The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator...

80
temperature calibration equipment & services The Industrial Laboratory Isotech’s Millennium Collection of Products for Industrial Temperature Calibration ISSUE 2

Transcript of The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator...

Page 1: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

temperature calibration equipment & services

TheIndustrial

LaboratoryIsotech’s Millennium

Collection of Productsfor Industrial

Temperature Calibration

ISSU

E 2

Page 2: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

CONTENTSIntroduction 3

Alternative methods of operating an Industrial Laboratory 4-5Selection Chart 6-7

Introduction to Liquid Baths 8-9Hyperion 10-11

Drago 12-13

Introduction to the Isocal-6 Calibration System 14-15Europa-6PLUS 16-17

VenusPLUS 2140 18-19CalistoPLUS 2250 20-21

580 Oceanus-6PLUS 22-23Slim Fixed Point Cells 24-25

Introduction to Dry Block Calibrators 26-27Apollo 1 & 2 28-29

JupiterPLUS 650 30-31GeminiPLUS 550 / 700 32-33

GeminiPLUS 550 / 700 Large Removable Insert 34-35PegasusPLUS 1200 36-37

510 MedusaPLUS 38-39426 Oberon Sodium Heatpipe Furnace 40-41

Introduction to Temperature Testers 42-43Quick-Cal Range 44-45

907 Portable 46903 Portable Temperature Test Unit 47

983 Small Hot Plate Surface Sensor Calibrator 48-49Model 862 24 Value Resistance Thermometer Simulator 50-51

Isocal Thermocouple Simulator 52-53

Indicators, Probes & Accessories Introduction 54-55TT-7 for thermocouples & Pt100’s 56-57

TTI-6 for 3 & 4 wire Pt100’s 58-59TTI accessories 60-61

Semi-standard Resistance Thermometers 62-63Semi-standard Thermocouples 64-65

Semi-standards Documentation 66-678 way Selector Switch 68-69

836 Miniature Fixed Resistor 70-71944 True Surface Temperature Measurement System 72-73

Introduction to Software 74-75Cal NotePad Software 76-77

I-cal Software 78-79

The company is always willing to givetechnical advice and assistance whereappropriate.

Equally because of the program ofcontinual development andimprovement, we reserve the right toamend or alter characteristics anddesign without prior notice.

This publication is for information only.

All registered product names andtrademarks belong to their respectivecompanies.

The print on the front cover shows one ofCount Lösers metallic thermometers.

For more information on the history of thethermometer the reader is referred to:-“A History of the Thermometer and its usein Metrology” by W. E. Knowles Middleton(The John Hopkins Press, Baltimore,Maryland).

Products in this catalogue are all Compliantthree/2

Page 3: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/3

INTRODUCTION: for industrial calibration

DataBook 3 contains Isotech products designed forIndustrial Temperature Calibration. This is an ambitiousclaim given that those faced with the task of calibratingindustrial sensors encounter sensors of widely varyingsize, temperature range, accuracies and type.

In this DataBook calibration solutions exist for accuraciesfrom 1ºC to less than 0.001ºC, temperature ranges frombelow 0ºC to 1200ºC, from handheld to bench topdevices. Products include Dry Block Calibrators, LiquidBaths, High Temperature Furnaces, PrecisionTemperature Indicators, Temperature Sensors withsupporting services, UKAS Calibration and software.

The industrial calibration products have been enhancedand upgraded benefiting from new instrumentation andsuperior performance. PC interfaces and supportingsoftware have been introduced as standard along withother enhancements making the products betterspecified, better value and easier to use.

Benefit from Isotech's premier UKASCalibration Laboratory

Many manufacturers of temperature calibrationequipment do not operate accredited laboratories. AtIsotech we remain unique in that in addition to amanufacturer of equipment we also operate a full scaleUKAS laboratory calibrating both our own equipmentand a full range of temperature equipment for our clients.

Benefit from Experience

Isotech fully understand the calibration needs of ourcustomers. We design and specify products accordingly.Isotech make available full evaluation reports withuncertainty calculations and publish uncertainty graphs.

Dry Blocks from other manufacturers are sometimesspecified in interesting ways, such as Minimum OperatingTemperature -30ºC but with an ambient of 5ºC, Isotechlabs and workshops are kept close to 20ºC! We don'tspecify performances of thermometers only at 0ºC wealso tell you what it is like at high temperatures. Isotechunderstand the difference between resolution andaccuracy, between accuracy and uncertainty, between"specmanship" and the firm evidence an auditor wouldexpect to see.

Benefit from Innovation

The international award winning ISOCAL-6 series set newstandards in temperature calibration, Dry Blocks nowinclude a PC interface and software as standard. TheI-CAL software can automatically capture temperaturestamped images with a camera, another first fromIsotech.

www.isotech.co.uk/industrial

Databook 3Introduction to Industrial Calibration

Page 4: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/4

A BASIC DRY BLOCK CALIBRATORThe thermometer under test is compared to the dry block controller value.Useful for moderate temperature ranges and quick testing.

AN ISO 9000 CALIBRATION SYSTEMA thermometer under test is compared to a calibrated standard,for true traceability and clearly meets the requirements of ISO9000

Many of the products can be used as a Dry Block Calibrator or as a Liquid Bath.

Thermometerunder test

of operating an Industrial Laboratory

Alternative Methods

BLACKBODY SOURCEScompared to standard

SURFACE SENSORcompared to standard

On the Site models the calibrated standard isconnected to an in-built indicator and a UKAScertificate issued for the system.

Thermometerunder test

CalibratedStandard

Page 5: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/5

USING AN EXTERNAL INDICATORSimilar to the previous configuration but an external indicator TTI is used - one TTI can be used with manycalibration baths - the bath or baths do not need a calibration certificate, but they need an evaluation report.

Many of the products can be used as a Dry Block Calibrator or as a Liquid Bath.

Here the B model is shown

Alternatively the S model may beused with the indicator free for usewith a unit under test.

of operating an Industrial Laboratory

Alternative Methods

AN EXAMPLE OF MULTIPLE SENSOR CALIBRATIONYou can add the TTI and Selector Switch for multi-probe calibration - and add I-cal software foraffordable automatic calibration systems.

Page 6: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/6

The Industrial Laboratory

Selection Chart

Isotech Industrial Temperature Sensor Calibration SolutionsIndustrial temperature sensors vary enormously in type, range and requirement, varying temperature ranges, sizesand accuracies. Isotech have developed a diverse range of equipment to meet these varying and frequentlydemanding requirements.

Testers

QuickCalsHandheld devices covering low to

moderate temperature ranges.These units are ideal for fast, portable

testing of temperature sensors.

pages three/44

903 and 907These models have larger calibration

volumes and a wider temperature rangebut retain portability

pages three/46

1

Portable Liquid Baths2

Drago andHyperion

Portable Liquid Baths have a large wellwhich is filled with a liquid avoiding the

need for specially drilled blocks.They are ideal for awkward shaped andshort sensors, accuracies are generally

better than for Dry Blocks. IsotechPortable Liquid Baths are unique in that

they have accessories for other modes ofuse, including Dry Block operation

pages three/10

Page 7: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/7

The Industrial Laboratory

Selection ChartISOCAL-6 Range

Europa, Venus,Calisto and

OceanusIsocal-6 is the flexible choice, this award

winning series covers a wide temperaturerange with a choice of two calibration volume

sizes. Uniquely the Isocal-6 can be used as aDry Block, a stirred Liquid Bath, for fixed point

calibration, infrared thermometer calibration,and surface sensor calibration.

pages three/14

3

Dry Block Calibrators4

Apollo, Jupiter,Gemini, Pegasus,Medusa, OberonThe Dry Block Calibrators complement the

Isocal-6 range, but also offer highertemperature ranges. Above temperature

ranges of approximately 250°C it is notpractical to use stirred liquids due to fumes,

risk of ignition and safety considerations.Isotech’s higher temperature calibration baths

incorporate as many of the Isocal-6 optionsthat are safe and practical to provide. A wide

range of calibration volumes are available.

pages three/26

Temperature Indicators, ReferenceThermometers and Software

5

These electronic temperature indicators andreference probes, both platinum resistance

thermometers and thermocouples, are ideal formost industrial calibration requirements.

Switchboxes allow for multiple channelcalibration and supporting software allows for

automatic temperature calibration.

pages three/56

Page 8: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/8

Liquid Baths:Temperature Range -25°C to 250°C

Portable Stirred Liquid Baths have a large well which isfilled with liquid. The liquid is stirred and heated orcooled to the desired temperature.

The temperature sensors are placed directly into theliquid avoiding the need for specially drilled blocks. Theyare ideal for awkward shaped and short sensors. Angledprobes will not readily fit into a metal block but can beplaced into the liquid.

Accuracies are in general better than dry block baths dueto lack of air gaps and the temperature uniformity of thestirred liquid.

Liquid Bath Advantages

Calibrate awkward shaped sensors.

High Accuracy.

Excellent temperature uniformity.

Sensors placed into liquid, no specially drilled insertsrequired.

Portable Liquid BathsTo calibrate temperature sensors to 250°C look nofurther than these portable stirred liquid baths. TheHyperion calibrates from -25°C (At ambient of 20°C) to140°C and the Drago from 30 to 250°C. The calibrationvolume is 65mm diameter and the useable depth of160mm gives more than twice the volume of alternativeproducts.

Stirred liquid baths are suitable for temperature sensorsof all types, shapes and sizes, angled probes, bentprobes and short probes with large mounting heads canall be readily accommodated, an advantage over DryBlock Calibrators.

If a laboratory standard temperature indicator is addedmuch greater accuracies than those from Dry Blocksalone can be achieved and with suitable referencethermometers performance of up to 0.005°C isachievable.

These models include a high accuracy inbuilt digitaltemperature indicator and a reference probe. Thisindependent measurement system is used as a referenceto which the thermometers under test are compared.This method gives good accuracy largely eliminatingtemperature gradient and loading errors and providestraceable temperature calibration.

Introduction to Liquid BathsLiquid Baths

Page 9: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/9

Introduction to Liquid BathsLiquid Bath Accessories

Thermometer Support Kit

Supplied as standard.

Simply place the probes in thebasket.

Part Number 936-06-02

Lid

Supplied as standard.

Seals the bath and prevents spillswhilst the unit is being moved.

Part Number 936-02-14

Support Kit

Optional.

Use in place of the basket tosuspend up to eight probes witha diameter range of 5 - 8mm.Useful with liquid in glassthermometers as well as RTDsand TCs. Fits both directly intothe block or into the optionalliquid containers.

Part Number 936-06-08

Metal Insert

Optional.

A Metal Insert can be placed intothe calibration volume to convertfor Dry Block use. The standardblock has eight 8mm pocketsand two 4.5mm pockets all157mm deep. Blank andspecially drilled inserts areavailable.

Standard Insert Part Number936-06-01

Liquid Baths

Page 10: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/10

140ºCto

45ºCbelow

ambient

936 Hyperion PLUS

To calibrate temperature sensors to 250°C look no further than Isotech’s portable stirred liquid baths. TheHyperion calibrates from 45°C below ambient to 140°C and the Drago from 30 to 250°C see page 12. Thelarge calibration volume is more than twice the size of alternative products. The useable area is a full 65mmdiameter with a depth of 160mm, the overall well depth of 190mm allows room for space at the top andthe bottom magnetic stirrer.

Stirred liquid baths are suitable for temperature sensors of all types, shapes and sizes. Accuracies are muchgreater than those from Dry Blocks alone and with suitable reference thermometers performance of up to0.005°C is achievable.

The Hyperion is available in two models. If the liquid is directly in the block then the controller only model,or Basic (B) model, can be selected. This model is also suitable where an external indicator and standardwill be used. Alternatively the site model (S) includes a built in temperature indicator for high accuracy orfor best accuracy an external indicator can be used, an ideal combination is a TTI model and a 935-14-16Standard Probe, see Indicators, Probes and Accessories section. The S model can be provided with UKAScertification.

The Hyperion can also be used with the supplied Cal NotePad software to automatically calibratethermostats.

When using a separate indicator and probe then different accessories can be added for Dry Block,Blackbody, Surface Sensor, Liquid Containers and even ITS-90 fixed point operation.

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature.Increased resolution of ±0.01 available throughout the range via the PC interface and from -19.99to +99.99 locally on the auto-ranging front display. The controller features multi-point block todisplay correction giving good absolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers,Thermocouples (types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including4-20mA current transmitters to be displayed on the inbuilt indicator. The indicator can beprogrammed with up to five calibration points to provide high accuracy digital probe matching.The indicator and controller are both addressable over the communications link.

New! The Site model can nowbe used with the supplied Cal NotePadsoftware to test thermostats.

Features

• 65mm Diameter Calibration Volume.

• Portable Liquid Bath for high accuracycalibration of awkward shaped sensors.

• Convertible for Dry Block Operationand more.

• Calibrate all Sensor types.

• Windows Software and PC Interface asstandard.

Range®

Page 11: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/11

Options HyperionPLUS Performance - Dry Block

936 Hyperion PLUS

The company is always willing to give technical advice and assistance whereappropriate. Equally because of the program of continual development and

improvement, we reserve the right to amend or alter characteristics and designwithout prior notice. This publication is for information only.

To see the ISOCAL-6 in use ask for our27 minute video supplied on CD-Rom

Note: Instead of putting liquids directly in the block liquidcontainers can be used to facilitate rapid change of liquids.When using a liquid container, Dry Block Insert, Blackbody Targetor the Surface Sensor Kit a separate reference themometer shouldbe used to compensate for the varying offset between thecontroller and the accessory temperature.Suitable choices include the SITE model with probe or a standalone indicator such as an Isotech TTI model.

Model No. HYPERION PLUS

Temperature 45 below ambient to +140ºCRange (absolute minimum -45ºC)

Absolute Stirred Liquid Bath ±0.025ºCstability over Dry Block Bath ±0.03ºC

30 minutes Ice / Water Bath ±0.001ºCBlackbody Source ±0.3ºCSurface Sensor Calibrator ±0.5ºCITS-90 Fixed Point Apparatus ±0.0005ºC

Computer Interface Included with Windows Software

Thermal As a liquid comparison bathPerformance Uniformity down to ±0.005ºC

over the full range

Uncertainties Refer to Uncertainties Graph

Calibration volume 65mm diameter by l6Omm deep

Display Resolution 0.01 -19.99 to 99.990.1 -55.0 to -20.0 and 100.0 to 140.0

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K

Power 108 to 120V (50/60 Hz) or208 to 240V (50/60 Hz)200 Watts

OveralI dimensions Height 302mmWidth 176mmDepth 262mm

Weight 12kg

How to Order Hyperion PLUS

Please specify model type requiredPlease specify voltage requiredPlease specify options required

Stirred Liquid Bath 936-06-02 Includes a container,Water / Ice Bath magnetic stirrer and

probe guide

Metal Block Insert 936-06-01a Standard Insert8 x 8mm + 2 x 4.5mmdiameter holes 157mmdeep

936-06-01b Blank Insert

936-06-01c Special Insert.Contact Isotech withyour requirements

Blackbody Target 936-06-03 Use with StandardProbe (935-14-61DB)

Surface Sensor Kit 936-06-04 Includes an Insert andan angled PRT.

Fixed Point Cells D8 Water Slim Cell17401 Slim Gallium Slim Cell

Standard Probe 935-14-61DB Platinum ResistanceThermometer

UKAS Calibration UKAS Calibration available to Order

Carrying Case 931-22-65 Sturdy caseaccommodates the unit

Liquid 936-06-07 1 Litre of C10 Oil(-35°C to +140°C)

Thermometer Holder 936-06-08 Fits into the top of theblock or liquidcontainer and supportsup to eightthermometers into theliquid. Suits probes of5mm to 8mm indiameter.

Range®

Page 12: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/12

250ºCto

30ºC

Range934 DRAGO PLUS

To calibrate temperature sensors to 250°C look no further than Isotech’s portable stirred liquid baths. TheDrago calibrates from 30 to 250°C and the Hyperion from 45°C below ambient to 140°C, see page 10. Thelarge calibration volume is more than twice the size of alternative products. The useable area is a full 65mmdiameter with a depth of 160mm, the overall well depth of 190mm allows room for space at the top andthe bottom magnetic stirrer.

Stirred liquid baths are suitable for temperature sensors of all types, shapes and sizes. Accuracies are muchgreater than those from Dry Blocks alone and with suitable reference thermometers performance of up to0.005°C is achievable.

The Drago is available in two models. If the liquid is directly in the block then the controller only model, orBasic (B) model, can be selected. This model is also suitable where an external indicator and standard willbe used. Alternatively the site model (S) includes a built in temperature indicator for high accuracy or forbest accuracy an external indicator can be used, an ideal combination is a TTI model and 935-14-16Standard Probe, see Indicators, Probes and Accessories section. The S model can be provided with UKAScertification.

The Drago can also be used with the supplied Cal NotePad software to automatically calibrate thermostats.

When using a separate indicator and probe then different accessories can be added for Dry Block,Blackbody, Surface Sensor, Liquid Containers and even ITS-90 fixed point operation.

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature.Increased resolution of ±0.01 available throughout the range via the PC interface and from 0.01to +99.99 locally on the auto-ranging front display. The controller features multi-point block todisplay correction giving good absolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers,Thermocouples (types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including4-20mA current transmitters to be displayed on the inbuilt indicator. The indicator can beprogrammed with up to five calibration points to provide high accuracy digital probe matching.The indicator and controller are both addressable over the communications link.

NEW! The Site model can now be used with the supplied Cal NotePad software totest thermostats.

Features

• 65mm Diameter Calibration Volume.

• Portable Liquid Bath for high accuracycalibration of awkward shaped sensors.

• Convertible for Dry Block Operationand more.

• Calibrate all Sensor types.

• Windows Software and PC Interface asstandard.

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

®

Page 13: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/13

Range934 DRAGO PLUS

Options DragoPLUS Performance - Dry Block

The company is always willing to give technical advice and assistance whereappropriate. Equally because of the program of continual development and

improvement, we reserve the right to amend or alter characteristics and designwithout prior notice. This publication is for information only.

