RRLC PPT
-
Upload
polisetty-guptha -
Category
Documents
-
view
258 -
download
5
Transcript of RRLC PPT
The underlying principles of this evolution are governed by the van Deemter equation, which is an empirical formula that describes the relationship between linear velocity (flow rate) and plate height (HETP or 1/column efficiency).
H=A+B/u+Cu
Since particle size is one of the variables, a van Deemter curve can be used to investigate chromatographic performance. According to the van Deemter equation, as the particle size decreases to less than 2.5 µm, there is a significant gain in efficiency.
HPLC Instrumentation and Techniques
Instrumentation
Mobile phase reservoirsConstruction pump material
PumpsRecriprocating
SyringeManual Injection
Automated Injection
Packings
Measure column performance
Column care and use
SolventDelivery
SampleIntroduction
Column Packingsand harware
Detectors
UV-Vis
Fluorescence
Refractive Index
ConductivityVolatametry/Amperometry
COLUMNS IN UPLC
ACQUITY UPLC™ BEH C18 and C8 columns were designed to be the universal columns of choice for most UPLC™ separations by providing the widest pH range
ACQUITY UPLC™ BEH C18 and C8 columns incorporate trifunctional ligand bonding chemistries which produce superior low pH stability and ultra-low column bleed.
This low pH stability is combined with the high pH stability of the 1.7 μm BEH particle to deliver the widest usable pH operating range.
These bonding chemistries and particle synthesis innovations produce the sharpest peaks, highest efficiencies.
ACQUITY UPLC™ BEH Shield RP18 Columns
Embedded polar group (Shield) RP columns contain stationary phases that combine the hydrophobicity of an alkyl ligand with the hydrophilicity of an embedded polar group.
Embedded polar group columns include alternate selectivity to that of alkyl reversed-phase columns, excellent peak shape for bases and aqueous mobile phase compatibility.
The unique selectivity of Shield RP phases, especially for polyphenolic compounds, has been attributed to the embedded polar groups acting as hydrogen bond acceptors.
ACQUITY UPLC™BEH Shield RP18 columns are designed to provide selectivities that complement the ACQUITY UPLC™ BEH C18 and C8 phases.
ACQUITY UPLC™ BEH Phenyl Columns
ACQUITY UPLC™ BEH Phenyl columns were intelligently designed to overcome these weaknesses and provide complementary selectivities, pH stability and excellent peak shape for all compounds
ACQUITY UPLC™ BEH Phenyl columns utilize a trifunctional C6 alkyl tether between the phenyl ring and the silyl functionality.
This ligand, combined with the same proprietary endcapping processes as the ACQUITY UPLC™ BEH C18 and C8 columns, provides long column lifetimes and excellent peak shape.
This unique combination of ligand and endcap on the 1.7 μm BEH particle creates a new dimension in selectivity and efficiency for challenging UPLC™ separations.
ACQUITY UPLC™ BEH columns provide column lifetimes under UPLC™ conditions which meet and/or exceed HPLC column lifetimes run under HPLC
COMPARISON OF HPLC AND UPLC charecteristics HPLC UPLC
particle size 3 to 5m less than 2m
Analytical column XTerra c18,Allitma c18
ACQUITY UPLC BEH C18,c8,RP SHEILD,phenyl
column Dimensions 150× 3.2 mm 150× 2.1mm
column temperature 30 c 65 c
Injection volume 5 µL 2µL
Flow Rate 3.0 mL/min 0.6mL/min
Needle wash methanol Strong Needle Wash: methanolWeak needle wash:ACN:H2o 10:90
APPLICATIONS OF UPLC
• Analysis of natural products and traditional herbal medicines• Identification of metabolite• Bioanalysis/bioequivalence studies• Impurity profiling
USE OF UPLC SYSTEM
1.Elevated-temperature chromatography also allows for high flow rates by lowering the viscosity of the mobile phase, which significantly reduces the column back pressure.
