SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 CALIBRATION · (A2LA Cert. No. 2357.15) Revised...
Transcript of SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 CALIBRATION · (A2LA Cert. No. 2357.15) Revised...
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 1 of 27
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005
& ANSI/NCSL Z540-1-1994
TEKTRONIX SERVICE SOLUTIONS 24301 Catherine Industrial Road, Suite 116
Novi, MI 48375 Rob Ciolino Phone: 248 305 5200
CALIBRATION
Valid to: December 31, 2017 Certificate Number: 2357.15 In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this laboratory to perform the following calibrations1: I. Chemical Quantities
Parameter/Equipment
Range CMC2 ()
Comments
pH3 Meters
4.01 pH 7.00 pH 10.0 pH
0.049 pH 0.031 pH 0.036 pH
Standard solutions
Conductivity Meters3
(1 to 10) µS (10 to 100) µS (100 to 1000) µS (1 to 20) mS
0.26 µS 1.0 µS 5.4 µS 0.067 mS
Standard solutions
II. Dimensional
Parameter/Equipment
Range CMC2, 5 ()
Comments
Micrometers
(0.05 to 4) in (4 to 12) in (12 to 48) in
(64 + 0.25L) µin (100 + 10L) µin (170 + 19.6L) µin
Grade 2 gage blocks
Calipers3
(0.05 to 4) in (4 to 12) in (12 to 48) in
(64 + 0.25L) µin (90 + 10L) µin (170 + 13L) µin
Grade 2 gage blocks
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 2 of 27
Parameter/Equipment
Range CMC2 ()
Comments
Length Indicators3 Up to 3 in
68 µin
Grade 1 gage blocks
Height Gages, & Depth Gages3
Up to 12 in (12 to 40) in
110 µin 320 µin
Grade 2 gage blocks
Surface Plate Flatness3
Up to 96 in 48 in 71 µin
Rahn 24-inch planekator
Rulers3
Up to 48 in 9000 µin
Gage blocks
Protractors, Inclinometers & Levels3
0 to 150
170 arc seconds
Sine plate and gage blocks
Angle3
1º 3º 5º 15º 30º 45º
0.00080º 0.00082º 0.00082º 0.00080º 0.00080º 0.00082º
Angle blocks
Surface Flatness3
Up to 24 in 4 µin
Mitutoyo Mu checker w/ head and gage blocks
Pins and Plugs3
Up to 1 in
16 µin
Lasermic w/ master pins
Threaded Plugs3 –
Pitch Diameter
Up to 3 in OD
77 µin
SIP machine
Extensometers3
Length/Displacement
Up to 0.5 in 0.5 to 50 in
31 µin 0.080 in
SIP machine red lion controls PAXI
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 3 of 27
III. Electrical – DC/Low Frequency
Parameter/Equipment
Range
CMC2, 4 ()
Comments
DC Voltage – Generate3
Up to 220 mV (0.22 to 2.2) V (2.2 to 11) V (11 to 22) V (22 to 220) V (220 to 1100) V 10 V 1.018 V 1.000 V
7.1 µV/V + 0.5 µV 6.5 µV/V 3.4 µV/V 3.3 µV/V 4.9 µV/V 6.6 µV/V 2.2 µV/V 7.7 µV/V 2.2 µV/V
Fluke 5720A Fluke 5720A Fluke 732A
DC Voltage – Measure3
Up to 200 mV (0.2 to 2) V (2 to 20) V (20 to 200) V (200 to 1050) V (1.1 to 40) kV
6.1 µV/V + 0.12 µV 3.0 µV/V 3.8 µV/V 3.8 µV/V 5.2µV/V 26 V
Fluke 8508A Fluke 80K40
DC Current – Generate3
Clamp-On Only
(0 to 2) pA (2 to 20) pA (20 to 200) pA (0.2 to 2) nA (2 to 20) nA (20 to 200) nA (0.2 to 2) A (0.1 to 220) A (0.22 to 2.2) mA (2.2 to 22) mA (22 to 220) mA (0.22 to 2.2) A (2.2 to 11) A (11 to 20.5) A (16.5 to 149.999) A (150 to 1025) A
1.3 mA/A + 0.40 fA 0.84 mA/A + 4.0 fA 0.59 mA/A + 40 fA 0.34 mA/A + 40 fA 0.34 mA/A + 25 fA 0.34 mA/A + 7.0 fA 0.34 mA/A + 4.0 fA 41 µA/A + 7.0 nA 35 µA/A 34 µA/A 43 µA/A 76 µA/A 0.32 mA/A 0.82 mA/A 0.60 % 0.56 %
Fluke 5720A w/ Keithley 5156 Fluke 5720A w/ 5725A Fluke 5520A Fluke 5520A w/5500A/ Coil
DC Power– Generate3
(0.01 to 330) W (0.33 to 2.9) kW (3.0 to 20.5) kW
0.24 mW/W 0.18 mW/W 0.56 mW/W
Fluke 5520A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 4 of 27
Parameter/Equipment
Range
CMC2, 4 ()
Comments
DC Current – Measure3
(0 to 200) A (0.2 to 20) mA (2 to 20) mA (20 to 200) mA (0.2 to 2) A (2 to 20) A (20 to 300) A (100 to 1000) A
