Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to...

12
156 Display ■■■■■■ Simultaneous display of conduc- tivity/resistance /salinity and tem- perature; large unit symbols. Icons provide operating instruc- tions and signal unusual operat- ing states. Status indicators pro- vide information on the current operating status. Eye-catching alarm signaling by red, flashing LED. Calibration ■■■■■■ Simple adaptation of the sensor by direct input of the cell con- stant or through manual deter- mination of the constant with a calibration solution. Sensocheck ® sensor monitoring ■■■■■■ Reliability due to continual moni- toring of polarization and cable influences, indication and alarm triggering. HART ® communication ■■■■■■ Integration in fully automatic process cycles is possible due to the remote control function via HART ® communication. The de- vice can also be fully configured via a handheld terminal or from the control room. Measured val- ues, messages and device identi- fication can be downloaded at any time. Also available for hazardous area applications. Measurement of lowest to high conductivity with 2 or 4-electrode sensors. Exact measurement of ultrapure and natural waters through special temperature compensa- tion. Limit contacts report inad- missible exceeding of conductivi- ty values according to monitor water quality in the pharma- ceutical industry. Stratos ® 2211 (X) Cond ■■■■■■ Stratos ® Robust, tightly sealed enclosure Wall, pipe/post or panel mounting Easy installation with pre-assembled enclosure and plug-in terminals Optical alarm signal by red LED Sensocheck ® continual sensor monitoring GainCheck ® automatic device self-test International user interface with icons Large, clear display °C /°F temperature scale Additional mode indicators for – Measuring mode – Calibration mode – Alarm – Wash contact – Configuration Cal timer Integrated current source for simple monitoring of the peripherals HART ® communication Hazardous and safe area applications ■■■■■■ The facts _____________________________

Transcript of Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to...

Page 1: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

156

Display

Simultaneous display of conduc-tivity/resistance /salinity and tem-perature; large unit symbols. Icons provide operating instruc-tions and signal unusual operat-ing states. Status indicators pro-vide information on the currentoperating status. Eye-catchingalarm signaling by red, flashingLED.

Calibration

Simple adaptation of the sensorby direct input of the cell con-stant or through manual deter-mination of the constant with a calibration solution.

Sensocheck® sensormonitoring

Reliability due to continual moni-toring of polarization and cableinfluences, indication and alarmtriggering.

HART® communication

Integration in fully automaticprocess cycles is possible due tothe remote control function viaHART® communication. The de-vice can also be fully configuredvia a handheld terminal or fromthe control room. Measured val-ues, messages and device identi-fication can be downloaded atany time.

Also available for hazardous area applications.

Measurement of lowest to high conductivity with 2 or 4-electrode sensors.Exact measurement of ultrapureand natural waters throughspecial temperature compensa-tion. Limit contacts report inad-missible exceeding of conductivi-ty values according to monitorwater quality in the pharma-ceutical industry.

Stratos® 2211 (X) Cond

Stratos®

• Robust, tightly sealedenclosure

• Wall, pipe/post or panel mounting

• Easy installation with pre-assembled enclosureand plug-in terminals

• Optical alarm signal byred LED

• Sensocheck® continual sensor monitoring

• GainCheck® automatic device self-test

• International user interfacewith icons

• Large, clear display

• °C/°F temperature scale

• Additional mode indicatorsfor– Measuring mode– Calibration mode– Alarm– Wash contact – Configuration

• Cal timer

• Integrated current sourcefor simple monitoring ofthe peripherals

• HART® communication

• Hazardous and safe areaapplications

The facts_____________________________

Page 2: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

157

Stratos® 2211 (X) Cond /2212 (X) CondI

Measurement of lowest to high conductivity withelectrodeless sensors.Insensitive to polarization andpollution. Concentration deter-mination of various acids, basesand salts in broad temperatureranges. Salinity determination.

Display

Simultaneous display of conduc-tivity / concentration / salinity andtemperature; large unit symbols.Icons supply operating messagesand signal unusual operatingstates. Status indicators provideinformation on the current oper-ating status. Eye-catching alarmsignaling by red, flashing LED.

