STRATEGIES FOR KVCBCT-GUIDED PATIENT SETUP OF...
Transcript of STRATEGIES FOR KVCBCT-GUIDED PATIENT SETUP OF...
-
1
STRATEGIES FOR KVCBCT-GUIDED PATIENT SETUP OF STEREOTACTIC RADIOTHERAPY AND ASSESSMENTS OF THEIR
DOSE CONSEQUENCES
By
BO LU
A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT
OF THE REQUIREMENTS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY
UNIVERSITY OF FLORIDA
2013
-
2
© 2013 Bo Lu
-
3
To my beloved wife, Ying, and son, Yiwen
-
4
ACKNOWLEDGMENTS
I would like to gratefully and sincerely thank my advisor, Dr. Chihray Liu, for his
guidance. His insights and suggestions are always inspiring. I really appreciate his
willingness to offer prompt help at any time. His diligence, perseverance and serious
attitude on both research and clinic set a good example for me. He is also the person
who introduced me into the exciting medical physics field as well as encouraged me to
come back to the academic world to pursue my PhD after two years clinic practice in
Mississippi. His unmatched enthusiasm inspired me to devote my career to medical
physics research and practice. I really thank him for his continuous encouragement and
support. My thanks extend to my committee members, Dr. Sanjiv Samant and Dr.
Jonathan Li and Dr. William M. Mendenhall for their input and many valuable
suggestions. I thank Dr. Darren L. Kahler for his comments, suggestions and hard
editing work on my papers.
I need to express my special gratitude to Ms. Kate Mittauer, Ms. Soyong Lee and
Dr. Yin Huang. Without their selfless help on the experiments, I would not achieve any
goal on the topic of the tumor alignment for lung SBRT, which is one of the critical
components of my PhD research. I enjoyed numerous inspiring discussions with them. I
would also like to sincerely thank Dr. Jean Peng from University of South Carolina for
her generous support on providing the code of image and contour rotation operation,
which I am really indebted to.
Most of all, I thank my wife and my son, my parents and my sister for their great
unconditional support, love and encouragement.
-
5
TABLE OF CONTENTS
page
ACKNOWLEDGMENTS .................................................................................................. 4
LIST OF TABLES ............................................................................................................ 8
LIST OF FIGURES .......................................................................................................... 9
ABSTRACT ................................................................................................................... 11
CHAPTER
1 INTRODUCTION .................................................................................................... 13
General Background ............................................................................................... 13 the Relationship between Cbct Scan Dose and Registration Accuracy ............ 14 Dose Consequence Due to Small Rotational Errors ......................................... 15 Cbct-Guided Setup Strategy for Lung Sbrt ....................................................... 17
Study Topics ........................................................................................................... 20
2 DECREASING THE KV-CBCT DOSE BY REDUCING THE PROJECTION NUMBER ................................................................................................................ 25
Abstract ................................................................................................................... 25 Background ............................................................................................................. 26 Methods and Materials............................................................................................ 30
X-ray Volume Imaging System ......................................................................... 30 Phantom and Patient Selection ........................................................................ 31
Catphan phantom ...................................................................................... 31 Rando phantom ......................................................................................... 32 Patient sites ............................................................................................... 32
Acquisition and Reconstruction of Cbct Images ............................................... 33 Imaging Quality Check with a Reduced Projection Number ............................. 35
3D spatial resolution .................................................................................. 35 3D low-contrast visibility ............................................................................. 35 3D uniformity .............................................................................................. 36
Registration Analysis ........................................................................................ 37 Results .................................................................................................................... 39
Reconstructed Images ..................................................................................... 39 Imaging-quality Check ...................................................................................... 39
3D spatial resolution check ........................................................................ 40 3D low-contrast visibility check .................................................................. 40 3D uniformity check ................................................................................... 41
Registration ...................................................................................................... 41 Discussion .............................................................................................................. 42
-
6
Summarization ........................................................................................................ 47
3 DOSE CONSEQUENCES CAUSED BY SMALL ROTATIONAL SETUP ERRORS ................................................................................................................ 57
Abstract ................................................................................................................... 57 Background ............................................................................................................. 58 Methods and Materials............................................................................................ 62
Descriptions of Selected Srt Cases .................................................................. 62 Case selection ........................................................................................... 62 Contour delineation .................................................................................... 62 Treatment techniques and plan evaluation criteria ..................................... 63 Patient setup .............................................................................................. 64
Rotation Simulations and Rotational Angle Selections ..................................... 65 Comparison between Dose Maps with and without Small Rotational Errors .... 66 Quantitative Comparisons among Original, Recalculated, and Odmorc-
generated Dvhs for Small Rotational Errors .................................................. 67 Ufpsas Workflow and a Fast Algorithm for Multi-contour Geometric Shift ........ 67
Results .................................................................................................................... 70 Comparison of Dose Maps for Rotated Rois .................................................... 70 Comparison of Dvhs for Ctvs and Oars ............................................................ 71 Summary of Calculation Speed for the Rotation Algorithm............................... 72
Discussion .............................................................................................................. 73 Summarization ........................................................................................................ 79
4 CENTROID-TO-CENTROID (CTC) STRATEGY FOR LUNG SBRT SETUP ......... 92
Abstract ................................................................................................................... 92 Background ............................................................................................................. 93 Methods and Materials............................................................................................ 96
Centroid-To-Centroid Registration Strategy ..................................................... 96 Case Selections ............................................................................................... 97 Clinic Workflow for Sbrt Lung ........................................................................... 98 Ctc Simulations and Dose Calculations ............................................................ 99 Deformable Registration ................................................................................. 100 Statistical Comparison among Various Registration Algorithms ..................... 102 Uncertainty Analysis ....................................................................................... 102
Uncertainty check of deformable registration results ............................... 102 ITV variation check during treatment course ............................................ 104 Concordance check of Itv between Cbct and A4dct ................................. 104
Results and Discussion......................................................................................... 105 Summarization ...................................................................................................... 113
5 THE EVALUATION OF CTC PERFORMANCE ON THE SCENARIOS WITH BREATHING VARIATIONS .................................................................................. 120
Abstract ................................................................................................................. 120
-
7
Background ........................................................................................................... 121 Methods and Materials.......................................................................................... 123
Phantom Description ...................................................................................... 123 Design and Simulation of Respiratory Variation ............................................. 124 Treatment Plan and Itv Delineations .............................................................. 127 Registration Comparison between T-only G Mode and Ctc method ............... 127
Results and Discussion......................................................................................... 129 the Comparison of Centroid Discrepancies between Two Methods ............... 129 DVH Analysis of Two Extreme Cases ............................................................ 133
Summarization ...................................................................................................... 134
6 CONCLUSIONS ................................................................................................... 142
Summarization ...................................................................................................... 142 Future Works ........................................................................................................ 143
LIST OF REFERENCES ............................................................................................. 145
BIOGRAPHICAL SKETCH .......................................................................................... 151
-
8
LIST OF TABLES
Table page 2-1 Technical parameters lists of scan for phantoms and patients ........................... 49
