The Development Of Nanosensors For In Vivo Detection Of ... · Nanosensors can be instrumental in...
Transcript of The Development Of Nanosensors For In Vivo Detection Of ... · Nanosensors can be instrumental in...
THE DEVELOPMENT OF NANOSENSORS FOR IN
VIVO DETECTION OF PHYSIOLOGICAL MOLECULES
Thesis Presented
By
Yi Luo
The Bouve’ Graduate School of Health Sciences
in Partial Fulfillment of the Requirements for the Degree of
Doctor of Philosophy in Pharmaceutical Sciences
May 2018
ii
TABLE OF CONTENTS
ABSTRACT ................................................................................................................................................. iii
ACKNOWLEGEMENT ............................................................................................................................... v
LIST OF TABLES ....................................................................................................................................... vi
LIST OF FIGURES .................................................................................................................................... vii
Chapter 1: Introduction and Dissertation Summary...................................................................................... 1
Chapter 2: Nanosensors for the Chemical Imaging of Acetylcholine Using Magnetic Resonance Imaging
.................................................................................................................................................................... 17
Chapter 3: Glucose-Sensitive Nanofiber Scaffolds with Improved Sensing Design for Physiological
Conditions ................................................................................................................................................... 61
Chapter 4: Conclusion and Future Direction .............................................................................................. 89
Reference .................................................................................................................................................... 92
iii
ABSTRACT
Nanosensors are an emerging tool for biomedical research and personalized medicine. By
incorporating a recognition moiety and a reporter on a nanoscale platform, the nanosensor has
displayed its ability to continuously monitor physiological molecules. To develop nanosensors for
in vivo applications, characteristics such as sensitivity, dynamic range, selectivity, reversibility,
biocompatibility, residency time, implantation, and clearance have to be considered. Choosing and
optimizing the recognition moiety is crucial for the fabrication of an in vivo nanosensor. Currently,
natural large molecules, such as proteins and nucleic acids, enzymes, and synthetic small
molecules have all been used as recognition moieties. In this thesis, we will demonstrate two
nanosensors using different recognition moieties to measure acetylcholine and glucose,
respectively. First, acetylcholine is a neurotransmitter associated with cognition, learning, and
memory. Previously, the detection of acetylcholine relied on microelectrode and microdialysis,
which suffer from invasiveness and limited spatial resolution respectively. To overcome these
issues, we developed a nanosensor to detect acetylcholine using magnetic resonance imaging
(MRI). Butyrylcholinesterase (BuChE) serving as the recognition moiety, and pH-sensitive
contrast agents serving as the reporter, are immobilized on the surface of a nanoparticle. Enzymatic
hydrolysis of acetylcholine created a pH-drop in the microenvironment close to the surface of the
nanoparticle, which was detected by the contrast agents leading to an increase in signal intensity.
Delivered to the brain of rats, this nanosensor detected drug-induced release of acetylcholine.
Second, we detected glucose as the first example of small molecule detection using an optode
format. The level of glucose is an important parameter to determine the dose of insulin therapy.
So far, the clinically used glucose monitors are all based on the enzymatic oxidation of glucose.
To extend the residency time of the glucose sensor, we developed a glucose-sensitive nanofiber
iv
based on a small molecule boronic acid as the recognition moiety. By incorporating boronic acids
containing an electro-withdrawing group, the nanofiber managed to continuously detect glucose
at physiological pH with an improved in vivo residency time.
v
ACKNOWLEGEMENT
This dissertation cannot be accomplished without the support and help from the following people
who I’d like to acknowledge. In Jan 2013, when I entered Dr. Clark’s lab, I was a student in the
master program and not sure about my future career. In the following years, my advisor Dr. Clark
guided me to the path to the degree of Ph.D. Not only she gave me the mentorship on the
development of nanosensor, but also trust and encouragement to help me overcome all the
difficulties and setbacks I encountered in the pursuit of science. I really appreciate all of the support
from her. I’m grateful to Dr. Mary Balaconis (Kate) and Eric Kim whom I learned a lot from when
we worked together on the projects presented in this dissertation. I want to thank all other past and
present members of our group including but not limited to Dr. Guoxin Rong, Wenjun Di, and
Jennifer Morales. I also want to thank Dr. Praveen Kulkarni and Dr. Chris Flask for their
instrumental help on MRI. Last but not least, I want to thank all my family members, including
my parents and my wife Dr. Yu Wang for their unconditional support and encouragement.
vi
LIST OF TABLES
Table 2.1. Relaxivity (r1) of contrast agents used in this study. ................................................... 40
Table 2.2. Relaxivity (r1) of ACh-MRNS in different pH used in this study. .............................. 40
vii
LIST OF FIGURES
Figure 1.1. Proposed general structure and sensing mechanism of nanosensor. ............................ 3
Figure 1.2. DNA-based nanosensor for optical detection of acetylcholine. ................................... 7
Figure 1.3. Sodium dependency of nanosensor. ........................................................................... 10
Figure 1.4. Boronic acid-based glucose nanosensor. .................................................................... 12
Figure 2.1. Schematic of the nanosensor structure and mechanism. ............................................ 32
Figure 2.2. Fabrication of the ACh-MRNS and pH-MRNS. ........................................................ 34
Figure 2.3. Structure and 1H NMR spectrum of pH-sensitive chelator. ....................................... 35
Figure 2.4 Characterization, in vitro calibration and selectivity of nanosensors. ......................... 37
Figure 2.5. TEM image of the ACh-MRNS using NanoVan stain. .............................................. 38
Figure 2.6. In vitro nanosensor to pH dependence. ...................................................................... 41
Figure 2.7. Characterization, in vitro calibration and selectivity of nanosensors. ........................ 43
Figure 2.8. pH change in the mixture of the nanosensor and acetylcholine. ................................ 45
Figure 2.9. Xylenol orange test. .................................................................................................... 46
Figure 2.10. Kinetics of BuChE. ................................................................................................... 47
Figure 2.11. Relative Signal intensity at different TR. ................................................................. 49
Figure 2.12. In vivo sensor contrast. ............................................................................................. 51
Figure 2.13. Histology. ................................................................................................................. 52
Figure 2.14. Acetylcholine detection in vivo. ............................................................................... 54
Figure 2.15. Diffusion of pH-MRNS in phantom brain................................................................ 57
Figure 3.1. Boronic acids incorporated into glucose-sensitive sensors. ....................................... 66
Figure 3.2. Response of glucose-sensitive macrosensors containing functionalized boronic acids
with increasing length of alkyl chains. ......................................................................................... 72
viii
Figure 3.3. Fluorescence decay of macrosensors with different boronic acids. ........................... 73
Figure 3.4. Response of macrosensors against different concentrations of glucose. ................... 75
Figure 3.5. Comparison of sensor response to two sugars, glucose and fructose. ....................... 76
Figure 3.6. Electrospun glucose-sensitive scaffolds. .................................................................... 79
Figure 3.7. Response of glucose-sensitive nanofibers containing different functionalized boronic
acids. ............................................................................................................................................. 80
Figure 3.8. In vivo comparison of glucose-sensitive nanoparticles and nanofiber scaffolds. ....... 81
Figure 3.9 Fluorescence measurements of glucose-sensitive nanoparticles and nanofiber
scaffolds over time in vivo. ........................................................................................................... 82
Figure 3.10. Response of glucose nanosensor embedded in alginate hydrogel. ........................... 86
Figure 3.11. The stability of hydrogel in PBS. ............................................................................. 88
1
Chapter 1: Introduction and Dissertation Summary
1.1 Introduction of in vivo application of nanosensors
Modern healthcare and biomedical research require imaging tools to measure critical physiological
analytes continuously in order to correlate them to biological activities for research, diagnosis and
personalized medicine. Nanosensors can be instrumental in revealing alterations of the
physiological analytes. After years of research, a wide variety of nanosensors have been developed
to detect health-related analytes in a well plate, cell culture, tissue and other biological systems.1-
3 Compared to molecular probes, such as fluorescent molecular indicators of ions and pH4-5, which
are also extensively used in biological systems for imaging and detection, nanosensors excel in the
capacity multiplexing and controllable modulation of physical and chemical properties.2
Functional moieties on nanosensor can be modified to detect other analytes. Size, surface potential,
and morphology of nanosensors can also be tuned to meet the metrics required for effective
biosensing. In vivo monitoring using nanosensors can measure physiological substances
continuously and detect complex activity and function in living organisms. Although some
nanosensors have been successfully applied in vitro, further modifications need to be made to
develop nanosensors for in vivo use. In vivo sensors are required to show desired sensitivity and
dynamic range to detect analytes, selectivity against potential interfering substances, fast response
time, and reversibility.6-7 Stability, residency time, biocompatibility, implantation, and clearance
also need to be considered in the development of a nanosensors for in vivo monitoring.8-9 For these
reasons, only a few types of nanosensors have been applied in live animals. In this dissertation, we
will focus on the development of nanosensors for in vivo detection of physiological molecules.
Nanosensors consist of a recognition moiety to selectively recognize analytes and a reporter to
transduce analyte-based biochemical changes into detectable electronic, optical and magnetic
2
signals. These elements are typically co-immobilized on a nano-platform, such as nanoparticles,
nanotubes, nanofibers, etc (Figure 1.1).1 Selection and modification of a recognition moiety in
this system are essential to determine if required sensitivity, dynamic range, and selectivity can be
achieved. The binding kinetics of the recognition element also impacts the reversibility and
response time of the sensor. Natural proteins, such as antibodies and receptors, can bind to some
physiological analytes selectively. If the binding can trigger a subsequent physical or chemical
change, these molecules can be potentially used as recognition moieties in nanosensors. An
enzymatic reaction is another suitable candidate mechanism to recognize analytes for its excellent
selectivity to substrates and high efficiency in generating detectable products. Also, organic
molecules have been synthesized to selectively bind to analytes. In this chapter, we will categorize
and summarize nanosensors for in vivo application by their mechanism of recognition: binding of
protein and nucleic acid, enzymatic reaction, and recognition by small synthetic molecules.
Recognition based on binding of large natural molecules
Natural large molecules, such as proteins and nucleic acids, selectively binds to analytes with their
unique 3D coformations. Antibodies have been widely used in in vitro assays, such as enzyme-
linked immunosorbent assays (ELISA), Western Blot, and Microchip assays. The other family of
large natural molecules, nucleic acids, can also bind to analytes via a process of evolution and
selection. In both cases, binding to the analytes needs to trigger a subsequent transduction to
detectable signal change. Although efficient and selective binding can be achieved using proteins
and nucleic acids without complex chemical synthesis or modification, the stability of the natural
molecules has restricted their application to sensing in living animals. Also, the high binding
affinity of antibodies make the recognition essentially irreversible and restrict itself as a
recognition moiety for continuous detection in vivo.10
3
Figure 1.1. Proposed general structure and sensing mechanism of nanosensor. The
nanosensor contains a recognition moiety and a reporter on a nanostructure. Binding to analytes
triggers a local chemical or physical change leading to a detectable signal.
4
Here we will briefly overview two types of nanosensors based on binding of proteins and nucleic
acids. As an example, to detect the cancer biomarker chorionic gonadotrophin (HCG), Kim et al.
covalently conjugated two monoclonal antibodies, anti-HCG-β95 and anti-HCG-β97, to iron
nanoparticles (CLIOs) separately to fabricate two types of sensors (CLIO-95 and CLIO-97).11
Once HCG was added to the blend of the two sensors, the antibodies bound to two different
epitopes crosslinking CLIOs, leading to a decrease in T2 signal in Magnetic Resonance Imaging
(MRI) study. This method can efficiently detect 0.5 – 5 µg/mL of HCG in solution. In a further
attempt, this group sealed the CLIO 95/97 blend in a medical device with a semi-permeable
membrane to prevent diffusion of nanoparticles in vivo and then implanted the device in a mouse
via a dorsal midline incision.12 When no tumor was induced, no T2 contrast was created by the
CLIOs in the device. When a tumor was induced in the mouse, the HCG released from tumor was
successfully sensed by the device resulting in a 15% drop in T2.
Single-stranded DNA and RNA show affinity to physiological analytes based on their 3-
dimensional conformation.13-14 A vast library of nucleic acids can be selected in a process termed
"systematic evolution of ligands by exponential enrichment," or SELEX to evolve aptamers with
high affinity to analytes.13 In this process, nucleic acids with random sequences are applied to
analytes, and only the successfully bound nucleic acid sequences are selected and reserved for the
next round of enrichment. After 8-15 cycles, the yielded nucleic acids can be used as a potential
recognition moiety in biosensors, resulting in a pico- to nanomolar affinity.
The massive diversity of nucleic acids and the process of SELEX lead to aptamers with a high
potential to detect targeted analytes. Although a broad range of in vitro applications have been
reported, the research on in vivo nanosensor containing aptamers is still limited. Yi et al. reported
the in vivo detection of adenosine triphosphate (ATP) by combining a two-photon (TP) dye, an
5
aptamer, and graphene oxide (GO).15 In this study, the TP dye was covalently conjugated to the
aptamer that selectively bound to ATP. Without the presence of ATP, the aptamer was adsorbed
on the surface of the GO via π-stacking and hydrogen bonds between nucleobases and the GO,16-
17 and the fluorescence from the tethered TP dye was quenched by the GO via Foster Resonance
Energy Transfer (FRET). When ATP was added, the aptamer was released from the GO when
bound to the analyte and “turned on” the fluorescence. The sensor demonstrated an increase in
fluorescence when 10 µM to 3 mM of ATP was added, and a lower limit of detection (LLOD) of
0.5 µM in cell culture. After the sensor was delivered to a zebrafish, a "turned on" fluorescence
indicated the presence of ATP in the animal at an imaging depth of 270 µm.
Recognition based on enzymatic reactions
Enzymatic reactions can also be used to sense physiological analytes. Compared to other
mechanisms, enzymatic reactions excel in high sensitivity and specificity towards analytes. Both
consumptions of substrates and creation of products can initiate chemical and physical changes
which can be further detected by reporters in nanosensors.
When the targeted analyte is the substrate of an enzymatic reaction, the enzyme can be
incorporated onto nanosensors as a recognition moiety. One commonly used example is the
monitoring of glucose by immobilizing glucose oxidase (GOx) or glucose dehydrogenase (GDH)
coupled with reporters. When glucose is present, the enzymes will oxidize glucose into glucono-
1,5-lactone. Glucose loses electrons in this redox reaction in the micro-environment close to the
surface of microelectrodes leading to a change in optical or electronic signal. This methodology
has been applied to fabricate research- and clinically-used electrodes18 and optical sensors19.
6
An acetylcholine nanosensor based on an enzymatic reaction has been successfully developed by
Walsh et al (Figure 1.2).20 In this study, a DNA-origami dendrimer was fabricated to form the
platform of this nanosensor. Both butyrylcholinesterase (BuChE) and the pH-sensitive dye
fluorescein were covalently conjugated to the DNA backbone. When acetylcholine was present,
the hydrolysis of the neurotransmitter catalyzed by the BuChE created a local pH drop protonating
fluorescein and triggering a decrease in fluorescence. This design enabled measuring micro- to
millimolar of acetylcholine in a well plate and successfully detected exogenously injected
acetylcholine in brain tissue. With a similar mechanism, a nanosensor for chemical imaging of
acetylcholine using MRI will be discussed in Chapter 2 in this thesis.
Cash et al. also reported a nanosensor using an enzymatic reaction to detect histamine in vivo.21 A
nanoparticle composed of a core of plasticized polyvinyl chloride (PVC) and a coat of amphiphilic
DSPE-PEG-lipids was fabricated as the nano-platform of the sensor. The oxygen sensitive
PtTPFPP was incorporated into the lipophilic core of the nanoparticle via sonication. The oxygen
sensitive nanoparticles were suspended in a buffered solution with free diamine oxidase (DAO).
When no histamine was present, the PtTPFPP was quenched by singlet oxygen in the solution.
When histamine was present, the enzymatic oxidation of histamine catalyzed by the DAO
consumed oxygen and unquenched the dye leading to an increase in phosphorescence. This design
can detect histamine in millimolar range with good reversibility. Combined with free DAO, an
increase in phosphorescence from subcutaneously administrated nanosensors was observed
starting at 5 mins post intraperitoneal injection of histamine. Other than incorporating enzyme as
a recognition moiety, an enzymatic reaction can be used to sense biomarkers in other ways.
Bhatia's group has been using substrate conjugated nanomaterial to detect the activity of protease
in vivo.22 Specifically, the substrate peptide of protease was conjugated to nanomaterial as a reco-
7
Figure 1.2. DNA-based nanosensor for optical detection of acetylcholine. Both the recognition
moiety butyrylcholinesterase and fluorescent reporter were conjugated to a DNA dendrimer. When
acetylcholine was hydrolyzed by the enzyme, the resulting local pH drop was detected by the pH-
sensitive fluorescein leading a decrease in fluorescence. Reprinted from permission from Ref 15
(Sci. Rep. 2015, 5, 14832. Copyright 2015, Macmillan Publishers Limited).
8
-gnition moiety. When the activity of protease was high, more peptide was cleaved by the
enzymatic hydrolysis. The hydrolyzed peptide coupled with a reporter was excreted in urine and
then analyzed to determine the relative activity of the protease. Although the final analysis was
carried out ex vivo, the enzymatic recognition happened in vivo. Using the same mechanism,
Kwong et al. reported a mass-encoded nanoworm sensor to detect the activity of protease
expressed from disease sites.23 The concentration of peptides cleaved by the protease was
determined by the analysis of urine using mass spectroscopy. They also showed that the
nanosensor could noninvasively monitor liver fibrosis and cancer in mice by assessing the level of
protease. A mathematical framework was also established to assess the activity of cancer-related
biomarkers using this mechanism.24
The same group further reported a nanoparticle-based sensor with a photolabile group blocking
the site of enzymatic cleavage.25 Only at the site of disease, a UV exposure photolyzed the blocking
molecule to unveil the peptide for the protease catalyzed cleavage. Again, more active protease
led to a higher amount of cleaved peptide in the urine of the animals. This method allowed specific
detection of the protease at the site of disease. Coupled with an immunoassay and microfluidics,
the subcutaneous delivery of the nanosensors, so-called “synthetic biomarkers”, provides a facile
point of care monitoring of thrombosis26-27 and noncommunicable diseases28, respectively.
Recognition based on binding of synthetic molecules
Synthetic organic or organometallics molecules can be screened and modified to recognize
analytes. Compared to natural proteins, it is relatively easier to synthesize and screen small
molecules to recognize the analyte. The binding affinity of the selected molecules can be further
optimized by chemical modifications. Up to now, a variety of synthetic molecules with desired
binding affinity and selectivity have already been reported or are even commercially available. Jin
9
et al. incorporated a sensing molecule, 4-amino-5-methylamino-2′,7′-difluorofluorescein (DAF-
FM) to a poly-(lactic-co-glycolic acid) (PLGA) nanoparticle to track nitric oxide in joint fluid in
vivo.29 In the presence of oxygen, DAF-FM reacted to nitric oxide inducing an increase in
fluorescent intensity. In another report, Zheng et al. immobilized an oxygen-sensitive iridium
complex to a poly(ε-caprolactone)-b-poly(N-vinylpyrrolidone) (PCL-PVP) nanoparticle to detect
hypoxia conditions in tumor tissue.30
Clark’s group has been fabricating optode-based nanosensors to detect critical physiological ions
(Na+, K+, Ca2+, etc.) in vivo using commercially available ionophores. The mechanism of this type
of nanosensor is based on a local protonation/deprotonation. The nanosensor is composed of a
hydrophobic core and an amphiphilic coating. Coated with negatively charged amphiphilic DSPE-
PEG-lipids, the nanosensor can maintain its stability in a buffered solution for days after
fabrication. All sensing components including pH-sensitive fluorescent chromoionophores, ion-
binding ionophores, plasticizer and additives are incorporated in the hydrophobic phase. When the
targeted ion is present, it will be selectively bound by the ionophore and then extracted into the
hydrophobic phase. The chromoionophore will then be protonated/deprotonated to balance the
charge to neutral leading to an alteration of fluorescent intensity. Using this mechanism, Dubach
et al. developed a nanosensor to detect sodium ion in its physiological range.31 In this study, a
higher concentration of sodium led to a lower fluorescent intensity. When this nanosensor was co-
injected with 0 to 500 mM sodium chloride solution to a mouse, a sodium dependency of
fluorescent intensity was observed (Figure 1.3).32 A similar mechanism was used by Cash et al.
to monitor therapeutic lithium in vivo.33 To achieve a high depth of imaging, a photoacoustic
chromoionophore was deprotonated triggered by the extraction of lithium ion. Injected into the
skin of a mouse, the nanosensor successfully detected intraperitoneal administrated lithium ions.
