STRATEGIES FOR KVCBCT-GUIDED PATIENT SETUP OF...

151
1 STRATEGIES FOR KVCBCT-GUIDED PATIENT SETUP OF STEREOTACTIC RADIOTHERAPY AND ASSESSMENTS OF THEIR DOSE CONSEQUENCES By BO LU A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY UNIVERSITY OF FLORIDA 2013

Transcript of STRATEGIES FOR KVCBCT-GUIDED PATIENT SETUP OF...

  • 1

    STRATEGIES FOR KVCBCT-GUIDED PATIENT SETUP OF STEREOTACTIC RADIOTHERAPY AND ASSESSMENTS OF THEIR

    DOSE CONSEQUENCES

    By

    BO LU

    A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT

    OF THE REQUIREMENTS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY

    UNIVERSITY OF FLORIDA

    2013

  • 2

    © 2013 Bo Lu

  • 3

    To my beloved wife, Ying, and son, Yiwen

  • 4

    ACKNOWLEDGMENTS

    I would like to gratefully and sincerely thank my advisor, Dr. Chihray Liu, for his

    guidance. His insights and suggestions are always inspiring. I really appreciate his

    willingness to offer prompt help at any time. His diligence, perseverance and serious

    attitude on both research and clinic set a good example for me. He is also the person

    who introduced me into the exciting medical physics field as well as encouraged me to

    come back to the academic world to pursue my PhD after two years clinic practice in

    Mississippi. His unmatched enthusiasm inspired me to devote my career to medical

    physics research and practice. I really thank him for his continuous encouragement and

    support. My thanks extend to my committee members, Dr. Sanjiv Samant and Dr.

    Jonathan Li and Dr. William M. Mendenhall for their input and many valuable

    suggestions. I thank Dr. Darren L. Kahler for his comments, suggestions and hard

    editing work on my papers.

    I need to express my special gratitude to Ms. Kate Mittauer, Ms. Soyong Lee and

    Dr. Yin Huang. Without their selfless help on the experiments, I would not achieve any

    goal on the topic of the tumor alignment for lung SBRT, which is one of the critical

    components of my PhD research. I enjoyed numerous inspiring discussions with them. I

    would also like to sincerely thank Dr. Jean Peng from University of South Carolina for

    her generous support on providing the code of image and contour rotation operation,

    which I am really indebted to.

    Most of all, I thank my wife and my son, my parents and my sister for their great

    unconditional support, love and encouragement.

  • 5

    TABLE OF CONTENTS

    page

    ACKNOWLEDGMENTS .................................................................................................. 4

    LIST OF TABLES ............................................................................................................ 8

    LIST OF FIGURES .......................................................................................................... 9

    ABSTRACT ................................................................................................................... 11

    CHAPTER

    1 INTRODUCTION .................................................................................................... 13

    General Background ............................................................................................... 13 the Relationship between Cbct Scan Dose and Registration Accuracy ............ 14 Dose Consequence Due to Small Rotational Errors ......................................... 15 Cbct-Guided Setup Strategy for Lung Sbrt ....................................................... 17

    Study Topics ........................................................................................................... 20

    2 DECREASING THE KV-CBCT DOSE BY REDUCING THE PROJECTION NUMBER ................................................................................................................ 25

    Abstract ................................................................................................................... 25 Background ............................................................................................................. 26 Methods and Materials............................................................................................ 30

    X-ray Volume Imaging System ......................................................................... 30 Phantom and Patient Selection ........................................................................ 31

    Catphan phantom ...................................................................................... 31 Rando phantom ......................................................................................... 32 Patient sites ............................................................................................... 32

    Acquisition and Reconstruction of Cbct Images ............................................... 33 Imaging Quality Check with a Reduced Projection Number ............................. 35

    3D spatial resolution .................................................................................. 35 3D low-contrast visibility ............................................................................. 35 3D uniformity .............................................................................................. 36

    Registration Analysis ........................................................................................ 37 Results .................................................................................................................... 39

    Reconstructed Images ..................................................................................... 39 Imaging-quality Check ...................................................................................... 39

    3D spatial resolution check ........................................................................ 40 3D low-contrast visibility check .................................................................. 40 3D uniformity check ................................................................................... 41

    Registration ...................................................................................................... 41 Discussion .............................................................................................................. 42

  • 6

    Summarization ........................................................................................................ 47

    3 DOSE CONSEQUENCES CAUSED BY SMALL ROTATIONAL SETUP ERRORS ................................................................................................................ 57

    Abstract ................................................................................................................... 57 Background ............................................................................................................. 58 Methods and Materials............................................................................................ 62

    Descriptions of Selected Srt Cases .................................................................. 62 Case selection ........................................................................................... 62 Contour delineation .................................................................................... 62 Treatment techniques and plan evaluation criteria ..................................... 63 Patient setup .............................................................................................. 64

    Rotation Simulations and Rotational Angle Selections ..................................... 65 Comparison between Dose Maps with and without Small Rotational Errors .... 66 Quantitative Comparisons among Original, Recalculated, and Odmorc-

    generated Dvhs for Small Rotational Errors .................................................. 67 Ufpsas Workflow and a Fast Algorithm for Multi-contour Geometric Shift ........ 67

    Results .................................................................................................................... 70 Comparison of Dose Maps for Rotated Rois .................................................... 70 Comparison of Dvhs for Ctvs and Oars ............................................................ 71 Summary of Calculation Speed for the Rotation Algorithm............................... 72

    Discussion .............................................................................................................. 73 Summarization ........................................................................................................ 79

    4 CENTROID-TO-CENTROID (CTC) STRATEGY FOR LUNG SBRT SETUP ......... 92

    Abstract ................................................................................................................... 92 Background ............................................................................................................. 93 Methods and Materials............................................................................................ 96

    Centroid-To-Centroid Registration Strategy ..................................................... 96 Case Selections ............................................................................................... 97 Clinic Workflow for Sbrt Lung ........................................................................... 98 Ctc Simulations and Dose Calculations ............................................................ 99 Deformable Registration ................................................................................. 100 Statistical Comparison among Various Registration Algorithms ..................... 102 Uncertainty Analysis ....................................................................................... 102

    Uncertainty check of deformable registration results ............................... 102 ITV variation check during treatment course ............................................ 104 Concordance check of Itv between Cbct and A4dct ................................. 104

    Results and Discussion......................................................................................... 105 Summarization ...................................................................................................... 113

    5 THE EVALUATION OF CTC PERFORMANCE ON THE SCENARIOS WITH BREATHING VARIATIONS .................................................................................. 120

    Abstract ................................................................................................................. 120

  • 7

    Background ........................................................................................................... 121 Methods and Materials.......................................................................................... 123

    Phantom Description ...................................................................................... 123 Design and Simulation of Respiratory Variation ............................................. 124 Treatment Plan and Itv Delineations .............................................................. 127 Registration Comparison between T-only G Mode and Ctc method ............... 127

    Results and Discussion......................................................................................... 129 the Comparison of Centroid Discrepancies between Two Methods ............... 129 DVH Analysis of Two Extreme Cases ............................................................ 133

    Summarization ...................................................................................................... 134

    6 CONCLUSIONS ................................................................................................... 142

    Summarization ...................................................................................................... 142 Future Works ........................................................................................................ 143

    LIST OF REFERENCES ............................................................................................. 145

    BIOGRAPHICAL SKETCH .......................................................................................... 151

  • 8

    LIST OF TABLES

    Table page 2-1 Technical parameters lists of scan for phantoms and patients ........................... 49

    2-2 Projection number of different schemes for phantoms and patients used in study ................................................................................................................... 49

    2-3 Xvi computing time of reconstructions for patients and phantoms with different projection-number schemes ................................................................. 49

    3-1 Summary of diagnosis information and plan technique for the 15 cases in this study ................................................................................................................... 81

    3-2 Roi statistical results of dose differences between original doses and recalculated doses on rotated images. ............................................................... 82

    4-1 Case characteristics ......................................................................................... 114

    4-2 Statistic results of geometric discrepancies among ctc, b and g methods ........ 115

    4-3 Uncertainty check ............................................................................................. 116

    4-4 Itv volumetric variations from a4dct to cbct ....................................................... 116

    5-1 The list of scanned cases with combination of various scenarios ..................... 135

    5-2 Itv Registration results in terms of centroid alignment discrepancy for ctc and t-only methods .................................................................................................. 136