To see the ISOCAL-6 in use ask for our27 minute video supplied on CD-Rom

Note: Instead of putting liquids directly in the block liquidcontainers can be used to facilitate rapid change of liquids.When using a liquid container, Dry Block Insert, Blackbody Targetor the Surface Sensor Kit a separate reference themometer shouldbe used to compensate for the varying offset between thecontroller and the accessory temperature.Suitable choices include the SITE model with probe or a standalone indicator such as an Isotech TTI model.

Model No. DRAGO PLUS

Temperature + 30ºC to + 250ºCRange in an ambient of 25ºC or below

Absolute Stirred Liquid Bath ±0.025ºCstability over Dry Block Bath ±0.03ºC

30 minutes Ice / Water Bath ±0.001ºCBlackbody Source ±0.3ºCSurface Sensor Calibrator ±0.5ºCITS-90 Fixed Point Apparatus ±0.0005ºC

Computer Interface Included with Windows Software

Thermal As a liquid comparison bathPerformance Uniformity down to ±0.005ºC

over the full range

Calibration volume 65mm diameter by l6Omm deep

Display Resolution 0.01 Up to 99.990.1 100.0 to 250.0ºC

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K

Power 100 to 120V (50/60 Hz) or200 to 240V (50/60 Hz)1000 Watts

OveralI dimensions Height 302mmWidth 176mmDepth 262mm

Weight 8kg

How to Order Drago PLUS

Please specify model type requiredPlease specify voltage requiredPlease specify options required

Stirred Liquid Bath 936-06-02 Includes a container,Water / Ice Bath magnetic stirrer and

probe guide

Metal Block Insert 936-06-01a Standard Insert8 x 8mm + 2 x 4.5mmdiameter holes 157mmdeep

936-06-01b Blank Insert

936-06-01c Special Insert.Contact Isotech withyour requirements

Blackbody Target 936-06-03 Use with StandardProbe (935-14-61DB)

Surface Sensor Kit 936-06-04 Includes an Insert andan angled PRT.

Fixed Point Cells 17401 Slim Gallium Slim Cell

Standard Probe 935-14-61DB Platinum ResistanceThermometer

UKAS Calibration UKAS Calibration available to Order

Carrying Case 931-22-65 Sturdy caseaccommodates the unit

C20 Oil 580-06-09 1 Litre of C20 Oil (fromambient to + 200ºC)

Very High Temp. Oil 915/09 1 Litre of V.H.T. Oil(from + 150ºC to 250ºC)

Thermometer Holder 936-06-08 Fits into the top of theblock or liquid containerand supports up toeight thermometers intothe liquid. Suits probesof 5mm to 8mm indiameter.

®

Page 14: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/14

Introduction to the Isocal-6 Calibration System

®

What is Isocal-6?

Isocal-6 is a range of calibration systems that cancalibrate all types of thermometer and share commonfeatures.

What is special about the Isocal-6 range?

Previously temperature calibration baths werespecified by type, e.g. A Dry Block Bath, A Liquid Bathetc.

With the Isocal-6 different accessories are added to theIsocal-6 to allow it to be used in many different modesproviding a solution for all temperature sensorcalibration requirements.

What are these different modes?

The Isocal-6 includes a removable insert as standard foruse as a Dry Block Calibrator, with optional accessoriesthe Isocal-6 can perform as:

A Stirred Liquid Bath

A Stirred Ice Bath 0ºC Reference (Cooling models only)

A Blackbody Source for IR thermometers

A Calibrator for Surface Sensors

An ITS-90 Fixed Point Apparatus

What if only one option is required?

One of the advantages is that you need only purchasethe desired option but other accessories can be addedat any time.

What temperature range is covered?

Different models cover the temperature range frombelow -40ºC to 250ºC, above this maximumtemperature it is not practical to use a stirred liquidbath. For higher temperatures see the other Dry BlockCalibrators that go to 1200ºC, many with accessoriesfor Surface Sensor and Blackbody use.

What Performance can be achieved?

This varies on the way it is used, for an ITS-90 fixedpoint apparatus uncertainties of better than 0.001ºCcan be achieved - see the uncertainty graphs for othermodes of operation.

International Support?

Isotech have a global network of authorised agents anddistributors providing both before and after salessupport.

Made to last?

All Isotech products are engineered for long life. If youhave ever tried to get inside the dashboard of a motorcar you will appreciate that cheapness goes withinaccessibility for repair and maintenance. Isotech’sproducts are designed for long life, easy repair,accessibility and maintenance.

The ISOCAL-6 is currently subject to a patentapplication (reference 9900158.8).

Page 15: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/15

Introduction to the Isocal-6 Calibration System

®

1

2 3

4 5 6

The ISOCAL-6 is the only calibration system in the World (patent application 9900158.8) to offer you sixcalibration baths in a single simple product.

Metal Block Bath

Stirred Liquid Bath

Stirred Ice/Water Bath

Black Body Source

Surface Sensor Calibrator

ITS-90 Fixed Point Apparatus

This schematicshows the six

options availablefor the

range ofcalibration baths.

®

®

1

2

3

4

5

6

Page 16: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/16

140ºCto

65ºCbelow

ambient

RangeEuropa-6PLUS

®

The EUROPAPLUS offers an exceptionally wide operating range in an easy to use portable package. Aspart of the ISOCAL-6 series the EUROPAPLUS can be used as a Dry Block, a Liquid Bath, a BlackbodySource for infrared thermometers, a Surface Sensor Calibrator and for performance to a few mK(0.001ºC) ITS-90 Fixed Points. The EUROPAPLUS is available in two models, the BASIC (B) and the SITE(S). The B model includes a sophisticated temperature controller with a dual display for SetTemperature and Dry Block Temperature, refer to page 27 for further details. The S model additionallyincludes an in-built digital indicator to which an external temperature sensor can be connected.

For Dry Block use the unit under test may be compared to the controller value, Isotech recommendthe use of an external probe connected to the indicator of the Site model (see page 23) whichcompensates for the temperature gradient and loading errors. This gives greater accuracy andtraceability meeting the requirements of ISO 9000 and other quality standards.

For liquid bath, surface sensor and blackbody use an external standard probe should always be used,either with the S models indicator or an external indicator, e.g. an Isotech TTI model. For laboratoryuse the EUROPAPLUS can be used with a laboratory performance temperature indicator such as one ofthe Isotech True Temperature Indicators (TTI).

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature.Increased resolution of ±0.01 available throughout the range via the PC interface and from-19.99 to +99.99 locally on the auto-ranging front display. The controller features multi-pointblock to display correction giving good absolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers,Thermocouples (Types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including4-20mA current transmitters to be displayed on the inbuilt indicator. The indicator can beprogrammed with up to five calibration points to provide high accuracy digital probe matching.The indicator and controller are both addressable over the communications link.

New! The Site model can now be used with the supplied Cal Notepad software to testthermostats.

CellsPages 24 - 25

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

Key Features

Extreme Low Temperature Operation-45ºC in an ambient of 20ºC - Use for Triple Pointof Mercury, 0ºC, 121ºC (250ºF), 135ºC.

Realise ITS-90 Fixed Points -the EUROPAPLUS can be used with Slim Mercury,Gallium and Triple Point of Water Fixed PointCells.

Calibrate as a Dry or Liquid Bath,thermocouples, platinum resistancethermometers, thermistors, liquid in glass,thermostats, semiconductor types.

S Model Includes Universal Input TemperatureIndicator allowing for up to five “correction”points to be programmed.

Windows Software and PC Interface asStandard.

Free Evaluation Report - ask for full data or visitwww.isotech.co.uk/industrial

Page 17: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/17

RangeEuropa-6PLUS

®1. Metal Block Insert 951-02-15 Standard InsertIncluded

951-02-15A Blank InsertInsert without pockets forlocal machining

951-02-15C Special InsertContact Isotech with yourrequirements

2. Stirred Liquid Bath 951-06-01 Includes a container,3. Water/Ice Bath magnetic stirrer and

probe guideThermometer 951-06-03 Allows three thermometersSupport Kit to be suspended in the

bath, including liquid inglass types

Oil 951-06-06 0.1L20ºC to 140ºC

4. Blackbody Target 951-06-04 Use with Standard Probe(935-14-82)

5. Surface Sensor Kit 951-06-02 Includes an Insert and anangled PRT.

6. Fixed Point Cells 17724 Slim Mercury Slim CellD8 Water Slim Cell17401 Slim Gallium Slim Cell

Standard Probe 935-14-82 Platinum ResistanceThermometer

UKAS Calibration UKAS Calibration available to Order

Carrying Case 931-22-64 Sturdy caseaccommodates the unitwith room for accessories

Calibration and Uncertainty

A certificate, traceable to National Standards, is included as standard.Recommended is an optional UKAS five-point calibration.

The accuracy will depend very much on the mode of use and the typesof sensor to be used. Please contact Isotech for tutorials anduncertainty calculations and comprehensive evaluation reports. TheEuropaPLUS meets the Calibration Capacity requirements of EA I0/13,“EA Guidelines on the Calibration of Temperature Block Calibrators”.

Options

Model No. EUROPAPLUS

Temperature 65ºC below ambient to +140ºCRange (absolute minimum -55ºC)

Absolute stability Dry Block Bath ±0.03ºCover 30 minutes Stirred Liquid Bath ±0.025ºC

Ice/Water Bath ±0.001ºCBlackbody Source ±0.3ºCSurface Sensor Calibrator ±0.5ºCITS-90 Fixed Point Apparatus ±0.0005ºC

Computer Interface Included with Windows Software (page 74)

Cools from 140ºC to 0ºC in 15 minutes

Heats from -30ºC to 140ºC in 15 minutes

Best Performance See Graph

Calibration volume 35mm diameter by 160mm deep

Standard Insert 6 pockets, 9.5mm, 8.0mm, 6.4mm,6.4mm, 4.5mm, 4.5mm, all 157mm deep

Uniformity ±0.018ºC

Display Resolution 0.01 -19.99 to 99.990.1 -55.0 to -20.0 and 100.0 to 140.0

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K

Power 100 to 120V (50 / 60 Hz) or200 to 240V (50 / 60 Hz)300 Watts

Overall dimensions Height 322mmWidth 176mmDepth 262mm

Weight 14kg

How to Order EUROPA-6PLUS

Please specify model type requiredPlease specify voltage requiredPlease specify options required

Alternative Metal Block Inserts

951-06-07 Standard Insert type B13mm , 10mm, 8mm, 5mmand 3.5mm diameter holes,all 157mm deep

951-06-08 Special Insert type C 8mm,6 x 6.5mm diameter holes,all 157 deep

To see the ISOCAL-6 in use ask for our27 minute video supplied on CD-Rom

Page 18: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/18

140ºCto

55ºCbelow

ambient

RangeVenusPLUS 2140

®

The VenusPLUS2140 range offers outstanding performance in an easy to use portable package. As partof the ISOCAL-6 series the VenusPLUS2140 can be used as a Dry Block, a Liquid Bath, a BlackbodySource for infrared thermometers, a Surface Sensor Calibrator and for performance to a few mK(0.001ºC) ITS-90 Fixed Points. The Venus is available in two models, the BASIC (B) and the SITE (S).The B model includes a sophisticated temperature controller with a dual display for Set Temperatureand Dry Block Temperature, refer to page 27 for further details. The S model additionally includes anin-built digital indicator to which an external temperature sensor can be connected.

For Dry Block use the unit under test may be compared to the controller value, Isotech recommendthe use of an external probe connected to the indicator of the Site model (see page 23) whichcompensates for the temperature gradient and loading errors. This gives greater accuracy andtraceability meeting the requirements of ISO 9000 and other quality standards

For Liquid Bath, Surface Sensor or Blackbody use an external thermometer should always be used -either with the S model or the B model and a separate stand-alone indicator. For laboratory use theVenus can be used with a laboratory performance temperature indicator such as one of the IsotechTrue Temperature Indicators (TTI).

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature.Increased resolution of ±0.01 available throughout the range via the PC interface and from-19.99 to +99.99 locally on the auto-ranging front display. The controller features multi-pointblock to display correction giving good absolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers,Thermocouples (Types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including4-20mA current transmitters to be displayed on the inbuilt indicator. The indicator can beprogrammed with up to five calibration points to provide high accuracy digital probe matching.The indicator and controller are both addressable over the communications link.

New! The Site model can now be used with the supplied Cal Notepad software to testthermostats.

Key Features

Calibrate Whole Measurement Loop -using a heat source rather than an electricalsimulator a test instrument and probe can becalibrated as a system.

Calibrate as a Dry or Liquid Bath,thermocouples, platinum resistancethermometers, thermistors, liquid in glass,thermostats, semiconductor types.

Covers key calibration points from below toabove ambient, 0ºC, 121ºC (250ºF), 135ºC.

Simple To Use - Outstanding Value for Money!

S Model Includes Universal Input TemperatureIndicator allowing for up to five “correction”points to be programmed.

Windows Software and PC Interface asStandard.

Free Evaluation Report - ask for full data or visitwww.isotech.co.uk/industrial.

CellsPages 24 - 25

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

Page 19: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/19

RangeVenusPLUS 2140

®

2. Stirred Liquid Bath 951-06-01 Includes a container,3. Water/Ice Bath magnetic stirrer and

probe guideThermometer 951-06-03 Allows three thermometersSupport Kit to be suspended in the

bath, including liquid inglass types

951-06-06 0.1L of Oil20ºC to 140ºC

4. Blackbody Target 951-06-04 Use with Standard Probe(935-14-82)

5. Surface Sensor Kit 951-06-02 Includes an Insert and anangled PRT.

6. Fixed Point Cells D8 Water Slim Cell17401 Slim Gallium Slim Cell

Standard Probe 935-14-82 Platinum ResistanceThermometer

UKAS Calibration UKAS Calibration available to Order

Carrying Case 931-22-64 Sturdy caseaccommodates the unitwith room for accessories

Calibration and Uncertainty

A certificate, traceable to National Standards, is included as standard.Recommended is an optional UKAS five-point calibration.

The accuracy will depend very much on the mode of use and the typesof sensor to be used. Please contact Isotech for tutorials and uncertaintycalculations and comprehensive evaluation reports. The VenusPLUS 2140meets the Calibration Capacity requirements of EA I0/13, “EAGuidelines on the Calibration of Temperature Block Calibrators”.

Options

Model No. VENUSPLUS 2140

Temperature 55ºC below ambient to +140ºCRange (absolute minimum -55ºC)

Absolute stability Dry Block Bath ±0.03ºCover 30 minutes Stirred Liquid Bath ±0.025ºC

Ice/Water Bath ±0.001ºCBlackbody Source ±0.3ºCSurface Sensor Calibrator ±0.5ºCITS-90 Fixed Point Apparatus ±0.0005ºC

Computer Interface Included with Windows Software (page 74)

Cools from 140ºC to 0ºC in 15 minutes

Heats from -30ºC to 140ºC in 15 minutes

Best Performance See Graph

Calibration volume 35mm diameter by 160mm deep

Standard Insert 6 pockets, 9.5mm, 8.0mm, 6.4mm,6.4mm, 4.5mm, 4.5mm, all 157mm deep

Uniformity ±0.018ºC

Display Resolution 0.01 -19.99 to 99.990.1 -55.0 to -20.0 and 100.0 to 140.0

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K

Power 100 to 120V (50 / 60 Hz) or200 to 240V (50 / 60 Hz)150 Watts

Overall dimensions Height 302mmWidth 176mmDepth 262mm

Weight 10.2kg

How to Order VenusPLUS 2140Please specify model type requiredPlease specify voltage requiredPlease specify options required

To see the ISOCAL-6 in use ask for our27 minute video supplied on CD-Rom

1. Metal Block Insert 951-02-15 Standard InsertIncluded

951-02-15A Blank InsertInsert without pockets forlocal machining

951-02-15C Special InsertContact Isotech with yourrequirements

Alternative Metal Block Inserts

951-06-07 Standard Insert type B13mm , 10mm, 8mm, 5mmand 3.5mm diameter holes,all 157mm deep

951-06-08 Special Insert type C 8mm,6 x 6.5mm diameter holes,all 157 deep

Page 20: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/20

250ºCto 30ºC

RangeCalistoPLUS 2250

®

The CalistoPLUS2250 range offers outstanding performance in an easy to use portable package. As partof the ISOCAL-6 series the CalistoPLUS2250 can be used as a Dry Block, a Liquid Bath, a BlackbodySource for infrared thermometers, a Surface Sensor Calibrator and for performance to a few mK(0.001ºC) ITS-90 Fixed Points. The CalistoPLUS2250 is available in two models, the BASIC (B) and theSITE (S). The B model includes a sophisticated temperature controller with a dual display for SetTemperature and Dry Block Temperature, refer to page 27 for further details. The S model additionallyincludes an in-built indicator to which an external temperature sensor can be connected.

For Dry Block use the unit under test may be compared to the controller value, Isotech recommendthe use of an external probe connected to the indicator of the Site model (see page 23) whichcompensates for the temperature gradient and loading errors. This gives greater accuracy andtraceability meeting the requirements of ISO 9000 and other quality standards.

For Liquid Bath, Surface Sensor or Blackbody use an external thermometer should always be used -either with the S model or the B model and a separate stand-alone indicator. For laboratory use theCalistoPLUS2250 can be used with a laboratory performance temperature indicator such as one of theIsotech True Temperature Indicators (TTI).

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature.Increased resolution of ±0.01 available throughout the range via the PC interface and from-19.99 to +99.99 locally on the auto-ranging front display. The controller features multi-pointblock to display correction giving good absolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers,Thermocouples (Types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including4-20mA current transmitters to be displayed on the inbuilt indicator. The indicator can beprogrammed with up to five calibration points to provide high accuracy digital probe matching.The indicator and controller are both addressable over the communications link.

New! The Site model can now be used with the supplied Cal Notepad software to testthermostats.

CellsPages 24 - 25

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

Key Features

Calibrate Whole Calibration Loop -using a heat source rather than an electricalsimulator a test instrument and probe can becalibrated as a system.

Calibrate as a Dry or Liquid Bath,thermocouples, platinum resistancethermometers, thermistors, liquid in glass,thermostats, semiconductor types.

Covers key calibration points37ºC, 121ºC (250ºF), 135ºC.

Simple To Use - Outstanding Value for Money!