14 µA/A+ 0.4 nA 15 µA/A 16 µA/A 51 µA/A 0.18 mA/A 0.42 mA/A 0.53 mA/A 0.25 %
Fluke 8508A 8508A w/ L&N 4361 shunt
Empro E1500 100
DC Resistance – Generate3
Fixed Points
(0 to 11) (11 to 33) (33 to 110) (110 to 330) (.33 to 1.1) k (1.1 to 3.3) k (3.3 to 11) k (11 to 33) k (33 to 110) k (110 to 330) k (.33 to 1.1) M (1.1 to 3.3) M (3.3 to 11) M (11 to 33) M (33 to 110) M (110 to 330) M (330 to 1100) M 1 k 10 k 100 k 1 M 10M 100 M 1 G 10 G 100 G 10 M 100 M 1000 M
44 µΩ/Ω + 1.0 mΩ 14 µΩ/Ω + 1.5 mΩ 34 µΩ/Ω 32 µΩ/Ω 33 µΩ/Ω 30 µΩ/Ω 26 µΩ/Ω 27 µΩ/Ω 26 µΩ/Ω 36 µΩ/Ω 30 µΩ/Ω 57 µΩ/Ω 0.12 mΩ/Ω 0.30 mΩ/Ω 0.55 mΩ/Ω 3.7 mΩ/Ω 13 mΩ/Ω 14 m 0.19 1.7 22 0.72 k 0.062 M 0.24 M 7.1 M 0.095 G 0.83 k 100 k 1.1 M
Fluke 5520A Biddle megadek PPM R3-1, 110M
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 5 of 27
Parameter/Equipment
Range CMC2,4 ()
Comments
DC Resistance – Generate3 (cont)
Fixed Values
0.01 0.1 0.001 1 1.9 10 19 100 190 1 k 1.9 k 10 k 19 k 100 k 190 k 1 M 1.9 M 10 M 19 M 100 M
0.41µΩ 4.2 µΩ 0.041 µΩ 7.0 µΩ 0.19 m 0.24 m 0.48 m 1.1 m 1.9 m 7.9 m 16 m 37 mΩ 0.15 1.0 2.7 50 60 4.0 k 8.5 k 26 k
L&N 4361 L&N 4360 L&N 4363 Fluke 742A-1 Fluke 5720A Fluke 742A-10K Fluke 5720A
DC Resistance – Measure3
(0 to 2) (2 to 20) (20 to 200) (0.2 to 2) k (2 to 20) k (20 to 200) k (0.2 to 2) M (2 to 20) M (20 to 200) M (0.2 to 2) G
19 µ/ + 5 µ 10 µ/ 8.2 µ/ 8.1 µ/ 8.2 µ/ 8.2 µ/ 11 µ/ 49 µ/ 81 µ/ 0.25 m/
Fluke 8508A
Rise Time – Generate3
< 300 pS 8.4 ps
Fluke 5520A-SC1100
Rise Time – Measure3
<350 pS 8.4 ps
TDS 684A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 6 of 27
Parameter/Equipment
Range CMC2, 4 ()
Comments
Leveled Sinewave Flatness3
(Reference to 50 kHz)
50 kHz to 100 MHz (100 to 300) MHz (300 to 600) MHz (600 to 1100) MHz
18 mV/V + 100 µV 21 mV/V + 100 µV 34 mV/V + 100 µV 42 mV/V + 100 µV
Fluke 5520A-SC1100
AC Voltage – Generate3
(0.2 to 2.2) mV (1 to 22) mV (22 to 220) mV (0.22 to 2.2) V
(10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz
2.2 mV/V + 4.0 µV 1.3 mV/V + 4.0 µV 0.099 % + 4.0 µV 1.7 mV/V + 4.0 µV 2.3 mV/V + 0.0 µV 4.1 mV/V + 10 µV 6.5 mV/V + 20 µV 8.5 mV/V + 20 µV 0.038 % + 5 µV 0.022 %+ 5 µV 0.026 % + 5 µV 0.038 % + 5 µV 0.066 % + 7 µV 1.5 mV/V + 12 µV 1.9 mV/V + 25 µV 3.4 mV/V + 25 µV 0.026 % + 15µV 0.01 % + 8 µV 94 µV/V + 8 µV 0.021 % + 8 µV 0.05 % + 20 µV 0.09 % + 25 µV 1.4 mV/V + 30 µV 2.8 mV/V + 60 µV 0.024 % + 50 µV 90 µV/V + 20 µV 50 µV/V + 10 µV 85 µV/V + 12 µV 0.013 % + 40 µV 0.042 % + 100 µV 0.097 % + 250 µV 1.8 mV/V + 400 µV
Fluke 5720A w/wideband Fluke 5720A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 7 of 27
Parameter/Range Frequency CMC2, 4 ()
Comments
AC Voltage – Generate3 (cont)
(2.2 to 22) V
(22 to 220) V
(200 to 1100) V
(220 to 750) V
(220 to 1100) V
(10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 20 kHz (20 to 50) kHz (50 to 100) kHz (100 to 300) kHz (300 to 500) kHz (0.5 to 1) MHz (15 to 50) Hz 50 Hz to 1 kHz 30 kHz to 50 kHz (50 to 100) kHz 40 Hz to 1 kHz (1 to 20) kHz (20 to 30) kHz
0.024 % + 500 µV 91 µV/V + 200 µV 50 µV/V + 70 µV 87 µV/V + 120 µV 0.013 % + 250 µV 0.032 % + 800 µV 1.0 mV/V + 2.5 mV 1.9 mV/V + 4.0 mV 0.025 % + 5 mV 93 µV/V + 2 mV 63 µV/V + 0.7 mV 0.011 % + 1.2 mV 0.017 % + 3 mV 0.086 % + 20 mV 4.2 mV/V + 50 mV 7.8 mV/V + 100 mV 0.028 % + 20 mV 77 µV/V + 4 mV 0.062 % + 11 mV 2.4 mV/V + 45 mV 98 µV/V + 4 mV 0.017 % + 6 mV 0.062 % + 11 mV
Fluke 5720A Fluke 5720 w/5725A
AC Voltage – Measure3
(0 to 200) mV
(10 to 40) Hz (40 to 100) Hz (0.1 to 2) kHz (2 to 10) kHz (10 to 30) kHz (30 to 100) kHz
0.016 % + 5 µV 0.012 % + 5 µV 0.013 % + 2 µV 0.015 % + 5 µV 0.034 % + 10 µV 0.074 % + 24 µV
Fluke 8508A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 8 of 27
