Issue No: : 2 Issue Dt 30.06.2003 Page No : 10 OF 19 Issued : Q A Sl.No. Parameters CPRI Specification/Requirements Guaranteed Technical Particulars (GTP) Deviation/Remark s Specify if any 1. OBJECTIVE: To provide facilities for carrying out testing of Routine, Acceptance and Certification (Type Test) test on all types of Electro Mechanical /Electronics 1 Phase / 3 Phase whole current & 3 phase CT/PT operated energy Meters (For all the tests mentioned in this specification ) RSS meters , ERS meters in Active, Reactive and apparent Energy Mode. 2. SCOPE: Design, Engineering, Manufacture, Supply, Installation And Commissioning of Fully Automatic Portable Integrated Three Phase Current And Voltage Source With 0.05 % Reference Standard For Wh / LTCT /HTCT Energy Meters. In addition to the above the bidder shall also be required to provide Operations & Maintenance Manuals including drawings. Training : Three days training excluding duration of Installation and commissioning to CPRI Engineers on all aspects of operation and maintenance at CPRI. Warrantee: Warrantee of the test system is required covering all the supplies for period of Three (3) year from the date of successful installation and commissioning. Continued technical support during warranty period to be provided. FORMAT NO.:PUR/TENDDOC/P10/06 Name of the Vendor To be completed by the Bidder CPRI www.tenderwizard.com/CPRI Section II -Technical Specification Tender Enquery No : PUR/ID-05/12-13 Offer Number and Date Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection PURCHASE DIVISION DEPARTMENT QUALITY MANUAL Revision No. : 02 Dt of Revision : 12.9.2012 Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS) Section : 0 Document : DQM-01 1

Transcript of PURCHASE DIVISION DEPARTMENT QUALITY MANUAL 13010,IS13779,IS14697&IS15884 Test of Meter constant...

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

1.      OBJECTIVE: To provide facilities for carrying out testing of Routine, Acceptance and Certification (Type Test) test on all types of Electro Mechanical /Electronics 1 Phase / 3 Phase whole current & 3 phase CT/PT operated energy Meters (For all the tests mentioned in this specification ) RSS meters , ERS meters in Active, Reactive and apparent Energy Mode.

2.      SCOPE: Design, Engineering, Manufacture, Supply, Installation And Commissioning of Fully Automatic Portable Integrated Three Phase Current And Voltage Source With 0.05 % Reference Standard For Wh / LTCT /HTCT Energy Meters. In addition to the above the bidder shall also be required to provide Operations & Maintenance Manuals including drawings.

Training : Three days training excluding duration of Installation and commissioning to CPRI Engineers on all aspects of operation and maintenance at CPRI.

Warrantee: Warrantee of the test system is required covering all the supplies for period of Three (3) year from the date of successful installation and commissioning. Continued technical support during warranty period to be provided.

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

1

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

3 QUALIFYING REQUIREMENTS (FOR MANUFACTURERS):Shall have at least 10 years of experience in design and manufacture and supply of Meter Test Equipment. Should have supplied at least five similar equipment during the last five years in India including GOI organisations which should be in successful operation at ISO/IEC/ 17025 laboratory on the date of offer.

Bidder must give details like Copy of P.O, name of user, contact person, addressand phone no. of user who is using similar system in support of above.

All major equipment like Power Source, Reference Meter, and (Software) Measuring Unit shall be manufactured by original manufacturer for seamless integration, functioning and after sales & service. (Make of Major equipment shall be mentioned)The Bidder should declare himself that his firm is not black listed by SEB’s/any utility in India. If found later order may be cancelled and compensation claimed.

2

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

The bids may be submitted by the manufacturer or their sole authorised representative duly supported by certificate of authorisation

The Bidder should have its own service centre and trained engineers dedicated for trouble shooting and technical support permanently posted in India.