S Model Includes Universal Input TemperatureIndicator allowing for up to five “correction”points to be programmed.

Windows Software and PC Interface asStandard.

Free Evaluation Report - ask for full data or visitwww.isotech.co.uk/industrial

Page 21: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/21

RangeCalistoPLUS 2250

®

2. Stirred Liquid Bath 951-06-01 Includes a container,magnetic stirrer andprobe guide

Thermometer 951-06-03 Allows three thermometersSupport Kit to be suspended in the

bath, including liquid inglass types

C20 Oil 951-06-06 0.1L (Ambient to +200ºC)Oil 953-04-01 0.1L (+150ºC to +250ºC)

4. Blackbody Target 951-06-04 Use with Standard Probe(935-14-82)

5. Surface Sensor Kit 951-06-02 Includes an Insert and anangled PRT.

6. Fixed Point Cells 17401 Slim Gallium Slim Cell17156M Indium Slim Cell

Standard Probe 935-14-82 Platinum ResistanceThermometer

UKAS Calibration UKAS Calibration available to Order

Carrying Case 931-22-64 Sturdy caseaccommodates the unitwith room for accessories

Calibration and Uncertainty

A certificate, traceable to National Standards, is included as standard.Recommended is an optional UKAS five-point calibration.

The accuracy will depend very much on the mode of use and the types ofsensor to be used. Please contact Isotech for tutorials and uncertaintycalculations and comprehensive evaluation reports. The CalistoPLUS 2250meets the Calibration Capacity requirements of EA I0/13, “EA Guidelineson the Calibration of Temperature Block Calibrators”.

Model No. CALISTOPLUS 2250

Temperature +30ºC to +250ºCRange

Absolute stability Dry Block Bath ±0.03ºCover 30 minutes Stirred Liquid Bath ±0.025ºC

Blackbody Source ±0.3ºCSurface Sensor Calibrator ±0.5ºCITS-90 Fixed Point Apparatus ±0.0005ºC

Computer Interface Included with Windows Software (page 74)

Cools from 250ºC to Ambient in 25 minutes

Heats from Ambient to 250ºC in 25 minutes

Best Performance See Graph

Calibration volume 35mm diameter by 160mm deep

Standard Insert 6 pockets, 9.5mm, 8.0mm, 6.4mm,6.4mm, 4.5mm, 4.5mm, all 157mm deep

Uniformity ±0.018ºC

Display Resolution 0.01 Up to 99.990.1 100.0 to 250.0ºC

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K

Power 100 to 120V (50 / 60 Hz) or200 to 240V (50 / 60 Hz)300 Watts

Overall dimensions Height 302mmWidth 176mmDepth 262mm

Weight 8kg

How to Order CalistoPLUS 2250Please specify model type requiredPlease specify voltage requiredPlease specify options required

Options

To see the ISOCAL-6 in use ask for our27 minute video supplied on CD-Rom

1. Metal Block Insert 951-02-15 Standard InsertIncluded

951-02-15A Blank InsertInsert without pockets forlocal machining

951-02-15C Special InsertContact Isotech with yourrequirements

Alternative Metal Block Inserts

951-06-07 Standard Insert type B13mm , 10mm, 8mm, 5mmand 3.5mm diameter holes,all 157mm deep

951-06-08 Special Insert type C 8mm,6 x 6.5mm diameter holes,all 157 deep

Page 22: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/22

140ºCto

-25ºC

Range580 Oceanus-6PLUS

®

The Oceanus-6PLUS has all the advantages of the Isocal-6 models but with a substantially largercalibration volume, 52mm diameter by 300mm deep. The Oceanus-6 can be used as a Dry Block, aLiquid Bath, a Blackbody Source for infra red thermometers, a Surface Sensor Calibrator and forperformance to a few mK (0.001ºC) ITS-90 Fixed Points. The Oceanus-6PLUS is available in two models,the BASIC (B) and the SITE (S). The B model includes a sophisticated temperature controller with adual display for Set Temperature and Dry Block Temperature, refer to page 27 for further details.

The S model includes a built in digital thermometer to which an external standard thermometer canbe connected, for Dry Block use this will give greater accuracy eliminating temperature gradient andloading errors.

For Liquid Bath, Surface Sensor or Blackbody use an external thermometer should always be used -either with the S model or the B model and a separate stand-alone indicator. For Lab use theOceanus-6PLUS can be used with a laboratory performance temperature indicator such as one of theIsotech True Temperature Indicators with performance, for similar sensors, down to hundredths of adegree. The Oceanus-6PLUS offers unprecedented accuracies of ±0.0002ºC (2 Sigma) at the Water triplepoint and the Gallium melt temperature of 29.7646ºC and up to ±0.005ºC in the stirred liquid bathoption (by comparison).

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature.Increased resolution of ±0.01 available throughout the range via the PC interface and from-19.99 to +99.99 locally on the auto-ranging front display. The controller features multi-pointblock to display correction giving good absolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers,Thermocouples (Types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including4-20mA current transmitters to be displayed on the inbuilt indicator. The indicator can beprogrammed with up to five calibration points to provide high accuracy digital probe matching.The indicator and controller are both addressable over the communications link.

New Improved Temperature Range and the Site model can now be used with the suppliedCal Notepad software to test thermostats. Cells

Pages 24 - 25

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

Key Features

S Model Includes Universal Input TemperatureIndicator allowing for up to five "correction"points to be programmed.

Realise ITS-90 Fixed Points - the Oceanus-6PLUS

can be used with the Gallium and Triple Point ofWater Fixed Point Cells.

Calibrate as a Dry or Liquid Bath,thermocouples, platinum resistancethermometers, thermistors, liquid in glass,thermostats, semiconductor types.

Windows Software and PC Interface asStandard.

High Capacity Deep Block50mm diameter x 300mm Deep.

Use for Comparison and Fixed PointCalibration.

Use with very long thermometers.

Page 23: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/23

Range580 Oceanus-6PLUS

®

1. Metal Block Insert 580-06-03 Standard InsertIncluded

580-06-04 Blank InsertInsert without pockets forlocal machining

580-06-05 Special InsertContact Isotech with yourrequirements

580-06-06 Adjustable EqualisingBlock

2. Stirred Liquid Bath 580-06-07 Includes a container,magnetic stirrer andprobe guide

Thermometer 580-06-00 Allows three thermometersSupport Kit to be suspended in the

bath, including liquid inglass types

C20 Oil 580-06-09 1 Litre(Ambient to +150ºC)

4. Blackbody Target 580-02-12 Use with Standard Probe(935-14-85)

5. Surface Sensor Kit 580-06-08 Includes an Insert andangled thermocouple

6. ITS-90 Fixed Point C12 Water Triple Point Cell17401 Gallium Cell

Standard Probe 935-14-85 Platinum ResistanceThermometer for use upto 250ºC

UKAS Calibration UKAS Calibration available to Order

Carrying Case 931-22-58 Large sturdy caseaccommodates the unitwith room for accessories

Calibration and Uncertainty

A certificate, traceable to National Standards, is included asstandard. Recommended is an optional UKAS five-pointcalibration.

The accuracy will depend very much on the mode of use andthe types of sensor to be used. Please contact Isotech fortutorials and uncertainty calculations and comprehensiveevaluation reports. The 580 Oceanus-6PLUS meets theCalibration Capacity requirements of EA I0/13, “EA Guidelineson the Calibration of Temperature Block Calibrators”.

Model No. OCEANUS-6

Temperature 45ºC below ambient to +140ºCRange

Absolute stability Dry Block Bath ±0.03ºCover 30 minutes Stirred Liquid Bath ±0.025ºC

Ice/Water Bath ±0.001ºCBlackbody Source ±0.03ºCSurface Sensor Calibrator ±0.5ºCITS-90 Fixed Point ±0.0002ºC

Computer Interface Included with Windows Software (page 74)

Cools from 20ºC to -10ºC in 90 minutes

Heats from -10ºC to 80ºC in 60 minutes

Best Performance See Graph

Calibration volume 50mm diameter by 300mm deep

Standard Insert 6 pockets, all 8.0mm diameter and250mm deep

Display Resolution 0.01 -19.99 to 99.990.1 100.0 to 140.0

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K

Power 108 to 130V (50 / 60 Hz) or208 to 240V (50 / 60 Hz)300 Watts

Overall dimensions Height 430mmWidth 310mmDepth 300mm

Weight 17kg

How to Order OceanusPLUS 580Please specify model type requiredPlease specify voltage requiredPlease specify options required

Options

To see the ISOCAL-6 in use ask for our27 minute video supplied on CD-Rom

Page 24: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/24

Seetemperature

chart

Proven QualitySlim Fixed Point Cells

The International Temperature Scale of 1990 applies World-wide; it includes specified fixed points -fundamental constants of nature. The majority of which relate to the freezing point of a pure substance,e.g. Pure Aluminium freezes at 660.3230ºC. These standards are generally kept by Primary andNational Laboratories. Isotech's Databook 1 includes these Primary Standards and associatedequipment. Databook 2 includes a range of more affordable fixed points that are used in SecondaryLaboratories and for extreme accuracy industrial calibration.

Many of the products featured in this Databook can be used with Fixed Point Cells enabling a modestlyconceived industrial laboratory to be upgraded if and as the requirements arise. These Fixed Points arecalled Slim Cells because they are somewhat smaller than the cells normally used for Primary Standards.

These more economic cells are available in a number of constructions including metal clad as well asquartz glass clad cells. The metal cells are pure to 6N, 99.9999% pure and are sealed with 99.9999%pure argon at one atmosphere.

The Water Triple Point is perhaps the most important of the fixed points and there is a choice of twomodels, the smaller for the ISOCAL-6 Range, and the larger cell for the Oceanus-6.

The table opposite shows the Slim Cells available for the products in this databook. For furtherinformation relating to fixed point calibration please refer to Databook 1, Databook 2 and visit ourweb-site.

Isotech's Slim Fixed Point Cells can be used to calibrate temperature sensors, either as the metal withinthem melts, or as it freezes. Typically, says CCT/96-8, a 6N pure cell will melt over 80% of its plateauwithin ±1mK. Exceptionally Isotech's Gallium Cell will melt over ±0.2mK.

Allowance must be made for the sensor being calibrated, sensors with short sensing lengths will addno additional errors to the above but other types with longer sensing lengths will add additionaluncertainties. An article is available from Isotech detailing stem conduction errors, please ask for yourcopy, free with our compliments.

Two examples showing the cell/apparatus/sensor performance are illustrated on the facing page.

Key Features

Affordable solution for extreme accuracyIndustrial calibration.

6N (99.9999%) purity metals.

Sealed with 6N (99.9999%) pure argon at thefreeze temperature.

Smaller versions of ITS-90 optimal Fixed Points.

Ideal for shorter sensors.

10 years of proven history of use in manylaboratories throughout the World.

Page 25: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/25

Proven Quality

The graph shows a Standard Platinum ResistanceThermometer's temperature firstly as measured in a 2week old large Water Triple Point Cell and then asmeasured in a Slim Cell placed in an Isotech metalblock bath such as Europa-6 or Venus 2140. After 16hours the apparatus was switched off (we wanted togo home).

The graph shows a Standard Platinum ResistanceThermometer's temperature firstly as measured in alarge 7N pure Gallium Cell and then as recorded in aSlim Gallium Cell placed in an Isotech metal blockbath such as Europa-6 or Venus 2140.

(1) Depth from metal surface to the bottom of the re-entrant quartz tube(2) Immersion errors depend on total depth of immersion in the apparatus which for

Oceanus, Medusa and Oberon is 300mm, for Europa is 160mm.Please ask for a free article titled 'Depth of Immersion Errors' for more details.

(3) Assuming the law of Dilute Solutions (ref. CCT/96-8 Page 6). The re-entrant tubein all the above cells is at least 8mm inside diameter.

(4) This cell is smaller and fits into the Calisto from the Isocal-6 range, metal clad.How to Order Slim Cells

Please state Type and Model Number

Slim Fixed Point Cells

Cell

Mercury

Water

Water

Gallium

Gallium

Indium

Indium (4)

Tin

Zinc

Aluminium

Silver

Gold

Copper

Temperature ITS-90Value (ºC)

-38.8344ºC

0.01ºC

0.01ºC

29.7646ºC

29.7646ºC

156.5985ºC

156.5985ºC

231.928ºC

419.527ºC

660.323ºC

961.78ºC

1064.78ºC

1084.62ºC

Length Immersed(mm) (1) (2)

120

130

260

250

120

130

110

130

130

130

130

130

130

Expected depressionfrom ITS-90 (mK) (3)

0.04

0.1

0.04

0.04

0.04

0.5

0.5

0.3

0.5

0.7

1.1

2

2

Type

Metal Clad

Glass

Glass

Metal Clad

Metal Clad

Metal Clad

Metal Clad

Metal Clad

Metal Clad

Metal Clad

Quartz Clad

Quartz Clad

Quartz Clad

Apparatus

Europa-6

Europa-6Venus 2140

Oceanus-6

Oceanus-6

Europa-6Venus 2140Calisto 2250

Medusa

Calisto 2250

Medusa

Medusa

Oberon / Medusa 3

Oberon

See Databook 2Model 469

See Databook 2Model 469

Model Number

17724 Slim

D8

C12

17401

17401 Slim

17668M Slim

17156M Slim

17669M Slim

17671M Slim

17672M Slim

17673 Slim

17675 Slim

17674 Slim

Page 26: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/26

Introduction to Dry Block CalibrationDry Block Calibrators

Isothermal Technology manufacture a full range oftemperature calibration equipment for PrimaryStandards used in National and Primary Labs, forSecondary Lab Equipment used in accreditedcalibration laboratories and through to the Dry Blocksfeatured here. Customers include the worlds largestPrimary Standards Laboratories, AccreditedLaboratories (UKAS, DKD etc), large multinationalcompanies, research organisations, manufacturingetc. Eighty percent of Nations rely on Isotech tosupply their country's standards. This is not aresponsibility taken lightly and Isotech constantlyinvests in its own full scale UKAS (formerly NAMAS)accredited laboratory. Isotech issues UKAS certificatesfor fixed point cells, thermometers, indicators and dryblocks. Isotech have issued several thousandcalibration certificates and carried out severalthousand measurements on Dry Blocks. We calibrateall types, not just our own.

The benefit of this experience, and the knowledge ofyears of manufacturing Dry Blocks is invested backinto these products with the goal of constantimprovement. In recent years the number ofproducers of Dry Blocks has increased dramatically,whilst many look similar (and some look surprisinglysimilar to our established models) they are often verydifferent inside and can perform very badly.

The Dry Block Calibrators complement with theISOCAL-6 range. Above temperatures ofapproximately 250ºC it is not practical to use stirredliquids due to fumes, risk of ignition and safetyconsiderations.

Isotech's higher temperature calibration bathsincorporate as many of the Isocal-6 options as is safeand practical to provide.

Dry Block Check ListBefore you purchase check:-

1 - Does the Producer Have anAccredited Laboratory?UKAS accreditation, “is the means by which, in thepublic interest, the integrity and competence ofindependent evaluators is confirmed and declared”.Isotech can issue a UKAS certificate with theperformance expressed in the manner that you willneed, not to some confusingly expressed specificationthat is made with no confirmation of integrity andcompetence.

2 ExperienceDoes the producer have experience? Do theyunderstand the difference between accuracy anduncertainty? Can they tell you how to calculate theuncertainty of a probe being calibrated in the block?Isotech can.

3 ExpandableCan the Dry Block be used with other sensors? Arethere accessories available for future expansion? WithIsotech products they are.

4 - PC SupportCan it be connected to a computer? Is there softwareavailable, can it be automated? Isotech Dry BlockCalibrators have a range of software options.

5 - DocumentedIs the bath fully documented? Can you download afull evaluation report from the Web Site? Does itcome with a comprehensive handbook and tutorial? Istraining available? Isotech provide all of these free ofcharge.

6 PracticalIsotech Dry Blocks are practically designed with astrong metal case, and are a compact portable size. Ifyou are going to carry it around don't forget to checkthe size and weights. It is surprising how large someother blocks are, even though they take the samenumber of probes. Beware if the specification doesnot include the weight.

7 ValueCheck the prices, all the above come at an amazinglycompetitive price if you buy from Isotech.

Page 27: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/27

Introduction to Dry Block CalibrationDry Block Calibrators

ImprovementsNew in this Databook are the PLUS models. Isotech workedwith a world leader in temperature control technology todevelop easy to use Dry Block controllers and Indicators.Now Isotech plug-in controllers are exceptionally easy touse with a clear user interface. Power feedback is usedto stabilise against supply voltage changes, a digital filtercircuit ensures high integrity of measurement withoutdrift, rejecting 50/60Hz pick up and filtering out othersources of noise. Resolution is increased. The indicatorshave universal thermocouple inputs and we have addeda PC interface and software as standard. Check theindividual models for full details. Windows software isnow provided as standard, with expandable options tocalibrate up to 16 sensors at a time. The new I-CALsoftware can event capture images for automaticcalibration of hand-helds etc.

Inbuilt IndicatorThe SITE (S) models include an electronic temperatureindicator that can be used with a 100Ω ResistanceThermometer, Thermocouples, (K, N, R, S, L, B, PL2, T,J and E) and DC process inputs including 4-20mAcurrent transmitters . A reference thermometer can beconnected or for complete flexibility the in-builtindicator can be used to show the value from a sensorbeing calibrated.

Using Isotech's Dry Blocks TraceableCalibrationFor best practice the recommendation is that acalibrated probe is placed into the Dry Block Insertand the thermometers under test “can be related toappropriate standards, generally international ornational standards, through an unbroken chain ofcomparisons”. Thus meeting many quality systemsincluding requirements of ISO 9000.

Using the Dry Bock itself as the Reference (orstandard) raises a number of issues, such as how is theuncertainty of the Dry Block calculated. In practice, itcan vary significantly, and there are some poordesigns from many suppliers where it is not possibleto achieve this in a satisfactory manner. Recently,International Guidelines have been published fromthe EA (European co-operation for Accreditation) thatgive guidance, and requirements, for the calibrationof Dry Blocks (EA10-13). For the most demandingapplications we continue to recommend that areference probe is used, the same method as used insecondary temperature laboratories, but for lessdemanding calibration, and the quick testing ofsensors the Dry Block can be used without a referenceprobe, refer to the Dry Block’s Evaluation Report fortypical performance.

