Small fields dosimetry in the commissioning of medical linacs
Transcript of Small fields dosimetry in the commissioning of medical linacs
Small fields dosimetry
in the commissioning of medical linacs
Kinga Marta Jakub Agnieszka Aneta
Polaczek-Grelik Stefańczyk Derus Giłka Kawa-Iwanicka
Department of Medical Physics University of Silesia
Katowice, Poland
Cancer Diagnosis and Treatment Center Katowice, Poland
Small-field issues
Physical aspects
Occlusion of primary radiation source
Overlapping penumbra regions (dose profile broadening as a consequence)
No plateau in the field center
Lateral charged particles disequilibrium
Technical problems
Detector perturbation of photon fluence
Volume-averaging effect
• Underestimation of the dose in the center of field – output factors affected
• Washed out the penumbra region – profile shape affected
No enough space for a reference detector
In-air measurements instead of water
IPEM Report 103
Das et al. Med. Phys 35(1), 2008.
Introduction
Linac commissioning
• Field sizes: 10x10, 30x30 [cm2]
• In-plane and cross-plane scans
• Dmax on the base of PDD curves
• Open fields measurements
• Particular detector type not stated
Oncentra External Beam modelling
• Field sizes: 5x5, 10x10, 15x15, 20x20, 30x30, max x max [cm2]
• In-plane and cross-plane scans
• Output factors in water and in air
• PDD curves
• Diagonal and star profiles
• Open and wedge fields measurements
• As small detector as available, preferably diamond or diode detector
The aim: Investigation of the influence of detector type and orientation on profile parameters examined during linac commissioning and important for TPS modelling.
Material: Three Elekta Precise linear accelerator with Agility system Photon beams: 6 MV, 10 MV and 15 MV
Radiation Detectors
Ionization chambers
less quantum noise
more tissue-equivalent materials
decrease of response in high dose rates
Semiconductor diodes
energy dependence of signal in low-energy range
more uncertain for large fields and depths
overresponse in high dose rates
Feature CC13 CC01 SFD3G PFD3G
Outer electrode C-552 C-552 ― ―
Inner electrode C-552 Steel ― ―
Active material air air pSi pSi
Enclosure material ― ― ABS ABS
Stem material PEEK PEEK Steel Steel
Diameter [mm] 6.0 2.0 0.6 2.0
Length [mm] 5.8 3.6 0.5 0.5
Active volume [mm3] 0.13 0.01 0.0004 0.001
Sensitivity [nC/Gy] ~3.8 ~0.4 ~4 ~100
Detector orientation
Ionization chambers
upward – parallel to CAX
perpendicular to the beam axis
Semiconductor diodes
upward – parallel to CAX
Geometrical conditions
Field sizes at isocentre
10x10 cm2, 5x5 cm2, 3x3 cm2
Depth of measurements
1.5 cm, 5 cm, 10 cm, 20 cm
SSD = 90 cm
Method
Air-filled ionisation chambers
CC01 ion chamber reproduces regions of field edges as good as diodes. Whereas dose gradient regions seen by CC13 are washed out.
Outside the field of irradiation CC01 indications are higher than other investigated detectors, therefore the field size based on its readings is larger than for other detectors (from 0.6% to 3%)
TPS based on CC01 scan profiles might give overestimated doses outside the PTV. This is important for combination of several adjacent fields in radiotherapy process.
Results & Conclusions
Semiconductor diodes
Unshielded photon diode tends to overresponse outside the field of radiation.
This process is energy-dependent and field size-dependent, as the contribution of scattered radiation increases.
3x3 cm2 beam proflies as seen by stereotactic and shielded photon diodes at isocentre in water phantom
Results & Conclusions
6MV
15MV
CC13 vs. PFD
Profile scans for 5x5cm2 field size at different depths and for various beam energies.
6 MV 15 MV
Results & Conclusions
Transversal scans (limited by MLC)
Longitudinal scans (limited by solid jaws)
Profile parameters
Field size Beam energy CC13 CC01 SFD3G PFD3G
10 x 10 cm2 6 MV 100.3 100.3
100.9 101.4
100.6 100.3
10 MV 100.4 100.5
101.1 101.5
100.4 100.3
15 MV 100.6 100.6
101.1 101.3
100.6 100.7
5 x 5 cm2 6 MV 100.4 100.4
101.0 100.6
100.7 100.2
10 MV 100.5 100.5
101.2 101.9
100.7 100.3
15 MV 100.6 100.7
102.9 102.6
100.9 100.4
3 x 3 cm2 6 MV 100.1 100.2
100.6 101.2
100.7 100.5
10 MV 100.4 100.5
100.4 100.9
100.5 100.6
15 MV 100.6 100.6
100.5 101.3
100.2 100.3
Symmetry parameter for in-line (Y direction) profile scans by various detectors in upward orientation. Data for orthogonal location of ion chambers for comparison purposes are shown.
Depth Beam Energy CC13 S F
PFD S F
1.5 cm 6 MV 100.2 100.8 100.1 100.3
10 MV 100.6 100.8 100.6 101.1
15 MV 100.2 101.2 100.5 100.9
5.0 cm 6 MV 100.2 101.4 100.2 100.9
10 MV 100.4 101.8 100.4 100.7
15 MV 100.3 102.8 100.6 101.8
10.0 cm 6 MV 100.4 102.0 100.3 101.5
10 MV 100.5 102.3 100.3 101.2
15 MV 100.3 103.5 100.5 102.3
20.0 cm 6 MV 100.5 102.8 100.4 102.0
10 MV 100.7 103.1 100.4 101.9
15 MV 100.5 104.6 100.7 102.9
Symmetry and flatness parameters for in-line (Y direction) profile scans of 5x5cm2 field as seen by ionisation chamber and shielded photon diode.
Results & Conclusions
Penumbra region
- Diodes show good reproducibility in small fields (below 3x3cm2).
- Shielded diode could be used even for standard field sizes, whereas stereotactic diode is not appropriate for this purpose
- Small ion chambers (CC01) has an advantage over CC13 for fields below 5x5cm2, but its orientation has to be carefuly chosen.
- For large/standard fields ion chambers of various dimensions give similar results
- CC13 ion chamber overestimates the penumbra size of about 55% regardless of the measurement depth .
Penumbra regon as a function of in-water depth for the field 5x5 cm2 at isocentre for 10 MV beam.
10 x 10 5 x 5 3 x 3
CC13 parallel 7.4 6.8 6.3
CC13 orthogonal 7.5 6.8 6.4
CC01 parallel 7.8 6.4 5.6
CC01 orthogonal 6.8 6.0 5.4
PFD 4.8 4.4 4.1
SFD 5.6 4.7 4.2
Penumbra region as seen by various detectors as a function of field size. Scans in Y direction.
Depth [cm] PFD CC13
1.5 3.6 5.7
5.0 4.1 6.3
10.0 4.4 6.8
20.0 4.9 7.6
Results & Conclusions
Penumbra region
CC13 CC01
PFD
SFD
10 x 10 cm2
5 x 5 cm2
3 x 3 cm2
Results & Conclusions
- Diodes show good reproducibility in small fields (below 3x3cm2).
- Shielded diode could be used even for standard field sizes, whereas stereotactic diode is not appropriate for this purpose
- Small ion chambers (CC01) has an advantage over CC13 for fields below 5x5cm2, but its orientation has to be carefuly chosen.
- For large/standard fields ion chambers of various dimensions give similar results
- CC13 ion chamber overestimates the penumbra size of about 55% regardless of the measurement depth .
Thank you for your attention