  • 9

    LIST OF FIGURES

    Figure page 2-1 Catphan reconstruction results.. ......................................................................... 50

    2-2 Rando phantom reconstruction results.. ............................................................. 51

    2-3 Patient reconstruction results.. ........................................................................... 52

    2-4 Results of a low-contrast visibility check with different reconstruction schemes. ............................................................................................................ 53

    2-5 Results of a uniformity check with different reconstruction schemes. ................. 53

    2-6 The scatter charts of mean registration errors versus reconstruction schemes for Phantoms with “bone mode” registration algorithm.. ..................................... 54

    2-7 The scatter charts of mean registration errors versus reconstruction schemes for all patient sites with “bone mode” registration algorithm. ............................... 55

    2-8 The scatter charts of mean registration errors versus reconstruction schemes for non-rigid patient sites with “grey value” registration algorithm. ...................... 56

    3-1 Clinical setup for srt treatment ............................................................................ 84

    3-2 Definition of the four layers for investigating various degrees of dose variations due to rotations. ................................................................................. 85

    3-3 Flowcharts of contour geometric rotation algorithm. ........................................... 86

    3-4 A 2-Dimensional example of coupling and decoupling processes for multi-contour rotation operation. .................................................................................. 87

    3-5 Distribution of the percent dose difference between the original dose map and recalculated dose map on the rotated image using rotation-simulated ct images and Rois in axial view cut ....................................................................... 88

    3-6 Dvh curves of the ctv and the oars of case 1 for odoc, odrc and rdrc with rotations of +5o in all three directions .................................................................. 88

    3-7 Absolute percentage differences of doses for ctvs at 99%, 95% and 20% of

    ctv volume among odoc, rdrc and odrc with ±5⁰ rotations .................................. 89

    3-8 Number of cases cataloged by magnitudes of dose variations at oars dose limit volume among odoc, rdrc and odrc with rotations. ...................................... 90

    3-9 Schematic drawing of the treatment depth change due to head rotation. ........... 91

  • 10

    4-1 Comparison of ITV registration results by rigid registration .............................. 117

    4-2 Schematic draw of tumor motion patterns and Itv positions during 4dct and cbct scans for registration uncertainty check on a respiration motion phantom. .......................................................................................................... 117

    4-3 Registration shifts in three directions among b, g and ctc methods. ................ 118

    4-4 Statistic results of dose coverage differences of Itv by b, g and ctc methods ... 118

    4-5 Dvh illustrations of three extreme cases ........................................................... 119

    5-1 The Illustration of components of cirs dynamic thorax phantom (model 008). .. 137

    5-2 Simulated respiratory profiles with five different inspiration to expiration ratios. ................................................................................................................ 138

    5-3 Coronal and transvers views of a4dct image data sets from three different cases ................................................................................................................ 139

    5-4 Registration workflows and results for case no.5 for both ctc and t-only methods ............................................................................................................ 140

    5-5 Dvhs of two extreme cases with iso-center determined by different strategies and beam aperture defined by different sizes of ptv ......................................... 141

  • 11

    Abstract of Dissertation Presented to the Graduate School of the University of Florida in Partial Fulfillment of the Requirements for the Degree of Doctor of Philosophy

    STRATEGIES FOR KVCBCT-GUIDED PATIENT SETUP OF STEREOTACTIC RADIOTHERAPY AND ASSESSMENTS OF THEIR DOSE CONSEQUENCES

    By

    Bo Lu

    August 2013

    Chair: Chihray Liu Major: Biomedical Engineering

    Stereotactic radiotherapy (SRT) technique has been widely utilized for various

    treatments sites over the past two decades. To be able to escalate target dose while

    keeping normal tissue dose under toxicity threshold during every session of the entire

    treatment course, high accuracy and reproducibility of patient positioning were required.

    Until recent, Kilo-Voltage (kV) CBCT systems integrated into the gantries of

    linear accelerators have been pervasively employed as an advanced image-guided

    radiotherapy (IGRT) modality to acquire high-resolution volumetric images of patients

    for treatment localization purposes. However, in the current research regime, it lacks

    specific strategies for various sites on how to achieve the optimal patient setup and

    pertinent assessment of their dose consequence.

    Most of the registration strategies are based upon the methodology of intensity

    matching between reference and localization images for 6 degree freedom shift, which

    is not always practical for all the clinical scenarios and even possess intrinsic flaws for

    some specific site. Meanwhile the relationship between the patient’s dose received from

    the treatment (beam delivery and CBCT scan itself) and setup alignment strategies had

    not been sufficiently addressed.

  • 12

    This work provided specific strategies for CBCT-guided setup procedures to

    achieve optimal dose effects for various sites as well as assess the dose consequence

    of those strategies. It includes the following works: 1.the assessment of the patient’s

    dose generated by CBCT scan itself, which involved with the investigation of the effect

    of CBCT on registration accuracy and image qualities with a reduced number of planar

    projections used in volumetric imaging reconstruction, 2. the investigation of the impact

    of small rotational errors on the magnitudes and distributions of spatial dose variations

    for intracranial stereotactic radiotherapy (SRT) treatment setups, 3. the development of

    a quick online evaluation method to assess the dose consequence due to small

    rotational errors, 4. the invention of a novel method (centroid-to-centroid (CTC)) to

    facilitate the lung SBRT setup for translational-only scenarios in order to achieve

    optimal tumor dose coverage, 5. the assessment of CTC method’s performance on

    breathing variation scenarios using a phantom study.

  • 13

    CHAPTER 1 INTRODUCTION

    General Background

    The treatment effects of stereotactic radiotherapy, such as intracranial SRT, lung

    SBRT and etc., are heavily relied on the patient setup accuracy. The invention of IGRT

    provides the possibility to facilitate such need. While all radiation therapy procedures

    are image-guided per se. traditionally, imaging technology has primarily been used in

    producing 3D scans of the patient’s anatomy to identify the location of the tumor prior to

    treatment. The verification of a treatment plan is typically done at the level of beam

    portals relative to the patient’s bony anatomy before patient treatment. In contemporary

    radiation therapy field, the term of IGRT is refer to newly emerging radiation planning,

    patient setup, and delivery procedures that integrate cutting-edge image-based tumor

    definition methods, patient positioning devices, and/or radiation delivery guiding tools.

    These techniques combine new imaging tools, which interface with the radiation

    delivery system through hardware and/or software, and state-of-the-art 3DCRT or

    IMRT, and allow physicians to optimize the accuracy and precision of the radiotherapy

    by adjusting the radiation beam based on the true position of the tumor and critical

    organs. With IGRT, it is also possible to take tumor motion into account during radiation

    therapy planning and treatment setup. Because IGRT improves precision, it raises the

    possibility of shortening the duration of radiation therapy by reducing the number of

    treatment sessions for some forms of cancer, such as lung SBRT.

    Regarding patient setup in the IGRT workflow, KVCBCT-guided setup is surely

    the most popular approach implemented currently. It allows physician to utilize the

    volumetric image information to verify and correct the planning patient setup in the linac

  • 14

    coordinates by comparing with the patient position defined in treatment plan. In this

    work, we are interested in several questions regarding the strategies on how to use

    CBCT system to efficient and accurately guide the patient setup for various treatment

    sites as well as their corresponding dose consequence (beam delivery and CBCT scan

    itself).

    The Relationship between Cbct Scan Dose and Registration Accuracy

    While the dose from a single CBCT image acquisition is small compared to the

    therapy dose, repeated daily image-guidance procedures can not only lead to

    substantial dose to normal tissue, but also generate a high integral dose for the

    scanning area of the patient 1. Radiation therapy patients are already being exposed to

    very high and localized radiation; therefore, the additional radiation from imaging has an

    associated risk and should be kept low 2,3. How to balance imaging-quality improvement

    and dose escalation risk over broad clinical scenarios is still a controversial topic in the

    radiation therapy community and would be beyond the scope of this study. In this study,

    however, we attempt to reveal the impact of image information and registration

    accuracy resulting from dose-reduction strategies.

    In general, dose can be minimized during CBCT scan through two different

    means: one is by reducing the tube current; the other is by reducing the projections

    used in volumetric imaging reconstruction 3. The first method would result in a reduced

    signal-noise ratio due to less incident photons (reduced mAs) interacting with detectors,

    which ultimately degrade the image quality from several perspectives (spatial resolution,

    soft-tissue contrast, edge preservation, etc.). Those degrading effects have been

    studied by several investigators, and solutions to alleviate them have been proposed4-6.

