Neck Preserving Hip System - Smith & Nephe...Figure 7: Post-op x-ray coxa vara 4 Surgical...
Transcript of Neck Preserving Hip System - Smith & Nephe...Figure 7: Post-op x-ray coxa vara 4 Surgical...
Surgical Technique
Legal Manufacturer
OHST Medizintechnik AG
Neck Preserving Hip System
2 Surgical Instructions – NANOS™ femoral neck prosthesis
NANOS™
Table Of Contents
Introduction and Product Description ............................................................................ 3
Femoral Neck Preservation ................................................................................................ 4
Indications ........................................................................................................................... 5
Contraindications ................................................................................................................ 5
Risks and conditions that can impair the success of the operation ................................... 5
Surgical Technique ............................................................................................................ 6
Preoperative Planning ......................................................................................................... 6
Case Studies ....................................................................................................................... 7
Surgery Step 1 ..................................................................................................................... 9
Surgery Step 2 ..................................................................................................................... 9
Surgery Step 3 ....................................................................................................................10
Surgery Step 4 ....................................................................................................................10
Postoperative Treatment ......................................................................................................11
Notes Regarding Implant Unpacking ..................................................................................11
Femoral Heads ....................................................................................................................11
Implants ..............................................................................................................................11
NANOS femoral neck prosthesis ........................................................................................11
Instrumentation ................................................................................................................ 12
NANOS Instrument Tray ..................................................................................................... 12
Optional Instruments ...................................................................................................... 14
NANOS Tray Rasps ............................................................................................................ 14
X-ray Templates ................................................................................................................ 15
DisclaimerThese Surgical Instructions are intended exclusively for professionals in the medical technology field. They are expressly not intended for informational purposes of medical lay persons. The product explanations contained in this pamphlet are of a general nature and do not constitute medical advice. The instructions were created and compiled by medical experts and technically qualified employees of OHST AG, based on their best knowledge. We do not assume liability or guarantee the topicality, accuracy and completeness of the information provided herein. Any liability for damages, material or immaterial, incurred through the use of this information shall be excluded.
Surgical Instructions – NANOS™ femoral neck prosthesis 3
The NANOS™ femoral neck prosthesis provides metaphyseal anchoring and load transfer. The implant requires a gentle and minimal bone resection. The cancellous bone around the metaphysis and the greater trochanter are retained to ensure load distribution and transfer.1,2 This upholds the principle of the ‚second line of defense‚ for this prosthesis concept.1,2,3,4,5,6
Matching sizes as well as a clearly structured assortment of instruments make it easier to determine and select the suitable implant intraoperatively.
A total of eleven fully compatible sizes are available, designed to complement each other. The part of the femoral neck below the conus has been tapered to increase the range of motion.
The development is based on past clinical experience of various short stem prostheses, biomechanical explorations and physiological optimization. NANOS™ is the third generation of this type of prosthesis.
Introduction and Product Description
Figure 1: NANOS Neck Preserving Hip Prostesis
Coxa valga
Figure 2: Illustration coxa valga
Figure 4: Illustration coxa normala
Figure 6: Illustration coxa vara
Figure 3: Post-op x-ray coxa valga
Figure 5: Post-op x-ray coxa normala
Figure 7: Post-op x-ray coxa vara
4 Surgical Instructions – NANOS™ femoral neck prosthesis
Coxa normala
Coxa vara
Femoral Neck Preservation The resection is done with maximum bone retention with a straight cross-cut osteotomy of the femoral neck. The arched stem follows the cancellous channel of the femoral neck into the femur and thus accommodates anatomical anteversion and retroversion as well as varus and valgus position of the femoral neck.
Retention of femoral neck = retention of offsetThe implant follows the natural CCD angle as shown in the example below.
Surgical Instructions – NANOS™ femoral neck prosthesis 5
Indications• Primary and secondary coxarthrosis
• Dysplasia coxarthrosis
• Avascular and post-traumatic femoral head necrosis
• Patients with good bone density and intact femoral neck
Contraindications• Marked osteoporosis
• Dysmorphia and loss of bone substance, as well as prior surgical procedures that do not guarantee the intended support
• Marked coxa valga
• Marked coxa vara with an implant position CCD < 125°
• Intolerance to the materials
If femoral heads with a neck XL are used, the range of motion is reduced by about 30°.
Risks and conditions that can impair the success of the operationPotential risks associated with the operation are:
• Obesity (body mass index BMI > 30)
• Anticipated extreme stresses (e. g. through sports or work), especially in patients with a body weight over 100 kg
• Disorders of bone metabolism (osteoporosis, osteomalacia)
• Occurrence of fissures, in rare cases fracture of the femoral bone
• Circulatory disorders of the affected limb
• Neurological disorders of the affected limb
• Muscle malfunction in the affected limb
• Overweight
• Alcohol or substance abuse
• Patient groups with mental disorders or addictions
• Pregnancy
• Growth in children and adolescents
• Anticipated extreme loading e.g. due to work and sport
• Epilepsy or other reasons for repeated trauma with an increased risk of fracture
• Joint deformities that make fixation of the implant difficult
• Weakening of the bearing structures by tumour
• High-dose ingestion of cortisone or cytostatics
• Previous or threatening infectious diseases with possible joint involvement
• Deep vein thrombosis and/or history of pulmonary embolism
• All general surgical risks
Figure 8: Example of preoperative planning for the NANOS™ NeckPreserving Hip Prostesis
6 Surgical Instructions – NANOS™ femoral neck prosthesis
The information provided under surgical technique is intended as recommendations and references. However, the detailed application or feasibility of application depends on the skill and experience levels of the individual practitioner.
