Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside...

144
www.knick.de Latest Product Information: Stratos®Pro A2... PH Instruction Manual

Transcript of Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside...

Page 1: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

www.knick.deLatest Product Information:

Stratos®Pro A2... PHInstruction Manual

Page 2: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

2

Warranty WarrantyDefects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender).Sensors, fi ttings, and accessories: 1 year.

Subject to change without notice.

Return of Products Under WarrantyPlease contact our Service Team before returning a defective device. Ship the cleaned device to the address you have been given. If the device has been in contact with process fl uids, it must be decontaminated/disinfected before shipment. In that case, please attach a corresponding certifi cate, for the health and safety of our service personnel.

DisposalPlease observe the applicable local or national regulations concerning the disposal of “waste electrical and electronic equipment”.

Page 3: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

3

Documents Supplied

Stratos® Pro A411/A211 pHQuickStart

Other languages: www.knick.de

Short Instructions ............................3

Kurzübersicht ................................. 15

Quick Start ...................................... 27

Быстрый старт .............................. 39

Inicio rápido ................................... 51

Início rápido ................................... 63

快速启动 ........................................ 75

CD-ROMComplete documentation:• Instruction manuals• Safety instructions • Short instructions

Safety InformationIn offi cial EU languages and others.• ATEX / IECEX / FM / CSA• EC Declarations of Conformity

Short InstructionsIn German, English, French, Russian,Spanish, Portuguese, Swedish, and Dutch. More languages on CD-ROM and on our website: www.knick.de• Installation and commissioning• Operation• Menu structure• Calibration• Error messages and recommended actions

Stratos® Pro SeriesSafety Instructions

www.knick.de

Specifi c Test Report

Page 4: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

4

ContentsDocuments Supplied ..................................................................... 3Introduction .................................................................................... 7

Intended Use ....................................................................................................7

Safety Information ......................................................................... 8Overview of Stratos Pro A2... PH ...............................................10Assembly ........................................................................................11

Package Contents .........................................................................................11Mounting Plan, Dimensions .....................................................................12Pipe Mounting, Protective Hood ............................................................13Panel Mounting .............................................................................................14

Installation .....................................................................................15Installation Instructions .............................................................................15Rating Plates / Terminal Assignments...................................................16Wiring of Stratos Pro A201/211 PH ........................................................17Wiring Examples ...........................................................................................19

User Interface, Keypad ................................................................30Display ............................................................................................31

Signal Colors (Display Backlighting) ......................................................31

Measuring Mode...........................................................................32Selecting the Mode / Entering Values ......................................33Operating Modes ..........................................................................34

Menu Structure of Modes and Functions ............................................35HOLD Mode ....................................................................................................36Alarm .................................................................................................................37

Configuration ................................................................................38Menu Structure of Configuration ...........................................................38Parameter Set A/B ........................................................................................40Configuration (Original for Copy) ...........................................................44Sensor ...............................................................................................................46Current Output 1 ..........................................................................................54Current Output 2 ..........................................................................................60

Page 5: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

5

ContentsTemperature Compensation ....................................................................62Alarm .................................................................................................................64Time and Date ...............................................................................................66Tag Number ...................................................................................................66

Digital Sensors ..............................................................................68Operation ........................................................................................................68Connecting a Digital Sensor .....................................................................69Sensor Replacement ...................................................................................70

Calibration .....................................................................................72Selecting a Calibration Mode ..................................................................73Zero Adjustment (ISFET) ............................................................................74Automatic Calibration (Calimatic) ..........................................................76Manual Calibration with Buffer Entry ...................................................78Data Entry of Premeasured Sensors ......................................................80Product Calibration (pH) ............................................................................82ORP ( Redox) Calibration ...........................................................................84Temp Probe Adjustment ............................................................................86

Measurement ................................................................................87Diagnostics ....................................................................................88Service ............................................................................................93Operating States ...........................................................................96Product Line and Accessories ....................................................97A201/A211X: Supply Units and Connection ...........................98Specifications ................................................................................99Buffer Tables ................................................................................105

-U1- Specifiable Buffer Set .....................................................................114

Error Handling .............................................................................117Error Messages ............................................................................118Sensoface .....................................................................................121

Page 6: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

6

ContentsEC Declaration of Conformity ..................................................124Stratos Pro A201X/A211X PH: Control Drawing ...................126FM Control Drawing ...................................................................128CSA Control Drawing .................................................................129FDA 21 CFR Part 11 ....................................................................130

Electronic Signature – Passcodes ........................................................130Audit Trail .....................................................................................................130

Glossary ........................................................................................131Index .............................................................................................135

Trademarks ..................................................................................................143

Passcodes .....................................................................................144

Page 7: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

7

Introduction Intended UseThe Stratos Pro A2... PH is used for pH/mV, ORP, and temperature measurement in industry, environment, food processing, and sewage treatment. The sturdy molded enclosure can be fi xed into a control panel or mounted on a wall or at a post. The protective hood, which is avail-able as accessory, provides additional protection against direct weather exposure and mechanical damage.The device has been designed for application with commercially available sensors with a nominal zero point at pH 7, ISFET sensors, or digital sensors.Plain-text messages in a large, backlit display allow intuitive opera-tion. The colored display backlighting signals alarm messages (red) or HOLD mode (orange).The “Sensocheck“ automatic monitoring of glass and reference electrode and the “Sensoface“ function for clear indication of the sensor condition provide excellent diagnostics. The internal logbook can handle up to 100 entries – up to 200 with AuditTrail (TAN).The device provides two parameter sets which can be switched manually or via a control input for diff erent process adaptations or diff erent process conditions (e.g. beer and CIP).Password protection for granting access rights during operation can be confi gured.Two fl oating, digital control inputs (“Hold“ and “Control“) are available for external control.The device provides two current outputs (for transmission of measured value and temperature, for example).

Approvals for Measurement in Hazardous Locations:

Stratos Pro A201/A211N PH: General Safety.

Stratos Pro A201/A211X PH: Approved for operation in hazardous locations according to IECEx / ATEX / FM / CSA.

Page 8: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

8

Safety InformationSafety information –Be sure to read and observe the following instructions!The device has been manufactured using state of the art technology and it complies with applicable safety regulations.When operating the device, certain conditions may nevertheless lead to danger for the operator or damage to the device.

See also separate document:“Safety Instructions“• (EC Declaration of Conformity, FM, CSA, ATEX (if applicable) Certifi cates)

CAUTION!

Please note:Before commissioning it must be proved that the device may be connected with other equipment.

Commissioning must only be performed by trained personnel autho-rized by the operating company! Whenever it is likely that protection has been impaired, the device shall be made inoperative and secured against unintended operation. The protection is likely to be impaired if, for example:

the device shows visible damage• the device fails to perform the intended measurements• after prolonged storage at temperatures above 70°C• after severe transport stresses•

Before recommissioning the device, a professional routine test must be performed. This test must be carried out at the manufacturer's factory.

Page 9: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

9

Safety Information

Terminals:Screw terminal, suitable for single wires / fl exible leads up to 2.5 mm2 (AWG 14). Recommended torque for the terminal screws: 0.5 ... 0.6 Nm.

Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

Information for Installation in Hazardous Locations(Stratos Pro A201/A211X PH)

Be sure to observe the stipulations of EN 60079-10 / EN 60079-14 • or the corresponding local regulations during installation and commissioning. See also separate “Safety Instructions“ document.

Approvals for application in hazardous locations(Stratos Pro A201/A211X PH)

according to IECEx in Zone 0, 1, 20, 21• according to ATEX in Zone 0, 1, 2 , 20, 21• according to • cCSAus in Class I Div 1, 2 / Zone 1according to FM in Class I, Div 1, 2 / Zone 1•

When the device provides diff erent types of protection, the operator must specify the applied type of protection during installation. To do so, use the checkboxes on the rating plate:

Important Notice: The operator must indicate the type of protection!

Page 10: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

10

OverviewOverview of Stratos Pro A2... PH

17

9

8

C

H

I

F

G

D

E

K

5

6

13

14

B

A Output 1Measure el.

Reference el.

SG

RTD

RTD

ISM GND

ISM data

ISFET

ISFET+3 V

–3 V

Shield

Input +

Input –

CONTROL

CONTROL

+ Output 1, 2 / HART

– Output 1 / HART

– Output 2

pH / mV / temp input

+3 V /–3 Vfor ISFET adapter

input

Control

10

11

HOLD

HOLD

input

Current

input

HOLD

Output 2

1

2

3

4

RS 485 Power supply

RS 485 A

RS 485 B

GND/Shield

Page 11: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

11

Assembly Package ContentsCheck the shipment for transport damage and completeness!The package should contain:

Front unit, rear unit, bag containing small parts• Specific test report• Documentation (cf Pg 3)• CD-ROM•

Fig.: Assembling the enclosure

Jumper (3 x)1) Washer (1 x), for conduit 2) mounting: Place washer between enclosure and nutCable tie (3 x)3) Hinge pin (1 x), insertable 4) from either sideEnclosure screw (4 x)5)

Sealing insert (1 x)6) Rubber reducer (1 x)7) Cable gland (3 x)8) Filler plug (3 x)9) Hexagon nut (5 x)10) Sealing plug (2 x), for sealing 11) in case of wall mounting

11

10

9

8

7 6 5 4

1

2

3

Page 12: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

12

Assembly Mounting Plan, Dimensions

Fig.: Mounting plan (All dimensions in mm!)

Cable gland (3 x)1) Knockouts for cable gland or 2) ½" conduit, 21.5 mm dia. (2 knockouts) Conduits not included!Knockout for pipe mounting 3) (4 x)Knockout for wall mounting 4) (2 x)

148

148

42 4242

14

11741

21431

2

34 80

746,

2

4

3

Page 13: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

13

Assembly Pipe Mounting, Protective Hood

Fig.: ZU 0737 protective hood for wall and pipe mounting(All dimensions in mm!)

Fig.: ZU 0274 pipe-mount kit (All dimensions in mm!)

147199

91

185

1

2

3

4

Hose clamp with worm gear 1) drive to DIN 3017 (2 x)Pipe-mount plate (1 x)2) For vertical or horizontal posts 3) or pipesSelf-tapping screw (4 x)4)

ø40...ø60

Page 14: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

14

Assembly Panel Mounting

Fig.: ZU 0738 panel-mount kit (All dimensions in mm!)

Circumferential sealing 1) (1 x)Screw (4 x)2) Position of control panel3) Span piece (4 x)4) Threaded sleeve (4 x)5)

Cutout 138 x 138 mm (DIN 43700)

<30 76 31

1...221

23451...221

Page 15: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

15

InstallationInstallation Instructions

Installation of the device must be carried out by trained experts in • accordance with this instruction manual and as per applicable local and national codes.Be sure to observe the technical specifi cations and input ratings • during installation!Be sure not to notch the conductor when stripping the insulation!• The supplied current must be galvanically isolated. If not, connect • an isolator module.All parameters must be set by a system administrator prior to • commissioning!

Terminals: suitable for single wires / fl exible leads up to 2.5 mm2 (AWG 14)

Additional safety precautions have to be taken for applications in hazardous locations according to IECEx, ATEX, FM, CSA! (See also separate “Safety Instructions“ document.)

Page 16: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

16

Installation Rating Plates / Terminal Assignments

Fig.: Outer rating plate Stratos Pro A2...N at bottom of front

Fig.: Terminal assignments of Stratos Pro A2...

Fig.: Stratos Pro A2...X rating plate at outside bottom of front

Important Notice: The operator must indicate the type of protection!When the device provides diff erent types of protection, the operator must specify the applied type of protection during installation. To do so, use the checkboxes on the rating plate.See also “Safety Information“ chapter.

Fig.: Example of additional approval plate (cCSAus, FM)The specifi cations refer to the respective device.

Page 17: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

17

Installation Wiring of Stratos Pro A201/211 PH

Sensor connection MK-PH moduleA meas. el.

B ref. el.

C SG

D +3 V

E –3 V

F ISM GND

G ISM data

H RTD (GND)

I RTD

K Shield

Fig.: MK-PH module terminal assignments

Fig.: Terminals, device opened, back of front unit

Areas for placing the screwdriver to pull out the terminals

1 9 10 18HART

Terminal row 11 Power supply

2 RS 485 A

3 RS 485 B

4 GND/shield

5 + input

6 – input

7 PA (equip. bonding)

8 +out 1,2/HART

9 – out 1/HARTIn addition: 2 HART pins (between terminal row 1 and 2

Terminal row 210 hold

11 hold

12 n.c.

13 contr

14 contr

15 n.c.

16 n.c.

17 – out 2

18 n.c.

Page 18: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

18

Page 19: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

19

Wiring Examples Example 1:

Measuring task: pH, temperature, glass impedanceSensors ( example): HA 405-DXK-58 (Mettler-Toledo)Cable (example): AS9 ZU 0318 (Knick)

Jumper!

Core

Shie

ld

Sens

or(s

)Ca

ble

Dev

ice

Page 20: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

20

Wiring ExamplesExample 2:Measuring task: pH/ORP, temp, glass impedance, ref. impedanceSensors (example): pH: HA 405-DXK-58 (Mettler-Toledo),

Pt: ZU 0073 (Knick)Cable (example): 2x AS9 ZU 0318 (Knick)

ISM

(GN

D)

B C D E F G HA I K

mea

s. e

l

ref.

el.

