Overview of MSWS Control of 212-LC July 15, 2006.
-
Upload
augustus-george -
Category
Documents
-
view
218 -
download
1
Transcript of Overview of MSWS Control of 212-LC July 15, 2006.
Overview ofMSWS Control of 212-LC
July 15, 2006
2
Software Overview
Functionality• Derived from ProStar/Dynamax system driver used for 210s
• Only supports Binary system at this point.
• RS232 connection to Pump A only, requires port in COM1-4 range
• Method flow and composition ramps (up to 100 steps).
• Optional ramp on method activation and completion
• Manual mode
• GLP features (serial numbers, solvent info, stroke counters)
Distribution• 212-LC driver available for MSWS 6.6 SP1 and above
• “212-LC Control Software” CD to add driver to MSWS 6.6 SP1 ships with pumps – only needed with MSWS 6.6 SP1.
• Driver included in “LC Drivers” in MSWS 6.7 and above (installed when selecting (1200 LC or 500-MS during installation)
3
212-LC Software Configuration (1)
In System Control, select
Instruments->Setup Comm Ports…
For the serial port to which the 212-LC System is connected, pull down the Combo Box and select 212-LC
4
212-LC Software Configuration (2)
• In the configuration screen, the 212-LC module icon appears in the ‘Available Modules’ area.
• Drag it to the instrument window as shown.
• Switch to the Instrument screen and open the 212-LC window.
5
212-LC Software Configuration (3)
• In the 212-LC Control and Status window, push the ‘System Setup’ button.
• The System Setup dialog shows the current configuration and allows you to set solvents and change Pmin & Pmax
• Clicking on one of the Min or Max Pressure cells enables the ‘Set Pressure Limits’ button.
• Clicking on one of the ‘Solvent’ cells enables the ‘Select Solvent’ button.• Check boxes control whether or not composition and pressure traces are displayed.
6
212-LC Software Configuration (4)
• In ‘System Setup’, click on one of the cells showing Pressure Limits
• Press the ‘Set Pressure Limits’ button.
• Make a meaningful change to Pmax and press OK
• The selected values will be set for both pumps.
• In ‘System Setup’, click on the name of Solvent B
• Press the ‘Select Solvent’ button.
• Select a new solvent from the list and press ‘OK’.
• Press ‘Save’ to exit the System Setup dialog.
• Depending on changes made, a warning will come up saying that you need to change the Active Method or activate a new method. Opt to edit the active method.
7
212-LC Method Editor
• When the method editor opens up, select the 212-LC section
• As you do so, the editor may detect that the configuration does not match that of the instrument last connected at that address and offer to update the method. Press ‘Yes’.
8
212-LC Method: Window Layout
Time Program Spreadsheet
•Program Editing Buttons
Flow Profile
Composition Profile
•Activation & Completion
Ramps
•HW Configuration
9
Method Editor: Time Program Spreadsheet
• Click and type in cells to edit
• Enter time as mm:sec or fractional minutes (0:30 or 0.5)
• Enter %B, %A is computed
• Enter flow in uL/min
• “Blank” value means same as previous
• Option to show or Hide %A
10
Method Editor: Time Program Edit Buttons
Basic Editing
• Add, Insert, Delete – all apply to the current selection in the spreadsheet
Advanced Editing
• Scale Run Time: Recompute all the steps to scale the whole program to a different length.
• Add / Subtract Run Time from specific location in program.
11
Method Editor: Set Configuration
Configuration holds
• Pump Info
• Solvent info
The configuration is used to
• Know the pump info when editing off-line
• Verify at download time that the method was meant to be used with the
current solvents and pump options
12
Method Editor: Ramps
Ramp on Activation
• Ramp Time – Upon method activation, ramp from the current flow/composition to the time zero conditions of the method.
• Hold Time – equilibration time to follow the ramp.
Ramp on Completion
• Ramp Time – At the end of the method, ramp from the final flow/composition to the time zero conditions of the method.
• Hold Time – equilibration time to follow the ramp.
