Sterilizing Grade Filter & Filter Validation - Mr. Ajeet Singh - 09-03-2007
-
Upload
ajeets3627 -
Category
Documents
-
view
941 -
download
13
Transcript of Sterilizing Grade Filter & Filter Validation - Mr. Ajeet Singh - 09-03-2007
STERILIZING GRADE FILTER &
FILTER VALIDATION
Ajeet Kumar Singh
STERILIZING GRADE FILTER
“ A filter which, when challenged with the
micro-organism Pseudomonas diminuta
at a minimum concentration of 107/cm2of
filter surface, will produce a sterile
effluent” ………… (1987)
“ A filter that, when appropriately
VALIDATED, will remove all micro-
organisms from a fluid stream,
producing a sterile effluent”
………..FDA(2004)
PARAMETERS
USP 28 FDABP2007 &
EC-GMP
Nominal pore size
Retentive membrane with a nominal pore size 0.2µm or 0.22µm
Retentive membrane with a nominal pore size 0.2µm or less
Bacteria retentive membrane with a nominal pore size of 0.22 mm or less
Pre-filtration Bioburden
NoPre-filtration is not recommended but setting of pre-filtration limit for bioburden is to be established
Yes. Pre-filtration bioburden limits – Not More Than 10 cfu/100 ml
Integrity testBefore useAfter use
YesYes
YesYes
YesYes
Cited Integrity test
Bubble pointPressure hold
YesYes
YesNo
YesYes
Microbial challenge test
A challenge of not less than 107
P.dimunata suspension per sq. com of filter surface area
A challenge concentration of at least 107 organisms per cm2 of effective filtration area of B. diminuta
A challenge of at least 107 cfu of P. diminuta per cm2 of active filter surface
STERILIZING GRADE FILTER REQUIREMENTS
PORE SIZE RATING
o 0.22 µm sterilizing filter membranes :
Capable of retaining 100% of a culture of
107 P.diminuta /cm2 of membrane surface
under a pressure of NLT 30 psi (2.0 bar).
………….USP-28
Use of Multiple Sterilizing GradeFilters in Aseptic Processes ……..
US PERSPECTIVE
o Use of redundant sterilizing filters should be
considered in many cases. This additional filter
must be satisfactorily tested before use, but
does not require post-use integrity testing
unless the primary filter fails……………PDA
Technical report
Use of Multiple Sterilizing GradeFilters in Aseptic Processes ……..
EU PERSPECTIVE
o Potential additional risks as compared with other
sterilization processes can do second filtration
(sterilized microorganism retaining filter)
immediately prior to filling.
o The final sterile filtration should be as close
as possible to the filling point.………(EU GMP)
Multiple filters US Vs EU
US EU
Filter #1 passes integrity
testing post-use, batch can be
released
Filter #2 is discarded without
testing
Filter #1 (Bioburden
reduction filter)
Must pass the post integrity
test
Filter #2 (Final Sterilizing
filter)
Must pass the post integrity
test
If filter #1 fails integrity
testing post-use, than filter
#2 is tested
If filter #2 passes, batch can
be released
FDA GUIDANCE NOTE
21 CFR 211.113(b) states that "Appropriate
written procedures, designed to prevent
microbiological contamination of drug products
relate to be sterile, shall be established and
followed. Such procedures shall include
validation of any sterilization process."
VALIDATION FO ASEPTIC
PROCESSING AND
STERILLIZATION
Parameter
Filter Manufacturer
Filter UserData Provided by
Millipore
Sterilization
Procedure
Recommend sterilizing
procedure with temperature and cycle time limits
Operate within the prescribed limits.
Validate the process
application
-
Integrity Testing
Provide filter usage limits
Validate for their process and
solution
Filter usage limits with water provided in the
guide. Product specific limit to be set by filter
user
FibresMeet non-fibre
release claim [21 CFR 210.(3b)]
Document the appearance of the
filtered product
Manufacturer certifies that the filter/cartridge is
non-fiber shedding
Sterilizing grade filter
Provide information on release criteria
Validate with their process solution
--
Requirements/Responsibility for Filter Validation
Parameter
Filter Manufacturer
Filter UserData Provided by Millipore
ExtractablesProvide extractables
data – type and amount, Pass USP test
Evaluate with their system and solution
Data provided with water. Done on model stream
systems
CompatibilityProvide compatibility
tablesEvaluate with their
solutionData provided for different solvents
Toxicity
Provide test data showing product meets
USP plastic test requirements
Document -
AdsorptionProvide known
information
Evaluate with their solution (both for active as well as preservatives)
-
Requirements/Responsibility for Filter Validation (contd….)