Calibration

Simple standardization of thesensor by direct input of the cellfactor or through manual deter-mination of the cell factor with acalibration solution. Zero pointdeviations can be calibrated.

Sensocheck® sensor monitoring

Reliability due to continual mo-nitoring of sensor function andconnection cable, indication andalarm triggering.

HART® communication

Integration in fully automaticprocess cycles is possible due to the remote control functionvia HART® communication. Thedevice can also be fully config-ured via a handheld terminal orfrom the control room. Mea-sured values, messages and de-vice identification can be down-loaded at any time.

Also available for hazardous areaapplications.

Stratos® 2212 (X) CondI

_

• Robust, tightly sealed enclosure

• Wall, pipe/post or panel mounting

• Easy installation with pre-assembled enclosureand plug-in terminals

• Optical alarm signal byred LED

• Sensocheck® continualsensor monitoring

• GainCheck® automatic device self-test

• International user interfacewith icons

• Large, clear display

• °C/°F temperature scale

• Additional mode indicatorsfor– Measuring mode– Calibration mode– Alarm– Wash contact – Configuration

• Cal timer

• Integrated current sourcefor simple monitoring ofthe peripherals

• HART® communication

• Hazardous and safe areaapplications

The facts_______________________________

Warranty

3 years!Warranty

Defects occurring within 3 years from delivery date shall be remediedfree of charge at our works (carriage

and insurance paid by sender).

Page 3: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

158

Product line 2211 (X) Cond / 2212 (X) CondI_________________________________________________________________________________

Order No.l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Stratos® 2211 (X) Stratos® (2-wire technology / HART®) 2211 CondCond l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Stratos® (2-wire technology / HART®) 2211 X Cond–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Stratos® 2212(X) Stratos® (2-wire technology / HART®) 2212 CondICondI l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Stratos® (2-wire technology / HART®) 2212 X CondI

l–––––––––––––––––––––––– l––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Mounting accessories See Accessories chapter (page 121)

Repeater power supplye. g. WG 20or WG 21

Handheldterminal

Stratos® e2211 X Cond

mS, MΩ, SALSensocheck®

mA/HART®

mS

4 ... 20 mA

EEx ia IIC EEx ib IIC

TempPt 100/Pt 1000/NTC 30 kΩ

ϑ

Cond measuring point Control room

Repeater power supplye. g. WG 20or WG 21

Handheldterminal

Stratos® e2212 X CondI

mS, %, SALSensocheck®

mA/HART®

mS

4 ... 20 mA

EEx ia IIC EEx ib IIC

Temp

Control roomCondl measuring point

Connection 2211 X Cond / 2212 X CondI_________________________________________________________________________________

Stratos®

Page 4: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

159

Terminal assignment 2211 (X) Cond_________________________________________________________________________________

Terminal assignment 2212 (X) CondI_________________________________________________________________________________

Stratos® 2211 (X) Cond /2212 (X) CondI

Page 5: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

160

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Cond input Operation with 2- or 4-electrode sensors

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Conductivity ranges With 4-el sensor: 0.2 µS · c ... 1000 mS · c

With 2-el sensor: 0.2 µS · c ... 200 mS · c(Display range limited to 3500 mS)l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Effective ranges*) Conductivity 0.000 ... 9.999 µS/cm00.00 ... 99.99 µS/cm000.0 ... 999.9 µS/cm0000 ... 9999 µS/cm0.000 ... 9.999 mS/cm00.00 ... 99.99 mS/cm000.0 ... 999.9 mS/cm0.000 ... 9.999 S/m00.00 ... 99.99 S/m

Resistivity 00.00 ... 99.99 MOhm cmConcentration 00.00 ... 9.99 %Salinity 0.0 ... 45.0 ‰ (0 ... 35 °C)l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Measurement error1, 2, 3) <1 % meas. val + 0.4 µS · c (c = cell constant)l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensor – Input of the cell constant with simultaneous display of conductivity value and standardization temperature

– Input of conductivity of calibration solution with simultaneous display of cell constant and temperature

– Product calibration– Temperature probe adjustmentl–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Adm. cell constant 00.0050 ... 19.9999 cm–1