10
Figure 1.3. Sodium dependency of nanosensor. (A) Nanosensors responded to sodium in
aqueous buffer. (B) Nanosesnor responded to co-injected sodium chloride solution. Fluorescence
overlaid with brightfield is shown with seven different subcutaneous injections of sensors in
different sodium concentration. On the right is the corresponding sodium concentration inmM.
Reprinted from permission from Ref 27 (Integr. Biol. 2011, 3, 142-148. Copyright 2011, The
Royal Society of Chemistry.)
11
In addition, reported by Cash et al., an ionophore for histamine was used to track in vivo
concentration of histamine.34
Another optode-based nanosensor reported by Clark's group used small-molecule components to
detect glucose in vivo. Compared to the traditionally used enzyme-based sensing mechanism, this
design avoids local depletion of resources, such as oxygen when glucose oxidase is used, and the
dysfunction of the sensor caused by the degradation of biological components. Billingsley et al.
developed an optode sensor using a combination of alizarin dye and boronic acid to monitor
glucose concentration in vivo (Figure 1.4).35 Without glucose, the complex of alizarin and boronic
acid was fluorescent. When glucose is present, it competitively bound to boronic acid and left
alizarin free. The free alizarin loses its fluorescence. Thus, a high concentration of glucose was
indicated by a weak fluorescence. Implanted in the skin of mice, the optode sensor can detect
glucose administrated by oral gavage.36 To improve the implantation and in vivo residency time of
the optode sensor, glucose-sensitive nanoparticles were fabricated and then embedded in
commercially available hydrogels.37 A glucose-sensitive nanofiber to improve sensitivity and
residency time will be discussed in chapter 3 in this thesis. In this thesis, two recognition
mechanisms are applied to detect physiological molecules: cholinesterase catalyzed hydrolysis is
used in the design of nanosensor for acetylcholine, while boronic acid mediated small molecule
binding is used to sense glucose. The two projects will be discussed in detail in Chapter 2 and 3,
respectively.
12
Figure 1.4. Boronic acid-based glucose nanosensor. (A) Components and mechanism of boronic
acid-based glucose nanosensor. Reprinted from permission from Ref 32 (J. Diabetes. Sci. Technol.
2011, 5, 68-75. Copyright 2011, Diabetes Technology Society.) (B) Image of mouse injected at
four locations with glucose-sensitive nanosensors. Image was obtained with an IVIS-Spectrum
imaging system. Excitation and emission wavelengths were 500 and 600 nm, respectively.
Intensity bar displays the normalized fluorescence efficiency, which represents the fractional ratio
of fluorescent emitted photons per incident excitation photon. Residual background fluorescence
was attributed to remaining fur. (C) The representative response to oral gavage of the blood
glucose (red) and fluorescence of the glucose nanosensors (black). Mean ±SD for one mouse is
shown. Reprinted from permission from Ref 31 (Anal. Chem. 2010, 82, 3707-3713. Copyright
2010, American Chemical Society.)
13
1.2 Acetylcholine and its detection
Acetylcholine is a neurotransmitter expressed to transmit signals between neurons38-39 and in
neuromuscular junctions40. This neurotransmitter is released by cranial nerves, motor neurons, and
preganglionic neurons that exit the central nervous system (CNS). In the CNS, acetylcholine is
also found in the basal forebrain complex that projects to the hippocampus and neocortex, and the
pontomesencephalic cholinergic complex that projects to the dorsal thalamus and forebrain.41
Synthesis of acetylcholine from choline is catalyzed by choline acetyltransferase (ChAT) in
cholinergic neurons. Then the neurotransmitter is stored in synaptic vesicles, released from the
pre-synaptic membrane to bind to nicotinic or muscarinic receptors at either post-synaptic
membrane or axons.41 The nicotinic receptor is a ligand-gate ion channel. Once bound to
acetylcholine, the ion channel will open and allows sodium, potassium, and calcium to pass
through.42 On the other hand, five subtypes of muscarinic (M1-M5) are G-protein-coupled receptor
(GPCR).43 Binding of acetylcholine triggers a cascade of signal transduction involving Inositol
trisphosphate (IP3) or cyclic adenosine monophosphate (cAMP) as second messengers. After
signal transmission, acetylcholine will be hydrolyzed by cholinesterase into acetic acid and choline
which will be recycled by pre-synaptic neurons.44
In the CNS, acetylcholine is closely related to learning and memory.45-46 Behavioral pharmacology
reveals that scopolamine, a muscarinic antagonist, blocked sensory or attentional processes and
affected short-time memories in brains of rodents, monkeys, and humans.47 Also, a neurotoxin 192
IgG saporin created cholinergic lesions by decreasing activity of ChAT and impaired learning and
memory performance of rats.48 Deficiency of acetylcholine in brain is also associated with aging
and dementia especially Alzheimer’s disease (AD).49 Symptomized by memory loss, confusion
with time or place, decreased or poor judgement, etc., this disease is now the 6th leading cause of
14
death in the United States. According to the facts from Alzheimer’s Association, 5.7 million (2017)
Americans are now affected by this disease, and this number is expected to keep rising in next
decades.50
Research and clinical reports have unveiled that the decline of cholinergic function in basal
forebrain plays an important role in causing AD.49 Also, the formation of pathological beta-
amyloid (Aβ) is increased by stimulation of muscarinic acetylcholine receptors.51 There are reports
that Aβ interacts with nicotinic acetylcholine receptors.52 Thus, investigation of the relationship
between the concentration of acetylcholine and AD will render a crucial tool for research and
diagnosis of this disease.
Currently, functional positron emission tomography (PET), computed tomography (CT),
electroencephalogram (EEG) and magnetic resonance imaging (MRI) have been applied to explore
the alteration of brain function in animal models with the AD.53 More specific detection of
acetylcholine still relies on microelectrode54-55 and microdialysis.56-57 Though these methods
provide tools for direct measuring of acetylcholine, the invasive nature and limited spatial
resolution restrict their applications.
Chapter 2 will report our innovation to develop a nanosensor for in vivo detection of acetylcholine
using MRI. This nanosensor uses BuChE as the recognition moiety and a pH-sensitive MRI
contrast agent as the reporter. When acetylcholine is hydrolyzed by the enzyme, the local decrease
in pH will be sensed by the reporter and transduced to an increase in signal intensity in MR image.
This design enables selective imaging of acetylcholine in deep brain for the first time. Compared
to previously reported fluorescent nanoprobe for acetylcholine and CEST-based MR spectroscopic
imaging, the nanosensor we developed is advantageous for its high depth of imaging and
15
selectivity. More details of both in vitro and in vivo application of this nanosensor will be discussed
in detail in Chapter 2.
1.3 Glucose and its detection
Diabetes is a disease associated with the disorder of glucose metabolism. Currently, more than 29
million Americans (2016)58 and 388 million Chinese (2013)59 adults are diabetic, and much more
population have prediabetes. This epidemic has sparked interest in detecting and monitoring the
concentration of glucose which is an important reference to determine the dose of insulin
therapy.60-61 Although a lot of efforts have been made to detect glucose in tears62, sweat63 and other
body fluids64, the gold standard to measure glucose is still the finger-prick test. A series of point-
of-care glucose meters have been commercially available for patients to self-monitor blood
glucose.65 All of these devices still require blood samples for discrete measurement. Continuously
monitoring glucose can track the change of the blood glucose levels for hours and days and provide
more information on diet control, physical activity, and dose of medicine. Now there are four types
of continuous glucose monitors (CGM) available on market to invasively measure glucose in
interstitial fluid.66 All of these meters and monitors coat enzymes such as GOx or GDH on an
electrode to gain the desired sensitivity within the physiological range of glucose in the blood.
However, the degradation of biological elements in these monitors hinders the long-term
application of these devices. To minimize invasiveness, extend sensor residency time, and enhance
accuracy, the development of novel glucose sensors is still an area of interest.
The Clark lab has been using a non-biological small molecule boronic acid as a recognition moiety
to detect glucose. Coupled with a fluorescent dye alizarin as the reporter, a small molecule-based
nanosensor can be used to continuously measure glucose in vivo with minimal invasiveness.
Compared to enzyme-based glucose sensors, the application of small molecule boronic acids
16
avoids the potential degradation of biological elements in vivo. A nano-optode based sensor was
firstly reported in 2010 showing desired dynamic range, sensitivity, and selectivity.35 In vivo
experiment demonstrated this sensor can detect glucose administrated by oral gavage in mice. The
result was consistent with the gold-standard blood test. Later, the nanoparticle-based sensor was
embedded into commercially available hydrogels to extend its residency time at injection sites.37
In chapter 3 of this thesis, we will illustrate a new glucose sensitive nanofiber. The nanofiber was
fabricated to prolong the in vivo liftetime of the boronic acid-based nanosensors for the purpose of
coniuously monitoring glucose. We synthesized and screened boronic acids with lower pKa to
improve its sensitivity at physiological pH. Then we incorporated the selected boronic acids to a
nanofiber using electrospinning. The final sensor had an increase residency time in vivo.
17
Chapter 2: Nanosensors for the Chemical Imaging of Acetylcholine
Using Magnetic Resonance Imaging
2.1 Introduction
Imaging tools that enable real-time visualization of molecular neural events, namely
neurotransmitter release, are highly valuable for understanding the basis of brain function and
disease. Cellular-level methods such as electrophysiology67 and optical imaging68 offer recordings
of neural activity with precision and high specificity, however, are often limited to sampling a
relatively small area of the mammalian brain. Conversely, among modern imaging techniques,
magnetic resonance imaging (MRI)69 is a powerful tool that provides advantages for in vivo
analysis as it can be applied noninvasively, with unlimited tissue penetration and mapping
capabilities of the whole-brain. Despite these advantages, resolving target-specific detection of
neurotransmitters with molecular specificity for functional neuroimaging is yet to be fully
established, and only a limited number of successful studies have been implemented for in vivo
monitoring in the brain. This is in part due to a low concentration of neurotransmitters in the brain,
as well as the low intrinsic sensitivity and resolving power of the MRI. Consequently, synthesizing
a highly sensitive, stable and non-toxic probe is always a great challenge for effective application
of molecular neuroimaging in the brain. Currently, a number of MRI molecular contrast agents
have been explored for imaging of neurotransmitters in the brain. For example, detection of
dopamine70-71 and serotonin72 have been developed from engineered forms of flavocytochrome
P450-BM3 with a detectable T1 signal. A contrast agent to detect glutamate73 has also been
developed based on the displacement of mGluR5 receptor as well as agents based on a crown ether
cation-binding motif chelating a gadolinium (Gd) to target glutamate, GABA, and glycine albeit
18
with millimolar affinity.74 An additional approach, using the chemical exchange saturation transfer
(CEST) effect, has been demonstrated to detect glutamate in human subjects.75-76 While there is a
growing repertoire of neurotransmitter-sensitive MRI probes, there remains considerable
opportunities for the development of imaging agents that can respond to neural activity with high
chemical specificity and sensitivity in order to overcome the inherently low signal-to-noise ratio
of MRI.
In particular, the development of nanostructured sensors as MRI contrast agents is a promising
direction for the detection and activity-dependent in vivo monitoring of neurotransmitters in the
brain. Unlike traditional molecular organometallic chelates, the nanostructured probes offer
advantages of flexibility and modularity to modify their physicochemical properties and
functionalities.2 To date, MRI contrast agents have been packaged in nanoparticles in order to
amplify magnetic relaxivity (r1),77 improve penetration and retention in tumor,78 monitoring of
enzymatic activity79 and for theranostic applications.80 However, the development of nano-scale
MRI contrast agents for the chemical imaging of neurotransmitters has yet been realized.
In the present study, we developed a nanoparticle sensing platform for the imaging
neurotransmitter acetylcholine in the living brain tissue using MRI. As an important molecular
messenger, acetylcholine is involved in regulating chemical communication between cells in the
brain. In particular, the cholinergic system is one of the most important modulatory
neurotransmitter systems in the brain, in which both synaptic38 and volume39 transmission govern
activities that depend on selective attention,81 formation of working memories82 and cognitive
behavior.83 Additionally, perturbations of the cholinergic system are implicated in schizophrenia,84
depression85 and Alzheimer’s disease.86 Previously, direct measurement of choline using 1H MR
spectroscopy and its application in malignant breast tumor was reported.87-88 Our method differs
19
by detecting acetylcholine directly, since it is unlikely that choline could be used a surrogate in
the brain.89
The design of the nanosensor involves co-immobilizing the enzyme butyrylcholinesterase (BuChE)
and pH-sensitive gadolinium contrast agents on a nanoparticle to create a pH drop triggered by the
enzymatic hydrolysis of acetylcholine which detected by the proximate contrast agents within the
nanoparticle microenvironment leading to an increase in T1 relaxation rate (1/T1). The nanosensor
platform presented here prevents the sensing components from diffusing away in vivo and provides
a required proximity between the contrast agent and BuChE. As such, this design is well-suited
for real-time imaging of acetylcholine in the living brain, in addition to filling a need in the field
of MRI by providing chemical specificity to neuroimaging.
Here, we show: (1) synthesis and characterization of the nanosensor for detection of acetylcholine
(ACh-MRNS); (2) response of the nanosensor to increase in acetylcholine levels with suitable
sensitivity and selectivity in vitro; and (3) in vivo detection of endogenous release of acetylcholine
in the rat medial prefrontal cortex (mPFC), which is known to receive dense cholinergic inputs
from the basal forebrain and the hippocampal formation90 stimulated by systemic administration
of clozapine.
2.2 Materials and Methods
Materials
2-hydroxyl-5-nitrobenzyl bromide, acetic anhydride, bis-(2-ethylhexyl)sebacate (DOS),
butyrylcholinesterase from equine serum (EC 3.1.1.8) (BuChE), clozapine, 5,5′-Dithiobis(2-
nitrobenzoic acid) (DTNB), dopamine hydrochloride, fluorescein sodium salt, gadolinium(III)
nitrate hexahydrate, γ-aminobutyric acid (GABA), glutamic acid, glycine, methanol, N-(3-
dimethylaminopropyl)-N’-ethylcarbodiimide hydrochloride (EDC), N,N’-dimethylaminopyridine
20
(DMAP), N,N’-dimethylformamide (DMF), N-hydroxysuccinimide (NHS), N,N,N’,N’-
tetramethyl-O-(N-succinimidyl)uroniumtetrafluoroborate (TSTU), potassium carbonate (K2CO3),
triethylamine (TEA) trifluoro acetic acid (TFA) and xylenol orange tetrasodium salt were
purchased from Sigma Aldrich (St Louis, MO, USA). 15-azido-4,7,10,13-tetraoxa pentadecanoic
acid was purchased from Alfa Aesar. DO3A tert-butyl ester (t-BOC DO3A) was purchased from
Macrocyclics (Plano, TX, USA). Phosphate Buffered Saline (PBS) (1×, pH 7.4) and sterilized
0.9% saline solution were purchased from Invitrogen (Carlsbad, CA, USA). Hydrochloric acid
(1.0 N) and sodium bicarbonate were purchased from Fisher Scientific (Fair Lawn, NJ, USA). 1,2-
distearoyl-sn-glycero-3-phosphoethanolamine-N-[amino(polyethylene glycol)-2000] (ammonium
salt) (DSPE-PEG-amine), 1,2-distearoyl-sn-glycero-3-phosphoethanolamine-N-
[dibenzocyclooctyl (polyethylene glycol)-2000] (ammonium salt) (DSPE-PEG-DBCO), 1,2-
distearoyl-sn-glycero-3-phosphoethanolamine-N-[methoxy(polyethylene glycol)-550]
(ammonium salt) (DSPE-PEG-methoxy) and 1,2-distearoyl-sn-glycero-3-phosphoethanolamine-
N-[poly(ethylene glycol)2000-N'-carboxyfluorescein] (ammonium salt) (DSPE-PEG-fluorescein)
were purchased from Avanti Polar Lipids, Inc. (Alabaster, AL, USA), and octadecyl Rhodamine
chloride (R18) was purchased from Thermo Fisher Scientific (Waltham, MA, USA).
Synthesis of pH sensitive contrast agents
t-BOC DO3A (200 mg, 0.4 mmol) and 2-hydroxyl-5-nitrolbenzyl bromide (360 mg, 2.0 mmol)
were dissolved and stirred in a mixture of 2 mL DCM and 2 mL DMF for 1 h at room temperature.
After addition of 600 mg K2CO3, the resulting suspension was stirred overnight. Supernatant was
then collected after centrifugation, reduced by vacuum, and 3 mL TFA was added dropwise at 0
oC, then stirred overnight after it was allowed to warm to room temperature. TFA was removed by
21
vacuum and the residue was purified by flash chromatography to yield product (34% after two
steps). 1H NMR spectrum of product was compared91 to verify success of synthesis.
EDC (7.6 mg, 0.04 mmol) and NHS (4.6 mg, 0.04 mmol) in 200 µL 0.1× PBS, pH 6 was added to
pH-sensitive chelates (20 mg, 0.04 mmol) in 3 mL 0.1× PBS (pH 6) in 10 aliquots. The solution
was stirred for 30 min and then added to a solution of DSPE-PEG-amine (6 mg, 0.002 mmol) in 3
mL 0.1× PBS (pH 7.4). After the pH was adjusted to 7.4, the reaction mixture was stirred overnight
at room temperature. Gd(NO3)3·6H2O (36 mg, 0.08 mmol) in 200 µL DI water was added to the
resulting solution in 5 aliquots, and the pH of the reaction mixture was maintained between 4 and
6 during the addition. The resulting solution was stirred overnight at 40 oC, then diluted to 24 mL
with DI water and stored at 4 oC.
Fabrication of pH-sensitive nanoparticles
DSPE-PEG-DBCO (0.2 mg, 0.067 µmol) in 20 µL chloroform was dried in a glass vial before
addition of 4 mL stock solution of DSPE-PEG attached pH sensitive contrast agent (0.33 µmol).
The mixture was sonicated at 20% amplitude for 30 s to re-dissolve dried DSPE-PEG-DBCO using
a Branson digital sonifier (Danbury, CT). Following addition of a mixture of 3 mg PVC and 6 mg
DOS in 50 µL THF and 80 µL DCM, the solution was sonicated at 20% amplitude for 3 min. The
resulting nanoparticle suspension was filtered using Acrodisc syringe filter with 0.45 µm
membrane (Pall Cooperation, Ann Arbor, MI, USA), washed by DI water (1 mL × 5) in 100 kD
molecular weight cut off (MWCO) Amicon Ultra centrifugal filters (EMD Millipore, Billerica,
MA, USA), and concentrated to 0.1 mL in DI water by the same filters.
The concentrated suspension was treated with 1.5 µL TEA, vortexed for 3 h, then treated with 1
µL acetic anhydride, and vortexed overnight. The resulting nanoparticle suspension was washed
22
(1 mL × 5) in 100 kD MWCO spin-filters as mentioned above and then diluted to 0.5 mL with
PBS, pH 8.
Conjugation of BuChE to nanoparticles
Solution of 15-azido-4,7,10,13-tetraoxa-pentadecanoic acid (30 nmol in 3 µL DMF), TSTU (1.8
mg, 6.0 µmol) and DMAP (1.4 mg, 11 µmol) were all dissolved in 8 µL DMF. The solution was
vortexed for 1 h, and added to solution of BuChE (0.3 nmol in 292 µL PBS, pH 8). The mixture
was vortexed for another hour, washed (1 mL × 5) and diluted to 0.5 mL with PBS, pH 8 in 100
kD MWCO spin-filters as mentioned above. The modified enzyme solution and 0.5 mL
nanoparticle suspension were combined and incubated for 72 h at 4 oC. The suspension was
concentrated to 30 µL with 100 kD MWCO spin-filters as mentioned above before in vitro
calibration.
Particle sizing, zeta-potential and concentration of nanoparticles measurements
The conjugated nanoparticles above were characterized for measurement of particle size and zeta-
potential by dynamic light scattering (DLS) using a 90 Plus particle size analyzer (Brookhaven
Instruments Corporation). The concentration of nanoparticles was measured using nCS1TM
nanoparticle analyzer (Spectradyne, Torrance, CA).
Inductively coupled plasma mass spectrometry (ICP-MS)
To identify the relaxivity (r1) of nanoparticles, we used a Bruker Aurora M90 inductively coupled
plasma-mass spectrometer (Bruker Scientific Instruments, Billerica, MA, USA) to determine the
amount of Gd(III) bound to the surface of nanoparticles. Standard solutions were prepared by
dissolving Gd(NO3)3 in DI water with different concentrations (0, 0.5, 5, 20, 50 µg/L). The
concentrated nanoparticle suspension from previous step was diluted by DI water in ratios of 0.2,
23
2, 8, and 20 µL/L, and compared with the standard solution to identify the exact amount of bound
Gd(III).