Page 28: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/28

400ºCto

37ºC

Dry Block CalibratorApollo 1 & 2

The Apollo Dry Block range offer a complete self contained and simple solution for the checking andcalibrating of temperature sensors. The sensors under test are placed in the large block and one of thefive preset temperatures is selected, the temperature from the sensor is then recorded against thevalue from the Apollo's UKAS (formerly NAMAS) Calibration Certificate. This calibration certificate isincluded as standard. The moderate temperature range, deep immersion depth and fixed block allowit to be issued without the need for an external standard.

The Apollo is widely used as a reference standard in hospitals, local government and food industriesetc. The Apollo has been developed from Isotech's very first dry block and has a long history ofsuccessful use. See the calibration history chart of a typical bath.

The block has four 8mm pockets and four 19.5mm pockets all 160mm deep.

The Apollo can be provided with customer selected temperatures in addition to the two standardmodels which have the following temperatures preset.

Apollo 1 37ºC 65ºC 100ºC 121ºC 130ºC

Apollo 2 100ºC 150ºC 200ºC 250ºC 300ºC

Key Features

Easy to use temperature reference, selfcontained no further equipment necessary.

High Capacity Block.

Gold Plated Selector Switch.

Exceptionally Low Drift.

Includes UKAS Certificate as Standard.

Option for customer specified presettemperatures in the range 35ºC to 400ºC.

This graph showsthe actual

calibration historyover 12 years of

an Apollo, thedeviation over 12years is ±0.05ºC

and all values fallcomfortably

within theuncertainty band,

0.15ºC

Page 29: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/29

Dry Block CalibratorApollo 1 & 2

Calibration and Uncertainty

A UKAS calibration certificate is included as standard.

The Apollo meets the Calibration Capacity requirements ofEA-10/13, “EA Guidelines on the Calibration of TemperatureBlock Calibrators”.

Model No. APOLLO 1 & APOLLO 2

Temperature APOLLO 1 37, 65, 100, 121, 130ºCRange APOLLO 2 100, 150, 200, 250, 300ºC

Absolute stability At 100ºC ±0.02ºCover 30 minutes

Cools from Apollo 1 Apollo 2130ºC to 65ºC 70 minutes N/A

300ºC to 100ºC N/A 160 minutes

Heats from Apollo 1 Apollo 237ºC to 130ºC 20 minutes N/A

100ºC to 300ºC N/A 30 minutes

Uncertainties Refer to Uncertainties Graph

Calibration volume Four 8mm pockets, 160mm deep andfour 19.5mm pockets, 160mm deep

Display LED indicator of At TemperatureUnder Temperature andOver Temperature

Power 100 to 120V (50 / 60 Hz) or200 to 240V (50 / 60 Hz)500 Watts

Overall dimensions Height 302mmWidth 176mmDepth 262mm

Weight Apollo 1 9.5kgApollo 2 9.5kg

How to Order Apollo 1 or Apollo 2Please specify model type requiredPlease specify voltage requiredPlease specify options required

Options

Customer Specified Temperatures Select Five Pointsbetween 35 and 400ºC

Point Nominal Temperature Value

1.

2.

3.

4.

5.

Carrying Case 931-22-64 Sturdy caseaccommodates the unitwith room foraccessories

Page 30: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/30

650ºCto

35ºC

Dry Block CalibratorJupiterPLUS 650

The JupiterPLUS650 Dry Block range offers industry-leading performance in an easy to use portablepackage - ideal for the calibration of thermocouples and platinum resistance thermometers. It hasbeen designed for fast heating and cooling for convenient field use. For flexibility surface sensor andinfrared thermometer accessories can be added.

The standard insert can hold up to six thermometers. For larger blocks see the Gemini and Medusamodels. The JupiterPLUS650 is available in two models, the BASIC (B) and the SITE (S). The B modelincludes a sophisticated temperature controller with a dual display for Set Temperature and Dry BlockTemperature.

The S model includes a built in digital indicator to which an external standard thermometer can beconnected giving greater accuracy eliminating temperature gradient and loading errors. For SurfaceSensor and Blackbody use an external thermometer should always be used. For lab accuracy theJupiterPLUS650 can be used with a high-end temperature indicator such as one of the Isotech TrueTemperature Indicators (TTI).

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature.Increased resolution of ±0.01 available throughout the range via the PC interface and from0.01 to +99.99 locally on the auto-ranging front display. The controller features multi-pointblock to display correction giving good absolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers,Thermocouples (Types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including4-20mA current transmitters to be displayed on the inbuilt indicator. The indicator can beprogrammed with up to five calibration points to provide high accuracy digital probe matching.The indicator and controller are both addressable over the communications link.

New! The Site model can now be used with the supplied Cal Notepad software to testthermostats.

CellsPages 24 - 25

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

Key Features

Calibrate Whole Calibration Loop - using a heatsource rather than an electrical simulator a testinstrument and probe can be calibrated as asystem.

Good Temperature UniformityCareful design and research allows use of a highconductivity heater block to give the besttemperature uniformity.

Wide Operating RangeFast response from 35ºC to 650ºC ideal for fielduse.

Simple To Use - Outstanding Value for Money!

S Model Includes Universal Input TemperatureIndicator allowing for up to five “correction”points to be programmed.

Windows Software and PC Interface asStandard.

Free Evaluation Report - ask for full data or visitwww.isotech.co.uk/industrial.

Page 31: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/31

Dry Block CalibratorJupiterPLUS 650

Blackbody Kit 852-09-01 Includes a Blackbodytarget and Sensor

Surface Sensor Kit 852-07-15 Includes an Insert andangled thermocouple

Air Cooling 853-04-02 For use with acompressor thisaccessory allows air tobe blown into the blockfor rapid cooling.

Standard Probe 935-14-72 Platinum ResistanceThermometer for useup to 650ºC

UKAS Calibration UKAS Calibration available to Order

Carrying Case 931-22-64 Sturdy caseaccommodates the unitwith room for accessories

Calibration and Uncertainty

A certificate, traceable to National Standards, is included asstandard. Recommended is an optional UKAS five-pointcalibration.

The accuracy will depend very much on the mode of use andthe types of sensor to be used. Please contact Isotech fortutorials and uncertainty calculations and comprehensiveevaluation reports. The JupiterPLUS 650 meets the CalibrationCapacity requirements of EA I0/13, “EA Guidelines on theCalibration of Temperature Block Calibrators”.

Options

Model No. JUPITERPLUS 650

Temperature 35ºC to 650ºCRange

Absolute stability At 50ºC ±0.02ºCover 30 minutes At 250ºC ±0.02ºC

At 650ºC ±0.03ºCBlackbody Source ±0.3ºCSurface Sensor Calibrator ±0.5ºC

Computer Interface Included with Windows Software (page 74)

Cools from 650ºC to 150ºC in 60 minutes

Heats from 30ºC to 650ºC in 20 minutes

Best Performance See Graph

Calibration volume 35mm diameter by 148mm deep

Standard Insert 6 pockets, 2 x 4.5mm, 2 x 6.4mm,1 x 8.0mm, 1 x 9.5mm diameter,all 140mm deep

Display Resolution 0.01 to 99.990.1 100.0 to 650.0

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K

Power 100 to 120V (50 / 60 Hz) or200 to 240V (50 / 60 Hz)1000 Watts

Overall dimensions Height 302mmWidth 176mmDepth 262mm

Weight 8.5kg

How to Order JupiterPLUS 650Please specify model type requiredPlease specify voltage requiredPlease specify options required

Metal Block Insert 852-07-11 Standard InsertIncluded

852-07-07 Blank InsertInsert without pockets forlocal machining

852-07-07C Special InsertContact Isotech with yourrequirements

Alternative Metal Block Inserts

852-09-03 Standard Insert type B13mm , 10mm, 8mm, 5mmand 3.5mm diameter holes,all 140mm deep

852-09-04 Special Insert type C 8mm,6 x 6.5mm diameter holes,all 140 deep

Page 32: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/32

700ºCto

50ºC

550ºCto

35ºC

Dry Block CalibratorGeminiPLUS 550/700

The GeminiPLUS Dry Block range offers industry-leading performance in an easy to use portablepackage - ideal for the calibration of thermocouples and platinum resistance thermometers. It hasbeen designed to have a large block suitable for more probes, or larger probes than can beaccommodated in the Jupiter and Calisto models.

The GeminiPLUS is available with a fixed block with 8 pockets or with a large removable block, seeGemini LRI overleaf. The GeminiPLUS is available in two models, the BASIC (B) and the SITE (S). The Bmodel includes a sophisticated temperature controller with a dual display for Set Temperature and DryBlock Temperature. The large block has the benefit of higher capacity but with longer heating andcooling times compared to the fast response models such as the Jupiter 650.

The S model includes a built in digital indicator to which an external standard thermometer can beconnected giving greater accuracy eliminating temperature gradient and loading errors. For labaccuracy the GeminiPLUS can be used with a high-end temperature indicator such as one of the IsotechTrue Temperature Indicators (TTI).

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature.Increased resolution of ±0.01 available throughout the range via the PC interface and from0.01 to +99.99 locally on the auto-ranging front display. The controller features multi-pointblock to display correction giving good absolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers,Thermocouples (Types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including4-20mA current transmitters to be displayed on the inbuilt indicator. The indicator can beprogrammed with up to five calibration points to provide high accuracy digital probe matching.The indicator and controller are both addressable over the communications link.

New! The Site model can now be used with the supplied Cal Notepad software to testthermostats.

CellsPages 24 - 25

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

Key Features

Calibrate Whole Calibration Loop - using a heatsource rather than an electrical simulator a testinstrument and probe can be calibrated as asystem.

High Capacity Block - Large Calibration volumefor many sensors and low loading errors.

Two Operating Ranges - Fast response from35ºC to 550ºC or 50ºC to 700ºC.

Simple To Use - Outstanding Value for Money!

S Model Includes Universal Input - TemperatureIndicator allowing for up to five “correction”points to be programmed.

Windows Software and PC Interface asStandard.

Free Evaluation Report - ask for full data or visitwww.isotech.co.uk/industrial

Page 33: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/33

Dry Block CalibratorGeminiPLUS 550/700

Metal Block SleevesGeminiPLUS 550 Please Set of four Sleeves to suit

specify the block.Optional single holesizes 4, 6, 8, 10, 12,14mm diameterall 150mm deep.

857-07-01 Undrilled sleeves forlocal machining

857-07-03 1 sleeve with 2 holes4.5mm x 150mm deep

GeminiPLUS 700 Please Set of four Sleeves to suitspecify the block.

Optional single holesizes 4, 6, 8, 10, 12,14mm diameterall 150mm deep

857-07-02 Undrilled sleeves forlocal machining

857-07-04 1 sleeve with 2 holes4.5mm x 150mm deep

Note: The use of sleeves will introduce an additionalthermal gradient into the block. This can be avoidedby using the LRI model with a block drilled forspecific probes

Standard Probe 935-14-72 Platinum ResistanceThermometer for useup to 650ºC

935-14-63 Type N Thermocouplefor use up to 700ºC

UKAS Calibration UKAS Calibration available to Order

Carrying Case 931-22-64 Sturdy caseaccommodates the unitwith room for accessories

Calibration and Uncertainty

A certificate, traceable to National Standards, is included asstandard. Recommended is an optional UKAS five-pointcalibration.

The accuracy will depend very much on the mode of use andthe types of sensor to be used. Please contact Isotech fortutorials and uncertainty calculations and comprehensiveevaluation reports. The GeminiPLUS 550/700 meets theCalibration Capacity requirements of EA I0/13, “EA Guidelineson the Calibration of Temperature Block Calibrators”.

Model No. GEMINIPLUS 550 / 700

Temperature GEMINIPLUS 550 35ºC to 550ºCRange GEMINIPLUS 700 50ºC to 700ºC

Absolute stability At 35ºC ±0.05ºCover 30 minutes At 275ºC ±0.05ºC

At 550ºC ±0.05ºCAt 700ºC ±0.05ºC

Computer Interface Included with Windows Software (page 74)

Cools from GeminiPLUS 550550ºC to 275ºC 35 minutes550ºC to 60ºC 345 minutes

Heats from GeminiPLUS 550 GeminiPLUS 70030ºC to 550ºC 35 minutes N/A50ºC to 700ºC N/A 110 minutes

Best Performance See Graph

Calibration volume Four 8mm pockets, four 19.5mm pockets,Standard Block all 160mm deep

Display Resolution 0.01 to 99.990.1 100.0 to 700.0

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K

Power 100 to 120V (50 / 60 Hz) or200 to 240V (50 / 60 Hz)600 Watts

Overall dimensions Height 302mmWidth 176mmDepth 262mm

Weight GeminiPLUS 550 8.5kgGeminiPLUS 700 14kg

How to Order GeminiPLUS 550 or GeminiPLUS 700Please specify model type requiredPlease specify voltage requiredPlease specify options required

Options

Page 34: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/34

700ºCto

50ºC

550ºCto

35ºC

Dry Block CalibratorGeminiPLUS 550/700 LRI

The GeminPLUS LRI (Large Removable Insert) Dry Block range offers industry-leading performance in aneasy to use portable package - ideal for the calibration of thermocouples and platinum resistancethermometers.

It differs from the GeminiPLUS models in that it has been designed to have a large removable block witheight 8mm pockets, supplied as standard. Alternatively a block can be custom drilled allowing manyprobes to be calibrated simultaneously without the need for inserts and ensuring excellenttemperature uniformity.

GeminiPLUS LRI is available in two models, the BASIC (B) and the SITE (S). The B model includes asophisticated temperature controller with a dual display for Set Temperature and Dry BlockTemperature. The large block has the benefit of higher capacity but with longer heating and coolingtimes compared to the fast response models such as the Jupiter 650.

The S model includes a built in digital indicator to which an external standard thermometer can beconnected giving greater accuracy eliminating temperature gradient and loading errors. For labaccuracy the GeminiPLUS LRI can be used with a high-end temperature indicator such as one of theIsotech True Temperature Indicators (TTI).

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature.Increased resolution of ±0.01 available throughout the range via the PC interface and from0.01 to +99.99 locally on the auto-ranging front display. The controller features multi-pointblock to display correction giving good absolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers,Thermocouples (Types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including4-20mA current transmitters to be displayed on the inbuilt indicator. The indicator can beprogrammed with up to five calibration points to provide high accuracy digital probe matching.The indicator and controller are both addressable over the communications link.

New! The Site model can now be used with the supplied Cal Notepad software to testthermostats. Cells

Pages 24 - 25

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

Key Features

Calibrate Whole Calibration Loop - using a heatsource rather than an electrical simulator a testinstrument and probe can be calibrated as asystem.

High Capacity Insert - Large removable insertcan hold many sensors simultaneously.

Custom Inserts Available - Inserts can be drilledto customer specifications to hold many sensorswithout the need for packing sleeves or inserts.

Simple To Use - Outstanding Value for Money!

S Model Includes Universal Input - TemperatureIndicator allowing for up to five “correction”points to be programmed.

Windows Software and PC Interface asStandard.

Free Evaluation Report - ask for full data or visitwww.isotech.co.uk/industrial

Page 35: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/35

Dry Block CalibratorGeminiPLUS 550/700 LRI

Removable InsertsGeminiPLUS 550 LRI 976-07-01a Included as Standard

Removable insert witheight 8mm pockets

976-07-01b Blank InsertInsert without pockets forlocal machining

976-07-01c Special InsertContact Isotech with yourrequirements

GeminiPLUS 700 LRI 976-07-02a Included as StandardRemovable block witheight 8mm pockets

976-07-02b Blank InsertInsert without pockets forlocal machining

976-07-02c Special InsertContact Isotech with yourrequirements

Standard Probe 935-14-72 Platinum ResistanceThermometer for useup to 650ºC

935-14-63 Type N Thermocouplefor use up to 700ºC

UKAS Calibration UKAS Calibration available to Order

Carrying Case 931-22-65 Sturdy caseaccommodates the unitwith room for accessories

Calibration and Uncertainty

A certificate, traceable to National Standards, is included asstandard. Recommended is an optional UKAS five-pointcalibration.

The accuracy will depend very much on the mode of use andthe types of sensor to be used. Please contact Isotech fortutorials and uncertainty calculations and comprehensiveevaluation reports. The GeminiPLUS 550/700 LRI meets theCalibration Capacity requirements of EA I0/13, “EA Guidelineson the Calibration of Temperature Block Calibrators”.

Model No. GEMINIPLUS 550 / 700 LRI

Temperature GEMINIPLUS 550 LRI 35ºC to 550ºCRange GEMINIPLUS 700 LRI 50ºC to 700ºC

Absolute stability At 35ºC ±0.02ºCover 30 minutes At 250ºC ±0.03ºC

At 550ºC ±0.04ºCAt 700ºC ±0.05ºC

Computer Interface Included with Windows Software (page 74)

Cools from GeminiPLUS 550550ºC to 275ºC 132 minutes550ºC to 60ºC 420 minutes

Heats from GeminiPLUS 550 GeminiPLUS 70030ºC to 550ºC 60 minutes N/A50ºC to 700ºC N/A 120 minutes

Best Performance See Graph

Calibration volume 64mm diameter by 160mm deep

Standard Insert Eight 8mm diameter pocketsall 154mm deep

Display Resolution 0.01 to 99.990.1 100.0 to 700.0

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K

Power 100 to 120V (50 / 60 Hz) or200 to 240V (50 / 60 Hz)1000 Watts

Overall dimensions Height 302mmWidth 176mmDepth 262mm

Weight GeminiPLUS 550 8.5kgGeminiPLUS 700 18kg

How to Order GeminiPLUS 550 or GeminiPLUS 700Please specify model type requiredPlease specify voltage requiredPlease specify options required

Options

Page 36: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/36

1200ºCto

150ºC

Portable Calibration FurnacePegasusPLUS 1200

The PegasusPLUS range offers extreme high temperature calibration in an easy to use portable package- ideal for the calibration of high temperature thermocouples. It has been designed for fast heatingand finds applications in the glass, electrical power, automotive and material processing industries. ABlackbody target can be added for the calibration of infrared thermometers.