2-2 Projection number of different schemes for phantoms and patients used in study ................................................................................................................... 49
2-3 Xvi computing time of reconstructions for patients and phantoms with different projection-number schemes ................................................................. 49
3-1 Summary of diagnosis information and plan technique for the 15 cases in this study ................................................................................................................... 81
3-2 Roi statistical results of dose differences between original doses and recalculated doses on rotated images. ............................................................... 82
4-1 Case characteristics ......................................................................................... 114
4-2 Statistic results of geometric discrepancies among ctc, b and g methods ........ 115
4-3 Uncertainty check ............................................................................................. 116
4-4 Itv volumetric variations from a4dct to cbct ....................................................... 116
5-1 The list of scanned cases with combination of various scenarios ..................... 135
5-2 Itv Registration results in terms of centroid alignment discrepancy for ctc and t-only methods .................................................................................................. 136
-
9
LIST OF FIGURES
Figure page 2-1 Catphan reconstruction results.. ......................................................................... 50
2-2 Rando phantom reconstruction results.. ............................................................. 51
2-3 Patient reconstruction results.. ........................................................................... 52
2-4 Results of a low-contrast visibility check with different reconstruction schemes. ............................................................................................................ 53
2-5 Results of a uniformity check with different reconstruction schemes. ................. 53
2-6 The scatter charts of mean registration errors versus reconstruction schemes for Phantoms with “bone mode” registration algorithm.. ..................................... 54
2-7 The scatter charts of mean registration errors versus reconstruction schemes for all patient sites with “bone mode” registration algorithm. ............................... 55
2-8 The scatter charts of mean registration errors versus reconstruction schemes for non-rigid patient sites with “grey value” registration algorithm. ...................... 56
3-1 Clinical setup for srt treatment ............................................................................ 84
3-2 Definition of the four layers for investigating various degrees of dose variations due to rotations. ................................................................................. 85
3-3 Flowcharts of contour geometric rotation algorithm. ........................................... 86
3-4 A 2-Dimensional example of coupling and decoupling processes for multi-contour rotation operation. .................................................................................. 87
3-5 Distribution of the percent dose difference between the original dose map and recalculated dose map on the rotated image using rotation-simulated ct images and Rois in axial view cut ....................................................................... 88
3-6 Dvh curves of the ctv and the oars of case 1 for odoc, odrc and rdrc with rotations of +5o in all three directions .................................................................. 88
3-7 Absolute percentage differences of doses for ctvs at 99%, 95% and 20% of
ctv volume among odoc, rdrc and odrc with ±5⁰ rotations .................................. 89
3-8 Number of cases cataloged by magnitudes of dose variations at oars dose limit volume among odoc, rdrc and odrc with rotations. ...................................... 90
3-9 Schematic drawing of the treatment depth change due to head rotation. ........... 91
-
10
4-1 Comparison of ITV registration results by rigid registration .............................. 117
4-2 Schematic draw of tumor motion patterns and Itv positions during 4dct and cbct scans for registration uncertainty check on a respiration motion phantom. .......................................................................................................... 117
4-3 Registration shifts in three directions among b, g and ctc methods. ................ 118
4-4 Statistic results of dose coverage differences of Itv by b, g and ctc methods ... 118
4-5 Dvh illustrations of three extreme cases ........................................................... 119
5-1 The Illustration of components of cirs dynamic thorax phantom (model 008). .. 137
5-2 Simulated respiratory profiles with five different inspiration to expiration ratios. ................................................................................................................ 138
5-3 Coronal and transvers views of a4dct image data sets from three different cases ................................................................................................................ 139
5-4 Registration workflows and results for case no.5 for both ctc and t-only methods ............................................................................................................ 140
5-5 Dvhs of two extreme cases with iso-center determined by different strategies and beam aperture defined by different sizes of ptv ......................................... 141
-
11
Abstract of Dissertation Presented to the Graduate School of the University of Florida in Partial Fulfillment of the Requirements for the Degree of Doctor of Philosophy
STRATEGIES FOR KVCBCT-GUIDED PATIENT SETUP OF STEREOTACTIC RADIOTHERAPY AND ASSESSMENTS OF THEIR DOSE CONSEQUENCES
By
Bo Lu
August 2013
Chair: Chihray Liu Major: Biomedical Engineering
Stereotactic radiotherapy (SRT) technique has been widely utilized for various
treatments sites over the past two decades. To be able to escalate target dose while
keeping normal tissue dose under toxicity threshold during every session of the entire
treatment course, high accuracy and reproducibility of patient positioning were required.
Until recent, Kilo-Voltage (kV) CBCT systems integrated into the gantries of
linear accelerators have been pervasively employed as an advanced image-guided
radiotherapy (IGRT) modality to acquire high-resolution volumetric images of patients
for treatment localization purposes. However, in the current research regime, it lacks
specific strategies for various sites on how to achieve the optimal patient setup and
pertinent assessment of their dose consequence.
Most of the registration strategies are based upon the methodology of intensity
matching between reference and localization images for 6 degree freedom shift, which
is not always practical for all the clinical scenarios and even possess intrinsic flaws for
some specific site. Meanwhile the relationship between the patient’s dose received from
the treatment (beam delivery and CBCT scan itself) and setup alignment strategies had
not been sufficiently addressed.
-
12
This work provided specific strategies for CBCT-guided setup procedures to
achieve optimal dose effects for various sites as well as assess the dose consequence
of those strategies. It includes the following works: 1.the assessment of the patient’s
dose generated by CBCT scan itself, which involved with the investigation of the effect
of CBCT on registration accuracy and image qualities with a reduced number of planar
projections used in volumetric imaging reconstruction, 2. the investigation of the impact
of small rotational errors on the magnitudes and distributions of spatial dose variations
for intracranial stereotactic radiotherapy (SRT) treatment setups, 3. the development of
a quick online evaluation method to assess the dose consequence due to small
rotational errors, 4. the invention of a novel method (centroid-to-centroid (CTC)) to
facilitate the lung SBRT setup for translational-only scenarios in order to achieve
optimal tumor dose coverage, 5. the assessment of CTC method’s performance on
breathing variation scenarios using a phantom study.
-
13
CHAPTER 1 INTRODUCTION
General Background
The treatment effects of stereotactic radiotherapy, such as intracranial SRT, lung
SBRT and etc., are heavily relied on the patient setup accuracy. The invention of IGRT
provides the possibility to facilitate such need. While all radiation therapy procedures
are image-guided per se. traditionally, imaging technology has primarily been used in
producing 3D scans of the patient’s anatomy to identify the location of the tumor prior to
treatment. The verification of a treatment plan is typically done at the level of beam
portals relative to the patient’s bony anatomy before patient treatment. In contemporary
radiation therapy field, the term of IGRT is refer to newly emerging radiation planning,
patient setup, and delivery procedures that integrate cutting-edge image-based tumor
definition methods, patient positioning devices, and/or radiation delivery guiding tools.
These techniques combine new imaging tools, which interface with the radiation
delivery system through hardware and/or software, and state-of-the-art 3DCRT or
IMRT, and allow physicians to optimize the accuracy and precision of the radiotherapy
by adjusting the radiation beam based on the true position of the tumor and critical
organs. With IGRT, it is also possible to take tumor motion into account during radiation
therapy planning and treatment setup. Because IGRT improves precision, it raises the
possibility of shortening the duration of radiation therapy by reducing the number of
treatment sessions for some forms of cancer, such as lung SBRT.
Regarding patient setup in the IGRT workflow, KVCBCT-guided setup is surely
the most popular approach implemented currently. It allows physician to utilize the
volumetric image information to verify and correct the planning patient setup in the linac
-
14
coordinates by comparing with the patient position defined in treatment plan. In this
work, we are interested in several questions regarding the strategies on how to use
CBCT system to efficient and accurately guide the patient setup for various treatment
sites as well as their corresponding dose consequence (beam delivery and CBCT scan
itself).
The Relationship between Cbct Scan Dose and Registration Accuracy
While the dose from a single CBCT image acquisition is small compared to the
therapy dose, repeated daily image-guidance procedures can not only lead to
substantial dose to normal tissue, but also generate a high integral dose for the
scanning area of the patient 1. Radiation therapy patients are already being exposed to
very high and localized radiation; therefore, the additional radiation from imaging has an
associated risk and should be kept low 2,3. How to balance imaging-quality improvement
and dose escalation risk over broad clinical scenarios is still a controversial topic in the
radiation therapy community and would be beyond the scope of this study. In this study,
however, we attempt to reveal the impact of image information and registration
accuracy resulting from dose-reduction strategies.
In general, dose can be minimized during CBCT scan through two different
means: one is by reducing the tube current; the other is by reducing the projections
used in volumetric imaging reconstruction 3. The first method would result in a reduced
signal-noise ratio due to less incident photons (reduced mAs) interacting with detectors,
which ultimately degrade the image quality from several perspectives (spatial resolution,
soft-tissue contrast, edge preservation, etc.). Those degrading effects have been
studied by several investigators, and solutions to alleviate them have been proposed4-6.
