The Anesthesia Machine and Breathing Systems

120
Trey Bates, MD Special Thanks to Judson Mehl The Anesthesia Machine and Breathing Systems

description

The Anesthesia Machine and Breathing Systems. Trey Bates, MD Special Thanks to Judson Mehl. A quick word on medical gas - . All those hoses: Oxygen Air Nitrous Vacuum WAGD Waste Anesthetic Gas Disposal. 2000 PSI (FULL). H (7000L). E (700L). Pressure Reduction Pathway. - PowerPoint PPT Presentation

Transcript of The Anesthesia Machine and Breathing Systems

Page 1: The Anesthesia Machine and Breathing Systems

Trey Bates, MDSpecial Thanks to Judson Mehl

The Anesthesia Machine and Breathing Systems

Page 2: The Anesthesia Machine and Breathing Systems

• All those hoses:• Oxygen• Air• Nitrous• Vacuum• WAGD

• Waste Anesthetic Gas Disposal

A quick word on medical gas -

2000 PSI(FULL)

H (7000L) E (700L)

Page 3: The Anesthesia Machine and Breathing Systems

• H-Cylinder 2000 psi

• Manifold 55 psi

• Hospital line55 psi

Pressure Reduction Pathway

O2 - 2000 psi, 625 L N2O - 750 psi, 1590 L Co2 - 838 piso, 1590 L Air - 1800 psi, 625 L

Page 4: The Anesthesia Machine and Breathing Systems

2000 psi !!!

45 psi

Oxygen Failure Protection Device

Flow of nitrous-oxide is dependent on oxygen pressure.

If oxygen pressure is lost then the other gases cannot flow past their regulators

Page 5: The Anesthesia Machine and Breathing Systems

• Liquid oxygen must be stored below its critical temperature of -119 C

• In oxygen tanks, the pressure falls in proportion to the remaining volume of oxygen• If a full E-cylinder at 2000 psi contains 700 L O2, then

a half full tank at 1000 psi contains ?• What about an H-cylinder at 1000 psi?• What about an E-cylinder at 500 psi?

Key points for ITE:

Page 6: The Anesthesia Machine and Breathing Systems

• ICU transport with an E-cylinder with 700 psi.

• Need to run a NRB at 10 lpm.

• Tulane elevator breaks down. How much sh*t are we in?

More math . . . for fun

Page 7: The Anesthesia Machine and Breathing Systems

• Nitrous is NOT an ideal gas. Thus it has several unique properties:

• Transition between liquid and gas states does not lead to huge increases in pressure

• It is easy to compress, so the cylinders hold a lot moregas.• Its critical temp is 36.5 C, so it doesn’t need refrigeration

Nitrous

Page 8: The Anesthesia Machine and Breathing Systems

• N2O is vaporized at the same rate it is utilized• The pressure in the tank never changes

• You don’t know what you’ve got til its gone (400 L/1600L = 25% remaining))

• The only way to tell how much N2O is left, is to measure the tank. • Consult the tare weight on the bottle• I have never been asked to determine how much N2O was left

based on weight on an ITE.

More on N2O

Page 9: The Anesthesia Machine and Breathing Systems

• This stuff matters because:• Oxygen is pretty important• Agent delivery is pretty important• Getting rid of CO2 is pretty important

• And these ALWAYS show up on the ITE. ALWAYS !!

Breathing Systems

Page 10: The Anesthesia Machine and Breathing Systems

• Is a . . . • Cardiac Depressant

• And it . . .• Increases CBF• Increases bleeding• Causes acidosis which . . .