For detailed information about the implant system and the instruments please refer to their respective instructions for use (Lit. No. 50000697 and 50000354).
Preoperative PlanningX-ray templates with 15% enlargement in analogous form are available for preoperative planning. As a standard, digital x-ray templates with a scale of 1:1 for use with the planning software MediCAD* are also available for download. The following illustration shows the preoperative planning for the NANOS™ femoral neck prosthesis.
The size and position of the prosthesis are planned using the anterior/posterior and the axial x-ray image. Depending on the system requirements, attachment is advised in the anterior/posterior region on the calcar femorale and the lateral corticalis. In the axial x-ray a proximal pressfit in the ventro/dorsal region and support of the distal tip of the prosthesis is achieved. Leg length and offset can be adjusted during the planning by setting the resection height and the appropriate choice of prosthesis.
Surgical Technique
*Upon request, the manufacturer OHST Medizintechnik AG, Rathenow will provide digital x-ray templates in databases of other providers for the digital planning software.
Figure 9: Preoperative x-ray Figure 10: Postoperative x-ray
Case 1Left hip, 49 years of age, male
Surgical Instructions – NANOS™ femoral neck prosthesis 7
Case Studies
Figure 11: Preoperative x-ray
Figure 12: Postoperative x-ray
8 Surgical Instructions – NANOS™ femoral neck prosthesis
Case 2Left hip, 48 years of age, male
Figure 13: Resection of the femoral neck
Figure 14: Resection of the femoral neck in case of a narrow bone situation or a fracture of the femoral neck
Figure 15: Preparation with the opening rasp
Figure 16: Preparation with rasp size 0
Figure 17: Compactor in situ
Surgical Instructions – NANOS™ femoral neck prosthesis 9
Surgery Step 1Resection of the femoral neckA maximum amount of bone is preserved during the resection of the femoral neck, in accordance with the planning. Usually 0.5 – 1.0 cm subcapital with a straight transverse osteotomy of the femoral neck.
In case of a narrow bone situation or a fracture of the femoral neck, the resection can be performed as shown in Figure 14.
Surgery Step 2Preparing the prosthesis siteThe opening rasp is used to prepare the path for the forming rasps. The opening rasp is inserted with a slight curved motion inside the interior cortex until the lateral cortical bone is reached at the height of the bottom edge of the lesser trochanter.
For minimally invasive preparation, the size 0 rasp is connected to the curved rasp handle and moved forward along the calcar ridge. This rasp can later be used for the finishing preparation at the lateral cortical ring.
This is followed by finishing with the cancellous compactors. The aim of the preparation with compactors is to compact cancellous bone.
The preparation is carried out in stages until the planned size is reached or until the compactor is in cortical contact in the load-bearing zones. Cortical contact can be determined by mechanical stability and when the sound changes.
The rasps or compactors should be used in a slightly curved motion to ensure that there are no gaps that won't be filled by the prosthesis. The compactor should be beyond or flush with the resection line. If needed, the next size will be fitted.
NoteRasps of all sizes can be used for individual applications as desired.
Figure 18: Trial reduction using a compactorand trial head
Figure 19: Intraoperative sizing and X-ray control with compactor in situ
Figure 20: Implantation of the stem
Figure 21: Stem extractor
10 Surgical Instructions – NANOS™ femoral neck prosthesis
Surgery Step 3Trial positioningOnce the correctly fitting compactor is properly positioned, remove the rasp handle and add the trial head. Next the trial positioning takes place. The range of motion should be checked for possible restrictions, along with clinical review of telescoping and leg length.
Trial heads are available in diameters of 22*, 28, 32 and 36 mm with a neck length of XS*, S, M, L and XL.
*optional
After reduction, it is recommended to use a c-arm x-ray unit to assess the fit, center of rotation and offset of the implants.
Remove the trial head and compactor.
If needed, rework the prosthesis site with the rasps and a larger compactor. Lateral, between the compactor and the cortical femoral neck ring where a bone conflict can occur should the opening rasp be used, until the compactor can be put in place with the desired fit.
Surgery Step 4Implanting the stemManually insert the NANOS™ original prosthesis that corresponds to the size of the last used compactor. Subsequently impact it to the same depth as the compactor had previously, using the inserter.
Carefully clean the conus, position and impact the original ball head. Reposition, check stability and movement.