    Alternatively, the second strategy to reduce the patient’s dose underlies the reduction of

  • 15

    sampling frequency during image acquisition. The major advantage of the second

    method versus the first one is that it is a time saver. Faster scan speed and shorter

    reconstruction time could ultimately limit setup uncertainty as a result of less patient on-

    couch time. Unfortunately, like the mA-reduction method, the disadvantage of the

    second strategy is that image quality diminishes. More specifically, the soft-tissue

    contrast degrades and streak-shaped artifacts appear. In IGRT, patient alignment

    information derived from imaging registration is the main purpose of onboard therapy

    imaging, which leads to a major concern with IGRT imaging – registration accuracy

    between reference CT and CBCT rather than general image-quality critiques. The

    alignment information generated by volumetric imaging registration will eventually

    provide clinicians with patient setup accuracy and, subsequently, position-adjustment

    guidance, if necessary. What are the effects of imaging quality degradation under a

    dose-decreasing strategy of reducing the projection number? And what is the optimal

    projection-number needed for reconstruction to balance between the dose and

    registration accuracy for different sites? This work provided answers to all those

    questions

    Dose Consequence Due to Small Rotational Errors

    The characteristic that is common to all of the current IGRT systems is that they

    are capable of providing target geometric correction information in terms of 6 degrees-

    of-freedom movements prior to and/or during treatment, which is adequate for setup

    corrections to rigid anatomy, such as the brain. While translational corrections can be

    easily achieved by shifting the table, performing rotational corrections is not always an

    easy task. On the one hand, manually adjusting the head position according to the

    IGRT system guidance is not always an option for the therapists since the restriction of

  • 16

    the immobilization device and/or the patient’s comfort level can sometimes restrict the

    freedom of the rotation and consequently leave a large discrepancy between the

    expected and achieved correction. On the other hand, most of the regular linear

    accelerator treatment couches can only be rotated in one direction (axis perpendicular

    to the couch surface)7, which makes a complete geometric correction (pitch, yaw, and

    roll) using current couch systems not possible7. Several approaches have been

    introduced to compensate for uncorrected rotational errors7-9. These methods either

    require a beam’s eye view (BEV) adjustment7,8 or need an additional rotational device9

    to offset rotational errors, which is not always practical in all clinical scenarios. Also,

    introducing new devices or altering treatment parameters can potentially complicate the

    entire system and may ultimately aggravate treatment uncertainties and increase the

    chance of setup errors.

    Dose consequence due to setup errors for various sites has been studied

    extensively10-19. Some investigators10-15 suggest that the dosimetric influence of small

    rotational errors is minimal for most cases in head and neck treatment, prostate

    treatment, and spine treatment, whereas larger rotational errors may dramatically affect

    clinical target volume (CTV) coverage and/or organs-at-risk (OAR) sparing, such as in

    cases of irregular target shape and/or OAR positions that are proximate to the target.

    Our previous study16 demonstrated similar dosimetric consequences in DVH form for

    SRT of intracranial tumors. All of these studies indicate that dosimetric consequences

    due to uncorrected rotational errors vary on a case by case basis and depend on

    rotational angles and orientations, tumor size and shape, treatment site, and other

    related factors. In this work, we will not restrict our visions in the level of evaluating DVH

  • 17

    changes for PTV and OARs for intracranial stereotactic radiotherapy treatment setups

    due to small rotational errors. Instead, we thoroughly investigated the impact of small

    rotational errors on the magnitudes and distributions of dose map variations.

    For cases in which rotational compensation is untenable, online systems that

    allow for instantaneous evaluation of the changes between planned dose and

    regenerated dose accounting for uncorrected rotational errors are desirable. Such

    systems can help clinicians make judgments immediately prior to the treatment and can

    ultimately improve clinical efficacy. However, online evaluation systems that regenerate

    3D dose distributions using current stand-alone treatment planning systems (TPS) are

    not practical for multiple reasons, such as the unavailability of rotation-simulation

    modules for dose regeneration, inadequate dose computational speed, costly hardware

    and software, etc. A practical on-line algorithm which can estimate the new DVHs due

    to rotational errors was developed in our work.

    Cbct-Guided Setup Strategy for Lung Sbrt

    For lung SBRT setup, the alignment between reference CT and localization CT

    are typically initiated with an automatic registration algorithm and finalized according to

    the clinician’s judgment. Commercial software for automatic online registration of KV-

    CBCT and planning CT (PCT) often provides two kinds of similarity measures (“bone”

    mode and “gray-value” mode) with six degrees of freedom for application of various

    clinical scenarios20. The bone mode, which uses a Chamfer matching algorithm to

    match the edge information of two image data sets, is relatively insensitive to noise and

    has a fast computation time21-24. It is especially useful when the matching objects are

    the rigid bony structures, but it performs poorly with lower contrast subjects. The gray-

    value mode, which handles the auto registration by matching voxel gray-scale intensity

  • 18

    values throughout the entire selected image volume of interest, is known as the “cross-

    correlation” technique25. In addition to edge information, soft tissue information is also

    included in the gray-value mode and can potentially average out the registration error

    caused by anatomy variance between the two image sets. It performs well with

    appropriate selection of registration area for non-rigid anatomy possessing lower

    contrast objects (like soft tissues). In terms of calculation time, the gray-value mode

    generally takes much longer than the bone mode.

    Although automatic registrations are able to generate reasonable alignment

    parameters, which can help clinicians perform accurate patient setups for most of

    cases, several challenges and issues still exist: (1) Automatic registration methods can

    potentially result in a mismatch of lung tumors if: (i) these tumors have a large variation

    in size and/or location or (ii) the region of registration (ROR) is inappropriately selected.

    Since variations in tumor shape and position relative to the rib cage occur during the

    preparation period (after the CT scan and prior to the first treatment), they are often

    unknown or difficult to quantify, and bony registration results can potentially lead into an

    inaccurate or incorrect position26. The gray-value method is capable of alleviating such

    problems to some extent with an appropriately defined ROR. However, finding a generic

    rule of ROR selection is not an easy task due to the tremendous variety of tumor

    shapes and locations. Our clinical experience indicates that gray-value registration with

    too large a ROR selection possesses almost no advantage compared with bone mode

    registration because the contribution of intensity weighting over tumor area to the

    overall objective is negligibly small. With too small a ROR selection (only focusing on

    tumor region), large rotation corrections can be induced due to the symmetric shape

  • 19

    and/or the deformation of the tumor. Such parameters of rotation correction are invalid

    and cannot be implemented. (2) For cases in which the isocenter is located away from

    tumor (e.g., multi-lesions with one isocenter), translational correction only (a

    conventional linear accelerator couch has only translational shift and yaw rotation

    capability) with six-degree automatic registration can potentially result in a significant

    setup error since rotation offset can be amplified by the distance between the isocenter

    and the corresponding lesion27. (3) Manual registrations lacking systematic guidelines

    can lead to inconsistent or incorrect setups, while automatic registration results are not

    reliable as the first-level alignment. The registration accuracy of a manual adjustment

    can be influenced by the clinician’s knowledge of the anatomy as well as by image

    quality and window/level settings, especially for those cases where large tumor

    variations are presented between CBCT and reference CT. (4) The inconsistency of

    image data sets (between PCT and CBCT) for registration, which has not drawn a great

    deal of attention in the radiation oncology society, is another issue. Most clinics register

    volumetric average CT images acquired by slow CBCT scanning with free breathing

    (arbitrary phase) CT images acquired through fast scanning to obtain setup correction

    parameters. Such methods can cause low-accuracy alignment due to difference in

    tumor appearance between the two image sets.

    In order to achieve the optimal dosimetric consequence for tumors with the

    strategy of couch translational corrections of patient setup, Yue et al. 28 presented an

    approach using a two-step optimization algorithm. Since their method needs physician

    involvement for tumor delineation post to CBCT scanning and immediately prior to the

    treatment for setup corrections, it is obviously impractical for real clinical situations. In

  • 20

    addition, the optimization process is underlain by the application of model-based dose

    computation algorithms (inverse square and tissue maximum ratio corrections) to

    compensate for anatomy variance between planning CT and CBCT, which is only

    suitable for dose calculation in relative homogeneity anatomies (such as pelvic region

    for prostate treatment) but inappropriate to be applied to the anatomy with highly

    heterogeneous density (such as lung tumors and lung tissues in its vicinity).