Alternatively, the second strategy to reduce the patient’s dose underlies the reduction of
-
15
sampling frequency during image acquisition. The major advantage of the second
method versus the first one is that it is a time saver. Faster scan speed and shorter
reconstruction time could ultimately limit setup uncertainty as a result of less patient on-
couch time. Unfortunately, like the mA-reduction method, the disadvantage of the
second strategy is that image quality diminishes. More specifically, the soft-tissue
contrast degrades and streak-shaped artifacts appear. In IGRT, patient alignment
information derived from imaging registration is the main purpose of onboard therapy
imaging, which leads to a major concern with IGRT imaging – registration accuracy
between reference CT and CBCT rather than general image-quality critiques. The
alignment information generated by volumetric imaging registration will eventually
provide clinicians with patient setup accuracy and, subsequently, position-adjustment
guidance, if necessary. What are the effects of imaging quality degradation under a
dose-decreasing strategy of reducing the projection number? And what is the optimal
projection-number needed for reconstruction to balance between the dose and
registration accuracy for different sites? This work provided answers to all those
questions
Dose Consequence Due to Small Rotational Errors
The characteristic that is common to all of the current IGRT systems is that they
are capable of providing target geometric correction information in terms of 6 degrees-
of-freedom movements prior to and/or during treatment, which is adequate for setup
corrections to rigid anatomy, such as the brain. While translational corrections can be
easily achieved by shifting the table, performing rotational corrections is not always an
easy task. On the one hand, manually adjusting the head position according to the
IGRT system guidance is not always an option for the therapists since the restriction of
-
16
the immobilization device and/or the patient’s comfort level can sometimes restrict the
freedom of the rotation and consequently leave a large discrepancy between the
expected and achieved correction. On the other hand, most of the regular linear
accelerator treatment couches can only be rotated in one direction (axis perpendicular
to the couch surface)7, which makes a complete geometric correction (pitch, yaw, and
roll) using current couch systems not possible7. Several approaches have been
introduced to compensate for uncorrected rotational errors7-9. These methods either
require a beam’s eye view (BEV) adjustment7,8 or need an additional rotational device9
to offset rotational errors, which is not always practical in all clinical scenarios. Also,
introducing new devices or altering treatment parameters can potentially complicate the
entire system and may ultimately aggravate treatment uncertainties and increase the
chance of setup errors.
Dose consequence due to setup errors for various sites has been studied
extensively10-19. Some investigators10-15 suggest that the dosimetric influence of small
rotational errors is minimal for most cases in head and neck treatment, prostate
treatment, and spine treatment, whereas larger rotational errors may dramatically affect
clinical target volume (CTV) coverage and/or organs-at-risk (OAR) sparing, such as in
cases of irregular target shape and/or OAR positions that are proximate to the target.
Our previous study16 demonstrated similar dosimetric consequences in DVH form for
SRT of intracranial tumors. All of these studies indicate that dosimetric consequences
due to uncorrected rotational errors vary on a case by case basis and depend on
rotational angles and orientations, tumor size and shape, treatment site, and other
related factors. In this work, we will not restrict our visions in the level of evaluating DVH
-
17
changes for PTV and OARs for intracranial stereotactic radiotherapy treatment setups
due to small rotational errors. Instead, we thoroughly investigated the impact of small
rotational errors on the magnitudes and distributions of dose map variations.
For cases in which rotational compensation is untenable, online systems that
allow for instantaneous evaluation of the changes between planned dose and
regenerated dose accounting for uncorrected rotational errors are desirable. Such
systems can help clinicians make judgments immediately prior to the treatment and can
ultimately improve clinical efficacy. However, online evaluation systems that regenerate
3D dose distributions using current stand-alone treatment planning systems (TPS) are
not practical for multiple reasons, such as the unavailability of rotation-simulation
modules for dose regeneration, inadequate dose computational speed, costly hardware
and software, etc. A practical on-line algorithm which can estimate the new DVHs due
to rotational errors was developed in our work.
Cbct-Guided Setup Strategy for Lung Sbrt
For lung SBRT setup, the alignment between reference CT and localization CT
are typically initiated with an automatic registration algorithm and finalized according to
the clinician’s judgment. Commercial software for automatic online registration of KV-
CBCT and planning CT (PCT) often provides two kinds of similarity measures (“bone”
mode and “gray-value” mode) with six degrees of freedom for application of various
clinical scenarios20. The bone mode, which uses a Chamfer matching algorithm to
match the edge information of two image data sets, is relatively insensitive to noise and
has a fast computation time21-24. It is especially useful when the matching objects are
the rigid bony structures, but it performs poorly with lower contrast subjects. The gray-
value mode, which handles the auto registration by matching voxel gray-scale intensity
-
18
values throughout the entire selected image volume of interest, is known as the “cross-
correlation” technique25. In addition to edge information, soft tissue information is also
included in the gray-value mode and can potentially average out the registration error
caused by anatomy variance between the two image sets. It performs well with
appropriate selection of registration area for non-rigid anatomy possessing lower
contrast objects (like soft tissues). In terms of calculation time, the gray-value mode
generally takes much longer than the bone mode.
Although automatic registrations are able to generate reasonable alignment
parameters, which can help clinicians perform accurate patient setups for most of
cases, several challenges and issues still exist: (1) Automatic registration methods can
potentially result in a mismatch of lung tumors if: (i) these tumors have a large variation
in size and/or location or (ii) the region of registration (ROR) is inappropriately selected.
Since variations in tumor shape and position relative to the rib cage occur during the
preparation period (after the CT scan and prior to the first treatment), they are often
unknown or difficult to quantify, and bony registration results can potentially lead into an
inaccurate or incorrect position26. The gray-value method is capable of alleviating such
problems to some extent with an appropriately defined ROR. However, finding a generic
rule of ROR selection is not an easy task due to the tremendous variety of tumor
shapes and locations. Our clinical experience indicates that gray-value registration with
too large a ROR selection possesses almost no advantage compared with bone mode
registration because the contribution of intensity weighting over tumor area to the
overall objective is negligibly small. With too small a ROR selection (only focusing on
tumor region), large rotation corrections can be induced due to the symmetric shape
-
19
and/or the deformation of the tumor. Such parameters of rotation correction are invalid
and cannot be implemented. (2) For cases in which the isocenter is located away from
tumor (e.g., multi-lesions with one isocenter), translational correction only (a
conventional linear accelerator couch has only translational shift and yaw rotation
capability) with six-degree automatic registration can potentially result in a significant
setup error since rotation offset can be amplified by the distance between the isocenter
and the corresponding lesion27. (3) Manual registrations lacking systematic guidelines
can lead to inconsistent or incorrect setups, while automatic registration results are not
reliable as the first-level alignment. The registration accuracy of a manual adjustment
can be influenced by the clinician’s knowledge of the anatomy as well as by image
quality and window/level settings, especially for those cases where large tumor
variations are presented between CBCT and reference CT. (4) The inconsistency of
image data sets (between PCT and CBCT) for registration, which has not drawn a great
deal of attention in the radiation oncology society, is another issue. Most clinics register
volumetric average CT images acquired by slow CBCT scanning with free breathing
(arbitrary phase) CT images acquired through fast scanning to obtain setup correction
parameters. Such methods can cause low-accuracy alignment due to difference in
tumor appearance between the two image sets.
In order to achieve the optimal dosimetric consequence for tumors with the
strategy of couch translational corrections of patient setup, Yue et al. 28 presented an
approach using a two-step optimization algorithm. Since their method needs physician
involvement for tumor delineation post to CBCT scanning and immediately prior to the
treatment for setup corrections, it is obviously impractical for real clinical situations. In
-
20
addition, the optimization process is underlain by the application of model-based dose
computation algorithms (inverse square and tissue maximum ratio corrections) to
compensate for anatomy variance between planning CT and CBCT, which is only
suitable for dose calculation in relative homogeneity anatomies (such as pelvic region
for prostate treatment) but inappropriate to be applied to the anatomy with highly
heterogeneous density (such as lung tumors and lung tissues in its vicinity).