The equipment shall be CE compliant4 OPERATING & SAFETY REQUIREMENTS:

The Portable Power Source shall be suitable for giving an uninterrupted service in following conditions:Ambient temperature (+) 10°C to (+) 40°C for operation. Relative humidity up to 90%.Mains voltage shall be 3x240V 10% for three phase supply. Frequency 50Hz ± 5 %.The Equipment must be Dust proof.The Portable Power Source shall draw purely sinusoidal current from mains without polluting the mains and keeping the power factor of mains nearly unity

3

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

5 CONFIGURATIONS & TESTS TO BE PERFORMED Meter Configurations (Minimum requirements):1 PH 2 W (1 No with closed link):Active & Reactive Energy)3 PH 4 W (1 No with closed link):Active & Reactive & Apparent Energy3 PH 3 W (1 No):Active, & Reactive & Apparent Energy The offered Portable Power Source shall be capable to perform the following tests on the meters as per IEC 62052-11, 62053-11,21,22,23, is 13010,IS13779,IS14697&IS15884Test of Meter constantTest of Starting ConditionTest of No Load Test of Power ConsumptionTest of Self heatingTest of HeatingTest of Immunity to Earth faultTest of Influence quantities:Accuracy Test for Active & Re active Energy :

4

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

Limits of Error ( Balanced & Un balanced in all 4 Quadrants)Test of repeatability of errorVoltage variationFrequency variationReverse Phase sequence test Voltage Unbalance testAll other accuracy tests of Electro mechanical metersCalibration/ testing of ERS Meters

6 CONSTRUCTION AND COMPONENT OF SYSTEM Protection against under voltage and over voltage of mains supply shall be provided.System shall be easily programmable to give reference output frequency independent of mains from 45 Hz to 65 Hz in steps of 0.01 Hz. THE SYSTEM SHALL COMPRISE OF :Voltage Source. (Refer 6.1)Current Source . (Refer 6.2)Three Phase Reference Meter of 0.05 Accuracy class. (Refer 6.3)

5

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

Clip – on CT’s(Refer 6.4)Connecting cables (Refer 6.5)Scanning head with holding arrangement and Connection cable. (Refer 6.6)

Software features(Refer 6.7)6.1 VOLTAGE SOURCE CAPABILITIES: :

It should have output VA burden rating not less than 30 VA per phase.Electronic protection against O/L and Short Circuit. LED indication for fault shall be provided . Stability 300 ppm / h, or better Distortion factor < 1.0 %.Test voltage range : 40 - 480 V (Phase-Neutral)

6.2 CURRENT SOURCE CAPABILITIES:It should have output VA burden rating not less than 60 VA per phase .Electronic protection against open circuit and over load.Stability 300 ppm / h with integration time of 60 sec

6

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

Distortion factor <1.0 % Test Current range:10 mA to 120A & facility to generate starting current in the range of 1mA to 10 mA Test voltage and test current system should be freely selectable as symmetrical & Un symmetrical with change in phase sequence         Star systems (phase angle 120°)         Any non -symmetrical system          Any non balanced system       PHASE ANGLE:   0 TO 360 deg         Angle setting in step of 0.01° or better         Resolution : Shall be stated by the bidder

FREQUENCY :45 to 65 Hz         Resolution : Shall be stated by the bidder

6.3 THREE PHASE REFERENCE METER OF 0.05 ACCURACY CLASS Voltage range : 20 mV -480 V ( Phase - neutral ) Working Current range: 1 mA to 120 A Direct connected.