The standard insert has four 8mm pockets 80mm deep. The metal insert is strategically placedbeneath 50mm of insulation to provide optimal performance over the radiant temperature range. Forlarger blocks see the Oberon model. The Pegasus is available in two models, the BASIC (B) and theSITE (S). The B model should be used with an external reference probe and indicator, such as the TTI7. The thermocouples under test should be calibrated by comparison to the external probe.

The S model includes a built in digital indicator to which an external standard thermometer should beconnected, giving greater accuracy eliminating temperature gradient and loading errors. Therecommended probe is a platinum Type R thermocouple. The optional Blackbody target is used witha specially angled Type R thermocouple that sits immediately behind the target area.

Includes as standard: Windows Software and a Computer Interface. Increased resolution of ±0.1available throughout the range via the PC interface and from 150.0 to 999.9 locally on the auto-ranging front display. The controller features multi-point block to display correction giving goodabsolute accuracy.

The S model has universal sensor input allowing Platinum Resistance Thermometers,Thermocouples (Types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including4-20mA current transmitters to be displayed on the inbuilt indicator. The indicator can beprogrammed with up to five calibration points to provide high accuracy digital probe matching.The indicator and controller are both addressable over the communications link.

New! The Site model can now be used with the supplied Cal Notepad software to testthermostats.

CellsPages 24 - 25

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

Key Features

Calibrate Whole Calibration Loop - using a heatsource rather than an electrical simulator a testinstrument and probe can be calibrated as asystem.

Custom furnace design - provides fast heatingwith optimised thermal profile.

High Operating RangePortable field use up to 1200ºC.

Simple To Use - Outstanding Value for Money!

S Model Includes Universal Input TemperatureIndicator allowing for up to five “correction”points to be programmed.

Windows Software and PC Interface asStandard.

Free Evaluation Report - ask for full data or visitwww.isotech.co.uk/industrial

Page 37: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/37

Portable Calibration FurnacePegasusPLUS 1200

Metal Block Insert 853-06-01 Standard InsertIncluded

853-06-02 Blank InsertInsert without pockets forlocal machining

853-06-02b Special InsertContact Isotech with yourrequirements

Standard Probe 935-14-91 Type R PlatinumThermocouple for useup to 1200ºC

UKAS Calibration UKAS Calibration available to Order(S models only)

Air Cooling 853-04-02 For use with a compressorthis accessory allows air tobe blown into the blockfor rapid cooling.

Ceramic Insulators 853-06-04 Spare insulation packIncludes 2 x standard topsand 2 x standard bottoms

Carrying Case 931-22-64 Sturdy caseaccommodates the unitwith room for accessories

Calibration and Uncertainty

A certificate, traceable to National Standards, is included asstandard. Recommended is an optional UKAS five-pointcalibration.

The accuracy will depend very much on the mode of use andthe types of sensor to be used. Please contact Isotech fortutorials and uncertainty calculations and comprehensiveevaluation reports. The PegasusPLUS 1200 meets the CalibrationCapacity requirements of EA I0/13, “EA Guidelines on theCalibration of Temperature Block Calibrators”.

Model No. PEGASUSPLUS 1200

Temperature 150ºC to 1200ºCRange

Absolute stability At 150ºC ±0.1ºCover 30 minutes At 1200ºC ±0.2ºC

Blackbody Source ±0.3ºC

Computer Interface Included with Windows Software (page 74)

Cools from 1200ºC to 800ºC in 50 minutes*1200ºC to 200ºC in 180 minutes**substantially reduced by the cooling adaptor

Heats from 100ºC to 1200ºC in 20 minutes

Best Performance See Graph

Calibration volume 33.5mm diameter by 130mm deep

Standard Insert 4 pockets, 8.0mm diameterby 80mm deep

Display Resolution 0.1 to 999.91 1000 to 1200

PC can display 0.1 across wholerange with the software included

Indicator units ºC, ºF, K

Power 100 to 120V (50 / 60 Hz) or200 to 240V (50 / 60 Hz)2300 Watts

Overall dimensions Height 302mmWidth 176mmDepth 262mm

Weight 8.5kg

How to Order PegasusPLUS 1200Please specify model type requiredPlease specify voltage requiredPlease specify options required

Options

Blackbody Kit 853-06-03 Includes a Blackbodytarget and Sensor

The Pegasus 1200 has to be used with the targetvertical. Databook 4 includes the Pegasus 1200Rwhich is suitable for horizontal operation

Page 38: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/38

700ºCto

30ºC

Dry Block Calibrator510 Medusa & 511 Medusa 3

Isotech have a wide range of Dry Blocks to suit probes requiring a large immersion depth. These products featurelarge and deep calibration volumes. As such they are less portable then the earlier Dry Blocks in this databookbut have higher capacities and retain outstanding temperature uniformity, this uniformity is so good that theselarger products are also featured in DataBook 2 as apparatus for Secondary Laboratories to realize the FixedPoints of ITS-90.

Medusa 510 has a maximum operating temperature of 550°C. The Medusa 3 Model 511 can be used to 700°Cand features three zone control. In addition to the main heating zone there are additional top and bottom heaterswhich compensate for the end losses creating a constant temperature zone across the well.

For Comparison Calibration the MedusaPLUS should be used with an insert, the standard insert has six 8mmpockets 250mm deep. Also available is an insert 44mm diameter x 170mm deep which is suspended from thetop of the block so that the height is user adjustable. For flexibility the MedusaPLUS can also be used withaccessories for infrared thermometers and surface sensors. The MedusaPLUS is available in two models, theBASIC (B) and the SITE (S). The B model includes a sophisticated temperature controller with a dual display forSet Temperature and Dry Block Temperature.

The S model includes a built-in digital thermometer to which an external standard thermometer can beconnected giving greater accuracy, eliminating temperature gradient and loading errors. Also included in theSite model is a timer which can set the bath between two temperatures, and automate ITS-90 fixed pointoperation. For Surface Sensor and Blackbody use an external thermometer is recommended. For laboratoryaccuracy the MedusaPLUS can be used with a high-end temperature indicator such as an Isotech TTI model.

Includes as standard: Windows Software, Computer Interface and a Ramp to Set Point Feature. Increasedresolution of ±0.01 available throughout the range via the PC interface and from 0.01 to +99.99 locally on theauto-ranging front display. The controller features multi-point block to display correction giving good absoluteaccuracy.

New in the S model is universal sensor input allowing Platinum Resistance Thermometers, Thermocouples(types K, N, R, S, L, B, PL2, T, J and E) along with Linear Process Inputs including 4-20mA currenttransmitters to be displayed on the in-built indicator. The indicator can be programmed with up to fivecalibration points to provide high accuracy digital probe matching. The indicator and controller are bothaddressable over the communications link.

New ! The Site model can now be used with the supplied Cal NotePad software to test thermostats.

CellsPages 24 - 25

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

Key Features

High Capacity Deep Block45mm diameter x 285mm Deep.

Use for Comparison and Fixed PointCalibration.

Use with very long thermometers.

Fixed Point Cells Available

Material Temperature Uncertainty

Gallium 29.7646ºC ±0.001ºC

Indium 156.598ºC ±0.001ºC

Tin 231.928ºC ±0.002ºC

Zinc 419.527ºC ±0.005ºC

Lead 327.462ºC ±0.010ºC

Zinc 419.527ºC ±0.005ºC

Aluminium 660.323ºC ±0.010ºC

Page 39: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/39

Dry Block Calibrator510 Medusa & 511 Medusa 3

510 Metal Block Insert 510-06-01 Standard InsertIncluded

510-06-02 Blank InsertInsert without pockets forlocal machining

510-06-03 Special InsertContact Isotech with yourrequirements

510-06-04 Adjustable EqualisingBlock

Standard Probe 935-14-95 Platinum ResistanceThermometer for use upto 650ºC

UKAS Calibration UKAS Calibration available to Order

ITS-90 Fixed Points ITL17401M Gallium Slim CellITL17668M Indium Slim CellITL17669M Tin Slim CellITL17670M Lead Slim CellITL17671M Zinc Slim CellITL17672M Slim Aluminium Cell510-05-00 Slim Cell Holder

Carrying Case 931-22-58 Sturdy caseaccommodates the unitwith room for accessories

Calibration and Uncertainty

A certificate, traceable to National Standards, is included asstandard. Recommended is an optional UKAS five-pointcalibration.

The accuracy of the Medusa will depend very much on themode of use, see the Uncertainty Graph for typicaluncertainties. NTPL calculate the uncertainties to UKASrequirements. The Medusa meets the Calibration Capacityrequirements of EA-10/13, “EA Guidelines on the Calibrationof Temperature Block Calibrators.”

Model No. 510 MEDUSAPLUS 511 MEDUSA 3

Temperature 30ºC to 550ºC 50ºC to 700ºCRange

Absolute stability Metal Block Bath ±0.03ºCover 30 minutes Blackbody Source ±0.1ºC

Surface Sensor Calibrator ±0.5ºCITS-90 Fixed Point ±0.001ºC

Computer Interface Included with Windows Software (page 74)

Cools from 550ºC to 30ºC in 5 hours

Heats from 30ºC to 550ºC in 90 minutes

Uncertainties Refer to Uncertainties Graph

Calibration volume 45mm diameter by 285mm deep

Standard Insert Six 8mm pockets all 250mm deep

Display Resolution 0.01 to 99.990.1 100.0 to 650.0

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K ºC, ºF, K

Power 100 to 120V 108 to 130V (50 / 60 Hz) or (50 / 60Hz) or200 to 240V 208 to 240V (50 / 60 Hz) (50 / 60Hz)1000 Watts 1800 Watts

Overall dimensions Height 430mm Height 430mmWidth 310mm Width 310mmDepth 300mm Depth 300mm

Weight 17kg 25kg

How to Order 510 MedusaPLUS or 511 Medusa 3Please specify model type requiredPlease specify voltage requiredPlease specify options required

Options

510 Blackbody Kit 510-06-05 Includes a Blackbodytarget and Sensor

511 Blackbody Kit 511-06-05 Includes a Blackbodytarget and Sensor

510 Surface Sensor Kit 510-06-06 Includes an insert andangled thermocouple

511 Surface Sensor Kit 511-06-06 Includes an insert andangled thermocouple

511 Metal Block Insert 511-06-01 Standard InsertIncluded

511-06-02 Blank InsertInsert without pockets forlocal machining

511-06-03 Special InsertContact Isotech with yourrequirements

511-06-04 Adjustable EqualisingBlock

For Evaluation Reports, Uncertainty Budgets and Calculations with regard to EA 10-13UKAS etc, please contact Isotech - also http://www.isotech.co.uk/refer.html

510 MedusaPLUS Performance

Page 40: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/40

1090ºCto

450ºC

Sodium Heatpipe Furnace426 Oberon

Isotech has a wide range of Dry Blocks to suit probes requiring a large immersion depth. Theseproducts feature large and deep calibration volumes. As such they are less portable then the earlierDry Blocks but have higher capacities and retain outstanding temperature uniformity, this uniformityis so good that these larger products are also featured in DataBook 2 as apparatus for SecondaryLaboratories to realise the Fixed Points of ITS-90.

For comparison calibration the Oberon should be used with an insert. The insert has six pockets 8mmdiameter by 160mm deep which is suspended from the top of the heatpipe so that the height is useradjustable. For flexibility an accessory can be added for blackbody use.

The Oberon uses a specially designed stress free isothermal heat-pipe as its furnace that ensures avery low thermal gradient along the furnace core.

The heat-pipe is designed so that the inner wall is not subject to thermal expansion stresses from theouter wall before the heat-pipe reaches conduction temperature. The working fluid is permanentlyand safely sealed within the plasma-arc welded enclosure.

Connections are provided for a water supply of 0.5 to 1 litre per minute to cool the case of theOberon. Water supply and waste connections are provided. Use without cooling is not recommendedabove 700ºC.

CellsPages 24 - 25

SoftwarePages 74 - 79

ProbesPages 62 - 67

IndicatorsPages 54 - 59

Key Features

Uses a Heat-pipe to eliminate verticaltemperature gradients.

Use for Comparison and Fixed PointCalibration.

Use with very long thermometers.

Fixed Point Cells Available

Material Temperature Uncertainty

Aluminium 660.323ºC ±0.02ºC

Silver 961.78ºC ±0.04ºC

Page 41: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/41

Sodium Heatpipe Furnace426 Oberon

Metal Block Insert 426-06-01 Standard Insert426-06-02 Blank Insert

Insert without pockets forlocal machining

426-06-03 Special InsertContact Isotech with yourrequirements

426-06-05 Adjustable EqualisingBlock

Transformer 935-19-43 230V/110V Transformer

ITS-90 Fixed Points 17672 Slim Aluminium Cell17673 Slim Silver Cell426-04-00 Slim Cell Holder

Model No. OBERON

Temperature 450ºC to 1090ºCRange

Absolute stability Metal Block Bath ±0.05ºCover 30 minutes Blackbody Source

Computer Interface Included with Windows Software (page 72)

Display Resolution 0.1 to 999.9PC can display 0.1 across wholerange with the software included

Power 108 to 130V (50 / 60 Hz)1000 Watts(230V Transformer available as an option)

Overall dimensions Height 410mmWidth 415mmDepth 280mm

Weight 30.5kg

How to Order 426 Oberon Please specify model type requiredPlease specify options required

Options

Blackbody Kit 426-06-04 Includes a Blackbodytarget and Sensor

Page 42: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/42

Introduction to Temperature TestersTesters

Isotech Dry Blocks have the features and performanceto meet most calibration needs. Sometimes there areapplications where a compromise has to be made,with regard to size or price. An engineer forced tocarry a calibrator up a ladder and into a confinedspace might have to sacrifice performance for physicalsize. Where sensors of small dimensions and largeuncertainties are used.

Isotech have developed a range of products called“Testers”. They don't have the same immersiondepths that larger baths have, the blocks are smallerand the dry blocks do not have a PC interface.

What they do have is:

Outstanding Value

Compact Size with true handheld models

Wide Operating Ranges, from less than-35ºC to 650ºC in three models

Fast Response

QuickCalsThere are two QuickCal models, handheld, portableand capable of operating from below -10ºC to 350ºC.

35 to 650ºC Model 903Model 903 is slightly larger than the QuickCals butoffers a larger block and operation to 650ºC.

55ºC below Ambient to 140ºC Model 907Model 907 like the 903 is larger than the QuickCal butoffers a larger block and wider temperature range.

Small Hot PlateA small portable hot plate for testing surface sensors,use up to 350ºC.

SimulatorsAlso included in this section are two devices thatsimulate a temperature sensors output, Isotech'soriginal “Isocal” an adjustable thermocouplesimulator and our Model 863 ResistanceThermometer Simulator, Model 862.

Page 43: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/43

Introduction to Temperature TestersTesters

Immersion depth is very important

When selecting a Dry Block Calibrator depth of immersion is very important. The chart below providesguidance in selecting a bath for immersion depth. Note that sensors with a long sensing length willrequire greater immersion. The chart is general and applies equally to all dry blocks - not to a particularmodel or manufacturer.

Immersion depths for various diameter thermocouples or thermistors in a dry block bath

Example shows 0.3% stem error for a 15mm diameter thermocouple immersed 160mm in a dry block.

Note 1. For sensors immersed in stirred liquids the diameter of sensor can be doubled, or the minimum depth halved.

Note 2. The sensing length must be added to the above immersion depth calculation

N.B. The above gives a good guide, however each sensor will be slightly different.

For full information on selecting the correct immersion depth for all types of sensors ask for our free 8 page immersion tutorial.

Page 44: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/44

+140ºCto

-12ºC

+350ºCto

+30ºC

Temperature Test UnitQuick-Cal Range

Isotech's Quick Cal range of testers are ideally suited to the less exacting applications where ourlarger, better specified baths are unnecessary.

The Low Temperature Quick Cal covers the temperature range of -15ºC to +140ºC, making it ideal forthe medical, sterilisation and food industries.

The High Temperature Quick Cal covers the temperature range from 30ºC to 350ºC.

Both have interchangeable inserts, so you don't have to buy another calibrator each time you find anew diameter sensor to calibrate. The Low Temperature Quick Cal is particularly exciting since it canwork from a small 12 volt battery, making it truly portable, alternatively a mains supply is available with100V - 240V, 50/60Hz input.

The Low Temperature Quick Cal comes complete with two inserts, one blank and one drilled 6.8mmdiameter 120mm deep. Alternatively we can provide pre-drilled inserts to suit your needs. Eachcalibration well is 13mm diameter and 120mm deep.

The High Temperature Quick Cal has a single larger well and interchangeable inserts, 25mm indiameter by 120mm deep. Blank and other special drillings are available to order. The HighTemperature Quick Cal is supplied 100-130 or 200-250V AC, 50/60Hz.

The handbook which accompanies the Quick Cal contains helpful advice on how to get the best fromyour calibrator. Accessories for the above include a carry case, alternative inserts and a rechargeablebattery for the Low Temperature Quick Cal.

Key Features

Interchangeable inserts - to suit differentdiameter sensors.

Wide Temperature Range - 15ºC to 350ºC in tworanges.

Faster - 9 minutes to temperature.

Compact - 65mm x 142mm x 175mm notincluding power entry.

Lightweight - weighs 1.5kg (3.3 pounds).

Two models - Low Temperature Quick Cal andHigh Temperature Quick Cal.