Electron microscopy
Four microliters of sample was pipetted onto a C-Flat (Protochips) holey carbon film then plunge
frozen using a Gatan Cp3 CryoPlunge unit. Prepared grids were stored under liquid nitrogen until
loaded for imaging in an FEI Arctica CryoFEG-TEM with autoloader. Images were collected
using low-dose techniques at 200 kV. Images were analyzed using ImageJ software92 to measure
diameter of nanoparticles.
To stain lipid coating of the nanoparticles, diluted nanosensors (5 µL) were placed on a 300 mesh
carbon film coated copper grid (Electron Microscopy Sciences) for 1 min. The excessive liquid
was removed by a piece of filter paper. The remaining sample on carbon film was stained using 5
µL of methylamine vanadate (Nanovan) for 1 min, and then the excessive liquid was blotted by a
filter paper. After two rounds of staining, the images were acquired at 200 kV accelerating voltage
using FEI Arctica CryoFEG-TEM.
Relaxivity test in low magnetic field
ACh-MRNS corresponding to 0.43, 0.21, 0.12. 0.11 and 0.043 mM Gd(III) was suspended in 500
µL 1× PBS in NMR tubes and then analyzed by a 1.5 T Bruker Minispec mq60 NMR analyzer (60
MHz, Bruker Inc., Billerica, MA) at 37 oC to yield T1 of each sample. 1/T1 was plotted as a function
of concentration of Gd(III) and the slope of the plotted curve is r1.
pH calibration
We added 2 µL of a nanosensor suspension to 198 µL PBS, pH 6, 6.5, 7, 7.4 and 8 in well plates.
The well plate was scanned in a 7 T Bruker Biospec MRI scanner for small animals (Bruker
24
Scientific Instrument, Billerica, MA, USA). A T1-weighted Rapid Acquisition with Relaxation
Enhancement with Variable TR (RARE-VTR) sequence (1 slice; 1.0 mm; TE = 12.5 ms, TR = 70,
291, 576, 976, 1651, and 5000 ms, FOV = 40 mm × 40 mm; data matrix 64 × 64) was used to
generate a T1 map in about 9.5 min. The signal intensities at different TRs were also collected
using Matlab code.
In vitro calibration
Sensor calibration was performed in a Bruker coil with an inner diameter of 7.5 cm. A Tripilot
scan was initially conducted followed by addition of 2 µL of a nanosensor suspension to 198 µL
solution of acetylcholine (0, 50, 100, 250, 500 and 1000 µM) in 1× PBS, pH 7.4 in well plates.
The well plate was scanned in a 7 T Bruker Biospec MRI scanner for small animals (Bruker
Scientific Instrument, Billerica, MA, USA). A T1-weighted Rapid Acquisition with Relaxation
Enhancement with Variable TR (RARE-VTR) sequence (1 slice; 1.0 mm; TE = 12.5 ms, TR = 70,
291, 576, 976, 1651, and 5000 ms, FOV = 40 mm × 40 mm; data matrix 64 × 64) was used to
generate a T1 map in about 9.5 min. The signal intensities at different TRs were also collected
using Matlab code. To verify that the pH change is a local effect, we also suspended 2 µL pH
sensitive nanoparticles without conjugation of enzyme (pH-MRNS) and 8.7 units free BuChE in
198 µL solution of acetylcholine (0, 10, 50, 100, and 500 µM) in PBS, pH 7.4 in well plates. The
same sequence was used for the T1 map.
In vitro calibration at pH 7.2, 7.4 and 7.8
ACh-MRNS corresponding to 0.04 mM Gd(III) was suspended in 500 µL 1× PBS, pH 7.2, 7.4 and
7.8 with a concentration of acetylcholine varying from 0 to 100 µM in NMR tubes and then
analyzed by a 1.5 T Bruker Minispec mq60 NMR analyzer (60 MHz, Bruker Inc., Billerica, MA)
at 37 oC to yield T1 of each sample.
25
Measurements of overall pH change in the mixture of nanosensors and acetylcholine
Fluorescein sodium (5 µg/mL) was mixed with ACh-MRNS (0.2 mM Gd(III)) and pH-MRNS (0.2
mM Gd(III) with 8.7 units of BuChE in 100 µL 1× PBS, pH 7.4, respectively. Then 100 µL of
acetylcholine solution with concentrations varying from 0 to 1 mM in 1× PBS, pH 7.4 was added
to the mixture. The mixture was excited at 460 nm and read the emission at 520 nm using a
SpectraMax Gemini EM plate reader.
Xylenol orange test
The test was performed following the procedure reported in the previous literature.93 Specifically,
the ACh-MRNS corresponding to 0.1 mM conjugated Gd(III) was incubated with concentrations
of acetylcholine varying from 0 to 5 mM in 100 µL HEPES buffer, pH 7.4 for 10 min and then
was added with 100 µL of 0.6% (mg/mL) xylenol orange in 50 mM acetic acid buffer, pH 5.4. The
absorbance at 573 nm and 433 nm was read using a SpectraMax Gemini EM plate reader. No
significance change in A573/A433 was observed. We also plotted a work curve of the test: 100 µL
of 0.6% (w/v) xylenol orange in 50 mM acetic acid buffer, pH 5.4 was added to 100 µL of
Gd3(NO3)3 solution in HEPES buffer, pH 7.4 to make the final concentrations of Gd3+ varying
from 0 to 100 µM. The absorbance at 573 nm and 433 nm was read using a SpectraMax Gemini
EM plate reader.
Enzymatic kinetics studies
DSPE-PEG-methoxy was used to replace pH-sensitive contrast agents conjugated DSPE-PEG-
amine to coat nanoparticles to avoid interference from pH-sensitive contrast agent’s absorbance at
400 nm in this study. Ellman’s assay was used to test the catalytic kinetics of conjugated and free
BuChE. 8.7 units of BuChE in each form was suspended in 1 mM DTNB in 100 µL PBS, pH 7.4
26
in a 96 well plate. Acetylthiocholine in 100 µL PBS, pH 7.4, was added to the mixture to make
final concentration of acetylthiocholine 0, 5, 10, 50 and 100 µM. The absorbance at 412 nm was
recorded right after the addition every 5 s for 2 min using a SpectraMax Gemini EM plate reader.
To study the overall change of pH led by the enzymatic hydrolysis, we suspended 2 µL
nanoparticles coated by DSPE-PEG-fluorescein and 8.7 units of azide modified BuChE in 100 µL
1× PBS, pH 7.4. Acetylcholine in 100 µL 1× M PBS, pH 7.4, was added to the mixture to make
final concentration of acetylthiocholine 0, 5, 10, 50, 100, 500 and 5000 µM. The fluorescent signal
(excitation: 490 nm, emission: 520 nm, cut-off: 515 nm) was recorded right after the addition every
10 s for 10 min using a SpectraMax Gemini EM plate reader.
Selectivity studies
2 µL of concentrated nanosensors were suspended in 198 µL of either glutamate (5 mM), dopamine
(5 mM), GABA (5 mM), or glycine (5 mM) in PBS, pH 7.4, and scanned with the same coil and
sequence used for in vitro calibration. The resulting 1/T1 was compared with the 1/T1 of
nanosensors in PBS, pH 7.4 and acetylcholine solution (0.1 mM in 1× PBS, pH 7.4).
Animal care and stereotaxic surgery
Adult male Sprague-Dawley rats (230-300 g) were obtained from Charles River Laboratories
(Wilmington, MA). The rats were maintained on a 12:12 h light:dark cycle and allowed access to
food and water ad libitum. All procedures were approved by the Northeastern University
Institutional Animal Care and Use Committee and were in accordance with the National Institutes
of Health guidelines.
Three days prior to MRI experiments, unilateral implantation of 26-gauge plastic guide cannula
(Plastics One) aimed at the medial prefrontal cortex (mPFC) was performed on animals using a
27
stereotaxic device (Kopf Instruments) under isoflurane anesthesia. A small incision was made to
expose the dorsal surface of the skull and wiped clean to reveal the position of lambda and bregma
landmarks. A small hole was drilled into the skull at the coordinate position necessary to gain
access to the prefrontal cortex (bregma: +2.8 mm anterior, +0.8 mm lateral, +4.0 mm below the
surface of the skull). The cannula was placed in the brain and anchored using plastic screws and
dental acrylic. The head-wound was then sutured closed and topical antibiotic ointment was
applied to the wound area. Buprenorphine (0.5 mg/Kg) was administrated to reduce pain.
In vivo nanosensor injection and MRI
Animals were first anesthetized with 1-2% isoflurane and placed in a plastic positioning device
and a head holder built-in with quadrature transmit/receive volume coil. Infusion of nanosensors
into the mPFC was performed by lowering and placing the internal cannula attached to a 10 µL
Hamilton syringe via polyethylene tubing filled with nanosensors through the guide cannula,
delivering a final volume of 2 µL of nanosensors. The air was first removed prior to nanosensor
delivery. The internal-injector cannula protruded 1 mm beyond the guide cannula toward the
mPFC.
After delivery, MRI experiments were conducted using a 7 T Bruker Biospec 300 MHz MRI
scanner for small animals (same as above for in vitro studies). The design of the positioning device
and head holder coil provided complete coverage of the brain from olfactory bulbs to brain stem
with excellent B1 field homogeneity. At the beginning of each imaging session, a high-resolution
anatomical data set was collected using the RARE-VTR sequence (25 slices; 1.0 mm; TE = 12.5
ms, TR = 513, 800, 1400, 2200, and 6000 ms, FOV = 40 mm x 40 mm, data matrix 128 x 128) to
assess time-lapse nanosensor response, followed by acquisition of same sequence at multiple time-
points (0, 23, 46 min post-nanosensor injection). Each scan took about 23 minutes. For detection
28
of drug-evoked cholinergic transients, subcutaneous injection of clozapine (20 mg/Kg) dissolved
in PBS, pH 6.5 into the back of rat was administered at the time of nanosensor injection, i.e., just
prior to T1 scan at 0 min time point. MRI signal was quantified in regions of interest (ROIs)
covering the injection sites. Each ROI volume was defined by a cylinder with a diameter of 1.2
mm in the coronal plane and a thickness of 1 mm along the rostrocaudal axis at the site of injection,
centered on the tip of the internal cannula, and computing signal amplitudes normalized with
respect to identical control ROIs placed on respective coordinates on the contralateral side of the
brain without sensor delivery.
Histological analysis
To verify cannula placement following MRI contrast agent injection experiments, animals were
anesthetized with carbon-dioxide and transcardially perfused with a solution of PBS (pH 7.4) with
1% sodium nitrite, followed by 4% wt/vol paraformaldehyde in 0.1 M phosphate buffer (pH 7.4).
Brains were removed, postfixed for 90 min in perfusion fixative, and cryoprotected in a series of
20% and 30% sucrose in 0.1 M phosphate buffer each overnight at 4 oC. Coronal sections of 40
µm thickness across a range extending ~2 mm anterior and posterior to the cannula insertion site
was sectioned on a cryostat (Microm HM 550). Standard protocols were used for choline
acetyltransferase (ChAT) immunohistochemical and cresyl violet (Nissl) histological staining.
Briefly, for ChAT staining, free-floating sections in well-plates were first incubated in 1%
H2O2/50% methanol solution for 10 min, followed by serum-blocking buffer for 60 min at room
temperature. Sheep polyclonal anti-ChAT antibody (ab18736, Abcam) diluted 1:1,000 in
immunobuffer containing 1% normal rabbit serum (16120107, Thermo Fisher) in PBS-0.2%
Triton-X100 (PBS-T) was applied and incubated overnight at 4 oC. The sections were then
incubated with horse-radish peroxidase (HRP)-conjugated rabbit anti-sheep IgG secondary
29
antibody (818620, Thermo Fisher) diluted 1:1,000 in immunobuffer for 2 h at room temperature,
and developed in 3,3-diaminobenzidine tetrahydrochloride (DAB) solution (34002, Thermo
Fisher). The sections were washed between each step (3× 5 min) in PBS-T. The sections were then
mounted onto 0.5% gelatin/0.05% chrom alum coated glass slides, allowed to air-dry, and
dehydrated through a series of alcohols (75%, 85%, 95%, 100% twice 5 min each), cleared with
xylene, and coverslipped with Permount (Fischer Scientific, Pittsburgh, PA). The sections were
viewed with Olympus BX51 light microscope. Sections processed to determine non-specific
staining by following the same procedures, but with omission of the primary antibody, showed no
immunohistochemical labeling.
Diffusion of the fluorescent nanosensors
The fluorescent nanosensors were fabricated as the pH-MRNS except for 0.001 mg R18 was
incorporated to the mixture of PVC, DOS and THF. 2 µL of nanosensors containing 47 µM of Gd
was delivered into a phantom made of 0.6% agarose. The follow-up imaging was performed on
IVIS Lumina II (Perkin Elmer) small animal imager in fluorescence mode with a 535 nm excitation
filter and DSRed emission filter at 0, 2, 5, 10, 15, 20, 30, 40, 50 and 60 min post injection for 1 h.
To obtain quantitative information about nanosensor diffusion, circular ROI was positioned
centered near the tip of injection site and fluorescence signal amplitudes per unit area (mm2),
normalized with respect to signal intensity at t = 0 was acquired using Image J software.92 Relative
intensity from three sets of identical ROIs were obtained and averaged for the each time point.
Acetylcholine signal analysis and group sizes
Data for final analysis was extracted from 19 animals divided into three main groups: ACh-MRNS
(N = 6); control (N = 6), and pH-MRNS (N = 3) groups. Only animals with correctly placed cannula
tips into the mPFC, as judged after surgery from MRI scans and histological analyses, were
30
included in the data analysis. Based on this criteria, four animals with incorrectly placed cannula
were excluded from the final analysis. No randomization was conducted to determine allocation
of different animal groups for the MRI scan procedure. Instead, animals were split to both
experimental and control groups in each procedure. Prior to ROI image analyses, individual cases
were assigned with a random number to ensure analyses were conducted in a blinded fashion.
Image and data analysis
All data analysis and image processing was performed with Bruker Paravision 5.1 software
(Billerica, MA, USA), Matlab (Mathworks, Natick, MA), and itk-SNAP.94 Images were
reconstructed and analyzed using custom routines running in Matlab. Relaxivities were calculated
from T1 obtained from itk-SNAP and concentration of Gd(III). Graphs and illustrations were
compiled using Origin (OriginLab, Northampton, MA, USA) and Illustrator (Adobe, San Jose,
CA, USA), respectively.
Statistical analysis
All the relaxivities, T1, sizes, and zeta-potentials collected from the in vitro experiments are
average of three separate studies using different batches of ACh-MRNS or pH-MRNS. The T1
collected from the in vivo experiments are average of T1 collected from 6 rats in Control group, 6
rats in Experiment group and 3 rats in pH-MRNS group using different batches of ACh-MRNS or
pH-MRNS.
The differences between groups for the selectivity study was calculated using one-way analysis of
variance (ANOVA) and adjusted using Tukey’s HSD for multiple comparisons. Kolmogorov-
Smirnov normality test was used to test for normality. Differences between each group obtained
from in vivo data were calculated using Student’s t- test. The ɑ level for all statistical analyses was
31
set at 0.05. Sample size was chosen based on previous reports on MRI method development70 and
drug concentrations.95 When P < 0.05, the difference was considered as significant.
Errors (Table 2.1, 2.2, Figures 2.7B and C, 2.12B, 5C and D, Figure 2.14) were propagated from
standard deviation (S.D.) of T1 (σT1) or normalized T1 (σ(normalized T1) using formula derived
from previous reports.96
The equation used for Table 1 and 2 is defined as:
1 1( ). .
[ ]
T average of TS D
Gd
=
where, [Gd]: concentration of Gadolinium in mM.
The equation used for Figures 2.7B, C, and 2.12B is defined as:
2
1 1. . ( )S D T average of T −=
The equation used for Figures 2.14C, D, and 2.6 defined as:
2
1 1. . ( )( )S D normalized T average of normalized T −=
where, normalized T1: T1 of each animal normalized to 0 min post nanosensor injection.
2.3 Results and Discussion
Nanosensor mechanism and particle characterization
In our nanosensor design we fabricated highly-plasticized polymer nanoparticles as a platform and
functionalized the surface to co-immobilize both transduction and signaling moieties (Figure 2.1).
The mechanism is based on the enzymatic hydrolysis of acetylcholine by BuChE into choline and
acetic acid, and the resulting reduction in local pH alters the water coordination97 of the pH-
sensitive contrast agent leading to an increase in r1 and T1 relaxation rate (1/T1).
32
Figure 2.1. Schematic of the nanosensor structure and mechanism. (A) The pH-MRNS. Only
pH-sensitive contrast agents were covalently conjugated to the DSPE-PEG lipids and coated on
the surface of the lipophilic core. Without co-immobilized BuChE, acetylcholine will not be
hydrolyzed to alter local pH. (B) The ACh-MRNS. Both pH-sensitive contrast agents and BuChE
were covalently conjugated to the DSPE-PEG lipids and coated on the surface of the lipophilic
core. The The BuChE catalyzes the hydrolysis of acetylcholine to choline and acetic acid, and the
resulting drop in local pH triggers a conformational switch of the contrast agent: There is one more
water molecule coordinated to one Gd(III) chelate in acidic conditions compared to its structure in
basic conditions, which leads to increased T1 relaxation rate (1/T1) of the contrast agent.
33
The close proximity of the enzymes and pH-sensitive contrast agents on a nanoscale platform
ensures MR-signal changes are generated by the contrast agent at the origin of the pH changes
created by the enzymatic hydrolysis of acetylcholine. The H+ will be generated in the
microenvironment close to the surface of the ACh-MRNS during the enzymatic hydrolysis and
detected by the pH-sensitive contrast agent nearby. The localized pH changes result from a
gradient as acetylcholine is consumed and H+ ions are released at a fast rate. The signal is
reversible, dependent on acetylcholine concentration, and achieves a limit of detection (LOD) and
sensitivity in the physiological concentration of acetylcholine (nanomolar to micromolar in the
extracellular space of the brain).39
We characterized two types of nanosensors intended for the in vivo study: (1) ACh-MRNS and (2)
a pH-sensitive sensor (pH-MRNS) (Figure 2.1 and Figure 2.2). The ACh-MRNS contained both
the pH-sensitive contrast agents and BuChE conjugated to nanoparticles for specific detection of
acetylcholine; and the pH-MRNS was conjugated with pH-sensitive contrast agents only, i.e.,
without enzyme. The pH-MRNS served as a control due to potential interference from systemic
changes in pH. The structure of nanoparticle is similar to the optode based nanosensors previously
reported by our lab.98-99 The nanoparticle is composed of a core of highly-plasticized polyvinyl
chloride (PVC) and a coat of amphiphilic DSPE-PEG-lipid. The high molecular weight PVC
(Sigma catalog No.: 81392; Mw ~90,000) was dissolved in bis-(2-ethylhexyl)sebacate (DOS) to
form a highly lipophilic nanoparticle. The surface of the particles is derivatized through the use of
DSPE-PEG lipids as a biocompatible coating on the lipophilic surface (Figure 2.2). Separately,
pH-sensitive contrast agents (Compound 1) and BuChE (Compound 2) were covalently linked
sequentially to corresponding DSPE-PEG-lipid derivatives (see Materials and methods).
34
Figure 2.2. Fabrication of the ACh-MRNS and pH-MRNS. To fabricate the pH-MRNS,
contrast agents conjugated DSPE-PEG lipids and DSPE-PEG-DBCO were coated on the lipophilic
core of the nanoparticle. The pH-MRNS react with azide modified BuChE to form the ACh-
MRNS.
35
Figure 2.3. Structure and 1H NMR spectrum of pH-sensitive chelator. Structure (inset) and
the spectrum was obtained from a Varian Inova 500 MHz NMR spectrometer. 1H NMR (500 MHz,
D2O): δ = 8.35 ppm (s, 1H), 8.22 ppm (m, 1H), 7.01 ppm (m, 1H), 4.97 ppm (m, 1H), 3.66 ppm
(s, 2H), 3.36 ppm (m, 16H), 3.05 ppm (m, 6H).
36
We synthesized the transduction element, Gd(NP-DO3A), a pH-sensitive analogue of the
clinically-used Gd(DOTA) based on the procedure by Woods et al.,91 (Figure 2.3). In this
structure, paramagnetic Gd(III) is chelated within a twelve-member ring containing a p-
nitrophenol. Under basic conditions, the probe is coordinated to a single water molecule, but under
acidic conditions, two water molecules will coordinate to Gd(III), leading to increase in 1/T1
(Figure 2.1).