2.Monolithic columns contain a polymerized porous support structure that provide lower flow resistance than conventional particle-packed columns
ULTRA FAST LIQUID CHROMATOGRAPHY
Ten times higher speed and three times
better separation. Uflc offers outstanding
speed and separation even at normal
pressure levels, objectives that were
difficult for conventional systems. By
maximizing the column and performance
of the entire system,uflc minimizes the
deviation from the van deemter theory
UFLC COLUMNS• Uflc silica column• Available as 1.8 and 2.2µm particle sizes
•Available in analytical and preparative HPLC scale column dimensions and particle sizes•Wide pore size selection - 60, 120, 200, 300, 500, 1000 & 2000 Å
• HP Silica phase is made of activated hydroxyl (-OH) functional group. Used for both normal phase and HILIC applications e.g. LC/MS analysis of nitrogen-containing compounds.
• C18 uflc column• C18 reverse-phase UFLC/UHPLC Column• 2.5µm partcle size• Significantly lower backpressure compared to other 2µm
columns• Columns to support easy scale-up of UFLC separations to
preparative scale
• Cyano column• Available as 1.8 and 2.2um particle sizes• Available in analytical and preparative HPLC scale
column dimensions and particle sizes• Monomeric and fully endcapped phase structure; %
carbon load - 7%• Propyl cyano functional group• pH stability 2 - 8.5• amino column• Available as 1.8 and 2.2µm particle sizes
•Available in analytical and preparative HPLC scale column dimensions and particle sizes•Monomeric and fully endcapped phase structure; % carbon load - 4%•Propyl amino functional group•pH stability 2 - 8.5
APPLICATIONS OF UFLC• Analysis of isoflavones in soy column: shimpack XR-ODS mobile phase : A. 0.1% formic acid aqueous
solution B. 0.1% formic acetonitrile solution flow rate : 1.2ml per min column temperature : 400 c injection volume : 5µl column pressure : 28.8 mpa
Analysis of catechins in green tea
Column : shimpack XR- ODS
Mobile phase : A. 0.1%formic acid aqueous
solution/THF = 95/5
B.Acetonitrile
Flow rate :0.5ml/min
Column temperature : 50o c
Injection volume :2µl
column pressure : 19mpa
RAPID RESOLUTION LIQUID CHROMATOGRAPHY
• Typically RRLC (Rapid Resolution Liquid Chromatography )columns have lower particle size - 1.8 microns compared with 2 to 10 micron
conventional columns.
• Because of their lower particle sizes they
operate at higher pressure (600 Bar ) levels than
normal columns (400 Bar).
• Rapid Resolution Liquid Chromatography (RRLC) has become an increasingly useful approach to achieve higher throughput, improve sensitivity and reduce costs.
“Rapid Resolution” LC system enables faster
analysis (theoretically up to 20x) than with conventional HPLC while maintaining equivalent resolution.
This is achieved by using sub-2 micron column particle chemistry and high flow rates. Often higher temperatures are employed to minimize system back-pressure.
• High resolution chromatography – 90,000 plates
in 4 minutes• Ultra-fast separations – up to 20 times faster• Full compatibility with existing HPLC methods• More detection capabilities – from UV-visible
and ELSD through LC/MS• Near-zero sample carryover – for
uncompromised data quality• Highest system flexibility – for automated
method development
COLUMNS IN RRLC
• Thermostatted Column Compartment SL• Temperature range: 10 °C below ambient to 100 °C• Two independent heat exchangers allow pre-column heating
and post-column cooling for lowest detection limits 400.
PARAMETERS IN HPLC AND RRLC
HPLC RRLC
column ID(mm) 2.1- 4.6 2.1- 4.6
Particle size (µm) 3,5 1.8
Pressure (psi) 3000 9000
Flow rate (ml/min) 0.6-1.2 0.2- 2.0
Temperature (0c) upto 40 upto 100
Faster Analyses with RRLC Quaternary Amines by HPLC-CAD
System: Agilent 1200 RRLCColumn: Shiseido MG C18, 4.6 x 250 mm (5 μm)
Mobile Phases A: Water, 0.1% Formic acid B: AcetonitrileFlow Rate : 1.00 mL/min
Gradient : T= 0 min 10% B, T= 15 min 90% B, T= 20 min 90% B, T= 22 min 10% B.