Parameter/Range Frequency CMC2, 4 () Comments
AC Voltage – Measure3
(cont)
(0.2 to 2) V
(2 to 20) V
(20 to 200) V (300 to 1050) V
(10 to 40) Hz (40 to 100) Hz (0.1 to 2) kHz (2 to 10) kHz (10 to 30) kHz (30 to 100) kHz (100 to 300) kHz (0.3 to 1) MHz (10 to 40) Hz (40 to 100) Hz (0.1 to 2) kHz (2 to 10) kHz (10 to 30) kHz (30 to 100) kHz (100 to 300) kHz (0.3 to 1) MHz (10 to 40) Hz (40 to 100) Hz (0.1 to 2) kHz (2 to 10) kHz (10 to 30) kHz (30 to 100) kHz (100 to 300) kHz (0.3 to 1) MHz (10 to 40) Hz (0.04 to 10) kHz (10 to 20) kHz (20 to 100) kHz
0.012 % + 24 µV 0.012 % + 24 µV 80 µV/V + 24 µV 0.012 % + 24 µV 0.023 % + 50 µV 0.057 % + 240 µV 2.6 mV/V + 2.4 mV 8.5 mV/V + 24 mV 0.013 % + 1.4 mV 99 µV/V + 0.24 mV 81 µV/V + 0.24 mV 0.012 % + 0.24 mV 0.022 % + 0.50 mV 0.056 % + 2.4 mV 2.6 mV/V + 24 mV 8.5 mV/V + 240 mV 0.017 % + 2.4 mV 0.011 % + 2.4 mV 90 µV/V + 2.4 mV 0.012 % + 2.4 mV 22 µV/V + 2.4 mV 0.056 % + 5 mV 2.5 nV/V + 240 mV 8.5 nV/V + 2.4 V 0.014 % + 80 mV 12 µV/V + 25 mV 0.023 % + 40 mV 0.059 % + 200 mV
Fluke 8508A
AC Current – Generate3
(9 to 220) A
(10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz
0.026 % + 16 nA 0.017 % + 10 nA 0.013 % + 8 nA 0.029 % + 12 nA 1.2 mA/A + 65 nA
Fluke 5720A w/ 5725A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 9 of 27
Parameter/Range Frequency CMC2, 4 () Comments
AC Current – Generate3 (cont)
(0.22 to 2.2) mA (2.2 to 22) mA
(22 to 220) mA (0.22 to 2.2) A (2.2 to 11) A (11 to 20.5) A Clamp-On Only (16.5 to 149.999) A (16.5 to 149.999) A (150 to 1025) A (150 to 1025) A
(10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (10 to 20) Hz (20 to 40) Hz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz 20 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz 40 Hz to 1 kHz (1 to 5) kHz (5 to 10) kHz (45 to 100) Hz (0.1 to 1) kHz (1 to 5) kHz (45 to 65) Hz (65 to 440) Hz (45 to 65) Hz (65 to 440) Hz
0.026 % + 40 nA 0.017 % + 35 nA 0.013 % + 35 nA 0.021 % + 110 nA 1.2 nA/A + 650 nA 0.026 % + 40 nA 0.017 % + 350 nA 0.013 % + 350 nA 0.021 % + 550 nA 1.5 nA/A + 5.0 µA 0.026 % + 4 µA 0.017 % + 3.5 µA 0.013 % + 2.5 µA 0.021 % + 3.5 µA 1.2 nA/A + 10 µA 0.027 % + 35 µA 0.047 % + 80 µA 7.3 nA/A + 160 µA 0.048 % + 170 µA 0.099 % + 380 µA 3.8 nA/A + 750 µA 1.1 mA/A + 5 mA 1.3 mA/A+ 5 mA 23 mA/A + 5 mA 0.68 % 1.0 % 0.32 % 0.81 %
Fluke 5720A w/ 5725A Fluke 5520A Fluke 5520A w/5500A/ coil Fluke 5520A w/5500A/ coil
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 10 of 27
Parameter/Range Frequency CMC2, 4 () Comments
AC Current – Measure3
Up to 200 A (0.2 to 2) mA
(2 to 20) mA (20 to 200) mA
(0.2 to 2) A
(2 to 20) A
10 Hz to 10 kHz (10 to 30) kHz (30 to 100) kHz 10 Hz to 10 kHz (10 to 30) kHz (30 to 100) kHz 10 Hz to 10 kHz (10 to 30) kHz (30 to 100) kHz 10 Hz to 10 kHz (10 to 30) kHz 10 Hz to 2 kHz (2 to 10) kHz (10 to 30) kHz 10 Hz to 2 kHz (2 to 10) kHz
0.038 % + 24 nA 0.038 % + 24 nA 3.4 nA/A + 24 nA 0.034 % + 0.24 µA 0.074 % + 0.24 µA 3.4 nA/A + 0.24 µA 0.033 % + 2.4 µA 0.069 % + 2.4 µA 4.1 nA/A + 2.4 µA 0.031 % + 24 µA 0.074 % + 24 µA 0.062 % + 0.24 mA 0.075 % + 0.24 mA 2.5 nA/A + 0.24 mA 0.078 % + 2.4 mA 2.1 nA/A + 2.4 mA
Fluke 8508A
AC Power – Generate3
(0.01 to 0.1) W (0.1 to 890) W (0.89 to 3) kW (3 to 11) kW (11 to 20.5) kW
(45 to 65) Hz, PF = 1
1.6 mW/W 1.1 mW/W 0.9 mW/W 0.78 mW/W 0.78 mW/W
Fluke 5520A
AC Level Flatness – Generate3
(0.45, 3, 6) V
10 Hz 100 Hz 10 kHz 30 kHz 100 kHz 300 kHz 1 MHz 3 MHz
5.1 mV/V 2.1 mV/V 2.1 mV/V 2.5 mV/V 3.1 mV/V 6.7 mV/V 4.2 mV/V 9.1 mV/V
Thermal converters EL1100 and 1395A w/ HP 3458A and Fluke 5700A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 11 of 27