The size and surface charge of fabricated nanoparticles were characterized by dynamic light
scattering (DLS) and zeta-potential in 1× PBS, pH 7.4, respectively. We measured the average size
of three batches of the ACh-MRNS and pH-MRNS using DLS. The average hydrodynamic
diameter of ACh-MRNS was 96 ± 26 nm (mean ± standard deviation) and the size of the pH-
MRNS was 77 ± 23 nm. Figure 2.4A and 2.4B display the size distribution of a single nanosensor
batch, which is representative of a typical batch. The surface charge of the ACh-MRNS and pH-
MRNS was -41 ± 1.9 mV and -29 ± 3.4 mV, respectively. The negative charge of BuChE may
contribute to the difference in zeta-potential.100 We also performed transmission electron
microscopy (TEM) using one batch of the ACh-MRNS as an example, and the generated image in
Figure 2.4C indicated spherical ACh-MRNS with a size of 118 ± 25 nm and distribution shown
in Figure 2.4D. The DSPE-PEG lipid coat was also observed in the TEM (Figure 2.5). The
concentration of nanoparticles (2.62 ± 0.13 × 1014 particles/mL) was also identified using an
nCS1TM nanoparticle analyzer (Spectradyne, Torrance, CA). By combining the concentration and
information from the amount of Gd bound on the surface of the nanoparticle quantified by
Inductively Coupled Mass Spectrometry (ICP-MS), we estimated the density of Gd(III) on the
surface of the nanoparticle to be 0.99 ± 0.27 atoms of Gd/nm2.
37
Figure 2.4 Characterization, in vitro calibration and selectivity of nanosensors. Dynamic light
scattering (DLS) analysis showing size distribution of nanosensors, (A) ACh-MRNS, and (B) pH-
MRNS. (C) TEM image of the ACh-MRNS. (D) Distribution of sizes of the ACh-MRNS from the
TEM image.
38
Figure 2.5. TEM image of the ACh-MRNS using NanoVan stain. The coated lipid layer was
positively stained and shown in bright at the surface of the nanoparticle.
39
The r1 of ACh-MRNS and pH-MRNS was determined as 6.36 and 6.91 mM-1 s-1 (pH 7.4),
respectively, from the concentration of Gd(III) and the T1 collected from a 7 T Bruker Biospec
small animal MRI scanner (Bruker Inc., Billerica, MA). In addition, the r1 of ACh-MRNS was
also measured in the low field with a 1.5 T Bruker Minispec mq60 NMR analyzer (Bruker Inc.,
Billerica, MA) as 12.1 mM-1 s-1 (pH 7.4) at 37 oC. The r1 of both ACh-MRNS and pH-MRNS were
approximately twice the value obtained from the clinically used DOTA-Gd101 and the free Gd(NP-
DO3A) (Table 2.1), indicating suitable contrast for subsequent studies. Compared to free
molecules, the density of bound contrast agents and the slower tumbling rate of the particles may
restrict their internal and overall motion leading to a longer rotational correlation time of the water-
bound contrast agent (τR), which would theoretically account for the increased r1 at the same pH.102-
103
Sensor pH dependence
To evaluate the pH-dependence of the r1 and the corresponding 1/T1, we suspended ACh-MRNS
in PBS at pH 6, 6.5, 7, 7.4 and 8, and scanned using MRI. When the pH decreased from 8 to 6, the
signal intensity (1/T1) increased by 24% (Figure 2.6A). The corresponding r1 of ACh-MRNS at
different pH are indicated in Table 2.2. The extracellular pH in the brain is approximately 7.3.104
Within the physiological pH range 7.2 to 7.8, the 1/T1 of the ACh-MRNS changed by less than 5%
(Figure 2.6A). The moderate response in this range means that the ACh-MRNS are minimally
affected by global pH fluctuations in the brain. To further assess the effect of pH on the response
of ACh-MRNS’s, we calibrated the acetylcholine response of the sensor in a background of varied
pH using a 1.5 T Bruker Minispec mq60 NMR analyzer (Bruker Inc., Billerica, MA). We did not
observe significant differences in the calibration against acetylcholine at pH 7.2, 7.4 and 7.8
(Figure 2.6B). According to the literature,105 BuChE is active between pH 6 to 8.
40
Table 2.1. Relaxivity (r1) of contrast agents used in this study.
Contrast agent r1 (mM-1 s-1) at pH 7.4
DOTA-Gd 3.2191
Gd(DO3A-NP) 3.15 ± 0.47
ACh-MRNS 6.36 ± 0.44
pH-MRNS 6.91 ± 0.65
*Errors were calculated from S.D. of T1 of three independent tests using error propagation.
Table 2.2. Relaxivity (r1) of ACh-MRNS in different pH used in this study.
pH r1 (mM-1 s-1)
8 6.14 ± 0.57
7.4 6.36 ± 0.44
7 7.13 ± 0.32
6.5 7.87 ± 0.35
6 8.55 ± 0.56
*Errors were calculated from S.D. of T1 of three independent tests using error propagation.
41
Figure 2.6. In vitro nanosensor to pH dependence. (A) Response of nanosensors to varying pH
was examined in 1× PBS at pH 6, 6.5, 7, 7.5 and 8. (B) The ACh-MRNS were exposed to solutions
of 0, 5, 10, 2, 30, 40, 50 and 100 µM of acetylcholine in 1× PBS, pH 7.2, 7.4 and 7.6. The 1/T1
was calculated from an average T1 of three independent measurements. Error bars were calculated
from S.D. of T1 using error propagation (error bars are small in (B) to be observed).
42
When acetylcholine is hydrolyzed, the enzyme will remain active until the pH drops below 6, at
which point the activity will be reduced. Thus, we believe that the capability of the ACh-MRNS
to measure acetylcholine in this range is not affected by endogenous changes in pH. It will be
important to continue to use the pH-MRNS as a control to further mitigates the risk of interference
from physiological pH changes.
Sensor calibration
We calibrated the nanosensors by suspending nanoparticles corresponding to 0.079 mM
conjugated Gd(III) with concentrations of acetylcholine varying from 0 to 100 µM in 1× PBS, pH
7.4. The ionic strength of this buffer is 162.7 mM which is consistent with the ionic strength of
artificial cerebrospinal fluid (ACSF) (152.8 mM). Since the sensors were delivered to the CSF
instead of blood, we did not incorporate serum proteins to our measurements. A clear gradient in
brightness was observed from the T1-weighted MR image (Figure 2.7A): the higher concentrations
of acetylcholine led to increasingly brighter images, which represents a higher 1/T1. A calibration
curve of 1/T1 as a function of concentration of acetylcholine (Figure 2.8B), showed an
enhancement of 1/T1 by more than 10% when the concentration of acetylcholine increased from 0
to 6 µM, and 20% when the concentration increased to 10 µM (blue line). Using the sigmoidal fit
of the calibration curve to calculate the analytical properties of the ACh-MRNS, (Figure 2.7B) we
found a lower limit of detection (LLOD) of 2.64 µM and a sensitivity of ±4 µM (calculated from
the fitted curve and error bar in Figure 2.7B) at an acetylcholine concentration of 10 µM. In
comparison, a control study using an equivalent amount of pH-MRNS with the same concentration
of Gd(III) and free BuChE (not conjugated to the particle) led to no significant change in 1/T1
when the concentration of acetylcholine was increased from 0 to 500 µM (grey). This result
indicates that the amount of free enzyme was not sufficient to create T1 contrast by the pH-MRNS.
43
Figure 2.7. Characterization, in vitro calibration and selectivity of nanosensors. (A) Higher
concentrations of acetylcholine led to brighter MR images. (B) 1/T1 of ACh-MRNS was enhanced
when higher concentrations of acetylcholine were present. The ACh-MRNS (blue) were exposed
to solutions of 0, 1, 2, 6, 10, 30 and 100 µM of acetylcholine. The pH-MRNS and free BuChE
(grey) were exposed to solution of 0, 10, 50, 100 and 500 µM of acetylcholine. (C) The selectivity
towards acetylcholine of ACh-MRNS. The nanosensors were exposed to PBS buffer or to solutions
of acetylcholine (0.1 mM), glutamate (5 mM), dopamine (5 mM), GABA (5 mM) and glycine (5
mM). 1/T1 in solutions of acetylcholine was significantly higher than other groups (one-way
ANOVA, *P < 0.005, for N = 3). The 1/T1 in panel (B) and (C) were calculated from an average
T1 of three independent measurements. Error bars in panel (B) and (C) were calculated from
standard deviation (S.D.) of T1 using error propagation.
44
Our design theory was that the nanoscale particle is a necessary component of the sensing
mechanism. In short, adding the individual components without immobilization on the scaffold
would not be sufficient for two reasons: First, the sensing components would diffuse away from
each other in vivo and would not remain in proximity for sensing. Second, the enzyme and pH
indicator are preconcentrated on the sensor, which sets up a localized pH environment that would
not be seen if the enzyme was not attached to the particle. In order to validate the pH effect
confined to a localized microenvironment, we measured the pH in the mixture of 0 to 5 mM of
acetylcholine with the ACh-MRNS (Figure 2.8A) and pH-MRNS (with free BuChE) (Figure
2.8B), respectively, using fluorescein, a commonly used pH indicator. The results showed that no
detectable pH changes of the bulk buffered system were detected until the level reached 5 mM of
acetylcholine. Thus, we extrapolate these findings to the enhancement of 1/T1 by locally generated
protons (H+) from the enzymatic hydrolysis of acetylcholine in the microenvironment at the
surface of the ACh-MRNS. Similar localized pH effects have been observed in biosensors and
microelectrodes, as reported by our lab and other groups.20, 106-108 Also, previous report suggested
that within the hydrophilic microenvironment close to the surface of an electrode, the pH can be
inhomgeneously distributed depending on the local chemical environment.109-111 This discovery
suggested the local pH-drop created within the microenvironment can be detected by the contrast
agents in the vicinity. To confirm that the change in 1/T1 was not initiated by free Gd3+, a xylenol
orange test showed that the A573/A433 did not increase when higher concentrations of
acetylcholine were hydrolyzed by the ACh-MRNS and no free Gd3+ was generated in this process
(Figure 2.9). Thus, the increase of 1/T1 was not caused by the alteration of free Gd3+.
To determine the reaction time of the enzymatic hydrolysis, we performed an Ellman’s assay
which indicated that the conjugated BuChE consumed 5-100 µM of acetylcholine between 40 s
45
Figure 2.8. pH change in the mixture of the nanosensor and acetylcholine. Changes of
fluorescence at 520 nm led by the hydrolysis of acetylcholine with concentrations varying from 0
to 5000 µM by (A) ACh-MRNS and (B) pH-MRNS and unconjugated BuChE using fluorescein
as a pH indicator. Error bars were calculated from S.D. of three separate tests of fluorescence using
error propagation.
46
Figure 2.9. Xylenol orange test. (A) Xylenol orange test of the mixture of the ACh-MRNS and
acetylcholine. The A573/A433 didn’t increase when higher concentrations of acetylcholine were
present, indicating that increasing acetylcholine is not removing Gd from the sensor. (B) For
reference, xylenol orange test of a solution of increasing Gd(NO3)3 levels. The A573/A433 increased
when higher concentrations of Gd(NO3)3 were present. The value of A573/A433 is an average of
three independent measurements. Error bars were calculated from the standard deviation (S.D.) of
recorded absorbance at 573 nm and 433 nm, respectively, using error propagation (error bars are
small to be observed).
47
Figure 2.10. Kinetics of BuChE. Changes of absorbance at 412 nm led by hydrolysis of 0, 5, 10,
50, and 100 µM of acetylthiocholine by (A) free BuChE and (B) conjugated BuChE in 2 min after
the start of the reaction in the Ellman’s assay was plotted. Error bars were S.D. of three separate
tests of absorbance.
48
and 2 min (Figure 2.10A and B). For MRI measurements, though, our acquisition times are longer
than this two-minute timeframe. For the acquisition, we used a Rapid Acquisition with Relaxation
Enhancement with Variable TR (RARE-VTR) sequence to obtain the T1 in a total time of 9.5 min.
This sequence was chosen as it minimizes distortion during in vivo imaging.112 During the scan,
signal intensities at six different TRs were collected sequentially. The relative changes of signal
intensity elicited by increasing acetylcholine levels decreased from 20% at the first TR (0-4.3 s)
to 0 at last TR (236.3-570.1 s) scan sequence (Figure 2.11). These results indicate that the in vitro
calibration mainly represented the response of ACh-MRNS to acetylcholine in the first 2 min of
the scan during which the enzymatic hydrolysis of acetylcholine took place. Thus, although the
nine-minute scan is essential for obtaining a reliable T1, the final value correctly reflects the shorter
timeframe of acetylcholine hydrolysis.
Sensor selectivity
Since ACh-MRNS is enzyme-based, we expect high specificity against other neurotransmitters
(Figure 2.7C). To verify the selectivity, we suspended nanoparticles (corresponding to 0.079 mM
conjugated Gd) in 200 µL solutions of either PBS buffer, acetylcholine, glutamate, dopamine,
GABA or glycine. By measuring 1/T1, we found that only acetylcholine solution led to a more than
20% increase compared to PBS (P = 0.0001, F-value = 21.68, df = 17; ANOVA with Tukey’s
post-hoc test), and none of the potential interfering neurotransmitters elicited any significant
difference (P > 0.5). These results are consistent with the assumption that BuChE selectively
hydrolyzes acetylcholine, and supports our nanosensors to selectively respond to acetylcholine.
Acetylcholine detection in the rat brain in vivo
We investigated the ability of ACh-MRNS to detect endogenous acetylcholine release in the rat
medial prefrontal cortex (mPFC).
49
Figure 2.11. Relative Signal intensity at different TR. Signal intensities in in vitro calibration
at TR = 70, 291, 576, 976, 1651, and 5000 ms were collected sequentially in 4.3, 19.1, 38.1, 64.8,
110.0 and 333.8 s, and normalized to the corresponding signal intensity elicited by 0 µM of
acetylcholine, respectively. The normalized data was plotted and fitted in sigmoidal curves. Error
bars were calculated from S.D. of T1 using error propagation.
50
First, to assess the amount contrast provided by ACh-MRNS in the living brain tissue, 2 µL
solution (47 µM of Gd(III)) of the nanosensors were injected into the mPFC through implanted
cannula of anesthetized rat subjects (Figure 2.12). An increase of 60% in 1/T1 was produced
(Figure 2.12B) and a strong T1-weighted contrast was observed at the site of injection (Figure
2.12C). Placement of the cannula in the mPFC was verified during the anatomical MRI scans and
after surgery from histological analysis (Figure 2.13).
Next, time-course changes in acetylcholine-dependent 1/T1 were acquired by stimulating the
release of acetylcholine in the mPFC using a pharmacological agent, clozapine. Clozapine, an
atypical antipsychotic drug, has been shown to induce 2-3 fold increase in acetylcholine
concentration from the basal level, which peaks after 30 min, and is sustained for over one hour in
the rat mPFC.95 Briefly, the procedure for in vivo imaging included three consecutive scans after
nanosensor delivery on a 7 T MRI scanner at an interval of every 23 min (MRI scan procedure: 0,
23, 46 min post nanosensor injection; Figure 2.14A). The experimental group (N = 6) consisted
of ACh-MRNS infused through the cannula and a concurrent subcutaneous injection of clozapine
into the hind of the animal. For the control group (N = 6), ACh-MRNS were delivered without
clozapine treatment. T1 was quantified in regions of interest (ROIs) covering the injection sites.
Each ROI volume was defined by a cylinder with a diameter of 1.2 mm and a thickness of 1 mm
centered at the injection site, and T1 was measured and normalized with respect to control ROIs
identically placed on the contralateral side of the brain with no sensor delivery. Examination of
the T1 and 1/T1 time courses in Figures 2.14B-D clearly displayed a difference in 1/T1 after 30 min
post sensor delivery in the experimental group (blue) compared to the control group (green). For
both groups, a slight decrease in 1/T1 was observed at the 23min time period.
51
Figure 2.12. In vivo sensor contrast. (A) Schematic diagram of placement of cannula and ACh-
MRNS infusion (blue) aimed at rat mPFC, and MRI data acquired from sensor encompassing slices.
(B) ROI-averaged MR signal intensity showing increase in 1/T1 for >60% in the post-nanosensor
injection in comparison to pre-injection slice. (C) Coronal MR image brain slice (bregma: +2.8
mm) of pre- (left) and post-nanosensor injection (right), with circle ROI (blue) defined for analysis.
Arrowhead (white) indicates the position of cannula, and the circle (blue) defines the ROI for
analysis. Error bars were calculated from S.D. of T1 using error propagation. Diagram adapted
from Paxinos and Watson 113.
52
Figure 2.13. Histology. Representative photomicrographs showing placement of cannula in the
rat cortex. Coronal rat brain sections in the bottom left panel shows choline acetyltransferase
(ChAT) immunoreactivity and cresyl violet (Nissl) histological staining at the level of mPFC.
Diagram adapted from Paxinos and Watson. 113
53
A markedly evident 1/T1, however, were observed at 46 min time point with a significant
difference of more than 13% 1/T1 in the experimental group compared to the controls (P = 0.018,
Student’s t-test). With microdialysis, Ichikawa et al. uncovered that clozapine elicited a 2.5-times
increase in the concentration of acetylcholine in the mPFC half hour after the drug administration.
.114 The peak time recorded by the microdialysis is consistent with the result from the ACh-MRNS.
In addition, the pattern of 1/T1 decrease over time in the control groups were consistent with the
effects of particle diffusion, as indicated by a similar ~19% decrease in 1/T1 by observing the time
course of similarly injected fluorescent nanoparticles in the 0.6% agarose phantoms (Figure 2.15).
The particle diffusion was significantly slower than a molecular dye (data not shown). Previous
reports show that nanoparticles can be cleared by microglia/macrophages in the CNS.115-117 The
rate of sensor diffusion and phagocytosis is a potential factor for in vivo measurements, and we
reason that the ACh-MRNS diffused at the site of injection would lead to a decrease in 1/T1,
however, this effect was offset by the response of ACh-MRNS against endogenously released
acetylcholine in the experiment group as observed by the significant difference detected in the scan
at 46 min compared to controls.
To ascertain whether global pH changes in the brain were induced by clozapine administration and
interfered with the sensor response, identical scanning procedures were performed on another
cohorts (N = 3). In this group, the pH-MRNS (grey, Figure 2.14C) were delivered with a
concurrent clozapine injection, procedurally identical to the experimental group above. A similar
1/T1 was observed after injection of nanosensors (0 and 23 min post-injection period). However,
at a subsequent time point at 46 min post-injection, a difference of >15% 1/T1 was observed in
comparison to the experimental group (P = 0.005) which indicates that clozapine does not cause
significant acidic conditions in the brain.
54
Figure 2.14. Acetylcholine detection in vivo. (A) Experimental procedure: Delivery of
nanosensors (ACh-MRNS or pH-MRNS) through cannula followed by subcutaneous
administration of drug (clozapine) and then three consecutive MR scans 23 min apart denoted, t =
0, 23, 46 min. (B) Coronal brain slices showing time-courses of acetylcholine detection. In the
experimental group (top panels, N = 6), ACh-MRNS were injected through the cannula with
clozapine administration, while the control group (bottom panels, N = 6) comprised ACh-MRNS
55
delivery without clozapine. For the purpose of display, the heatmap of the top layer (T1-registered
map) was generated between 600 and 1600 ms and overlaid onto the image. (C) Distinct
acetylcholine signal changes accompanied by a difference of 13% in 1/T1 (P = 0.018; Student’s t-
test) between experimental (blue) and control groups (green) was observed at 46 min, indicating
distinct detection of acetylcholine driven by local enzymatic hydrolysis by the ACh-MRNS
induced by clozapine, as shown by higher 1/T1 in the experimental group. Identical MR scanning
procedures were conducted in a new cohort (pH-MRNS group, grey, N = 3) to study the localized
effects of enzymatic hydrolysis of acetylcholine to trigger changes in 1/T1. In this group, pH-
MRNS (nanosensors with pH-sensitive Gd and without conjugated enzymes) were delivered along
with clozapine administration (grey). After 46 min, 1/T1 also showed a difference of >15% (P =
0.005) in comparison to the experimental group, showing global pH changes were not detected
and induced by clozapine administration and confirming the validity of specific detection of
acetylcholine driven by local enzymatic hydrolysis by the ACh-MRNS. (D) The individual data
points of relative 1/T1 in the ACh-MRNS (blue, N = 6), control (green, N = 6) and pH-MRNS
(grey, N = 3) group at 0, 23, 46 min post-injection were plotted. Error bars were calculated from
S.D. of normalized T1 using error propagation.
56
The individual data points of relative 1/T1 in this group is available in Figure 2.14D. This result
suggests that ACh-MRNS reliably detect acetylcholine levels as governed by the predicted
mechanism of the nanosensor, and not due to a global change in pH in the brain. The individual
difference in activation of microglia and rate of diffusion between each animal may explain the
variance in the control and pH-MRNS cohort.
Discussion
The use of MR-active nanosensors to image acetylcholine is particularly attractive for several
reasons. First, the nanoscale platform incorporates all sensing components together, and hence the
close proximity between cholinesterase and pH-sensitive contrast agent creates a localized effect
and facilitates specific detection of acetylcholine. Also, the nanosensors are based on a modular
design that can be extended to the detection of other neurotransmitters and physiological analytes,
by simple substitution of enzymes or functional contrast agents.