• Shifts the Carboxy-Hgb curve• Shifts Ca2+ and K+ out of the cell

CO2

Page 11: The Anesthesia Machine and Breathing Systems

Anesthesia at Tulane

Page 12: The Anesthesia Machine and Breathing Systems

• Gas delivery under a drape

• Serious CO2 accumulation without high gas flow

• Lets be honest, everybody in the room is breathing this stuff

Insufflation anesthesia

Page 13: The Anesthesia Machine and Breathing Systems

Open Drop Anestheisa

Schimmelbusch mask

Page 14: The Anesthesia Machine and Breathing Systems

• Hose serves as an open-ended reservoir

• Addition of oxygen possible• 1 lpm 30-40%• 4 lpm 60-80%

• Simple• Portable• No scavenging

Draw-over anesthesia

Page 15: The Anesthesia Machine and Breathing Systems

• Ingredients:• Breathing tube• Fresh gas inlet• APL valve• Reservoir bag

The Mapleson Circuit

The only real difference is the order in which the ingredients occur

Page 16: The Anesthesia Machine and Breathing Systems

• Important points• There are NO one-way valves • There is no CO2 absorber

• Some rebreathing is prevented by venting through the APL before the next inspiration

Mapleson Circuits

Page 17: The Anesthesia Machine and Breathing Systems

Basic Mapleson A

During spontAneous ventilation, the Mapleson A is most efficient. That long breathing tube full of fresh gas is a great reservoir for the patients next breath.

Page 18: The Anesthesia Machine and Breathing Systems

• Giving positive pressure is going to require me to partially close that APL valve. When I ventilate, half of my FGF is going to exit the partially open APL valve.

• During exhalation, all that exhaled air is going to fill the breathing tube because the APL is now closed and the only way it is going to vent is if the gas flows are really high.

Why?

Page 19: The Anesthesia Machine and Breathing Systems

• The FGF is happening right at the patient’s face.

• Now, watch this . .

Mapleson D

Page 20: The Anesthesia Machine and Breathing Systems

When I positive pressure ventilate, I close the APL and use the old air in the reservoir to generate the force to blow the fresh air into the patient. Anything I lose out of the APL will be old air.

Between ventilations the new air is pushing the old air out of the APL and away from the patient.

Page 21: The Anesthesia Machine and Breathing Systems

• The Bain circuit deposits the FGF in the same place as Mapleson D, but it traveled through the warm, exhaled air on the way in, so the FGF is warmed.

Bain

Page 22: The Anesthesia Machine and Breathing Systems

• If not, and you probably wont on the day of the ITE, then check out this aswesome memory aid. Its pretty complex:

• Ventilation is most efficient in a

• Mapleson A during spontAneous ventilation• There is no D in spontaneous

• Mapleson D during controleD ventilation• There is no A in controled

Get it now?

Page 23: The Anesthesia Machine and Breathing Systems

• Lose all the heat and humidity• High FGF to prevent rebreathing• All that agent is ventilated out to the room

The downfalls of the Mapleson

Page 24: The Anesthesia Machine and Breathing Systems

• And then we added:• CO2 absorbers• Unidirectional valves• Scavenging

• And voila, we have the CIRCLE SYSTEM !!

So, science happened

Page 25: The Anesthesia Machine and Breathing Systems

• How they work:• So why is it bad that the CO2 absorber “dries out?”

• Well, here is why:

• CO2 + H2O → H2CO3 (this is carbonic acid)

• Then the hydroxide salts in the CO2 absorber do this:• H2CO3 + 2NaOH → Na2CO3 + 2H2O + heat (this is why they get

warm)

• Then all that Na2CO3 (sodium hydroxide) produced in the first reaction does this:• Na2CO3 + Ca(OH)2 → CaCo3 + 2NaOH (We just regenerated our starting

reagent)

CO2 absorbers

Page 26: The Anesthesia Machine and Breathing Systems

• As the absorbent is used up, it becomes more acidic.• That purple color change is a pH indicator

• When 50-70% has changed color, its time to change the absorber.

CO2 absorbers

Granule size is a trade off:

Larger granules minimize resistance to airflow

Smaller granules maximize surface area for more absorption

Page 27: The Anesthesia Machine and Breathing Systems

• Inspiratory• Expiratory• Valve

incompetence is usually due to unseated or warped disc

• Note what is in the reservoir bag

And what about these unidirectional valves?