SG +3 V

–3 V

ISM

(Dat

a)

RTD

(GN

D)

RTD

shie

ld

Core

Shie

ld

Core

Shie

ld

Sens

or(s

)Ca

ble

Dev

ice

Page 21: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

21

Wiring ExamplesExample 3:

Measuring task: pH, temp, glass impedanceSensors (example): SE 533 (Knick)Cable (example): VP6 ZU 0313 (Knick)

B C D E F G HA I K

mea

s. e

l

ref.

el.

SG +3 V

–3 V

ISM

(GN

D)

ISM

(Dat

a)

RTD

(GN

D)

RTD

shie

ld

Gre

en

Whi

te

Jumper!

Core

Shie

ld

Shie

ld

Clea

r

Red

Yello

w/G

reen

Sens

or(s

)Ca

ble

Dev

ice

Page 22: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

22

Wiring Examples

B C D E F G HA I K

mea

s. e

l

ref.

el.

SG +3 V

–3 V

ISM

(GN

D)

ISM

(Dat

a)

RTD

(GN

D)

RTD

shie

ld

Example 4:

Measuring task: pH/ORP, temp, glass impedance, ref. impedanceSensors (example): pH: SE 533 (Knick)

Pt: ZU 0073 (Knick)Cable (example): VP6 ZU 0313 (Knick) or AS9 ZU 0318 (Knick)

Gre

en

Whi

te

Core

Shie

ld

Shie

ld

Clea

r

Red

Yello

w/G

reen

Core

Shield

Sens

or(s

)Ca

ble

Dev

ice

Page 23: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

23

Wiring Examples

B C D E F G HA I K

mea

s. e

l

ref.

el.

SG +3 V

–3 V

ISM

(GN

D)

ISM

(Dat

a)

RTD

(GN

D)

RTD

shie

ld

Example 5:

Measuring task: pH/ORP, temp, glass impedance, ref. impedanceSensors (example): InPro 4260 (Mettler-Toledo)Cable (example): VP6 ZU 0313 (Knick)

Gre

en

Whi

te

Core

Shie

ld

Shie

ld

Clea

r

Red

Yello

w/G

reen

Blue

Sens

or(s

)Ca

ble

Dev

ice

Page 24: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

24

Wiring Examples

B C D E F G HA I K

mea

s. e

l

ref.

el.

SG +3 V

–3 V

ISM

(GN

D)

ISM

(Dat

a)

RTD

(GN

D)

RTD

shie

ld

Example 6:

Measuring task: pH, temp (safe areas only)Sensors (example): InPro 4260 ISFET (Mettler-Toledo)Cable (example): ZU 0582 (Knick)

Pink

Brow

n

Gre

en

Whi

te

Yello

w/G

reen

Gra

y

Yello

w

Jumper!

Sens

or(s

)Ca

ble

wit

h am

plifi

erD

evic

e

Page 25: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

25

Wiring ExamplesExample 7:

Caution!Do not connect an additional analog sensor!Measuring task: pH/ORP, temp, glass impedance, ref. impedanceSensors (example): ISM digital InPro 4260i (Mettler-Toledo)Cable (example): AK9 (Mettler-Toledo)

Core

Shie

ld

Sens

or(s

)Ca

ble

Dev

ice

Page 26: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

26

Wiring ExamplesExample 8 - Note: Switch off Sensocheck!

Measuring task: ORP, temp, glass impedance, ref. impedance Sensors (example): ORP: SE 535 (Knick)Cable (example): AS9 ZU 0318 (Knick)

Jumper!

Core

Shie

ld

Shie

ld

Sens

or(s

)Ca

ble

Dev

ice

Page 27: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

27

Wiring Examples

1 2 3 4

3,3

V

RS 4

85 (A

)

RS 4

85 (B

)

GN

D/s

hiel

d

Example 9:

Measuring task: pH/ORP, temp, glass impedance, ref. impedanceSensors (example): SE 533/1-ADIN (Knick)Cable (example): CA/003-NAADIN11 (Knick)

Indu Con

Yello

w

Whi

te/C

lear

Shie

ld (b

lue)

Brow

n

Gre

en

Sens

or(s

)Ca

ble

Dev

ice

Page 28: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

28

Page 29: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

29

Wiring Examples

1 2 3 4

3,3

V

RS 4

85 (A

)

RS 4

85 (B

)

GN

D/s

hiel

d

Example 10: Memosens

Measuring task: pH/ORP, temp, glass impedance, ref. impedanceSensors (example): Orbisint CPS 11 D MemosensCable (example): CYK 10

Yello

w

Whi

te

Shie

ld (c

lear

)

Brow

n

Gre

en

Sens

or(s

)Ca

ble

Dev

ice

Page 30: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

30

User Interface, Keypad

1 IrDA transmitter/receiver2 Display3 Keypad4 Rating plate (bottom)

Key Functionmeas Return to last menu level•

Directly to measuring mode (press > 2 s)• info Retrieve information•

Show error messages• enter Confi guration: Confi rm entries, •

next confi guration stepCalibration:• Continue program fl owMeasuring mode: Display output current•

Arrow keysup / down

Measuring mode: Call menu• Menu: Increase/decrease a numeral• Menu: Select•

Arrow keysleft / right

Measuring mode: Call menu• Menu: Previous/next menu group• Number entry: Move between digits•

1

3

4

2

Page 31: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

31

Display

1 Temperature2 Sensocheck3 Interval/response time4 Sensor data5 Digital sensor devaluated6 Limit values7 Alarm8 Service9 Parameter sets A/B10 Calibration11 Digital sensor12 Waiting time running

13 Info available14 HOLD mode active15 Main display16 Secondary display17 Proceed with enter18 Digital sensor19 Diagnostics20 Confi guration mode21 Calibration mode22 Measuring mode23 Sensoface24 Measurement symbol

1 2 3 4 5 6 7 8 9 10 11

171819202122

121314

15

16

2324

Signal Colors ( Display Backlighting)

Red AlarmOrange HOLD mode (Calibration, Confi guration, Service)Turquoise DiagnosticsGreen InfoPurple Sensoface message

Page 32: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

32

Measuring ModeAfter the operating voltage has been connected, the device auto-matically goes to “Measuring“ mode. To call the measuring mode from another operating mode (e.g. Diagnostics, Service): Hold meas key depressed (> 2 s).

Process variableTemperature

Sensoface indicator(sensor status)

Time

Mode indicator (measuring)

Hold meas key depressed to call the measuring mode(pressing once more switches the display)

Active parameter set

enter key

In measuring mode the display indicates:Measured value and time (24/12 h AM/PM) as well as temperature in • °C or °F (formats selected during confi guration)

By pressing the meas key in measuring mode you can view the following displays (for approx. 60 sec):

Measured value and selection of parameter set A/B • (if set to “Manual“)Measured value and tag (point of measurement designation • – entered during confi guration)Time and date•

Pressing the enter key shows the output currents. They are displayed as long as enter is held depressed, then the measured-value display will return after 3 sec.

The device must be confi gured for the respective measurement task!

Page 33: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

33

Selecting the Mode / Entering Values

To enter a value:Select numeral: left / right arrow key5) Change numeral: up / down arrow key6) Confi rm entry with 7) enter

To select the operating mode:Hold 1) meas key depressed (> 2 s) (measuring mode)Press any arrow key: the selection menu appears2) Select operating mode using left / right arrow key3) Press 4) enter to confi rm the selected mode

Selection menu

Selected mode (blinks)

1 2

3

4

5 6

7

Page 34: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

34

DiagnosticsDisplay of calibration data, display of sensor data, performing a device self-test, viewing the logbook entries, display of hardware/software versions of the individual components. The logbook can store 100 events (00...99). They can be displayed directly on the device. The logbook can be extended to 200 entries using a TAN (Option).

HOLDManual activation of HOLD mode, e.g. for replacing a digital sensor. The signal outputs adopt a defi ned state.

CalibrationEvery sensor has typical characteristic values, which change in the course of the operating time. Calibration is required to supply a cor-rect measured value. The device checks which value the sensor delivers when measuring in a known solution. When there is a deviation, the de-vice can be “adjusted“. In that case, the device displays the “actual“ value and internally corrects the measurement error of the sensor. Calibration must be repeated at regular intervals. The time between the calibration cycles depends on the load on the sensor. During calibration the device is in HOLD mode. During calibration the analyzer remains in the HOLD mode until it is stopped by the operator.

Confi gurationThe analyzer must be confi gured for the respective measurement task. In the “Confi guration“ mode you select the connected sensor, the mea-surement range to be transmitted, and the conditions for warning and alarm messages. During confi guration the device is in HOLD mode. Confi guration mode is automatically exited 20 minutes after the last keystroke. The device returns to measuring mode.

ServiceMaintenance functions (monitor, current source), IrDA operation, passcode assignment, reset to factory settings, enabling of options (TAN).

Operating Modes

Page 35: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

35

Menu Structure of Modes and Functions

Measuring mode

TAG display

Display of calibration data

Display of sensor data

Self test: RAM, ROM, EEPROM, module

100 events with date and time

Display of measured values (mV_pH, mV_ORP, RTD, resistances of glass electrode, reference electrode)

Display of software version, model designation, serial number

Manual activation of HOLD mode, e.g. for sensor replacement.The signal outputs behave as configured (e.g. last measured value, 21 mA)

pH adjustment (as configured)

ORP adjustment

Product calibration

Zero adjustment (for ISFET only)

Adjustment of temperature probe

Configuring parameter set A

Configuring parameter set B

Display of measured values for validation (simulators)

Current source, output 1

Current source, output 2

Activating the IrDA interface

Specifying access codes for operating modes

Reset to factory setting

Enabling an option via TAN

Pressing any arrow key opens the selection menu. Select the menu group using the left/right arrow keys.Press enter to open a menu. Press meas to return.

(Access via code, factory setting:5555)

measCLK display

meas meas

after 60 s after 60 s

Page 36: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

36

HOLD ModeThe HOLD mode is a safety state during confi guration and calibration. Output current is frozen (Last) or set to a fi xed value (Fix).The HOLD mode is indicated by orange display backlighting.

Terminating the HOLD Mode The HOLD mode is ended by switching to measuring mode (hold meas key depressed). The display reads “Good Bye“, after that, the HOLD mode is exited. When the calibration mode is exited, a confi rmation prompt ensures that the installation is ready for operation (e.g.: sensor reinstalled, located in process).

HOLD mode, display icon:

HOLD active

21

4HOLD active

Output current[mA]

Output signal HOLDSetting FIX = 21.0 mA

Output signal HOLDSetting LAST

Output Signal During HOLD:

Output Signal ResponseLast:• The output current is frozen at its last value. Recommended for short confi guration procedures. The process should not change decisively during confi guration. Changes are not noticed with this setting!Fix:• The output current is set to a value that is noticeably diff erent from the process value to signal the control system that the device is being worked at.

Page 37: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

37

Alarm

Manual Activation of HOLDThe HOLD can be activated manually from the HOLD menu. This allows checking or replacing a sensor, for example, without provoking unintended reactions of outputs or contacts.Press meas key to return to selection menu.

AlarmWhen an error has occurred, Err xx is displayed immediately.Only after expiry of a user-defi ned delay time will the alarm be registered and entered in the logbook.During an alarm the display blinks, the display backlighting turns red.

Error messages can also be signaled by a 22 mA output current (see Confi guration).2 sec after the failure event is corrected, the alarm status will be deleted.

External Activation of HOLDThe HOLD mode can be activated from outside by sending a signal to the Hold input (e.g. from the process control system).

10

11

Process control system (PCS)

Power supply12...24 V AC/DC

Stratos Pro A2...

input

HOLD

HOLD inactive 0...2 V AC/DCHOLD active 10...30 V AC/DC

Page 38: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

38

Configuration Menu Structure of Confi gurationThe device provides 2 parameter sets “A“ and “B“. By switching between the parameter sets you can adapt the device to diff erent measurement situations, for example. Parameter set “B“ only permits setting of process-related parameters.The confi guration steps are assigned to diff erent menu groups.Using and , you can jump between the individual menu groups.Each menu group contains menu items for setting the parameters.Pressing enter opens a menu item. The values are edited using and . Pressing enter confi rms/stores the settings. Return to measurement: Press meas.

Select menu group Menu group Code Display Select

menu item

Sensor selection SNS:

Menu item 1

...

Menu item ...

Current output 1 OT1:

Current output 2 OT2:

Compensation COR:

Alarm mode ALA:

Setting the clock CLK:

Tag number TAG:

enter

enter

enter

enter

Page 39: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

39

Confi guration Parameter Set A/B: Confi gurable Menu Groups

(Some parameters are identical in A and B. They are confi gured in parameter set A only.)