In our studies, a 1/T1 of ~13% at 46 min indicates a micromolar increase in the acetylcholine levels,
as estimated according to the in vitro calibration. This finding was comparable to other groups’
attempts to measure acetylcholine in the brain, although there is no direct method of comparison
in the literature. As cited above, our studies reflect the temporal increase in acetylcholine in
response to clozapine stimulation, as had been observed previously via microdialysis.95 Other
examples include those that used coated microelectrodes, nicotine118 or KCl89 as local stimuli to
trigger an increase of acetylcholine of up to 25 µM.55, 119 However, we emphasize the difficulty of
direct comparison between methods of analysis and pharmacological stimulation. Our future
studies will focus on in vivo calibration and rigorous comparison as an assessment of the
advantages and disadvantages of various methodologies.
57
Figure 2.15. Diffusion of pH-MRNS in phantom brain. Rhodamine 18 incorporated pH-MRNS
diffusion profile in 0.6% agarose phantom imaged using IVIS fluorescence imager at 0, 2, 5, 10,
15, 20, 30, 40, 50 and 60 min post-injection. The relative intensity per area data at each injection
site was collected using Image J, and normalized to the relative intensity per area at t = 0 min.
Error bars indicate S.D. derived from three ROI intensity measurements at each time point
indicated by the blue circle (shown on top right phantom).
58
In our in vivo study, we expect that nanoparticles diffuse similarly in both Experimental and
Control groups, hence a sustained 1/T1 signal is attributed to the molecular changes in the brain
rather than other natural variation occurring in the brain, such as pH changes, anesthetic effect,
temperature differences, as the two groups differ only by administration of the drug otherwise all
other experimental conditions were identical. We did observe larger differences in the Control
group (Figure 2.14D), however, the variation becomes more pronounced at later scanning stages,
i.e., 46 min, than earlier scan time points. This is observable in both Control and pH-MRNS groups,
but evidently less in the ACh-MRNS group. An alternative faster pulse sequence paired with
higher-resolution imaging could be performed to improve temporal resolution and increase the
fidelity of acetylcholine detection.120-121 The current methodology demonstrates relative changes
in acetylcholine levels, and would require in-situ calibration to be performed before quantitative
results could be achieved.
Due to the difficulty of delivering our nanosensors noninvasively through the blood-brain barrier
(BBB),122 which remains a key challenge for molecular neuroimaging applications in live animals,
we have implemented a cannula placement for delivery of nanosensors aimed at the mPFC. For
ongoing applications in animal studies, disruption of BBB using hyperosmotic shock or ultrasound
methods which have been used to deliver small molecules123 and nanoparticles124 into the brain,
will be explored to improve probe delivery. Also, non-disruptive methods to deliver nanoparticles
to the CNS have been established in pre-clinical studies. Polymeric nanoparticles are coated with
artificial amphiphilic polymers or protein-based antibodies, peptides, or receptors to overcome the
BBB via transcytosis for therapy against stroke, Alzheimer’s disease, or Parkinson’s disease.125-
126 Thanks to the modular feature of the ACh-MRNS, modification of the coatings with targeting
protein will also be considered as a strategy to deliver the nanosensors. As with all measurement
59
methods, the risk of distorting the biological environment by the removal of analyte for detection
is a real possibility. In these studies, due to the size of the sensors, we believe they are located in
the extrasynaptic microdomain. Thus, the volume transmission i.e., “spillover” acetylcholine are
the species that are predominantly detected, rather than acetylcholine molecules directly involved
in the synaptic cleft. Hence, the acetylcholine in the synaptic cleft will be readily available for
recycling process, reducing the buffering effect by the sensors. In the future, as we strive to reduce
the size of the sensors and target them to the synaptic cleft, the possible consumption of analyte
may become a greater issue. Lastly, enhanced sensitivity for detection of acetylcholine can be
achieved by exploring a more active enzyme such as acetylcholinesterase.127 These steps will
facilitate the application of MRI and nanosensors for chemical imaging of neurotransmitters
fundamental to the understanding of brain function and disease.
2.4 Summary
In summary, we have developed and characterized a neurotransmitter-sensitive MR-active
nanosensor for the detection of acetylcholine in the brain. Acetylcholine is a neurotransmitter
known to play a prominent role in mammalian social behaviors and neural processes that govern
cognition and memory. As such, there is a considerable interest for imaging this molecule. In this
study, we firstly demonstrated the ACh-MRNS was capable of measuring acetylcholine in low
micromolar concentration by co-immobilizing cholinesterase and pH-sensitive contrast agents on
a nanoparticle. And we further proved that the ACh-MRNS can detect clozapine induced
endogenous release of acetylcholine in the rat brain. The ACh-MRNS we report here is the first
nanosensor for the detection of acetylcholine using MRI in living brain, as characterized and
implemented both in vitro and in vivo. Also, the modular design of sensors offers a sensing
platform which can be extended to integrate different components for detection of other
60
neurotransmitters and physiological analytes, by substituting the enzymes or functional contrast
agents to achieve better specificity and sensitivity.
2.5 Acknowledgement
This work was supported by the National Institutes of Health through Grant R01NS08164. We
thank K. Bardon, P. Larese-Casanova, C. Marks for technical assistance in flash column
chromatography, ICP-MS, and TEM, respectively. We also thank C. Ferris and P. Kulkarni for
help in MRI setup and pulse sequences.
61
Chapter 3: Glucose-Sensitive Nanofiber Scaffolds with Improved
Sensing Design for Physiological Conditions
3.1 Introduction
Continuously monitoring physiological analytes such as electrolytes and glucose may
revolutionize disease diagnosis and management by enabling patients and physicians to accurately
track an individual’s analyte levels and fluctuation patterns. Implantable nanosensors offer a
promising platform for physiologic monitoring because their small size makes implantation
minimally-invasive, and the small suite of biocompatible polymers already FDA-approved for
implant coatings and catheters provide a safe starting point for material selection. Optode-based
nanosensors are robust tools for continuous and reversible physiological analyte measurements,
and several designs have successfully monitored glucose, histamine, and sodium in vivo.21, 32, 35 In
optode-based nanosensors, reviewed extensively elsewhere128-130, a hydrophobic plasticized
polymer matrix provides support for hydrophobic analyte recognition elements and hydrophobic
reporters. When the recognition element binds to its target analyte, the binding event causes a
change in the local environment (e.g.; pH change, charge movement, oxygen consumption) and
the reporter’s optical properties change concomitantly. Nanosensors designed around optodes are
essentially nanoparticles that incorporate the recognition and reporting chemistries. The
components are contained within the hydrophobic nanoparticle and the resulting nanosensors’
analytical properties can be tuned by changing the relative ratio of sensing components within the
nanoparticle.
Previous works evaluating optode-based nanosensors for bio-analyte monitoring have used
platforms such as a sliver sensor, which contained individual sensing capsules on a cellulose
62
acetate support,131-132 Others, such as McShane, have encapsulated sensing and reporting
chemistries for glucose contained within alginate microspheres and subcutaneously injected those
microspheres into rats for glucose monitoring.133 The eventual clinical utility of any nanosensor
will depend on the nanosensors’ sensitivity, selectivity, biocompatibility, reversibility, response
time, appropriate residency and clearance time.134 To date, no implantable nanosensor system
meets the clinical requirements for all of those factors.
Any sensor will have a recognition element and a reporting element, and personal glucometers
often use the enzyme glucose oxidase and then use electrochemistry to detect the enzyme’s activity
in response to blood glucose from a finger prick. Alternatively, non-enzymatic recognition
elements such as concanavalin A, a lectin that specifically and reversibly binds to polysaccharides
via hydrogen bonds and van der Waals interaction135, or boronic acids, which reversibly bind to
diols through boronate ester formation136-137 can detect glucose. Boronate ester formation increases
through the addition of electron withdrawing groups to the boronic acid, strengthening diol
binding.138-141 Asher and coworkers used this approach by incorporated a fluoro- electron-
withdrawing group onto their boronic acid derivative and were able to monitor glucose at pH 7.4
with their photonic crystal glucose sensing material.139, 142 Using carbon nanotube-based sensors,
Strano and coworkers also showed that boronic acids with electron-withdrawing groups such as
chloro- and cyano- groups were optimal for their sensing design.143 Thus, we hypothesize that the
sensitivity of boronic acids to glucose at physiological pH can be tuned by increasing or decreasing
the electro-withdrawing ability of functional groups on a boronic acid derivative. We aim to design
optode-based nanosensors that respond to physiologic glucose concentrations by synthesizing
hydrophobic boronic acids with electron-withdrawing groups and fabricating nanosensors with
those boronic acids. The reporter in this design is alizarin, which has a diol group, allowing it to
63
compete with glucose for reversible binding to boronic acids. When bound to a boronic acid, it
fluoresces very strongly, but its fluorescence decreases when displaced from the boronic acid by
glucose. Higher lever of glucose will lead to less binding between boronic acid and alizarin, and
cause larger drop of fluorescence. Based on this mechanism, concentration of glucose can be
measured by the decrease of fluorescence.
A variety of nanosensors have been developed for in vivo glucose monitoring, but many of them
have a limited residence time at the site of injection.35 Despite their short residency time, the in
vivo experiments showed that fluorescent glucose-responsive nanosensors are able to track
changes in glucose levels for up to one hour.35 Similar results were observed with sodium-sensitive
nanosensors, and short in vivo residency times were attributed to particle migration away from and
cellular uptake at the injection site.32 Various approaches have been used to overcome these issues
by immobilizing nanosensors within gels144 or producing high aspect-ratio sensor geometry.145
Gel immobilization improved sensor residence time at the injection site over the course of one
hour, but did not provide a long-term solution to sensor migration because nanosensors are small
enough to diffuse out of the gels.144 Our group previously demonstrated that encapsulating
nanosensors into worm-like geometries with chemical vapor deposition prevented the signal loss
associated with diffusion away from the injection site,145 though the chemical vapor deposition
fabrication methods used in that study have low batch yields. Electrospinning is a high-yield
process that can fabricate continuous polymer nanofibers of optode material. With nanofiber
geometries, implanted nanosensors may achieve a residency time in conjunction with a high
throughput and scalable production technique while retaining advantages of nano-scale sensors.146
Although other groups have utilized electrospinning to fabricate sensors for detecting silver,147
64
mercury,148 nitroaromatics,149 and glucose,150-151 none have shown that their sensor designs
function in vivo.
In this work, to improve stability of our previously presented1 glucose-sensitive nanosensors, we
fabricated nanofiber scaffolds from plasticized polycaprolactone and incorporated the best of the
boronic acid derivatives with alizarin to show that this sensor platform exhibits extended in vivo
sensor residency time.
In addition, we functionalized 4-carboxy-3-fluorophenyl boronic acid with hydrophobic alkyl side
chains of varying lengths to increase the nanosensors’ stability to leaching and sensitivity to
glucose, as compared to previous formulations.1
3.2 Material and Methods
Materials: Carboxylated poly(vinyl chloride) (>97% GC) (PVC-COOH), bis-(2-
ethylhexyl)sebacate (DOS), polycaprolactone (Mn 70,000-90,000) (PCL),
tridodecylmethylammonium chloride (TDMAC), alizarin, 4-carboxy-3-fluorophenylboronic acid
(1), 3-fluoro-4-methoxycarbonylphenylboronic acid (2a), D-(+)-glucose, tetrahydrofuran (≥
99.9%) (THF), dicyclohexylcarbodiimide solution (60% w/v in xylene) (DCC), N-
hydroxysuccinimide (NHS), aniline(≥ 99.5%), 1-propanol (anhydrous, 99.7%), 1-butanol (HPLC,
99.7%), 1-hexanol (98%), cyclohexanol (99%),sodium sulfate (anhydrous,≥ 99.9%), sodium
chloride, ethyl acetate (anhydrous, 99.8%), hexane (anhydrous, 99.5%), N, N’-dimethylformamide
(DMF) and N, N’-dimethylaminopyridine (DMAP) were purchased from Sigma Aldrich (St Louis,
MO, USA). Octylboronic acid (>97%) and Citroflex A-6 were acquired from Synthonix (Wake
Forest, NC, USA) and Vertellus (Indianapolis, IN, USA), respectively. Phosphate Buffered Saline
(PBS) (1x, pH = 7.4) was purchased as a solution from Invitrogen (Carlsbad, CA, USA).
65
Hydrochloric acid (1.0 N) and sodium bicarbonate were purchased from Fisher Scientific (Fair
Lawn, NJ, USA). 1,2-distearoyl-sn-glycero-3-phosphoethanolamine-N-[methoxy(polyethylene
glycol)-550] (ammonium salt) (DSPE-mPEG550) was purchased from Avanti Polar Lipids, Inc.
SKH1-E mice were acquired from Charles River Laboratories International Inc. (Wilmington,
MA).
Boronic Acid Synthesis. To control response, we systematically functionalized BA1 with alkyl
chains of various lengths (Figure 3.1). The synthesis protocol has been previously developed by
Steglich and coworkers. 152 Specifically, 200 mg BA1 (1.09 mmol, 1 Eq.) was mixed with 40 mg
DMAP (0.33 mmol, 0.3 Eq.) and alcohol 2 (3.27 mmol, 3 Eq.) in 4 mL DMF. DCC solution in
xylene (60% w/v) (1.09 mmol, 1 Eq.) (220 µL) was added dropwise to the reaction mixture at 0o
C, which was then warmed to room temperature and stirred overnight. The urea precipitate was
removed by centrifugation and then the supernatant was extracted with 20 mL ethyl acetate and
0.5 M HCl aqueous solution. This process was repeated three times. The product was washed
with saturated NaHCO3 aqueous solution and then brine (saturated sodium chloride solution). The
organic phase was dried over Na2SO4 and further purified by flash column chromatography. The
product was characterized by 1H NMR recorded on a Varian Inova 500 MHz NMR spectrometer.
1H NMR data is available in the supplementary information.
Optode Composition. Macrosensors, nanofiber scaffolds, and nanoparticle-based sensors were
formed from optode cocktails containing all sensing components. Macrosensors were made from
the following components: 30 mg PCL, 60 μl Citroflex A6, 82.3 µmol of a boronic acid (BA)
derivative (BA2b – BA2c), 2.0 mg (2.49 µmol) TDMAC, and 1.0 mg (4.16 µmol) alizarin. These
materials were placed into a glass vial and then dissolved in 500 μl THF. The boronic acids
incorporated into these formulations were BA1, and BA2a-c.
66
Figure 3.1. Boronic acids incorporated into glucose-sensitive sensors. (A) Structures and (B)
synthesis of boronic acids with different alkyl chain lengths and ring structures.
67
For production of electrospun scaffolds, the general optode cocktail was made with a solution of
12% (weight/volume) of PCL in Citroflex A-6 and THF. Of this weight percentage, 10% was
Citroflex A-6. Specifically, the optode formulation was: 216 mg PCL, 24.0 μL Citroflex A-6, 2.0
mg (2.49 µmol) TDMAC, 1.0 mg (4.16 µmol) alizarin, and 82.3 µmol boronic acid in 2 ml THF.
Three boronic acids, 2a, 2b and 2c, were tested in electrospun scaffolds. Nanoparticle-based
sensors were fabricated with an optode formulation previously described and include: 30 mg high
molecular weight PVC-COOH, 60 μl DOS, 2.0 mg octylboronic acid, 4.0 mg TDMAC, and 1.0
mg alizarin.35 These materials were transferred into a glass vial and then dissolved in 500 μl THF.
Response of Macrosensors to Glucose. Prior to miniaturization to the nanoscale, each new BA
was assessed as a glucose-sensitive macrosensor. The method for testing macrosensor responses
has been described previously.35 Briefly, macrosensors are formed by pipetting 2 µL of optode
onto glass discs adhered to the bottom of an optical bottom 96-well plate. The optodes were then
allowed to dry at least 15 minutes forming thin film macrosensors. A Spectramax Gemini EM
micro plate fluorometer (Molecular Devices, Sunnyvale, CA, USA) acquired fluorescence data
(ex/em: 460/570 nm). After forming macrosensors, each macrosensor was hydrated in 200 µL
PBS (pH=7.4) for 45 minutes. This process was repeated 4 times until the fluorescence intensity
stabilized. After the macrosensors were hydrated, the PBS solution was removed from all wells
and 200 μl of 0.1 M glucose in PBS was pipetted into half of the wells to determine macrosensor
response to glucose. The remaining wells acted as controls and contained fresh, glucose-free PBS.
Changes in fluorescence response were monitored for 60 minutes at a sampling rate of 5 minutes.
The fluorescence intensity of each sensor was normalized to time zero and then the mean was
taken for both the experimental and control groups. The average of the experimental group was
68
subtracted from the control group and multiplied by 100 to obtain a percent change. The error of
percent change was calculated using error propagation.
Fabrication of Fibrous Scaffolds: Electrospinning was performed on a Nanospinner NE 200
(Inovenso, Istanbul, Turkey) equipped with a syringe pump. The optode solution was spun at a
distance of 10 cm from the collector with a rate of 3 ml/hr and at an applied voltage of 15 kV. The
fibers were spun onto either aluminum foil or silanized glass discs attached to aluminum foil for
imaging and testing scaffold response.
Nanofiber Scaffold Responses to Glucose: To determine scaffold response to glucose, scaffolds
spun onto glass discs were removed from the aluminum foil using a 6 mm biopsy punch (Miltex,
Inc., Plainsboro, NJ, USA) and placed in a 96-well optical bottom well plate. PBS (200 µL) was
added to each well and the sensors were hydrated in PBS overnight to stabilize the fluorescence
intensity. All fluorescence measurements (ex/em: 460/570 nm) were acquired using a SpectraMax
Gemini EM plate reader. After hydration, the PBS was replaced with 200 µL of fresh PBS (pH
7.4) as a control or 0.1 M glucose in PBS (pH 7.4). The fluorescent responses were measured for
60 minutes at 5-minute intervals. Fluorescence measurements were normalized to the first time
point and averaged for each experimental group. The average response of the experimental group
was subtracted from the control group and then plotted over time. Error was determined using
error propagation.
Fluorescence Imaging: Images of scaffolds were acquired on a Zeiss Confocal Microscope
(Thornwood, NY) using a 488 nm laser and 10x air objective (PlanApo, NA = 0.17).The laser
intensity was set to 1% (10 mW full power)
69
SEM Acquisition: Images of scaffolds were acquired on a Hitachi S4800 with a 5 kV accelerating
voltage. Samples were not sputter coated. Fiber diameters were measured using Quartz
PCI (Quartz Imaging Corp.) software. Magnification – 10X, NA – 0.45 in
Fabrication of Nanoparticle-based Sensors: The fabrication of nanoparticle-based sensors is
described previously. 35 Briefly, optode was dried overnight on a glass plate, and then transferred
into a scintillation vial. Then 5 ml of PBS (pH=7.4) and 5 mg of DSPE-mPEG (550) in 500 µL of
chloroform was added. The mixture was sonicated for 3 minutes at 40% amplitude using a
Branson digital sonifier (Danbury, CT). The nanosensor solution was pipetted out from vial
leaving residual optode.
In Vivo Studies: Animal procedures were approved by Northeastern University’s Institutional
Animal Care and Use Committee. To determine whether nanofiber scaffolds minimized sensor
diffusion in vivo, glucose nanosensors and scaffolds were prepared as above. Scaffolds were cut
into circular pieces using a 6 mm diameter biopsy punch and sterilized by soaking in 70% ethanol
and then sterile PBS (pH=7.4). SKH1-E mice were anesthetized and then injected with 20 µL of
either nanosensors or scaffolds along their back. To determine the injection volume, the amount
of sensor material in a 6 mm diameter scaffold was estimated and then approximated to the same
amount of material in the nanosensor formulation. Nanosensors were injected with 31G insulin
syringes (BD Biosciences, Franklin Lakes, NJ). Scaffolds were injected using an indwelling
needle assembly.153 The assembly consisted of a 20 G outer needle and a 25 G inner needle with
a blunted tip that acted as the plunger. 3M Vetbond ™ tissue adhesive (3M Animal Care Products,
St. Paul, MN) was then applied to the injection site. Imaging was performed on an IVIS Lumina
II (Perkin Elmer) small animal imager in fluorescence mode with a 465/30 excitation filter and
580/20 emission filter. Mice were imaged every 5 minutes for 1 hour and then at 3 hours post-
70
injection. Fluorescence measurements were analyzed by selecting a region of interest around each
injection spot to obtain the total radiant efficiency of the area. The background-subtracted total
radiant efficiency from each region of interest containing either scaffolds or nanosensors was
measured at each time point and then normalized to the total radiant efficiency at time 0. The
normalized values were then averaged across three mice for both the scaffolds and nanosensors.