13
System Control: Activation
• In the 212-LC window, press the ‘Activate’ button. This activates the method and starts the pumps.
• If a ramp on activation is specified, the window title will read ‘Equilibrating’, then ‘Stabilizing’ while the ramp and hold time are satisfied.
• During the ramp, the Runtime field counts down from ramp time + hold time.
• Once Activation is complete, the window title reads ‘Ready’ and the ‘Run’ button is enabled.
14
System Control: Running the 212-LC
• If you are set up with an AutoSampler, use ‘Inject Single Sample’ to initiate a run. Otherwise, just press the ‘Run’ button in the 212-LC window.
• As the program executes, the Runtime is updated in min:sec, and the time program and pressure traces are shown if specified in ‘System Setup’..
• At the end of the run, the display is ‘frozen’ until the start of the next run. Press ‘Resume Plot’ to un-freeze the display.
15
System Control: ‘Run Status’ controls
The ‘Hold’ button is active during the run. It suspends the time program and extends the end time accordingly. Press ‘Continue’ to resume the program.
See trace for held/resumed program below.
The ‘Method’ button calls the method editor with the module’s method section selected. When the editor is closed, System Control will offer to reactivate the module method if it has been modified..
16
System Control: ‘Pump Status’ and Keypad
The ‘Unlock’ button temporarily disconnects the pumps and returns functionality to the pump’s keypad. Press ‘Lock’ to reconnect the pumps.
‘Hide Keypad’ hides most of the keypad to give more space to the graphics.
Press ‘View Keypad’ to restore it.
STOP PUMPS immediately stops the pumps.
17
System Control: 212-LC Manual Mode
• In the 212-LC Window, press ‘Manual Control’
• Select the desired composition by editing %B
• Edit the desired flow and ramp time. If no ramp is needed, set the time to zero.
• Press ‘Start Ramp’ to execute the command. e system will ramp flow and composition from the original settings to the new settings.
• When done with the Manual Mode, press ‘End Manual Control’
18
System Control: GLP Information
• Displayed as a result of pressing the ‘GLP’ button in the 212-LC window.
• Keeps track of serial numbers, firmware versions, solvent info, and stroke counters. The bottom three counters are associated with consumable items and can be reset when the items are replaced.
• The GLP information is stored in each data file produced by the instrument including the 212-LC System.
19
212-LC Run Log
• MS and Chromatography data files acquired by an instrument include a full copy of the instrument method.
• The 212-LC method used can be seen in the Run Log in the Standard Report Application or in the MS Data Review Application.
• The Run Log also includes a formatted version of the GLP info.
************************************
212-LC Solvent Delivery System
************************************
Module Address: 24
GLP Information For 2 Pump System:
Pump A is Online:
Serial Number: 3961
Firmware Version: 5.39
Solvent: H2O
Compressibility: 100
Lifetime Strokes: 5069 since June 27 2006
Seal Strokes: 5069 since June 27 2006
Inlet CV Strokes: 5069 since June 27 2006
Outlet CV Strokes: 5069 since June 27 2006
Pump B is Online:
Serial Number: 3962
Firmware Version: 5.39
Solvent: H2O
Compressibility: 100
Lifetime Strokes: 2036 since June 27 2006
Seal Strokes: 2036 since June 27 2006
Inlet CV Strokes: 2036 since June 27 2006
Outlet CV Strokes: 2036 since June 27 2006
Hardware Configuration:
Solvent Pump A: 212-LC; Headsize: 2 ml; Solvent: H2O
Solvent Pump B: 212-LC; Headsize: 2 ml; Solvent: H2O
Ramp on Activation: Ramp Time: 0:00 min; Hold Time: 0:00 min
Ramp on Completion: Ramp Time: 0:00 min; Hold Time: 0:00 min
Time %A %B Flow Wait Alarm
0:00 100.0 0.0 0.200
5:00 100.0 0.0
5:06 90.0 10.0
10:00 90.0 10.0
10:06 80.0 20.0
15:00 80.0 20.0
[…]
20
212-LC On-Line Help
For more information, refer to the System Control and Method Editor On-Line Help.