7

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

Frequency OUTPUT/INPUT proportional to the power to calibrate reference standard against High or Lower precision reference standard.RS 232 serial communication port for communicating with PC. Reference standard should have auto-range selection facility. Measuring modes :2 wire active &Reactive3 wire active & reactive & apparent mode4 wire active & reactive & apparent modeFrequency Range : 45... 65 Hz Fundamental. Indicate harmonic measurement capabilities.Accuracy of Parameters:Voltage : 0.05 %Current (Direct Connected ) : 0.05 % ( 50 mA to 120 A ) ( 10mA to <50 mA ) Shall be stated by the manufacturerCurrent with clamp On CT 100 Amps : Shall be stated by the manufacturer Power / Energy ( For active and reactive measurement ) 0.05 % or better at cos f =1 or sin f =1

8

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

0.05 % or better at cos f =0.5 or sin f =0.5 Phase Angle Accuracy 0.01 Voltage , Current & Power measurement drift shall be better than 300 PPM / year . Temperature drift for voltage & current measurement shall be 5 PPM/ºC definedclearly in the offer.Temperature drift for power measurement shall be 10PPM/ K or better. Display of reference standard :The reference meter shall have following display parameters. True RMS value of each voltage & current input Phase angle between voltage / current and defined reference Power factor of each phaseActive, reactive & apparent power of each phase Total active, reactive & apparent powerPhase Sequence Frequency Integration time

9

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

Facility to select integration time between 1 to 99 second shall be provided in the RSM .Meter constantVector representation of voltage & currentRepresentation of Harmonic voltage & Current in Bar Chart mode & THDReference Channel The RSM shall have facility to select reference for phase angle measurement. Selection of reference shall be provided manually & automatically.

Frequency out put for calibration of reference standard: This shall provide frequency output proportional to power to calibrate the reference standard against high precision reference standard. BNC type socket is preferred for output .Frequency Input for calibration of substandard meters: Frequency input connections shall be provided preferably with BNC socket to receive electrical pulses from substandard meters. It shall be possible to calibrate/ test substandard meters on offered system .

10

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

6.4 ELECTRONICALLY COMPENSATED CLAMP ON CT Required Accuracy from 5A To 50 A : At least 0.2 %

6.5 CONNECTION CABLES:Apart from connection cable required to install and commission the meter supplier shall provide the following set of connection cable for meter under test.

Requirement for simultaneous testing of three meters

Current connection cable set for of 3 phase 4 wire CTVT operated meters up to 15 amps continuous current capacity : 6 Nos Any other Cable type shall be stated by the manufacturer

High Current connection cable set up to 120 amps continuous current rating fortesting of 3 phase 4 wire whole current meter Any other type shall be stated bythe manufacturer : 6 NosVoltage connection cable for testing ERSS : 4 Nos

Voltage connection cable for phase & Neutral (2.5 sqmm /one side will be connected to meter under test by Screwing). 4 Nos Any other type shall be stated by the manufacturer

11

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

Current connection cable for testing ERSS : 6 NosScanning Head Connection cable : 1 SetSCANNING HEAD WITH HOLDING ARRANGEMENT AND CONNECTIONCABLE.

6.6 1 photoelectric scanning head suitable for reading the marking from the disc ofFerraris wheel meters without opening the cover of the meter. Same scannershould also be suitable for reading the pulse output of electronic meters with LED& LCD Display.

Suitable Arrangement shall be provided with the equipment to test Energy Meter in EMI/EMC environment using Optical Pulses of Energy Meter

The scanning heads must be insensitive to ambient light and shall meet therequirements of IEC 62052-11,Clause 5.11.The pulse frequency shall be minimum 500 Hz & actual frequency shall be stated by the manufacturer.

6.7 SOFTWARE FEATURES: Shall have provisions for

12

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

v   both automatic and manual modes.v   User friendly operation with pull down menusv   Setting of parameters for source v   Setting of display parameters for RSMv   Selection of only desired testv   Creation of batch test sequence / test tablev   Creation of test procedurev   Supervision / control of test procedure

v   Facility to define test parameters in terms of percentage and absolute term.

v   Display of errors in percentage.v   Setting test parameters like CT,PT Ratio ,Meter constant, Max-Min V / I, Errorlimits for each position and any other relevant parametersv   Display of V,I, PF, F, phase sequence, three Phase power & energies,demand and other related parameters as per user choice.v   Vector diagram for the chosen mode of connection with valuesv   Tripping source to MUT in case of abnormal test conditions

13

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

v   Selecting print and save options.User configurable print formats

Facility to limit maximum current and voltage of the meter under test for protectionof the meter. Software shall have adequate protection system to protect meterfrom over voltage & over current in manual mode and automatic mode.