Menu group Parameter set A Parameter set BSENSOR Sensor selection ---OUT1 Current output 1 Current output 1OUT2 Current output 2 Current output 2CORRECTION Compensation CompensationALARM Alarm mode Alarm modePARSET Parameter set

selection---

CLOCK Setting the clock ---TAG Tag number ---

External Switchover of Parameter Sets A/BYou can switch between parameter sets A and B by applying a signal to the CONTROL input. For confi guration, see Pg 43.

13

14

e.g. process control system

Max. 30 V AC/DC

Stratos Pro A2...

input

Control

Parameter set A active 0...2 V AC/DCParameter set B active 10...30 V AC/DC

Page 40: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

40

Confi gurationParameter Set A/B Manual selection

Display Action RemarkTo switch between parameter sets:Press meas(must have been selected in CONFIG)

Wrong settings change the measurement properties!

PARSET blinks in the lower line.Select parameter set using and keys

Select PARSET A / PARSET B

Confi rm with enterCancel with meas

Page 41: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

41

Confi guration

Confi guration Choices DefaultSensor (SENSOR)SNS: STANDARD

ISFETINDUCONISMMEMOSENS

STANDARD

RTD TYPE 100 PT1000 PT30 NTC

100 PT

TEMP UNIT °C / °F °C

TEMP MEAS AUTOMANEXT (only if enabled via TAN)

AUTO

MAN –20...200 °C(–4...392 °F)

025.0 °C(077.0 °F)

TEMP CAL AUTOMANEXT (only if enabled via TAN)

AUTO

MAN –20...200 °C(–4...392 °F)

025.0 °C(077.0 °F)

CAL MODE AUTOMANDAT

AUTO

AUTOBUFFER SET

-00-...-09-, -U1-Please note:Pressing info displays nominal buffer values and manufacturers

-00-

U1(specifiable buffer set)

EDIT BUFFER 1 (NO, YES) NO

Enter values for buffer 1

EDIT BUFFER 2 (NO, YES) NO

Enter values for buffer 2

Page 42: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

42

Confi guration

Confi guration Choices DefaultSensor (SENSOR)SNS: CAL TIMER OFF

FIXADAPT

OFF

ON CAL-CYCLE 0...9999 h 0168 h

DS(InduCon)

CIP COUNT ON/OFF OFF

ON CIP CYCLES 0...9999 CYC 0025 CYC

SIP COUNT ON/OFF OFF

ON SIP CYCLES 0...9999 CYC 0025 CYC

Output 1 (OUT1)OT1: CHANNEL PH/ORP/TMP PH

PH BEGIN –2.00...16 PH 00.00 PH

END –2.00...16 PH 14.00 PH

ORP BEGIN –1999...1999 mV

END –1999...1999 mV

TMP°C

BEGIN –20...300 °C

END –20...300 °C

TMP°F

BEGIN –4...572 °F

END –4...572 °F

FILTERTIME 0...120 SEC 0000 SEC

22mA-FAIL ON/OFF OFF

HOLD MODE LAST/FIX LAST

FIX HOLD-FIX 4...22 mA 021.0 mA

Page 43: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

43

Confi guration

Confi guration Choices DefaultOutput 2 (OUT2)OT2: CHANNEL PH/ORP/TMP TMP

... other steps like output 1

Temperature compensation (CORRECTION)COR: TC LIQUID –19.99...19.99%/K 00.00%/K

TEMP EXT* ON/OFF OFF

ON I-INPUT 0...20 mA/4...20 mA

4...20 mA

°C BEGIN 4 mA –20...200 °C 000.0 °C

END 20 mA –20...200 °C 100.0 °C

°F BEGIN 4 mA –4...392 °F 032.0 °F

END 20 mA –4...392 °F 212.0 °F

Alarm (ALARM)ALA: DELAYTIME 0...600 SEC 0010 SEC

SENSOCHECK ON/OFF OFF

Parameter set (PARSET)PAR: Select fixed parameter set

(A) or switch between A/B via control input or manu-ally in measuring mode

PARSET FIX /CNTR INPUT /MANUAL

PARSET FIX(parameter set A fixed)

Real-time clock (CLOCK)CLK: FORMAT 24 h / 12 h

24 h TIME hh/mm 00..23:00...59 00:00

12 h TIME hh/mm 00...11 AM/PM:00...59

00.00

DAY/MONTH 01...31/01...12 31.12.

YEAR 2000...2099 2006

Tag number (TAG)TAG: (Input in text line) XXXXXXXXXX

*) is only displayed if enabled and SENSOR TEMP EXT has been selected.

Page 44: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

44

Two complete parameter sets are stored in the EEPROM. As delivered, the two sets are identical but can be edited.

Please note:Fill in your confi guration data on the following pages or use them as original for copy.

Configuration (Original for Copy)

Parameter Parameter set A Parameter set BSNS: Sensor type --- *)

SNS: RTD type ---SNS: Temperature unit ---SNS: Temp measurement ---SNS: Manual meas. temp ---SNS: Calibration temp ---SNS: Manual cal temp ---SNS: Calibration mode ---SNS: Buffer set selection(-U1-, see Appendix)

---

SNS: Calibration timer ---SNS: Calibration cycle ---SNS: CIP counter ---SNS: CIP cycles ---SNS: SIP counter ---SNS: SIP cycles ---OT1: Process variable

OT1: Current start

OT1: Current end

*) These parameters cannot be adjusted in parameter set B, the values are the same as in parameter set A.

Page 45: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

45

(Original for Copy) Confi gurationParameter Parameter set A Parameter set BOT1: Filter time

OT1: 22 mA error current

OT1: HOLD mode

OT1: HOLD-FIX current

OT2: Process variable

OT2: Current start

OT2: Current end

OT2: Filter time

OT2: 22 mA error current

OT2: HOLD mode

OT2: HOLD-FIX current

COR: Temp coefficient

COR: Ext. temp input

COR: Current range

COR: Current start

COR: Current end

ALA: Alarm on/off

ALA: Delay

ALA: Sensocheck on/off

PAR: Parameter set selection

---

CLK: Time format ---CLK: Time hh/mm ---CLK: Day/month ---CLK: Year ---TAG: Tag number ---

*) These parameters cannot be adjusted in parameter set B, the values are the same as in parameter set A.

Page 46: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

46

1

Confi guration Sensor Select: sensor type, temperature probe, temperature unit,

temp detection during measurement

Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 keys,

press enter. Select 4 SENSOR menu using keys, press

enter. All items of this menu group are indicated by 5

the “SNS:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas key until the [meas] mode indicator is displayed.

2

3

4

Select sensor type

Select type of temp probe

Temperature unit

Temp detection during measurement

(Man. temperature)

Temp detection during calibration

(Man. temperature)

Calibration mode

(AUTO: Buffer set)

Calibration timer

Calibration cycle

Cleaning cycles

Cleaning cycle counter

Sterilization cycles

Sterilization cycle counter

5enter

enter

6

enter

5

enter

meas

Page 47: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

47

Confi gurationMenu item Action ChoicesSelect sensor type Select sensor type using

keys.

Confirm with enter

STANDARDISFETINDUCON (DS)ISM-DISO (DS)MEMOSENS (DS)

Select type of temp probe

(not for digital sensors)Select type of temperature probe using keys.

Confirm with enter

100 PT1000 PT30 NTC

Temperature unit Select °C or °F using keys.

Confirm with enter

°C / °F

Temp detection during measurement

Select mode using :AUTO: Measured by s ensor MAN: Direct input of temperature, no measure-ment (see next step)EXT: Temperature speci-fied via current input (only if TAN E enabled)Confirm with enter

AUTOMANEXT

(Manual temp) Modify digit using , select next digit using

keys.Confirm with enter

–20...200 °C(–4...+392 °F)

5

Page 48: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

48

Confi guration

1

SensorSelect: temp detection during calibration, calibration mode

Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 keys,

press enter. Select 4 SENSOR menu using keys, press

enter. All items of this menu group are indicated by 5

the “SNS:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas key until the [meas] mode indicator is displayed.

2

4

Select sensor type

Select type of temp probe

Temperature unit

Temp detection during measurement

(Man. temperature)

Temp detection during calibration

(Man. temperature)

Calibration mode

(AUTO: Buffer set)

Calibration timer

Calibration cycle

Cleaning cycles

Cleaning cycle counter

Sterilization cycles

Sterilization cycle counter

5enter

enter

6

enter

5

enter

meas

3

Page 49: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

49

Confi gurationMenu item Action ChoicesTemp detection during calibration

Select mode using :AUTO: Measured by sensor MAN: Direct input of temperature, no measure-ment (see next step)EXT: Temperature speci-fied via current input (only if TAN E enabled)Confirm with enter

AUTOMANEXT

(Manual temp) Modify digit using , select next digit using

keys.Confirm with enter

–20...200 °C(-4...+392 °F)

Calibration Mode Select CALMODE using keys:

AUTO: Calibration with Calimatic buffer set recognitionMAN: Manual input of buffer solutions.DAT: Input of adjustment data of premeasured sensorsConfirm with enter

AUTOMANDAT

(AUTO: Buff er set) Select buffer set using keys (see buffer

tables for nominal values).

Confirm with enter

-00-...-09-,-U1- (see Appendix)

Pressing the info key dis-plays the manufacturer and nominal values in the lower line.

5

Page 50: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

50

Confi guration

1

SensorAdjust: calibration timer, calibration cycle

Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 keys,

press enter. Select 4 SENSOR menu using keys, press

enter. All items of this menu group are indicated by 5

the “SNS:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas key until the [meas] mode indicator is displayed.

2

4

Select sensor type

Select type of temp probe

Temperature unit

Temp detection during measurement

(Man. temperature)

Temp detection during calibration

(Man. temperature)

Calibration mode

(AUTO: Buffer set)

Calibration timer

Calibration cycle

Cleaning cycles

Cleaning cycle counter

Sterilization cycles

Sterilization cycle counter

5enter

enter

6

enter

5

enter

meas

3

Page 51: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

51

Confi gurationMenu item Action Choices Calibration timer Adjust CALTIMER using

keys:OFF: No timerADAPT: Maximum cal cycle (adjust in the next step)FIX: Fixed cal cycle (adjust in the next step)Confirm with enter

OFF/ADAPT/FIX

With ADAPT, the calibra-tion cycle is automatically reduced depending on the sensor load (high temperatures and pH values) and for digital sensors also depending on the sensor wear

Calibration cycle Only with FIX/ADAPT:Modify digit using , select next digit using

keys.Confirm with enter

0...9999 h

5

Note for the calibration timer:When Sensocheck has been activated in the Confi guration > Alarm menu, the expiration of the calibration interval is indicated by Sensoface:

Display Status

+

+

Over 80% of the calibration interval has al-ready past.

The calibration interval has been exceeded.

The calibration timer settings apply to both parameter sets A and B.

The time remaining until the next due calibration can be seen in the diagnostics menu (see Diagnostics chapter).

Page 52: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

52

Confi guration

1

SensorAdjust: CIP cleaning cycles, SIP sterilization cycles

Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 keys,

press enter. Select 4 SENSOR menu using keys, press

enter. All items of this menu group are indicated by 5

the “SNS:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas key until the [meas] mode indicator is displayed.

2

4

Select sensor type

Select type of temp probe

Temperature unit

Temp detection during measurement

(Man. temperature)

Temp detection during calibration

(Man. temperature)

Calibration mode

(AUTO: Buffer set)

Calibration timer

Calibration cycle

Cleaning cycles

Cleaning cycle counter

Sterilization cycles

Sterilization cycle counter

5enter

enter

6

enter

5

enter

meas

3

Page 53: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

53

Confi gurationMenu item Action Choices CIP / SIPThe following adjustments are possible for digital sensors (INDUCON, ISM-DIGITAL):Cleaning cycle counter

Select ON or OFF using keys.

Activates/deactivates log-ging in extended logbook and counters (if provided)

Confirm with enter

ON/OFF

Cleaning cycles Only with CIP COUNT ON:Enter value using

keys.

Confirm with enter

0...9999 CYC(0025 CYC)

Sterilization cycle counter

Select ON or OFF using keys.

Activates/deactivates log-ging in extended logbook and counters (if provided)

Confirm with enter

ON/OFF

Sterilization cycles Only with SIP COUNT ON:Enter value using

keys.

Confirm with enter

0...9999 CYC(0025 CYC)

The cleaning and sterilization cycles are counted to measure the load on the sensor. Suitable for biochemical applications (process temp approx. 0...50 °C, CIP temperature > 55 °C, SIP temperature > 115 °C).

5

Page 54: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

54

Process variableCurrent startCurrent endTime constant of output fi lterOutput current during error messageOutput current during HOLDOutput current for HOLD FIX

Confi guration Current Output 1 Process variable, current start, current end

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 keys,

press enter. Select 4 OUT1 menu using keys, press

enter. All items of this menu group are indicated by 5

the “OT1:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas key until the [meas] mode indicator is displayed.

2

4

5enter

enter

6

enter

5

enter

meas

3

Page 55: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

55

Confi gurationMenu item Action Choices Process variable Select using keys:

PH: pH valueORP: Redox potentialTMP: TemperatureConfirm with enter

PH/ORP/TMP

Current start Modify digit using , select next digit using

keys.

