1 Size-Exclusion Ultra Performance Liquid Chromatography for the Analysis of Covalent High Molecular Weight Insulin Stephan Koza, Paula Hong, and Kenneth J. Fountain Waters Corporation, 34 Maple Street, Milford, MA, USA WATERS SOLUTIONS ACQUITY UPLC ® H-Class Bio System ACQUITY UPLC BEH125 SEC, 1.7 µm Column Empower 2 Software KEY WORDS Size-exclusion chromatography, UPLC, insulin, USP, EP, method development, aggregates APPLICATION BENEFITS Improved separation of insulin monomer and covalent high molecular weight forms Faster analysis times and high throughput SEC separation Increased resolution of low molecular weight insulin degradant Reduced acetonitrile containing waste-stream volumes INTRODUCTION According to the World Diabetes Foundation there were an estimated 285 million people world-wide with diabetes in 2010, which is predicted to grow to 438 million people by 2030 and insulin is the primary treatment for this affliction. In 1982, insulin was the first recombinant biopharmaceutical, and by 2009, world-wide insulin and insulin analog sales were $13.1 billion. 1 As part of the US and European Pharmacopoeia (USP and EP), one of the critical quality control attributes for injectable insulin is the control of covalent high molecular weight (HMW) proteins. 2, 3 The current USP 34 (p. 3134) and EP (Volume 5, p. 1800) monograph methods for this HMW determination are based on HPLC size-exclusion chromatography (SEC). The USP method prescribes an L20 packing (dihydroxypropane groups chemically bonded to porous silica particles, 3 to 10 micro-m in diameter) in a 7.8-mm x 30-cm geometry and the EP method prescribes the use of a “hydrophilic silica gel for chromatography R (5-10 μm) with a pore size of 12-12.5 nm, of a grade suitable for the separation of insulin monomer from dimer and polymers” with a length of 30 cm and a minimum internal diameter of 7.5 mm. Presented in this application are the advantages that may be realized using a 125Å pore size, sub-2-µm ethylene-bridged hybrid (BEH) silica packing material and Waters Ultra Performance Liquid Chromatography (UPLC ® ) instrumentation for this traditional analysis. Among these advantages are faster run times, higher sensitivity, and higher resolving separations of insulin and covalent insulin HMW, while at the same time greatly reducing acetonitrile containing waste-stream volumes.

Transcript of Size-Exclusion Ultra Performance Liquid Chromatography for ... · Size-Exclusion Ultra Performance...

Page 1: Size-Exclusion Ultra Performance Liquid Chromatography for ... · Size-Exclusion Ultra Performance Liquid Chromatography for the Analysis of Covalent High Molecular Weight Insulin

1

Size-Exclusion Ultra Performance Liquid Chromatography for the Analysis of Covalent High Molecular Weight InsulinStephan Koza, Paula Hong, and Kenneth J. Fountain Waters Corporation, 34 Maple Street, Milford, MA, USA

WAT E R S SO LU T IO NS

ACQUITY UPLC® H-Class Bio System

ACQUITY UPLC BEH125 SEC,

1.7 µm Column

Empower™ 2 Software

K E Y W O R D S

Size-exclusion chromatography, UPLC,

insulin, USP, EP, method development,

aggregates

A P P L I C AT IO N B E N E F I T S ■■ Improved separation of insulin

monomer and covalent high

molecular weight forms

■■ Faster analysis times and high

throughput SEC separation

■■ Increased resolution of low molecular

weight insulin degradant

■■ Reduced acetonitrile containing

waste-stream volumes

IN T RO DU C T IO N

According to the World Diabetes Foundation there were an estimated 285 million

people world-wide with diabetes in 2010, which is predicted to grow to 438 million

people by 2030 and insulin is the primary treatment for this affliction. In 1982,

insulin was the first recombinant biopharmaceutical, and by 2009, world-wide

insulin and insulin analog sales were $13.1 billion.1 As part of the US and European

Pharmacopoeia (USP and EP), one of the critical quality control attributes for

injectable insulin is the control of covalent high molecular weight (HMW) proteins.2, 3

The current USP 34 (p. 3134) and EP (Volume 5, p. 1800) monograph methods for

this HMW determination are based on HPLC size-exclusion chromatography (SEC).