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Concentration -01- NaCI 0.00 ... 9.99 % by wt (0 ... 60 °C)determination -02- HCI 0.00 ... 9.99 % by wt (–20 ... 50 °C)

-03- NaOH 0.00 ... 9.99 % by wt (0 ... 100 °C)-04- H2SO4 0.00 ... 9.99 % by wt (–17 ... 110 °C)-05- HNO3 0.00 ... 9.99 % by wt (–17 ... 50 °C)

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensocheck® Polarization detection and monitoring of the cable capacitancel–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensoface® Provides information on the condition of the sensor, evaluation of Sensocheck®

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensor monitor Direct display of measured values from sensor for validation, resistance/ temperaturel–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Temperature input*) Pt100/Pt1000/NTC 30 kOhm/NTC 8.55 kOhm (Betatherm); 2-wire connection,

adjustablel–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Measurement ranges Pt 100/Pt 1000 –20.0 ... +200.0 °C (–4 ... +392 °F)NTC 30 kOhm –20.0 ... +150.0 °C (–4 ... +302 °F)NTC 8.55 kOhm –10.0 ... +130.0 °C (+14 ... +266 °F)l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Resolution 0.1 °C / 1 °Fl–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Measurement error1,2,3) <0.5 K (<1 K with Pt100; <1K with NTC >100 °C)

*) User-defined 1) According to IEC 746 part 1, at nominal operating conditions2) ± 1 count 3) Plus sensor error

Specifications 2211 (X) Cond_________________________________________________________________________________

Stratos®

Page 6: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

161

Stratos® 2211 (X) Cond /2212 (X) CondI

Specifications for 2211 (X) Cond, continued_________________________________________________________________________________

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Temperature (OFF) Withoutcompensation*) (Lin) Linear characteristic 00.00 ... 19.99 %/K(reference tempera- (NLF) Natural waters to EN 27888ture 25 °C) (NaCI) Ultrapure water with NaCl traces (0 ... 120 °C)

(HCI) Ultrapure water with HCl traces (0 ... 120 °C)(NH3) Ultrapure water with NH3 traces (0... 120 °C)

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Loop current 4 ... 20 mA floating

Supply voltage 14 ... 30 Vl–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Measured variable*) Conductivity, concentration or salinityCharacteristic Linear or logarithmicOverrange*) 22 mA in the case of error messagesOutput filter*) PT1 filter, filter time constant 0 ... 120 sMeas. error1) < 0.3 % current value + 0.05 mAStart /end of scale As desired within rangeMin. span LIN 5 % of selected range

LOG 1 decadeCurrent source function 3.8 mA to 22.00 mA

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Hart® communication Digital communication via FSK modulation of the loop current,

reading of device identification, measured values, status and messagesreading and writing parameters, starting product calibration,signaling configuration amendment according to FDA 21 CFR 11

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Display LC display, 7 segment with symbols

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Main display Character height 17 mm, unit symbols 10 mm

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Secondary display Character height 10 mm, unit symbols 7 mm

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensoface® 3 status indicators (friendly, neutral, sad Smiley)

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Mode indication 4 mode indicators ‘meas’, ‘cal’, ‘alarm’, ‘config’

18 further icons for configuration and messagesl–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Alarm indication Red LED with alarm and HOLD, user-definedl–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Keypad 5 keys: [cal] [conf] [❿] [] [enter]l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Service functionsCurrent source Loop current, specifiable 3.8 ... 22 mA

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Device self-test Automatic memory testing (RAM, ROM, EEPROM)

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Display test Display of all segments

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Last Error Display of last error occurred

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensor monitor Display of the direct sensor signal (resistance/ temperature)l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Data retention Parameters and calibration data >10 years (EEPROM)

*) User-defined 1) According to IEC 746 part 1, at nominal operating conditions

Page 7: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

162

Stratos®

Specifications for 2211 (X) Cond, continued_________________________________________________________________________________

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––EMC EN 61326

Emitted interference: Class B (residential area) Class A: for mains > 60 V DCImmunity to interference: Industry

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Explosion protection II 2 (1) G EEx ib [ia] IIC T6(2211 X) FM IS, Cl.1, Div1, Group A,B,C & D T4