Page 28: The Anesthesia Machine and Breathing Systems
Page 29: The Anesthesia Machine and Breathing Systems

In a closed scavenging system, what happens to the reservoir bag during expiration and inspiration?

What does it mean with the opposite happens?

Page 30: The Anesthesia Machine and Breathing Systems

The reservoir bag expands during expiration and deflates during inspiration. During inspiration in MV, the ventilator pressure relief valve closes, directing ventilator bellows into patient breathing circuit.

If the PRV is incompetent, there will be a direct communication between breathing circuit and scavenging circuitand the reservoir bag would inflate during inspiration.

Page 31: The Anesthesia Machine and Breathing Systems

• A size E compressed-gas cylinder completely filled with N2O contains how many litres?

• A. 1160• B. 1470• C.1590• D. 1640• E. 1750

A few questions

Page 32: The Anesthesia Machine and Breathing Systems

• C• Size E compressed gas cylinders completely filled

contain 1590 L gas

Answer:

Page 33: The Anesthesia Machine and Breathing Systems

• The pressure gauge on a size E compressed-gas cylinder containing O2 reads 1600 psi. How long could O2 be delivered from this cylinder at 2 LPM?

• A. 90 min• B. 140 min• C. 280 min• D. 320 min• E. Cannot be calculated

Question

Page 34: The Anesthesia Machine and Breathing Systems

• C

Answer:

Page 35: The Anesthesia Machine and Breathing Systems

• If the anesthesia machine is discovered Monday morning having run with 5L/min of O2 all weekend, the most reasonable course of action to take before administering the next anesthetic would be:

• A. Turn the machine off for 30 min• B. Place a humidifier in the expiratory limb• C. Avoid the use of Sevoflurane• D. Change the CO2 absorbent• E. Use N2O for the first hour of the case

Question

Page 36: The Anesthesia Machine and Breathing Systems

• D – of course, but why change it if its not purple?

Answer

Page 37: The Anesthesia Machine and Breathing Systems

• A mechanically ventilated patient is transported from the OR to the ICU using a portable ventilator that consumes 2L/min of O2 to run the ventilator itself. The patient gets 100% O2 and tidal volumes of 500 ml at a rate of 10/min. You have an E-cylinder with 2000 psi. The vent will shut off below 200 psi. How long do you have?

• A. 10 min• B. 30 min• C. 60 min• D. 90 min• E. 100 min

One last painful question

Page 38: The Anesthesia Machine and Breathing Systems

The DISS The PISS

Page 39: The Anesthesia Machine and Breathing Systems
Page 40: The Anesthesia Machine and Breathing Systems

Lets take a quick look at these so-called “OFPDs”

Page 41: The Anesthesia Machine and Breathing Systems

The OFPD

Page 42: The Anesthesia Machine and Breathing Systems

Lets look again

Page 43: The Anesthesia Machine and Breathing Systems

• One thing I want you to note. We have discussed the HIGH-PRESSURE CIRCUIT to this point.• Gas lines proximal to the flow valves (knobs) are

considered the high-pressure circuit• Distal to the knobs (eg. In the Thorpe tubes and

onward) you are in the low-pressure circuit.

• To be continued . . .

Lets stop for today

Page 44: The Anesthesia Machine and Breathing Systems
Page 45: The Anesthesia Machine and Breathing Systems

• Oxygen is universally on the right

• The knob is larger and fluted

• Why?

Flowmeter sequence:

Page 46: The Anesthesia Machine and Breathing Systems

It is a safety feature, but not 100% fool proof. You can still make a hypoxic gas mixture.

The less circuit AFTER the O2 joins, the less chance of a leak in the post-O2 part of the circuit.

Page 47: The Anesthesia Machine and Breathing Systems

• These are called “constant-pressure variable-orifice” flowmeters.