The USP method prescribes an L20 packing (dihydroxypropane groups chemically

bonded to porous silica particles, 3 to 10 micro-m in diameter) in a 7.8-mm x 30-cm

geometry and the EP method prescribes the use of a “hydrophilic silica gel for

chromatography R (5-10 μm) with a pore size of 12-12.5 nm, of a grade suitable for

the separation of insulin monomer from dimer and polymers” with a length of 30 cm

and a minimum internal diameter of 7.5 mm.

Presented in this application are the advantages that may be realized using a

125Å pore size, sub-2-µm ethylene-bridged hybrid (BEH) silica packing material

and Waters Ultra Performance Liquid Chromatography (UPLC®) instrumentation

for this traditional analysis. Among these advantages are faster run times,

higher sensitivity, and higher resolving separations of insulin and covalent

insulin HMW, while at the same time greatly reducing acetonitrile containing

waste-stream volumes.

Page 2: Size-Exclusion Ultra Performance Liquid Chromatography for ... · Size-Exclusion Ultra Performance Liquid Chromatography for the Analysis of Covalent High Molecular Weight Insulin

2 Size Exclusion UPLC Analysis of Insulin

E X P E R IM E N TA L

LC Conditions

System: ACQUITY UPLC H-Class Bio

with TUV and titanium

flow cell

Wavelength: 276 nm

Column Temp.: 25 °C

Sample Temp.: 10 °C

Injection Volume: 10 µL

(unless specified otherwise)

Flow Rate: 0.4 mL/min

(unless specified otherwise)

Mobile Phase: L-arginine (1.0 g/L)/acetic

acid (99%)/acetonitrile;

65/15/20 (v/v/v)

Wash and Injector

Needle Purge: 10% acetonitrile

Seal Wash: 10% methanol

Sample Diluent: 0.01 N hydrogen chloride

Columns

ACQUITY UPLC BEH125 SEC, 1.7 µm,

4.6 x 150 mm (part number 186006505)

ACQUITY UPLC BEH125 SEC, 1.7 µm,

4.6 x 300 mm (part number 186006506)

ACQUITY UPLC BEH200 SEC, 1.7 µm,

4.6 x 300 mm (part number 186005226)

BioSuite™ 125 UHR SEC, 4 µm,

4.6 x 300 mm (part number 186002161)

Insulin HMWP HPLC, 10 µm, 7.8 x 300 mm

(part number WAT201549)

Data Management

Software: Empower 2 with Auto•Blend Plus™

SAMPLE PREPARATION

The insulin control sample (Sigma, I2643) was reconstituted and diluted

to 4.0 mg/mL in a 0.01 N hydrochloric acid solution. The injectable insulin

product samples were analyzed past expiry.

RESULTS AND DISCUSSION

The focus of this application is an evaluation of the performance of the ACQUITY

UPLC BEH125 SEC, 1.7 µm, 4.6 x 300 mm column (part number 186006506)

under the conditions provided by the USP and EP monographs for the analysis of

HMW protein in therapeutic insulin samples. The acidic mobile phase prescribed by

both of these pharmacopoeial methods is comprised of 0.65 g/L L-arginine, 15%

acetic acid, and 20% (v/v) acetonitrile. This mobile phase provides an assessment

of the levels of covalent HMW present in these preparations while disrupting

non-covalent insulin self-association and column interactions. The column factors

that were evaluated for this method include pore-size, particle diameter, and

column length.

Selected columns, all of which posses a hydrophilic coating to minimize undesired

secondary effects between analyte and particle surface chemistry, were evaluated

based on resolution and HMW quantitation. All columns were configured on

the same ACQUITY UPLC H-Class Bio System using Waters Empower 2 Software.