Confirm with enter

–2...16 pH (PH)–1999...1999 mV (ORP)–20...300 °C /–4...572 °F (TMP)

Current end Enter value using keys.

Confirm with enter

–2...16 pH (PH)–1999...1999 mV (ORP)–20...300 °C /–4...572 °F (TMP)

5

Assignment of Measured Values: Current Start and Current End

Output current

[pH]

20 4

14

020 4

7

5

[mA]

Example 1: Range pH 0...14

[mA]

Output current

75

Example 2: Range pH 5...7 Advantage: Higher resolution in range of interest[pH]

Page 56: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

56

Confi gurationCurrent Output 1Adjust time constant of output fi lter

Current rangeProcess variableCurrent startCurrent end Time constant of output fi lterOutput current during error messageOutput current during HOLDOutput current for HOLD FIX

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 keys,

press enter. Select 4 OUT1 menu using keys, press

enter. All items of this menu group are indicated by 5

the “OT1:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas key until the [meas] mode indicator is displayed.

2

4

5enter

enter

6

enter

5

enter

meas

3

Page 57: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

57

Confi gurationMenu item Action ChoicesTime constant of output fi lter

Enter value using keys.

Confirm with enter

0...120 SEC(0000 SEC)

Time Constant of Output FilterTo smoothen the current output, a low-pass fi lter with adjustable fi lter time constant can be switched on. When there is a jump at the input (100 %), the output level is at 63 % after the time constant has been reached. The time constant can be set from 0 to 120 sec. If the time constant is set to 0 sec, the current output directly follows the input. Please note: The fi lter only acts on the current output, not on the display!During HOLD the fi lter is not applied. This prevents a jump at the output.

Time constant 0...120 s

5

pH

0/4-20 mA

Time constant 0...120 s

Temp

Display

Page 58: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

58

Confi gurationCurrent Output 1 Output current during Error and HOLD

Current rangeProcess variableCurrent startCurrent endTime constant of output fi lterOutput current during error messageOutput current during HOLDOutput current for HOLD FIX

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 keys,

press enter. Select 4 OUT1 menu using keys, press

enter. All items of this menu group are indicated by 5

the “OT1:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas key until the [meas] mode indicator is displayed.

2

4

5enter

enter

6

enter

5

enter

meas

3

Page 59: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

59

Confi gurationMenu item Action ChoicesOutput current during error message

Select ON or OFF using keys.

Confirm with enter

ON/OFF

Output current during HOLD

LAST: During HOLD the last measured value is maintained at the output.FIX: During HOLD a value (to be entered) is main-tained at the output.Select using Confirm with enter

LAST/FIX

Output current for HOLD FIX

Only with FIX selected:Enter current which is to flow at the output during HOLDEnter value using

keys.

Confirm with enter

00.00...22.00 mA(21.00 mA)

HOLD active

21

4HOLD active

Output current[mA]

Output signal HOLDSetting FIX = 21.0 mA

Output signal HOLDSetting LAST

Output Signal During HOLD:

5

Page 60: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

60

Confi guration Current Output 2 Output current range, current start, current end

Process variableCurrent startCurrent endTime constant of output fi lterOutput current during error messageOutput current during HOLDOutput current for HOLD FIX

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 keys,

press enter. Select 4 OUT2 menu using keys, press

enter. All items of this menu group are indicated by 5

the “OT2:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas key until the [meas] mode indicator is displayed.

2

4

5enter

enter

6

enter

5

enter

meas

3

Page 61: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

61

Confi gurationMenu item Action Choices Process variable Select using keys:

PH: pH valueORP: Redox potentialTMP: TemperatureConfirm with enter

PH/ORP/TMP

.

.

.

All the following adjustments are made as for current output 1 (see there)!

5

Page 62: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

62

Confi guration

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 keys,

press enter. Select 4 CORRECTION menu using keys,

press enter. All items of this menu group are indicated by 5

the “COR:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas key until the [meas] mode indicator is displayed.

2

4

enter

enter

6

enter

5

enter

meas

Temperature compensation, process mediumCurrent input for ext. temperature measurementCurrent rangeCurrent startCurrent end

5

3

Temperature CompensationTC process medium, current input for temp measurement

Page 63: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

63

Confi gurationMenu item Action Choices Temp compensation, process medium

For pH measurement only: Enter temperature compensation of the process medium.Enter value using

keys.Confirm with enter

–19.99...+19.99 %/K

Current input for ext. temperature measurement

Only if enabled via TAN and selected during configuration (SENSOR).Select ON or OFF using

keys.

Confirm with enter

ON/OFF

Current range Select desired range using keys.

Confirm with enter

4-20 mA / 0-20 mA

Current start Modify digit using ,select next digit using

keys.

Confirm with enter

Input range:–20...200 °C /–4...392 °F

Current end Enter value using keys.

Confirm with enter

Input range:–20...200 °C /–4...392 °F

5

Page 64: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

64

Confi guration Alarm Alarm delay, Sensocheck

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 keys,

press enter. Select 4 ALARM menu using keys, press

enter. All items of this menu group are indicated by 5

the “ALA:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas key until the [meas] mode indicator is displayed.

2

4

enter

6

enter

enter

meas

3

enterDelay Sensocheck

5

5

Page 65: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

65

Error messages can be signaled by a 22 mA output current (see Error Messages and Confi guration of Output 1/Output 2).

The alarm delay time delays the color change of the display back-lighting to red and the 22 mA signal (if confi gured).

Confi gurationMenu item Action ChoicesAlarm delay Only with Alarm ON

selected:Enter value using

keys.Confirm with enter

0...600 SEC(010 SEC)

Sensocheck Select Sensocheck (continuous monitoring of glass and reference electrode)Select ON or OFF using

keys.Confirm with enter.(At the same time, Sensoface is activated. With OFF, Sensoface is also switched off.)

ON/OFF

5

Page 66: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

66

Confi guration Time and Date Tag Number

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set A using 3 keys,

press enter. 4 Press enter. Select 5 CLOCK or TAG using keys, press

enter. All items of this menu group are indicated by 6

the “CLK:” or “TAG” code. Press enter to select menu, edit with arrow keys (see next page).Confirm (and proceed) with enter.

End: Press 7 meas key until the [meas] mode indicator is displayed.

2

3

4

enter

6

enter

5

enter

meas

Time formatTimeDay and monthYear

Tag number

enter

5

Page 67: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

67

Time and DateControl of the calibration and cleaning cycles is based on the time and date of the integrated real-time clock. In measuring mode the time is shown in the lower display. When using digital sensors, the calibration data is written in the sensor head.In addition, the logbook entries (cf Diagnostics) are provided with a time stamp.

Please note:After prolonged power outage (> 5 days) the time display is • replaced by dashes and cannot be used for processing. Enter the correct time.There is no automatic switchover from winter to summer time!• Be sure to manually adjust the time!

Confi guration

Menu item Action ChoicesTag number Select character using

keys,select next digit using

keys.

Confirm with enter

A...Z, 0...9, – + < > ? / @

The first 10 characters are seen in the display with-out scrolling.

Tag Number (“ TAG“)You can enter a designation for the point of measurement (tag number) in the lower display line. Up to 32 digits are possible.Pressing meas (repeatedly) in the measuring mode indicates the tag number.Being part of the device confi guration, the “TAG“ can be read out via IrDA.A standardized tag number helps, for example, to correctly re-install a device after repair.

5

Page 68: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

68

Digital SensorsOperation

Stratos Pro can be operated with three diff erent types of digital sensors: Memosens, InduCon, ISM.The following display examples refer to a transmitter and a digital pH sensor (slight variations for other combinations).

The sensor type is selected during confi guration, the selected type is indicated by a display icon:

Indu Con

The device only switches to measuring mode when the connected sensor corresponds to the type confi gured (Sensoface is happy):

Otherwise, an error message is released. The info icon is displayed. You can display the error text in the bottom line using the keys. Sensoface is sad (see table of error messages and Sensoface in the Appendix):

Page 69: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

69

Connecting a Digital SensorStep Action/Display Remark

Connect sensor

Before a digital sensor is connected, the error message “No sensor“ is displayed.

Wait until the sensor data are displayed.

The hourglass in the display blinks.

Check sensor data

View sensor informa-tion using keys, confi rm with enter.

The display color changes to green, the InduCon or ISM icon is displayed, if applicable.

Sensoface is happy.

Go to measuring mode

Press meas, info, or enter

After 60 sec the device automatically returns to measuring mode (timeout).

Possible error messagesSensor devaluatedReplace sensor

When this error message appears, the sensor can-not be used any more. Sensoface is sad.

Sensor defectiveReplace sensor

When this error message appears, the sensor cannot be used.Sensoface is sad.

Digital Sensors

Page 70: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

70

Digital Sensors Sensor Replacement

A digital sensor should only be replaced during HOLD mode to pre-vent unintended reactions of the outputs or contacts. When you fi rst want to calibrate the new sensor, it can also be replaced in calibration mode.

Step Action/Display RemarkSelect HOLD mode Press any key to call

the selection menu, select HOLD using the keys, press enter to confi rm.

Now the device is in HOLD mode. The HOLD mode can also be acti-vated externally via the HOLD input.During HOLD the output current is frozen at its last value or set to a fixed value.

Disconnect and remove old sensorInstall and connect new sensor.

Temporary messages which are activated dur-ing the replacement are indicated but not output to the alarm contact and not entered in the log-book.

Wait until the sensor data are displayed.

Page 71: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

71

Digital Sensors

Step Action/Display RemarkCheck sensor data

View sensor informa-tion using keys, confi rm with enter.

You can view the sensor manufacturer and type, serial number, and last calibration date.

Check measured valuesExit HOLD Hit meas key:

Return to selection menu. Hold meas key depressed: Device switches to measuring mode

The sensor replacement is entered in the extended logbook.

Page 72: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

72

CalibrationPlease note:

All calibration procedures must be performed by trained person-• nel. Incorrectly set parameters may go unnoticed, but change the measuring properties.

The response time of the sensor and temperature probe is • considerably reduced when the sensor is fi rst moved about in the buff er solution and then held still.

The device can only operate properly when the buff er solutions • used correspond to the confi gured set. Other buff er solutions, even those with the same nominal values, may demonstrate a diff erent temperature response. This leads to measurement errors.

When using ISFET sensors or sensors with a zero point other than pH 7, the nominal zero point must be adjusted each time a new sensor is connected. This is important if you want to obtain reliable Sensoface messages. The Sensoface messages issued during all further calibrations are based on this basic calibration.

Page 73: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

73

Calibration is used to adapt the device to the individual sensor characteristics, namely asymmetry potential and slope. Access to calibration can be protected with a passcode (SERVICE menu).First, you open the calibration menu and select the calibration mode:CAL_PH Depending on confi guation setting:

AUTO Automatic buff er recognition (Calimatic)MAN Manual buff er inputDAT Input of premeasured electrode data

CAL_ORP ORP calibrationP_CAL Product calibration (calibration with sampling)ISFET-ZERO Zero adjustment. Required for ISFET sensors, subse-

quently you can conduct either a one or a two-point calibration.

CAL_RTD Temperature probe adjustment

Selecting a Calibration Mode

To preset CAL_PH (CONF menu / confi guration):Hold 1) meas key depressed (> 2 s) (measuring mode)Press any arrow key: the selection menu appears2) Select CONF mode using left / right arrow key3) Select “SENSOR“ – “CALMODE“: AUTO, MAN, or DAT.4) Press enter to confi rm

Selection menu

Selected mode (blinks)

1 2

3

4

Page 74: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

74

Zero Adjustment (ISFET)This adjustment allows the use of ISFET sensors with diff ering nominal zero (pH only). The function is available when Sensor selection = ISFET has been set during confi guration. Zero adjustment is disabled for any other sensors.The adjustment is made using a zero buff er (pH 7.00). Permitted range for buff er value: pH 6.5 ... 7.5. Temperature-corrected input. Maximum zero off set: ± 200 mV.

Display Action RemarkSelect Calibration.Press enter to proceed.

Ready for calibration.Hourglass blinks.

Display (3 sec)Now the device is in HOLD mode.

Immerse sensor in a pH 7.00 buff er. Enter the temperature-corrected pH value in the range 6.50 to 7.50 using the arrow keys (see buff er table).Confi rm with enter.

If the zero off set of the sensor is too large (> ± 200 mV), a CAL ERR error message is gener-ated. In that case the electrode cannot be calibrated.

Stability check.The measured value [mV] is displayed.The “hourglass” icon is blinking.

Please note: Stability check can be stopped (by pressing enter). However, this re-duces calibration accuracy.

Zero adjustment

Page 75: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

75

Display Action RemarkAt the end of the ad-justment procedure the zero off set [mV] of the sensor is displayed (based on 25 °C).Sensoface is active.Press enter to proceed.

This is not the fi nal calibration value of the sensor! Asym-metry potential and slope must be determined with a complete 2-point calibration.

Use the arrow keys to select:

Repeat • (repeat calibration) or Measuring. •

Confi rm with enter.Place sensor in process.End zero calibration with enter.

After end of calibra-tion, the outputs re-main in HOLD mode for a short time.

Note for Zero AdjustmentAfter having adjusted the zero off set, be sure to calibrate the sensor following one of the procedures as described on the next pages.