• Conductive coating to reduce effect of static electricity

• Calibrated to be gas-specific **

This is a Thorpe Tube

** Flow rate across a constriction depends on the gas’s viscosity at low laminar flows and its density at high turbulent flows.

Page 48: The Anesthesia Machine and Breathing Systems

• Draeger utilizes this little gem:

Oxygen/Nitrous Oxide ratio controllers

Page 49: The Anesthesia Machine and Breathing Systems

• But, Datex-Ohmeda got it right

Page 50: The Anesthesia Machine and Breathing Systems

• A couple key points on vaporization• Anesthetics have a

vapor pressure, which is the propensity to come out of solution and form a . . . Vapor.

• Vapor pressure is temp-dependent.

• Higher temp = vapor pressure

On to vaporizers

The energy required for vaporization is manifested as loss of heat from the anesthetic solution

As the anesthetic vaporizes, the solution becomes colder . . . And when the temp drops, so does the

vapor pressure !!!

Page 51: The Anesthesia Machine and Breathing Systems

Copper kettlesCopper has a high specific heat

Copper has high thermal conductivity

Resistant to the temperature drop from vaporization

Best material to maintain a constant temperature

Page 52: The Anesthesia Machine and Breathing Systems

• Measured-flow vaporizer• Separate flowmeter

for the gas flowing through the kettle

• Gas passing through the kettle becomes fully saturated

• Then you dilute it out to the proper percentage with the other flowmeter

Copper Kettle

The math gets a little funny here, and they will throw you a copper kettle equation on the ITE, so watch this . . .

Page 53: The Anesthesia Machine and Breathing Systems

• Vapor pressure of Halothane (and ISO) is about 243 mmHg at 20 C• So 243/760 = 32%

The math of a copper kettle

Page 54: The Anesthesia Machine and Breathing Systems

• At atmospheric pressure, if we put 100 ml O2 through the kettle, we will get 150 ml of FGF on the other side.• 50 mL of that will be volatile

• If we keep the flow at 5LPM total we can do this• 100 ml to the kettle → 150 ml, 50 of which is halothane• 4850 ml to the dilution limb

• 50 ml halothane / 5000 ml total FGF = 1% halothane

So here is what we know

Page 55: The Anesthesia Machine and Breathing Systems

• For the ITE, in a copper kettle there should always be a total of 5L FGF• Add 50 ml vapor to every 100 ml you put through the

kettle• Keeping the totals at 5L/min, every extra 100 ml

through the kettle increases the agent concentration 1%

• Eg. 100 → kettle → 150 (50ml agent) + 4850 = 1% agent

• 200 → kettle → 300 (100ml agent) + 4700 = 2% agent• 300 → kettle → 450 (150ml agent) + 4550 = 3% agent

For the ITE

Remember, Iso and Halothane have similar vapor pressures, so this applies to Iso too

Page 56: The Anesthesia Machine and Breathing Systems

• These are modern vaporizers

• Tec 4,5,6 all have similar mechanisms

• Aladin is very different.

Well, that sucked. Lets move on.

Tec 4 Tec 5

Tec 6

Aladin Vaporizers

Page 57: The Anesthesia Machine and Breathing Systems

• Recall saturated vapor pressures:• So, at atmospheric pressure and 20 C, if I let all the

FGF flow through the vaporizer, it would saturate and produce:

• Halo 243 243/760 = 32%• Iso 238 238/760 = 32%• Sevo 157 157/760 = 20%• Des 672 672/760 = 88%

• These are “slightly” above clinically relevant concentrations

Lets jump back to physics for one slide

Page 58: The Anesthesia Machine and Breathing Systems

• Well, then we have a variable bypass vaporizer.

But if we split the FGF between the vaporizer and a bypass channel . . .

Page 59: The Anesthesia Machine and Breathing Systems

• Note the bi-metalic strip

• This serves to compensate for temperature changes

• Vapor pressure is temperature dependent.

