silice ulr.pdf
-
Upload
lilibeth-maria-canas-mercado -
Category
Documents
-
view
218 -
download
0
Transcript of silice ulr.pdf
-
8/10/2019 silice ulr.pdf
1/8
Silica
Page 1 of 8
Silica, 8282
How to use instrument-specific information
The Instrument-specific informationtable displays requirements that may vary between
instruments. To use this table, select an instrument then read across to find the corresponding
information required to perform this test.
Silica DOC316.53.01130Heteropoly Blue Method1
1 Adapted fromStandard Methods for the Examination of Water and Wastewater.
Method 8282
ULR(3 to 1000 g/L as SiO2) Pour-Thru Cell
Scope and Application: For testing trace levels of soluble silica in pure and ultrapure water
Test preparation
Table 383 Instrument-specific information
Instrument Pour-Thru Kit Cell Orientation Adapter
DR 6000 LQV175.99.20002 Arrow faces right
DR 5000 LZV479
DR 3900LQV157.99.10002 Align cell flow arrows with arrows on cell
compartment
DR 3800, DR 2800, DR 27005940400 1-inch (round) path aligned with arrow on
the adapter
LZV585 (B)
Before starting the test:
See the user manual for Pour-Thru Module installation instructions.
Clean the Pour-Thru cell and all labware as specified in Labware.
Protect the Pour-Thru Cell from contamination when not in use by inverting a small beaker over the top of the glass funnel.
See Reagent preparationfor instructions on preparing the Amino Acid F Reagent.
The four-minute reaction time in step 10is for samples at 20 C; for samples at 10 C, wait eight minutes; for samples at
30 C, wait two minutes.
The one-minute reaction time in step 12is for samples at 20 C; for samples at 10 C, wait two minutes; for samples at
30 C, wait 30 seconds.
Make sure the PourThru cell is completely seated in the sample cell compartment.
Collect the following items:
Description Quantity
Amino Acid F Reagent Solution 1 mL
Citric Acid F Reagent 1 mL
Molybdate 3 Reagent 1 mL
Cylinder, graduated, 50-mL, poly 1
Flask, Erlenmeyer, 250-mL, PMP, with cap 2
Pipet, TenSette, 0.1 to 1.0 mL with tips 1
http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/http://doc316.53.01000.pdf/ -
8/10/2019 silice ulr.pdf
2/8
Silica
Page 2 of 8
Silica
See Consumables and replacement itemsfor reorder information.
Heteropoly Blue method
1. Select the test.
Insert an adapter if
required (see Instrument-
specific information).
2. Activate the reagent
blank option to account forthe Molybdate 3 reagent
blank. The value is printed
on the reagent blank.
3. Use the numeric
keypad on the instrumentto manually adjust the
reagent blank value.
4. Fill two clean 250-mL
Erlenmeyer flasks tooverflowing with sample.
5. Fill a clean 50-mL
plastic graduated cylinder
with sample from one of
the flasks; then discard the
contents of the cylinder.
Repeat three times.
6. Fill the rinsed cylinder
to the 50-mL mark with
sample from the same
flask. Discard any
remaining sample in the
flask.
7. Pour the contents of
the 50-mL cylinder back
into the original flask.
8. Repeat steps 5
through 7for the second
flask containing sample.
Collect the following items:
Description Quantity
645 Silica ULR
Stored Programs
Start
Reagent Blank
ON
Repeat
Steps 57
-
8/10/2019 silice ulr.pdf
3/8
Silica
Silica
Page 3 of 8
9. Use a TenSettePipet
to add 1.0 mL of
Molybdate 3 Reagent to
each flask. Swirl to mix.
10. Start the instrument
timer.
A four-minute reaction
period will begin.
11. When the timer
expires, add 1.0 mL of
Citric Acid F Reagent to
each flask. Swirl to mix.
12. Start the instrument
timer.
A one-minute reaction
period will begin. The
destruction of possible
phosphate interference
occurs during this period.
13. When the timer
expires, pour the contentsof one flask into the
Pour-Thru Cell.
14. After the flow stops,
ZEROthe instrument.
The display will show:
0 g/L SiO2
15. Add 1.0 mL of Amino
Acid F Reagent to theremaining flask. Swirl to
mix.
A faint blue color will
develop if silica is present.
16. Wait at least 15
seconds, then pour thecontents of the second
flask into the Pour-Thru
Cell.
17. READthe results in
mg/L SiO2.
18. Flush the Pour-Thru
Cell with at least 50 mL of
deionized water
immediately after use.