Column Temperature: 40 °C
Injection Volume: 2 μL
Run Time: 25.00 minutes
Corona CAD: N2 Pressure: 35.0 psi
Filter: High
Nebulizer Temperature: 30 °C
APPLICATIONS OF RRLC
Quaternary Amines by RRLC-CADSystem: Agilent 1200 RRLCColumn: Waters UPLC BEH C18, 2.1 x 50 mm (1.7 μm)Mobile Phases A: Water, 0.1% Formic acid B: Acetonitrile
Flow Rate : 0.65 mL/minGradient: T= 0 min 10% B, T= 3 min 90% B, T= 4 min 90% B, T= 4.5 min 10% B
Column Temperature: 50 °C (pre-col.), 30°C (post-col.)
Injection Volume: 2 μL
Run Time: 6.00 minutes
Corona ultra2 Pressure: 35.0 psi: N
Filter: High
Nebulizer Temperature: 30 °C
NANO LIQUID CHROMATOGRAPHY
• Optimized for ultra-low flow rates of first dimension 1-20 µL/min, second dimension 50-1000 nL/min.
• Novel pumping system eliminates the need for flow splitting, providing excellent run-to-run reproducibility.
• Use automated peak parking for extended MS/MS analysis to improve peptide sequence coverage.
• Take advantage of nanospray sensitivity. • Active flow control of each mobile phase
ensures precise gradient delivery. • Quickly responds to changes in flow rate set
point in less than two seconds. • Reduce solvent consumption by 99%!
COLUMNS IN NANO LC Three basic types of capillary columns used in nano-liquid
chromatography: packed, monolithic, and open tubular.
• Packed Capillary ColumnPacked columns are made by “stuffing” the capillary with silica-modified particles of 3–5 μm. Though recently, particles of even smaller sizes 1.5–1.8 μm were successfully employed in ultra performance LC (UPLC).
• Such a small particle size provides nano-liquid chromatography systems with higher efficiency, resolution, selectivity, and shorter analysis time; however, it does increase the backpressure.
• So far, the application of packed capillary columns is the most
explored option in nano-liquid chromatography.
• Monolithic Capillary ColumnMonoliths are a block of continuous materials made of highly porous rods with two types of pore structures (macropores and mesopores of different sizes), which allow the use of higher flow rates and thus reduces the analysis time.
• Presently four types of monolithic capillary columns can be found: particle fixed, silica based, polymer based, and molecular imprinted monolith.
• Up-to-date there is not much research information on application of monolithic capillary in nano-LC.
• Open Tubular (OT) Capillary ColumnIn open tubular liquid chromatography column, the capillary wall is coated with highly permeable porous material that serves as the stationary phase.
• The OT capillary has lower sample loading capacity of the column, because only a small surface area is available for analyte interaction that can result in column overloading causing peak asymmetry and poor efficiency.
APPLICATIONS OF NANO LC
• Protein identification and functional analysis require new analytical tools and techniques that offer high throughput and enhanced
detection sensitivity for identification of low-abundance proteins.
ADVANTAGES 1.Significantly reduces solvent consumption and subsequent waste production. 2.ID diameter reduction increases sensitivity and/or less sample requirement. 3.Decrease in column bead size (packing) narrowers the peak width of chromatogram due to better separation efficiency. 4.Does not increase system pressure
REFERENCESIJPQA 2010, JAN-MAR VOL.2,ISSUE 1(19-25)
WWW.SHIMADZU.NET
WWW.ESAINC.COM
WWW.EKSIGENT.COM
WWW.AGILENTED.COM