Zero Adjustment (ISFET)

Page 76: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

76

Adjustment (Calimatic)

Automatic Calibration ( Calimatic)The AUTO calibration mode and the type of temperature detection are selected during confi guration. Make sure that the buff er solutions used correspond to the confi gured buff er set. Other buff er solutions, even those with the same nominal values, may demonstrate a diff er-ent temperature response. This leads to measurement errors.

Display Action RemarkSelect Calibration.Press enter to proceed.

Ready for calibration.Hourglass blinks. Select calibration method: CAL_PH Press enter to proceed.

Display (3 sec)Now the device is in HOLD mode.

Remove the sensor and temperature probe, clean them, and im-merse them in the fi rst buff er solution (in any order). Start with enter

When manual input of temperature has been confi gured, the temp value in the display blinks and can be edited using the arrow keys.

Buff er recognition.While the “hourglass” icon is blinking, the sensor and temperature probe remain in the fi rst buff er solution.

The response time of the sensor and temperature probe is considerably reduced when the sensor is fi rst moved about in the buff er solution and then held still.

Buff er recognition terminated, the nomi-nal buff er value is displayed.

Page 77: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

77

Display Action RemarkAt the end of the stabil-ity check, the value will be saved and the asym-metry potential will be displayed.Calibration with the fi rst buff er is terminated. Remove the sensor and temp probe from the fi rst buff er solution and rinse them thoroughly.Use the arrow keys to select:

END (1-point cal)• CAL2 (2-point cal)• REPEAT•

Press enter to proceed.

Please note:Stability check can be stopped after 10 sec (by pressing enter). However, this reduces calibration accuracy. Display for 1-point cal:

Sensoface is active. End with enter

2-point calibration:Immerse sensor and temperature probe in the second buff er solu-tion. Start with enter

The calibration pro-cess runs as for the fi rst buff er.

Retract sensor and temp probe out of second buff er, rinse off , re-install.Press enter to proceed.

The slope and asym-metry potential of the sensor (based on 25 °C) are displayed.

Use the arrow keys to select:

MEAS (end)• REPEAT•

Press enter to proceed.End: HOLD is deacti-vated with delay.

When 2-point cal is ended:

Automatic Calibration (Calimatic)

Page 78: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

78

Manual Calibration with Buffer EntryThe MAN calibration mode and the type of temperature detection are selected during confi guration. For calibration with manual buff er specifi cation, you must enter the pH value of the buff er solution used in the device for the proper temperature. Any desired buff er solution can be used for calibration.

Display Action RemarkSelect Calibration.Press enter to proceed.

Ready for calibration.Hourglass blinks.

Display (3 sec)Now the device is in HOLD mode.

Remove the sensor and temperature probe, clean them, and im-merse them in the fi rst buff er solution. Press enter to start.

When manual input of temperature has been confi gured, the temp value in the display blinks and can be edited using the arrow keys.

Enter the pH value of your buff er solution for the proper temperature. While the “hourglass” icon is blinking, the sensor and temperature probe remain in the buff er solution.

The response time of the sensor and temperature probe is considerably reduced when the sensor is fi rst moved about in the buff er solution and then held still.

Page 79: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

79

Display Action RemarkAt the end of the stabil-ity check, the value will be saved and the asym-metry potential will be displayed.Calibration with the fi rst buff er is terminated. Remove the sensor and temp probe from the fi rst buff er solution and rinse them thoroughly.Use the arrow keys to select:

END (1-point cal)• CAL2 (2-point cal)• REPEAT•

Press enter to proceed.

Please note:Stability check can be stopped after 10 sec (by pressing enter). However, this reduces calibration accuracy. Display for 1-point cal:

Sensoface is active. End with enter

2-point calibration:Immerse sensor and temperature probe in the second buff er solution.Enter pH value. Press enter to start.

The calibration pro-cess runs as for the fi rst buff er.

Rinse sensor and tem-perature probe and reinstall them.Press enter to proceed.

Display of slope and new asymmetry potential (based on 25 °C).

Use the arrow keys to select:

MEAS (end)• REPEAT•

Press enter to proceed.End: HOLD is deacti-vated with delay.

When 2-point cal is ended:

Manual Calibration with Buff er Entry

Page 80: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

80

Data Entry of Premeasured SensorsThe DAT calibration mode must have been preset during confi guration. You can directly enter the values for slope and asymmetry potential of a sensor. The values must be known, e.g. determined beforehand in the laboratory.

Display Action RemarkSelect Calibration.Press enter to proceed.

“Data Input“Ready for calibration.Hourglass blinks.

Display (3 sec)Now the device is in HOLD mode.

Enter asymmetry potential [mV].Press enter to proceed.

Enter slope [%].

The device displays the new slope and asymmetry potential (at 25 °C).Sensoface is active.Use the arrow keys to select:

MEAS (end)• REPEAT•

Press enter to proceed.

End: HOLD is deactivated with delay.

Page 81: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

81

Converting Slope to mV

Converting slope [%] to slope [mV/pH] at 25 °C

% mV/pH78 46,280 47,482 48,584 49,786 50,988 52,190 53,392 54,594 55,696 56,898 58,0100 59,2102 60,4

Converting asymmetry potential to sensor zero point

= Sensor zero

= Asymmetry potential

= Slope

ZERO = 7 - ZERO

VAS

S

VAS

[mV]

S [mV / pH]

Page 82: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

82

Product calibration

Product Calibration (pH)Calibration by sampling (one-point calibration). During product calibration the sensor remains in the process. The measurement process is only interrupted briefl y.

Procedure: The sample is measured in the lab or directly on the site using a por-1) table meter. To ensure an exact calibration, the sample temperature should correspond to the measured process temperature. During sampling the device saves the currently measured value and then returns to measuring mode. The “calibration“ mode indicator blinks.In the second step you enter the measured sample value in the 2) device. From the diff erence between the stored measured value and entered sample value, the device calculates the new asymmetry potential.

If the sample is invalid, you can take over the value stored during sampling. In that case the old calibration values are stored. Afterwards, you can start a new product calibration.

Display Action Remark Select product calibration: P_CAL.Press enter to proceed.

Ready for calibration.Hourglass blinks.

Display (3 sec)Now the device is in HOLD mode.

Take sample and save value. Press enter to proceed.

Now the sample can be measured.

Page 83: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

83

Display Action RemarkThe device returns to measuring mode.

From the blinking CAL mode indicator you see that product calibration has not been terminated.

Product calibration step 2

Display (3 sec)Now the device is in HOLD mode.

The stored value is displayed (blinking) and can be overwritten with the measured sample value.Press enter to proceed.Display of new asym-metry potential (based on 25°C).Sensoface is active.To end calibration:Select MEAS, then enter

To repeat calibra-tion: Select REPEAT, then enter

End of calibration After end of calibration, the outputs remain in HOLD mode for a short time.

Product Calibration (pH)

Page 84: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

84

ORP ( Redox) CalibrationThe potential of a redox sensor is calibrated using a redox (ORP) buff er solution. In the course of that, the diff erence between the measured potential and the potential of the calibration solution is determined according to the following equation. During measurement this diff er-ence is added to the measured potential.

= displayed ORP

= direct sensor potential

= delta value, determined during calibration

mVORP

= mVmeas

+ ΔmV mVORP

mVmeas

ΔmV

The sensor potential can also be related to another reference system – e.g. the standard hydrogen electrode. In that case the temperature-corrected potential (see table) of the reference electrode used must be entered during calibration. During measurement, this value is then added to the ORP measured. Please make sure that measurement and calibration temperature are the same since the temperature behavior of the reference electrode is not automatically taken into account.

0102025304050607080

Temperature[°C]

Ag/AgCl/KCl1 mol/l[∆mV]

Ag/AgCl/KCl3 mol/l[∆mV]

Thalamid[∆mV]

Mercurysulfate[∆mV]

249244240236233227221214207200

224217211207203196188180172163

-559-564-569-571-574-580-585-592-598-605

672664655651647639631623613603

Temperature Dependence of Commonly Used Reference Systems Measured Against SHE

Page 85: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

85

ORP CalibrationDisplay Action Remark

Select ORP calibration, proceed with enter

Remove the sensor and temperature probe, clean them, and im-merse them in the redox buff er.

Display (3 sec)Now the device is in HOLD mode.

Enter setpoint value for redox buff er.Press enter to proceed.

The ORP delta value is displayed (based on 25°C).Sensoface is active.Press enter to proceed.

To repeat calibration: Select REPEAT. To end calibration:Select MEAS, then press enter

After end of calibra-tion, the outputs re-main in HOLD mode for a short time.

Page 86: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

86

Temp Probe AdjustmentDisplay Action Remark

Select temp adjust-ment. Press enter to proceed.

Wrong settings change the mea-surement proper-ties!

Measure the tempera-ture of the process me-dium using an external thermometer.

Display (3 sec)Now the device is in HOLD mode.

Enter the measured temperature value.Maximum diff erence: 10 K.Press enter to proceed.

Display of actual temperature (un-compensated) in the lower display.

The corrected tempera-ture value is displayed. Sensoface is active.To end calibration: Select MEAS, then enterTo repeat calibration: Select REPEAT, then enter After calibration is ended, the device will switch to measuring mode.

After end of calibra-tion, the outputs re-main in HOLD mode for a short time.

Temp probe adjustment

Page 87: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

87

MeasurementDisplay Remark

or AM/PM and °F:

From the confi guration or calibration menus, you can switch the device to measuring mode by pressing the meas key.In the measuring mode the main display shows the confi gured process variable (pH, ORP [mV], or temperature), the second-ary display shows the time and the second confi gured process variable (pH, ORP [mV], or temperature). The [meas] mode indicator lights and the active parameter set (A/B) is indicated.

By pressing the enter key you can briefl y display the output currents.By pressing the meas key you can step through the following displays. When no key has been pressed for 60 sec, the device returns to the standard display.

Selecting the1) parameter set (if set to “manual“ in the confi guration). Select the desired parameter set using the arrow keys (PARSET A or PARSET B blinks in the lower display line). Confi rm with enter.

Further displays (each with meas).

Display of tag number (“TAG“)2) Display of 3) time and date

Page 88: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

88

DiagnosticsIn the Diagnostics mode you can access the following menus without interrupting the measurement:CALDATA Viewing the calibration dataSENSOR Viewing the sensor dataSELFTEST Starting a device self-testLOGBOOK Viewing the logbook entriesMONITOR Displaying currently measured valuesVERSION Displaying device type, software version, serial number

Access to diagnostics can be protected with a passcode (SERVICE menu).

Please note:HOLD is not active during Diagnostics mode!

Action Key RemarkActivate Diagnostics

Press any arrow key to call the selection menu.(Display color changes to turquoise.)Select DIAG using keys, confirm with enter

Select diagnostics option

Use keys to select from:CALDATA SENSOR SELFTEST LOGBOOK MONITOR VERSIONSee next pages for further proceeding.

End meas End with meas.

Diagnostics (DIAG)

Page 89: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

89

DiagnosticsDisplay Menu item

Display of calibration dataSelect CALDATA using , confirm with enter.Use the keys to select the desired parameter from the bottom line of the display (LAST_CAL ISFET-ZERO ZERO SLOPE NEXT_CAL).The selected parameter is shown in the main display.

Press meas to return to measurement.

Display of sensor dataFor analog sensors, the type is displayed (STANDARD / ISFET). Not applicable for digital transmitters (-MSPH).For digital sensors, the manufacturer, type, serial number, and last calibration date is displayed.In each case Sensoface is active.

Display data using keys,return with enter or meas.

Page 90: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

90

DiagnosticsDisplay Menu item

Device self-test(To abort, you can press meas.)

1) Display test: Display of all segments with changing background colors white/green/red.Proceed with enter

2) RAM test: Hourglass blinks, then display of --PASS-- or --FAIL-- Proceed with enter

3) EEPROM test: Hourglass blinks, then display of --PASS-- or --FAIL-- Proceed with enter

4) FLASH test: Hourglass blinks, then display of --PASS-- or --FAIL-- Proceed with enter

5) Module test: Hourglass blinks, then display of --PASS-- or --FAIL-- Press enter or meas to return to measuring mode.

Page 91: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

91

DiagnosticsDisplay Menu item

Display of logbook entriesSelect LOGBOOK using , confirm with enter.

With the keys, you can scroll backwards and forwards through the logbook (entries -00-...-99-), -00- being the last entry.

If the display is set to date/time, you can search for a particular date using the keys.Press to view the corresponding message text.

If the display is set to the message text, you can search for a particular message using the keys.Press to display the date and time.

Press meas to return to measurement.

Extended logbook / Audit Trail (via TAN)With the keys, you can scroll backwards and forwards through the extended logbook (entries -000-...-199-), -000- being the last entry.Display: CFRAudit Trail also records function activations (CAL CONFIG SERVICE), some Sensoface messages (cal timer, wear), and opening of the enclosure.