To account for sensor degradation over time, scaffolds and nanosensors were prepared as above
and placed into a 96-well plate with a total volume of 200 µL of either PBS or PBS and nanosensors.
Their total radiant efficiency was tracked using the same imaging parameters and data analysis as
the in vivo studies.
3.3 Results and Discussion
Boronic Acid (BA) Selection. The clinical utility of glucose-responsive nanosensors depends on
their ability to exhibit proper dynamic range and sensitivity.7 In the sensors presented here, the
boronic acid sensing moiety governs the sensor response to glucose. The sensors respond to
glucose by a competitive binding interaction between boronic acids and diols on either alizarin or
glucose. In the absence of glucose, the boronic acid binds to the diol on alizarin, statically
quenching its fluorescence. As local glucose concentrations increase, those molecules displace the
alizarin, allowing it to fluoresce.
We derived phenylboronic acids containing fluoro- and carboxyl- groups that withdraw electrons
in order to improve sensor response compared to octylboronic acid, which was used previously.1
Comparing boronic acid used in this paper to cotylboronic acid, the fluorescence was enhanced by
10%. In addition to acting as an electron-withdrawing group, carboxyls provide a site for the alkyl
chain additions performed herein and other chemical modifications.
71
The initial screen for glucose-responsiveness showed that macrosensors with 4-carboxy-3-
fluorophenyl boronic acid (BA1) increased fluorescence 13% from baseline in response to 100
mM glucose (Figure 3.2). This compound’s reactivity derives from having both fluoro- and
carboxyl groups withdrawing electrons from the boronic acid group, however this increases the
compound’s polarity. Consequently, BA1 readily leached from the hydrophobic sensor platform
over time (Figure 3.3), which leads to signal degradation and loss of sensitivity to glucose. We
then produced a new set of boronic acid molecules with varying polarities by systematically
converting the carboxyl group into esters with various alkyl chain lengths to find responsive and
stable sensors.
Adding a methyl ester to BA1 produced BA2a, which leached out of the macrosensors
significantly less than BA1, and longer alkyl chains (BA2b &BA2c) produced no significant
reduction in leaching compared to the methyl ester (Figure 3.3). Improvement in stability when
replacing the carboxylate ligand to an ester suggests that leaching of boronic acid may play in an
important role. Increasing the alkyl length decreased the resulting boronic acid’s reactivity; the
magnitude of macrosensor responses to glucose when formulated with BA2a, BA2b, and BA2c
were all less than compared to macrosensors made with BA1. Macrosensors with BA2a were still
relatively sensitive at physiological pH, exhibiting a 10% increase in fluorescence in response to
100 mM glucose. By contrast, macrosensors made with BA2b and BA2c only increased by 3%
and less than 1%, respectively (Figure 3.2).
The nanosensors’ competitive binding mechanism depends on the boronic acid diffusing within
the hydrophobic matrix and interacting with glucose molecules at the sensor-environment interface.
The result that longer alkyl chains reduced the magnitude of sensor responses suggests that long
alkyl chains inhibited boronic acid diffusion within the polymer matrix.
72
Figure 3.2. Response of glucose-sensitive macrosensors containing functionalized boronic
acids with increasing length of alkyl chains. The macrosensors contain Boronic Acids 1 (Ncontrol
= 7, Nglucose = 8), 2a (Ncontrol = 7, Nglucose = 7), 2b (Ncontrol = 7, Nglucose = 7), or 2c (Ncontrol = 8, Nglucose
= 8). Macrosensors were exposed to either PBS as a control or 100 mM glucose in PBS for 60
minutes. The percent change in fluorescence response was calculated as the average normalized
difference between the control and glucose groups. Error bars were calculated using error
propagation.
73
Figure 3.3. Fluorescence decay of macrosensors with different boronic acids. The
macrosensors contained Boronic Acids 1 (N = 7), 2a (N = 7), 2b (N = 7) or 2c (N = 8) and were
exposed to PBS for 60 minutes. Fluorescence intensities were normalized to time 0 and error bars
represent standard deviations.
74
While leaching is much less problematic for those derivatives such as BA2c, the increased
hydrophobicity may impart too high of an affinity to polymer matrix, causing sluggish diffusion
and small sensor responses.
Analysis of the calibration curve displays 10 fold change between 30 mM and 100 mM glucose
(Figure 3.4). To examine selectivity of the formulation as compared to other sugars fructose was
tested in vitro. The signal intensity of the 1 mM fructose solution was about half that of the 100
mM glucose after 1 h of incubation (Figure 3.5). We chose to compare 1 mM fructose because it
is present in the body at a concentration of about 8 µM30, so we tested it at 100-fold excess, as we
did with the glucose. Considering the difference in the comparative concentrations of these sugars
in the body, the interference is at an acceptable, if not optimal, level.
Our previous glucose-sensitive nanosensors included octylboronic acid, a hydrophobic aliphatic
derivative, as the sensing moiety,1,31 because it was stable in the hydrophobic nanosensor core.
14,32 Despite its stability, nanosensors with octylboronic acid were not sufficiently sensitive to
glucose. From studies on optodes, we discovered that 4-carboxyl-3-fluoroboronic acid 1 and its
derivatives are more sensitive to glucose due to their fluoro- and carboxyl groups. With the results
showing that BA2a leaches significantly less than BA1 and is must more responsive to glucose
than BA2b and BA2c, BA2a was selected as the lead candidate for nano-scale sensor fabrication.
Glucose-Sensitive Nanofibers. In addition to improvements in nanosensor sensitivity, nanosensor
systems need new design strategies for increasing residency time at the implantation site, ideally
with minimally-invasive delivery methods. Glucose nanosensors with BA2a were electrospun to
produce nanosensors with nanofiber architectures, requiring a plasticizer content of 10%.
For comparison, spherical nanosensors were also made using the fabrication method described in
the Materials and Methods section.
75
Figure 3.4. Response of macrosensors against different concentrations of glucose. The
macrosensors contained Boronic Acids 1a were exposed to 0 mM(N = 4), 30 mM (N = 4), 50 mM
(N = 4), 80 mM (N = 4) or 100 mM (N = 4) glucose solution in PBS for 60 minutes. The percent
change in fluorescence was calculated as the average normalized difference between the control
and glucose groups. Error bars were calculated using error propagation.
76
Figure 3.5. Comparison of sensor response to two sugars, glucose and fructose. Macrosensors
containing Boronic Acid 1a were exposed to PBS as a control ( N = 4), 100 mM glucose (N = 4)
or 1 mM fructose (N = 4) solution in PBS a for 60 minutes. The percent change in fluorescence
response was calculated as the average normalized difference between the control and glucose
groups. Error bars were calculated using error propagation.
77
Electrospinning optodes with 70 – 90 kDa PCL successfully produced continuous polymer
nanofibers, as confirmed with SEM images for high resolution fiber measurements and with
confocal images to show homogenous fluorescence from the alizarin within the fibers (Figure 3.6).
Measurements from the SEM images indicate that fiber diameters were 374 ± 142 nm and were
continuous without beading or wetting. Optode-based sensors are typically highly plasticized to
aid the mobility of sensor components and analytes within the sensor.33 Nanofibers that were
electrospun with PCL and 30% or 60% plasticizer increased the glucose-sensitivity by 6%, as
expected. However, even the 30% plasticized scaffolds showed signs of electrospinning instability
with discontinuous fibers and areas of pools of plasticizer (data not shown). Therefore, in order
to maintain the nanofibrous structure, we used 10% plasticizer content at the trade-off of sensor
response.
Glucose-sensitive nanofibers with BA2a, 2b and 2c responded 2% less, 2% more and 1% less than
their macrosensor counterparts respectively. Boronic acids with longer alkyl chains decreased the
sensitivity to glucose. To test the electrospun nanosensor response times, fluorescence intensity
was monitored over one hour after placing scaffolds in 100 mM glucose in PBS. Sensors
containing BA2b reached 95% of their maximum response within 12 minutes, but sensors
containing BA2a did not level off within an hour (Figure 3.7). The slow response times are likely
due to the low plasticizer content as well as the static flow conditions for the experimental
configuration. Low plasticizer content would restrict components from diffusing to the sensor-
environment interface. An experiment conducted in a flow cell would have enhanced the rate of
solution diffusion throughout the porous scaffold and decreased the response time. Despite these
slow response times, it is important to note that physiologic glucose levels change over the course
78
of tens of minutes,34 meaning that the BA2b formulation in nanofiber form responds sufficiently
fast to capture these changes.7
Higher molecular weight PCL or other polymers can support higher plasticizer percentage; for
example, electrospun nanofibers fabricated with ethyl cellulose were able to support up to 40%
plasticizer.23 Such strategies offer additional ways to improve the sensitivity and response times
of future nanosensor designs.
In Vivo Residency time Studies. In previous in vivo studies, nanoparticle-based sensors diffused
away from the implantation site within one hour. To show that nanofiber nanosensors improve
residency times at the implantation site, either spherical nanosensors1 or nanofiber nanosensors
were implanted subdermally (Figure 3.8) and their signal loss was directly compared to their in
vitro signal loss. Similar to previous experiments, the spherical nanosensors lost radiant efficiency
at the injection site significantly greater than the signal loss observed in vitro. In vitro signal loss
is attributed to boronic acid leaching from the hydrophobic core, and the difference between in
vivo and in vitro signal loss is attributed to nanosensor diffusion away from the implantation site.
By contrast, nanofiber scaffolds exhibited very closely matched signal loss between the in vivo
and in vitro experiments after one hour, and they were nearly equal after three hours (Figure 3.9).
The spherical nanosensors experienced a ~30% difference in total radiant efficiency loss when
compared to the in vitro control, whereas the decay constants for nanofiber scaffolds differed only
by 6%. Several factors accelerated the signal loss for spherical nanosensors in vivo compared to in
vitro, most notably sensor diffusion, cellular uptake, and the potential for facilitated transport of
components (either alizarin or boronic acid) out of the nanosensors due to amphiphilic serum
components in the in vivo environment.
79
Figure 3.6. Electrospun glucose-sensitive scaffolds. (A) Confocal image, (B) SEM image and
(C) size distribution of glucose-sensitive nanofibers. The average fiber diameter was 374 ± 142
nm (n=49). The width of histogram columns represents 50 nm.
80
Figure 3.7. Response of glucose-sensitive nanofibers containing different functionalized
boronic acids. Glucose-sensitive nanofibers contained fluorinated boronic acid derivatives 2a
(Ncontrol = 6, Nglucose = 8), 2b (Ncontrol = 5, Nglucose = 7), and 2c (Ncontrol = 7, Nglucose = 8). Increasing
alkyl chain lengths on fluorinated boronic acid derivatives effected the response of glucose-
sensitive nanofibers. Nanofibers were exposed to either PBS as a control or 100 mM glucose in
PBS for 60 minutes. The percent change in fluorescence response was calculated as the average
normalized difference between the control and glucose groups. Error bars were calculated using
error propagation.
81
Figure 3.8. In vivo comparison of glucose-sensitive nanoparticles and nanofiber scaffolds.
Mice were injected with glucose-sensitive nanoparticles and nanofiber scaffolds along the back
and then imaged with a fluorescent small animal imager for one hour and then at 3 hours post-
injection. Shown here are the fluorescent images from one mouse over this time frame.
82
Figure 3.9 Fluorescence measurements of glucose-sensitive nanoparticles and nanofiber
scaffolds over time in vivo. The average normalized total radiant efficiency of glucose-sensitive
(A) nanoparticles and (B) nanofiber scaffolds both in vivo (○) and in vitro control (■) were plotted
over time. The normalized in vivo average for nanoparticles and nanofiber scaffolds was
calculated across 3 different mice with Nnanoparticles= 8 and Nnanofiber scaffolds= 6 injection spots.
Similarly, the normalized in vitro average was calculated from Nnanoparticles= 8 and Nnanofiber scaffolds=
7. Error bars represent standard deviations.
83
Since the in vivo residency time of the nanofiber scaffold compared to the nanoparticles was
increased almost to the levels observed with the nanofibers in vitro, we could conclude that the
new sensor geometry maintained sensor residency at the injection site and would allow for longer
monitoring times.
3.4 Summary
In this study, we developed optode-based glucose nanosensors that were more sensitive to glucose
and more stable at the site of in vivo implantation.20 The initial macrosensor screen showed that
electron-withdrawing groups on BA1 and its derivatives facilitated a response to glucose under
physiological conditions, which is a major improvement over previous hydrophobic boronic acid
derivatives. Using the most responsive hydrophobic boronic acid derivative, BA2a, nanosensors
were electrospun into nanofibers and the nanofiber format was significantly more stable in vivo
than spherical nanosensors. Future work will focus on further increasing sensitivity and stability
by red-shifting the reporters and adding a reference signal for quantitative measurements.
3.5 Acknowledgments
This work was supported by the National Institute of Health under award number 5RO1GM084366
and Northeastern University’s internal funding Tier 1 support. Additionally, we thank Chris
Skipwith for his help in obtaining SEM images, Roger Kautz for his help in obtaining NMR spectra,
and Ganesh Thakur for his help with flash column chromatography.
3.6 Appendix: Glucose nanosensors emebedded in an alginate hydrogel
Previously, our group incorporated glucose-sensitive nanoparticles into commercially available
hydrogels containing thiol-modified hyaluronate and gelatin.37 Acrylate was used to crosslink the
thiol groups in the polymer. Incorporating glucose nanosensors to this polymer prolonged the in
84
vivo residency time of the nanosensor from less than 1 hour to 3 hours but decreased its sensitivity
towards glucose as well. To improve the sensitivity, response time, and residency time of the
glucose nanosensor, we incorporated the optimized recognition moiety, BA2a, into a glucose-
sensitive nanoparticle and then embedded the nanoparticle into a matrix of alginate hydrogel this
time. Alginic acid is a polysaccharide containing guluronic acid and mannuronic acid monomers.
Calcium ions can crosslink the polymers by the chelation with the carboxylates on guluronic
groups forming a hydrogel structure in aqueous buffer.154 The alginate hydrogel has been used
extensively applied in oral especially enteric drug delivery for its biocompatible feature and
responsiveness to pH.155 We proposed that the hydrogel structure will protect embedded
nanosensors from diffusion and clearance at the injection site to improve the in vivo residency time.
On the other hand, the cavities created by un-crosslinked monomers and hydrophilic environment
within the hydrogel facilitate glucose molecules to approach the sensors.
The glucose-sensitive nanoparticle was fabricated as shown in the Material and methods section
in this chapter. BA2a was used to prepare the optode. We followed procedure reported by Dong
et al. to fabricate hydrogels.156 Specifically, 0.5 mL nanoparticles suspended in DI water was
mixed with 0.5 mL HEPES buffer containing 4% sodium alginate and 4% gelatin. For the study
conducted using plate reader, 50 µL mixture in a well in 96 well plate was treated with 50 µL 10%
CaCl2 in DI water and then incubated for 10 mins. After the gelation was completed, the liquid in
the well was removed using a pipette, and the gel was washed three times with PBS before 100 µL
PBS was added to each well. For the study conducted using IVIS, 200 µL mixture in a PetriDish
was treated with 200 µL 10% CaCl2 and then incubated for 10 mins. After the gelation was
completed, the liquid in the well was removed using a pipette, and the gel was washed three times
85
with PBS before 3 mL PBS was added to each dish. The gelatin was used to improve the mechanic
property of the gel.
We firstly tested the response of the embedded sensor to glucose. Hydrogels containing glucose-
sensitive nanoparticles were soaked in PBS for 2 hours before the test in a SpectraMax Gemini
EM plate reader. After 50 mM glucose was added to the solution, the relative difference between
the experiment group and control group increased by more than 10% in 30 mins (Figure 3.10).
The difference became distinguishable in first couple minutes. Comparing to glucose-sensitive
nanofiber, this design improved sensitivity and response time of the detection. We reason that the
hydrophilic cavities in the hydrogel matrix facilitate the glucose to approach the sensors leading
to the improvements.
We also examined the stability of the hydrogel-embedded nanosensor. Since the crosslink of the
alginate hydrogel depends on the chelation of Ca2+ by carboxylates, we propose that if the
nanoparticle can be coated with carboxylates which may participate in chelation, the stability of
the nanosensor within the matrix can be enhanced. In this study, we used two types of amphiphilic
PEG-lipids to coat nanoparticles: the DSPE-mPEG550 (with a methoxy group at the hydrophilic
end), and DSPE-PEG-carboxylate (with a carboxylate at the hydrophilic end). The study was
conducted using IVIS with a 465/30 excitation filter and 580/20 emission filter. The hydrogels
were soaked in PBS in PetriDish and measured at different time points. Fluorescence
measurements were analyzed by selecting a region of interest around each injection spot to obtain
the total radiant efficiency of the area. The background-subtracted total radiant efficiency from
each region of interest containing either hydrogels was measured at each time point and then
normalized to the total radiant efficiency at time 0. In first two hours, the fluorescence from
hydrogels containing both types of nanoparticles dropped.
86
Figure 3.10. Response of glucose nanosensor embedded in alginate hydrogel. Hydrogels
containing glucose-sensitive nanoparticles were exposed to either PBS as a control (N = 5) or 50
mM glucose (N = 6) in PBS for 30 minutes. The percent change in fluorescence response was
calculated as the average normalized difference between the control and glucose groups. Error
bars were calculated using error propagation.
87
The total radiant efficiency from the nanoparticle coated by DSPE-mPEG550 dropped by 12% and
the one coated by DSPE-PEG-carboxylate dropped by 7% (Figure 3.11A). In 8 days after the
gelation, the fluorescence from the nanoparticle using DSPE-mPEG550 dropped by 20%, while
the one coated by DSPE-PEG-carboxylate dropped by less than 10% (Figure 3.11B). This result
indicated that the coating of the nanoparticle played a role in the stability of nanosensors within
the hydrogel matrix. The possible chelation of Ca2+ by carboxylates on nanoparticles and alginates
may contribute to this change.
In future, more studies need to be carried out to further exploit this design of using alginate
hydrogels as a matrix. The ratio of nanoparticles, alginate, and gelatin can be trialed and screened.
On the other hand, the condition of crosslink, such as the concentration of Ca2+ and the time for
crosslink can also be optimized to yield a glucose nanosensor with improved sensitivity, response
time and physical strength suitable for in vivo application. Then, the hydrogel can be applied in
vivo for continuous detection of glucose
88
Figure 3.11. The stability of hydrogel in PBS. The alginate hydrogels containing DSPE-
mPEG550 coated (N = 3) and DSPE-PEG-Carboxylate coated (N = 3) nanosensors were exposed
to PBS. (A) The normalized decay of total radiant efficiency in first 2 hours after fabrication. (B)
The normalized dacay of total radiant efficiency in 8 days after fabrication. Error bars represent
S.D.
89
Chapter 4: Conclusion and Future Direction
The development of personalized medicine, health monitoring devices, and early-stage diagnosis
requires more accurate, precise, and fast biological and chemical nanosensors to detect
physiological substances in vivo. Selection of the recognition moiety plays a significant role in
achieving the desired sensitivity, dynamic range, selectivity, and reversibility of the nanosensor.
In this thesis, we have demonstrated two nanosensors using different recognition mechanisms: 1)
ACh-MRNS using enzymatic hydrolysis to recognize acetylcholine, and 2) glucose sensitive
nanofibers utilizing boronic acids to detect analytes. In both cases, we co-immobilized the
recognition moiety as well as a reporter to create local physical and chemical changes leading to
an alteration in detectable signal intensities. In vitro calibration demonstrated that the dynamic
range, limit of detection, sensitivity and selectivity of ACh-MRNS are suitable for measuring
acetylcholine in brain; Delivered to the brain of rats, the ACh-MRNS detected drug-induced
endogenously released acetylcholine. To improve the sensitivity of the glucose-sensitive nanofiber,
we screened purchased and newly synthesized boronic acids to optimize the binding affinity
between the recognition moiety and glucose. The nanofiber with the optimized boronic acid
responded to glucose at physiological pH with an improved in vivo residency time.
The recognition moieties used in this dissertation can be further modified to improve the
nanosensors for in vivo application. In the ACh-MRNS, BuChE can be replaced by
acetylcholinesterase (AChE) to achieve a better sensitivity since the AChE is 2.5 times more active
than the BuChE to catalyze the hydrolysis of acetylcholine. This modification needs to be paired
with a faster MRI pulse sequence to improve temporal resolution and increase the fidelity of
acetylcholine detection. On the other hand, boronic acids with a higher affinity to glucose can also
be explored. For example, cyano and nitro groups can substitute fluoride or carboxylate to render
90
a stronger electron withdrawing effect to increase the affinity between glucose and boronic acids.