The resolution results were determined based on the tangent peak widths USP

determination and by peak-to-valley ratio (P/V) as directed by both the EP and

USP monographs as part of the determination of system suitability. The P/V ratio is

obtained by dividing Hp (height above the baseline of the peak due to the covalent

dimer) by Hv (height above the baseline of the lowest point of the curve separating

this peak from the peak due to the monomer). The requirement for this system

suitability parameter is not less than (NLT) 2.0 for a sample containing more than

0.4% HMW in both the EP and USP monographs.

Pore Size

The insulin size-exclusion HMW determination method in the EP monograph

prescribes the use of an SEC pore size of 120Å to 125Å (12 - 12.5 nm), however,

the USP monograph does not specifically state a pore size requirement. The

ACQUITY UPLC BEH125 SEC, 1.7 µm particles have a pore size of 125Å and,

therefore, meet the EP monograph requirements. Figure 1 shows the comparison

between the BEH125 SEC, 1.7 µm particle column and the 200Å pore diameter

BEH200 SEC 1.7 µm particle column in order to demonstrate the importance of

using the appropriate pore diameter for a SEC separation. The principal benefit

that can be realized by selecting the optimal pore-size is increased resolution. The

critical separation shown in these chromatograms is between the HMW species

and the insulin monomer and an 8% improvement in that resolution is observed

for 125Å pore-size particles as compared to the 200Å pore-size particles

for the standard sample. The improvement observed in the two therapeutic

samples, which were analyzed past expiry, is lessened due to the increased

Page 3: Size-Exclusion Ultra Performance Liquid Chromatography for ... · Size-Exclusion Ultra Performance Liquid Chromatography for the Analysis of Covalent High Molecular Weight Insulin

3SEC-MS Using Non-Denaturing Conditions

extent of peptide degradation. In addition to an improvement in resolution, the average P/V determined for the two

therapeutic samples, which had HMW peak area percentages above 0.4%, was higher for the 125Å pore-size particles

(P/V=37) versus the 200Å pore-size particles (P/V=11). However, both columns were able to surpass the US and

EP monograph system suitability criterion of a P/V of NLT 2.0 for a sample containing more than 0.4% HMW.

Additional improvements in the separation between the covalent insulin dimer HMW form and the multimeric

HMW forms, and between the insulin monomer and an insulin fragment, are also observed in the therapeutic

samples analyzed using the 125Å pore-size particles that are not observed using the 200Å pore-size particles,

thereby providing an additional assessment of sample quality.

ACQUITY UPLC BEH125, 1.7 µm (4.6 x 300 mm)

ACQUITY UPLC BEH200, 1.7 µm(4.6 x 300 mm)

Fragment

HMW

ControlRs=3.37

Sample 1Rs=2.63

Sample 2Rs=2.63

ControlRs=3.13

Sample 1Rs=2.38

Sample 2Rs=2.55

4 5 6 7 87 min 9 min

UV A

bsor

banc

e (2

76 n

m)

Figure 1. Zoomed view of insulin HMW, monomer, and low molecular weight fragment SEC separations using a 125Å pore-size ACQUITY UPLC BEH125 column and a 200Å pore-size ACQUITY UPLC BEH200 column. Samples 1 and 2 were therapeutic insulin samples analyzed past expiry.