To facilitate the problems mentioned above for lung SBRT setup, a practical
strategy for setup alignment of SBRT lung treatments was introduced. This strategy can
guarantee the best tumor dose coverage using DVH for evaluation for translational-only
setup correction scenarios. The uncertainty of this method for different clinical scenarios
was also evaluated using patient study and phantom study. The performance of this
method, under the scenarios that the variation of breathing pattern, tumor size and
tumor position occurred between CBCT and planning CT, was assessed using a
phantom study.
Study Topics
Stereotactic radiotherapy (SRT) technique has been widely utilized for various
treatments sites over the past two decades. 3D conformal radiation therapy, intensity-
modulated radiation therapy (IMRT) and arc therapy have been routinely applied for
various SRT treatments in order to escalate target dose while keeping normal tissue
dose under toxicity threshold 29,30. However, the full potential of these technologies in
radiation treatment can be achieved only if the patient can be positioned accurately and
reproducibly during every session of the entire course of treatment delivery 1.
The need for more precise patient positioning has increased the interest in
developing 3D imaging techniques that can verify the patient setup immediately before
-
21
and after treatment. The availability of large area flat-panel detectors has facilitated the
development of integrated cone-beam computed tomography (CBCT) systems on linear
accelerators2-5. Until now, Kilo-Voltage (kV) CBCT systems integrated into the gantries
of linear accelerators have been pervasively employed as an advanced image-guided
radiotherapy (IGRT) modality to acquire high-resolution volumetric images of patients
for treatment localization purposes. Using on-line kV CBCT software and hardware, a
patient’s position can be determined with a high degree of precision and, subsequently,
setup parameters can be adjusted to deliver the intended treatment.
However, in the current research regime, it lacks specific strategies for various
sites on how to achieve the optimal patient setup though the guidance of CBCT system
and pertinent assessment of their dose consequence. Most of the registration strategies
are based upon the methodology of intensity matching between reference and
localization images for 6 degree freedom shift, which is not always practical for all the
clinical scenarios and even possess intrinsic flaws for some specific site. Meanwhile the
relationship between the patient’s dose received from the treatment (beam delivery and
CBCT scan itself) and setup alignment strategies had not been sufficiently addressed.
This work provided specific strategies for CBCT-guided setup procedures to
achieve optimal dose effects for various sites as well as assess the dose consequence
of those strategies. We started with the assessment of the patient’s dose generated by
CBCT scan itself. It involved with the investigation of the effect of CBCT on registration
accuracy and image qualities with a reduced number of planar projections used in
volumetric imaging reconstruction as well as ultimately seeking for the optimal
projection-number needed for reconstruction to balance between the dose and
-
22
registration accuracy. Then we addressed the impact of small rotational errors on the
magnitudes and distributions of spatial dose variations for intracranial stereotactic
radiotherapy (SRT) treatment setups. Furthermore, a quick online evaluation method to
assess the dose consequence due to small rotational error was introduced. Eventually,
in lieu of intensity matching based approaches, we developed a novel method to
facilitate the lung SBRT setup for translational-only scenarios in order to achieve
optimal tumor dose coverage. The thorough evaluation and efficiency improvement of
this method were also addressed using a phantom study.
Topic 1: Evaluate the effect of CBCT on registration accuracy and image quality
with a reduced number of planar projects.
The image qualities of 3D volumetric images reconstructed with a decreasing
number of projections was quantitatively evaluated using a CT phantom. The effects of
imaging-registration accuracy under a dose-decreasing strategy of reducing the project
number of different clinical sites were examined. Optimal frame reduction scheme which
can balance between the dose and registration accuracy for various sites was provided
for various sites.
Topic 2: Investigate the impact of small rotational errors on the magnitudes and
distributions of dose variations for intracranial SRT treatment setup.
On top of DVH analysis of dose consequence resulting from systematic rotational
setup errors, we further investigated the dose impact on the magnitudes and distribution
of the dose map variations for intracranial SRT treatment setups. The comparison
between planned dose and regenerated dose through rotation-simulated imaging was
-
23
presented. It included the mean and standard deviations of differential dose maps in
CTVs, OARs and different layers of rotated brain anatomy.
Topic 3: Develop an DVH-based online evaluation algorithm for dose
consequences caused by small rotational setup errors in intracranial stereotactic
radiation therapy.
The feasibility of using the original dose map overlaid with rotated contours
(ODMORC) method as a fast, online evaluation tool to estimate dose changes to CTVs
and OARs caused by small rotational setup errors will be assessed. A Fast algorithm for
multi-contour rotation, which can expedite the ODMORC method speed drastically, was
developed
Topic 4: Develop a CBCT-guided patient setup strategy based on deformable
registration technique for lung SBRT in order to achieve the optimal tumor dose.
A new registration strategy for translation-only correction scenarios of lung SBRT
setups was developed. Its effect will be evaluated by comparing ITV’s dosimtric
coverage of this method with bony and grey value methods, especially for those cases
as the tumor position varies greatly from its position on the original CT relative to the rib
cage.
Topic 5: Compare the uncertainties of the method from AIM 4 with other new
methods for the scenarios of breathing pattern, tumor position and tumor size change
using phantom study.
Various scenarios of breathing pattern change between planning CT and CBCT
scans, which involved exhalation/inhalation ratios change, motion trajectory length
change, were simulated by a 4D lung phantom. The scenarios with tumor position
-
24
change and tumor size change were also included. The corresponding scans for each
scenario were performed with 4DCT, CBCT and 4DCBCT. The comparisons on
registration accuracy of ITV centroids as well as dosimetric consequence of ITVs
among method suggested by Topic 4 and Translational-only Grey value registration
were performed.
-
25
CHAPTER 2 DECREASING THE KV-CBCT DOSE BY REDUCING THE PROJECTION NUMBER
Abstract
The objective of this study was to evaluate the effect of Kilo-Voltage cone-beam
computed tomography (CBCT) on registration accuracy and image qualities with a
reduced number of planar projections used in volumetric imaging reconstruction. The
ultimate goal is to evaluate the possibility of reducing the patient dose while maintaining
registration accuracy under different projection-number schemes for various clinical
sites. An Elekta Synergy Linear accelerator with an on-board CBCT system was used in
this study. The quality of the Elekta XVI cone-beam 3Dimensional volumetric images
reconstructed with a decreasing number of projections was quantitatively evaluated by a
Catphan phantom. Subsequently, we tested the registration accuracy of imaging data
sets on 3 rigid anthropomorphic phantoms and 3 real patient sites under the reduced
projection-number (as low as 1/6th) reconstruction of CBCT data with different rectilinear
shifts and rotations. CBCT scan results of the Catphan phantom indicated the CBCT
images got noisier when the number of projections was reduced, but their spatial
resolution and uniformity were hardly affected. The maximum registration errors under
the small amount transformation of the reference CT images were found to be within 0.7
mm translation and 0.3º rotation. However, when the projection number was lower than
one-fourth of the full set with a large amount of transformation of reference CT images,
the registration could easily be trapped into local minima solutions for a non-rigid
anatomy. We concluded, by using projection-number-reduction strategy under
conscientious care, imaging-guided localization procedure could achieve a lower patient
dose without losing the registration accuracy for various clinical sites and situations. A
-
26
faster scanning time is the main advantages compared to mAs decrease-based dose-
reduction method, which ultimately reduce the patient-on-couch time to alleviate the
setup uncertainty.
Background
Advances in 3Dimensional (3D) radiation therapy technologies have led to the
safe delivery of escalated doses to tumors, improving local control and sparing dose to
healthy tissues for improving quality of life31. However, the full potential of these
technologies in radiation treatment can be achieved only if the patient can be positioned
accurately and reproducibly during every session of the entire course of treatment
delivery1. Mega-voltage (MV) portal imaging has been widely implemented for the past
2 decades as patient treatment-positioning tools. Due to the inherent low-contrast and
2-dimensional nature of the projection images, the precision of MV portal imaging is
limited so far as accurately defining the patient’s position32. The need for more precise
patient positioning has increased the interest in developing 3D imaging techniques that
can verify the patient setup immediately before and after treatment. The availability of
large-area flat-panel detectors has facilitated the development of integrated cone-beam
computed tomography (CBCT) systems on linear accelerators33-36. Up until now,
kilovoltage (kV) CBCT systems integrated into the gantries of linear accelerators have
been widely used as an advanced image-guided radiotherapy (IGRT) modality to
acquire high-resolution volumetric images of patients for treatment localization
purposes. Using on-line kV CBCT software and hardware, a patient’s position can be
determined accurately with a high degree of precision, and, subsequently, set-up
parameters can be adjusted to deliver the intended treatment.