    To facilitate the problems mentioned above for lung SBRT setup, a practical

    strategy for setup alignment of SBRT lung treatments was introduced. This strategy can

    guarantee the best tumor dose coverage using DVH for evaluation for translational-only

    setup correction scenarios. The uncertainty of this method for different clinical scenarios

    was also evaluated using patient study and phantom study. The performance of this

    method, under the scenarios that the variation of breathing pattern, tumor size and

    tumor position occurred between CBCT and planning CT, was assessed using a

    phantom study.

    Study Topics

    Stereotactic radiotherapy (SRT) technique has been widely utilized for various

    treatments sites over the past two decades. 3D conformal radiation therapy, intensity-

    modulated radiation therapy (IMRT) and arc therapy have been routinely applied for

    various SRT treatments in order to escalate target dose while keeping normal tissue

    dose under toxicity threshold 29,30. However, the full potential of these technologies in

    radiation treatment can be achieved only if the patient can be positioned accurately and

    reproducibly during every session of the entire course of treatment delivery 1.

    The need for more precise patient positioning has increased the interest in

    developing 3D imaging techniques that can verify the patient setup immediately before

  • 21

    and after treatment. The availability of large area flat-panel detectors has facilitated the

    development of integrated cone-beam computed tomography (CBCT) systems on linear

    accelerators2-5. Until now, Kilo-Voltage (kV) CBCT systems integrated into the gantries

    of linear accelerators have been pervasively employed as an advanced image-guided

    radiotherapy (IGRT) modality to acquire high-resolution volumetric images of patients

    for treatment localization purposes. Using on-line kV CBCT software and hardware, a

    patient’s position can be determined with a high degree of precision and, subsequently,

    setup parameters can be adjusted to deliver the intended treatment.

    However, in the current research regime, it lacks specific strategies for various

    sites on how to achieve the optimal patient setup though the guidance of CBCT system

    and pertinent assessment of their dose consequence. Most of the registration strategies

    are based upon the methodology of intensity matching between reference and

    localization images for 6 degree freedom shift, which is not always practical for all the

    clinical scenarios and even possess intrinsic flaws for some specific site. Meanwhile the

    relationship between the patient’s dose received from the treatment (beam delivery and

    CBCT scan itself) and setup alignment strategies had not been sufficiently addressed.

    This work provided specific strategies for CBCT-guided setup procedures to

    achieve optimal dose effects for various sites as well as assess the dose consequence

    of those strategies. We started with the assessment of the patient’s dose generated by

    CBCT scan itself. It involved with the investigation of the effect of CBCT on registration

    accuracy and image qualities with a reduced number of planar projections used in

    volumetric imaging reconstruction as well as ultimately seeking for the optimal

    projection-number needed for reconstruction to balance between the dose and

  • 22

    registration accuracy. Then we addressed the impact of small rotational errors on the

    magnitudes and distributions of spatial dose variations for intracranial stereotactic

    radiotherapy (SRT) treatment setups. Furthermore, a quick online evaluation method to

    assess the dose consequence due to small rotational error was introduced. Eventually,

    in lieu of intensity matching based approaches, we developed a novel method to

    facilitate the lung SBRT setup for translational-only scenarios in order to achieve

    optimal tumor dose coverage. The thorough evaluation and efficiency improvement of

    this method were also addressed using a phantom study.

    Topic 1: Evaluate the effect of CBCT on registration accuracy and image quality

    with a reduced number of planar projects.

    The image qualities of 3D volumetric images reconstructed with a decreasing

    number of projections was quantitatively evaluated using a CT phantom. The effects of

    imaging-registration accuracy under a dose-decreasing strategy of reducing the project

    number of different clinical sites were examined. Optimal frame reduction scheme which

    can balance between the dose and registration accuracy for various sites was provided

    for various sites.

    Topic 2: Investigate the impact of small rotational errors on the magnitudes and

    distributions of dose variations for intracranial SRT treatment setup.

    On top of DVH analysis of dose consequence resulting from systematic rotational

    setup errors, we further investigated the dose impact on the magnitudes and distribution

    of the dose map variations for intracranial SRT treatment setups. The comparison

    between planned dose and regenerated dose through rotation-simulated imaging was

  • 23

    presented. It included the mean and standard deviations of differential dose maps in

    CTVs, OARs and different layers of rotated brain anatomy.

    Topic 3: Develop an DVH-based online evaluation algorithm for dose

    consequences caused by small rotational setup errors in intracranial stereotactic

    radiation therapy.

    The feasibility of using the original dose map overlaid with rotated contours

    (ODMORC) method as a fast, online evaluation tool to estimate dose changes to CTVs

    and OARs caused by small rotational setup errors will be assessed. A Fast algorithm for

    multi-contour rotation, which can expedite the ODMORC method speed drastically, was

    developed

    Topic 4: Develop a CBCT-guided patient setup strategy based on deformable

    registration technique for lung SBRT in order to achieve the optimal tumor dose.

    A new registration strategy for translation-only correction scenarios of lung SBRT

    setups was developed. Its effect will be evaluated by comparing ITV’s dosimtric

    coverage of this method with bony and grey value methods, especially for those cases

    as the tumor position varies greatly from its position on the original CT relative to the rib

    cage.

    Topic 5: Compare the uncertainties of the method from AIM 4 with other new

    methods for the scenarios of breathing pattern, tumor position and tumor size change

    using phantom study.

    Various scenarios of breathing pattern change between planning CT and CBCT

    scans, which involved exhalation/inhalation ratios change, motion trajectory length

    change, were simulated by a 4D lung phantom. The scenarios with tumor position

  • 24

    change and tumor size change were also included. The corresponding scans for each

    scenario were performed with 4DCT, CBCT and 4DCBCT. The comparisons on

    registration accuracy of ITV centroids as well as dosimetric consequence of ITVs

    among method suggested by Topic 4 and Translational-only Grey value registration

    were performed.

  • 25

    CHAPTER 2 DECREASING THE KV-CBCT DOSE BY REDUCING THE PROJECTION NUMBER

    Abstract

    The objective of this study was to evaluate the effect of Kilo-Voltage cone-beam

    computed tomography (CBCT) on registration accuracy and image qualities with a

    reduced number of planar projections used in volumetric imaging reconstruction. The

    ultimate goal is to evaluate the possibility of reducing the patient dose while maintaining

    registration accuracy under different projection-number schemes for various clinical

    sites. An Elekta Synergy Linear accelerator with an on-board CBCT system was used in

    this study. The quality of the Elekta XVI cone-beam 3Dimensional volumetric images

    reconstructed with a decreasing number of projections was quantitatively evaluated by a

    Catphan phantom. Subsequently, we tested the registration accuracy of imaging data

    sets on 3 rigid anthropomorphic phantoms and 3 real patient sites under the reduced

    projection-number (as low as 1/6th) reconstruction of CBCT data with different rectilinear

    shifts and rotations. CBCT scan results of the Catphan phantom indicated the CBCT

    images got noisier when the number of projections was reduced, but their spatial

    resolution and uniformity were hardly affected. The maximum registration errors under

    the small amount transformation of the reference CT images were found to be within 0.7

    mm translation and 0.3º rotation. However, when the projection number was lower than

    one-fourth of the full set with a large amount of transformation of reference CT images,

    the registration could easily be trapped into local minima solutions for a non-rigid

    anatomy. We concluded, by using projection-number-reduction strategy under

    conscientious care, imaging-guided localization procedure could achieve a lower patient

    dose without losing the registration accuracy for various clinical sites and situations. A

  • 26

    faster scanning time is the main advantages compared to mAs decrease-based dose-

    reduction method, which ultimately reduce the patient-on-couch time to alleviate the

    setup uncertainty.

    Background

    Advances in 3Dimensional (3D) radiation therapy technologies have led to the

    safe delivery of escalated doses to tumors, improving local control and sparing dose to

    healthy tissues for improving quality of life31. However, the full potential of these

    technologies in radiation treatment can be achieved only if the patient can be positioned

    accurately and reproducibly during every session of the entire course of treatment

    delivery1. Mega-voltage (MV) portal imaging has been widely implemented for the past

    2 decades as patient treatment-positioning tools. Due to the inherent low-contrast and

    2-dimensional nature of the projection images, the precision of MV portal imaging is

    limited so far as accurately defining the patient’s position32. The need for more precise

    patient positioning has increased the interest in developing 3D imaging techniques that

    can verify the patient setup immediately before and after treatment. The availability of

    large-area flat-panel detectors has facilitated the development of integrated cone-beam

    computed tomography (CBCT) systems on linear accelerators33-36. Up until now,

    kilovoltage (kV) CBCT systems integrated into the gantries of linear accelerators have

    been widely used as an advanced image-guided radiotherapy (IGRT) modality to

    acquire high-resolution volumetric images of patients for treatment localization

    purposes. Using on-line kV CBCT software and hardware, a patient’s position can be

    determined accurately with a high degree of precision, and, subsequently, set-up

    parameters can be adjusted to deliver the intended treatment.