Parameter/Range
Frequency CMC2, 4 ()
Comments
AC Level Flatness – Generate3 (cont)
(0.45, 3, 6) V
8 MHz 10 MHz 20 MHz 30 MHz 50 MHz 70 MHz 80 MHz 100 MHz
9.1 mV/V 1.0 mV/V 1.6 mV/V 2.2 mV/V 4.4 mV/V 16 mV/V 17 mV/V 17 mV/V
Thermal converters EL1100 and 1395A w/ HP 3458A and Fluke 5700A
AC Resistance – Generate3
Fixed Values
10 100 1 k 10 k 100 k
DC to 1 MHz (1 to 2) MHz (2 to 3) MHz (3 to 4) MHz (4 to 5) MHz (5 to 10) MHz (10 to 13) MHz DC to 1 MHz (1 to 2) MHz (2 to 3) MHz (3 to 4) MHz (4 to 5) MHz (5 to 10) MHz (10 to 13) MHz DC to 3 MHz (3 to 4) MHz (4 to 5) MHz (5 to 10) MHz (10 to 13) MHz DC to 1 MHz DC to 1 MHz
5.2 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 4.3 mΩ/Ω 5.2 mΩ/Ω 5.9 mΩ/Ω 6.6 mΩ/Ω 2.0 mΩ/Ω 3.0 mΩ/Ω 5.2 mΩ/Ω 5.2 mΩ/Ω
HP 42030 resistor set
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 12 of 27
Parameter/Range Frequency CMC2, 4 ()
Comments
Capacitance – Generate3
Fixed Values 1 pF 10 pF 100 pF 1000 pF (10, 100) nF 1 F
100 Hz to 1 kHz 1 kHz to 1 MHz (1 to 2) MHz (2 to 3) MHz (3 to 4) MHz (4 to 5) MHz (5 to 10) MHz (10 to 13) MHz 100 Hz to 1 kHz 1 kHz to 1 MHz (1 to 2) MHz (2 to 3) MHz (3 to 4) MHz (4 to 5) MHz (5 to 10) MHz (10 to 13) MHz 100 Hz to 1 kHz 1 kHz to 1 MHz (1 to 2) MHz (2 to 3) MHz (3 to 4) MHz (4 to 5) MHz (5 to 10) MHz (10 to 13) MHz 100 Hz to 1 kHz 1 kHz to 1 MHz (1 to 2) MHz (2 to 3) MHz (3 to 4) MHz (4 to 5) MHz (5 to 10) MHz (10 to 13) MHz (100 to 120) Hz 120 Hz to 1 kHz (1 to 10) kHz (10 to 100) kHz (100 to 120) Hz 120 Hz to 1 kHz (1 to 10) kHz (10 to 100) kHz
7.1 µF/F 7.1 µF/F 7.4 µF/F 8.2 µF/F 9.5 µF/F 1.1 µF/F 2.6 µF/F 3.8 µF/F 7.1 µF/F 7.1 µF/F 7.1 µF/F 7.1 µF/F 7.1 µF/F 7.1 µF/F 7.1 µF/F 7.2 µF/F 7.1 µF/F 7.1 µF/F 7.1 µF/F 7.1 µF/F 7.1 µF/F 7.2 µF/F 7.8 µF/F 8.6 µF/F 7.1 µF/F 7.1 µF/F 7.2 µF/F 7.6 µF/F 8.3 µF/F 9.4 µF/F 2.0 µF/F 2.9 µF/F 1.9 µF/F 1.9 µF/F 1.9 µF/F 1.9 µF/F 1.9 µF/F 1.9 µF/F 1.9 µF/F 2.0 µF/F
HP 1638XX standard capacitor
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 13 of 27
Parameter/Range Frequency CMC2, 4 ()
Comments
Capacitance – Generate3 (cont)
(0.19 to 1.09) nF (1.1 to 3.29) nF (3.3 to 10.9) nF (11 to 109.9) nF (110 to 329.9) nF (0.33 to 1.09) F (1.1 to 3.29) F (3.29 to 10.9) F (11 to 32.9) F (33 to 109.9) F (110 to 329.9) F (0.33 to 1.09) mF
10 Hz to 10 kHz 10 Hz to 3 kHz 10 Hz to 1 kHz 10 Hz to 1 kHz 10 Hz to 1 kHz (10 to 600) Hz (10 to 300) Hz (10 to 150) Hz (10 to 120) Hz (10 to 80) Hz (10 to 50) Hz (10 to 20) Hz
4.4 mF/F + 0.01 nF 4.1 mF/F + 0.01 nF 2.1 mF/F + 0.01 nF 2.1 mF/F + 0.1 nF 2.1 mF/F + 0.3 nF 2.1 mF/F + 1nF 2.1 mF/F + 3 nF 2.1 mF/F + 10 nF 3.2 mF/F + 30 nF 3.7 mF/F + 100 nF 3.7 mF/F + 300 nF 3.7 mF/F + 1 µF
Fluke 5520A
Capacitance – Measure3
(10 to 100) pF 100 pF to 10 F 10 F to 10 mF
(0.1 to 100) kHz
3.4 mF/F 1.8 mF/F 2.0 mF/F
HP 4284A
Inductance – Generate3
(1 to 100) mH 1 H 10 H
(0.1 to 100) kHz
0.026 % 0.053 % 2.0 mH/H
GenRad 1482 Series
Inductance – Measure3
1 H to 1 H (1 to 10) H
(0.1 to 100) kHz
1.8 mH/H 3.8 mH/H
HP 4284A
Distortion – Measure3
20 Hz to 100 kHz
20 Hz to 20 kHz (20 to 100) kHz
1.2 dB of reading 2.3 dB of reading
HP 8903B
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 14 of 27
Parameter/Range Range CMC2 ()
Comments
Electrical Calibration of Thermocouple Indicators3 –
Type B Type C Type E Type J
Type K Type N Type R
600 C to 800 C 800 C to 1000 C 1000 C to 1550 C 1550 C to 1820 C 0 C to 150 C 150 C to 650 C 650 C to 1000 C 1000 C to 1800 C 1800 C to 2316 C -250 C to -100 C -100 C to -25 C -25 C to 350 C 350 C to 650 C 650 C to 1000 C -210 C to -100 C -100 C to -30 C -30 C to 150 C 150 C to 760 C 760 C to 1200 C -200 C to -100 C -100 C to -25 C -25 C to 120 C 120 C to 1000 C 1000 C to 1372 C -200 C to -100 C -100 C to -25 C -25 C to 120 C 120 C to 410 C 410 C to 1300 C 0 C to 250 C 250 C to 400 C 400 C to 1000 C 1000 C to 1767 C
