Seite 1 © 2015 Schön Klinik
Clinical experience with more than 4.800 CoC
P. Dufek
Schön Klinik Neustadt, Lehrkrankenhaus
der Uni S-H, Campus Lübeck, Germany
Praha 2015
Seite 2 © 2015 Schön Klinik
Stable fixation- cementless, new technics of cementing
Minimal wear of contact surfaces-
highly crosslinked PE, ceramics
Tribology, the basic investigation
Long term results of THR
Seite 4 © 2015 Schön Klinik
Young oldies - baby boomer - problem of PE wear, ostelysis
Seite 5 © 2015 Schön Klinik
0
0.02
0.04
0.06
0.08
0.1
0.12
0.14
Me/PE20a
Ce/PE20a
Me/XPE10a
Ce/XPE5a
Me/Me15a
Ce/Ce15a
0.137
0.072
0.042
0.031
0.005 < 0.001
We
ar
Ra
te [
mm
/a]
Clinical Wear Rates
20a – observation for 20 years; benign reaction of ceramic
particles (PE – inflammatory, Me – toxic, allergic)
Seite 6 © 2015 Schön Klinik
Corrosion: ceramic 10x less; no Co, Cr & Ni release
Co/Cr/Ni
Seite 7 © 2015 Schön Klinik
Why I use ceramics
Discussion:
medical - economic management
Prices of the implants
Best alternative for the patients
in endoprosthetics must be used!!
Effectively lower cost, more profit, good for the patients!
Seite 8 © 2015 Schön Klinik
Silony Medical Int.
own company for endoprothesis
MicroPort Profemur Preserve
BIOLOX®delta (CeramTec)-
use in our clinic since 2005
Seite 9 © 2015 Schön Klinik
Ce/Ce 36-36 ....... 2.914
Ce/Ce 32-32 ....... 1.896
Combination CoP 32-32 …… n: 3.950
Ceramic-on-ceramic bearing surface
Inlay 36 - head 36
Inlay 32 - head 32 (acc. cup-size)
Material Orthopedics Neustadt
Jan 2005- April 2015
Combination CoC Biolox®delta ...... n: 4.810
Seite 10 © 2015 Schön Klinik
Use of CoC Orthop/ Traumatology NEU since 2010
2010 4 354 358
2011 1 593 594
2012 6 747 753
2013 5 842 847
2014 3 973 976
2015 5 401 406
Total 24 3910 3.934
Year Traumatology Orthop
Seite 11 © 2015 Schön Klinik
Advantage -
Good stability and ROM of the hip after dysplastic-arthrosis
CoC 36 mm
Seite 12 © 2015 Schön Klinik
Complications
- Fractures - head & insert 1x
- chipping of the rim of insert 2x
- Subluxation of insert (implantation failure) 2x
- Squeaking - benign 7x
Head fracture - periop. status
Patient 140 kg, fell down the staircase
Problem S-ROM Cone
Seite 13 © 2015 Schön Klinik
Follow-up
Range of motion ROM , dislocation
2002-2004 cup Duraloc, head 28 mm
2004- today cup PINNACLE, head 32 – 36 mm
Material, Method
Retrospective study of ROM of heads 28, 32 &
36 mm based on the medical report- discharge from
REHAB
Each group of 100 patients, btw. 50-70y
Statistics of luxation 6 months after operation acc.
to the med. reports from orth. colleagues in the praxis
Seite 14 © 2015 Schön Klinik
ROM
head 36 mm 15 dg. more than head 32
and 20 dg. more than head 28
Luxation- 6 M after op/ pro y (ca. 1500 TEPs pro y)
Head 28 mm MoP 0,9 %
Head 32 mm MoP 0,17%
Head 32 mm CoC
Head 36 mm CoC till today no luxation
Best combination CoC
Follow-up
results
Seite 15 © 2015 Schön Klinik
• 126c bilateral 32 mm
– CoC 6x significantly less dislocation (p < 0.001)
• Meta-analysis
– CoC has significantly decreased dislocation rate in comparison to MoP (1.2% versus 5.0%, p < 0.01)
Hernigou, 2013
Hu, 2015
Less dislocation with CoC
Seite 16 © 2015 Schön Klinik
Revisions for Infection (500’749 pats)
Streicher, 2014
Seite 17 © 2015 Schön Klinik ©
CoC in Early Infection therapy with PE
2005-2014 n:47 cases
Prospective study- 2 recidive cases
OP revision
Debridement of the joint
Exchange of PE liner to Cer insert mit Delta Option
Head , in the case of C/C comb. Only debridement
Lavage with antiseptics
Local application of GENTA- COLL resorb
Antibiotics therapy till CRP normalisation
Best solution for the infection-revision?
More studies necessary !
Seite 18 © 2015 Schön Klinik
One week after THR with CoC & fast track rehabilitation
Seite 19 © 2015 Schön Klinik
Discussion C on C (delta)
- Negligible wear, benign particles
- No influence of head size on wear
- No corrosion and metal release (Co, Cr, Ni)
- Only 1 head breakage (0,011%; forte 0,1%) - Fracture rate: austral. register delta = 0.0017% (all)
- Low incidence of rumors - benign
- Lower incidence of PJI
- Lower dislocation risk
- Low revision rate even at long-term
Take home message
Biolox®delta articulation with a ceramic head of 32 and 36mm
are very stabile, have a better range of motion and the risk of
long-term wear is basically zero compared to other
articulations
Seite 20 © 2015 Schön Klinik
Thank you and you are welcome in Neustadt in Holstein