Heteropoly Blue method (continued)
Zero
Read
-
8/10/2019 silice ulr.pdf
4/8
Silica
Page 4 of 8
Silica
Interferences
Table 384 Interfering substances
Interfering substance Interference level
Color Eliminated by zeroing the instrument with the blank (follow procedure).
Iron Interferes at high levels.
pH (extreme) Adjust pH to less than 7.
Phosphate (PO43) Interferes at levels greater than 50 mg/L PO4
3.
Sulfides Interfere at all levels.
Turbidity Eliminated by zeroing the instrument with the blank (follow procedure).
-
8/10/2019 silice ulr.pdf
5/8
Silica
Silica
Page 5 of 8
Sample collection, preservation and storage
Use only plastic containers with tight-fitting closures. Do not use glass containers; they will
contaminate the sample with silica.
Soak sampling containers with a solution made of one part Molybdate 3 Reagent to 50 parts of
high quality deionized water of low silica concentration. Fill the containers completely and let
stand for several hours. Rinse thoroughly with low-level silica water, drain and close. Repeat
this cleaning periodically.
Allow the sample stream to flow for 12 minutes before collection. Do not adjust the flow
during the sampling period as this may introduce particulates.
Rinse the container well with sample before collecting the portion for analysis.
Analyze samples as soon as possible.
Reagent preparation
Amino Acid F Reagent Solution is available in either 100-mL bottles or a package of 20 unit-dose
ampules. The bottled reagent is stable for up to one year if the bottle is kept closed when not in
use. The ampuled reagent is sealed under argon and is more stable with a shelf life greater than 1
year. Reduced sensitivity at high concentrations (1000 g/L) indicates reagent instability. Checkthe bottled reagent on a routine basis by performing an analysis on a 1-mg/L (1000 g/L) Silica
Standard Solution. If the concentration is less than 950 g/L, use a fresh bottle of Amino Acid F
Reagent Solution.
Prepare larger or smaller volumes of Amino Acid F Reagent by dissolving Amino Acid F Reagent
Powder in Amino Acid F Reagent Solvent at a ratio of 11 grams per 100 mL of reagent solvent.
These reagents are available as the Amino Acid F Reagent Package. This prepared solution has
limited stability; test routinely with the 1-mg/L Silica Standard Solution.
Labware
All containers used in this test must be cleansed thoroughly to remove any traces of silica. Use
plastic containers for all analysis and storage because glass can contaminate the sample withsilica. Small bottles or flasks with screw-type closures work well.
1. Clean containers by normal means (do not use phosphate detergents), then rinse with high
quality deionized water of low-level silica concentration.
2. Soak for 10 minutes with a 1:50 dilution of Molybdate 3 Reagent in low-level silica water.
3. Rinse repeatedly with either low-level silica water or the sample before use. Keep containers
tightly closed when not in use.
4. Fill the Pour-Thru Cell with this same mixture of Molybdate 3 and water and let stand for
several minutes before use.
5. Rinse with low-level silica water.
Cleaning the Pour-Thru Cell
The Pour-Thru Cell may accumulate a buildup of colored products, especially if the reacted
solutions are allowed to stand in the cell for long periods after measurement.
1. Remove the color by rinsing with a 1:5 dilution of ammonium hydroxide and deionized water.
2. Rinse several times with deionized water.
3. Cover the Pour-Thru Cell when it is not in use.
-
8/10/2019 silice ulr.pdf
6/8
Silica
Page 6 of 8
Silica
Accuracy check
Standard additions method (sample spike)
Required for accuracy check:
1-mg/L (1000-g/L) Silica standard
TenSette Pipet and Pipet tips.
250-mL plastic Erlenmeyer flasks (3)
1. After reading test results, leave the sample cell (unspiked sample) in the instrument. Verify the
chemical form.
2. Select standard additions from the instrument menu: OPTIONS>MORE>STANDARD
ADDITION.
3. Accept the default values for standard concentration, sample volume and spike volumes. After
the values are accepted, the unspiked sample reading will appear in the top row. See the user
manual for more information.
4. Prepare three samples. Fill three plastic Erlenmeyer flasks with 50 mL of prepared sample.
5. Use the TenSette Pipet to prepare spiked samples: add 0.2 mL, 0.4 mL and 0.6 mL of the1-mg/L standard to each flask and mix thoroughly.
6. Follow the Heteropoly Blue methodtest procedure for each of the spiked samples, starting
with the 0.2 mL sample spike. Measure each of the spiked samples in the instrument.
7. Select GRAPHto view the results. Select IDEAL LINE (or best-fit) to compare the standard
addition results to the theoretical 100% recovery.