0.34° C 0.27° C 0.24° C 0.26° C 0.24° C 0.21° C 0.25° C 0.39° C 0.65° C 0.39° C 0.13° C 0.11° C 0.13° C 0.17° C 0.25° C 0.13° C 0.12° C 0.14° C 0.18° C 0.26° C 0.14° C 0.13° C 0.20° C 0.31° C 0.31° C 0.17° C 0.15° C 0.14° C 0.21° C 0.45° C 0.66° C 0.26° C 0.32° C
Fluke 5520A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 15 of 27
Parameter/Equipment
Range CMC2 ()
Comments
Electrical Calibration of Thermocouple Indicators3 – (cont)
Type S Type T Type U
0 C to 250 C 250 C to 1000 C 1000 C to 1400 C 1400 C to 1767 C -250 C to -150 C -150 C to 0 C 0 C to 120 C 120 C to 400 C -200 C to 0 C 0 C to 600 C
0.38° C 0.28° C 0.29° C 0.36° C 0.50° C 0.19° C 0.13° C 0.11° C 0.44° C 0.21° C
Fluke 5520A
Electrical Calibration of RTDs3 –
Pt 385, 100 Pt 3926, 100 Pt 3916, 100
-200 C to -80 C -80 C to 0 C 0 C to 100 C 100 C to 300 C 300 C to 400 C 400 C to 630 C 630 C to 800 C -200 C to -80 C -80 C to 0 C 0 C to 100 C 100 C to 300 C 300 C to 400 C 400 C to 630 C -200 C to -190 C -190 C to -80 C -80 C to 0 C 0 C to 100 C 100 C to 260 C 260 C to 300 C 300 C to 400 C 400 C to 600 C 600 C to 630 C
0.09° C 0.12° C 0.12° C 0.11° C 0.10° C 0.11° C 0.20° C 0.06° C 0.08° C 0.08° C 0.09° C 0.10° C 0.11° C 0.21° C 0.06° C 0.07° C 0.07° C 0.08° C 0.08° C 0.09° C 0.13° C 0.19° C
Fluke 5520A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 16 of 27
Parameter/Equipment
Range CMC2 ()
Comments
Electrical Calibration of RTDs3 (cont) –
Pt 385, 200 Pt 385, 500
Pt 385, 1000 PtNi 385, 120 Cu 427, 10
-200 C to -80 C -80 C to 0 C 0 C to 100 C 100 C to 260 C 260 C to 300 C 300 C to 400 C 400 C to 600 C 600 C to 630 C -200 C to -80 C -80 C to 0 C 0 C to 100 C 100 C to 260 C 260 C to 300 C 300 C to 400 C 400 C to 600 C 600 C to 630 C -200 C to -80 C -80 C to 0 C 0 C to 100 C 100 C to 260 C 260 C to 300 C 300 C to 400 C 400 C to 600 C 600 C to 630 C -80 C to 0 C 0 C to 100 C 100 C to 260 C -100 C to 260 C
0.04° C 0.04° C 0.04° C 0.05° C 0.11° C 0.11° C 0.11° C 0.13° C 0.04° C 0.04° C 0.04° C 0.05° C 0.07° C 0.07° C 0.07° C 0.09° C 0.03° C 0.03° C 0.04° C 0.04° C 0.05° C 0.18° C 0.06° C 0.18° C 0.08° C 0.11° C 0.11° C 0.69° C
Fluke 5520A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 17 of 27
Parameter/Equipment
Frequency CMC2, 4 ()
Comments
Phase – Measure3
-180o to +360o 0.03º
HP 53131A
Phase – Generate3, 0 to 180º
(10 to 65) Hz (65 to 500) Hz (0.5 to 1) kHz (1 to 5) kHz (5 to 10) kHz (10 to 30) kHz
0.14º 0.28º 0.52º 3.0º 5.4º 11º
Fluke 5520A
Electrical Calibration of RPM – Measure3
(0 to 100 000) RPM
0.35 RPM
HP 53131A
Oscilloscope –
Amplitude – DC3 50 Ω 1 MΩ
Amplitude – Square Wave3
50 Ω
1 MΩ
Time Marker3 Rise Time3 Bandwidth3 Bandwidth w/ Power Sensor
(-6.6 to 6.6) V (-130 to 130) V ±1 mV to ±6.6 V(p-p) 10 Hz to 10 kHz ±1 mV to ±6.6 V(p-p) 10 Hz to 1 kHz 5 s to 50 ms 20 ms to 1 ns >300 ps 50 kHz to 100 MHz (100 to 300) MHz (300 to 600) MHz (600 to 1100) MHz (1.1 to 18) GHz
1.9 mV/V + 31 µV 0.39 mV/V + 31 µV 2.0 mV/V + 31 µV 0.78 mV/V + 31 µV (19 + 39t) µs/s 2.1 µs/s 16 ps 2.8 % 3.0 % 4.1 % 4.7 % 7.5 %
Fluke 5520A/SC1100 t is time in seconds
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 18 of 27
VI. Electrical – RF/Microwave
Parameter/Equipment
Frequency
CMC2 ()
Comments
LISN –
Insertion Loss
Impedance – Magnitude
Impedance – Phase
150 kHz to 100 MHz (100 to 400) MHz 150 kHz to 100 MHz (100 to 400) MHz 150 kHz to 200 MHz (200 to 400) MHz
0.86 dB 2.2 dB 1.4 Ω 2.5 Ω 6.2 º 9.2 º
CISPR 25, C63.4-2003,CISPR 16-1-2, HP 4195A