NI, Cl.1, Div2, Group A,B,C & D T4l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Nominal operatingconditionsAmbient temperature –20 ... +55 °C

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Transport/ storage temp. –20 ... +70 °C

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Relative humidity 10 ... 80 % not condensing

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Supply voltage 14 ... 30 Vl–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Enclosure Molded enclosure made of PBT (polybutylene terephtalate)

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Color Bluish gray RAL 7031

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Assembly – Wall mounting

– Pipe mounting: Ø 40 ... 60 mm, 30 ... 45 mm– Panel mounting, cutout to DIN 43 700 Sealed against panell–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Dimensions H 144 mm, W 144 mm, D 105 mml–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Protection IP 65/NEMA 4X (USA, Canada: indoor use only)l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Cable glands 3 breakthroughs for cable glands M20x1.52 breakthroughs for NPT 1/2” or Rigid Metallic Conduitl–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Weight Approx. 1 kg

Page 8: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

163

Stratos® 2211 (X) Cond /2212 (X) CondI

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Cond input Input for electrodeless conductivity sensors: SE 655/656

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Conductivity ranges Conductivity: 0.000 ... 1999 mS/cm

Saturation: 0.00 ... 100.0 % by wtSalinity: 0.0 ... 45.0 ‰ (0 ... 35 °C)l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Effective ranges*) Conductivity 0.000 ... 9.999 mS/cm (only with 2212)00.00 ... 99.99 mS/cm000.0 ... 999.9 mS/cm0000 ... 1999 mS/cm0.000 ... 9.999 S/m00.00 ... 99.99 S/m

Concentration 0.00 ... 9.99 % 10.0 ... 100.0 %Salinity 0.0 ... 45.0 ‰ (0 ... 35 °C)l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Measurement error1, 2, 3) <1 % meas. val. +0.02 mS / cml–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensor – Input of the cell factor with simultaneous display of conductivity value and temperaturestandardization – Input of conductivity of calibration solution with simultaneous display of cell factor and

temperature– Product calibration– Zero point adjustment– Temperature probe adjustmentl–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Permissible cell factor 00.100 ... 19.999l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Adm. transfer ratio 001.00 ... 199.99l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Adm. zero point adjustment ± 0.5 mS/cml–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Concentration -01- NaCI 0 – 26 % by weight (0 °C) ... 0 – 28 % by weight (100 °C)determination*) -02- HCI 0 – 18 % by weight (–20 °C) ... 0 – 18 % by weight (50 °C)

-03- NaOH 0 – 13 % by weight (0 °C) ... 0 – 24 % by weight (100 °C)-04- H2SO4 0 – 26 % by weight (–17 °C) ... 0 – 37 % by weight (110 °C)-05- HNO3 0 – 30 % by weight (–20 °C) ... 0 – 30 % by weight (50 °C)-06- H2SO4 94 – 99 % by weight (–17 °C) ... 89 – 99 % by weight (115 °C)-07- HCI 22 – 39 % by weight (–20 °C) ... 22 – 39 % by weight (50 °C)-08- HNO3 35 – 96 % by weight (–20 °C) ... 35 – 96 % by weight (50 °C)-09- H2SO4 28 – 88 % by weight (–17 °C) ... 39 – 88 % by weight (115 °C)-10- NaOH 15 – 50 % by weight (0 °C) ... 35 – 50 % by weight (100 °C)

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensocheck® Monitoring of the primary coil and lines for short circuits and the secondary coil

for open circuitl–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensoface® Provides information on the condition of the sensor, evaluation of zero point,

and Sensocheck®

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensor monitor Direct display of measured values from sensor for validation, resistance/ temperature

*) User-defined 1) According to IEC 746 part 1, at nominal operating conditions2) ± 1 count 3) Plus sensor error

Specifications 2212 (X) CondI_________________________________________________________________________________

Page 9: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

164

Stratos®

Specifications for 2212 (X) CondI, continued_________________________________________________________________________________

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Temperature input*) Pt100/Pt1000/NTC 30 kOhm/NTC 100 kOhm; 2-wire connection, adjustable

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Measurement range Pt 100/Pt 1000 –20.0 ... +150.0 °C (–4 ... +302 °F)