Facility to check meters for short circuit and open circuit conditions prior to start oftesting in automatic mode for each sequence.

7 DOCUMENTATION The bidder shall submit detailed General arrangement drawing for Source, withdifferent terminal required for voltage and current circuit connection,,

Two sets of following documents shall be supplied along with each test system.

Operating manual of each components of test equipment like reference standard, amplifier, etc.Wiring diagram

14

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

Service manualCalibration certificate in the name of CPRI for RSM from the laboratory accredited as per ISO/IEC/17025 . Name of the laboratory shall be given in the bid.Test certificate of complete test systemProcedure to validate software shall be provided.

8 INSTALLATION AND COMMISSIOING The supplier shall be responsible to install & commission the meter test equipment at the purchaser location. The supplier shall submit the layout plan, installation proposal and electric supply requirements with in 4 weeks after receiving the purchase order.

TRAINING 9 The supplier shall provide three days training on operation and maintenance of the

meter test equipment to engineers of purchaser. ( Excluding the period ofinstallation and commissioning)

10 ADDITIONAL ACCESSORIES SHALL BE QUOTED INDIVIDUALLY

15

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

Requirement & QtySCANNING HEAD WITH HOLDING ARRANGEMENT AND CONNECTIONCABLE. : 3 NosVoltage connection cable for phase & Neutral (2.5 sqmm /one side will beconnected to meter under test by Screwing).Any other type shall bestated by the manufacturer : 20 NosCurrent connection cable set for of 3 phase 4 wire CTVT operated meters up to15 amps continuous current capacity .Any other Cable type shall be stated by themanufacturer : 30 NosHigh Current connection cable set up to 120 amps continuous current rating fortesting of 3 phase 4 wire whole current meter Any other type shall be stated bythe manufacturer : 30 NosVoltage connection cable for testing ERSS : 8 NosCurrent connection cable for testing ERSS :12 Nos Fuses :10 NOS EACH

PN : 1. A detailed technical catalogue/literature/phamplet and any other technical details shall be sent in hard copy in a sealed cover super scribing tender

16

Issue No: : 2Issue Dt 30.06.2003 Page No : 10 OF 19Issued : Q A

Sl.No. Parameters CPRI Specification/RequirementsGuaranteed

Technical Particulars

(GTP)

Deviation/Remarks Specify if any

FORMAT NO.:PUR/TENDDOC/P10/06

Name of the Vendor

To be completed by the Bidder

CPRIwww.tenderwizard.com/CPRI

Section II -Technical SpecificationTender Enquery No : PUR/ID-05/12-13

Offer Number and Date

Name of the Equipment:FULLY AUTOMATIC PORTABLE INTEGRATED THREE PHASE PORTABLE POWER SOURCE WITH 0.05% REFERENCE STANDARD FOR WH / LTCT /HTCT ENERGY METERS

Note :The Offers should be submitted only in this format otherwise the offer will be liable for rejection

PURCHASE DIVISIONDEPARTMENT QUALITY MANUAL

Revision No. : 02Dt of Revision : 12.9.2012

Topic : FORMAT 6 OF PROCUREMENT PROCEDURE OF CPRI (NON-WORKS)Section : 0 Document : DQM-01

enquiry number and due date so as to reach the following address within the due date and time. This is very much essential to evaluate your offer.

2. Letter of authorization issued by the foreign Principal shall also be sent by the Indian agents who have offered on behalf of their Principals.

Address : Joint Director (Purchase) Central Power Research Institute, Sir C,V.Raman Road, P.B.No.8066, BANGALORE-560 080. PH. No.080-23602919/23602829. Extn: 2271/2274. Fax : 080 - 2360 4446

17

18

19

20

21

22

23

24

25

26

27

28

29

30

31

32

33

34