Page 45: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/45

Temperature Test UnitQuick-Cal Range

A. Extra Sleeves 560-06-01A Standard Sleeve560-06-01B Blank Sleeve560-06-01C Special Sleeve

(maximum bore 10mm)

B. Carrying Case 931-22-66

C. UKAS Certification

D. Rechargeable Battery & ChargerBattery is 12V 12Ah.On full charge battery: 60 minutes when cooling

90 minutes when heatingLonger life available with larger batteries - consult IsotechTemperature Range is -10ºC to 123ºC with battery operation

E. Fahrenheit OperationUnit can be supplied for ºF Operation

A. Extra Inserts 550-06-01A Standard Insert550-06-01B Blank Insert550-06-01C Special Insert

B. Carrying Case 931-22-66

C. UKAS Certification

E. Fahrenheit OperationUnit can be supplied for ºF Operation

F. Semi Standard Platinum Thermometer, model 935-14-98

Model No. High Temperature Quick-Cal

Temperature +30ºC to +350ºCRange In an ambient of 20ºC

Set Point 0.1ºC over rangeResolution

Accuracy ±0.1 to ±0.4ºC using ComparisonTechniques<100ºC ±0.4ºC against display temperaturewith a single probe in the 4.5mm pocket>100ºC ±1.5ºC against display temperaturewith a single probe in the 4.5mm pocket

Stability ±0.05ºC

Time to 9 minutes fromSet Point Ambient to 350ºC

Additional Time Typically 5 minutesfor best stability

Calibration 25mm diameter 120mm deepvolume Supplied with insert, 5 pockets, 8mm,

6.5mm, 6.5mm, 4.5mm, 4.5mm,all 115mm deep

Power 100-130V or 200-240V,(50 / 60 Hz)

Overall dimensions Height 65mmWidth 152mmDepth 175mm

Weight 1.5kg

How to Order High Temperature Quick-CalPlease specify Voltage requiredPlease specify Options required

Model No. Low Temperature Quick-Cal

Temperature -12ºC to +140ºCRange In an ambient of 20ºC

The Quick-Cal will achieve lower temperatureif used in a lower ambient temperature

Set Point 0.1ºC over rangeResolution

Accuracy ±0.1 to ±0.4ºC using ComparisonTechniques±0.4ºC against display temperature with asingle probe in the 4.5mm pocket

Stability ±0.05ºC

Time to 9 minutes fromSet Point 0ºC to 100ºC

Additional Time Typically 5 minutesfor best stability

Calibration Two 13mm diameter pockets 120mm deepvolume Supplied with two sleeves, one blank and

one drilled 6.8mm diameter 120mm deep

Power 15V DC or 100-240V, Switch modepower supply(50 / 60 Hz)

Overall dimensions Height 65mmWidth 152mmDepth 175mm

Weight 1.5kg

How to Order Low Temperature Quick-CalPlease specify Options required

High Temperature Quick-Cal OptionsLow Temperature Quick-Cal Options

Page 46: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/46

Temperature Test Unit907 Portable

The Isotech Portable Temperature Test Unit is a model designed specifically for ease of use and for onsite application where a small, light, portable unit is required. The 907 reaches its maximum temperatureof 140°C in 18 minutes and features a best accuracy of 0.2°C. The unit is economically priced yet sturdyand rugged, eliminating the need for a separate carrying case.

Two models are available, the Site model, which has both a controller and a built in indicator for PRT andThermocouple inputs (details page 27), and the Basic model, which has only the controller.

Calibration volume is 25mm diameter by 148mm deep and the 907 is provided with an insert drilled with4 holes, 2 x 4.5mm diameter, 1 x 6.5mm diameter and 1 x 9.5mm diameter. Other inserts can be providedwith specially drilled calibration pockets to suit your special requirements.

A certificate of test traceable to National Standards is provided with each purchase. As an option a UKASInternationally recognised certificate of calibration can be provided at extra cost.

Features

Wide temperature range - 55ºC below ambientto 140ºC.

Fast response time - minutes to temperature.

Portable

Includes a PC Interface

Metal Block Insert 907-02-03 Standard InsertIncluded

907-02-03B Blank InsertInsert without pockets forlocal machining

907-02-03C Special InsertContact Isotech with yourrequirements

Standard Probe 935-14-82 Platinum ResistanceThermometer

UKAS Calibration UKAS Calibration available to Order

Options

140ºCto

55ºCbelow

ambient

Calibration and Uncertainty

A certificate, traceable to National Standards, is included asstandard. Recommended is an optional UKAS five-pointcalibration.

The accuracy of the 907 Test Unit will depend very much on themode of use. NTPL calculate the uncertainties to UKASrequirements. The 903 Test Unit meets the Calibration Capacityrequirements of EA-10/13, ”EA Guidelines on the Calibration ofTemperature Block Calibrators.”

Model No. 907 Portable Temperature Test Unit

Temperature 55ºC below ambient to + 140ºCRange

Stabilization 10 minutesRange

Cools from 25ºC to -30ºC in 25 minutes

Heats from 25ºC to 140ºC in 18 minutes

Calibration 25mm diameter by 148mm deepVolume 4 pockets, 2 x 4.5mm diameter, 6.5mm

diameter and 9.5mm diameter.

Display Resolution 0.01 -19.99 to 99.990.1 -55.0 to -20.0 and 100.0 to

140.0 PC can display 0.01 acrosswhole range with the software included.

Indicator units ºC, ºF, K

Power 100 to 120V (50 / 60 Hz) or200 to 240V (50 / 60 Hz)150 Watts

Overall dimensions Height 228mmWidth 248mmDepth 143mm

Weight 6kg

How to Order 907 Portable Temperature Test UnitPlease specify model type requiredPlease specify voltage requiredPlease specify options required

Page 47: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/47

Temperature Test Unit903 Portable

The Isotech Portable Temperature Test Unit is a model designed specifically for ease of use and for on siteapplication where a small, light, portable unit is required. The 903 reaches its maximum temperature of 650°C in20 minutes and features a best accuracy of 0.25°C. The unit is economically priced yet sturdy and rugged,eliminating the need for a separate carrying case.

The 903 is available in two models, the Site model, which has both a controller and a built in indicator for use withtype ’N’ thermocouples, and the Basic model, which has only the controller. The unit features a shutter device onthe front panel to protect the user from heated inserts falling out whilst the unit is being moved or carried, andan internal fan to aid in the cooling of the block.

An optional cooling accessory is available which connects to an air supply and speeds up the cooling of the unit.Calibration volume is 25mm diameter by 148mm deep and the 903 is provided with an insert drilled with 3 holes,1 x 3mm diameter, 1 x 6.5mm diameter and 1 x 9.5mm diameter. Other inserts can be provided with speciallydrilled calibration pockets to suit your special requirements.

A certificate of test traceable to National Standards is provided with each purchase. As an option a UKASInternationally recognised certificate of calibration can be provided at extra cost.

Metal Block Insert 903-06-01 Standard InsertIncluded

903-06-02 Blank InsertInsert without pockets forlocal machining

903-06-03 Special InsertContact Isotech with yourrequirements

Standard Probe 935-14-63 Type N Thermocouple foruse up to 650ºC

UKAS Calibration UKAS Calibration available to Order

Options

650ºCto

35ºC

Calibration and Uncertainty

A certificate, traceable to National Standards, is included asstandard. Recommended is an optional UKAS five-pointcalibration.

The accuracy of the 903 Test Unit will depend very much on themode of use. NTPL calculate the uncertainties to UKASrequirements. The 903 Test Unit meets the Calibration Capacityrequirements of EA-10/13, ”EA Guidelines on the Calibration ofTemperature Block Calibrators.”

Key Features

Wide temperature range - +35ºC to +650ºC

Fast response time - 9 minutes to temperature.

Compact - 65mm x 142mm x 175mm notincluding power entry

Lightweight - weighs just 1.5kg (3.3 pounds)

Model No. 903 Portable Temperature Test Unit

Temperature 35ºC to 650ºCRange

Stabilization 10 minutesRange

Cools from 650ºC to 50ºC in 20 minutes

Heats from 50ºC to 650ºC in 20 minutes

Calibration 25mm diameter by 148mm deepVolume 3 pockets, 3mm diameter, 6.5mm

diameter and 9.5mm diameter.

Display Resolution 0.01 to 99.990.1 100 to 650

PC can display 0.01 across whole range with the software included

Indicator units ºC, ºF, K

Power 100 to 120V (50 / 60 Hz) or200 to 240V (50 / 60 Hz)750 Watts

Overall dimensions Height 185mmWidth 120mmDepth 232mm

Weight 4.9kg

How to Order 903 Portable Temperature Test UnitPlease specify model type requiredPlease specify voltage requiredPlease specify options required

Page 48: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/48

350ºCto

50ºC

Surface Sensor Calibrator983 Small Hot Plate

The Isotech Small Hotplate is a lightweight portable calibration system purpose designed for surfacemounted sensors. The flat surface plate is made from precision-machined aluminum. The sensor to betested is simply placed on the surface, for higher accuracy a calibrated surface sensor can be placedalongside and the two compared, see page 68 for details.

Good thermal contact is ensured by the flat disk that is recessed to allow the optional use of a heattransfer paste or fluid. Uniform heat distribution is achieved with a flat spiral heater clamped to anintegrating block below the surface of the plate. The typical accuracy that can be achieved 1ºC butthis will be influenced by the type of sensor to be calibrated.

The internal control sensor is located immediately below the plate’s surface.

A protective cover that can fit over the block is included along with a comprehensive handbook. TheHot Plate can be supplied configured for Fahrenheit operation, please specify on order.

The temperature range is from 35 to 350ºC, which is set by an advanced, but easy to use temperaturecontroller. The controller has been upgraded to have 0.01 resolution below 100ºC (0.1º above 100º).A PC interface is included as standard along with an RS232 converter lead and Windows software.

Key Features

Low Cost Portable Hot Plate.

PC Interface and Software.

Stable to +/- 0.1C.

Isotech has a True SurfaceTemperature Measurement System(see page 72) which is ideal for usewith this surface sensor calibrator.

Page 49: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/49

Surface Sensor Calibrator983 Small Hot Plate

Notes:

A similar model but with a black high emissivity surface isavailable, please refer to databook 4 for details.

Many of the dry block calibrators featured earlier within thisdatabook have accessories available for surface sensorcalibration.

Model No. 983 Small Hot Plate

Temperature 50ºC to 350ºCRange

Stabilisation Time 10 minutes

Cools from 350ºC to 100ºC in 125 minutes

Heats from 50ºC to 350ºC in 20 minutes

Uncertainties Dependant on sensorsand method of use1ºC typical

Calibration volume Flat Plate 71mm diameter

Display Resolution 0.01 to 99.990.1 100 to 350.0

PC can display 0.01 across wholerange with the software included

Indicator units ºC, ºF, K

Power 100 to 115V (50 / 60 Hz) or200 to 230V (50 / 60 Hz)200 Watts

Overall dimensions Height 115mmWidth 230mmDepth 225mm

Weight 3.9kg

How to Order 983 Small Hot PlatePlease specify voltage required

Page 50: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/50

Seeresistance

values

Resistance Thermometer SimulatorModel 862 - 24 Value

Isotech's well established Resistance Thermometer Simulator has set the standard for calibratinginstruments reading from Industrial Platinum Resistance Thermometers.

Each of the 24 temperatures simulated comprise a high stability network permitting the exactresistance to be set during calibration at our Laboratory.

The high stability (10ppm/ºC) of the Resistance Thermometer Simulator reduces drift and hencere-calibration costs.

Thick gold plating of all contacts ensures long life and high reliability.

TEMPERATURES SIMULATED

OTHER TEMPERATURES AVAILABLE TO SPECIAL ORDER

Key Features

Low temperature coefficient of resistance -10ppm/ºC (0ºC to 40ºC).

Maximum current - 10mA.

Resistance tolerance - (Initial resistance accuracy)±0.005%.

24 resistances equivalent to the full range ofindustrial temperature sensors.

Each resistance adjustable during calibration.

Traceable certificate with each simulator.

-100ºC

0ºC

50ºC

125ºC

150ºC

350ºC

-50ºC

10ºC

75ºC

130ºC

200ºC

400ºC

-25ºC

25ºC

100ºC

135ºC

250ºC

500ºC

-10ºC

37ºC

120ºC

140ºC

300ºC

600ºC

Page 51: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/51

Resistance Thermometer Simulator

PLEASE NOTE:

We offer other Resistor Ranges including the models 456,836, the SRA and the SRB ranges.

For more information please contact Isotech, or visit ourwebsite www.isotech.co.uk

Model No. 862 24 ValueResistance Thermometer Simulator

Temperatures Refer to previous page for standardSimulated temperatures, others to special order

Devices Pt 100Simulated 100Ω at 0ºC

Uncertainties ±110 ppm

Stability 10ppm/ºC

Switch Heavy duty gold platedcontact resistance less than 0.01Ω

Overall dimensions Height 110mmWidth 110mmDepth 140mm

Weight 0.62kg

How to Order Model 862Please state to either IEC751

or SAMAPlease state ifUKAS Certification is required

Diagram of model 86224 Value Resistance Thermometer Simulator

Model 862 Resistance Thermometer Simulator to IEC751

Model 862 Resistance Thermometer Simulator to SAMA

UKAS Calibration UKAS Calibration available to Order

Options

Model 862 - 24 Value

Page 52: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/52

SeeTable

Thermocouple Simulator870 Isocal

A method of accurately calibrating any thermocouple temperature measurement or control system isavailable from Isothermal Technology Limited. Replacing the normal calibration “Trolley” containingMicrovolt Generator Ice Point Reference Chamber and Digital Voltmeter by a pocket sized calibrationunit complete with its own battery and cold junction compensation.

To use the unit, simply follow the steps outlined below:

1. Disconnect the normal Measurement Thermocouple.

2. Substitute the Isocal unit.

3. Select one of ten switch preset temperatures on the front of the Isocal.

4. Check that the measurement or indicator system reads the preset temperature.

The ability of the Isocal unit to simulate any preset temperature with and without cold junctioncompensation, enables the user to substitute the electrical equivalent of the thermocouple output towithin a fraction of a degree and inject the signal into almost any kind of measuring system. Modelsare available for all common thermocouple material to IEC 584 tables.

Key Features

10 position Thermocouple Simulator.

Simple to use.

Certified.

Built-in Cold Junction.

Page 53: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/53

Thermocouple Simulator870 Isocal

Model No. ISOCALThermocouple Simulator

Reference Effectively 0ºCTemperature

Accuracy Better than 0.05%of full scale

Stabilisation Time 5 minutes

Long Term Stability Better than 0.05%

Ambient 0ºC to 55ºCTemperature

Temperature ±0.01% per ºCStability

Power PP3 9V Battery1 year intermittent operationConsumption less than 5mA

Overall dimensions Height 45mmWidth 90mmDepth 160mm

Weight 300g

How to Order 870 IsocalPlease specify thermocouple type(for example Isocal K)Please state ifUKAS Certification is required

UKAS Calibration UKAS Calibration available to Order

Options

Standard Combination Offered Designation Preset Values ºC1 2 3 4 5 6 7 8 9 10

Nickel Chromium vs. Nickel Aluminium* K 0 100 200 300 400 500 600 700 800 1000

Nickel Chromium vs. Copper Nickel* E -100 0 100 200 300 400 500 600 700 800

Iron vs. Copper Nickel* J 0 50 100 150 200 300 400 500 600 700

Copper vs. Copper Nickel* T -50 0 20 37 60 100 121 150 200 300

Platinum vs. Platinum 10% Rhodium S 0 800 900 1000 1100 1200 1300 1400 1500 1600

Platinum vs. Platinum 13% Rhodium R 0 800 900 1000 1100 1200 1300 1400 1500 1600

* Conformance certificate available

Page 54: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/54

& Accessories - IntroductionIndicators, Probes

PRT andPRT / Thermocouple Indicators

Our TTI range are True Temperature Indicators, thismeans individual calibrated sensors can have theircalibration values programmed into the TTI. Theinstrument then digitally 'calibrates' to suit the probeand the TTI shows the corrected value accordingly.Additional manual measurements are not required tocompute the precise temperature readings since theseare displayed correctly, miscalculation errors areavoided.

These instruments both have two channel operation,which allows for a calibrated reference probe to beused on one channel, with the unit under test on thesecond, permitting simultaneous comparisoncalibration which eliminates many errors compared to asingle channel system that calibrates sequentially.

When several sensors need to be switched Isotech hasa range of Selector Switches to accompany theindicators, along with software to ease and automatethe instruments use.

True SurfaceTemperature Measuring System

This is a true temperature indicator for use with surfacetemperature measurement, ideal for use with the smallhot-plate model 983, see page 44.

Page 55: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/55

& Accessories - IntroductionIndicators, Probes

Semi Standard Resistance Thermometersand Thermocouples

This section includes a range of “Semi Standard”thermometers that can be used with the TTI’s. Theseprecision semi standards are more rugged andaffordable than the standard thermometers featured inDatabooks 1 and 2 – and ideal for most industrialapplications.

FixedResistors

Isotech have a miniature resistor with an ultra lowtemperature coefficient, model 836, with accuracies of±0.005% and temperature coefficients of less than1ppm. These resistors offer an outstanding cost toperformance ratio and will find use alongside a modelfrom our TTI range and in other areas of industrialcalibration.

Page 56: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/56

2315ºCto

-200ºC

High Accuracy PRT and Thermocouple Thermometer

TTI-7

The TTI-7 is a very high accuracy multi purpose digital thermometer for both platinum resistancethermometers and thermocouples. Laboratory users will welcome the features to eliminate Thermal EMFErrors and Self Heating Errors along with provision to store the calibration data of up to 20 PRT probes.The rugged aluminum case, internal battery pack and integrated power supply ensure reliable portablefield use for demanding measurement applications all at great value for money.

Dual Channel input allows a probe on Channel B to be calibrated against a standard on Channel A -directly compare any combination of PRT and Thermocouple. The TTI-7 supports ten thermocoupleTypes, B,C, D, E, J, K, N, R, S and T and Pt100 thermometers. Connect up to 16 sensors via the optionalswitchboxes, Model 954 and 958.

Data Logging and Statistical AnalysisThe TTI-7 includes an inbuilt data logger internally storing up to 4,000 date and time stamped readings. Recall thedata from the front panel or send to a PC or Printer via the PC interface which is included as standard. The powerfulmath function enables statistical analysis of the captured data, mean, max, min, peak and standard deviation.

UsabilityEase of use, password protected digital calibration and a large clear backlit LCD graphics panel ensure the TTI-7 is adelight to use. Resistance thermometer connections are via LEMO connectors. Both sub miniature thermocouple andstandard thermocouple plugs are accepted directly into the thermocouple inputs with no need for further adapters.

Why the TTI-7?The TTI-7 has the features you need for high accuracy temperature measurement. With resistance thermometers usedat high temperatures unwanted thermal EMFs are generated, the TTI-7 can take two measurements switching thepolarity then computing the average to eliminate this error source. Many other instruments lack the ability to eliminatethermal EMFs. The thermal EMF error can be greater than the quoted accuracy of an instrument, if you need smallmeasurement uncertainty for high temperature PRT work you need this feature.