-
27
While the dose from a single CBCT image acquisition is small compared to the
therapy dose, repeated daily image-guidance procedures can not only lead to
substantial dose to normal tissue, but also generate a high integral dose for the
scanning area of the patient.1 Radiation therapy patients are already being exposed to
very high and localized radiation; therefore, the additional radiation from imaging has an
associated risk and should be kept low.2 Concerns about the stochastic risk of inducing
cancer or genetic defects have already been accounted for in the limits on leakage from
the primary therapy beam, which should not exceed 0.2% of the absorbed dose rate on
the central axis at the treatment distance.3 This recommendation suggests that for a
total treatment dose of 60 Gy delivered over 30 fractions, only 120 mGy or less should
be contributed by MV radiation leakage. Although the energy and field of exposure differ
for MV beam leakage with CBCT, making direct comparison problematic, the imaging
doses are clearly not negligible. The health and safety considerations that underlie the
limit on therapy-beam background dose should, therefore, be considered as relevant for
imaging exposures as well.3 How to balance imaging-quality improvement and dose-
escalation risk over broad clinical scenarios is still a controversial topic in the radiation
therapy community and would be beyond the scope of this article. In this study, we
wanted to focus on the impact of image information resulting from dose-reduction
strategies.
In general, dose can be minimized during CBCT scan through 2 different means:
one is by reducing the tube current; the other is by reducing the projections used in
volumetric imaging reconstruction.3 The first method would result in a reduced signal-
noise ratio due to less incident photons (reduced mAs) interacting with detectors, which
-
28
ultimately degrade the image quality from several perspectives: spatial resolution, soft-
tissue contrast, edge preservation, etc. Those degrading effects have been studied by
several investigators, and solutions to alleviate them have been proposed.4-6
Alternatively, the second strategy to reduce the patient’s dose underlies the reduction of
sampling frequency during image acquisition. The major advantage of the second
method versus the first one is that it is a time saver: faster scan speed and shorter
reconstruction time could ultimately limit setup uncertainty as a result of less patient-on-
couch time. Unfortunately, like the mA-reduction method, the disadvantage of the
second strategy is that image quality diminishes. More specifically, the soft-tissue
contrast degrades and the streak-shaped artifacts appear. In diagnostic imaging, there
is a direct relationship between the exposure level and image quality. The demand for
high-contrast, low-noise images pushes the exposure levels up. In IGRT, patient-
alignment information derived from imaging registration is the main purpose of on-board
therapy imaging, which leads to a major concern with IGRT imaging: registration
accuracy between reference CT and CBCT rather than general image-quality critiques.
The alignment information generated by volumetric imaging registration will eventually
provide clinicians with patient-setup accuracy and subsequently position-adjustment
guidance if necessary. The goal of this article is to examine the effects of imaging-
registration accuracy under a dose-decreasing strategy of reducing the projection
number for different clinical sites.
A preliminary study on the strategy to decrease the radiotherapy dose by
reducing the projection number was conducted by Sykes et al6 using a head-and-neck
phantom. Their study showed that for a specific phantom, using a reduced number of
-
29
projections (one-fifth of a full set) registration accuracy could be preserved as the on-
board imaging was reconstructed. However, several aspects of their study could
potentially be revisited or further engaged with. First, a quantitative analysis of the
degrading effects of imaging quality due to the reduced projection number was not
included in their study. Since the clinician’s judgment of image registration heavily
depends on imaging quality, evaluating the registration accuracy with any proposed
strategy is as important as assessing the degrading effects of the images, especially
when automatic registration cannot provide the correct alignment information. Secondly,
although head-and-neck phantom data demonstrate the effects of registration accuracy
for a rigid body, real patient imaging can be deformed with respect to the reference CT
due to setup variance, internal organ motion, and/or biological changes in the patient. It
is therefore worthwhile to exam whether a proposed strategy can still be applied to non-
rigid patient imaging. Furthermore, in Sykes et al’s study, only the cross-correlation
information of the pixel gray value between image sets was used as an objective
function for registration calculation due to the availability of commercial software at that
time. Today, integrated software is available for most CBCT sites, with both edge and
cross-correlation information included in the commercial software as customer options.
The effects of the proposed strategy should be assessed for all of today’s optional
modes. Another limitation of Sykes et al’s study is that they failed to provide the results
of full sets of projection-number-reduction schemes except for a specific scheme (one-
fifth of the projection number). We believe a more valuable study should demonstrate
the registration effects with different schemes of projection-number reduction to at least
provide the researcher with opportunities to seek optimum schemes that balance image
-
30
quality with patient dose. In this paper, we provide a thorough study for a dose-
decreasing method based on projection-number reduction.
The aim of this study is to evaluate registration accuracy and image quality with a
reduced number of planar projections for volumetric imaging reconstruction with an
Elekta on-board CBCT. The image qualities of 3D volumetric images reconstructed with
a decreasing number of projections were quantitatively evaluated using a CT phantom.
Subsequently, 3 different sites of phantoms and patients’ imaging data with different
projection-number schemes were analyzed, respectively.
Methods and Materials
X-ray Volume Imaging System
The X-ray volume imaging (XVI) system (version 3.5) used in this study is an on-
board kV CBCT imaging system integrated into the Elekta Synergy platform linear
accelerator (Elekta Corp., Stockholm). The XVI system consists of a kV radiation-
generation system that produces a cone-shaped X-ray beam, an image receptor
consisting of a kV imaging panel that acquires projected planar images from the X-ray
beam, and a control system with computer workstation hardware and XVI software. The
radiation generator and detector panel are both mounted on retractable arms extending
from the accelerator’s drum structure in an orthogonal direction to the MV system and
they share the same rotation center with a gantry-head MV imaging-panel rotation
system. The X-ray generator’s focus to the iso-center distance is 1,000 mm and the
distance from the focus to the receptor is 1,536 mm.
The kV generator is a 40-kilowatt unit capable of producing single and
continuous radiographic exposures of varying energy and dose. The energy range is
from 40 to 150 kV. The X-ray-tube filtration is 3.5 mm of aluminum and 0.1 mm of
-
31
copper. This equates to a total of approximately 6 mm of aluminum equivalence at 100
kV and 7 mm of aluminum equivalence at 120 kV. Tube-current options are 25, 32, and
40 mA while the pulse length can be set within the range of 4 to 80 ms.
The images for this study were acquired through 360° (full fan) or 200° (half fan)
rotations. The pulse rate for full fan and half fan were 2.7 Hz and 5.4 Hz respectively.
The fast gantry speed was 60 seconds. The regular scanning speed was about 120
seconds. Pixel size was 0.8 mm2. Each projection image was sampled with 512 x 512
pixels. The maximum image volume was defined by the size of the collimator, and the
maximum correspondent field of view was 26 cm.
XVI software was running on a clinical workstation with 2.4 GHz CPU and 2 GB
of memory. The reconstruction algorithm provided by the XVI is Feldkamp’s back-
projection algorithm37 considering flex-map correction in rotation orbits.38
Phantom and Patient Selection
Catphan phantom
Image quality drastically impacts the clinician’s judgment of registration accuracy
when auto-registration cannot provide correct registration results, especially for non-
rigid body registration.3 Typically, quantitative image-quality tests are performed by
analyzing a specially designed CT phantom scan. For instance, Catphan phantom
(model CTP 503, The Phantom Laboratory, Salem) has been recommended by the
Elekta XVI system as the acceptance test phantom. Reconstructed Catphan phantom
images with varying projection-number schemes were accessed in our study through 3
major criteria specified in the acceptance test guide. The Catphan phantom is cylindrical
and has multiple layers embedded with different shapes and materials of inserts. It
-
32
could be used to examine CT number uniformity, spatial resolution, low-contrast
resolution, and geometric accuracy, among other factors.
