Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type...

16
Smart Calibration Unit Mk4 Operator manual 431-441-01 January 2014

Transcript of Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type...

Page 1: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

Smart Calibration Unit Mk4

Operator manual

431-441-01

January 2014

Page 2: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

2014 © Mecmesin Ltd, supplied with Mecmesin test systems and not for redistribution

Part no. 431-441-01

Page 3: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

SCU Mk4 Mecmesin iii

Contents

1. Overview 1

1.1 Before use 1

1.2 Maintenance and repair 1

1.3 Mounting the SCU 1

2. Powering the gauge 2

2.1 Replacing batteries 2

2.2 Mains operation 2

2.3 Fitting of alkaline batteries 3

2.4 Battery safety information 3

3. Adjustment and calibration stages 4

3.1 Powering up 4

3.2 Start-up screen 4

3.3 Inversion of loadcell signal 4

3.4 Unit of measurement 5

3.5 Loadcell type select 5

3.6 Initial zero screen (‘IO’) 7

3.7 Compression zero screen (‘CO’/’ACWO’) 7

3.8 Compression max screen (‘CM’/’ACWM’) 8

3.9 Tension Zero Screen (‘TO’/’CWO’) 8

3.10 Tension max screen (‘TM’/’CWM’) 9

4. Linearisation 10

4.1 Diagnostic screen 10

4.2 Error messages 10

Page 4: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell
Page 5: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

SCU Mk4 Mecmesin 1

1. Overview

This instrument’s firmware only allows the adjustment and calibration of Smart cells. It does not communicate with an internal loadcell.

As with the previous SCU, this instrument allows the user to skip different stages depending on whether the Smart is calibrated or not.

• With an uncalibrated Smart, the user must enter loadcell capacity, initial offset and at least one of the compression/tension stages.

• With a calibrated Smart, provided the loadcell capacity or initial offset settings are not modified, the user can skip straight to the ‘linearisation’ stage and back again.

No data is written to the EEPROM until all the stages have been completed, so if the user skips back to an earlier stage, any calibration readings taken will be lost.

Data is automatically stored after the ‘adjustment’ stage. Once the data is committed to memory, during the ‘linearisation’ stage, the instrument can display the contents of the Smart’s EEPROM.

1.1 Before use

On receiving the unit, check that no physical damage has occurred to the packaging, plastic case or the instrument. If any damage is evident, notify Mecmesin immediately.

1.2 Maintenance and repair

When cleaning the keypad, care must be taken to avoid liquids, especially alcohols, seeping around the edge of the membrane. We therefore recommend the use of a lightly-dampened cloth to avoid liquid spillage onto the membrane.

Mecmesin test instruments contain no user-serviceable parts. If the gauge fails, or appears to behave in an unusual manner, contact your local supplier for support. Do not continue to use the gauge until it has been checked and if necessary, repaired and returned to a safe working condition. Do not attempt any repairs without consulting Mecmesin or your local supplier.

1.3 Mounting the SCU

On the rear of the gauge there are two M5 threaded holes, used for mounting the gauge. Ensure that the screws used are threaded into the gauge to a maximum depth of 12 mm. If screws are fitted beyond this depth, damage to the internal PCB may occur.

Page 6: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

2 Mecmesin SCU Mk4

2. Powering the gauge

The AFG is supplied with a set of five Nickel Metal Hydride AAA rechargeable batteries, which are supplied fully charged to allow use straight from the box. Do not use any other battery charger other than that supplied with the force gauge.

2.1 Replacing batteries

To replace the batteries you must first remove the battery cover on the upper part of the rear of the gauge by removing the two retaining screws. Remove the fitted batteries and fit the five new batteries in the battery holder, ensuring that you observe polarity. Place the batteries on top of the ‘release tag’ so they can be freed easily next time from the spring-loaded contacts.

Refit the battery cover and tighten the two retaining screws.

Connect the mains adaptor/charger to the AFG charger socket located at the right-hand side of the gauge next to the display, and charge the batteries for 14 to 16 hours. Only use the adaptor/charger supplied. A fully-charged battery pack will provide approximately 20 hours constant use between charges.