High accuracy, highest accuracy is for Pt100 inputs, the TTI-7 Uncertainty of Measurement (1 Year) in the range -100°Cto 500°C is 0.01°C. Watch for specifications that quote the value at -100°C and then get larger as the temperaturerises. The TTI-7 is optimized over the most frequently used and useful temperature range. For thermocouplemeasurements the automatic CJC is far better than 0.1°C at 20°C. Great design care was taken, both thermocoupleinputs are measured with separate Pt100 sensors. This approach gives outstanding CJC performance, again a pointto check against other instruments which can have significantly less performance.

Key Features

Accuracy to 0.01ºC, 0.001ºC Resolution (Pt100Inputs).

PRT and Thermocouple Inputs True Dual Input.

Eliminate Unwanted Thermal EMFs withcurrent reversal.

Store 4000 Measurements.

PC Interface and Software Included, RS232.

Portable 10 Hours Use from Internal Battery.

For more details please ask for ourBuyers Guide or download fromhttp://www.isotech.co.uk/tti-7

Page 57: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/57

High Accuracy PRT and Thermocouple Thermometer

TTI-7

Inputs Thermocouples via sub miniature andstandard connectors. Reference JunctionCompensation - Automatic with internalsensor, or with external Pt100 probe. PRTsLemo Socket.

Working 0ºC … 40ºC rel. humidityTemperature 80% max non condensing

Storage -20ºC … +50ºC.Temperature

Main Supply 100/120/220/240 Volts +10% -13%47 … 63Hz max. 40VA

Dimensions Height 110.3mmWidth 219mmDepth 315mm

Weight 8kg

Battery Sealed lead acid, rechargeable cell givingapproximately 10 hours continuousoperation. Internal battery charger.

Model No. TTI-7

Temperature Depending on SensorRange -200 to 2315ºC

Indicator units ºC, ºF, K

Display LCD Graphics Panel, 240 x 64 Dot withLED backlight contrast control viakeyboard

Maths Display Min / Max, Peak to Peak andStandard Deviation

PC Interface RS232 and Software Included

Data Logging Includes a data logging function, enablingup to 4000 single channel (2000 dualchannel) readings to be stored togetherwith a date and time stamp.

The stored values can be recalled to theinstrument display, downloaded to a PCfile or printer.

Pt100 Range Resistance Current Resolution Uncertainty 1 year(ºC) (Ohm) ºC ºF K @20ºC ±5ºC

-200 … -100 18 … 60 1mA 0.001 0.02ºC

-100 … +500 60 … 280 1mA 0.001 0.01ºC

+500 … +800 280 … 450 1mA 0.001 0.02ºC

Type Range Resolution Display Uncertainty Uncertainty TemperatureºC ºC ºF K Resolution µV @20ºC ±5ºC 1 Year @20ºC ±5ºC 60 Days Coefficient / ºC

B +250ºC to +1820 0.01 1.0 ±(0.025% Rdg + 0.006%FS)* ±(0.02% Rdg + 0.006%FS)* 7ppm Rdg + 6ppm FS

C 0 to +2315 0.01 1.0 ±(0.075% Rdg + 0.005%FS) ±(0.05% Rdg + 0.005%FS) 7ppm Rdg + 6ppm FS

D 0 to +2315 0.01 1.0 ±(0.075% Rdg + 0.005%FS) ±(0.05% Rdg + 0.005%FS) 7ppm Rdg + 6ppm FS

E -200 to +1000 0.01 1.0 ±(0.026% Rdg + 0.004%FS) ±(0.01% Rdg + 0.004%FS) 7ppm Rdg + 6ppm FS

J -210 to +1200 0.01 1.0 ±(0.03% Rdg + 0.005%FS) ±(0.008% Rdg + 0.005%FS) 7ppm Rdg + 6ppm FS

K -200 to +1372 0.01 1.0 ±(0.035% Rdg + 0.006%FS) ±(0.01% Rdg + 0.006%FS) 7ppm Rdg + 6ppm FS

N -200 to +1300 0.01 1.0 ±(0.035% Rdg + 0.005%FS) ±(0.01% Rdg + 0.005%FS) 7ppm Rdg + 6ppm FS

R -50 to +1768 0.01 1.0 ±(0.02% Rdg + 0.015%FS) ±(0.005% Rdg + 0.015%FS) 7ppm Rdg + 6ppm FS

S -50 to +1768 0.01 1.0 ±(0.02% Rdg + 0.015%FS) ±(0.005% Rdg + 0.015%FS) 7ppm Rdg + 6ppm FS

T -200 to +400 0.01 1.0 ±(0.025% Rdg + 0.015%FS) ±(0.005% Rdg + 0.015%FS) 7ppm Rdg + 6ppm FS

UNCERTAINTY OF MEASUREMENT

TC input for external CJC, automatic CJC is better than 0.1ºC at 20ºC, typically 0.01ºC / ºC over the range 0ºC to100ºC *Apply to readings above 600ºC

Page 58: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/58

849.99ºCto

-199.99ºC

For 3 & 4 Wire PT100’sTTI-6

The TTI-6 is a high precision portable thermometer for metrology and other exacting laboratoryapplications, the TTI-6 is a proven instrument used World-wide as a laboratory and site standard. It isparticularly suitable as the reference standard for temperature calibration baths.

The TTI-6 Digital Thermometer utilises highly advanced microprocessor circuit design to achieveexceptional measuring accuracy, linearisation conformity and stability in a versatile but easy to useconfiguration.

Based on a high resolution 20 Bit Analogue to Digital convertor, all measurement computations areperformed digitally without drift. The 5 LED display provides a readout to 0.01ºC over the entire-199.99ºC to +849.99ºC range; alternatively ºF, Kelvin or Ohms values can be displayed up to 999.99units.

Single or dual Pt100 3 or 4 wire sensors are accepted, the TTI-6 will automatically recognise and select3 or 4 wire mode. Display of input A, B or A - B (differential) can be selected; a differential “zerofunction” allows sensor accuracy differences to be eliminated for accurate differential readings.

Calibrated sensors can have their calibration points programmed into the TTI-6 via a personalcomputer using Isotech Software. The instrument then digitally self calibrates to the probe(s)providing corrected temperature readout to give optimum system accuracy.

A common application is for the comparison calibration of a sensor on channel B against a referencestandard on channel A. With one of the Selector Switches several sensors can easily be compared toa standard probe on A.

The instrument includes a UKAS certificate as standard (simulation calibrated). Isotech also offers acomplete system calibration (probe and instrument combined). Refer to Databook 5 for details.

Key Features

Two channel RTD indicator.

RS232 and analogue output as standard.

Correction facility for optimum systemaccuracy.

Includes UKAS certificate and Cal NotePadSoftware.

TTI-6 shown with optionalTerminal Adaptor and Probe

Page 59: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/59

For 3 & 4 Wire PT100’sTTI-6

Measurement units ºC, ºF, K, Ω,

Measurement Input A or B or A-B (differential)modes Null facility in A-B mode

Custom calibration Up to 10 calibration values can beallocated to (via PC DOS software)channels A and B.Values are retained in non-volatilememory until replaced by user

Null Function Corrects differential temperature readoutbetween two sensors to zero

Display 14mm LED, 5 digit, 999.99 range

Front panel 5 x membrane keys for user functionscontrols

Input 2 x Pt100 via ‘Lemo’ connectorsConnections

Analogue output Analogue 0 to 1 Volt d.c. Between(standard) programmable (standard) lower and upper

set limits representing channel A, B orA-B. Accuracy 0.5% of reading.Non-isolated

Power Internal chargeable batteries.Supply Mains 220/240V 50/60Hz, adaptor

included. Battery life up to 12 hours withfull charge dependent on pattern of usage.Charger requirement 10-11.5V d.c.1A.10W nominal, max 20W

Overall dimensions Height 145mmWidth 66mmDepth 240mm

Weight 1.5kg

How to Order TTI-6

Model No. TTI-6

Linearisation ±0.01ºCConformity

Stability 0.0025ºC / ºC ambient change(vs ambient

temperature)

Pt100 Sensor 0.5mA nominalCurrent

Resolution of 0.01ºC, K, ºF, Ωdata display

Note: All values arevalid for a nominal240V 50Hz supplyand 20ºC ambienttemperature (±2ºC)

Input Type Temperature Range Overall accuracy

Pt100 -199.99ºC to 849.99ºC ±0.02ºC ±1 digit for range -200ºC to +500ºC To IEC 751, Amendment 2, ±0.005% reading ±1 digit for range 500ºC to 850ºC(1995. -199.99 to 849.99ºC. Ro=100Ω.Two input channels, each 3 or 4 wire connectionwith automatic recognition (with manual override)

Note:Refer to accessory pagesfor probe types and

system performance.

Carrying Case 931-22-51 Soft vinyl carrying case

Options

Page 60: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/60

True Temperature Indicator AccessoriesTTI Accessories

TTI-7

TTI-6

SYSTEM UNCERTAINTYT.T.I.7 ISOTECHS UKAS UNCERTAINTIES (2 sigma)

Using semistandards

UK

AS

UN

CE

RTA

INTI

ES

±m

ºC

TEMPERATURE RANGE OF VARIOUSSEMISTANDARDS THAT CAN BE USED WITH T.T.I.7

SYSTEM UNCERTAINTYT.T.I.6 ISOTECHS UKAS UNCERTAINTIES (2 sigma)

Using semistandards

TEMPERATURE RANGE OF VARIOUSSEMISTANDARDS THAT CAN BE USED WITH T.T.I.6

T.T.I.6 USINGPRT SEMISTANDARDSwith UKAS CALIBRATION

Page 61: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/61

True Temperature Indicator AccessoriesTTI Accessories

Overall dimensions Height 68mm (including feet)Width 140mmDepth 185mm (including connectors)

Weight 0.660kg

How to Order 956 Terminal Adaptor

Terminal Adaptor

Our TTI range use high quality 'Lemo' connectors for the Pt100 inputs. This simple accessory provides 4mm TerminalPosts for the connection of bare wires, spade terminals or 4mm plugs - useful if a lot of probes are going to be usedwith the instrument.

Page 62: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/62

Platinum Resistance ThermometersSemi-Standards

670ºCto

-50ºC

In most industrial laboratories the costly and mechanically fragile SPRT is unnecessary. For theseapplications the uncertainties may be larger, and hence the Isotech range of semi-standards may bea more practical choice. Usually made in stainless steel sheaths 6mm in diameter, with an Ro value of100Ω and an (R100-R0)/R0 of 0.385. A number of configurations are available since particularcharacteristics might be required such as short sensing length, fast response, or applicability to aparticular temperature range.

In this collection of semi-standards Isotech has examined not only the uses that customers put themto but also the equipment that is used to calibrate them. Most units are now a little longer and havea slightly wider temperature range than our previous products. The stability of the semi-standard willdepend upon its range of usage; your semi-standard should remain with those uncertainties betweensuccessive annual calibrations, unless you drop it or otherwise mechanically shock it. Our semi-standards are now fitted with a handle and strain relief bush for the cable, except where otherwisestated. All are made with our own high quality 100 ohm detectors and completely assembled andcalibrated in-house in our UKAS accredited and supervised Laboratory. Calibration is available withfixed points for the best uncertainties and to give the highest confidence possible.

Please note that stem conduction accounts for the biggest sources of errors during calibration.Remember that for a given immersion depth stem conduction is dependent on the temperaturedifference between the sensing element and ambient temperature, and on the total conductance ofthe thermometer assembly. It is sometimes thought (mistakenly) that a shorter thermometer will besubject to less stem conduction.

Advantages

Ideal for all Industrial applications.

High Stability.

Low Drift.

Can be supplied with UKAS Calibration.

Selection of Temperature Ranges.

Choice of lengths and diameters.

Specials available to special order.

Long life.

Maximum accuracy for many of our Industrialproducts.

Page 63: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/63

Platinum Resistance ThermometersSemi-Standards

Termination OptionsBW Bare WireTTI Lemo Connector to suit TTI-5, TTI-6, TTI-7DB Connector for Dry Block Calibrator Site Indicator

For independent use (without Dry Block Calibrator) please referto Databook 5 for Calibration ranges and uncertainties

Model Number

Diameter (A)Length (B)

Sensing LengthHandle (C)Cable (D)

TemperatureRange (ºC)

Application

Features

935-14-61

4mm300mm6mmYes

2m PTFE4 wire

-50ºC to+250ºC

FastResponse

Small StemConduction

935-14-13

6mm350mm25mm

Yes2m PTFE

4 wire

-196ºC to+250ºC

LowTemperature

Use

Copper EarthScreen

Model Number

Diameter (A)Length (B)

Sensing LengthHandle (C)

Cable (D)

TemperatureRange (ºC)

Application

Features

935-14-82

4mm210mm6mmNo

(Angled Pot)1.5m PTFE

4 wire

-50ºC to+250ºC

EuropaVenusCalisto

GeneralPurpose

935-14-85

6mm420mm25mm

No(Angled Pot)540mm PTFE

4 wire

-50ºC to+250ºC

Oceanus-6

GeneralPurpose

935-14-16

6mm450mm25mm

Yes2m PTFE

4 wire

-100ºC to+450ºC

GeneralUse

316 stainlesssteel sheath

935-14-98

4mm300mm8mmNo

2mm PTFE4 wire

-50ºC to+350ºC

FastResponse

Small StemConduction

935-14-72

6mm375mm25mm

No2m PTFE

4 wire

-50ºC to+670ºC

GeminiJupiter

Fits lid ofCarry-case

935-14-95

6mm450mm25mm

Yes2m PTFE

4 wire

-100ºC to+670ºC

GeneralUse

Medusa

Metal alloysheath

Model No. Refer to Chart

Temperature Refer to ChartRange

Ro 100 ohms ±0.05 ohms

Alpha 0.003850 ±0.000005

Stability 0.010 ohm/year

Calibration A UKAS Calibration Certificate can be provided at extra cost

Recommended 1mACurrent

Self-heating error 0.004ºCat 1mA

Overall dimensions Refer to Chart

How to Order Please Specify Model Type andTermination Option(for example 935-14-13/BW)Please state whetherUKAS Certification is required

Page 64: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/64

ThermocouplesSemi-Standard

1300ºCto

0ºC

Semi-standard thermocouples without cold junction.

This range is based on our popular standard thermocouples. We receive many requests for differentsizes, length and thermocouple materials. From our experience in calibration Isotech have been ableto select those thermocouple combinations and sizes which give the best and most stable results.

Below is an extract from Isotech's UKAS schedule that shows the laboratory's best capability tocalibrate thermocouples. The uncertainty of a particular thermocouple will depend on factorsincluding how it is calibrated. System uncertainties will also vary with which indicator and calibrationbath that the sensor is used with. Most of the products shown here include system uncertainty graphs;please contact Isotech for details of any combination not shown. There is also a range of laboratorygrade and standard thermocouples, Databooks 1 and 2.

These semi-standards do not have a cold-junction. This feature is usually incorporated in the indicatorto which the semi-standard is connected. We offer the type N, R or S semi-standards either with1 meter leads, with standard plug or miniature plug.

Key Features

Type N and Noble metal thermocouples givethe best long life, stability and reproducibility.

No Hysteresis.

Wide temperature range.

Ideal for all Industrial applications.

Minimum uncertainties in many of ourIndustrial products.

Can be supplied with UKAS Calibration.

(K=2)Item Measured Quantity Temperature Range Best measurement capability

Instrument or Gauge expressed as an uncertainty (±)

1 Temperature -50ºC to +660ºC 0.5KPlatinum Thermocouples 0ºC to 1100ºC 0.8K

Above 1100ºC to 1300ºC 2.2K

2 Other Thermocouples -196ºC 0.3K-80ºC to 300ºC 0.25KAbove 232ºC to 420ºC 0.3KAbove 420ºC to 660ºC 0.4KAbove 660ºC to 1100ºC 0.8KAbove 1100ºC to 1300ºC 2.2K

3 Compensating and -25ºC to 200ºC 1KExtension cable

Page 65: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/65

ThermocouplesSemi-Standard

For independent use (without Dry Block Calibrator) please referto Databook 5 for Calibration ranges and uncertainties

Model No. Refer to Chart

Temperature Refer to ChartRange

Calibration A UKAS Calibration Certificate canbe provided at extra cost

Overall dimensions Refer to Chart

How to Order Please Specify Model Type(for example 935-14-65)Please state whetherUKAS Certification is required

Model No. Diameter (A) Length (B) Termination Range (ºC) Application Type

935-14-91 5mm 300mm 1M Cable 0 to 1300ºC Pegasus R*Miniature Plug

935-14-88 5mm 300mm 1M Platinum 0 to 1300ºC Higher R*Cable Accuracy

935-14-63 3mm 300mm 1M Cable 0 to 1300ºC Gemini 700 NMiniature Plug Jupiter 650

935-14-64 3mm 300mm Miniature 0 to 1300ºC General NPlug Purpose

935-14-65 3mm 300mm Standard 0 to 1300ºC General NPlug Purpose

935-14-66 3mm 500mm 1M Cable 0 to 1300ºC General NMiniature Plug Purpose

935-14-67 3mm 500mm Miniature 0 to 1300ºC General NPlug Purpose

935-14-68 3mm 500mm Standard 0 to 1300ºC General NPlug Purpose

* Type S available to special order

Page 66: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/66

DocumentationSemi-Standards

Isotech have generated a long history of many of our semi-standards.

Here are a few documented facts:

The 935-14-95 model has the widest temperature range and in consequence is likely to suffer the largest changes incharacteristics.

Guy Snelling sent the following email about the 935-14-95.

Page 67: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/67

DocumentationSemi-Standard

The most severe test of a thermometer is to cycle it from its minimum temperature of use to its maximum temperature.

Below is tabulated the annealing drifts of a typical 935-14-95.

Annealing Range - 0ºC to +670ºCwith 4 hour dwell at 670ºC

Cycle Number Shift (mK)

1 -0.8

2 -2.2

3 +0.5

4 -0.5

5 -2.3

6 -1.2

7 -0.2

8 +0.6

Total Shift over 8 Cycles: -6.1mK

Mean Shift per Cycle: -0.76mK

Page 68: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/68

8 Way Selector SwitchSelector Switch

Isotech produces two eight way selector switches, one for resistance thermometers Model 954 andModel 958 for thermocouples. These switches have been designed for use with a model from our TTIrange.

