20 … +70 °C

Relative humidity 10 ... 95 % not condensing

EMC EN 61326-1, EN 61326-2-3 Emitted interference: Class B (residential area) Class A: for mains > 60 V DC Immunity to interference: Industry

Explosion protection FM NI, Class I Div. 2, Groups A,B,C & D T4, Ta = 55 °C; Type 2 NI, Class I Zone 2, Group IIC T4, Ta = 55 °C; Type 2 CSA Class I Div. 2, Groups A, B, C and D, T4 Ex nA IIC T4

Stratos® Eco 2405 Cond

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Specifications (continued)

Page 31: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

31

Enclosure Molded enclosure made from PBT (polybutylene terephtalate)

Color Bluish gray RAL 7031

Assembly – Wall mounting – Pipe mounting ZU 0274 Ø 40 … 60 mm, _ 30 … 45 mm – Panel mounting ZU 0275

cutout to DIN 43 700, sealed against panel

Dimensions H 144 mm, W 144 mm, D 105 mm

Ingress protection IP 65 / NEMA 4X

Cable glands 3 knockouts for cable glands M20x1.5 2 knockouts for 1/2" NPT or rigid metallic conduit

Weight Approx. 1 kg

*) User-defined 1) To IEC 746 Part 1, at nominal operating conditions 2) ±1 count 3) Plus sensor error

Specifications (continued)

Page 32: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

32

78 27

10527

2143

144

144

4284

1

2

3

32 80

72

132

, Dimension Drawings

App

rox.

14

6.2

M 20 x 1.5 (3 x)

2 dia. 21.5 mm knockouts for M 20 x 1.5 cable glands or 1/2” conduit (not included)

ZU 0274 pipe-mount kitconsisting of:

1 Self-tapping screw (4 x)2 Pipe-mount plate

(1 x)3 Hose clamp with

worm gear drive to DIN 3017 (2 x)

ZU 0276 protective hood ZU 0275 panel-mount kit

Max. 25

Ø 40 ... 60

For vertical or horizontal posts or pipes

1 ... 22 mm

Control panel cutout138 x 138 mm (DIN 43700)

Stratos® Eco

All You Need For Measuring.

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings

Page 33: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

33

ZU 0274 pipe-mount kitFor mounting on vertical or horizontal posts or pipes

ZU 0276 protective hoodAdditional protection against direct weather exposure and mechanical damage

ZU 0275 panel-mount kitFor mounting in standardized panel cutout 144 x 144 mm (DIN 43700), sealed against panel

, Mounting Examples

Page 34: The Price Advantage.ppm-systems.ru/resources/files/stratoseco_prospekt_e.pdf · • Sensocheck® continuous sensor monitoring • SensoFace® preventive maintenance indication •

TP-1

94.1

30-E

0708

onl

ine

Sub

ject

to

chan

ge w

ithou

t no

tice!

Knick Elektronische Messgeräte GmbH & Co. KGBeuckestraße 22, D-14163 Berlin, GermanyPhone: +49 (0)30 - 801 91 - 0 Fax: +49 (0)30 - 801 91 - [email protected]

Isol

atio

n Am

plifi

ers

Tran

smitt

ers

Indi

cato

rs

Proc

ess

Ana

lyti

cs

Port

able

Met

ers

Labo

rato

ry M

eter

s

Sens

ors

Fitt

ings