Standalone VeloPix Simulation

Jianchun Wang

4/30/10

Page 2: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Introduction

VeloPix performance after irradiation affects our current design. We want to create tools to study these effects.

As a first step I modify the standalone simulation package and look at performance of pixel detector before irradiation. More features are to be added. More sophisticated electric field based ISE-TCAD simulation.

Charge trapping.

Finite integration time.

Disclaim: Some of the electronics properties may be too good to be true. If it is need, more realistic parameters can be added to provide input to other studies.

04/30/10 Jianchun Wang 2

Besides of many interesting features, I am particular interested in: The charge collection efficiency will be reduced due to insufficient bias

and/or charge trapping, etc.

Reconstruction of angled tracks are biased due charge collection inefficiency.

Non-uniform irradiation dose on a single chip/sensor makes it difficult bias.

Page 3: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Sensor and Electronics Properties

Silicon sensor Thickness = 150 mm. Charge carrier = electron Pixel size = 55 mm x 55mm. Full depletion voltage = 30 V Bias HV = 50 V

Electronics Charge collection efficiency = 100% Noise = 100 e Gain uncertainty = 10 % Crosstalk between adjacent pixels = 0 Threshold = 1000 e Non-uniformity of thresholds = 0 % ADC (TDC) bits = 8 ADC range = 1000 – 24000 e Non-linearity = 0

04/30/10 Jianchun Wang 3M

ore

real

istic

para

met

ers

will be

add

ed w

hen

they

are

avai

labl

e

Page 4: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Normal Incident Tracks

04/30/10 Jianchun Wang 4

Track: 20 GeV .p Row X, Column Y With VeloPix detector row and column have same pitches.

Angle X = 0Angle Y = 0

<N> =1.55 <N> =1.26

MPV ~ 11 Ke

Above ADC range

Page 5: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Eta Correction

04/30/10 Jianchun Wang 5

Pixel border

Linear charge weighting

~75%, no chargesharing info available

~25%, very narrow after eta correction partially due to small portion

eta correction

Page 6: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Tracks At Different Angles

04/30/10 Jianchun Wang 6

Tracks at 0 degree in Y/column direction

For threshold = 1000 e, the best resolution is at 18 – 20 degree.

Page 7: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

With Plane Tilted

04/30/10 Jianchun Wang 7

Tracks at 0 or 20 degree in Y/column direction

There are more charge sharing in column direction, thus slightly less charge sharing for normal incident track in X direction.

Page 8: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Different Thresholds and Noise

04/30/10 Jianchun Wang 8

Threshold = 1000 eNoise = 100 e

Threshold = 2000 eNoise = 100 e

Threshold = 1000 eNoise = 300 e

Just to show how thresholds and noise affect the resolution.

With increasing of noise, the resolution is affected the most at small angles where the shared charges are less.

Threshold affects the resolution, especially for thin detector. Trim DAC in each cell may be necessary to reduce the non-uniformity of threshold, and thus reduce the overall threshold level.

Page 9: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Plan

Use more realistic electronics parameters from TimePix studies, and generate inputs for other studies.

Add irradiation dose dependent effects

More sophisticated electric field based ISE-TCAD simulation.

Charge trapping.

Finite integration time.

May integrate it to more general simulation, depending CPU consumption ( ~10ms /hit ).

More interesting studies.

04/30/10 Jianchun Wang 9

Page 10: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Testbeam of Radiation

Hard Sensor

Page 11: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Telescope Configuration

11

TypeThickness(mm)

Size(mm2)

HV Comment

Telescope N-type Si 300 ?16 x 2432 x 16

220

DUT

sCVD 500 4x 4 250 At 0, 10, & 20, HV scan at 20

MCZ Si 300 ? 16 x 24 500 0, threshold scan

Purdue 3D 200 8x8 40 3d_4e_wb5_8, failed

200 8x8 40 3d_2e_wb216_6, HV & threshold scan

285 8x8 40 3d_2e_wb5_2, HV & threshold scan

DUTYY

120 GeV proton beam

Scint

X

Z

Y

XX XXYY

a: –22 b: +22 a: +22b: –22b: 0, –10, –20

Lab frame

04/30/10 Jianchun Wang

Page 12: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Diamond HV Scan at Angle What we want to extract from the testbeam for different bias HVs:

Total charge collected per particle hit in terms of MPV of the Landau distributions.