Particle Size

The insulin SEC HMW determination method in the EP monograph prescribes an SEC particle size of

5 to 10 µm while the USP monograph specifies a particle size of 3 to 10 µm. As part of this study, a

comparison was performed among three SEC columns with particle sizes of 1.7, 4, and 10 µm that are of

comparable 125Å pore size and equivalent 300 mm length (Figure 2). The 1.7-µm column demonstrates

a significant increase (>35%) in resolution as compared to the 4- and 10-µm columns. The average

HMW P/V system suitability criterion for the two therapeutic samples was also higher for the 1.7-µm

column (P/V=37) than for the 4-µm (P/V=9) and 10-µm (P/V=8) columns. All three columns met the

EP and USP HMW P/V system suitability criterion of NLT 2. In addition to increased resolution, the 1.7-µm

column also provides a smaller total elution volume (~5 mL) than the 4-µm (~6 mL) and 10-µm (~13 mL)

columns. This decrease in total elution volume provides the analyst with increased sample throughput in

addition to a decrease in mobile-phase use. Shown in Figure 3 is an expanded base-line view comparison

of the chromatograms obtained from the 1.7- and 10-µm particle size columns that highlights the dramatic

improvements in resolution that are observed for the sub-2-µm UPLC column configuration.

Page 4: Size-Exclusion Ultra Performance Liquid Chromatography for ... · Size-Exclusion Ultra Performance Liquid Chromatography for the Analysis of Covalent High Molecular Weight Insulin

4 Size Exclusion UPLC Analysis of Insulin

Fragment

ACQUITY UPLC BEH125, 1.7 µm (4.6 x 300 mm)

ControlRs=3.37

HMW

ControlRs=2.21

ControlRs=2.08

Sample 1Rs=2.63

Sample 1Rs=1.93

Sample 1Rs=1.95

Sample 2Rs=2.63

Sample 2Rs=1.92

Sample 2Rs=1.88

BioSuite125 UHR, 4 µm(4.6 x 300 mm)

HMWP, 10 µm(7.8 x 300 mm)

4 5 6 9 14 16 18 20877 min 10 min 22 min

UV A

bsor

banc

e (2

76 n

m)

ACQUITY UPLC BEH125(4.6 x 300 mm)

0.83% HMW

0.85% HMW

InsulinFragment

Waters HMWP(7.8 x 300 mm)

0.008 0.40

0.20

0.00

0.00

0.20

0.40

0.008

0.010

0.007

0.006

0.006

0.005

0.004

0.004

0.003

0.002

0.002

0.001

0.000

3.8

12 13 14 15 16 17 18 19 20 21 min

4.0

0 2 4 6 8 10 min

20 min10

4.2 4.4 4.6 4.8 5.0 5.2 5.4 5.6 5.8 6.0 6.4 min6.2

0.000

AUAU

UV A

bsor

banc

e (2

76 n

m)

Figure 2. Zoomed view of insulin HMW, monomer, and low molecular weight fragment SEC separations using 125Å pore-size columns with particle sizes of 1.7, 4 and 10 µm. The sample load volume and flow rate for the 10-µm particle size column were 100 µL and 0.5 mL/minute in accordance with the EP monograph method. Samples 1 and 2 were therapeutic insulin samples analyzed past expiry.

Figure 3. Zoomed view of chromatograms from Figure 2 of a biotherapeutic insulin sample (analyzed past expiry) generated by the ACQUITY UPLC BEH125 and the Waters HMWP columns. The complete chromatographic profiles are presented in the insets. The sample load volume and flow rate for the Waters HMWP column were 100 µL and 0.5 mL/minute in accordance with the EP monograph method. The ACQUITY UPLC BEH column provides improved resolution, improved sensitivity, significantly shorter analysis time, and reduced mobile-phase use.

Page 5: Size-Exclusion Ultra Performance Liquid Chromatography for ... · Size-Exclusion Ultra Performance Liquid Chromatography for the Analysis of Covalent High Molecular Weight Insulin

5Size Exclusion UPLC Analysis of Insulin

Column Length

The effect of column length was also evaluated by comparing the insulin separation on both a 4.6 x 150 mm and

a 4.6 x 300 mm ACQUITY UPLC BEH125 column (Figure 4). Chromatographic principles predict that resolution

should be linearly proportional to the square-root of column length and that is what was observed. The 300 mm

column provided a 41% to 43% increase in resolution as compared to the 150 mm column, consistent with

the predicted increase of 41% (√2). The average HMW peak-to-valley ratio system suitability criterion for the

two therapeutic insulin samples was also higher for the 300 mm column (P/V=37) than for the 150 mm column