  • 27

    While the dose from a single CBCT image acquisition is small compared to the

    therapy dose, repeated daily image-guidance procedures can not only lead to

    substantial dose to normal tissue, but also generate a high integral dose for the

    scanning area of the patient.1 Radiation therapy patients are already being exposed to

    very high and localized radiation; therefore, the additional radiation from imaging has an

    associated risk and should be kept low.2 Concerns about the stochastic risk of inducing

    cancer or genetic defects have already been accounted for in the limits on leakage from

    the primary therapy beam, which should not exceed 0.2% of the absorbed dose rate on

    the central axis at the treatment distance.3 This recommendation suggests that for a

    total treatment dose of 60 Gy delivered over 30 fractions, only 120 mGy or less should

    be contributed by MV radiation leakage. Although the energy and field of exposure differ

    for MV beam leakage with CBCT, making direct comparison problematic, the imaging

    doses are clearly not negligible. The health and safety considerations that underlie the

    limit on therapy-beam background dose should, therefore, be considered as relevant for

    imaging exposures as well.3 How to balance imaging-quality improvement and dose-

    escalation risk over broad clinical scenarios is still a controversial topic in the radiation

    therapy community and would be beyond the scope of this article. In this study, we

    wanted to focus on the impact of image information resulting from dose-reduction

    strategies.

    In general, dose can be minimized during CBCT scan through 2 different means:

    one is by reducing the tube current; the other is by reducing the projections used in

    volumetric imaging reconstruction.3 The first method would result in a reduced signal-

    noise ratio due to less incident photons (reduced mAs) interacting with detectors, which

  • 28

    ultimately degrade the image quality from several perspectives: spatial resolution, soft-

    tissue contrast, edge preservation, etc. Those degrading effects have been studied by

    several investigators, and solutions to alleviate them have been proposed.4-6

    Alternatively, the second strategy to reduce the patient’s dose underlies the reduction of

    sampling frequency during image acquisition. The major advantage of the second

    method versus the first one is that it is a time saver: faster scan speed and shorter

    reconstruction time could ultimately limit setup uncertainty as a result of less patient-on-

    couch time. Unfortunately, like the mA-reduction method, the disadvantage of the

    second strategy is that image quality diminishes. More specifically, the soft-tissue

    contrast degrades and the streak-shaped artifacts appear. In diagnostic imaging, there

    is a direct relationship between the exposure level and image quality. The demand for

    high-contrast, low-noise images pushes the exposure levels up. In IGRT, patient-

    alignment information derived from imaging registration is the main purpose of on-board

    therapy imaging, which leads to a major concern with IGRT imaging: registration

    accuracy between reference CT and CBCT rather than general image-quality critiques.

    The alignment information generated by volumetric imaging registration will eventually

    provide clinicians with patient-setup accuracy and subsequently position-adjustment

    guidance if necessary. The goal of this article is to examine the effects of imaging-

    registration accuracy under a dose-decreasing strategy of reducing the projection

    number for different clinical sites.

    A preliminary study on the strategy to decrease the radiotherapy dose by

    reducing the projection number was conducted by Sykes et al6 using a head-and-neck

    phantom. Their study showed that for a specific phantom, using a reduced number of

  • 29

    projections (one-fifth of a full set) registration accuracy could be preserved as the on-

    board imaging was reconstructed. However, several aspects of their study could

    potentially be revisited or further engaged with. First, a quantitative analysis of the

    degrading effects of imaging quality due to the reduced projection number was not

    included in their study. Since the clinician’s judgment of image registration heavily

    depends on imaging quality, evaluating the registration accuracy with any proposed

    strategy is as important as assessing the degrading effects of the images, especially

    when automatic registration cannot provide the correct alignment information. Secondly,

    although head-and-neck phantom data demonstrate the effects of registration accuracy

    for a rigid body, real patient imaging can be deformed with respect to the reference CT

    due to setup variance, internal organ motion, and/or biological changes in the patient. It

    is therefore worthwhile to exam whether a proposed strategy can still be applied to non-

    rigid patient imaging. Furthermore, in Sykes et al’s study, only the cross-correlation

    information of the pixel gray value between image sets was used as an objective

    function for registration calculation due to the availability of commercial software at that

    time. Today, integrated software is available for most CBCT sites, with both edge and

    cross-correlation information included in the commercial software as customer options.

    The effects of the proposed strategy should be assessed for all of today’s optional

    modes. Another limitation of Sykes et al’s study is that they failed to provide the results

    of full sets of projection-number-reduction schemes except for a specific scheme (one-

    fifth of the projection number). We believe a more valuable study should demonstrate

    the registration effects with different schemes of projection-number reduction to at least

    provide the researcher with opportunities to seek optimum schemes that balance image

  • 30

    quality with patient dose. In this paper, we provide a thorough study for a dose-

    decreasing method based on projection-number reduction.

    The aim of this study is to evaluate registration accuracy and image quality with a

    reduced number of planar projections for volumetric imaging reconstruction with an

    Elekta on-board CBCT. The image qualities of 3D volumetric images reconstructed with

    a decreasing number of projections were quantitatively evaluated using a CT phantom.

    Subsequently, 3 different sites of phantoms and patients’ imaging data with different

    projection-number schemes were analyzed, respectively.

    Methods and Materials

    X-ray Volume Imaging System

    The X-ray volume imaging (XVI) system (version 3.5) used in this study is an on-

    board kV CBCT imaging system integrated into the Elekta Synergy platform linear

    accelerator (Elekta Corp., Stockholm). The XVI system consists of a kV radiation-

    generation system that produces a cone-shaped X-ray beam, an image receptor

    consisting of a kV imaging panel that acquires projected planar images from the X-ray

    beam, and a control system with computer workstation hardware and XVI software. The

    radiation generator and detector panel are both mounted on retractable arms extending

    from the accelerator’s drum structure in an orthogonal direction to the MV system and

    they share the same rotation center with a gantry-head MV imaging-panel rotation

    system. The X-ray generator’s focus to the iso-center distance is 1,000 mm and the

    distance from the focus to the receptor is 1,536 mm.

    The kV generator is a 40-kilowatt unit capable of producing single and

    continuous radiographic exposures of varying energy and dose. The energy range is

    from 40 to 150 kV. The X-ray-tube filtration is 3.5 mm of aluminum and 0.1 mm of

  • 31

    copper. This equates to a total of approximately 6 mm of aluminum equivalence at 100

    kV and 7 mm of aluminum equivalence at 120 kV. Tube-current options are 25, 32, and

    40 mA while the pulse length can be set within the range of 4 to 80 ms.

    The images for this study were acquired through 360° (full fan) or 200° (half fan)

    rotations. The pulse rate for full fan and half fan were 2.7 Hz and 5.4 Hz respectively.

    The fast gantry speed was 60 seconds. The regular scanning speed was about 120

    seconds. Pixel size was 0.8 mm2. Each projection image was sampled with 512 x 512

    pixels. The maximum image volume was defined by the size of the collimator, and the

    maximum correspondent field of view was 26 cm.

    XVI software was running on a clinical workstation with 2.4 GHz CPU and 2 GB

    of memory. The reconstruction algorithm provided by the XVI is Feldkamp’s back-

    projection algorithm37 considering flex-map correction in rotation orbits.38

    Phantom and Patient Selection

    Catphan phantom

    Image quality drastically impacts the clinician’s judgment of registration accuracy

    when auto-registration cannot provide correct registration results, especially for non-

    rigid body registration.3 Typically, quantitative image-quality tests are performed by

    analyzing a specially designed CT phantom scan. For instance, Catphan phantom

    (model CTP 503, The Phantom Laboratory, Salem) has been recommended by the

    Elekta XVI system as the acceptance test phantom. Reconstructed Catphan phantom

    images with varying projection-number schemes were accessed in our study through 3

    major criteria specified in the acceptance test guide. The Catphan phantom is cylindrical

    and has multiple layers embedded with different shapes and materials of inserts. It

  • 32

    could be used to examine CT number uniformity, spatial resolution, low-contrast

    resolution, and geometric accuracy, among other factors.