A low battery symbol will appear in the display approximately two minutes before the gauge powers down:

2.2 Mains operation

The SCU can also be powered directly from the mains. This can be achieved with or without the rechargeable batteries being fitted. Connect the mains adaptor/charger to your mains supply. Only use the adaptor/charger supplied. If rechargeable batteries are fitted, a trickle charge will be applied to the batteries with the display switched on.

Page 7: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

SCU Mk4 Mecmesin 3

2.3 Fitting of alkaline batteries

The SCU can also be powered by AAA 1.5 V alkaline batteries (not supplied). For the fitting of alkaline batteries, follow fitting instructions as per rechargeable batteries above. Warning: When alkaline batteries are fitted, the mains adaptor/charger must NEVER be connected to the AFG due to the risk of acid leakage, which could damage the instrument.

2.4 Battery safety information

NEVER:

• short circuit

• disassemble or deform cells

• heat or incinerate

• immerse in water

• solder anything to the battery terminals

• reverse individual cell polarity

• use alternative chargers other than those supplied by Mecmesin

• use replacement parts other than those supplied by Mecmesin

Never dispose of batteries with ‘normal’ refuse. Contact your local Environmental Authority to determine the location of your appropriate disposal facility

Page 8: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

4 Mecmesin SCU Mk4

3. Adjustment and calibration stages

3.1 Powering up

Warning! The SCU must be powered down when connecting or disconnecting smart transducers.

To power up the gauge press the red key.

3.2 Start-up screen

The start-up screen displays the following:

• MECMESIN

• SCU in large letters

• Firmware version: (1.13 here was current at the time of publishing)

• Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated)

3.3 Inversion of loadcell signal

For a calibrated loadcell, the sign bit will be replicated and shown on the calibration screens. A ‘+’ will indicate what is considered correct operation i.e. Compression reduces count value. A ‘–’ will indicate that the load cell has reversed operation.

For uncalibrated Smart cells, the gauge will default to correct operation and a ‘+’ will be displayed.

It may be necessary to invert the output signal of some manufacturers single direction load cells as their convention could be that the loadcell’s output signal is positive for a compressive load.

Page 9: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

SCU Mk4 Mecmesin 5

If the operator changes the sign during calibration and subsequently cancels, reverting back to the loadcell selection, the sign will revert back to the original settings.

The sign bit can only be changed during the initial offset (IO) stage by pressing the Zero/Down button.

3.4 Unit of measurement

During the loadcell selection stage, the loadcell capacity is specified in newtons (N) for force and newton-metres (N.m) for torque. During the initial offset stage, the calibration unit of measurement can be changed by pressing the Units/Menu button but becomes fixed for subsequent stages (the user can skip back to change them).

3.5 Loadcell type select

Display

Line 1 (Left) Force / Torque 1 / Torque 2; (Right) blank / C / T

Line 2 Loadcell capacity

Line 4 Units N for Force, / N.m for Torque

Bidirectional calibration Unidirectional compression only Unidirectional tension only

Pressing Max/Esc allows the user to cycle through adjustment for: both spans, compression only, tension only.

Buttons

Max/Esc Allows single sided adjustment C or T

Units/Menu Switches between Torque/Force cell

Txd/Up Increase load cell capacity

Zero/Down Reduce load cell capacity

Reset/Enter Select the load cell capacity and move onto next step

Pressing the Max/Esc button allows the user to adjust a compression or tension only load cell. The span that is not selected is automatically set to the adjusted span. By default both spans must be calibrated. Pressing Max/Esc allows the user to cycle through adjustment for: both spans, compression only, tension only.