Page 92: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

92

DiagnosticsDisplay Menu item

Display example:

Display of currently measured values ( sensor monitor)Select MONITOR using , confirm with enter. Use the keys to select the desired parameter from the bottom line of the display: mV_PH mV_ORP RTD R_GLASS R_REF I-INPUT (for digital sensors also: OPERATION TIME SENSOR WEAR LIFETIME CIP SIP AUTOCLAVE). The selected parameter is shown in the main display.Press meas to return to measurement.

Display mV_pH(for validation, sensor can be immersed in a calibra-tion solution, for example, or the device is checked by using a simulator)

Display of remaining dynamic lifetime(only for digital sensors, however not for MEMOSENS)

Display of sensor operating time(for digital sensors only)

Display of sensor wear(only for digital sensors of the InduCon type)

VersionDisplay of device type, software/hardware version, and serial number for all device components.Use the keys to switch between software and hardware version. Press enter to proceed to next device component.

Page 93: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

93

ServiceService (SERVICE) A

The Service mode allows displaying the currently measured values with the sensor monitor• performing a device sef-test• testing the two current outputs• activating and communicating via the IrDA interface• assigning and editing passcodes• resetting the device to factory settings• enabling options via TAN.•

Please note:HOLD is active during Service mode!

Action Key/Display RemarkActivate Service

Press any arrow key to call the selection menu.(Display color changes to green)Select SERVICE using keys, confirm with enter

Passcode Enter passcode “5555“ for service mode using the keys.

Confirm with enter.

Display In service mode the following icons are displayed:

[diag] mode indicator• HOLD triangle• Service (wrench)•

End meas End with meas.

Page 94: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

94

ServiceMenu item Remark

Display example:

Display of currently measured values ( sensor monitor) with HOLD mode activated:Select MONITOR using , confirm with enter.Select variable in the bottom text line using .

The selected parameter is shown in the main display.As the device is in HOLD mode, you can perform validations using simulators without influencing the signal outputs.

Press meas to return to the service menu.Return to measurement: Press meas once more.

Specify current at outputs 1 and 2:Select OUT1 or OUT2 using the keys, confirm with enter.Enter a valid current value for the respective output using keys.Confirm with enter.For checking purposes, the actual output current is shown in the bottom right corner of the display.End with enter or meas.

Page 95: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

95

ServiceMenu item Remark

IrDA communication:Select IRDA using , confirm with enter.

When IrDA communication is active, the device remains in the HOLD mode for reasons of safety. Further operation is performed via IrDA.

End communication with meas.

Exception: Firmware update (must not be interrupted!)

Assigning passcodes:In the “SERVICE - CODES“ menu you can assign pass-codes to DIAG, HOLD, CAL, CONF, and SERVICE modes (Service preset to 5555).When you have lost the Service passcode, you have to request an “ Ambulance TAN“ from the manufac-turer specifying the serial number of your device. To enter the “Ambulance TAN“, call the Service func-tion and enter passcode 7321. After correct input of the ambulance TAN the device signals “PASS“ for 4 sec and resets the Service passcode to 5555.

Reset to factory settings:In the “SERVICE - DEFAULT“ menu you can reset the device to factory settings.Not affected: calibration data

Release of options: Options come with a “transaction number“ ( TAN). This TAN must be entered and confirmed with enter to release the option.

Page 96: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

96

Operating States

Operating status

OU

T 1

OU

T 2

Tim

e ou

t

Measuring -

DIAG 60 s

CAL No

CONF 20 min

SERVICE 20 min

SERVICE OUT 1 20 min

SERVICE OUT 2 20 min

HOLD No

Explanation: as configured (Last/Fix or Last/Off )

active

manual

Page 97: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

97

Product Line and Accessories Order Code Stratos Pro A 2...

TANExample A 2 1 1 X - PH - 1

2-wire / 4-20 mA A 2 B,C,E

CommunicationWithout (HART retrofittable via TAN) 0 A

Version numberVersion 1

ApprovalsGeneral Safety NATEX / IECEX Zone 2 BATEX / IECEX / FM / CSA Zone 1 / Cl 1 Div 1 XOther approvals Z

Measuring channelMemosens pH / Redox digital MSPHMemosens Cond digital MSCONDMemosens Oxy digital MSOXYDual COND (2x2-electrode sensors, analog) N CCpH / ORP value(ISM digital per TAN)

Measuring module PH F

Cond, 2-/4-electrode Measuring module CONDConductivity, electrodeless Measuring module CONDIOxygen (ISM digital and traces per TAN)

Measuring module OXY D, F

OptionsWithout 2nd current output 0With 2nd current output 1

TAN optionsHART SW-A001 (A)Logbook SW-A002 (B)Extended logbook (Audit Trail) SW-A003 (C)Trace oxygen measurement SW-A004 (D)Current input + 2 digital inputs SW-A005 (E)ISM digital SW-A006 (F)

Mounting accessoriesPipe-mount kit ZU 0274Protective hood ZU 0737Panel-mount kit ZU 0738

Page 98: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

98

Recommended Power Supply Units: Order No.:

Repeater power supply, IS, 24 V AC/DC,output 0/4...20 mA

WG 20 A2

Repeater power supply, IS, 90...253 V AC, output 4...20 mA

WG 21 A7

Repeater power supply, IS, 90...253 V AC, HART, output 4...20 mA

WG 21 A7 Opt. 470

Repeater power supply, IS, 24 V AC/DC,output 4...20 mA

WG 21 A7 Opt. 336

Repeater power supply, IS, 24 V AC/DC, HART, output 4...20 mA

WG 21 A7 Opt. 336, 470

Repeater power supply, non-IS, 24 V DC,output 4...20 mA

IsoAmp PWR B 10116

Repeater power supply, non-IS, 24 V DC, HART, output 0/4...20 mA / 0...10 V

IsoAmp PWR A 20100

A201/A211X: Supply Units and Connection

Connection to Supply Units

8

9

17

+ out

- out 1

- out 2

+

-

+

-

Supply unit 1

Supply unit 2

Stratos ProA 2...X

Hazardous-area requirements for cable:Between the wires:Solid insulation ≥ 0.5 mmTest voltage ≥ 500 V

Page 99: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

99

SpecificationspH/mV input Input for pH or ORP sensors or ISFET

Input Glass electrode or ISFET

Input Reference electrode

Input ORP electrode (e.g. platinum) or auxiliary electrode for impedance measurement

Measuring range (MR) -1500 ... +1500 mV

Display range pH value -2.00 ... 16.00

ORP -1999 ... +1999 mV

Glass electrode input4) Input resistance > 1 x 1012 Ω

Input current < 1 x 10-12 A 2)

Impedance range 0.5 ... 1000 MΩ (± 20%)

Reference electrode input4) Input resistance > 1 x 1010 Ω

Input current < 1 x 10-10 A 2)

Impedance range 0.5 ... 200 kΩ (± 20%)

Meas. error1,2,3) pH value < 0.02 TC: 0.002 pH/K

(Display) mV value < 1 mV TC: 0.1 mV/K

pH sensor standardization * pH calibration

Operating modes BUF Calibration with Calimatic automatic buff er recognition

MAN Manual calibration with input of individual buff er values

DAT Data entry of pre-measured electrodes

Product calibration

Calimatic buff er sets * -00- Knick 2.00/4.01/7.00/9.21

-01- Mettler-Toledo 2.00/4.01/7.00/9.21

-02- Merck/Riedel 2.00/4.00/7.00/9.00/12.00

-03- Ciba (94) 2.06/4.00/7.00/10.00

-04- NIST technical 1.68/4.00/7.00/10.01/12.46

-05- NIST standard 1.679/4.006/6.865/9.180

-06- HACH 4.00/7.00/10.01

-07- WTW techn. buff ers 2.00/4.01/7.00/10.00

-08- Hamilton 4.01/7.00/10.01

-09- Reagecon 2.00/4.00/7.00/9.00/12.00

-U1- Specifi able buff er set

Page 100: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

100

Specifi cations

Zero adjustment ± 200 mV (for ISFET)

Max. calibration range Asymmetry potential ±60 mV

Slope 80 ... 103 % (47.5 ... 61 mV/pH)

(possibly restricting notes from Sensoface)

ORP sensor standardization* ORP calibration (zero adjustment)

Max. calibration range -700 ... +700 ΔmV

Adaptive cal timer * Interval 0000 ... 9999 h

(German patent DE 101 41 408)

Sensocheck Automatic monitoring of glass and reference electrode, can be disabled

Delay Approx. 30 s

Sensoface Provides information on the sensor condition, evaluation of zero/slope, response time, calibration interval, wear, Sensocheck, can be switched off

Temperature input Pt100 / Pt1000 / NTC 30 kΩ *

2-wire connection, adjustable

Measuring range Pt 100/Pt 1000 -20.0 ... +200.0 °C / -4 ... +392 °F

NTC 30 kΩ -20.0 ... +150.0 °C / -4 ... +302 °F

Adjustment range 10 K

Resolution 0.1 °C / 1 °F

Meas. error1,2,3) < 0.5 K (< 1K for Pt100; < 1K for NTC > 100 °C)

Temperature compensation of process medium

Linear -19.99 ... +19,99 %/K

Reference temp 25 °C

ISM input “One wire“ interface for operation with ISM (digital sensors)

(6 V / Ri= approx. 1.2 kΩ)

I input Current input 0/4 ... 20 mA / 50 Ω for external temperature signal

Start/end of scale Confi gurable within the measurement range for °C (°F)

Characteristic Linear

Measurement error 1.3) < 1% current value + 0.1 mA

Page 101: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

101

Specifi cations

HOLD input Galvanically separated (OPTO coupler)

Function Switches device to HOLD mode

Switching voltage 0 ... 2 V (AC/DC) Inactive

10 ... 30 V (AC/DC) HOLD active

CONTROL input Galvanically separated (OPTO coupler)

Function Selecting parameter set A/B

Switching voltage 0 ... 2 V (AC/DC) Parameter set A

10 ... 30 V (AC/DC) Parameter set B

Output 1 Current loop 4 ... 20 mA, fl oating, protected against inverse polarity

HART communication

Supply voltage 14 ... 30 V

Process variable* pH, ORP, or temperature

Characteristic Linear

Overrange * 22 mA in the case of error messages

Output fi lter * PT1 fi lter, time constant 0 ... 120 s

Measurement error 1) < 0.25 % current value + 0.025 mA

Start/end of scale * Confi gurable within the measurement ranges for pH, mV, °C, °F

Min. span pH 2.00 / 200 mV / 20 K / 36 °F

Output 2 Current loop 4 ... 20 mA, fl oating, protected against inverse polarity

Supply voltage 14 ... 30 V

Process variable* pH, ORP, or temperature

Characteristic Linear

Overrange * 22 mA in the case of error messages

Output fi lter * PT1 fi lter, time constant 0 ... 120 s

Measurement error 1) < 0.25 % current value + 0.05 mA

Start/end of scale * Confi gurable within the measurement ranges for pH, mV, °C, °F

Min. span pH 2.00 / 200 mV / 20 K / 36 °F

Page 102: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

102

Specifi cations

Power output for operating an ISFET adapter

+3 V / 0,5 mA

–3 V / 0.5 mA

Real-time clock Diff erent time and date formats selectable

Power reserve > 5 days

Display LC display, 7-segment with icons

Main display Character height approx. 22 mm, unit symbols approx. 14 mm

Secondary display Character height approx. 10 mm

Text line 14 characters, 14 segments

Sensoface 3 status indicators (friendly, neutral, sad face)

Mode indicators meas, cal, conf, diag

Further icons for confi guration and messages

Alarm indication Red backlighting in case of alarm

Keypad Keys: meas, info, 4 cursor keys, enter

HART communication HART Version 6Digital communication by FSK modulation of output current 1

Device identifi cation, measured values, status and messages, parameter setting, calibration, records

IrDA interface Infrared interface for transmission of records and logbook, parameter setting, calibration, fi rmware update

FDA 21 CFR Part 11 Access control by editable passcodesLogbook entry and fl ag via HART in the case of confi guration changesMessage and logbook entry when enclosure is opened

Diagnostics Functions

Calibration data Calibration date, zero, slope, response time

Device self-test Displaytest, automatic memory test (RAM, FLASH, EEPROM), module test

Logbook 100 events with date and time

Extended logbook (TAN) AuditTrail: 200 events with date and time

Page 103: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

103

Specifi cations

Service functions

Sensor monitor Display of direct sensor signals (mV/temperature/resistance ...)