    Rando phantom

    Three body parts of a male RANDO phantom (The Phantom Laboratory, Salem,

    NY) were selected as rigid imaging objects for our registration-accuracy study. We refer

    them as the head-and-neck area, thoracic area, and pelvic area of the RANDO phantom

    as R-H&N, R-Thoracic, and R-Pelvis, respectively. The phantom was constructed with a

    natural human skeleton cast inside of soft-tissue-simulating material. In the thoracic

    area, lungs are molded to fit the contours of a natural rib cage. The air space of the

    head, neck, and stem bronchi are duplicated. The phantom is sliced at 2.5-cm intervals.

    The whole-grid patterns can be drilled into the sliced sections to enable the insertion of

    dosimeters. Two tissue-simulating materials are used to construct the RANDO

    Phantom: the RANDO soft-tissue material and the RADNO lung material. Both of these

    are designed to have the same absorption as human tissue at the normal radiotherapy

    and radiology exposure levels. The RANDO’s similarities to a real human structure

    make it an attractive and widely used imaging and dosimetry substitute for simulation

    among the radiology community.

    Patient sites

    Three patients’ image sets at different sites were used for our registration study.

    They were a brain-cancer site, a lung-cancer site, and a prostate-cancer site, which are

    referred to as P-Brain, P-Lung and P-Prostate, respectively. The reference CT images

    were taken a week before treatments for planning purpose whereas CBCT images were

    taken right prior to patient treatment to register with reference CT images for localization

    purpose. Unlike rigid phantom imaging, the real patients imaging data could be

  • 33

    deformed between reference CT and CBCT image sets. For example, patient’s prostate

    could be stretched or squeezed by its surrounding tissues and organs (Different amount

    of urine inside bladder and/or different amount of gas inside rectum between the

    moments of acquiring reference CT and CBCT can cause different shapes of the

    prostate in these two image sets); chest cage could be enlarged or shrank depending

    on at which phase of breathing while acquiring images. Since the registration solution of

    non-rigid imaging is much easier to trap into a local minima solution, assessing the

    robustness of the registration algorithm with non-rigid patient data would make the study

    more valuable and thorough.

    Acquisition and Reconstruction of Cbct Images

    The positioning procedure used with the Catphan phantom followed the

    acceptance test guide of the XVI. The RANDO phantom and patients were placed

    centrally in the field of imaging view prior to the scan. All CBCT images of phantoms

    and patients were acquired by the standard technique recommended by XVI protocol

    systems. The technique parameters, including beam energy, tube current, duration per

    pulse, projected image number, scan time, gantry rotation angle, and collimator type,

    are listed in Table 2-1. Since the recommended technique could generate near-optimum

    image quality for the current system, those 3D images reconstructed with full sets of

    projected planar images were treated as standard images for the registration study and

    referred to as “XVI-full.” The near-optimum image qualities for each site were achieved

    by choosing high tube currents and high energy beams to minimize noises and any

    reconstruction artifacts due to high-density materials inside the phantoms or patients. A

    deliberate, slow gantry speed also benefitted the image quality since it allowed enough

    projected images to be acquired during the slow gantry rotation in order to reconstruct

  • 34

    images with minimal artifacts and structural noise generated from the high-contrast

    edge. The term “optimum image quality” in our study refers to the best image quality

    achievable by the current system with minimal noise and artifacts, and thus minimal

    impact on the accuracy of 3D- image- registration.

    All 3D images were reconstructed using 1-mm3 voxels. The registration outcome

    of the XVI-full scan with the corresponding “reference planning CT scan” dataset was

    applied as the baseline to evaluate the registration accuracy of images reconstructed

    with fewer projections. In order to simulate a series of lower doses and faster scans, the

    reconstruction schemes with fewer projections consisted of half, one-third, one-fourth,

    one-fifth, and one-sixth of the full projections, referred to as XVI-1/2, XVI-1/3, XVI-1/4,

    XVI-1/5, and XVI-1/6, respectively. The image-quality degrading effects along with the

    fewer numbers of projections were analyzed. As the projection number became less

    than one-sixth of the full set, unacceptable image quality became an obstacle to most

    clinicians who could not provide sound judgments about registration results due to the

    low visibility in the soft-tissue contrast. Therefore, even reconstructed images with

    projections less than one-sixth of full set would potentially sustain the edge information,

    registration test with a projection number less than one-sixth of full set had been

    excluded from our study.

    3D imaging reconstructions were performed with XVI software. The schemes

    with fewer projection numbers were achieved by manually deactivating the undesirable

    projection frames in the XVI database. For instance, XVI-1/n was reconstructed only

    with frames whose index number could be divisible by the number n, and all other

    frames would not be used for reconstruction at all. Projection-number schemes used for

  • 35

    patients and phantoms are listed in Table 2-2. It should be noted that this was a

    feasibility study and XVI-1/n does not mean that the scan speed was increased by n

    times, but rather that it was limited by the gantry-speed limitations permitted by the

    International Electrotechnical Commission (IEC) and achievable tube firing frequency.

    The validity of simulating a low-dose fast scan will be discussed.

    Imaging Quality Check with a Reduced Projection Number

    The influence of a reduced number of projections on image quality was evaluated

    using the Catphan phantom. The evaluated items included 3D spatial resolution, 3D

    low-contrast visibility, and 3D uniformity. The “CAT-Image Quality” preset parameters

    (referred to in Table I) were used to acquire the volumetric images. A 360° rotation scan

    was performed and a total of 668 projections were registered in the XVI database. The

    reconstructed 3D images with all projection number schemes were tested separately. A

    detailed description and criteria of those tests follow.

    3D spatial resolution

    We located the spatial-resolution module slice of the reconstructed CBCT image,

    then magnified the image until the module filled the transverse view, and subsequently

    determined the highest numbered line pairs visible by adjusting the brightness and

    contrast to achieve the best view for the line pairs. The specification was greater or

    equal to 7 line pairs per centimeter. Figure 2-1(a) shows the reconstructed images with

    XVI-full, XVI-1/2, and XVI-1/6 schemes at the spatial resolution module slice,

    respectively.

    3D low-contrast visibility

    Figure 2-1(b) shows the contrast resolution module slice reconstructed with XVI-

    full, XVI-1/2, and XVI-1/6 schemes for the 3D low-contrast visibility test. This test

  • 36

    measures the mean pixel values of the polystyrene insert (7 o’clock, Figure 2-1(b)) and

    the LDPE insert (9 o’clock, Figure 2-1(b)), as well as the standard deviation of their pixel

    values. The low-contrast visibility value was defined mathematically as Equation (2-1),

    where “SD” represented the standard deviation of the pixel values; “Mean” represented

    the mean of the pixel value; the subscripts (“polystyrene” and “LDPE”) represented two

    different materials (polystyrene and low density polystyrene) respectively. If the

    calculated value of Equation (2-1) was large, it meant the difference of imaging noises

    between these two materials overwhelmed the difference of true imaging signals

    between them, which indicated poor low contrast visibility; vice-versa. The XVI

    specification for the low-contrast visibility value was lower than 2%.

    ]2

    [

    ][

    5.5

    LDPEepolystyren

    LDPEepolystyren

    SDSD

    MeanMean

    visiblitycontrastlow (2-1)

    3D uniformity

    The uniformity module slice of the Catphan phantom was used as the image for

    the 3D uniformity test. Four random locations in the uniformity module were chosen to

    measure the mean values of the pixels covered by a 1 x 1 cm2 probe window. Equation

    (2-2) was used to compute the maximum percentage difference between any 2 of the

    mean pixel values, where max(mean) and min(mean) are the largest and smallest mean

    pixel values among the 4 locations, respectively. The specification was less than or

    equal to 2%.

    )(

    )()(

    meanMax

    meanMinmeanMaxuniformity

    (2-2)

  • 37

    Registration Analysis

    The reference CT images of the phantoms and patients were acquired by a

    Philips multi-slice high-speed CT scanner (Amsterdam, The Netherlands). The

    placement of the phantoms allowed the scan origin to be near the iso-center used in

    XVI imaging, but no effort was taken to ensure that the center was identical or that the

    phantom was rotationally aligned in the CT scanner with respect to the XVI scans. Real

    patient setup, however, strictly followed department setup procedures to avoid any

    possible rotation variance between the reference CT scan and XVI scan for practical

    clinical purposes.