Page 10: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

6 Mecmesin SCU Mk4

Initial settings

Uncalibrated loadcell Force, 100 N, Torque, 6 N.m

Calibrated loadcell Type and capacity of loadcell

Selectable Force/Torque capacities:

Force (N) Torque (N.m) Table 1

Torque (N.m) Table 2

1 0.01 0.01

2 0.015 0.02

2.5 0.03 0.025

5 0.06 0.05

10 0.1 0.1

20 0.15 0.2

25 0.3 0.25

50 0.6 0.5

100 1 1

200 1.5 2

250 3 2.5

500 6 5

1000 10 10

2000 15 20

2500 30 25

5000 60 50

10000 100 100

20000 150 200

25000 300 250

50000 600 500

100000 1000 1000

200000 1500 2000

250000 3000 2500

500000 6000 5000

Page 11: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

SCU Mk4 Mecmesin 7

Additional

If the loadcell is calibrated, the screen will always default to the calibrated type (for example, when pressing escape on the initial offset screen). If the loadcell is not calibrated then it is initially set to the default value (100 N Force or 6 N.m Torque). Subsequently returning to the screen should display the default capacity for the cell type.

3.6 Initial zero screen (‘IO’)

Display

Header ‘+’ for correctly wired load cells, ‘-‘ for load cells that required inverting.

Middle line Load cell counts

Footer (middle) ‘IO’ indicates initial zero

Footer (right) Units Force: ‘N’ / ‘kgf’ / ‘lbf’ Torque: ‘N.m’ / ’kgf.cm’ / ‘lbf.in’

Buttons

Max/Esc Back to loadcell select screen

Units/Menu Selects units (see above)

Txd/Up If the Smart is calibrated it will advance to compression zero screen

Zero/Down Inverts load cell count value. Toggles ‘+’/’-‘ symbol

Reset/Enter Enters count value and moves to next screen

The ‘+’/’-‘ symbol can only be changed during this, the initial offset (IO) stage.

3.7 Compression zero screen (‘CO’/’ACWO’)

Display

Header ‘+’ for correctly wired loadcells, ‘-‘for load cells that required inverting.

Middle line Load cell counts

Footer (middle) ‘CO’/’ACWO’ indicates Compression zero or Anti-Clockwise zero

Footer (right) Units Force: ‘N’ / ‘kgf’ / ‘lbf’ Torque: ‘N.m’ / ’kgf.cm’ / ‘lbf.in’

Buttons

Max/Esc Back to previous stage

Units/Menu N/A

Txd/Up Skip stage (if possible)

Page 12: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

8 Mecmesin SCU Mk4

Zero/Down N/A

Reset/Enter Enters count value and moves to next stage

3.8 Compression max screen (‘CM’/’ACWM’)

Display

Header ‘+’ for correctly wired loadcells, ‘-‘for load cells that required inverting.

Middle line Load cell counts

Footer (middle) ‘CM’/’ACWM’ indicates Compression Max or Anti-Clockwise Max

Footer (right) Units Force: ‘N’ / ‘kgf’ / ‘lbf’ Torque: ‘N.m’ / ’kgf.cm’ / ‘lbf.in’

Buttons

Max/Esc Back to previous stage

Units/Menu N/A

Txd/Up Skip stage (if possible)

Zero/Down N/A

Reset/Enter Enters count value and moves to next stage

3.9 Tension Zero Screen (‘TO’/’CWO’)

Display

Header ‘+’ for correctly wired loadcells, ‘-‘for load cells that required inverting.

Middle line Load cell counts

Footer (middle) ‘TO’/’CWO’ indicates Tension zero or Clockwise zero

Footer (right) Units Force: ‘N’ / ‘kgf’ / ‘lbf’ Torque: ‘N.m’ / ’kgf.cm’ / ‘lbf.in’

Buttons

Max/Esc Back to previous stage

Units/Menu N/A

Txd/Up Skip stage (if possible)

Zero/Down N/A

Reset/Enter Enters count value and moves to next stage

Page 13: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

SCU Mk4 Mecmesin 9

3.10 Tension max screen (‘TM’/’CWM’)

Display

Header ‘+’ for correctly wired loadcells, ‘-‘for load cells that required inverting.

Middle line Load cell counts

Footer (middle) ‘TM’/’CWM’ indicates Tension Max or Clockwise Max

Footer (right) Units Force: ‘N’ / ‘kgf’ / ‘lbf’ Torque: ‘N.m’ / ’kgf.cm’ / ‘lbf.in’

Buttons

Max/Esc Back to previous stage

Units/Menu N/A

Txd/Up Skip stage (if possible)

Zero/Down N/A

Reset/Enter Enters count value and moves to next stage

The calibration data is automatically stored after the ‘adjustment’ stage.