For a fixed threshold how the charge sharing information the detector can deliver, in terms of number of rows, or columns per particle hit cluster.

Spatial resolution.

Shift of spatial position measurement due to partial charge collection and tracks at angle. This can give us some ideas on effective depth, and charge trapping.

Status of each task:

Need more work on readout electronics gain and pedestal calibrations. It is difficult to compare the absolute charge before that.

Numbers of pixels per hit vs bias HV qualitatively agree with expectation. We need to obtain precise thresholds from bench test for MC simulation. Then we can have quantitative comparison to test our understanding.

Current resolution is not as good as expectation. Need more work on gain curve and telescope alignment.

Shift of center residual shows correct trend. It will be revised after the spatial measurement optimization.

04/30/10 12Jianchun Wang

Page 13: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Diamond Sensor Charge Sharing vs HV

Number of Columns per Cluster

Number of Rows per Cluster

HV = -250 V

Sensor rotated to ~ 20 in row direction.

More charge collected with higher bias HV till saturation.

Need more work on gain calibration to extract the absolute charge (MPV of Landau distribution).

04/30/10 13

Preliminary

Preliminary

Jianchun Wang

Page 14: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Diamond Sensor Residual Center vs HV

Tracks are at ~ 20 with respect to normal of sensor plane in row direction.

Use the same set of telescope spatial configuration parameters.

With low bias HV, charges generated near readout electronics have more chance to be collected, equivalent to thinner effective sensor. Thus the residual center shifts.

In extreme case, the maximum possible shift ~ tan(q)*d/2 ~ 90 mm.

04/30/10 14

Preliminary

Jianchun Wang

Page 15: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Diamond Sensor Charge Sharing vs Angle

Diamond sensor is biased at -250 V.

Sensor was perpendicular to beam, or rotated by ~10 & ~20 in row direction.

Gain and threshold of the electronics are different from that of HV scan.

04/30/10 15

Preliminary

Preliminary

Jianchun Wang

Page 16: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Charge Distributions

04/30/10 Jianchun Wang 16

Charge (Ke)

Silicon Telescope

sCVD DUT

Plane 0

Plane 8

Plane 4

Charge (Ke)

Good

Bad

Weird

Page 17: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Problem with Diamond Gain Calibration

04/30/10 Jianchun Wang 17

sCVD DUT

Charge (Ke)

Double peaks belong to different cluster sizes, suggesting there is offset issue.

The difference between two peaks is too big.

The offset would have to be ~ –13 in order to have the same MP. And the MP would be ~ –3. So this is not a correct hypothesis.

MP=9.8

MP=22.7

Charge (Ke)

Charge (Ke)

Page 18: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Diamond Residual Distribution

04/30/10 Jianchun Wang 18

Xrecon – Xtrack (mm) Yrecon – Ytrack (mm)

s = 30.5 mm s = 46.7 mm

Tracks are at ~20 in X direction wrt the diamond.

Charging sharing information is not fully used yet due to calibration issues.

In comparison 100/12 = 28.9, 150/12 =43.3.

Page 19: Standalone VeloPix Simulation Jianchun Wang 4/30/10.

Summary

We had tested radiation hard sensors: diamond, MCZ silicon & 3D.

Some interesting results are produced from diamond test.

Gain calibration somehow becomes bottle neck.

We provide offline analysis and alignment program for this testbeam system as our promised contribution.

We may use the telescope to test our own sensors.

04/30/10 Jianchun Wang 19