VSWR – Measure
50 Ohms SWR < -3.5 dBm SWR > -3.5 dBm 0 < ρ ≤ 0.0476 0.0476 < ρ ≤ 0.1111 0.1111 < ρ ≤ 0.200 0.200 < ρ ≤ 0.3333 0.3333 < ρ ≤ 0.500
(0.1 to 2) GHz (0.1 to 2) GHz (0.1 to 2) GHz (0.1 to 2) GHz 50 MHz to 3 GHz (4 to 8) GHz (9 to 14) GHz (15 to 20) GHz (0.05 to 2) GHz (2 to 8) GHz (8 to 20) GHz (0.05 to 2) GHz (2 to 8) GHz (8 to 20) GHz (0.05 to 2) GHz (2 to 8) GHz (8 to 20) GHz (0.05 to 2) GHz (2 to 8) GHz (8 to 20) GHz
0.016 Γ 0.017 Γ 0.042 Γ 0.043 Γ 0.00077 ρ 0.0011 ρ 0.0022 ρ 0.0024 ρ 0.0040 ρ 0.0054 ρ 0.0078 ρ 0.0039 ρ 0.0043 ρ 0.0082 ρ 0.0086 ρ 0.0095 ρ 0.014 ρ 0.0086 ρ 0.0095 ρ 0.014 ρ
Wiltron 60A50 w/Agilent 8657A, HP8563E HP 8720C with 85052B
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 19 of 27
Parameter/Equipment
Frequency
CMC2 ()
Comments
Reflection Magnitude
(Into 50 Ohm) 0 < Γ ≤ 0.2 0.2 < Γ ≤ 0.4 0.4 < Γ ≤ 0.6 0.6 < Γ ≤ 0.8
0.8 < Γ ≤ 1.0
(0.05 to 20) GHz (0.05 to 2.0) GHz (2.0 to 20) GHz (0.05 to 2.0) GHz (2.0 to 20.) GHz (0.05 to 2.0) GHz (2.0 to 20.) GHz (0.05 to 2.0) GHz (2.0 to 20.) GHz
0.01Γ 0.01Γ 0.02Γ 0.01Γ 0.03Γ 0.02Γ 0.04Γ 0.03Γ 0.05Γ
HP 8720C
Transmission Magnitude (Into 50 Ohms)
(10 to 0) dB (0 to -10) dB (-10 to -20) dB (-20 to -30) dB (-30 to -40) dB (-40 to -50) dB (-50 to -60) dB (-60 to -70) dB
(10 to 0) dB (0 to -10) dB (-10 to -20) dB (-20 to -30) dB (-30 to -40) dB (-40 to -50) dB (-50 to -60) dB (-60 to -70) dB
50 MHz to 0.5 GHz (0.5 to 2.0) GHz
0.04 dB 0.04 dB 0.05 dB 0.07 dB 0.17 dB 0.47 dB 1.4 dB 3.5 dB 0.12 dB 0.05 dB 0.06 dB 0.06 dB 0.06 dB 0.07 dB 0.11 dB 0.25 dB
HP 8720C HP 8720C
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 20 of 27
Parameter/Equipment
Frequency CMC2 ()
Comments
Transmission Magnitude (Into 50 Ohms) (cont)
(10 to 0) dB (0 to -10) dB (-10 to -20) dB (-20 to -30) dB (-30 to -40) dB (-40 to -50) dB (-50 to -60) dB (-60 to -70) dB (10 to 0) dB (0 to -10) dB (-10 to -20) dB (-20 to -30) dB (-30 to -40) dB (-40 to -50) dB (-50 to -60) dB (-60 to -70) dB
(2.0 to 8.0) GHz (8.0 to 20) GHz
0.21 dB 0.11 dB 0.11 dB 0.11 dB 0.12 dB 0.12 dB 0.18 dB 0.32 dB 0.70 dB 0.21 dB 0.21 dB 0.21 dB 0.22 dB 0.22 dB 0.25 dB 0.37 dB
HP 8720C HP 8720C
Reflection Phase (Into 50 Ω, 0° to 360°)
0.0< Γ ≤ 0.2 0.2< Γ ≤ 0.4 0.0< Γ ≤ 0.2 0.2< Γ ≤ 0.4
(0.05 to 2.0) GHz (2.0 to 20) GHz
1.8º 1.3º 3.5º 3.0º
HP 8720C
Transmission Phase (Into 50 Ω, 0° to 360°)
(10 to 0) dB (0 to -10) dB (-10 to -20) dB (-20 to -30) dB (-30 to -40) dB (-40 to -50) dB (-50 to -60) dB (-60 to -70) dB (10 to 0) dB (0 to -10) dB (-10 to -20) dB (-20 to -30) dB (-30 to -40) dB (-40 to -50) dB
50 MHz to 0.5 GHz (0.5 to 2.0) GHz
0.31 dB 0.24 dB 0.28 dB 0.46 dB 1.1 dB 3.5 dB 11 dB 35 dB 0.59 dB 0.34 dB 0.37 dB 0.37 dB 0.37 dB 0.46 dB
HP 8720C HP 8720C
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 21 of 27
Parameter/Range
Frequency CMC2 () Comments
Transmission Phase (Into 50 Ω, 0° to 360°) (cont)
(-50 to -60) dB (-60 to -70) dB
(10 to 0) dB (0 to -10) dB (-10 to -20) dB (-20 to -30) dB (-30 to -40) dB (-40 to -50) dB (-50 to -60) dB (-60 to -70) dB (10 to 0) dB (0 to -10) dB (-10 to -20) dB (-20 to -30) dB (-30 to -40) dB (-40 to -50) dB (-50 to -60) dB (-60 to -70) dB
(0.5 to 2.0) GHz (2.0 to 8.0) GHz (8.0 to 20) GHz
0.72 dB 1.8 dB 1.3 dB 0.69 dB 0.69 dB 0.69 dB 0.69 dB 0.78 dB 1.0 dB 2.1 dB 5.9 dB 1.2 dB 1.3 dB 1.3 dB 1.3 dB 1.40 dB 1.8 dB 2.4 dB
HP 8720C HP 8720C
RF Power – Generate3
(+24 to -56) dBm (+13 to -127) dBm
0.01 Hz to 20 MHz (0.02 to 26.5) GHz
0.68 dB 0.19 dB
HP 3325A HP 83630B w/11792A on 8902A
RF Power – Measure3
(+30 to -20) dBm (-30 to +20) dBm
100 kHz to 2 GHz (2 to 26.5) GHz 50 MHz to 26.5 GHz
0.071 dBm 0.076 dBm 0.15 dBm