• If its warmer, the vapor pressure is higher, we need to slow the gas flow through the chamber.

The Datex-Ohmeda version

Page 60: The Anesthesia Machine and Breathing Systems

• Tilting old vaporizers could flood the bypass area, in which case you would deliver the full vapor pressure of the agent (Halothane 32%)

Tilting hazard !!

Page 61: The Anesthesia Machine and Breathing Systems

• The FGF does not actually flow through the sump.

• Instead fixed concentration Des vapor is added in proportion to the FGF

The Des Vaporizer -

Page 62: The Anesthesia Machine and Breathing Systems

• Well, the vapor pressure of DES at room temp is 672. • Problem #1 – The heat loss from that much vaporization

would rapidly cool the vaporizer and end up dropping the vapor pressure dramatically

• Problem #2 – 672/760 = 88%• It would take tremendous FGF through the bypass chamber to

dilute that down to a useable level.

• Just one more point on DES which is covered on the ITE – the old vaporizers for ISO, SEVO automatically compensate for changes in altitude. But, high altitude DECREASES the partial pressure for DES, so you will have to manually increase the concentration of DES at high altitudes.

Why vaporize DES this way?

Page 63: The Anesthesia Machine and Breathing Systems

• Because it is partial pressure that really matters

• Forget Volume% for a min• Partial pressure is measured in mmHg

• 2% ISO at 760mmHg = 15.2mmHg (partial pressure)• So, of the gas coming out of the machine at

760mmHg, 2% of it (or 15.2 of those mmHg) are Isoflurane.

Why?

Page 64: The Anesthesia Machine and Breathing Systems

• We know 2% @ 760 is 15.2 mmHg – but what if we climb??

• At higher altitude, the decreased Patm will allow more ISO to come out of solution. So, even if the dial is set at 2%, the actual concentration coming out is higher – again because the lower Patm lets more ISO come out of solution.

• But if we went to altitude with half the Patm, we would double the concentration coming out of the vaporizer, but the partial pressure remains the same

• 4% @ 380 is 15.2 mmHg – This is how these vaporizers self-equilibrate• % increases but Patm decreases so overall is same result

• The thing is, DES cant do this, because no matter how high you go in altitude, the % is exactly what you set on the dial . . . So you have to deliver higher concentrations to compensate for altitude

Page 65: The Anesthesia Machine and Breathing Systems

• This diagram sucks

• What I want you to take from this is that there is no bypass channel in the cassette itself

• As such, the cassette is not a tipping hazard.

• I bet you can stump the faculty with this one

The Aladin Cassette vaporizer

Page 66: The Anesthesia Machine and Breathing Systems

Ye olde breathing circuit

A couple things to notice here. This is the FG input. Note it is on the proximal side of the inspiratory valve. This valve would be closed during exhalation.

This is the old spirometer. It is just distal to the expiratory valve. So when it is taking a measurement at expiration, it is getting only exhaled tidal volume, and no FGF which would throw the measurement off.

Page 67: The Anesthesia Machine and Breathing Systems

As proven at lakeside, these spiromed spirometers really stand the test of time

Designing spiromed spirometer . . . In dog costume.

Page 68: The Anesthesia Machine and Breathing Systems

• This is the vane aneometer.• vane, like

weather vane. It spins in the wind.

• It was placed in line proximal to the exp. valve and the TV was calculated based on the spin.

Last few slides on spirometers, I promise

Page 69: The Anesthesia Machine and Breathing Systems

• This is a Pilot tube. It is used in aviation. Here is how it works:

Fixed orifice flowmeter

Pitot tubes are tube shaped and contain 2 holes. One hole faces the direction of movement and, measures the stagnation pressure of oncoming air. The other hole is on the side and measures static pressure. The difference between these two pressure types allows for the measurement of dynamic pressure, which is then used to calculate the aircraft's airspeed

Page 70: The Anesthesia Machine and Breathing Systems

• We can use those same principles to integrate flow over time to determine tidal volumes.