    Automatic 3D registration of the reference CT and the XVI scans was performed

    using XVI software. Two automatic registration modes were used for our study. The

    “bone mode” of the automatic registration uses a chamfer matching algorithm originally

    described by Borgefer.21 The chamfer algorithm performs registration by matching the

    edge information of 2 image sets. It is relatively insensitive to image noise and mild

    streak artifacts, and also has a quick computing time. It performed poorly with lower-

    contrast subjects only. Comparisons of the quality of registration with the bone mode

    with other techniques have been provided by several investigators.22,23 The “grey value

    mode” performs the auto-registration by matching voxel grey-scale intensity values

    throughout the entire interested image volume, which is well known as the “cross

    correlation” technique.25 In addition to the edge information, soft-tissue information has

    also been included in the grey-value algorithm and would potentially average out the

    registration error caused by anatomy variance between the 2 image sets. In terms of

    calculation time, the grey-value mode is much worse than the bone mode in general. In

  • 38

    our study, we had tried the bone mode for all patient and phantom images, and the

    grey-value mode only for patient images of pelvis and lung sites.

    The performance of the automatic registration was checked by repeatedly

    registering the reference CT planning scan with the XVI scan for each projection

    number scheme. We initially registered reference CT imaging with XVI-full imaging and

    considered the registration results as “Gold Standards;” the recorded translation and

    rotation values were then treated as baseline and were subtracted from the values of

    the registration results of other schemes to acquire the registration error values for all

    schemes accordingly. To simulate setup error in terms of translation and rotation, we

    manually entered the translation value in the x, y, and z axes and rotation angle in the x,

    y, and z directions to achieve the desirable transformed reference-CT datasets. Three

    independent translations with values of 2 mm, 5 mm and 20 mm along each axis of x, y,

    and z, and 2 independent rotations with values of 3° and 10° around each axis of x, y,

    and z, were applied to the reference CT scan. The transformed reference CT images

    were then registered with XVI images reconstructed with different numbers of

    projections to test the registration accuracy. Additionally, 2 combinations of translation

    and rotation were applied to the reference-CT planning scan for all 3 axes to access the

    registration capability with various scheme reconstructions for more realistic reference-

    CT transforms as well. To demonstrate that registration was reproducible, each

    registration procedure was performed 10 times repeatedly, and the mean and standard

    deviation values in terms of translation vector error and max rotation angle error of all

    directions were recorded.

  • 39

    Results

    Reconstructed Images

    The XVI reconstruction time of patients and phantoms with different projection-

    number schemes are listed in Table 2-3. Obviously, fewer projection numbers resulted

    in shorter computing times. XVI-1/6 reconstruction procedures lasted an average of 4.8

    seconds over all patients and phantoms, which is about 1/6th of the average time spent

    for all XVI-full reconstructions. The representative trans-axial, sagittal, and coronal

    slices of the XVI-full, XVI-1/2, and XVI-1/6 schemes for 3 RANDO phantom parts and 3

    different patient sites are shown in Figures 2-2 and 2-3, respectively. As predicted,

    detailed contrast of soft tissue in XVI imaging could not be visualized as well as with

    reference-CT imaging, even with full-projection reconstruction. XVI-full showed some

    low-frequency noise artifacts that created non-uniformity on the grey level in the tissue-

    equivalent material and some streak-shaped artifacts originating from high-contrast

    edges (like the couch top). It was noticed that XVI-1/2 images were slightly noisier than

    XVI-full; however, streak artifacts were not more severe than XVI-full, while edge

    information between the anatomy interfaces was well kept. XVI-1/6 images presented

    significant streak artifacts with the appearance of an interference pattern due to the

    under-sampling effect of the reconstruction. Reduced visibility in the overall soft-tissue

    contrast was also observed in the XVI-1/6 images. However, even in the XVI-1/6

    images, the gross bony anatomy and air cavity were geometrically undistorted and a

    surprising level of fine edge details still remained whereas the presence of degrading

    effects of soft-tissue contrast escalated.

    Imaging-quality Check

    The image quality results were provided quantitatively as follows.

  • 40

    3D spatial resolution check

    Figure 2-1(a) presents the image slice of the spatial-resolution module of the

    Catphan phantom with XVI-full, XVI-1/2, and XVI-1/6 reconstruction schemes

    separately. The highest number of visible line pairs of XVI-1/6 is 7. Increases in

    projection number led to better visibility of line pairs, especially for XVI-full and XVI-1/2

    for which line-pair 8 could easily be visualized. In summary, all tests satisfied the criteria

    (visibility of line-pair 7) provided by the XVI acceptance manual.

    3D low-contrast visibility check

    Figure 2-1(b) presents the 3D low-contrast visibility module slice of the Catphan

    phantom with XVI-full, XVI-1/2, and XVI-1/6 reconstruction schemes separately. Clearly,

    the imaging interference from noises became more severe as the number of projections

    used for 3D reconstruction decreased. Figures 2-4(a) and 2-4(b) show the mean CT

    number and standard deviation of pixels inside the 2 inserts regions (described in the

    Methods and Materials) versus different reconstruction schemes, respectively. The

    results show that mean pixel-value curves were almost flat, whereas the standard

    deviation increased as the number of projections decreased, indicating that increased

    noise due to under-sampling does significantly influence detailed contrast visibility.

    Figure 2-4(c) summarizes the calculated values of low-contrast visibility by Equation (2-

    1) versus the different reconstruction schemes. The XVI specification for the low-

    contrast visibility value was less than 2%. Only XVI-full and XVI-1/2 met the

    specification; others failed. More specific for clinic scenario, the degradation due to

    under-sampling effect would be the appearance of soft tissue contrast lost in CBCT

    images. For instance, muscle tissues could be clearly visualized in XVI-full image set of

    the prostate patient (referring figure 2-3(c)) whereas the appearance of muscle tissues

  • 41

    in XVI-1/6 images had been degraded (merged by surrounding tissues due to high

    noises).

    3D uniformity check

    All tests met the criteria. The results had been presented in Figure 2-5. No signs

    indicate that uniformity worsened with fewer projection numbers.

    Registration

    The mean registration errors of 10 repeated measurements in terms of

    translation vector error and max rotation-angle error versus different reconstruction

    schemes are presented as scatter charts in Figures 2-6 for Phantoms with “bone mode”

    registration algorithm (Fig. 2-6(a) head-and-neck phantom, Fig. 2-6(b) thoracic

    phantom, Fig. 2-6(c) pelvis phantom), Figure 2-7 for all patient sites with “bone mode”

    registration algorithm (Fig. 2-7(a) brain site, Fig. 2-7(b) lung site, Fig. 2-7(c) prostate

    site) and Figure 2-8 for non-rigid patient sites with “grey mode” registration algorithm

    (Fig. 2-8(a) lung site and Fig. 2-8(b) prostate site). Each chart located at left side of

    each subfigure showed the result data of translation vector error; the one at right side

    showed the result data of max rotation-angle error. Since each subfigure had same

    structures, we only described them in general but not repeated them individually. In the

    left-side chart of each subfigure, X axis represented the different schemes of XVI

    reconstruction; the Y axis represented the mean registration error of 10 repeated

    measurements for the translation vector only; 7 series labels with different shapes and

    colors represent 7 different translation and rotation combinations applied to the

    reference-CT images. To facilitate the visibility of overlapped error points across

    different series, the color lines between adjacent points of same series had been added

    to help reader distinguish those error points among different series. These combination

  • 42

    series consisted of 2 mm shifts along the x, y, and z axes called “T = 2 mm,” 5 mm

    shifts in the x, y, and z axes called “T = 5 mm,” 20 mm shifts in the x, y, and z axes

    called “T = 20 mm,” 3° rotations in x, y, and z directions called “R = 3°,” 10° rotations in x,

    y, and z directions called “R = 10°” 2 mm shifts along the x, y, and z axes plus 3°

    rotations in the x, y, and z directions called “T = 2 mm, R = 3°,” and 20 mm shifts along

    the x, y, and z axes plus 10° rotations in the x, y, and z directions called “T = 20 mm, R