Other than recognition moieties, nanoplatforms and reporters can also be optimized to improve the
in vivo use of nanosensors. Polyamindoamine (PAMAM) dendrimer can be used as the nano-
backbone for the ACh-MRNS: The small size of PAMAM allows the ACh-MRNS to diffuse into
the synapse where a higher concentration of acetylcholine is released and the synaptic transmission
takes place. To improve in vivo detection of glucose, we are still looking for new matrix material
such as a hydrogel to contain all sensing components as well as facilitate diffusion of glucose to
overcome kinetic barriers created within the matrix. Embedding nanoparticle-based sensors in
biocompatible hydrogel may be a route to meet both requirements of sensitivity and residency time.
In the future, different types of hydrogels made from peptides and organic polymers can be further
interrogated to find the best matrix.
Nanoscale biosensors have displayed the capability to detect physiological molecules with the
desired dynamic range, sensitivity, reversibility, and selectivity. However, to apply nanosensors
to in vivo measurement, more studies, including optimization of recognition moieties, are still an
ongoing challenge. The work demonstrated in this dissertation used two different types of
recognition moieties to detect two critical physiological molecules: acetylcholine and glucose. The
Ach-MRNS is the first nanosensor to selectively image and monitor acetylcholine in vivo. This
sensor can be used to uncover the neural activity in brain and mechanisms of acetylcholine-related
diseases. The glucose-sensitive nanofiber provides a candidate for continuously monitoring the
level of glucose in vivo. This design prolonged the residency time of the sensor making the long-
term self-monitoring by patients with diabetes possible. In the future, more sensitive recognition
moiety for in vivo application can be synthesized, screened and incorporated to improve the
nanosensors for in vivo application. Other components in the nanosensors, such as the
91
nanoplatform and reporter, can also be modified to meet the needs of the in vivo application. With
the groundwork complete, the nanosensors for acetylcholine, glucose and other physiological
molecules will pave a way for future biological and preclinical studies.
92
Reference
1. Rong, G.; Corrie, S. R.; Clark, H. A., In Vivo Biosensing: Progress and Perspectives. ACS
sensors 2017, 2 (3), 327-338.
2. Smith, B. R.; Gambhir, S. S., Nanomaterials for In Vivo Imaging. Chemical reviews 2017, 117
(3), 901-986.
3. Eckert, M. A.; Vu, P. Q.; Zhang, K.; Kang, D.; Ali, M. M.; Xu, C.; Zhao, W., Novel molecular
and nanosensors for in vivo sensing. Theranostics 2013, 3 (8), 583-94.
4. Tsien, R. Y., Fluorescent indicators of ion concentrations. Methods Cell Biol 1989, 30, 127-56.
5. Tsien, R. Y., Fluorescent probes of cell signaling. Annu Rev Neurosci 1989, 12, 227-53.
6. Borisov, S. M.; Wolfbeis, O. S., Optical Biosensors. Chem. Rev. 2008, 108, 423-461.
7. Wolfbeis, O. S., Fiber-optic Chemical Sensors and Biosensors. Analytical chemistry 2008, 80,
4269-4283.
8. Ruckh, T. T.; Clark, H. A., Implantable nanosensors: toward continuous physiologic monitoring.
Analytical chemistry 2014, 86 (3), 1314-23.
9. Soto, R. J.; Hall, J. R.; Brown, M. D.; Taylor, J. B.; Schoenfisch, M. H., In Vivo Chemical
Sensors: Role of Biocompatibility on Performance and Utility. Analytical chemistry 2017, 89 (1), 276-
299.
10. Lee, S. B.; Mitchell, D. T.; Trofin, L.; Nevanen, T. K.; Soderlund, H.; Martin, C. R., Antibody-
based bio-nanotube membranes for enantiomeric drug separations. Science 2002, 296 (5576), 2198-200.
11. Kim, G. Y.; Josephson, L.; Langer, R.; Cima, M. J., Magnetic Relaxation Switch Detection of
Human Chorionic Gonadotrophin. Bioconjugate Chem. 2007, 18, 2024-2028.
12. Daniel, K. D.; Kim, G. Y.; Vassiliou, C. C.; Galindo, M.; Guimaraes, A. R.; Weissleder, R.;
Charest, A.; Langer, R.; Cima, M. J., Implantable diagnostic device for cancer monitoring. Biosensors &
bioelectronics 2009, 24 (11), 3252-7.
93
13. Song, S.; Wang, L.; Li, J.; Fan, C.; Zhao, J., Aptamer-based biosensors. TrAC Trends in
Analytical Chemistry 2008, 27 (2), 108-117.
14. Acquah, C.; Danquah, M. K.; Yon, J. L.; Sidhu, A.; Ongkudon, C. M., A review on immobilised
aptamers for high throughput biomolecular detection and screening. Analytica chimica acta 2015, 888,
10-8.
15. Yi, M.; Yang, S.; Peng, Z.; Liu, C.; Li, J.; Zhong, W.; Yang, R.; Tan, W., Two-photon graphene
oxide/aptamer nanosensing conjugate for in vitro or in vivo molecular probing. Analytical chemistry
2014, 86 (7), 3548-54.
16. Varghese, N.; Mogera, U.; Govindaraj, A.; Das, A.; Maiti, P. K.; Sood, A. K.; Rao, C. N.,
Binding of DNA nucleobases and nucleosides with graphene. Chemphyschem 2009, 10 (1), 206-10.
17. Park, J. S.; Na, H. K.; Min, D. H.; Kim, D. E., Desorption of single-stranded nucleic acids from
graphene oxide by disruption of hydrogen bonding. The Analyst 2013, 138 (6), 1745-9.
18. Witkowska Nery, E.; Kundys, M.; Jelen, P. S.; Jonsson-Niedziolka, M., Electrochemical Glucose
Sensing: Is There Still Room for Improvement? Analytical chemistry 2016, 88 (23), 11271-11282.
19. Steiner, M. S.; Duerkop, A.; Wolfbeis, O. S., Optical methods for sensing glucose. Chemical
Society reviews 2011, 40 (9), 4805-39.
20. Walsh, R.; Morales, J. M.; Skipwith, C. G.; Ruckh, T. T.; Clark, H. A., Enzyme-linked DNA
dendrimer nanosensors for acetylcholine. Scientific reports 2015, 5, 14832.
21. Cash, K. J.; Clark, H. A., Phosphorescent nanosensors for in vivo tracking of histamine levels.
Analytical chemistry 2013, 85 (13), 6312-8.
22. Lin, K. Y.; Kwong, G. A.; Warren, A. D.; Wood, D. K.; Bhatia, S. N., Nanoparticles That Sense
Thrombin Activity As Synthetic Urinary Biomarkers of Thrombosis. ACS Nano 2013, 7, 9001-9009.
23. Kwong, G. A.; von Maltzahn, G.; Murugappan, G.; Abudayyeh, O.; Mo, S.; Papayannopoulos, I.
A.; Sverdlov, D. Y.; Liu, S. B.; Warren, A. D.; Popov, Y.; Schuppan, D.; Bhatia, S. N., Mass-encoded
synthetic biomarkers for multiplexed urinary monitoring of disease. Nature biotechnology 2013, 31 (1),
63-70.
94
24. Kwong, G. A.; Dudani, J. S.; Carrodeguas, E.; Mazumdar, E. V.; Zekavat, S. M.; Bhatia, S. N.,
Mathematical framework for activity-based cancer biomarkers. Proceedings of the National Academy of
Sciences of the United States of America 2015, 112 (41), 12627-32.
25. Dudani, J. S.; Jain, P. K.; Kwong, G. A.; Stevens, K. R.; Bhatia, S. N., Photoactivated
Spatiotemporally-Responsive Nanosensors of in Vivo Protease Activity. ACS Nano 2015, 9, 11708-
11717.
26. Dudani, J. S.; Buss, C. G.; Akana, R. T. K.; Kwong, G. A.; Bhatia, S. N., Sustained-Release
Synthetic Biomarkers for Monitoring Thrombosis and Inflammation Using Point-of-Care Compatible
Readouts. Advanced Functional Materials 2016, 26 (17), 2919-2928.
27. Warren, A. D.; Gaylord, S. T.; Ngan, K. C.; Dumont Milutinovic, M.; Kwong, G. A.; Bhatia, S.
N.; Walt, D. R., Disease detection by ultrasensitive quantification of microdosed synthetic urinary
biomarkers. Journal of the American Chemical Society 2014, 136 (39), 13709-14.
28. Warren, A. D.; Kwong, G. A.; Wood, D. K.; Lin, K. Y.; Bhatia, S. N., Point-of-care diagnostics
for noncommunicable diseases using synthetic urinary biomarkers and paper microfluidics. Proceedings
of the National Academy of Sciences of the United States of America 2014, 111 (10), 3671-6.
29. Jin, P.; Wiraja, C.; Zhao, J.; Zhang, J.; Zheng, L.; Xu, C., Nitric Oxide Nanosensors for
Predicting the Development of Osteoarthritis in Rat Model. ACS applied materials & interfaces 2017, 9
(30), 25128-25137.
30. Zheng, X.; Tang, H.; Xie, C.; Zhang, J.; Wu, W.; Jiang, X., Tracking Cancer Metastasis In Vivo
by Using an Iridium-Based Hypoxia-Activated Optical Oxygen Nanosensor. Angewandte Chemie 2015,
54 (28), 8094-9.
31. Dubach, J. M.; Balaconis, M. K.; Clark, H. A., Fluorescent nanoparticles for the measurement of
ion concentration in biological systems. Journal of visualized experiments : JoVE 2011, (53).
32. Dubach, J. M.; Lim, E.; Zhang, N.; Francis, K. P.; Clark, H., In vivo sodium concentration
continuously monitored with fluorescent sensors. Integrative biology : quantitative biosciences from nano
to macro 2011, 3 (2), 142-8.
95
33. Cash, K. J.; Li, C.; Xia, J.; Wang, L. V.; Clark, H. A., Optical Drug Monitoring: Photoacoustic
Imaging of Nanosensors to Monitor Therapeutic Lithium in Vivo. ACS Nano 2015, 9, 1692-1698.
34. Cash, K. J.; Clark, H. A., In vivo histamine optical nanosensors. Sensors 2012, 12 (9), 11922-32.
35. Billingsley, K.; Balaconis, M. K.; Dubach, J. M.; Zhang, N.; Lim, E.; Francis, K. P.; Clark, H. A.,
Fluorescent nano-optodes for glucose detection. Analytical chemistry 2010, 82 (9), 3707-13.
36. Balaconis, M. K.; Billingsley, K.; Dubach, M. J.; Cash, K. J.; Clark, H. A., The design and
development of fluorescent nano-optodes for in vivo glucose monitoring. J Diabetes Sci Technol 2011, 5
(1), 68-75.
37. Balaconis, M. K.; Clark, H. A., Gel Encapsulation of Glucose Nanosensors for Prolonged In Vivo
Lifetime. Journal of Diabetes Science and Technology 2013, 7, 53-61.
38. Gundelfinger, E. D.; Kessels, M. M.; Qualmann, B., Temporal and spatial coordination of
exocytosis and endocytosis. Nature reviews. Molecular cell biology 2003, 4 (2), 127-39.
39. Sarter, M.; Parikh, V.; Howe, W. M., Phasic acetylcholine release and the volume transmission
hypothesis: time to move on. Nature reviews. Neuroscience 2009, 10 (5), 383-90.
40. Katz, B.; Miledi, R., The Measurement of Synaptic Delay, and the Time Course of Acetylcholine
Release at the Neuromuscular Junction. Proc R Soc Lond B Biol Sci 1965, 161, 483-95.
41. Barrett, K. E.; Barman, S. M.; Boitano, S.; Brooks, H., Ganong's review of medical physiology.
24 ed.; McGraw-Hill Professional Publishing: New York, NY, 2012.
42. McGehee, D. S.; Role, L. W., Physiological diversity of nicotinic acetylcholine receptors
expressed by vertebrate neurons. Annu Rev Physiol 1995, 57, 521-46.
43. Caulfield, M. P.; Birdsall, N. J., International Union of Pharmacology. XVII. Classification of
muscarinic acetylcholine receptors. Pharmacol Rev 1998, 50 (2), 279-90.
44. Soreq, H.; Seidman, S., Acetylcholinesterase--new roles for an old actor. Nature reviews.
Neuroscience 2001, 2 (4), 294-302.
96
45. van der Zee, E. A.; Luiten, P. G., Muscarinic acetylcholine receptors in the hippocampus,
neocortex and amygdala: a review of immunocytochemical localization in relation to learning and
memory. Prog Neurobiol 1999, 58 (5), 409-71.
46. Blokland, A., Acetylcholine: a neurotransmitter for learning and memory? Brain Res Brain Res
Rev 1995, 21 (3), 285-300.
47. Hagan, J. J.; Morris, R. G. M., The cholinergic hypothesis of memory: a review of animal
experiments. In Handbook of Physcopharmacology, Iversen, L. L.; Iversen, S. D.; Snyder, S. H., Eds.
Plenum Press: New York, NY, 1988; pp 237-305.
48. Nilsson, O. G.; Leanza, G.; Rosenblad, C.; Lappi, D. A.; Wiley, R. G.; Bjorklund, A., Spatial
learning impairments in rats with selective immunolesion of the forebrain cholinergic system.
Neuroreport 1992, 3 (11), 1005-8.
49. Muir, J. L., Acetylcholine, aging, and Alzheimer's disease. Pharmacol Biochem Behav 1997, 56
(4), 687-96.
50. Alzheimer'sAssociation, 2018 Alzheimer's Disease Facts and Figures. Alzheimer's & Dementia:
The Journal of the Alzheimer's Association 2018, 14 (3), 367-429.
51. Nitsch, R. M.; Slack, B. E.; Wurtman, R. J.; Growdon, J. H., Release of Alzheimer amyloid
precursor derivatives stimulated by activation of muscarinic acetylcholine receptors. Science 1992, 258
(5080), 304-7.
52. Lombardo, S.; Maskos, U., Role of the nicotinic acetylcholine receptor in Alzheimer's disease
pathology and treatment. Neuropharmacology 2015, 96 (Pt B), 255-62.
53. McKhann, G.; Drachman, D.; Folstein, M.; Katzman, R.; Price, D.; Stadlan, E. M., Clinical
diagnosis of Alzheimer's disease: report of the NINCDS-ADRDA Work Group under the auspices of
Department of Health and Human Services Task Force on Alzheimer's Disease. Neurology 1984, 34 (7),
939-44.
97
54. Giuliano, C.; Parikh, V.; Ward, J. R.; Chiamulera, C.; Sarter, M., Increases in cholinergic
neurotransmission measured by using choline-sensitive microelectrodes: Enhanced detection by
hydrolysis of acetylcholine on recording site? Neurochemistry International 2008, 52, 1343-1350.
55. Parikh, V.; Kozak, R.; Martinez, V.; Sarter, M., Prefrontal acetylcholine release controls cue
detection on multiple timescales. Neuron 2007, 56 (1), 141-54.
56. Vinson, P. N.; Justice, J. B., Jr., Effect of neostigmine on concentration and extraction fraction of
acetylcholine using quantitative microdialysis. Journal of neuroscience methods 1997, 73 (1), 61-7.
57. Hedou, G.; Homberg, J.; Martin, S.; Wirth, K.; Feldon, J.; Heidbreder, C. A., Effect of
amphetamine on extracellular acetylcholine and monoamine levels in subterritories of the rat medial
prefrontal cortex. Eur J Pharmacol 2000, 390 (1-2), 127-36.
58. CDC, At a Glance 2016 Diabetes Working to Reverse the US Epidemic. 2016.
59. Wang, L.; Gao, P.; Zhang, M.; Huang, Z.; Zhang, D.; Deng, Q.; Li, Y.; Zhao, Z.; Qin, X.; Jin, D.;
Zhou, M.; Tang, X.; Hu, Y.; Wang, L., Prevalence and Ethnic Pattern of Diabetes and Prediabetes in
China in 2013. JAMA 2017, 317 (24), 2515-2523.
60. Hoeks, L. B.; Greven, W. L.; de Valk, H. W., Real-time continuous glucose monitoring system
for treatment of diabetes: a systematic review. Diabet Med 2011, 28 (4), 386-94.
61. Gandhi, G. Y.; Kovalaske, M.; Kudva, Y.; Walsh, K.; Elamin, M. B.; Beers, M.; Coyle, C.;
Goalen, M.; Murad, M. S.; Erwin, P. J.; Corpus, J.; Montori, V. M.; Murad, M. H., Efficacy of continuous
glucose monitoring in improving glycemic control and reducing hypoglycemia: a systematic review and
meta-analysis of randomized trials. J Diabetes Sci Technol 2011, 5 (4), 952-65.
62. Alexeev, V. L.; Das, S.; Finegold, D. N.; Asher, S. A., Photonic crystal glucose-sensing material
for noninvasive monitoring of glucose in tear fluid. Clin Chem 2004, 50 (12), 2353-60.
63. Gao, W.; Emaminejad, S.; Nyein, H. Y. Y.; Challa, S.; Chen, K.; Peck, A.; Fahad, H. M.; Ota, H.;
Shiraki, H.; Kiriya, D.; Lien, D. H.; Brooks, G. A.; Davis, R. W.; Javey, A., Fully integrated wearable
sensor arrays for multiplexed in situ perspiration analysis. Nature 2016, 529 (7587), 509-514.
98
64. McNichols, R. J.; Cote, G. L., Optical glucose sensing in biological fluids: an overview. J Biomed
Opt 2000, 5 (1), 5-16.
65. Ekhlaspour, L.; Mondesir, D.; Lautsch, N.; Balliro, C.; Hillard, M.; Magyar, K.; Radocchia, L.
G.; Esmaeili, A.; Sinha, M.; Russell, S. J., Comparative Accuracy of 17 Point-of-Care Glucose Meters. J
Diabetes Sci Technol 2017, 11 (3), 558-566.
66. NIDDK Continuous Glucose Monitoring. https://www.niddk.nih.gov/health-
information/diabetes/overview/managing-diabetes/continuous-glucose-monitoring.
67. Young, S. H.; Poo, M. M., Spontaneous release of transmitter from growth cones of embryonic
neurones. Nature 1983, 305 (5935), 634-7.
68. Nguyen, Q. T.; Schroeder, L. F.; Mank, M.; Muller, A.; Taylor, P.; Griesbeck, O.; Kleinfeld, D.,
An in vivo biosensor for neurotransmitter release and in situ receptor activity. Nature neuroscience 2010,
13 (1), 127-32.
69. Bartelle, B. B.; Barandov, A.; Jasanoff, A., Molecular fMRI. The Journal of neuroscience : the
official journal of the Society for Neuroscience 2016, 36 (15), 4139-48.
70. Shapiro, M. G.; Westmeyer, G. G.; Romero, P. A.; Szablowski, J. O.; Kuster, B.; Shah, A.; Otey,
C. R.; Langer, R.; Arnold, F. H.; Jasanoff, A., Directed evolution of a magnetic resonance imaging
contrast agent for noninvasive imaging of dopamine. Nature biotechnology 2010, 28 (3), 264-70.
71. Lee, T.; Cai, L. X.; Lelyveld, V. S.; Hai, A.; Jasanoff, A., Molecular-level functional magnetic
resonance imaging of dopaminergic signaling. Science 2014, 344 (6183), 533-5.
72. Brustad, E. M.; Lelyveld, V. S.; Snow, C. D.; Crook, N.; Jung, S. T.; Martinez, F. M.; Scholl, T.
J.; Jasanoff, A.; Arnold, F. H., Structure-guided directed evolution of highly selective p450-based
magnetic resonance imaging sensors for dopamine and serotonin. Journal of molecular biology 2012, 422
(2), 245-62.
73. Mishra, A.; Mishra, R.; Gottschalk, S.; Pal, R.; Sim, N.; Engelmann, J.; Goldberg, M.; Parker, D.,
Microscopic visualization of metabotropic glutamate receptors on the surface of living cells using
bifunctional magnetic resonance imaging probes. ACS chemical neuroscience 2014, 5 (2), 128-37.
99
74. Oukhatar, F.; Meme, S.; Meme, W.; Szeremeta, F.; Logothetis, N. K.; Angelovski, G.; Toth, E.,
MRI sensing of neurotransmitters with a crown ether appended Gd(3+) complex. ACS chemical
neuroscience 2015, 6 (2), 219-25.
75. Davis, K. A.; Nanga, R. P.; Das, S.; Chen, S. H.; Hadar, P. N.; Pollard, J. R.; Lucas, T. H.;
Shinohara, R. T.; Litt, B.; Hariharan, H.; Elliott, M. A.; Detre, J. A.; Reddy, R., Glutamate imaging
(GluCEST) lateralizes epileptic foci in nonlesional temporal lobe epilepsy. Sci Transl Med 2015, 7 (309),
309ra161.
76. Cai, K.; Haris, M.; Singh, A.; Kogan, F.; Greenberg, J. H.; Hariharan, H.; Detre, J. A.; Reddy, R.,
Magnetic resonance imaging of glutamate. Nature medicine 2012, 18 (2), 302-6.