• If you know the airway pressures you can integrate that info as well and produce the flow-volume loops which can tell you lots about airway and lung mechanics.

So how does that help us?

Page 71: The Anesthesia Machine and Breathing Systems

Like this . . .

Page 72: The Anesthesia Machine and Breathing Systems

• Older methods of external ventilation relied on generating negative pressure around the chest wall.

Ventilators

Page 73: The Anesthesia Machine and Breathing Systems

• 4 phases are identified• Inspiration• Transition from inspiration to expiration• Expiration• Transition from expiration to inspitation

• We classify a vent based on its inspiration and transition from inspiration to expiration characteristics

Phases of ventilation

Page 74: The Anesthesia Machine and Breathing Systems

Constant pressure generator

Constant flow generator

Non-constant generator

Note they are based on the way they handle inspiration

Page 75: The Anesthesia Machine and Breathing Systems

• You can terminate inspiration based on one of three parameters

• Time• Volume• Pressure

• I expect most of us understand the volume and pressure modes.

• Time mode: You set a time allotment for inspiration and you vary the gas inflow rate during that time allotment until you reach a tidal volume you are happy with. • You can really play with this mode in patients with terrible lung

compliance and adjust it to try and limit your Ppeak.

Transition from inspiration to expiration

Page 76: The Anesthesia Machine and Breathing Systems

• The expiratory phase is simple. • Return the lungs to atmospheric pressure, unless you

have set some PEEP• A completely passive process

Expiratory phase

Page 77: The Anesthesia Machine and Breathing Systems

• It is really this phase on which we base our nomenclature for vent modes

• Volume-Control → vent adjusts gas flow rate to deliver set tidal volume based on set vent rate and I:E ratio

• Pressure-Control → vent adjusts gas flow rate to deliver a constant pressure based on set vent rate and I:E ratio

• You already knew that.

Transition from Exp to Insp

Page 78: The Anesthesia Machine and Breathing Systems

• 2 main types of Ventilator Circuit Design:

• Double-Circuit

• Piston-Driven

Now its time for some meat and potatoes

Page 79: The Anesthesia Machine and Breathing Systems

This is a double circuit ventilator

Page 80: The Anesthesia Machine and Breathing Systems

Remember this guy?

He puts out O2 at hospital line pressure (?psi)

Page 81: The Anesthesia Machine and Breathing Systems

This is him too . . . sneaky

So it makes sense that it is delivering 50 psi of O2

Page 82: The Anesthesia Machine and Breathing Systems

• 1. In a double-circuit system the tidal volume is delivered from a bellow inside a plastic bucket

• Older vents used hanging bellows that were weighted. • But when the circuit disconnected the weight would pull the

bellow down and you may think you were still ventilating the patient.

• This is what they looked like:

Remember him, and lets make a few points

Obviously this was not too safe.

Page 83: The Anesthesia Machine and Breathing Systems

• Because for some reason, we tend to notice a lot faster when the bellows are just laying flat

So, we switched to ascending bellows

Before we flip to the next slide, I need a CA-1 to tell me where the air inside the bellows is coming from…

Page 84: The Anesthesia Machine and Breathing Systems

• It’s the patient’s last exhaled breath

• Just like the breathing bag, the bellows is simply pushing air around the circle system.

So, the vent is really just an automatic breathing bag

This is now the vent.

