PMP: SUB 23 NM MEASUREMENT RECOMMENDATIONS B. Giechaskiel, G. Martini Institute for Energy and...
-
Upload
gyles-mason -
Category
Documents
-
view
214 -
download
1
Transcript of PMP: SUB 23 NM MEASUREMENT RECOMMENDATIONS B. Giechaskiel, G. Martini Institute for Energy and...
PMP: SUB 23 NM
MEASUREMENT RECOMMENDATIONS
B. Giechaskiel, G. Martini
Institute for Energy and Transport Joint Research Centre
3 April 2014
Sub23nm particles
• Percentages around 30-40% (over a cycle)
Mamakos et al. 2011 ETHGiechaskiel, PMP 29
GDI, vehicle Khalek et al. 2010 Ronkko et al. 2014
GDI, produces nanoparticles even when not fueled, coming from oil
Monitoring GDI important
Formation of particles in the VPR
• Indications when sulfur exists• Formed solid particles were <10 nm• Difficult to justify (pyrolysis?)• Example is reaction of sucrose with strong sulfuric acid• Experimental suggestions:
• HC and sulfuric acid through PMP• Both as gaseous and liquid droplets• SMPS and CPC downstream PMP• Check PCRF effect
Volatile Removal efficiency (ET)
• Hydrocarbons• Tetracontane will re-nucleate at >107 p/cm3 or 3 mg/m3
• Sulfuric acid• Sulfuric acid will re-nucleate at 0.7-3.5 μg/m3
• Growth after the ET• The re-nucleated nucleis are 1-2 nm.• Organics are necessary for the nucleis to grow• Theoretical calculations and experimental results show that they can
grow to 6 nm (1011 p/cm3 organics) or 20 nm (1014 p/cm3 hexadecane, <5 ppm C3)
Sulfur compounds in DPFs
• Partial DPF• Alumina based coating
with Pt as active metal• No artifact from
sampling lines etc.• Formation of sulfuric
acid particles from SO2 to SO3 conversion
Kajalainen et al. 2014
SO2
DPF regeneration
Volatile artifact appears as solid particles <23 nm (CVS)PCRF: 100x10
No artifact >6 nm (tailpipe)PCRF: 15x15
Typically no artifact above 10 nm (Zheng et al. 2012, Mamakos et al. 2013)
DPF regeneration
• Gaseous HCs do not correlate with PN<23
• Probably heavy HCs• Sulfates probably
important• Artifact probably from
desorption after tailpipe• Pre-conditioning of
setup is important
PCRF selection
• PCRF 100x10• Still >3nm artifact• Lower solid surface
nucleation vs condensation
• Probably no issues with >10nm measurements
Moped
Volatile artifact even with high dilution (1000x10)
With EEPS connected to PMP (350x20) no detectable artifact.Similarly with a 10 nm CPC
Thus artifact around <6nm
Volatile Removal Efficiency (CS)
• Catalytic Stripper (CS)• Removes volatiles by oxidation
• Even >35 times highest HC concentration (eg moped)• Better removal efficiency of tetracontane
• Traps sulfur • Up to 10 mg/m3
• Might oxidize SO2 to SO3
• Particles might be formed if saturated• Higher particle losses
CS has also limitations
• Silica-alumina coating with Pt and Barium oxide
Amanatidis et al. 2013
Summary
• Monitoring of sub23nm is recommended• Decreasing the size to ~10 nm is possible. VPRs should be
used with high dilution (>1000:1)• Special attention to GPFs and DPFs (no soot)• The CS is an alternative solution that has to be
investigated:• Saturation with sulfur• Formation of particles
• SO2 to SO3 conversion
Summary
• To measure below 10 nm:• A dilution system with minimum losses of sub23 nm is
necessary.• A catalytic stripper well documented is necessary• A full flow CPC with low cut-off size is necessary
Next steps
• The setup has to be well pre-conditioned• Organics, sulfates would be useful• Measurements in parallel with PMP systems with ET
and CS devices with both 23nm and ~10 nm PNCs.• Support from instrument manufacturers needed
especially for CS, or setting a second CPC
Next steps
• Tests without CS should be conducted with as high PCRF as possible (at least 100x5)• Note that even with PCRF 10000, and PNC concentrations
<1000 p/cm3, artifacts can happen because they depend on the volatile species’ concentration.
• When high sub23nm concentrations are measured (>100%, at emissions 10^11 p/km) the tests should be repeated with 10x higher PCRF• Lower concentrations could be background noise