Current source Current specifi able for output 1 and 2 (00.00 ... 22.00 mA)

IrDA Activating the IrDA function

Passcodes Assigning passcodes for menu access

Factory setting Resetting all parameters to factory settingException: Calibration data

TAN Enabling optionally available additional functions

Data retention Parameters, calibration data, logbook > 10 years (EEPROM)

EMC EN 61326-1 (General Requirements)

Emitted interference Class B (residential area)

Immunity to interference IndustryEN 61326-2-3 (Particular Requirements for Transmitters)

Explosion protection Stratos Pro A201X/A211X PH(coming soon)

Europe: ATEX Zone 0,1,2

USA: FM Cl I Div 1,2 / Zone 0,1,2

Canada: CSA Cl I Div 1,2 / Zone 0,1,2

International: IECEx

Nominal operating conditions

Ambient temperature –20 ... +65 °C

Transport/Storage temperature –20 ... +70 °C

Relative humidity 10 ... 95% not condensing

Supply voltage 14 ... 30 V

Enclosure Molded enclosure made of PBT/PC, glass reinforced

Fastening Wall, pipe/post, or panel mounting

Color Gray, RAL 7001

Ingress protection IP 67

Flammability UL 94 V-0

Page 104: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

104

Specifi cations

Dimensions 148 mm x 148 mm

Control panel cutout 138 mm x 138 mm to DIN 43 700

Weight 1.2 kg (1.6 kg incl. accessories and packaging)

Cable glands 3 knockouts for M20 x 1.5 cable glands

2 knockouts for NPT ½” or rigid metallic conduit

Connections Terminals, conductor cross section max. 2.5 mm2

* User-defi ned

1) Acc. to EN 60746-1, at nominal operating conditions

2) ± 1 count

3) Plus sensor error

4) At room temperature

Page 105: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

105

-00- Knick technical buff ers (correspond to-01- Mettler-Toledo technical buff ers)

°C pH 0 5101520253035404550556065707580859095

2.032.022.012.002.002.001.991.991.981.981.981.981.981.991.992.002.002.002.002.00

4.014.014.004.004.004.014.014.024.034.044.064.084.104.134.164.194.224.264.304.35

7.127.097.067.047.027.006.996.986.976.976.976.986.986.997.007.027.047.067.097.12

9.529.459.389.329.269.219.169.119.069.038.998.968.938.908.888.858.838.818.798.77

Buffer Tables

Page 106: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

106

-02- Merck Titrisols, Riedel-de-Haen Fixanals

°C pH 0 5101520253035404550556065707580859095

2.012.012.012.002.002.002.002.002.002.002.002.002.002.002.012.012.012.012.012.01

4.054.044.024.014.004.014.014.014.014.014.004.004.004.004.004.004.004.004.004.00

7.137.077.057.027.006.986.986.966.956.956.956.956.966.966.966.966.976.987.007.02

9.249.169.119.059.008.958.918.888.858.828.798.768.738.728.708.688.668.658.648.64

12.5812.4112.2612.1012.0011.8811.7211.6711.5411.4411.3311.1911.0410.9710.9010.8010.7010.5910.4810.37

Buff er Tables

Page 107: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

107

-03- Ciba (94) buff ers Nominal values: 2.06 4.00 7.00 10.00

°C pH 0 5101520253035404550556065707580859095

2.042.092.072.082.092.082.062.062.072.062.062.052.082.07*2.072.04*2.022.03*2.042.05*

4.004.024.004.004.014.024.004.014.024.034.044.054.104.10*4.114.13*4.154.17*4.204.22*

7.107.087.057.026.986.986.966.956.946.936.936.916.936.92*6.926.92*6.936.95*6.976.99*

10.3010.2110.1410.06 9.99 9.95 9.89 9.85 9.81 9.77 9.73 9.68 9.66 9.61* 9.57 9.54* 9.52 9.47* 9.43 9.38*

* extrapolated

Buff er Tables

Page 108: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

108

Buff er Tables

-04- NIST technical buff ers

°C pH 0 5101520253035404550556065707580859095

1.671.671.671.671.6751.681.681.691.691.701.7051.7151.721.731.741.751.7651.781.791.805

4.004.004.004.004.004.0054.0154.0254.034.0454.064.0754.0854.104.134.144.164.184.214.23

7.1157.0857.067.047.0157.006.9856.986.9756.9756.976.976.976.986.997.017.037.057.087.11

10.3210.2510.1810.1210.0610.01 9.97 9.93 9.89 9.86 9.83 9.83* 9.83* 9.83* 9.83* 9.83* 9.83* 9.83* 9.83* 9.83*

13.4213.2113.0112.8012.6412.4612.3012.1311.9911.8411.7111.5711.4511.45*11.45*11.45*11.45*11.45*11.45*11.45*

* Values complemented

Page 109: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

109

Buff er Tables

-05- NIST standard buff ers NIST Standard (DIN 19266 : 2000-01)

°C pH 0 51015202530374045505560

70

80

9095

1.6681.6701.6721.6761.6801.6851.6941.6971.7041.7121.7151.723

1.743

1.766

1.7921.806

4.0044.0014.0014.0034.0084.0154.0284.0364.0494.0644.0754.091

4.126

4.164

4.2054.227

6.9506.9226.9006.8806.8656.8536.8416.8376.8346.8336.8346.836

6.845

6.859

6.8776.886

9.3929.3319.2779.2289.1849.1449.0959.0769.0469.0189.9858.962

8.921

8.885

8.8508.833

Please note:The actual pH values of the individual batches of the reference materials are documented in a certifi cate of an accredited laboratory. This certifi cate is supplied with the respective buff ers. Only these pH(S) values shall be used as standard values for the secondary refer-ence buff er materials. Correspondingly, this standard does not include a table with standard pH values for practical use. The table above only provides examples of pH(PS) values for orientation.

Page 110: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

110

Buff er Tables

-06- HACH buff ers Nominal values: 4.01 7.000 10.01 (± 0.02 at 25 °C)

°C pH 0 5101520253035404550556065707580859095

4.004.004.004.004.004.014.014.024.034.054.064.074.094.10*4.12*4.14*4.16*4.18*4.21*4.24*

7.1187.0877.0597.0367.0167.0006.9876.9776.9706.9656.9646.9656.9686.98*7.00*7.02*7.04*7.06*7.09*7.12*

10.3010.2310.1710.1110.0510.01 9.96 9.92 9.88 9.85 9.82 9.79 9.76 9.71* 9.66* 9.63* 9.59* 9.56* 9.52* 9.48*

* Values complemented

Page 111: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

111

-07- WTW technical buff ers

°C pH 0 5101520253035404550556065707580859095

2.032.022.012.002.002.001.991.991.981.981.981.981.981.992.002.002.002.002.002.00

4.014.014.004.004.004.014.014.024.034.044.064.084.104.134.164.194.224.264.304.35

7.127.097.067.047.027.006.996.986.976.976.976.986.986.997.007.027.047.067.097.12

10.6510.5210.3910.2610.1310.00 9.87 9.74 9.61 9.48 9.35

Buff er Tables

Page 112: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

112

-08- Hamilton Duracal buff ers

°C pH 0 5101520253035404550556065707580859095

4.014.014.004.004.004.014.014.024.034.044.054.064.084.10*4.12*4.14*4.16*4.18*4.21*4.24*

7.127.097.067.047.027.006.996.986.976.976.976.986.986.99*7.00*7.02*7.04*7.06*7.09*7.12*

10.2310.1910.1510.1110.0610.01 9.97 9.92 9.86 9.83 9.79 9.75 9.72 9.69* 9.66* 9.59* 9.59* 9.56* 9.52* 9.48*

* Values complemented

Buff er Tables

Page 113: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

113

-09- Reagecon buff ers

°C pH0°C *2.01 *4.01 *7.07 *9.18 *12.545°C *2.01 *4.01 *7.07 *9.18 *12.5410°C 2.01 4.00 7.07 9.18 12.5415°C 2.01 4.00 7.04 9.12 12.3620°C 2.01 4.00 7.02 9.06 12.1725°C 2.00 4.00 7.00 9.00 12.0030°C 1.99 4.01 6.99 8.95 11.8135°C 2.00 4.02 6.98 8.90 11.6340°C 2.01 4.03 6.97 8.86 11.4745°C 2.01 4.04 6.97 8.83 11.3950°C 2.00 4.05 6.96 8.79 11.3055°C 2.00 4.07 6.96 8.77 11.1360°C 2.00 4.08 6.96 8.74 10.9565°C *2.00 *4.10 *6.99 *8.70 *10.9570°C *2.00 *4.12 *7.00 *8.67 *10.9575°C *2.00 *4.14 *7.02 *8.64 *10.9580°C *2.00 *4.16 *7.04 *8.62 *10.9585°C *2.00 *4.18 *7.06 *8.60 *10.9590°C *2.00 *4.21 *7.09 *8.58 *10.9595°C *2.00 *4.24 *7.12 *8.56 *10.95

* Values complemented

Buff er Tables

Page 114: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

114

-U1- Specifiable Buffer Set

You can specify a buff er set with 2 buff er solutions in the temperature range of 0 ... 95 °C, step width: 5 °C. To do so, select buff er set -U1- in the confi guration menu. As delivered, the Ingold technical buff er solutions pH 4.01 / 7.00 are stored as buff er set and can be edited.

Conditions for the specifi able buff er set:

All values must lie in the range pH 0 ... 14• Maximum diff erence between two adjacent pH values • (5 °C step width) of the same buff er solution: pH 0.25The values of buff er solution 1 must be lower than those of • buff er solution 2:The diff erence between values for identical temperatures must be greater than 2 pH units.

Faulty entries are indicated in measuring mode by the “FAIL BUFFERSET -U1-“ message.

The 25 °C value is always used for buff er display during calibration.

Page 115: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

115

Step Action/Display Remark

Select buff er set -U1-(CONFIG / SNS menu)

Select buff er solution 1 for editing

Select “YES“ using up/down key.

You are prompted for confirmation to prevent accidental changes of the settings.

Editing the values of buff er solution 1

Edit using arrow keys, press enter to confi rm and proceed to next tempera-ture value.

Enter the values for the first buffer solution in 5°C steps.The difference to the next value must not exceed 0.25 pH unit.

Select buff er solution 2 for editing

The difference between buffer solutions for iden-tical temperatures must be greater than 2 pH units.

-U1- Specifi able Buff er Set

Page 116: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

116

Buff er Set U1:Fill in your confi guration data or use the table as original for copy.

Temperature (°C) Buff er 1 Buff er 25

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

-U1- Specifi able Buff er Set

Page 117: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

117

Error HandlingAlarm condition:

The display backlighting turns • redThe alarm icon • is displayedThe complete measured-value display blinks• „• ERR xxx“ is displayed in the lower menu line

Press the [info] key to view a short error text:The error text appears in the lower menu line• The main display reads “• InFo“.

Parameter errors:Confi guration data such as current range, limit values, etc are checked during the input.If they are out of range,

“• ERR xxx“ is displayed for 3 sec,the display backlighting fl ashes red,• the respective maximum or minimum value is shown,• input must be repeated.•

If a faulty parameter arrives through the interface (IrDA, HART),an error message will be displayed: “• ERR 100...199“the faulty parameter can be localized by pressing the [• info] key

Calibration errors:If errors occur during calibration, e.g. by using a wrong calibration solution,

an error message will be displayed• calibration will be restarted•

Sensoface:If the Sensoface becomes sad,

the display backlighting will turn purple• the cause can be seen by pressing the • info keythe calibration data can be seen in the Diagnostics menu•

Page 118: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

118

Error codes

Error Messages

Error Info text(is displayed in case of fault when the Info key is pressed)

ProblemPossible causes

ERR 99 DEVICE FAILURE Error in factory settingsEEPROM or RAM defectiveThis error message only occurs in the case of a total defect. The device must be repaired and recalibrated at the factory.

ERR 98 CONFIGURATION ERROR Error in confi guration or calibration dataConfiguration or calibration data defectiveReset device to factory settings (SERVICE/DEFAULT), then calibrate

ERR 97 NO MODULE INSTALLED No module(Not for digital devices) Please have the module installed in the factory.

ERR 96 WRONG MODULE Wrong modulePlease have the module replaced in the factory.

ERR 95 SYSTEM ERROR System errorRestart required.If error still persists, send in the device for repair.

ERR 01 NO SENSOR pH sensor *Sensor defectiveSensor not connectedBreak in sensor cable

ERR 02 WRONG SENSOR Wrong sensor *

ERR 03 CANCELED SENSOR Sensor devaluated *

Page 119: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

119

ErrorInfo text(is displayed in case of fault when the Info key is pressed)

ProblemPossible causes

ERR 04 SENSOR FAILURE Failure in sensor *ERR 05 CAL DATA Error in cal data *ERR 10 ORP RANGE ORP display range

violation< -1999 mV or > 1999 mV

ERR 11 PH RANGE pH display range violation< -2 or > 16

ERR 12 MV RANGE mV rangeERR 13 TEMPERATURE RANGE Temperature range

violationERR 15 SENSOCHECK GLASS-EL Sensocheck glassERR 16 SENSOCHECK REF-EL Sensocheck ref.ERR 60 OUTPUT LOAD Load errorERR 61 OUTPUT 1 TOO LOW Output current 1

< 3.8 mAERR 62 OUTPUT 1 TOO HIGH Output current 1

> 20.5 mAERR 63 OUTPUT 2 TOO LOW Output current 2

< 3.8 mAERR 64 OUTPUT 2 TOO HIGH Output current 2

> 20.5 mAERR 69 TEMP. OUTSIDE TABLE Temperature value outside

tableERR 100 INVALID SPAN OUT1 Confi guration error

Output range 1

ERR 101 INVALID SPAN OUT2 Confi guration errorOutput range 2

Error Messages

Page 120: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

120

ErrorInfo text(is displayed in case of fault when the Info key is pressed)

ProblemPossible causes

ERR 102 FAILURE BUFFERSET -U1- Confi guration errorSpecifi able buff er set U1

ERR 104 INVALID PARAMETER CONTROLLER

Confi guration errorController

ERR 105 INVALID SPAN I-INPUT Confi guration errorCurrent input

ERR 106...255

VOID PARAMETER Invalid parameter

*) Digital sensors (ISM, InduCon, Memosens)

Error Messages

Page 121: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

121

Sensoface(Sensocheck must have been activated during confi guration.)