Rando phantom
Three body parts of a male RANDO phantom (The Phantom Laboratory, Salem,
NY) were selected as rigid imaging objects for our registration-accuracy study. We refer
them as the head-and-neck area, thoracic area, and pelvic area of the RANDO phantom
as R-H&N, R-Thoracic, and R-Pelvis, respectively. The phantom was constructed with a
natural human skeleton cast inside of soft-tissue-simulating material. In the thoracic
area, lungs are molded to fit the contours of a natural rib cage. The air space of the
head, neck, and stem bronchi are duplicated. The phantom is sliced at 2.5-cm intervals.
The whole-grid patterns can be drilled into the sliced sections to enable the insertion of
dosimeters. Two tissue-simulating materials are used to construct the RANDO
Phantom: the RANDO soft-tissue material and the RADNO lung material. Both of these
are designed to have the same absorption as human tissue at the normal radiotherapy
and radiology exposure levels. The RANDO’s similarities to a real human structure
make it an attractive and widely used imaging and dosimetry substitute for simulation
among the radiology community.
Patient sites
Three patients’ image sets at different sites were used for our registration study.
They were a brain-cancer site, a lung-cancer site, and a prostate-cancer site, which are
referred to as P-Brain, P-Lung and P-Prostate, respectively. The reference CT images
were taken a week before treatments for planning purpose whereas CBCT images were
taken right prior to patient treatment to register with reference CT images for localization
purpose. Unlike rigid phantom imaging, the real patients imaging data could be
-
33
deformed between reference CT and CBCT image sets. For example, patient’s prostate
could be stretched or squeezed by its surrounding tissues and organs (Different amount
of urine inside bladder and/or different amount of gas inside rectum between the
moments of acquiring reference CT and CBCT can cause different shapes of the
prostate in these two image sets); chest cage could be enlarged or shrank depending
on at which phase of breathing while acquiring images. Since the registration solution of
non-rigid imaging is much easier to trap into a local minima solution, assessing the
robustness of the registration algorithm with non-rigid patient data would make the study
more valuable and thorough.
Acquisition and Reconstruction of Cbct Images
The positioning procedure used with the Catphan phantom followed the
acceptance test guide of the XVI. The RANDO phantom and patients were placed
centrally in the field of imaging view prior to the scan. All CBCT images of phantoms
and patients were acquired by the standard technique recommended by XVI protocol
systems. The technique parameters, including beam energy, tube current, duration per
pulse, projected image number, scan time, gantry rotation angle, and collimator type,
are listed in Table 2-1. Since the recommended technique could generate near-optimum
image quality for the current system, those 3D images reconstructed with full sets of
projected planar images were treated as standard images for the registration study and
referred to as “XVI-full.” The near-optimum image qualities for each site were achieved
by choosing high tube currents and high energy beams to minimize noises and any
reconstruction artifacts due to high-density materials inside the phantoms or patients. A
deliberate, slow gantry speed also benefitted the image quality since it allowed enough
projected images to be acquired during the slow gantry rotation in order to reconstruct
-
34
images with minimal artifacts and structural noise generated from the high-contrast
edge. The term “optimum image quality” in our study refers to the best image quality
achievable by the current system with minimal noise and artifacts, and thus minimal
impact on the accuracy of 3D- image- registration.
All 3D images were reconstructed using 1-mm3 voxels. The registration outcome
of the XVI-full scan with the corresponding “reference planning CT scan” dataset was
applied as the baseline to evaluate the registration accuracy of images reconstructed
with fewer projections. In order to simulate a series of lower doses and faster scans, the
reconstruction schemes with fewer projections consisted of half, one-third, one-fourth,
one-fifth, and one-sixth of the full projections, referred to as XVI-1/2, XVI-1/3, XVI-1/4,
XVI-1/5, and XVI-1/6, respectively. The image-quality degrading effects along with the
fewer numbers of projections were analyzed. As the projection number became less
than one-sixth of the full set, unacceptable image quality became an obstacle to most
clinicians who could not provide sound judgments about registration results due to the
low visibility in the soft-tissue contrast. Therefore, even reconstructed images with
projections less than one-sixth of full set would potentially sustain the edge information,
registration test with a projection number less than one-sixth of full set had been
excluded from our study.
3D imaging reconstructions were performed with XVI software. The schemes
with fewer projection numbers were achieved by manually deactivating the undesirable
projection frames in the XVI database. For instance, XVI-1/n was reconstructed only
with frames whose index number could be divisible by the number n, and all other
frames would not be used for reconstruction at all. Projection-number schemes used for
-
35
patients and phantoms are listed in Table 2-2. It should be noted that this was a
feasibility study and XVI-1/n does not mean that the scan speed was increased by n
times, but rather that it was limited by the gantry-speed limitations permitted by the
International Electrotechnical Commission (IEC) and achievable tube firing frequency.
The validity of simulating a low-dose fast scan will be discussed.
Imaging Quality Check with a Reduced Projection Number
The influence of a reduced number of projections on image quality was evaluated
using the Catphan phantom. The evaluated items included 3D spatial resolution, 3D
low-contrast visibility, and 3D uniformity. The “CAT-Image Quality” preset parameters
(referred to in Table I) were used to acquire the volumetric images. A 360° rotation scan
was performed and a total of 668 projections were registered in the XVI database. The
reconstructed 3D images with all projection number schemes were tested separately. A
detailed description and criteria of those tests follow.
3D spatial resolution
We located the spatial-resolution module slice of the reconstructed CBCT image,
then magnified the image until the module filled the transverse view, and subsequently
determined the highest numbered line pairs visible by adjusting the brightness and
contrast to achieve the best view for the line pairs. The specification was greater or
equal to 7 line pairs per centimeter. Figure 2-1(a) shows the reconstructed images with
XVI-full, XVI-1/2, and XVI-1/6 schemes at the spatial resolution module slice,
respectively.
3D low-contrast visibility
Figure 2-1(b) shows the contrast resolution module slice reconstructed with XVI-
full, XVI-1/2, and XVI-1/6 schemes for the 3D low-contrast visibility test. This test
-
36
measures the mean pixel values of the polystyrene insert (7 o’clock, Figure 2-1(b)) and
the LDPE insert (9 o’clock, Figure 2-1(b)), as well as the standard deviation of their pixel
values. The low-contrast visibility value was defined mathematically as Equation (2-1),
where “SD” represented the standard deviation of the pixel values; “Mean” represented
the mean of the pixel value; the subscripts (“polystyrene” and “LDPE”) represented two
different materials (polystyrene and low density polystyrene) respectively. If the
calculated value of Equation (2-1) was large, it meant the difference of imaging noises
between these two materials overwhelmed the difference of true imaging signals
between them, which indicated poor low contrast visibility; vice-versa. The XVI
specification for the low-contrast visibility value was lower than 2%.
]2
[
][
5.5
LDPEepolystyren
LDPEepolystyren
SDSD
MeanMean
visiblitycontrastlow (2-1)
3D uniformity
The uniformity module slice of the Catphan phantom was used as the image for
the 3D uniformity test. Four random locations in the uniformity module were chosen to
measure the mean values of the pixels covered by a 1 x 1 cm2 probe window. Equation
(2-2) was used to compute the maximum percentage difference between any 2 of the
mean pixel values, where max(mean) and min(mean) are the largest and smallest mean
pixel values among the 4 locations, respectively. The specification was less than or
equal to 2%.
)(
)()(
meanMax
meanMinmeanMaxuniformity
(2-2)
-
37
Registration Analysis
The reference CT images of the phantoms and patients were acquired by a
Philips multi-slice high-speed CT scanner (Amsterdam, The Netherlands). The
placement of the phantoms allowed the scan origin to be near the iso-center used in
XVI imaging, but no effort was taken to ensure that the center was identical or that the
phantom was rotationally aligned in the CT scanner with respect to the XVI scans. Real
patient setup, however, strictly followed department setup procedures to avoid any
possible rotation variance between the reference CT scan and XVI scan for practical
clinical purposes.