Page 85: The Anesthesia Machine and Breathing Systems

This is how they really work:

The inside of the bellows is continuous with the circle system

The 50 psi O2 from the oxygen power outlet enters here and pressurizes the bucket. This squeezes the bellow

Page 86: The Anesthesia Machine and Breathing Systems

• When you are running the vent, the APL is excluded from the circuit• But, luckily, the vent has its own APL – the “spill valve”

• Exhaled air beyond the capacity of the bellows and circle system opens the spill valve just like a pop-off and gets shunted out the WAGD

Spill valve

Page 87: The Anesthesia Machine and Breathing Systems

• Again, the drive gas is completely isolated from the circle system

• Drive mechanism is electrically powered

Piston Ventilators

Page 88: The Anesthesia Machine and Breathing Systems

• Don’t require much drive gas • More accurate tidal volumes

• Better for patients with poor lung compliance• Better for pediatrics s and small patients

• But, pistons have one big downfall:• During that downward stroke of the expiratory phase they

actually generate negative pressure in the circuit.• Look again:

Why pistons can be better:

Page 89: The Anesthesia Machine and Breathing Systems

So we modified the circle system:Ever wonder why the breathing bag in room 5 at children’s keeps moving while you are on the vent? Its not to amuse your during the 5th circumcision of the day . . .

Page 90: The Anesthesia Machine and Breathing Systems

When the piston generates that negative pressure

This valve closes to protect the patients lungs

This valve opens which allows . ..

Air to be pulled from the breathing bag So in a piston ventilator the bag is not excluded from the circle when on the vent.

Page 91: The Anesthesia Machine and Breathing Systems

• In the older machines, there was no mechanism to compensate for FGF during vent inspiration.

• So, tidal volumes were

higher than set values based upon the fresh gas flows.

Another frequently tested ITE topic

Page 92: The Anesthesia Machine and Breathing Systems

Look at this:

Vent

While the vent is delivering your 500 ml set tidal volume . . .

There is FGF still coming in through the inlet. Any FGF coming in during inspiration on the vent will be added to the tidal volume.

Page 93: The Anesthesia Machine and Breathing Systems

• They will give you a:• Rate (10 breaths/min)• TV (1000 ml) • I:E ratio (1:2)• FGF rate (6 lpm)

• With that info you will calculate this:• 10 breaths / min = 6 second breaths• I:E of 1:2 means 2 of those seconds spent in inhalation, 10 times per

minute.• A total of 20 seconds inhalation per min

• 20 seconds x (6000 ml FGF/ 60 seconds) = 2000 ml added TV over a minute

• At 10 breaths/ min that is 200 ml added to the TV of each breath

This is how you work through this problem on the test:

Page 94: The Anesthesia Machine and Breathing Systems

• Peak pressures → highest pressure during insp. cycle

• Reflects dynamic compliance

• Plateau pressure → pressure during insp. pause

• Reflects static compliance

• IN NORMAL HEALTHY LUNGS THESE TWO NUMBERS SHOULD BE VERY CLOSE.

Airway pressures:

Page 95: The Anesthesia Machine and Breathing Systems

• Increased peak and plateau pressures:• Increased TV• Worsening overall

compliance• T-berg• Pulm edema• Insufflation• Tension

Pneumo• Mainstem

• Isolated increased peak pressures• Increased insp flow

rate• Increased Airway

resistance• Bronchospasm• Tube kink• Secretions• Aw compression

Page 96: The Anesthesia Machine and Breathing Systems

• A SEVO vaporizer will deliver an accurate concentration of an unknown volatile anesthetic if the latter shares which property with sevoflurane?• A. Molecular weight• B. Viscosity• C. Vapor Pressure• D. Blood gas partition coefficient• E. Oil gas partition coefficient

Question time:

Page 97: The Anesthesia Machine and Breathing Systems

• C

Answer:

Page 98: The Anesthesia Machine and Breathing Systems

• Which of the following valves prevents transfilling between compressed gas cylinders? • A. Fail safe valve• B. Pop off valve• C. Pressure sensor shutoff valve• D. APL valve• E. Check valve

Page 99: The Anesthesia Machine and Breathing Systems

• E

Answer

Page 100: The Anesthesia Machine and Breathing Systems

• When the pressure gauge on a size E N20 cylinder begins to fall from 750 psi, how many liters of gas remain in the cylinder?• A. 200• B. 400• C. 600• D. 800• E. Cannot be calculated

