Small Animal Ventilator for Hyperpolarized Helium MRI
description
Transcript of Small Animal Ventilator for Hyperpolarized Helium MRI
![Page 1: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/1.jpg)
Small Animal Ventilator for Hyperpolarized Helium MRI
Ashley Anderson III, Micah Brown, Matt Smith, Chris Wegener
Client: Dr. Sean FainAdvisor: Dr. Willis Tompkins
![Page 2: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/2.jpg)
MRI compatible ventilator system Deliver 20% O2 and 80% He-3 in
same breath Allows constant scanning of the He-3 in
MRI rather than every 4 breaths
Client Requirements
![Page 3: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/3.jpg)
Background - Helium MRI
Newer technique Same MRI scanner
Convention is H Here, tuned to Helium
(48.6 MHz) Breath of Helium Displays respiratory
channels
![Page 4: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/4.jpg)
Project Concept
• Each syringe feeds off of independent gas reservoirs
• Gas mixing occurs right before entering small animal
![Page 5: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/5.jpg)
First Generation Prototype
Currently in use in animal studies Only using one syringe Not compatible with small syringes
for mouse studies
![Page 6: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/6.jpg)
Modifications to G1
Reinforced syringe clips
Larger motor Eliminated
O2 chaser breath
![Page 7: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/7.jpg)
Calibration
Results using manometer testing. Graph shows small deviation from entered volumes.
![Page 8: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/8.jpg)
Second Generation Concept
![Page 9: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/9.jpg)
Second Generation Prototype
Created in Solidworks
Smaller components
Interchangeable syringes
Adjustable axle support
![Page 10: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/10.jpg)
Second Generation Prototype
Large motor Motor shielding Precision
machining Can be
disassembled
![Page 11: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/11.jpg)
From Solidworks to Reality
![Page 12: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/12.jpg)
Does it move?
![Page 13: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/13.jpg)
Did we meet client requirements?
Gear ratios allow the appropriate 80/20 gas output ratio
MRI Safe materials Motor shielded for RF Rat/Mouse
compatible
![Page 14: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/14.jpg)
Future Work
Calibration testing Volume delivery Timing
Software development Simultaneous 3He/O2 delivery Inspired time based on %Insp and RR Valve switching
Mixing testing (imaging)
![Page 15: Small Animal Ventilator for Hyperpolarized Helium MRI](https://reader035.fdocuments.in/reader035/viewer/2022062411/568167a0550346895ddceb58/html5/thumbnails/15.jpg)
Questions?