HP 8902A w/11722A HP 8902A w/ 11722A and 11792A HP 8485A w/ 438A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 22 of 27
Parameter/Range
Frequency CMC2 () Comments
RF Attenuation – Tuned RF Power Measure3
(0 to -10) dB (-10 to -40) dB (-40 to -50) dB (-50 to -80) dB (-80 to -90) dB (-90 to -110) dB (-110 to -120) dB
100 kHz to 26.5 GHz
0.026 dB 0.064 dB 0.16 dB 0.20 dB 0.32 dB 0.36 dB 0.54 dB
HP 8902A HP 11722A HP 11792A HP 11793A
Phase Modulation – Measure3
Rate: 200 Hz to 10 kHz Rate: 200 Hz to 20 kHz
Rate: 200 Hz to 20 kHz
150 kHz to 10 MHz 10 MHz to 1.3 GHz (1.3 to 26.5) GHz
5.1 % 4.3 % 4.3 %
HP 8902A w/11722A HP 8902A w/ 11793A
Amplitude Modulation – Measure3
Rate: 50 Hz to 10 kHz Depths: 5 % to 99 % Rate: 20 Hz to 10 kHz Depths: to 99 % Rate: 50 Hz to 50 kHz Depths: 5 % to 99 % Rate: 20 Hz to 100 kHz Depths: to 99 % Rate: 20 Hz to 100 kHz Depths: to 99 % Rate: 20 Hz to 100 kHz Depths: to 99 %
150 kHz to 10 MHz 150 kHz to 10 MHz 10 MHz to 1.3 GHz 10 MHz to 1.3 GHz (1.3 to 26.5) GHz (1.3 to 26.5) GHz
2.5 % 3.8 % 1.3 % 3.8 % 1.3 % 3.4 %
HP 8902A w/11722A
w/ 11793A
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 23 of 27
Parameter/Range
Frequency CMC2 () Comments
Frequency Modulation – Measure3
Rate: 20 Hz to 10 kHz Dev.: ≤ 40 kHz peak Rate: 50 Hz to 100 kHz Dev.: ≤ 400 kHz peak Rate: 20 Hz to 200 kHz Dev.: ≤ 400 kHz peak
Rate: 50 Hz to 100 kHz Dev.: ≤ 400 kHz peak Rate: 20 Hz to 200 kHz Dev.: ≤ 400 kHz peak
250 kHz to 10 MHz 10 MHz to 1.3 GHz 10 MHz to 1.3 GHz (1.3 to 26.5) GHz (1.3 to 26.5) GHz
2.5 % 1.3 % 6.2 % 1.3 % 6.2 %
HP 8902A w/11722A
w/ 11793A
V. Mechanical
Parameter/Equipment
Range CMC2 ()
Comments
Mass
(0 to 100) g 100 g to 3 kg (3 to 4.8) kg (4.8 to 24) kg
0.22 mg 20 mg 90 mg 0.58 g
Mettler AE100 w/ Class S weights Mettler PJ3000 w/ Class 1 weights Mettler SB24001 w/ Class 1 and Class F weights
Scales & Balances3
(1 to 5) mg (5 to 10) mg (10 to 20) mg (20 to 50) mg (50 to 100) mg (100 to 200) mg (200 to 500) mg 500 mg to 1 g (1 to 2) g (2 to 5) g (5 to 10) g (10 to 20) g (20 to 50) g
6.4 µg 6.4 µg 6.4 µg 6.4 µg 6.4 µg 6.4 µg 6.4 µg 22 µg 20 µg 40 µg 0.018 mg 0.048 mg 0.048 mg
Class S1 weights
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 24 of 27
Parameter/Equipment
Range CMC2 ()
Comments
Scales & Balances3
(cont)
(50 to 100) g (100 to 200) g (200 to 500) g (500 to 1000) g (1 to 2) kg (2 to 5) kg (5 to 10) kg (10 to 20) kg (0.5 to 10) lb (10 to 25) lb (25 to 50) lb (50 to 100) lb (100 to 500) lb
0.090 mg 0.20 mg 0.21 mg 0.18 mg 1 mg 5.0 mg 7.7 mg 72 mg 0.0013 lb 0.0029 lb 0.058 lb 0.058 lb 0.061 lb
Class S1 weights Class F weights
Durometer Calibration3 – A, B, C, and D
Indentor extension and shape- Diameter Radius Angle Extension Indentor Display Spring Calibration- Force
0 to 100 duro 0.77 duro
Rex RDC-1 Durocalibrator Optical inspection
Torque – Measuring Equipment
100 in·ozf to 100 in·lbf100 in·lbf to 250 ft·lbf
0.00059 in·lbf/in·lbf 0.00030 ft·lbf/ft·lbf
CDI 2000-151-0 CDI 2000-152-0
Torque – Measure
(20 to 200) in·ozf 200 in·ozf to 250 ft·lbf (250 to 2000) ft·lbf
0.0081 in·ozf/in·ozf 0.0031 ft·lbf/ft·lbf 0.0013 ft·lbf/ft·lbf
CDI 2001 0-TTP w/ 950DT CDI TTPM-41 w/ 950DT AKO, Inc. TSD 2050 torque system
Pressure3
(-14.5 to 0) psig (0 to 100) psig (100 to 200) psig (200 to 500) psig (501 to 10 000) psig
0.012 psi 0.061 psi 0.068 psi 0.14 psi 2.1 psi
DPI 620 calibrator w/GE pressure sensor Pressurements DPT W2200/3
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 25 of 27
Parameter/Equipment
Range CMC2, 5 ()
Comments
Pressure3 (cont)
(0 to 0.25) inH2O (0.25 to 2) inH2O (2 to 10) inH2O
0.0017 inH2O 0.0029 inH2O 0.014 inH2O
Heisse PTE-1 w/ HQS-1