Automatic 3D registration of the reference CT and the XVI scans was performed
using XVI software. Two automatic registration modes were used for our study. The
“bone mode” of the automatic registration uses a chamfer matching algorithm originally
described by Borgefer.21 The chamfer algorithm performs registration by matching the
edge information of 2 image sets. It is relatively insensitive to image noise and mild
streak artifacts, and also has a quick computing time. It performed poorly with lower-
contrast subjects only. Comparisons of the quality of registration with the bone mode
with other techniques have been provided by several investigators.22,23 The “grey value
mode” performs the auto-registration by matching voxel grey-scale intensity values
throughout the entire interested image volume, which is well known as the “cross
correlation” technique.25 In addition to the edge information, soft-tissue information has
also been included in the grey-value algorithm and would potentially average out the
registration error caused by anatomy variance between the 2 image sets. In terms of
calculation time, the grey-value mode is much worse than the bone mode in general. In
-
38
our study, we had tried the bone mode for all patient and phantom images, and the
grey-value mode only for patient images of pelvis and lung sites.
The performance of the automatic registration was checked by repeatedly
registering the reference CT planning scan with the XVI scan for each projection
number scheme. We initially registered reference CT imaging with XVI-full imaging and
considered the registration results as “Gold Standards;” the recorded translation and
rotation values were then treated as baseline and were subtracted from the values of
the registration results of other schemes to acquire the registration error values for all
schemes accordingly. To simulate setup error in terms of translation and rotation, we
manually entered the translation value in the x, y, and z axes and rotation angle in the x,
y, and z directions to achieve the desirable transformed reference-CT datasets. Three
independent translations with values of 2 mm, 5 mm and 20 mm along each axis of x, y,
and z, and 2 independent rotations with values of 3° and 10° around each axis of x, y,
and z, were applied to the reference CT scan. The transformed reference CT images
were then registered with XVI images reconstructed with different numbers of
projections to test the registration accuracy. Additionally, 2 combinations of translation
and rotation were applied to the reference-CT planning scan for all 3 axes to access the
registration capability with various scheme reconstructions for more realistic reference-
CT transforms as well. To demonstrate that registration was reproducible, each
registration procedure was performed 10 times repeatedly, and the mean and standard
deviation values in terms of translation vector error and max rotation angle error of all
directions were recorded.
-
39
Results
Reconstructed Images
The XVI reconstruction time of patients and phantoms with different projection-
number schemes are listed in Table 2-3. Obviously, fewer projection numbers resulted
in shorter computing times. XVI-1/6 reconstruction procedures lasted an average of 4.8
seconds over all patients and phantoms, which is about 1/6th of the average time spent
for all XVI-full reconstructions. The representative trans-axial, sagittal, and coronal
slices of the XVI-full, XVI-1/2, and XVI-1/6 schemes for 3 RANDO phantom parts and 3
different patient sites are shown in Figures 2-2 and 2-3, respectively. As predicted,
detailed contrast of soft tissue in XVI imaging could not be visualized as well as with
reference-CT imaging, even with full-projection reconstruction. XVI-full showed some
low-frequency noise artifacts that created non-uniformity on the grey level in the tissue-
equivalent material and some streak-shaped artifacts originating from high-contrast
edges (like the couch top). It was noticed that XVI-1/2 images were slightly noisier than
XVI-full; however, streak artifacts were not more severe than XVI-full, while edge
information between the anatomy interfaces was well kept. XVI-1/6 images presented
significant streak artifacts with the appearance of an interference pattern due to the
under-sampling effect of the reconstruction. Reduced visibility in the overall soft-tissue
contrast was also observed in the XVI-1/6 images. However, even in the XVI-1/6
images, the gross bony anatomy and air cavity were geometrically undistorted and a
surprising level of fine edge details still remained whereas the presence of degrading
effects of soft-tissue contrast escalated.
Imaging-quality Check
The image quality results were provided quantitatively as follows.
-
40
3D spatial resolution check
Figure 2-1(a) presents the image slice of the spatial-resolution module of the
Catphan phantom with XVI-full, XVI-1/2, and XVI-1/6 reconstruction schemes
separately. The highest number of visible line pairs of XVI-1/6 is 7. Increases in
projection number led to better visibility of line pairs, especially for XVI-full and XVI-1/2
for which line-pair 8 could easily be visualized. In summary, all tests satisfied the criteria
(visibility of line-pair 7) provided by the XVI acceptance manual.
3D low-contrast visibility check
Figure 2-1(b) presents the 3D low-contrast visibility module slice of the Catphan
phantom with XVI-full, XVI-1/2, and XVI-1/6 reconstruction schemes separately. Clearly,
the imaging interference from noises became more severe as the number of projections
used for 3D reconstruction decreased. Figures 2-4(a) and 2-4(b) show the mean CT
number and standard deviation of pixels inside the 2 inserts regions (described in the
Methods and Materials) versus different reconstruction schemes, respectively. The
results show that mean pixel-value curves were almost flat, whereas the standard
deviation increased as the number of projections decreased, indicating that increased
noise due to under-sampling does significantly influence detailed contrast visibility.
Figure 2-4(c) summarizes the calculated values of low-contrast visibility by Equation (2-
1) versus the different reconstruction schemes. The XVI specification for the low-
contrast visibility value was less than 2%. Only XVI-full and XVI-1/2 met the
specification; others failed. More specific for clinic scenario, the degradation due to
under-sampling effect would be the appearance of soft tissue contrast lost in CBCT
images. For instance, muscle tissues could be clearly visualized in XVI-full image set of
the prostate patient (referring figure 2-3(c)) whereas the appearance of muscle tissues
-
41
in XVI-1/6 images had been degraded (merged by surrounding tissues due to high
noises).
3D uniformity check
All tests met the criteria. The results had been presented in Figure 2-5. No signs
indicate that uniformity worsened with fewer projection numbers.
Registration
The mean registration errors of 10 repeated measurements in terms of
translation vector error and max rotation-angle error versus different reconstruction
schemes are presented as scatter charts in Figures 2-6 for Phantoms with “bone mode”
registration algorithm (Fig. 2-6(a) head-and-neck phantom, Fig. 2-6(b) thoracic
phantom, Fig. 2-6(c) pelvis phantom), Figure 2-7 for all patient sites with “bone mode”
registration algorithm (Fig. 2-7(a) brain site, Fig. 2-7(b) lung site, Fig. 2-7(c) prostate
site) and Figure 2-8 for non-rigid patient sites with “grey mode” registration algorithm
(Fig. 2-8(a) lung site and Fig. 2-8(b) prostate site). Each chart located at left side of
each subfigure showed the result data of translation vector error; the one at right side
showed the result data of max rotation-angle error. Since each subfigure had same
structures, we only described them in general but not repeated them individually. In the
left-side chart of each subfigure, X axis represented the different schemes of XVI
reconstruction; the Y axis represented the mean registration error of 10 repeated
measurements for the translation vector only; 7 series labels with different shapes and
colors represent 7 different translation and rotation combinations applied to the
reference-CT images. To facilitate the visibility of overlapped error points across
different series, the color lines between adjacent points of same series had been added
to help reader distinguish those error points among different series. These combination
-
42
series consisted of 2 mm shifts along the x, y, and z axes called “T = 2 mm,” 5 mm
shifts in the x, y, and z axes called “T = 5 mm,” 20 mm shifts in the x, y, and z axes
called “T = 20 mm,” 3° rotations in x, y, and z directions called “R = 3°,” 10° rotations in x,
y, and z directions called “R = 10°” 2 mm shifts along the x, y, and z axes plus 3°
rotations in the x, y, and z directions called “T = 2 mm, R = 3°,” and 20 mm shifts along
the x, y, and z axes plus 10° rotations in the x, y, and z directions called “T = 20 mm, R
= 10°”. In each right-side chart of each subfigure, Y axis represented means of the max
angle error over all directions of 10 repeated measurements; other labels had same
meaning as the right-side chart and would not be repeated here. Standard deviations of
the registration error over 10 repeated measurements were 0.1 mm or less in translation
and 0.10 or less in rotation across the board. The superimposed standard deviation bars
were excluded from the charts to avoid the visibility interference of mean values. The
mean registration time by bone mode was 4.2 seconds and the standard deviation was
1.3 seconds; the mean registration time by grey-value mode was 15.6 seconds and the
standard deviation was 3.2 seconds
Discussion
The results indicate that, for image sets of rigid body (e.g. brain phantom, pelvis
phantom and brain site), when treatment volume can be determined by bone-tissue
and/or tissue-air interface, without clinician interference, auto-registration results using
reduced-projection-number (up to one sixth of projections) reconstruction can still
provide accurate enough alignment information with no obvious trends of error
escalation while projection number decreased. Error points were randomly scattered in
charts(shown in Fig.2-6(a), Fig 2-6(c) and Fig2-7(a)) with maximum errors of 0.7 mm in
translation and 0.3° in rotation except of the thoracic phantom for which a large
-
43
misalignment (10° rotations and/or 20 mm translation shown in Fig.2-6(b)) was applied
in the first place. For non-rigid anatomy sites like the lung and prostate, however, auto-
registration with too few (one fifth or one sixth) projection-number reconstructions could
lead to greater alignment errors (shown in Fig.2-7(b), Fig.2-7 (c), Fig. 2-8(a) and Fig. 2-
8(b)) unless manual pre-alignment was applied. Meanwhile, the overall reproducibility of
registration results was shown to be excellent (less than 0.1 mm; 0.1° standard
deviation), whereas other geometric uncertainties that exist in the radiotherapy
treatment process are much more significant.
