SSL-UCB, ALD MCP Test Progress

5
Space Sciences Laboratory, University of California, Berkeley SSL-UCB, ALD MCP Test Progress Sent two MCPs for Al 2 O 3 ALD coating of half surface Both MCPs standard glass MCPs ~ 90Meg Ω

description

SSL-UCB, ALD MCP Test Progress. Sent two MCPs for Al 2 O 3 ALD coating of half surface Both MCPs standard glass MCPs ~ 90Meg Ω. SSL-UCB, ALD MCP Test Progress. The two MCP assemblies were tightened differently, the 12nm sample (left) shows a shadow where the MCP was - PowerPoint PPT Presentation

Transcript of SSL-UCB, ALD MCP Test Progress

Page 1: SSL-UCB, ALD MCP Test Progress

Space Sciences Laboratory, University of California, BerkeleySSL-UCB, ALD MCP Test Progress

Sent two MCPs for Al2O3 ALD coating of half surface

Both MCPs standard glass MCPs ~ 90Meg Ω

Page 2: SSL-UCB, ALD MCP Test Progress

Space Sciences Laboratory, University of California, Berkeley

SSL-UCB, ALD MCP Test Progress

The two MCP assemblies were tightened differently, the

12nm sample (left) shows a shadow where the MCP was

clamped in, the 6nm (right) sample shows no deposits at all.

The 12nm MCP was a bit loose, and the 6nm was tight.

Page 3: SSL-UCB, ALD MCP Test Progress

Space Sciences Laboratory, University of California, BerkeleySSL-UCB, ALD MCP Test ProgressML201-B-2 12nm ALD shows some debris and smudging that was removed in cleaning

Before cleaning

After cleaning

Page 4: SSL-UCB, ALD MCP Test Progress

Space Sciences Laboratory, University of California, BerkeleySSL-UCB, ALD MCP Test ProgressML201-B-1 6nm ALD shows less debris and smudging, - was removed in cleaning

Before cleaning

After cleaning

Page 5: SSL-UCB, ALD MCP Test Progress

Space Sciences Laboratory, University of California, Berkeley

Initial Imaging Results on Standard MCPswith 6nm ALD coating

1150v MCP, 2800v Screen, UV floodPhotonis ML201-B-1 MCP, 86MΩ

1050v MCP, 2900v Screen, UV floodPhotonis ML201-B-1 MCP, 86MΩ

Shows definite enhancement of brightness in ALD areaNo degradation of image quality, except at the boundary zone for coating