Page 101: The Anesthesia Machine and Breathing Systems

• B

Answer

Page 102: The Anesthesia Machine and Breathing Systems

• For any given concentration of volatile anesthetic, the splitting ratio is dependent on which of the following characteristics of the anesthetic?• A. Vapor pressure• B. Barometric pressure• C. Molecular weight• D. Specific heat• E. MAC at 1 atmosphere

Page 103: The Anesthesia Machine and Breathing Systems

• Answer : A

Page 104: The Anesthesia Machine and Breathing Systems

• After induction and intubation with confirmation of tracheal placement the O2 sat begins to fall. The analyzer as well as the mass spectrometer show inspired O2 concentration of 4%. O2 line pressure is 55 psi and there is a E cylinder with 1600 psi. What do you do?

• A. Exchange the tank• B. Swith the O2 and N2O lines out• C. Disconnect the hospital O2 line• D. Extubate and mask ventilate• E. Replace the pulse ox probe

Page 105: The Anesthesia Machine and Breathing Systems

• C

Answer:

Page 106: The Anesthesia Machine and Breathing Systems

• Micrognathia• Glossoptosis

(retraction and posterior displaced tongue)

• Upper airway obstruction

• Frequent cleft palate

Pierre Robin

Page 107: The Anesthesia Machine and Breathing Systems

• Absent cheekbones

• Other craniofacial deformities

• Malformed or absent ears

• Conductive hearing loss

• Micrognathia

Treacher Collins

Page 108: The Anesthesia Machine and Breathing Systems

• Anomalous development of first and second branchial arches

• Hemifacial microsomia

• Ear, nose soft palate, vertebral abnormalities, often unilateral

Goldenhar

Page 109: The Anesthesia Machine and Breathing Systems

• Myopathy similar to central core disease

• Kids also have• Kyphosis/Lordosis• Micrognathia• Cryptorchidism• Pectus Carinatum• Other facial features

• THESE KIDS ALMOST ALWAYS HAVE MH !!!

King-Denborough

Page 110: The Anesthesia Machine and Breathing Systems

• Hypotonia• Mild developmental

delay• Skeletal abnormalities

(scoliosis)• Absent mitochondria

on light microscopy• VERY HIGH RISK OF

MH

• Can be autosomal dominant, recessive, but frequently De Novo mutation of RYR1

Central Core disease

Page 111: The Anesthesia Machine and Breathing Systems

• Trisomy 21• Hypothyroid• Small trachea

(reduce tube size)

• OA instability• Congenital heart

Dz

• NO INCREASED RISK OF MH

Downs

Page 112: The Anesthesia Machine and Breathing Systems

Gastroschisis

Page 113: The Anesthesia Machine and Breathing Systems

Omphalocoele

Page 114: The Anesthesia Machine and Breathing Systems
Page 115: The Anesthesia Machine and Breathing Systems

TEG

Page 116: The Anesthesia Machine and Breathing Systems

• Common peroneal nerve injury

• Foot drop• Loss of eversion• Loss of toe extension

• May also cause stretch on sciatic, causing sciatica

Lithotomy

Page 117: The Anesthesia Machine and Breathing Systems

• Which causes Meralgia paraesthetica• Lateral Femoral

Cutaneous Nerve

• A SENSORY ONLY NEUROPATHY

Do not confuse with excessive hip flexion

Page 118: The Anesthesia Machine and Breathing Systems

• Excessive retraction in lower abdominal surgery• Femoral nerve

• Quads weakness (decreased knee extension)

• Numbness anterior thigh, medial leg

Page 119: The Anesthesia Machine and Breathing Systems

• Obturator nerve• Decreased leg

adduction• Numbness at the

medial thigh

Difficult Forceps Delivery

Page 120: The Anesthesia Machine and Breathing Systems