Force3 – Tension and Compression3 Load Cells, Force Gages, Dynamometers
(0 to 500) lbf (501 to 5000) lbf (5001 to 50 000) lbf (50 001 to 100 000) lbf (100 001 to 200 000) lbf (0.125 to 1) lbf (1 to 10) lbf (10 to 50) lbf (50 to 100) lbf (100 to 500) lbf
0.039 % 0.063 % 0.056 % 0.068 % 0.074 % 0.019 % 0.014 % 0.013 % 0.018 % 0.0050 %
Interface 1600 Gold Series Class F Weights
Speed3
Up to 50 in/min 0.081 in/min
Stopwatch/ Red Lion Controls PAXI
Vibration
0.5 ≤ f < 1 1 ≤ f < 2 2 ≤ f < 5 5 ≤ f < 10 10 Hz to 99 Hz 100 Hz 100 Hz to 920 Hz 921 Hz to 5000 Hz 5001 Hz to 10 000 Hz 10 001 Hz to 15 000 Hz 15 001 Hz to 20 000 Hz
4.0 % 1.5 % 1.0 % 1.0 % 1.7 % 1.3 % 1.4 % 1.7 % 2.2 % 3.7 % 4.8 %
Modal Shop 9155
Impact Hammer
0 to 5000 lbs 2.8 % Model 9155C Hammer
Calibration System
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 26 of 27
Parameter/Equipment
Range CMC2, 5 ()
Comments
Shock
Up to 10 000 g 2.0 % Model 9525 Pneumatic
Shock Exciter
VI. Thermodynamics
Parameter/Equipment
Range CMC2 () Comments
Temperature – Measure3
-40 C to 400 C 400 C to 500 C
0.077 ºC 0.22 ºC
Hart 5618B PRT w/ 1502
IR Temperature – Measuring Equipment3
50C to 100 C 100 C to 200 C 200 C to 300 C 300 C to 400 C 400 C to 500 C
0.60 C 0.61 C 0.77 C 0.79 C 0.99 C
Hart 9132
Relative Humidity – Measure3
5 % RH to 95 % RH
1.3 % RH
Vaisala HM70 w/probe HMP46
VII. Time& Frequency
Parameter/Equipment
Range CMC2 () Comments
Frequency – Measuring Equipment3
DC to 20 MHz 10 MHz to 26.5 GHz 10 MHz
0.00058 parts in 106 0.29 parts in 106 5.1 parts in 1012
HP 3325B HP 83630B GPS
Frequency – Measure3
DC to 3 GHz 10 Hz to 26.5 GHz
0.00026 parts in 106 0.22 parts in 106
HP 53131A HP 5351B
Time Interval3
1 s to 3 hr 0.59 s
53131A w/GPS NIST SP960-12
____________________________________________
(A2LA Cert. No. 2357.15) Revised 01/28/2016 Page 27 of 27
1 This laboratory offers commercial and field calibration service. 2 Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMC’s represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
3 Field calibration service is available for this calibration and this laboratory meets A2LA R104 – General Requirements: Accreditation of Field Testing and Field Calibration Laboratories for these calibrations. Please note the actual measurement uncertainties achievable on a customer's site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration and for other possible adverse effects such as those caused by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC.
4 The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured value in the ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a percent or fraction of the reading plus a fixed floor specification.
5 In the statement of CMC, L is the numerical value of the nominal length of the device measured in inches.
In the statement of CMC, the value is defined as the percentage of reading, unless otherwise noted.
For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.
Accredited Laboratory
A2LA has accredited
TEKTRONIX, INC Novi, MI
for technical competence in the field of
Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005 General requirements for the competence of testing and calibration laboratories. This laboratory also meets the requirements of
ANSI/NCSLI Z540-1-1994 and any additional program requirements in the field of calibration. This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (refer to joint ISO-ILAC-
IAF Communiqué dated 8 January 2009).
Presented this 20th day of January 2016. _______________________ Senior Director, Quality and Communications For the Accreditation Council Certificate Number 2357.15 Valid to December 31, 2017