77. Andreozzi, E.; Wang, P.; Valenzuela, A.; Tu, C.; Gorin, F.; Dhenain, M.; Louie, A., Size-stable
solid lipid nanoparticles loaded with Gd-DOTA for magnetic resonance imaging. Bioconjugate chemistry
2013, 24 (9), 1455-67.
78. Zhao, J.; Zhang, B.; Shen, S.; Chen, J.; Zhang, Q.; Jiang, X.; Pang, Z., CREKA peptide-
conjugated dendrimer nanoparticles for glioblastoma multiforme delivery. Journal of colloid and
interface science 2015, 450, 396-403.
79. Zhang, H. W.; Wang, L. Q.; Xiang, Q. F.; Zhong, Q.; Chen, L. M.; Xu, C. X.; Xiang, X. H.; Xu,
B.; Meng, F.; Wan, Y. Q.; Deng, D. Y., Specific lipase-responsive polymer-coated gadolinium
nanoparticles for MR imaging of early acute pancreatitis. Biomaterials 2014, 35 (1), 356-67.
80. Park, Y. I.; Kim, H. M.; Kim, J. H.; Moon, K. C.; Yoo, B.; Lee, K. T.; Lee, N.; Choi, Y.; Park,
W.; Ling, D.; Na, K.; Moon, W. K.; Choi, S. H.; Park, H. S.; Yoon, S. Y.; Suh, Y. D.; Lee, S. H.; Hyeon,
T., Theranostic probe based on lanthanide-doped nanoparticles for simultaneous in vivo dual-modal
imaging and photodynamic therapy. Advanced materials 2012, 24 (42), 5755-61.
81. Harris, K. D.; Thiele, A., Cortical state and attention. Nature reviews. Neuroscience 2011, 12 (9),
509-23.
82. Hasselmo, M. E.; Bower, J. M., Acetylcholine and memory. Trends in neurosciences 1993, 16
(6), 218-22.
100
83. Everitt, B. J.; Robbins, T. W., Central cholinergic systems and cognition. Annual review of
psychology 1997, 48, 649-84.
84. Lewis, D. A.; Lieberman, J. A., Catching up on schizophrenia: natural history and neurobiology.
Neuron 2000, 28 (2), 325-34.
85. Manji, H. K.; Drevets, W. C.; Charney, D. S., The cellular neurobiology of depression. Nature
medicine 2001, 7 (5), 541-7.
86. Mattson, M. P., Pathways towards and away from Alzheimer's disease. Nature 2004, 430 (7000),
631-9.
87. Bolan, P. J.; Meisamy, S.; Baker, E. H.; Lin, J.; Emory, T.; Nelson, M.; Everson, L. I.; Yee, D.;
Garwood, M., In vivo quantification of choline compounds in the breast with 1H MR spectroscopy.
Magnetic resonance in medicine 2003, 50 (6), 1134-43.
88. Baik, H. M.; Su, M. Y.; Yu, H.; Mehta, R.; Nalcioglu, O., Quantification of choline-containing
compounds in malignant breast tumors by 1H MR spectroscopy using water as an internal reference at 1.5
T. Magma 2006, 19 (2), 96-104.
89. Mitchell, K. M., Acetylcholine and choline amperometric enzyme sensors characterized in vitro
and in vivo. Analytical chemistry 2004, 76 (4), 1098-106.
90. Conde, F.; Maire-Lepoivre, E.; Audinat, E.; Crepel, F., Afferent connections of the medial frontal
cortex of the rat. II. Cortical and subcortical afferents. The Journal of comparative neurology 1995, 352
(4), 567-93.
91. Woods, M.; Kiefer, G. E.; Bott, S.; Castillo-Muzquiz, A.; Eshelbrenner, C.; Michaudet, L.;
McMillan, K.; Mudigunda, S. D.; Ogrin, D.; Tircso, G.; Zhang, S.; Zhao, P.; Sherry, A. D., Synthesis,
relaxometric and photophysical properties of a new pH-responsive MRI contrast agent: the effect of other
ligating groups on dissociation of a p-nitrophenolic pendant arm. Journal of the American Chemical
Society 2004, 126 (30), 9248-56.
92. Schneider, C. A.; Rasband, W. S.; Eliceiri, K. W., NIH Image to ImageJ: 25 years of image
analysis. Nature methods 2012, 9 (7), 671-5.
101
93. Barge, A.; Cravotto, G.; Gianolio, E.; Fedeli, F., How to determine free Gd and free ligand in
solution of Gd chelates. A technical note. Contrast media & molecular imaging 2006, 1 (5), 184-8.
94. Yushkevich, P. A.; Piven, J.; Hazlett, H. C.; Smith, R. G.; Ho, S.; Gee, J. C.; Gerig, G., User-
guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and
reliability. NeuroImage 2006, 31 (3), 1116-28.
95. Ichikawa, J.; Dai, J.; O'Laughlin, I. A.; Fowler, W. L.; Meltzer, H. Y., Atypical, but not typical,
antipsychotic drugs increase cortical acetylcholine release without an effect in the nucleus accumbens or
striatum. Neuropsychopharmacology : official publication of the American College of
Neuropsychopharmacology 2002, 26 (3), 325-39.
96. Tellinghuisen, J., Statistical error propagation. J Phys Chem A 2001, 105 (15), 3917-3921.
97. Caravan, P.; Ellison, J. J.; McMurry, T. J.; Lauffer, R. B., Gadolinium(III) chelates as MRI
contrast agents: Structure, dynamics, and applications. Chemical reviews 1999, 99 (9), 2293-2352.
98. Cash, K. J.; Li, C.; Xia, J.; Wang, L. V.; Clark, H. A., Optical drug monitoring: photoacoustic
imaging of nanosensors to monitor therapeutic lithium in vivo. ACS Nano 2015, 9 (2), 1692-8.
99. Ruckh, T. T.; Skipwith, C. G.; Chang, W.; Senko, A. W.; Bulovic, V.; Anikeeva, P. O.; Clark, H.
A., Ion-switchable quantum dot Forster resonance energy transfer rates in ratiometric potassium sensors.
ACS Nano 2016, 10, 4020-4030.
100. Teng, T. L.; Harpst, J. A.; Lee, J. C.; Zinn, A.; Carlson, D. M., Composition and molecular
weights of butyrylcholinesterase from horse serum. Archives of biochemistry and biophysics 1976, 176
(1), 71-81.
101. Lee, S. M.; Song, Y.; Hong, B. J.; MacRenaris, K. W.; Mastarone, D. J.; O'Halloran, T. V.;
Meade, T. J.; Nguyen, S. T., Modular polymer-caged nanobins as a theranostic platform with enhanced
magnetic resonance relaxivity and pH-responsive drug release. Angewandte Chemie 2010, 49 (51), 9960-
4.
102
102. Grogna, M.; Cloots, R.; Luxen, A.; Jerome, C.; Desreux, J. F.; Detrembleur, C., Design and
synthesis of novel DOTA(Gd3+)-polymer conjugates as potential MRI contrast agents. J Mater Chem
2011, 21 (34), 12917-12926.
103. Caravan, P.; Farrar, C. T.; Frullano, L.; Uppal, R., Influence of molecular parameters and
increasing magnetic field strength on relaxivity of gadolinium- and manganese-based T1 contrast agents.
Contrast media & molecular imaging 2009, 4 (2), 89-100.
104. Chesler, M., Regulation and modulation of pH in the brain. Physiological reviews 2003, 83 (4),
1183-221.
105. Augustinsson, K. L., Butyryl- and propionylcholinesterases and related types of eserine-sensitive
esterases. In The enzymes, Second ed.; Boyer, P. D.; Lardy, H.; Myrback, K., Eds. Academic Press: New
York and London, 1960; Vol. 4, pp 521-540.
106. Hoa, D. T.; Kumar, T. N. S.; Punekar, N. S.; Srinivasa, R. S.; Lal, R.; Contractor, A. Q.,
Biosensor based on conducting polymers. Analytical chemistry 1992, 64 (21), 2645-2646.
107. Nishizawa, M.; Matsue, T.; Uchida, I., Penicillin sensor based on a microarray electrode coated
with pH-responsive polypyrrole. Analytical chemistry 1992, 64 (21), 2642-4.
108. Luo, X. L.; Xu, J. J.; Zhao, W.; Chen, H. Y., A novel glucose ENFET based on the special
reactivity of MnO2 nanoparticles. Biosensors & bioelectronics 2004, 19 (10), 1295-300.
109. Stuart, M. A.; Huck, W. T.; Genzer, J.; Muller, M.; Ober, C.; Stamm, M.; Sukhorukov, G. B.;
Szleifer, I.; Tsukruk, V. V.; Urban, M.; Winnik, F.; Zauscher, S.; Luzinov, I.; Minko, S., Emerging
applications of stimuli-responsive polymer materials. Nature materials 2010, 9 (2), 101-13.
110. Tagliazucchi, M.; Calvo, E. J.; Szleifer, I., Redox and acid-base coupling in ultrathin
polyelectrolyte films. Langmuir 2008, 24 (6), 2869-77.
111. Tagliazucchi, M.; Calvo, E. J.; Szleifer, I., Molecular Theory of Chemically Modified Electrodes
by Redox Polyelectrolytes under Equilibrium Conditions: Comparison with Experiment. Journal of
Physical Chemistry C 2008, 112, 458-471.
103
112. Ferris, C. F.; Kulkarni, P.; Sullivan, J. M., Jr.; Harder, J. A.; Messenger, T. L.; Febo, M., Pup
suckling is more rewarding than cocaine: evidence from functional magnetic resonance imaging and
three-dimensional computational analysis. The Journal of neuroscience : the official journal of the Society
for Neuroscience 2005, 25 (1), 149-56.
113. Paxinos, G.; Watson, C., The Rat Brain in stereotaxic coordinates. Compact sixth ed.; Academic
Press, Elsevier: 2009.
114. Ichikawa, J.; Li, Z.; Dai, J.; Meltzer, H. Y., Atypical antipsychotic drugs, quetiapine, iloperidone,
and melperone, preferentially increase dopamine and acetylcholine release in rat medial prefrontal cortex:
role of 5-HT1A receptor agonism. Brain research 2002, 956 (2), 349-57.
115. Harrison, J.; Bartlett, C. A.; Cowin, G.; Nicholls, P. K.; Evans, C. W.; Clemons, T. D.; Zdyrko,
B.; Luzinov, I. A.; Harvey, A. R.; Iyer, K. S.; Dunlop, S. A.; Fitzgerald, M., In vivo imaging and
biodistribution of multimodal polymeric nanoparticles delivered to the optic nerve. Small 2012, 8 (10),
1579-89.
116. Hutter, E.; Boridy, S.; Labrecque, S.; Lalancette-Hebert, M.; Kriz, J.; Winnik, F. M.; Maysinger,
D., Microglial response to gold nanoparticles. ACS Nano 2010, 4 (5), 2595-606.
117. Choi, J.; Zheng, Q.; Katz, H. E.; Guilarte, T. R., Silica-based nanoparticle uptake and cellular
response by primary microglia. Environ Health Perspect 2010, 118 (5), 589-95.
118. Giuliano, C.; Parikh, V.; Ward, J. R.; Chiamulera, C.; Sarter, M., Increases in cholinergic
neurotransmission measured by using choline-sensitive microelectrodes: enhanced detection by
hydrolysis of acetylcholine on recording sites? Neurochemistry international 2008, 52 (7), 1343-50.
119. Parikh, V.; Pomerleau, F.; Huettl, P.; Gerhardt, G. A.; Sarter, M.; Bruno, J. P., Rapid assessment
of in vivo cholinergic transmission by amperometric detection of changes in extracellular choline levels.
The European journal of neuroscience 2004, 20 (6), 1545-54.
120. Gao, Y.; Erokwu, B. O.; DeSantis, D. A.; Croniger, C. M.; Schur, R. M.; Lu, L.; Mariappuram, J.;
Dell, K. M.; Flask, C. A., Initial evaluation of hepatic T1 relaxation time as an imaging marker of liver
104
disease associated with autosomal recessive polycystic kidney disease (ARPKD). NMR in biomedicine
2016, 29 (1), 84-9.
121. Herrmann, K.; Erokwu, B. O.; Johansen, M. L.; Basilion, J. P.; Gulani, V.; Griswold, M. A.;
Flask, C. A.; Brady-Kalnay, S. M., Dynamic quantitative T1 mapping in orthotopic brain tumor
xenografts. Translational oncology 2016, 9 (2), 147-54.
122. Vykhodtseva, N.; McDannold, N.; Hynynen, K., Progress and problems in the application of
focused ultrasound for blood-brain barrier disruption. Ultrasonics 2008, 48 (4), 279-96.
123. Hynynen, K.; McDannold, N.; Sheikov, N. A.; Jolesz, F. A.; Vykhodtseva, N., Local and
reversible blood-brain barrier disruption by noninvasive focused ultrasound at frequencies suitable for
trans-skull sonications. NeuroImage 2005, 24 (1), 12-20.
124. Muldoon, L. L.; Sandor, M.; Pinkston, K. E.; Neuwelt, E. A., Imaging, distribution, and toxicity
of superparamagnetic iron oxide magnetic resonance nanoparticles in the rat brain and intracerebral
tumor. Neurosurgery 2005, 57 (4), 785-96; discussion 785-96.
125. Patel, T.; Zhou, J.; Piepmeier, J. M.; Saltzman, W. M., Polymeric nanoparticles for drug delivery
to the central nervous system. Adv Drug Deliv Rev 2012, 64 (7), 701-5.
126. Saraiva, C.; Praca, C.; Ferreira, R.; Santos, T.; Ferreira, L.; Bernardino, L., Nanoparticle-
mediated brain drug delivery: Overcoming blood-brain barrier to treat neurodegenerative diseases. J
Control Release 2016, 235, 34-47.
127. Darvesh, S.; Hopkins, D. A.; Geula, C., Neurobiology of butyrylcholinesterase. Nature reviews.
Neuroscience 2003, 4 (2), 131-8.
128. Bakker, E.; Buhlmann, P.; Pretsch, E., Carrier-Based Ion-Selective Electrodes and Bulk Optodes.
1. General Characteristics. Chem Rev 1997, 97 (8), 3083-3132.
129. Bakker, E.; Simon, W., Selectivity of ion-sensitive bulk optodes. Analytical Chemistry 1992, 64
(17), 1805-1812.
130. Buhlmann, P.; Pretsch, E.; Bakker, E., Carrier-Based Ion-Selective Electrodes and Bulk Optodes.
2. Ionophores for Potentiometric and Optical Sensors. Chem Rev 1998, 98 (4), 1593-1688.
105
131. Tohda, K.; Gratzl, M., A microscopic, continous, optical monitor for interstitial electrolytes and
glucose. Chemphyschem 2003, 4 (2), 155-160.
132. Tohda, K.; Gratzl, M., Micro-miniature autonomous optical sensor array for monitoring ions and
metabolites 1: Design, fabrication, and data analysis. Analytical Sciences 2006, 22 (3), 383-388.
133. Srivastava, R.; Jayant, R. D.; Chaudhary, A.; Mcshane, M. J., "Smart Tatoo" Glucose Biosensors
and Effects of Coencapsulated Anti-Inflammatory Agents. Journal of Diabetes Science and Technology
2011, 5 (1), 76.
134. Ruckh, T. T.; Clark, H. A. Implantable nanosensors: toward continuous physiologic monitoring
Analytical chemistry [Online], 2014, p. 1314-1323.
135. Loris, R.; Hamelryck, T.; Bouckaert, J.; Wyns, L., Legume lectin structure. Biochem. & Biophys.
Acta 1998, 1383, 9-36.
136. Springsteen, G.; Wang, B. H., A detailed examination of boronic acid-diol complexation.
Tetrahedron 2002, 58 (26), 5291-5300.
137. Yan, J.; Springsteen, G.; Deeter, S.; Wang, B. H., The relationship among pK(a),pH, and binding
constants in the interactions between boronic acids and diols - it is not as simple as it appears.
Tetrahedron 2004, 60 (49), 11205-11209.
138. Mulla, H. R.; Agard, N. J.; Basu, A., 3-Methoxycarbonyl-5-nitrophenyl boronic acid: high
affinity diol recognition at neutral pH. Bioorganic & Medicinal Chemistry Letters 2004, 14 (1), 25-27.
139. Das, S.; Alexeev, V. L.; Sharma, A. C.; Geib, S. J.; Asher, S. A., Synthesis and crystal structure
of 4-amino-3-fluorophenylboronic acid. Tetrahedron Letters 2003, 44 (42), 7719-7722.
140. Singhal, R. P.; Ramamurthy, B.; Govindraj, N.; Sarwar, Y., New Ligands for Boronate Affinity-
Chromatography - Synthesis and Properties. Journal of Chromatography 1991, 543 (1), 17-38.
141. Liu, X. C.; Scouten, W. H., New Ligands for Boronate Affinity-Chromatography Journal of
Chromatography A 1994, 687 (1), 61-69.
106
142. Alexeev, V. L.; Sharma, A. C.; Goponenko, A. V.; Das, S.; Lednev, I. K.; Wilcox, C. S.;
Finegold, D. N.; Asher, S. A., High ionic strength glucose-sensing photonic crystal. Analytical chemistry
2003, 75 (10), 2316-2323.
143. Yum, K.; Ahn, J. H.; McNicholas, T. P.; Barone, P. W.; Mu, B.; Kim, J. H.; Jain, R. M.; Strano,
M. S., Boronic Acid Library for Selective, Reversible Near-Infrared Fluorescence Quenching of
Surfactant Suspended Single-Walled Carbon Nanotubes in Response to Glucose. ACS Nano 2012, 6 (1),
819-830.
144. Balaconis, M. K.; Clark, H. A., Gel encapsulation of glucose nanosensors for prolonged in vivo
lifetime. J Diabetes Sci Technol 2013, 7 (1), 53-61.
145. Ozaydin-Ince, G.; Dubach, J. M.; Gleason, K. K.; Clark, H. A., Microworm optode sensors limit
particle diffusion to enable in vivo measurements. Proc. Natl. Acad. Sci. U. S. A. 2011.
146. Cash, K. J.; Clark, H. A., Nanosensors and nanomaterials for monitoring glucose in diabetes.
Trends in Molecular Medicine 2010, 16 (12), 584-593.
147. Kacmaz, S.; Ertekin, K.; Suslu, A.; Ozdemir, M.; Ergun, Y.; Celik, E.; Cocen, U., Emission based
sub-nanomolar silver sensing with electrospun nanofibers. Sensors and Actuators B-Chemical 2011, 153
(1), 205-213.
148. Kacmaz, S.; Ertekin, K.; Suslu, A.; Ergun, Y.; Celik, E.; Cocen, U., Sub-nanomolar sensing of
ionic mercury with polymeric electrospun nanofibers. Materials Chemistry and Physics 2012, 133 (1),
547-552.
149. Yang, Y. F.; Wang, H. M.; Su, K.; Long, Y. Y.; Peng, Z.; Li, N.; Liu, F., A facile and sensitive
fluorescent sensor using electrospun nanofibrous film for nitroaromatic explosive detection. Journal of
Materials Chemistry 2011, 21 (32), 11895-11900.
150. Manesh, K. M.; Santhosh, P.; Gopalan, A.; Lee, K.-P., Electrospun poly(vinylidene
fluoride)/poly(aminophenylboronic acid) composite nanofibrous membrane as a novel glucose sensor.
Analytical Biochemistry 2007, 360 (2), 189-195.
107
151. Zhou, C. S.; Shi, Y. L.; Ding, X. D.; Li, M.; Luo, J. J.; Lu, Z. Y.; Xiao, D., Development of a Fast
and Sensitive Glucose Biosensor Using Iridium Complex-Doped Electrospun Optical Fibrous Membrane.
Analytical chemistry 2013, 85 (2), 1171-1176.
152. Neises, B.; Steglich, W., Simple Method for the Esterification of Carboxylic Acids. Angewandte
Chemie International Edition in English 1978, 17 (7), 522-524.
153. Heo, Y. J.; Shibata, H.; Okitsu, T.; Kawanishi, T.; Takeuchi, S., Long-term in vivo glucose
monitoring using fluorescent hydrogel fibers. Proceedings of the National Academy of Sciences of the
United States of America 2011, 108 (33), 13399-13403.
154. Rowley, J. A.; Madlambayan, G.; Mooney, D. J., Alginate hydrogels as synthetic extracellular
matrix materials. Biomaterials 1999, 20 (1), 45-53.
155. George, M.; Abraham, T. E., Polyionic hydrocolloids for the intestinal delivery of protein drugs:
Alginate and chitosan - a review. J Control Release 2006, 114 (1), 1-14.
156. Dong, Z. F.; Wang, Q.; Du, Y. M., Alginate/gelatin blend films and their properties for drug
controlled release. J Membrane Sci 2006, 280 (1-2), 37-44.