(P/V=5). Both columns met the EP and USP HMW peak-to-valley ratio system suitability criterion of NLT 2. The

improved resolution is also apparent in the monomer peak tail, in which a small, lower molecular weight fragment

peak is partially resolved on the 300 mm but not on the 150 mm column. However, it should also be noted that

the improved resolution is accompanied by a two-fold increase in analysis time and mobile-phase use.

Depending on the method requirements, column length can be chosen to either provide improved resolution or

higher sample throughput. For example, for a registered quality test a longer column provides improved resolution

that can result in better quantitative reliability. While in discovery, development, or during real-time process

monitoring, a shorter column allows for faster analysis time and higher sample throughput.

ACQUITY UPLC BEH125, 1.7 µm (4.6 x 300 mm)

ACQUITY UPLC BEH125, 1.7 µm (4.6 x 150 mm)

Fragment

HMW

ControlRs=3.37

Sample 1Rs=2.63

Sample 2Rs=2.63

ControlRs=2.35

Sample 1Rs=1.87

Sample 2Rs=1.86

4 2.0 2.5 3.0 3.5 min5 6 7 min

UV A

bsor

banc

e (2

76 n

m)

Figure 4. A comparison of two ACQUITY UPLC BEH125 columns with lengths of 300 mm and 150 mm. The flow rates for this study were equivalent (0.5 mL/minute). The injection volumes were 10 µL and 5 µL for the 300 mm and 150 mm columns, respectively.

Page 6: Size-Exclusion Ultra Performance Liquid Chromatography for ... · Size-Exclusion Ultra Performance Liquid Chromatography for the Analysis of Covalent High Molecular Weight Insulin

Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com

CONCLUSIONS

Size exclusion chromatography is the USP and EP standard method

for the analysis of covalent HMW insulin in therapeutic preparations.

The chromatographic profiles demonstrating the performance of

this method using SEC columns of different pore size, particle size,

and length have been presented and these data are summarized

in Figure 5. Based on these results the use of 125Å pore size,

sub-2-µm ethylene-bridged hybrid (BEH) silica packing material

and Waters Ultra Performance Liquid Chromatography (UPLC)

instrumentation for this traditional SEC-based analysis provides

significant improvements in resolution compared to traditional

SE-HPLC methods while reducing analysis time and mobile-phase use.

REFERENCES

1. Walsh G. Biopharmaceutical benchmarks 2010. Nat Biotechnol.

2010; 28:917-24.

2. Insulin, Human, European Directorate for the Quality of

Medicines (EDQM), 2001.

3. Insulin, The United States Pharmacopoeia Convention, Inc.,

Rockville, MD, 2002.

ACQUITY U

PLC BEH

125,

1.7 µm, 3

00 mm

ACQUITY U

PLC BEH

125,

1.7 µm, 1

50 mm

HMWP,

10 µm, 3

00 mm

BioSuit

e BEH

125,

4 µm, 3

00 mm

ACQUITY U

PLC BEH

200,

1.7 µm, 3

00 mm

3.0Average Peak/Valley

Average Rs

40

35

30

25

20

15

10

5

0

USP/EP Limit(NLT 2)

Columns

2.5

2.0

1.5 RsP/V

1.0

0.5

0.0

Figure 5. A summary of the performance with respect to HMW peak-to-valley ratio (P/V) and USP resolution (Rs) for the columns evaluated. Shown for reference is the minimum P/V required by the EP and USP monograph methods.

Waters, UPLC, and ACQUITY UPLC are registered trademarks of Waters Corporation. Empower, BioSuite, Auto•Blend Plus, and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.

©2012 Waters Corporation. Produced in the U.S.A.March 2012 720004271EN IH-PDF