The switches allow easy selection of connected sensors. They can be operated from either the frontpanel switch or from an RS232 interface that is provided as standard. Channel status is indicated viafront panel LEDs. The Selector switches can be located adjacent to the sensors being calibrated,giving more flexibility than a permanently connected or stacked system.

The PRT Switch has 4mm Terminal Posts that can accept bare wires or 4mm plugs. The thermocoupleswitch has eight miniature thermocouple connectors. These thermocouple connectors are thermallybonded to a platinum resistance thermometer that measures the temperature of the connector andhence the “cold junction”. The TTI range temperature indicators feature the ability to measure aremote cold junction and this permits a mixture of thermocouple types to be connected through thebox.

The I-Cal software supports Switchbox models 954 and 958 and, for automatic operation, two boxescan be connected together with a “master / slave” lead allowing them to be controlled from a singleRS232 port and up to 16 sensors to be switched. The software can automatically switch between theboxes and connect the appropriate output to the TTI. This 16 channel operation is not convenientwithout the software and manual operation of two boxes together is not recommended. SeeDataBook 2 for a 16 channel manual switch, the Model 947.

Advantages

Use with TTI-6 and TTI-7 easily switch up toeight sensors manually or with RS232.

RTD and Thermocouple Models.

Use with I-Cal Software for automaticswitching and temperature calibration, add asecond box (either type) to calibrate up to 16Sensors.

Switches are stand-alone allowing them to bepositioned anywhere in a laboratory for mostefficient operation.

Page 69: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/69

8 Way Selector SwitchSelector Switch

Model No. 958 TC Selector Switch

Channels Eight - two wire (two pole)

Control Front panel switchAnd RS232(Also compatible with Isotech VLT system)

Connectors Miniature Thermocouple Connectors

Internal Circuit <250mΩResistance

Thermal EMF, 2µV after 1 minute of channel settypical 6µV after 30 minutes of channel set

Reference Junction 100Ω 1/10 Din Pt100Measuring Device

Thermal Coupling <0.2ºC**Basis of test. At ambient 20ºC ±2ºC the internalPt100 agreed with the connected thermocouples to±0.2ºC (including all measurement errors) usingIEC584-1995 and IEC751-1995. The uncertainty of thistest was ±0.3ºC which includes the reproducibility ofthe test thermocouples.

Power 5 VDC100-250 VAC, 50 / 60HzPower Supply Included

Overall dimensions Height 64mmWidth 141mmDepth 165mm

Weight 1kg

How to Order 958 TC Selector Switch

Model No. 954 RTD Selector Switch

Channels Eight - four wire (four pole)

Control Front panel switchAnd RS232(Also compatible with Isotech VLT system)

Connectors 4mm Terminal post

Internal Circuit <250mΩResistance

Thermal EMF, 2µV after 1 minute of channel settypical 6µV after 30 minutes of channel set

Power 5 VDC100-250 VAC, 50 / 60HzPower Supply Included

Overall dimensions Height 91mmWidth 141mmDepth 165mm

Weight 1kg

How to Order 954 RTD Selector Switch

Page 70: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/70

Miniature Fixed Resistor836 Resistor

SeeTableBelow

Isotech produces a miniature resistor with ultra-low temperature coefficient and ultra-high stability.

This is achieved because the resistors are oil filled and hermetically sealed.

The function of hermetic sealing is to eliminate the ingress of moisture and oxygen both of which playa role in both short and long term degradation of unsealed resistors. A further enhancement in bothshort and long term stability is achieved by oil filling. The oil also acts as a thermal conductor allowingthe device to accept short periods of overload without degradation.

With accuracies of ±0.005%, resistance range from 5 ohms to 3.3 megaohms and long term drift ofless than 5ppm, these devices are virtually secondary standards that can be carried in sets for daily orperiodic calibration of factory systems.

RESISTANCE VALUESWe can supply the value you choose ±0.1% between a minimum of 5 ohms and a maximum of 3.3megaohms. However we bulk buy and keep in stock the following standard values:10Ω, 25Ω, 100Ω, 1000Ω, 10,000Ω

Key Features

The most precise and stable resistors availablefor Industrial laboratories.

Impervious to harmful environments - oil filled.

Nominal temperature coefficients of resistance -+0.6ppm/ºC (0ºC to +25ºC); -0.6ppm/ºC (+25ºCto +60ºC).

Power Rating - 0.5 watt at +25ºC.

Resistance tolerance - (initial resistanceaccuracy): ±0.005%.

Page 71: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/71

Miniature Fixed Resistor836 Resistor

Please Note:

We offer other Resistor Ranges including the models 456, theSRA and the SRB ranges.

For more information please contact Isotech, or visit our websitewww.isotech.co.uk

UKAS CALIBRATIONFor the highest quality traceability we recommend that the 836 be UKAS Certified.

We can offer the following 2 Sigma Uncertainties:

Measured QuantityInstrument or GaugeRange

DC RESISTANCE0.1 Ω to 10 Ω10 Ω to 250 Ω250 Ω to 1000 Ω1 KΩ to 1 MΩ1 MΩ to 10 MΩ

AC RESISTANCE2.5 Ω to 400 Ω400 Ω to 1000 Ω

Frequency

75 Hz75 Hz

Best measurementCapability expressedas an ExpandedUncertainty (k=2)

±0.3ppm+0.1µΩ±0.3ppm+2.5µΩ±0.4ppm+10µΩ±20ppm±55ppm

±15ppm±100ppm

Remarks

Resistors suitable foroil immersion can bemeasured over therange 10ºC to 30ºC

The uncertainties canonly be realised forresistors of suitableAC characteristics

Model No. 836 Miniature Fixed Resistor

Power Rating 0.5 watt

Nominal +0.6ppm/ºC (0ºC to +25ºC)Temperature -0.6ppm/ºC (+25ºC to +60ºC)

Coefficientof Resistance

Resistance (Initial Resistance Accuracy)Tolerance ±0.005%

Resistance Range 5 ohms to 3.3 megaohms

Current Noise <0.010µV (RMS) / Volt of applied voltage

Thermal EMF 0.1µV/ºC maximum0.05µV/ºC typical

Connections Screw Terminal Posts

Stability Typically 1ppm per year at 1mA

Overall dimensions Height 30mmWidth 89mmDepth 58mm (including terminals)

Weight 90g

How to Order 836 Miniature Fixed ResistorPlease state Ohms Value RequiredPlease state if UKAS Certification isrequired

Page 72: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/72

Temperature Measurement System944 True Surface

350ºCto

30ºC

The fundamental problem with surface temperature measurement is that it is subject to large stemconduction errors, also because heat conducted from the surface of the hot-plate causes a localisedcold spot to be created which means that the temperature indicated by the hot plate is not necessarilythe temperature at the point of measurement.

An ideal system would not disturb the heat-flux from the hot-plate.

During 1993 such a system was described (ref. “Progress in Contact Thermometry” 1993 B. D. Foulis)and Isotech have the inventors permission to make and market the device World-wide.

Principal of Operation

A fine wire type N thermocouple is used as the surface temperature sensor, a second junction 2 to3mm along the thermocouple, senses the temperature difference due to heat flux along the sensor.A heater heats the thermocouple stem until the temperature gradient is zero, thus creating ameasurement without stem conduction, or disturbance of the hot-plate's surface.

The 944 can be used with the Isotech Small Hot Plate model 983, see pages 44 - 45. A traceablecalibration certificate is available to order.

Key Features

Indicates true surface temperaturemeasurements digitally to a resolution of0.1ºC or 0.1ºF.

Temperature range of 30ºC to 350ºC.

Page 73: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/73

Temperature Measurement System944 True Surface

Model No. 944 True Surface TemperatureMeasurement System

Temperature Range 30ºC to 350ºC

Resolution 0.1ºC or 0.1ºFof display

Stability ±1ºC

Accuracy ±2ºC with TRACEABLE Certification±5ºC without Certification

Probe Assembly Probe Diameter 7.5mmProbe Length 150mmLead Length 850mm

Power Supply 100V - 120V, 50 / 60Hz or200V - 240V, 50 / 60Hz

Overall dimensions Height 90mmWidth 153mmDepth 265mm (excluding plugs)

Weight 4kg

How to Order Model 944 & Probe 935-14-81Please state supply voltage requiredPlease state if Calibration is required

True surface temperature measurements

Page 74: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/74

IntroductionSoftware

SoftwareNewly developed software provides ease of use and routes to fully automatic temperature calibration.

CalNotePadIncluded as standard with the calibration baths and TTI indicators. This permits you to connect both an indicatorand the calibration bath to the PC. You can view data on clear configurable chart displays, log data to file andcontrol the calibration bath.

I-calA software application that can automatically calibrate up to 16 temperature sensors, it provides an expandablelow cost route to automatic calibration with I-cal Capture you can store pictures (along with time and temperaturedata) to enable, for example, dial thermometers to be calibrated automatically; return to a set of images of thedial captured at each calibration temperature!

Virtual Lab TechnicianVLT provides a turnkey solution for the automatic calibration of temperature sensors. Consisting of a PCWorkstation, Selector Switch, TTI Indicator and calibration bath(s) it provides a combination of computer hardware,calibration equipment and software to automate the calibration of temperature sensors, further details inDatabook 2, a demonstration disk is available on request and full details can be found at www.isotech.co.uk

The Dry Block Calibration Baths included in this DataBook have been enhanced to feature serialcommunications as standard. The baths have a serial interface (RS422) as standard, this allows anumber of calibration baths to be connected together in a daisy chain.

A special lead assembly is provided which converts the RS422 to a PC's standard RS232 port.

This innovative approach allows you a simple trouble free way of connecting a number of baths to asingle PC's RS232 port.

Page 75: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/75

IntroductionSoftware

Comparison Chart CalNotePad VLT I-cal I-cal Capture

Included as standard Yes No No No

Sensor Calibration With Manual Bath Operation Yes Yes Yes Yes

Automatic Sensor Calibration No Yes Yes Yes

Maximum Number of Sensors 4 16 16 16

Maximum Number of Calibration Baths 1 4 1 1for Automatic Operation

Internal Database with Certificate Printing No Yes No No

Save Results to File Yes Yes Yes Yes

Capture Images with Camera No No No Yes

Photograph of I-Calbeing used with a Gemini 700Dry Block Calibrator and PC

Screenshot ofCal NotePad

Page 76: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/76

SoftwareCal NotePad

The purpose of Cal NotePad (CNP) is to automatically log and display the temperature of an Isotechcalibration bath together with the unit under test. Cal NotePad can be used with baths (or indicators)without PC interfaces by the user typing in values from the keyboard. The Cal NotePad can be used toidentify the operator and the unit under test. With the click of a button data is logged with time information,it is also possible to log continually.

The calibration bath temperature can be changed from the PC or from the calibration baths controller - CalNotePad will display the temperature changes as they occur on the re-scalable chart display.

For traceable calibration the unit under test should be compared to a calibrated standard thermometer. CalNotePad can record the actual temperature of the bath from either the in-built indicator of an Isotech SITEmodel or from a variety of external instruments see list. If the external instrument has two channels (e.g.Isotech TTI) then the unit under test may be connected to channel B for logging with CNP. Alternatively thevalue can be typed in from the keyboard. Similarly the calibration bath controller value, actual temperature,SITE indicator value or unit under test value may also be entered manually.

Then the manually entered data is combined with that gathered automatically and the resultant file can beopened in an external application such as Excel for the preparation of reports, certificates etc.

Cal NotePad is designed for ease of use, it will give a chart of the system. When the operator determinesthe system is stable - easily seen from the chart then clicking a button will record time, operator, serialnumbers of unit under test along with controller and indicator values.

Cal NotePad can be used for semi-automatic calibration, see I-cal for a fully automated calibration solution.

New in Version 2 :-

Temperature Programming.

Program the bath to change temperature with time.

Ramp to Temperature.

Program the bath to ramp between limits ideal for liquid in glass thermometry.

Open new chart for individual channels.

Key Features

Included with the PLUS models, the Isocal-6range and Dry Block Calibrators.

Provides Sophisticated Real Time Chart Display.

Record Isotech Calibration Bath and IndicatorData.

Export data for Excel, Word, etc.

Page 77: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/77

SoftwareCal NotePad

How to Order Cal NotePadSupplied on CD Rom witha comprehensive manual

System Requirements CNP requires a PC with Windows 95/98/2000/XP 5Mb free hard drive space and a minimum of 8Mb

RAM CNP is compatible with the following Isotech

calibration equipment:-

Next Generation Dry Block Baths with RS232 orRS422 including:- Venus 2125 + 2150, Gemini 550 +700, Jupiter 650, Venus, Gemini, Jupiter, Pegasus LABmodels, ISOCAL-6 models with RS232 or RS422

Isotech's Professional Comparison Baths with RS232or RS422 - most models supported including:- FluidisedFurnace, Spherical Furnace, 915 + 814 Liquid Baths

Temperature Indicators: Isotech T.T.I.1, Isotech T.T.I.2,Isotech T.T.I.5, Isotech T.T.I.6 and T.T.I.7

In addition the following third party instruments aresupported:- Labfacility Tempmaster, Labfacility Labcal +,Fluke 45 DVM, in built conversion for type K, J, E, R, S, T,N, B (external CJC required) or Volts/Ohms display,Cropico, 3000 Series, and similar models with the samecommand protocol, e.g. ASL F150, Tinsley 5885 etc.

Manual Entry: - Any instrument data can be typed into Cal NotePad.

User Cal.Unit under Man Man Man ManDate Time Name BathTest No.

Controller Indicator ChnlA ChnlBCh1 Ch2 Ch3 Ch4Here Number

Type K TC1 May 1999 10:01:45 A User ITL12345-1 100 99.7 99.698 99.3 0 0 0 0Plant 45

Re-scalable chart shows values from calibration bath and indicator

Up to four values from connectedinstruments

Log a pre-set number of points

Enter manual data here

Log a single set of data

Sample Output Data

Page 78: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/78

SoftwareI-Cal

I-cal is a new software application designed for automatic temperature calibration using Isotech products. Itis affordable, easy to use and modular.

In the simplest form it can be used with a SITE Dry Block Calibrator to automatically calibrate a single sensoragainst the block temperature.

To calibrate multiple sensors simply add an Isotech Selector Switch - there are two models; one for up toeight resistance thermometers and one for up to eight thermocouples. You can have a total of two boxes,any combination, to calibrate up to 16 sensors!

I-cal also supports models from our TTI range of indicators and with these you can easily build an automaticcalibration system to calibrate up to 16 sensors on the TTI's channel B, with a standard probe on channel A.The calibration results are saved to a text file that can be opened in your preferred application to create acertificate or report.

I-cal CAPTURE is suitable for the automaticcalibration of hand-held thermometers andsimilar devices. Previously hand-helds could notbe easily calibrated automatically because mostmodels do not have provision for a connectionto a PC. I-cal CAPTURE solves this by using alow cost camera. I-cal, as usual, monitors the dryblock bath / standard for stability, it then grabsan image before setting the bath to the nexttemperature. The image is captured with thedate and time to make a complete visual record(as shown).

Key Features

Use ISOTECH equipment to automaticallycalibrate temperature sensors.

Affordable and Expandable - A Selector Switchand or True Temperature Indicator can be addedto the Dry Block at any time.

Calibrate up to 16 sensors - Use twoswitchboxes to calibrate up to 16 sensors at up to10 temperatures.

Save Time - Set I-cal going, go home and returnto the calibration results!

Results saved to Text File - the calibrationresults are stored to a text file for you to process.

I-cal CAPTURE can save images - of devices likehandheld thermometers that can not beconnected to the PC.

A caption is added showing dates,time and the standard temperature.

Page 79: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

three/79

SoftwareI-Cal

How to Order I-calI-cal CAPTURE

Desktop or Notepad PC 166Mhz Pentium Processor with mmx or greater 32Mb Ram (I-Cal Capture 64Mb Ram) 800 x 600 / 16 bit display

I-Cal Windows 95/98Windows 2000/XPWindows 98MEWindows NT4.0

I-Cal Capture Windows 95/98Windows 2000/XPWindows 98ME

I-Cal Capture - Supported Camera Most USB web cameras,check with Isotech forrecommended models150Mb free Hard disk space

Serial Ports A maximum of three portsare required, one for the dryblock, one for the TTI andone for the Switchboxes(Two Switchboxes can beoperated from a single port)

Hint

Extra serial ports can easily be

added to a PC with USB by

purchasing low cost “USB to

Serial Convertors”; Including

those from Belkin.

I-cal the affordable automatic calibration systemWin 9x - Expandable

Calibrate RTD's and thermocouples

Minimum System Requirements

Page 80: The Industrial Laboratory - meter.hu 3 Iss 2.pdf · 983 Small Hot Plate Surface Sensor Calibrator 48-49 Model 862 24 Value Resistance Thermometer Simulator 50-51 Isocal Thermocouple

Isothermal Technology Limited, Pine Grove, Southport, Merseyside, PR9 9AG, EnglandTelephone: +44 (0) 1704 543830 Fax: +44 (0) 1704 544799 E-mail: [email protected] Web-site: www.isotech.co.uk

databookoneRealising ITS-90

Includes:

• Fixed Point Cells• Primary Apparatus• Thermometers• Instruments & Software

databooktwoSecondary Laboratory Equipment

Includes:

• Slim Cells• Secondary Apparatus• Thermometers• Instruments & Software

databookthreeIndustrial Laboratory Equipment

Includes:

• Metal Block Baths• Furnaces• Thermometers• Instruments & Software

databookfourBlackbody & Thermocouple Referencing

Includes:

• Blackbody Cells• Blackbody Cavities• Blackbody Furnaces• Thermocouple Referencing

databookfiveCalibration Services & Information

Includes:

• U.K.A.S. Calibration• Journal of Thermometry• Video Courses• Training Courses

Please send me the following databooks:

databookonedatabooktwodatabookthreedatabookfourdatabookfive

Name:

Address:

Postcode:

Tel:

Fax:

Email:

Fax Back FormFax: +44 (0) 1704 544799