Standard solution methodNote: Refer to the instrument user manual for specific software navigation instructions.
Required for accuracy check:
500-g/L SiO2Standard Solution
1. Use a 500-g/L SiO2Standard Solution in place of the sample. Follow the Heteropoly Blue
methodtest procedure.
2. To adjust the calibration curve using the reading obtained with the standard solution, navigate
to Standard Adjust in the software: OPTIONS>MORE>STANDARD ADJUST.
3. Turn on the Standard Adjust feature and accept the displayed concentration. If an alternate
concentration is used, enter the concentration and adjust the curve to that value.
Method performance
Summary of method
A number of modifications are necessary to adapt the Low Range Silica method for analyzing
trace levels in the Ultra Low Range method. It is absolutely necessary to use the one-inch Pour-
Thru Cell and liquid reagents. The Pour-Thru Cell increases the reproducibility of the optics and
reduces the instability of the readings that result from moveable sample cells. Liquid reagents
produce more reproducible readings and lower blank values by eliminating slight turbidity that may
Program Instrument StandardPrecision95% Confidence
Limits of Distribution
SensitivityConcentration
per 0.010 Abs
645 DR 5000 500 g/L silica 496504 g/L silica 13 g/L silica
-
8/10/2019 silice ulr.pdf
7/8
Silica
Silica
Page 7 of 8
remain when using powdered reagents. Use of liquid reagents in continuous monitors for silica
provides a means of confirming the analyzer performance.
Silica and phosphate in the sample react with molybdate ions under acidic conditions to form
yellow silicomolybdic acid complexes and phosphomolybdic acid complexes. Addition of citric acid
destroys the phosphate complexes. Amino Acid F Reagent is then added to reduce the yellow
silicomolybdic acid to an intense blue color, which is proportional to the silica concentration. Test
results are measured at 815 nm.
Consumables and replacement items
Required reagents
Description Quantity/Test Unit Catalog number
ULR Silica Reagent Set (using Amino Acid F solution, 100 tests) 2553500
Includes: (2) 199532, (2) 2254232, (1) 2386442
ULR Silica Reagent Set (using Amino Acid F ampules, 40 tests) 2581400
Includes: (1) 199532, (1) 2254232, (2) 2386420
Amino Acid F Reagent Solution 1.0 mL 100 mL 2386442OR
Amino Acid F Reagent Solution, 1.2-mL Ampules 1 20/pkg 2386420
Citric Acid Reagent Solution 2 mL 500 mL 2254249
Molybdate 3 Reagent Solution 2.0 mL 500 mL 199549
Required apparatus
Description Quantity Unit Catalog number
Cylinder, graduated, 50-mL, poly 1 each 108141
Flask, Erlenmeyer, 250-mL, PMP, w/cap 2 each 2089846
Pipet, TenSette, 0.1 to 1.0 mL 1 each 1970001
Pipet Tips, for TenSette Pipet 19700-01 5 50/pkg 2185696
Recommended standards
Description Unit Catalog number
Silica Standard Solution, 1-mg/L SiO2 500 mL 110649
Silica Standard Solution, 500-g/L SiO2 3.78 L 2100817
Water, deionized 4 L 27256
http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2553500http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2581400http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2386442http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2386420http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2254249http://www.hach.com/hc/search.product.details.invoker/PackagingCode=199549http://www.hach.com/hc/search.product.details.invoker/PackagingCode=199549http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2254249http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2386420http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2386442http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2581400http://www.hach.com/hc/search.product.details.invoker/PackagingCode=2553500 -
8/10/2019 silice ulr.pdf
8/8
HACH COMPANYWORLD HEADQUARTERSTelephone: (970) 669-3050
FAX: (970) 669-2932
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:Call 800-227-4224
Contact the HACH office or distributor serving you.
www.hach.com [email protected]
In the U.S.A.
Outside the U.S.A.
On the Worldwide Web ; E-mail
toll-free
Hach Company, 2007, 2010, 2012. All rights reserved. Printed in the U.S.A. Edition 7
Optional reagents and apparatus
Description Unit Catalog number
Ammonium Hydroxide, 58% 500 mL 10649
Molybdate 3 Reagent 2.9 L 199503
Molybdate 3 Reagent 3.78 L 199517
Molybdate 3 Reagent 100 mL 199532
Molybdate 3 Reagent 1 L 199553
PourRiteampule breaker each 2484600
Amino Acid F reagent package each 2254117
Sampling Bottle with cap, low density polyethylene, 500 mL 12/pkg 2087079
Thermometer, Non-Mercury, -10 to 225 C each 2635700