Page 14: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

10 Mecmesin SCU Mk4

4. Linearisation

Middle line Load reading

Lower left Tension/compression or torque direction symbols

Lower right Units

Bottom Load bar

Max/Esc Back to previous stage

Units/Menu Change units

Txd/Up N/A

Zero/Down tare/zero

Reset/Enter Diagnostic screen

4.1 Diagnostic screen

During the ‘linearisation’ stage the instrument can display the contents of the Smart cell’s EEPROM by pressing the Reset/Enter button.

The EEPROM readings screen (detailed above) can also be accessed by holding down Max/Esc during start-up. Turn the gauge off to clear the screen.

4.2 Error messages

Error: No Smart

• Smart instrument not connected.

• Smart PCB has not been programmed.

• Smart PCB cannot be read.

Page 15: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

SCU Mk4 Mecmesin 11

Insufficient counts

The SCU requires the load cells output to move by 2500 counts between the zero and maximum load stages.

• Incorrect load applied for the loads cell capacity.

• Load cell output not 2mV/V +-0.2mV/V as required.

• Faulty Smart PCB.

Tensile load not applied

• A load in tension has not been applied at ‘TM’ stage.

• Load cell signal wired incorrectly to Smart PCB.

• Incorrect use of load cell inversion at ‘IO’ stage.

Compressive load not applied

• A load in compression has not been applied at ‘CM’ stage.

• Load cell signal wired incorrectly to Smart PCB.

• Incorrect use of load cell inversion at ‘IO’ stage.

CW torque not applied

• A load in the clockwise direction has not been applied at ‘CWM’ stage.

• Load cell signal wired incorrectly to Smart PCB.

• Incorrect use of load cell inversion at ‘IO’ stage.

ACW torque not applied

• A load in the anti-clockwise direction has not been applied at ‘ACWM’ stage.

• Load cell signal wired incorrectly to Smart PCB.

• Incorrect use of load cell inversion at ‘IO’ stage.

Page 16: Smart Calibration Unit Mk4 - archive.mecmesin.com SCU Mk4... · • Smart cell capacity and type (or XXXXN if a Smart cell is not present or uncalibrated) 3.3 Inversion of loadcell

General Operation.

Mecmesin : a world leader in affordable force and torque testing solutions

Since 1977, Mecmesin has assisted thousands of companies achieve enhanced quality control in design and production. The Mecmesin brand represents excellence in accuracy, build, service, and value. In production centres and

research labs worldwide, designers, engineers, operators, and quality managers endorse Mecmesin force and torque testing systems for their high performance across countless applications.

www.mecmesin.com

Algeria Argentina Australia Austria Bangladesh Belgium Brazil Bulgaria Cambodia Canada Chile China Colombia Costa Rica Croatia Czech Republic Denmark

Ecuador Egypt Estonia Finland France Germany Greece Hungary India Indonesia Iran Ireland Israel Italy Japan Korea Laos

Latvia Lebanon Lithuania Malaysia Mexico Morocco Myanmar Netherlands New Zealand Norway Peru Philippines Poland Portugal Romania Russia Saudi Arabia

Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sweden Switzerland Taiwan Thailand Tunisia Turkey UK USA Vietnam

The Mecmesin global distribution network guarantees your testing solution is rapidly delivered and efficiently serviced, wherever you are.

Head Office

Mecmesin Limited

w: www.mecmesin.com

e: [email protected]

France

Mecmesin France

w : www.mecmesin.fr

e: [email protected]

Germany

Mecmesin GmbH

w: www.mecmesin.de

e: [email protected] North America

Mecmesin Corporation

w: www.mecmesincorp.com

e: [email protected]

Asia

Mecmesin Asia Co. Ltd

w: www.mecmesinasia.com

e: [email protected]

China

Mecmesin (Shanghai) Pte Ltd

w: www.mecmesin.cn

e: [email protected]