Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 1 Declaration of Conflict of Interest or...
-
Upload
isabell-darke -
Category
Documents
-
view
217 -
download
1
Transcript of Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 1 Declaration of Conflict of Interest or...
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 1
Declaration of Conflict of Interest or Relationship
Speaker Name: Paul Tofts
I have no conflicts of interest to disclose with regard to the subject matter ofthis presentation.
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 2
Methods for quantitative relaxation parameter mapping:
measuring T1 and T2
Paul Tofts PhDChair in Imaging Physics
Brighton and Sussex Medical School, UK
Download this from http://www.paul-tofts-phd.org.uk/talks see also: qmri.org
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 3
Why measure relaxation times?
• T1 needed for Ktrans measurement in DCE MRI
• T1 needed for quantitative MT analysis
• T1 and T2 demonstrate subtle damage in ‘normal-appearing’ white matter
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 4
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 5
overview
• Principles of T1 measurement
• Principles of T2 measurement
• B1 effects
• Quality assurance
• Measurements on a clinical scanner
• Image analysis
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 6
T1 measurement
• Bloch eqn:
• Steady state solution:
Repeated θ pulses; Mz just before a pulse
1
)()(
T
tMM
dt
tdM zoz
1
1
/
/
0 cos1
1)0(
TTR
TTR
z e
eMM
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 7
T1 measurement: 1. saturation recovery
• Spin echo; repeated 90o pulses
• 2 (or more) TR’s gives T1
short and long TR’s (TR<0.5T1, TR>2T1) Needs good 90o pulse (need Mz(0+)=0) Slow (compared to gradient echo GE)
)1()( 1/0
TTRxy eMtM
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 8
T1 measurement: 2. spoilt gradient echo
• gradient echo; repeated θ pulses
• 2 (or more) FA’s gives T1 (nb FA=θ) Small and large θ
• (θ<θE: PDw; θ>θE: T1w; θE= Ernst angle) Needs accurate θ pulses (more later) Fast (can be 3D) Inaccuracies (at short TR):
• Incomplete spoiling• Relaxation during pulse
sincos1
1)(
1
1
/
/
0 TTR
TTR
xy e
eMtM
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 9
T1 measurement: 3. inversion recovery
• repeated [180o, TI, 90o , readout, TR-TI]
NB any TR
• arbitrary FA’s:
Fix TR, vary TI
Gives ACCURATE T1 (any TR, any FA)• This is gold standard, used in spectroscopy• MRI in ‘spectroscopy mode’ (small uniform sample; long TR)• Slow; interleaving slices
)21()2( 11 //0
TTRTTIz eeMM
1/):2( TTIz BeATIM
1
11
/21
/1
/1
0 coscos1
cos)cos1(1)2(
TTR
TTRTTI
z e
eeMM
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 10
T1 measurement: 4. specialist methods• There are a number of specialist
methods (see e.g. Gowland 2003), which differ quite substantially from the conventional methods described above
• include Look-Locker, DESPOT, DESPOT2, and TAPIR
• often very fast, and sometimes can function in the presence of B1 errors (although at the expense of worse SNR)
• may be vulnerable to off-resonance effects
• usually require intimate access to the MR imager to implement the sequences
Gowland PA, Stevenson VL. T1: the Longitudinal Relaxation Time (chapter 5). In: Paul Tofts, editor. Quantitative MRI of the brain: measuring changes caused by disease. Chichester: John Wiley, 2003: 111-141.
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 11
T2 measurement
• Bloch eqn:
• Steady state solution:
Needs TR-TE>>T1 (or else variable recovery after 180o pulse)
• If FA’s inaccurate,
T2 estimate is still ACCURATE
2
)()(
T
tM
dt
tdM xyxy
21 //0 )1(),( TTETTR
xy eeMTETRM
2/)( TTExy AeTEM
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 12
T2 measurement: 2. multi-echo
• Tempting to acquire >1 echo in 1 shot E.g. dual echo or FSE
• Selective 180o pulse is imperfect Slice thinning
• Even in a long T2 substance (e.g. pure water T2>2s) echoes decay with each refocusing
Simple QA on T2 accuracy
• Gold standard is: repeated single echo
long TR
few slices
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 13
Hippocampal T2 in epilepsy
• Conventional multi-slice dual spin echo TE=30,120ms
Figure 2 Unilateral focal anterior hippocampal sclerosis. The quantitative data show a left anterior involvement which was missed by visual inspection.