The smiley in the display (Sensoface) alerts to sensor problems ( defective sensor, sensor wear, defective cable, maintenance request). The permitted calibration ranges and the conditions for a friendly, neutral, or sad Sensoface are summarized in the following table. Additional icons refer to the error cause.

SensocheckContinuously monitors the sensor and its wiring. Critical values make the Sensoface “sad” and the corresponding icon blinks:

The Sensocheck message is also output as error message Err 15 (glass electrode) or Err 16 (reference electrode – for digital transmitters, however only with InduCon sensors with SG). The alarm contact is active, the display backlighting turns red, output current 1 is set to 22 mA (when confi gured correspondingly). Sensocheck can be switched off during confi guration (then Sensoface is also disabled). Exception: After a calibration a smiley is always displayed for confi rmation.

Please note:The worsening of a Sensoface criterion leads to the devaluation of the Sensoface indicator (Smiley becomes “sad”). An improvement of the Sensoface indicator can only take place after calibration or removal of the sensor defect.

Page 122: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

122

Sensoface

Display Problem StatusAsymmetry potential and slope

Asymmetry potential (zero) and slope of the sensor are still okay.The sensor should be replaced soon.

Asymmetry potential and slope of the sensor have reached values which no lon-ger ensure proper calibration. Replace sensor.

Calibration timer

Over 80% of the calibration interval has already past.

The calibration interval has been exceeded.

Sensor defect Check the sensor and its connections (see also Error Messages Err 15 and Err 16).

Response time

Sensor response time has increased.The sensor should be replaced soon.To achieve an improvement, clean the sensor and soak it in buffer.

Sensor response time signifi -cantly increased ( > 72 s, cali-bration aborted after 120 s)Replace sensor.

Page 123: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

123

Sensoface

Display Problem Status Sensor wear (for digital sensors only)

High temperatures and pH values have caused a wear of over 80%. The sensor should be replaced soon.

Wear is at 100%. Replace sensor.

Page 124: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

124

EC Declaration of Conformity

Page 125: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

125

EC Declaration of Conformity

Page 126: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

126

Stratos Pro A201X/A211X PH: Control Drawing

Page 127: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

127

Stratos Pro A201X/A211X PH: Control Drawing

Page 128: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

128

FM Control Drawing

Page 129: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

129

CSA Control Drawing

Page 130: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

130

FDA 21 CFR Part 11Conformity with FDA 21 CFR Part 11In their directive “Title 21 Code of Federal Regulations, 21 CFR Part 11, Electronic Records; Electronic Signatures“ the US American health agency FDA (Food and Drug Administration) regulates the production and processing of electronic documents for pharmaceutical devel-opment and production. This results in requirements for measuring devices used for corresponding applications. The following features ensure that the measuring devices of the M420 Series meet the de-mands of FDA 21 CFR Part 11:

Electronic Signature – PasscodesAccess to the device functions is regulated and limited by individually adjustable codes – “Passcodes“ (see SERVICE). This prevents unauthor-ized modifi cation of device settings or manipulation of the measure-ment results. Appropriate use of these passcodes makes them suitable as electronic signature.

Audit TrailEvery (manual) change of device settings can be automatically docu-mented. Each change is tagged with a “Confi guration Change Flag“, which can be interrogated and documented using HART communica-tion. Altered device settings or parameters can also be retrieved and documented using HART communication.

Extended logbookAudit Trail also records function activations (CAL, CONFIG, SERVICE), some Sensoface messages (cal timer, wear), and opening of the enclosure.

Page 131: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

131

Glossary Asymmetry potential

The voltage which a pH sensor provides at a pH of 7. The asymmetry potential is diff erent for each sensor and changes with age and wear.

Buff er set Contains selected buff er solutions which can be used for automatic calibration (Calimatic). The buff er set must be selected prior to the fi rst calibration.

Buff er solution Solution with an exactly defi ned pH value for calibrating a pH meter.

Calibration Adjustment of the pH meter to the current sensor characteristics. The asymmetry poten-tial and slope are adjusted. Either a one- or two-point calibration can be carried out. With one-point calibration only the asymmetry potential (zero point) is adjusted.

Calimatic Automatic buff er recognition. Before the fi rst calibration, the buff er set used must be activated once. The patented Calimatic then automatically recognizes the buff er solutions used during calibration.

Page 132: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

132

GlossaryCIP Cleaning In Place – CIP cycles are used for

cleaning the process-wetted parts in the process. They are performed for biotech ap-plications, for example. Depending on the application, one or more chemicals are used at temperatures above 70 °C. This extremely stresses the sensors. Digital sensors can release a message after preset number of CIP cycles. This allows replacing the sensor in time.

Combination electrode

Combination of glass and reference electrode in one body.

GainCheck Device self-test which runs automatically in the background at fi xed intervals. The memo-ry and measured-value transfer are checked. You can also start GainCheck manually in the diagnostics menu. In that case, also a display test will be performed.

ISFET adapter Adapter between ISFET sensor and trans-mitter. Here, the signal of the pH-sensitive FET is converted to voltage corresponding to the signal of a glass electrode. This volt-age is led to the pH input of the device and is processed further as usual. The adapter is directly supplied from the device.

Page 133: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

133

GlossaryISM® Intelligent Sensor Management –

ISM® sensors have an “electronic datasheet” which allows the storage of additional oper-ating parameters such as calibration date and settings directly in the sensor.

One-point calibration

Calibration with which only the asymmetry potential (zero point) is taken into account. The previous slope value is maintained. Only one buff er solution is required for a one-point calibration.

Passcode User-defi ned four-digit number to select certain operating modes.

pH sensor A pH sensor consists of a glass and a refer-ence electrode. If they are combined in one body, they are referred to as combination electrode. When the sensor has an additional platinum electrode, the oxidation-reduction potential (ORP) can be measured simultane-ously with the pH.

Response time Time from the start of a calibration step to the stabilization of the sensor potential.

Sensocheck Sensocheck continuously monitors the glass and reference electrodes. The resulting infor-mation is indicated by the Sensoface smileys. Sensocheck can be switched off .

Page 134: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

134

GlossarySensoface Provides information on the sensor condi-

tion. The zero point, slope, and response time are evaluated. In addition, the Sensocheck information is indicated.

SIP Sterilization In Place – CIP cycles are used for sterilizing the process-wetted parts in the process. They are performed for biotech applications, for example. Depending on the application, one or more chemicals are used at temperatures above 115 °C. This extremely stresses the sensors. Digital sensors can release a message after preset number of SIP cycles. This allows replacing the sensor in time.

Slope Is indicated in % of the theoretical slope (59.2 mV/pH at 25 °C). The sensor slope is diff erent for each sensor and changes with age and wear.

TAN Transaction number for releasing an additional function.

Two-point calibration

Calibration with which the asymmetry potential (zero point) and slope are deter-mined. Two buff er solutions are required for two-point calibration.

Zero adjustment Basic adjustment of the ISFET sensor to ensure reliable Sensoface information.

Zero point See asymmetry potential

Page 135: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

135

AAccess codes 95, 130, 144Accessories 97Alarm 37

Alarm delay 64Ambulance TAN 95Application in hazardous locations 15Approvals for application in hazardous locations 9, 103Assembly 11Asymmetry potential 81Audit Trail 91, 130Automatic calibration (Calimatic) 76

BBacklighting 31Buffer tables 105

CCalibration 34, 72

Automatic calibration (Calimatic) 76Calibration error 117Calibration timer 51, 122Configuration 48Data entry of premeasured sensors 80ISFET sensors 72Manual calibration with buffer entry 78Product calibration (pH) 82Redox calibration 84Temperature probe adjustment 86Zero adjustment 75

Calibration data 89Calibration error 117Calibration mode 49, 73Calimatic 76CD-ROM 3CIP 53Cleaning cycles 53Commissioning 8

Index

Page 136: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

136

Compulsory marking 9, 16Configuration 34

Alarm 64Calibration mode 48Calibration timer 50Cleaning cycles 52Current output 1 54Current output 2 60Individual configuration data 44, 116Menu groups 39Menu structure 38Output current during Error and HOLD 58Overview 41Sensocheck 64Sensor 46Sterilization cycles 52Tag number 66Temperature 46Time and date 66Time constant of output filter 56

Connection examples 19Connection to supply units 98Control Drawings 126Converting Slope to mV 81CSA Control Drawings 129

DData entry of premeasured sensors 80Date 67

Display 87Device self-test 90Device type, display 92Diagnostics 34, 88

Calibration data 89Device self-test 90Logbook 91Sensor data 89

Index

Page 137: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

137

IndexSensor monitor 92Version 92

Digital sensors 68Connection 69Sensor replacement 70Sensor type selection 47

Dimensions 12Display 31

Display test 90Display backlighting 31Disposal 2Documentation 3

EEC Declaration of Conformity 124EEPROM test 90Electronic Signature 130Enclosure 12Enclosure components 11Entering values 33Error handling 117Error messages 118Explosion protection 103Extended logbook 91, 130External temperature measurement 63

FFDA 21 CFR Part 11 130FLASH test 90FM Control Drawings 128

GGlossary 131

HHOLD 34, 36

End 36External activation of HOLD 37

Page 138: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

138

IndexManual activation of HOLD 37Output signal during HOLD 36, 59Output signal response 36

IInfo text 118Installation 15

Hazardous locations 9Intended use 7Introduction 7IrDA communication 95

KKeypad 30

LLogbook 91

MManual calibration with buffer entry 78Measured values, display 92Measurement 87Measuring 32Menu structure 35

Configuration 38Module test 90Mounting plan 12

OOperating modes 34Operating mode, selection 33Operating states 96Options 95, 97Order code 97ORP calibration 84Output current, fixed value 94Output current range 54, 60

Page 139: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

139

IndexOutput filter 56Output signal during HOLD 36, 59Overview 10

PPackage contents 3, 11Panel mounting 14Parameter error 117Parameter set A/B 39

Display 87Individual configuration data 44Manual selection 40

Passcodes 130, 144Assigning passcodes 95

Pipe mounting 13Point of measurement (TAG) 67Power supply units 98Presetting pH calibration 73Process variable 55, 61Product calibration 82Product line 97Protective hood 13

RRAM test 90Rating plates 16Redox calibration 84Registered trademarks 143Release of options 95Reset to factory settings 95Return of products under warranty 2

SSafety information 8Safety instructions 3Selection menu 33Sensocheck 64, 121

Configuration 65Sensoface 117, 121

Page 140: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

140

IndexSensor connection 17, 19Sensor data, display 89Sensor defect 122Sensor monitor 92, 94Sensor type selection 46Sensor wear 123Serial number, display 92Service 34, 93

Factory setting 95IrDA communication 95Passcodes 95Releasing options 95Sensor monitor 94Specifying current outputs 94

Service passcode lost 95Signal colors 31SIP 53Software version, display 92Specifications 99Start-up 8Sterilization cycles 53Supply units 98

TTag number (TAG) 67TAN options 95, 97Technical terms 131Temperature compensation 63Temperature dependence of reference systems measured against SHE 84Temperature detection 46

for calibration 49Temp specification via current input 47, 63

Temperature probe adjustment 86Temperature probe selection 47Terminal assignments 16Terminals 9, 15, 16

Page 141: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

141

IndexTime 67

Display 87Time constant of output filter 57Trademarks 143

UUser Interface 30

WWarranty 2Wiring 17

Examples 19Power supply units 98Sensor connection 17

ZZero adjustment (ISFET) 74

Page 142: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

142

Page 143: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

143

Trademarks

The following names are registered trademarks. For practical reasons they are shown without trademark symbol in this manual.Stratos®Sensocheck®Sensoface®Calimatic®GainCheck®InPro® is a registered trademark of Mettler-Toledo.Memosens® is a registered trademark of Endress+Hauser Conducta GmbH and Knick Elektronische Meßgeräte GmbH & Co. KG.HART® is a registered trademark of the HART Communication Foundation.

Page 144: Stratos®Pro A2 PH Instruction Manual - Elscolab · Stratos Pro A2...X rating plate at outside bottom of front with checkboxes for marking the respective application after installation

TA-212.115-KNE02 20091218 Software version: 1.x

KnickElektronische Messgeräte GmbH & Co. KGP.O. Box 37 04 15D-14134 Berlin

Tel: +49 (0)30 - 801 91 - 0Fax: +49 (0)30 - 801 91 - 200Internet: http://[email protected]

PasscodesIn the SERVICE – CODES menu you can assign passcodes to protect the access to certain functions.

Operating mode Passcode

Service (SERVICE) 5555

Diagnostics (DIAG)

HOLD mode

Calibration (CAL)

Confi guration (CONF)