    = 10°”. In each right-side chart of each subfigure, Y axis represented means of the max

    angle error over all directions of 10 repeated measurements; other labels had same

    meaning as the right-side chart and would not be repeated here. Standard deviations of

    the registration error over 10 repeated measurements were 0.1 mm or less in translation

    and 0.10 or less in rotation across the board. The superimposed standard deviation bars

    were excluded from the charts to avoid the visibility interference of mean values. The

    mean registration time by bone mode was 4.2 seconds and the standard deviation was

    1.3 seconds; the mean registration time by grey-value mode was 15.6 seconds and the

    standard deviation was 3.2 seconds

    Discussion

    The results indicate that, for image sets of rigid body (e.g. brain phantom, pelvis

    phantom and brain site), when treatment volume can be determined by bone-tissue

    and/or tissue-air interface, without clinician interference, auto-registration results using

    reduced-projection-number (up to one sixth of projections) reconstruction can still

    provide accurate enough alignment information with no obvious trends of error

    escalation while projection number decreased. Error points were randomly scattered in

    charts(shown in Fig.2-6(a), Fig 2-6(c) and Fig2-7(a)) with maximum errors of 0.7 mm in

    translation and 0.3° in rotation except of the thoracic phantom for which a large

  • 43

    misalignment (10° rotations and/or 20 mm translation shown in Fig.2-6(b)) was applied

    in the first place. For non-rigid anatomy sites like the lung and prostate, however, auto-

    registration with too few (one fifth or one sixth) projection-number reconstructions could

    lead to greater alignment errors (shown in Fig.2-7(b), Fig.2-7 (c), Fig. 2-8(a) and Fig. 2-

    8(b)) unless manual pre-alignment was applied. Meanwhile, the overall reproducibility of

    registration results was shown to be excellent (less than 0.1 mm; 0.1° standard

    deviation), whereas other geometric uncertainties that exist in the radiotherapy

    treatment process are much more significant.

    Three major image-quality parameters with different scheme reconstructions are

    shown in Figures 2-2, 2-3 and 2-4. As demonstrated, high-contrast details and spatial

    resolution were well preserved even with the lowest projection-number reconstruction

    (XVI-1/6) tested, whereas low-dose fast scans may not be sufficient to visualize low-

    contrast soft tissue with an ultra-low projection-number reconstruction due to increased

    noise and streak artifacts resulting from under-sampling effects. It could be concluded

    that for rigid-bony anatomy sites, like the brain, an ultra-low projection-number-reduction

    strategy may still be a trouble-free solution to tremendously reducing the patient dose,

    even if soft-tissue-contrast visibility was completely lost, whereas for non-rigid soft-

    tissue anatomy sites, like the prostate and lung, reconstruction with too few projections

    could trap the registration results into local minima, making the clinician’s interference

    for registration unavoidable. Thus, it is much more sensible to optimize soft-tissue

    visibility for rigid sites versus non-rigid sites. It is conceivable that imaging-processing

    techniques that could reduce noise and enhance soft-tissue contrast would be preferred

    when soft-tissue contrast becomes prominent for imaging registration. It was also

  • 44

    noticed that Figure 2-2(c) and 2-3(c) showed some ring artifacts caused by imperfect

    CBCT detector calibration in current version of XVI software. They are common artifacts

    for all the ELEKTA machines. In general, we assumed subtle ring artifacts should not

    influence registration accuracy since registration results were dominated by bony

    structure. However, to validate this assumption, a sophisticated comparison between

    registration outcomes with ring artifact removal and without removal is needed. Due to

    limited accessibility to the imaging data, such analysis is not feasible for this study.

    As mentioned, a local minima trapping situation occurred with the ultra-low

    projection reconstruction (XVI-1/4) for non-rigid sites (lung and prostate) due to the

    combined effects of increased noises, non-rigid anatomy, and large misalignments. For

    these 2 sites, as the 10° rotation and ultra-low reconstruction scheme were applied

    simultaneously, large translation and rotation errors led to the registration results failing

    regular clinical registration criteria (2 mm in translation and 3° in rotation), which

    indicates that reconstruction with too few projection numbers (XVI-1/4 or worse) will be

    the “scare factor” for some non-rigid anatomy sites, and initial human interference is

    imperative to overcome the local minima trapping effect. Another local minima trapping

    situation occurred while we were performing rigid thoracic phantom registrations with a

    setup simulation that had a large initial misalignment. As “R = 10°” or “T = 20 mm, R =

    10°” were applied to the reference CT, the registration results were mismatched by 1

    complete vertebra superiorly or inferiorly due to the similarity of adjacent vertebra and

    ribs in the thoracic region. Like the non-rigid cases, this mismatching could be

    conquered by initial manual alignment followed by the auto-registration procedure. For a

    real patient’s lung, auto-registration after applying a large misalignment performed fairly

  • 45

    well simply because a real patient has more irregular internal structures than a phantom

    structure (compare Figures 2-2(b) and 2-3(b)); therefore, the registration results are less

    likely to fall into the local minima and there is more time to seek a global minima

    solution. This mismatching occurred for all schemes from XVI-full to XVI-1/6 so that the

    projection number was irrelevant.

    In our study, bone mode and grey-value mode did not present any significant

    differences for all schemes of reconstruction in terms of registration accuracy. However,

    the computing speed was the key point distinguishing these 2 methods. On average,

    the bone mode ran almost 4-times faster than the grey-value mode. Thus, the bone

    mode would be recommended to save on procedure time for applicable cases.

    The reconstruction time for XVI-1/6 is 5.2 seconds on average using the XVI

    software. Considering that the average XVI-full reconstruction time was 30 seconds, we

    could save roughly 25 seconds for reconstruction, which is helpful but not significant.

    However, considering clinical settings, image reconstruction is typically concurrent with

    image acquisition and the reconstruction time is masked, therefore reconstruction time

    saving is not an important factor. The real time-saving factor could be generated by the

    fact of fewer projections needed during scanning as mentioned previously. However,

    even though XVI-1/6 reconstruction only needed 1/6th of the full-set projection number,

    it does not appear that that the potential scan time for the XVI-1/6th scheme could be

    escalated by 6 times. For example, a XVI-full scan of 652 projections needs an average

    of 2 minutes for the pelvis, whereas a 110-projection scan could not be accomplished in

    20 seconds due to the current limitations on the maximum gantry speed, the highest

    tube firing frequency, and patient-safety concerns. Since the maximum gantry speed is

  • 46

    limited to no greater than 60/s, in order to comply with IEC regulations,39 less than a 1-

    minute scanning time is impossible according to current safety regulations. Therefore

    the scanning time could be reduced from current 2 minutes to 1 minute. Since the

    system is designed for automatic registration integrated with software for XVI acquisition

    and reconstruction, then it should be possible to perform the complete image-guided

    process of image acquisition, reconstruction, registration, and correction of patient

    positioning within 3 minutes instead of the usual 5 minutes. This would make on-line

    image guidance a practical and realistic procedure

    Previous discussions indicate that an XVI-full scan is not necessary when auto-

    registration could provide accurate alignment information. For all the cases in our study,

    it appears that XVI-1/4 would allow a good balance between dose reduction and

    accurate-enough registration. Song’s study40 elucidates that for the average person, the

    brain, lung, and prostate doses to the patient’s surface (i.e., the highest dose to the

    patient) per XVI-full scan are 0.22 cGy, 4.6 cGy, and 2.4 cGy , respectively. Intuitively, a

    XVI-1/4 scan would give one-fourth of the dose of an XVI-full scan to the patient’s

    surface for any site. More specifically, 0.055 cGy would be the skin dose from a XVI-1/4

    brain-site scan, and for the lung site and prostate site they would be 1.15 cGy and 0.6

    cGy respectively. If we assume that the average prescription dose for IMRT or 3DCT is

    200 cGy, and we perform a cone-beam scan at each fraction, the percentage dose from

    XVI to MV would be 0.02%, 0.6%, and 0.3%. Adopting 0.2% of a MV dose as the dose

    limit recommended by Murphy, et al.,3 the XVI dose to the brain site by a XVI-1/4 scan

    would satisfy this dose limit; dose from XVI-1/4 scan for the other 2 sites would near the

    limit. In hypo-fractionation stereotactic body radiotherapy cases, the patient’s average

  • 47

    MV dose is 1000cGy, so the CBCT dose for all sites would be well below the limit.

    However, since no safe radiation level exists, the ALARA (“as low as reasonably

    achievable”) standard for non-target doses for concomitant exposures is always

    recommended by the radiology community. The radiation carcinogenesis risk for

    patients undergoing radiati