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 14
T2 measurement: 3. multiple-T2 s• There is no such thing as T2
• In real biological tissue the transverse magnetisation almost always decays with several T2 values
• 3 distinct compartments in brain Myelin water (10-50ms)
Intra-cellular + extra-cellular (70ms)
CSF (>1s)
• estimated (mono-exponential) T2 value depends on TE values chosen
e.g. 4ms range (white matter T2~80ms)
Whittall KP, MacKay AL, et al . In vivo measurement of T2 distributions and water contents in normal human brain. Magn Reson Med 1997;37:34-43.
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 15
B1 effects and FA errors
• Many RT measurement methods rely on accurate FA• sources of FA error
Scanner prescan procedure Slice selection (in 2D sequence) Dielectric resonance
• Worse at 3T Transmit coil nonuniformity
• Body coil better than head coil
• Solutions Use FA-insensitive methods Map FA
Dowell NG, Tofts PS. Fast, accurate, and precise mapping of the RF field in vivo using the 180 degrees signal null. Magn Reson Med 2007;58:622-630.
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 16
FA errorsand slice selection
Selective 90o pulse gives reasonable profile at correct B1 value
Poor profile at incorrect B1
Effective FA varies across slice
Out-of-phase component
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 17
Quality assurance for RT mapping
• Needs specific tests Distinct from e.g. resolution, SNR tests
Phantoms (test objects)
Healthy controls
Convenient to scan + -
Well characterised (measure accuracy)
+ -
need temperature control? Yes (2-3% / oC) +
realism - +
Manufacture/purchase - +
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 18
Accuracy and reproducibility• Accuracy
Closeness to truth Important for multi-centre studies Small systematic errors are acceptable in typical (well-designed)
cross-sectional and serial studies
• Reproducibility Test-retest Bland-Altmann analysis from pairs of measurements Instrumental variance usually
• sets limit to smallest biological change that can be measured• determines power of a study
Improved reproducibility gives smaller studies • faster, cheaper
Tofts PS, Collins DJ. Multicentre imaging measurements for oncology and in the brain. (in press) 2009.http:www.paul-tofts-phd.org.uk/CV/reprints/multicentre-2009.pdf
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 19
Cross-sectional study• Normal tissue value = 100• sd_normal = 3• disease effect = 5
• perfect instrument*:
sd << sd_normal
sd << 3
• Serial study:• perfect instrument
sd << sd_within subject
sd << 1
* An instrument which is so precise that it does not introduce any significant variation to the existing biological variation
Power calculation (G*Power3) effect=5; sd_norm=3; sd_disease=4.25; alpha=0.05; P=0.80; 2-tailed; N1=N2
0
20
40
60
80
100
120
140
160
0 2 4 6 8 10 12
sd_instrument
sample size
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 20
Measurement using a clinical MR imager
• Implementation not straightforward Commercial sequences for RT are undeveloped
• Speed; accuracy; B1 errors
Home-made sequence needs research agreement
• Literature Not much!
• cf MT, DTI, MRS, Ktrans, volume …
Scope for research
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 21
Measurement using a clinical MR imager – T1 • Spoilt Gradient Echo
Brain: 3D; 1.5 x 1.5 x 1.5 in 10 minutes Body: multi-slice; breath-hold? B1 and FA nonuniformity
• Body coil; nonuniformity fixed for a given subject
• Fast Inversion Recovery IR-prepared SPGR ; IR-EPI (MP-RAGE) Accurate (fixed readout; short readout, long TR) Typical 3% accuracy in 10 min
• Single slice, long time is gold standard Slow; limited coverage
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 22
Measurement using a clinical MR imager – T2 • Spin echo
Multi-slice (2D)• Beware slice cross-talk
Dual echo• Single echo more accurate but slower
Faster readout with RARE (multiple phase encodes per shot)
• FSE, turbo SE
• Multiple-T2 analysis No commercial software
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 23
Image analysis for relaxation time measurement
• Region of interest Test a specific location (prior information and hypothesis) biased
• Histogram Whole brain; unbiased good for diffuse disease in Normal Appearing Tissue
• Voxel-Based Morphometry VBM Unbiased testing of many locations Each location can be correlated with external score (clinical, genetic,
proteomic, cognitive)
• Texture ‘dirty white matter’ tissue often becomes more heterogeneous in disease
Paul Tofts ISMRM Hawaii 2009 - measuring T1 and T2 24
More…
• http://www.paul-tofts-phd.org.uk/talks This talk (presentation and document) Multicentre measurements 2009 (in press)
• qmri.org Out of print (2nd hand market £130 $288 $512) E-book ~ £110 through university library Book updates planned on website
• ISMRM list serve Standards in Quantitative MR http://ismrm.org/sqmr/sqmrrules.htm
Haacke et al 2009E-book