FILTER VALIDATION
o DEMONSTRATION OF BACTERIAL
RETENTION
o COMPATIBILITY STUDIES
o INTEGRITY TESTING
BACTERIAL RETENTION STUDY FOR STERILIZING GRADE FILTER
VIABILITY STUDY
Most appropriate method of delivering the challenge of
107/cm2Product viability study with B. Diminuta--------Direct Inoculation
Viability decrease greater than or equal to 1 log
Or
Oil based product
PRODUCT IS CONSIDERED ANTAGONIST
Surrogate fluid e.g. Saline lactose Broth ….for bacterial
retention
BACTERIAL RETENTION STUDY
PRODUCT CONDITIONING STAGEProduct is continuously filtered at desired flow rate for the max. processing time after that filter is rinsed to remove the product from the filter—Dead ended challenge filtration
CHALLENGED STAGEB. Diminuta is suspended in product/media and Product is continuously filtered at desired flow rate for the max. processing time
Analysis of filtrate for the the presence of B.diminuta
BACTERIAL RETENTION STUDY (contd…)
ACCEPTANCE CRITERIA
o 0.22µm filter must demonstrate complete retention. o 0.45µm filter (Positive control) must demonstrate
complete recovery of test organism.
o Sterility of the test system shall be demonstrated
o 0.22 & 0.45µm filter must not fail in the post integrity test.
COMPATIBILITY ASSESSMENT
Chemical and Physical resistance
Filter is subjected to:
o Microscopic inspection for the absence of
damage which might influence the filter
performance
o Dimensional measurements
o Extractable and Leachable analysis
STRUCTURAL COMPATIBILITY
SWELLING OF FILTER
Pore space tighterFlow rate decrease
Bubble point ( )
PLASTICIZING EFFECT
Release the polymer stress
Pore space openFlow rate increaseBubble point ( )
o To ensure the structural compatibility recording
of Bubble point is necessary after contact with
the product
ADSORPTION
o Formulations containing low
concentration of actives
o Product containing preservatives
o Specific nature of actives ingredients
INTEGRITY TESTINGObjective: To ascertain whether changes in the filter
may have occurred during filtration
Conform the correctness of the filter manufacturer’s
pore-size rating
If fails in the final integrity test ------filtrate should be
re-filter or discarded
Laplace law
P: Pressure necessary to expel the water
D: Diameter : Surface tension
: Angle of wetting
P =4 Cos
D
Bubble Point
o Liquid is held in the pores of the filter by
surface tension and capillary forces.
o The minimum pressure required to force
liquid out of the the pores………..Bubble
point
o Bubble point is a function of wetting liquid
surface tension, pore size Specific to
membrane polymer
Filter wetted with Purified water(approx. 200 ml (Temp.---, Time 15-20 mins)
Record the bubble point
Repeat the test with rewetting the membrane until stable bubble point in three consecutive
tests
Repeat the test with test fluid
Calculate the Bubble point ratio (BPR)
Procedure for Bubble point determination
Manually determination of Bubble point
Bubble Point
If Bubble point value lower than the
specification:
o Non-integral membrane
o High temperature
o Integral filter,but wrong pore size
o Incomplete wetted membrane
Diffusive flow/Forward flow test
o Pressure at which gas molecules that migrate
from water-filled pores of a wetted
membrane.
o Directly proportional to the differential
pressure and the total surface area of the
filter.
o Filter Wetted with liquid Upstream side of
the filter pressurized diffusive flow is
measured
Pressure Hold/ Decay Test
o Indirect method diffusive low/forward flow
testing
o Filter housing pressurized Gas flow across the
membrane quantitatively measured as decay in
the pressure over a specified period of time
o Conform the integrity of the entire filter
assembly
o Pressure decay below the max. allowable value
Passes the physical integrity test
Test Sample Size
Bubble Point200 ml from three
different lot
Viability study50 ml from one lot
(Sterile)
Bacterial retention study & Compatibility
2000 mL from one single lot
Filter validation sample requirements
DRF MFG. SITE
Establish lead formula based on Pre-formulation
Studies
Compatibility
Adsorption
Integrity Testing (200 ml
non-sterile
sample of 3 lots)
Batch Size, Operating Conditions
•FILTER MFR.Extractables – Model Solvents
Bacterial Retention
50 ml sterile sample for viability testing
2000 ml
MAPPING OF ACTIVITY
Product Description
o Aq/Non aq.o Biological/Non Biologicalo pHo Specificity Gravityo Viscosityo Detail unit formula along with CAS
No.o MSDSo Neutralizing agento Sensitivity if anyo pH modification possibleo Product description (Preservative)
INFORMATION NECESSARY FOR ACCESS QUESTIONNAIRE
Critical Process Description
o Processing Temp. for Filtrationo Minimum Batch Size (Min, Actual
and max.)o Filtration (Continuous/Intermittent)o Flow Rate (mL/min)
Test gas o Compressed Air / Nitrogen/Other
Type o Membrane/Cartridge
Process type o Continous / intermittent
INFORMATION NECESSARY FOR ACCESS QUESTIONNAIRE
Contd...
Ensure that the following information is
available on the filter
o Compatibility
o Bacterial Challenge Testing
o Extractable Testing
o Product Specific Integrity Testing Values
CONCLUSION