Three major image-quality parameters with different scheme reconstructions are
shown in Figures 2-2, 2-3 and 2-4. As demonstrated, high-contrast details and spatial
resolution were well preserved even with the lowest projection-number reconstruction
(XVI-1/6) tested, whereas low-dose fast scans may not be sufficient to visualize low-
contrast soft tissue with an ultra-low projection-number reconstruction due to increased
noise and streak artifacts resulting from under-sampling effects. It could be concluded
that for rigid-bony anatomy sites, like the brain, an ultra-low projection-number-reduction
strategy may still be a trouble-free solution to tremendously reducing the patient dose,
even if soft-tissue-contrast visibility was completely lost, whereas for non-rigid soft-
tissue anatomy sites, like the prostate and lung, reconstruction with too few projections
could trap the registration results into local minima, making the clinician’s interference
for registration unavoidable. Thus, it is much more sensible to optimize soft-tissue
visibility for rigid sites versus non-rigid sites. It is conceivable that imaging-processing
techniques that could reduce noise and enhance soft-tissue contrast would be preferred
when soft-tissue contrast becomes prominent for imaging registration. It was also
-
44
noticed that Figure 2-2(c) and 2-3(c) showed some ring artifacts caused by imperfect
CBCT detector calibration in current version of XVI software. They are common artifacts
for all the ELEKTA machines. In general, we assumed subtle ring artifacts should not
influence registration accuracy since registration results were dominated by bony
structure. However, to validate this assumption, a sophisticated comparison between
registration outcomes with ring artifact removal and without removal is needed. Due to
limited accessibility to the imaging data, such analysis is not feasible for this study.
As mentioned, a local minima trapping situation occurred with the ultra-low
projection reconstruction (XVI-1/4) for non-rigid sites (lung and prostate) due to the
combined effects of increased noises, non-rigid anatomy, and large misalignments. For
these 2 sites, as the 10° rotation and ultra-low reconstruction scheme were applied
simultaneously, large translation and rotation errors led to the registration results failing
regular clinical registration criteria (2 mm in translation and 3° in rotation), which
indicates that reconstruction with too few projection numbers (XVI-1/4 or worse) will be
the “scare factor” for some non-rigid anatomy sites, and initial human interference is
imperative to overcome the local minima trapping effect. Another local minima trapping
situation occurred while we were performing rigid thoracic phantom registrations with a
setup simulation that had a large initial misalignment. As “R = 10°” or “T = 20 mm, R =
10°” were applied to the reference CT, the registration results were mismatched by 1
complete vertebra superiorly or inferiorly due to the similarity of adjacent vertebra and
ribs in the thoracic region. Like the non-rigid cases, this mismatching could be
conquered by initial manual alignment followed by the auto-registration procedure. For a
real patient’s lung, auto-registration after applying a large misalignment performed fairly
-
45
well simply because a real patient has more irregular internal structures than a phantom
structure (compare Figures 2-2(b) and 2-3(b)); therefore, the registration results are less
likely to fall into the local minima and there is more time to seek a global minima
solution. This mismatching occurred for all schemes from XVI-full to XVI-1/6 so that the
projection number was irrelevant.
In our study, bone mode and grey-value mode did not present any significant
differences for all schemes of reconstruction in terms of registration accuracy. However,
the computing speed was the key point distinguishing these 2 methods. On average,
the bone mode ran almost 4-times faster than the grey-value mode. Thus, the bone
mode would be recommended to save on procedure time for applicable cases.
The reconstruction time for XVI-1/6 is 5.2 seconds on average using the XVI
software. Considering that the average XVI-full reconstruction time was 30 seconds, we
could save roughly 25 seconds for reconstruction, which is helpful but not significant.
However, considering clinical settings, image reconstruction is typically concurrent with
image acquisition and the reconstruction time is masked, therefore reconstruction time
saving is not an important factor. The real time-saving factor could be generated by the
fact of fewer projections needed during scanning as mentioned previously. However,
even though XVI-1/6 reconstruction only needed 1/6th of the full-set projection number,
it does not appear that that the potential scan time for the XVI-1/6th scheme could be
escalated by 6 times. For example, a XVI-full scan of 652 projections needs an average
of 2 minutes for the pelvis, whereas a 110-projection scan could not be accomplished in
20 seconds due to the current limitations on the maximum gantry speed, the highest
tube firing frequency, and patient-safety concerns. Since the maximum gantry speed is
-
46
limited to no greater than 60/s, in order to comply with IEC regulations,39 less than a 1-
minute scanning time is impossible according to current safety regulations. Therefore
the scanning time could be reduced from current 2 minutes to 1 minute. Since the
system is designed for automatic registration integrated with software for XVI acquisition
and reconstruction, then it should be possible to perform the complete image-guided
process of image acquisition, reconstruction, registration, and correction of patient
positioning within 3 minutes instead of the usual 5 minutes. This would make on-line
image guidance a practical and realistic procedure
Previous discussions indicate that an XVI-full scan is not necessary when auto-
registration could provide accurate alignment information. For all the cases in our study,
it appears that XVI-1/4 would allow a good balance between dose reduction and
accurate-enough registration. Song’s study40 elucidates that for the average person, the
brain, lung, and prostate doses to the patient’s surface (i.e., the highest dose to the
patient) per XVI-full scan are 0.22 cGy, 4.6 cGy, and 2.4 cGy , respectively. Intuitively, a
XVI-1/4 scan would give one-fourth of the dose of an XVI-full scan to the patient’s
surface for any site. More specifically, 0.055 cGy would be the skin dose from a XVI-1/4
brain-site scan, and for the lung site and prostate site they would be 1.15 cGy and 0.6
cGy respectively. If we assume that the average prescription dose for IMRT or 3DCT is
200 cGy, and we perform a cone-beam scan at each fraction, the percentage dose from
XVI to MV would be 0.02%, 0.6%, and 0.3%. Adopting 0.2% of a MV dose as the dose
limit recommended by Murphy, et al.,3 the XVI dose to the brain site by a XVI-1/4 scan
would satisfy this dose limit; dose from XVI-1/4 scan for the other 2 sites would near the
limit. In hypo-fractionation stereotactic body radiotherapy cases, the patient’s average
-
47
MV dose is 1000cGy, so the CBCT dose for all sites would be well below the limit.
However, since no safe radiation level exists, the ALARA (“as low as reasonably
achievable”) standard for non-target doses for concomitant exposures is always
recommended by the radiology community. The radiation carcinogenesis risk for
patients undergoing radiati