NTC 100 kOhm –20.0 ... +130.0 °C (–4 ... +266 °F)NTC 30 kOhm –20.0 ... +130.0 °C (–4 ... +266 °F)l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Resolution 0.1 °C / 1 °Fl–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Measurement error1, 2, 3) <0.5 K (<1 K with Pt100; <1K with NTC >100 °C)l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Temperature (OFF) withoutcompensation*) (Lin) linear characteristic 00.00 ... 19.99 %/K(reference temp 25 °C) (NLF) natural waters to EN 27888l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Loop current 4 ... 20 mA floating

Supply voltage 14 ... 30 Vl–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Measured variable*) Conductivity, concentration or salinityCharacteristic Linear or logarithmicOverrange*) 22 mA in the case of error messagesOutput filter*) PT1 filter, filter time constant 0 ... 120 sMeas. error1) < 0.3 % current value + 0.05 mAStart/end of scale As desired within rangeMin. span LIN 5 % of selected range

LOG 1 decadeCurrent source function 3.8 mA ... 22.00 mA

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Hart® communication Digital communication via FSK modulation of the loop current,

reading of device identification, measured values, status and messagesreading and writing parameters, starting product calibration,signaling configuration amendment according to FDA 21 CFR 11

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Display LC display, 7 segment with symbols

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Main display Character height 17 mm, unit symbols 10 mm

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Secondary display Character height 10 mm, unit symbols 7 mm

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensoface® 3 status indicators (friendly, neutral, sad Smiley)

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Mode indication 4 mode indicators ‘meas’, ‘cal’, ‘alarm’, ‘config’

18 further icons for configuration and messagesl–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Alarm indication Red LED with alarm and HOLD, user-definedl–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Keypad 5 keys: [cal] [conf] [❿] [] [enter]

*) User-defined 1) According to IEC 746 part 1, at nominal operating conditions

Page 10: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

165

Stratos® 2211 (X) Cond /2212 (X) CondI

Specifications for 2212 (X) CondI, continued_________________________________________________________________________________

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Service functionsCurrent source Loop current, specifiable 3.8 ... 22 mA

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Device self-test Automatic memory testing (RAM, ROM, EEPROM)

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Display test Display of all segments

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Last Error Display of last error occurred

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Sensor monitor Display of the direct sensor signal (resistance/ temperature)l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Data retention Parameters and calibration data >10 years (EEPROM)l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––EMC EN 61326

Emitted interference: Class B (residential area) Class A: for mains > 60 V DCImmunity to interference: Industry

l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Explosion protection II 2 (1) G EEx ib [ia] IIC T6(2211 X) FM IS, Cl.1, Div1, Group A,B,C & D T4

NI, Cl.1, Div2, Group A,B,C & D T4l–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Nominal operatingconditionsAmbient temperature –20 ... +55 °C

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Transport / storage temp –20 ... +70 °C

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Relative humidity 10 ... 80 % not condensing

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Supply voltage 14 ... 30 Vl–––––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Enclosure Molded enclosure made of PBT (polybutylene terephtalate)

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Color Bluish gray RAL 7031

l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––Assembly – Wall mounting

– Pipe mounting: Ø 40 ... 60 mm, 30 ... 45 mm– Panel mounting, cutout to DIN 43 700 Sealed against panell–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Dimensions H 144 mm, W 144 mm, D 105 mml–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Protection IP 65/NEMA 4X (USA, Canada: indoor use only)l–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Cable glands 3 breakthroughs for cable glands M20x1.52 breakthroughs for NPT 1/2” or Rigid Metallic Conduitl–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––

Weight Approx. 1 kg

Page 11: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

166

EC Type Examination Certificates 2212 (X) Cond_________________________________________________________________________________

Stratos®

Page 12: Stratos 2211 (X) Cond - Tech · 157 Stratos® 2211 (X) Cond/2212 (X) CondI Measurement of lowest to high conductivity with electrodeless sensors. Insensitive to polarization and pollution.

167

EC Type Examination Certificates 2212 (X) CondI_________________________________________________________________________________

Stratos® 2211 (X) Cond /2212 (X) CondI