The wound is sutured in the standard way according to the surgeon’s preference.
Prosthesis RemovalIf it is necessary to intraoperatively remove a NANOS™ original prosthesis, the provided extraction instrument can be used.
This can be fitted to the cone and connected to the rasp handle to remove the implant.
Surgical Instructions – NANOS™ femoral neck prosthesis 11
BONIT® is a Registered Mark from DOT (Medical Implant Solutions).*Special size
NANOS™ femoral neck prosthesis (ISO 5832-3 Ti6Al4V, TPS/Bonit®)
Implants
Implants Art. No. SAP No.
NANOS™ femoral neck prosthesis size 0 ISO 5832-3 Ti6Al4V, TPS/Bonit 425000 75008154
NANOS™ femoral neck prosthesis size 1 ISO 5832-3 Ti6Al4V, TPS/Bonit 425001 75008155
NANOS™ femoral neck prosthesis size 2 ISO 5832-3 Ti6Al4V, TPS/Bonit 425002 75008156
NANOS™ femoral neck prosthesis size 3 ISO 5832-3 Ti6Al4V, TPS/Bonit 425003 75008157
NANOS™ femoral neck prosthesis size 4 ISO 5832-3 Ti6Al4V, TPS/Bonit 425004 75008158
NANOS™ femoral neck prosthesis size 5 ISO 5832-3 Ti6Al4V, TPS/Bonit 425005 75008159
NANOS™ femoral neck prosthesis size 6 ISO 5832-3 Ti6Al4V, TPS/Bonit 425006 75008160
NANOS™ femoral neck prosthesis size 7 ISO 5832-3 Ti6Al4V, TPS/Bonit 425007 75008161
NANOS™ femoral neck prosthesis size 8 ISO 5832-3 Ti6Al4V, TPS/Bonit 425008 75008162
NANOS™ femoral neck prosthesis size 9 ISO 5832-3 Ti6Al4V, TPS/Bonit 425009 75008163
NANOS™ femoral neck prosthesis size 10* ISO 5832-3 Ti6Al4V, TPS/Bonit 75103109 75103109
Postoperative TreatmentThe follow-up treatment depends on the surgery outcome. Mobilization can be started very quickly after a minimally traumatic implantation. The postoperative procedure with partial or full load bearing, walking aids, three point gait, four point gait will be determined by the surgeon or the person in charge of postoperative treatment on an individual basis. The bone quality and condition of the patient should be taken into consideration. Directed behavior, muscle building through physiotherapy and gait training have a positive impact on the postoperative course.
Notes Regarding Implant Unpacking The implants are packaged in 3 layers of clear pouches made of plastic film (radiation sterilization 25 kGy minimum) with cardboard. The outer pouch of the 3-layer clear pouch packaging should be removed together with the cardboard by non-sterile staff. The second pouch should be opened in a manner that will not endanger the sterility of the inner pouch. The innermost pouch will be pulled out and opened by sterile staff. The implant needs to be handed to the surgeon in this way to enable direct removal of the sterile implant.
Femoral HeadsOnly femoral heads that are compatible with the system may be used for the assembly with the implant. The approved combinations with Smith & Nephew femoral heads are listed in a compatibility matrix. The matrix is available at: www.smith-nephew.com/compatibilitymatrix (Lit. No. 04758).
12 Surgical Instructions – NANOS™ femoral neck prosthesis
NANOS™ Instrument Tray (Set SAP No. 75200170)
Instrumentation
No. Description Art. No. SAP No.
1 Opening rasp 90° rotated 506-1155 7500-8173
2 NANOS™ rasp size 0 506-1170 7500-8185
3 NANOS™ compactor size 0 506-1160 7500-8175
4 NANOS™ compactor size 1 506-1161 7500-8176
5 NANOS™ compactor size 2 506-1162 7500-8177
6 NANOS™ compactor size 3 506-1163 7500-8178
7 NANOS™ compactor size 4 506-1164 7500-8179
8 NANOS™ compactor size 5 506-1165 7500-8180
9 NANOS™ compactor size 6 506-1166 7500-8181
10 NANOS™ compactor size 7 506-1167 7500-8182
11 NANOS™ compactor size 8 506-1168 7500-8183
12 NANOS™ compactor size 9 506-1253 7501-8378
13 NANOS™ compactor size 10* 7510-3110 7510-3110
14 Rasp handle, curved 506-1169 7500-8184
15 NANOS™ impactor with silicone 7510-2877 7510-2877
16 Stem extractor 506-073 7500-8168
17 Handle adapter, hook coupling 506-075 7500-8169
18 Trial head 28 S/+0 7510-3045 7510-3045
19 Trial head 28 M/+4 7510-3046 7510-3046
20 Trial head 28 L/+8 7510-3047 7510-3047
21 Trial head 28 XL/+12 7510-3048 7510-3048
22 Trial head 32 S/+0 7510-3051 7510-3051
23 Trial head 32 M/+4 7510-3052 7510-3052
24 Trial head 32 L/+8 7510-3053 7510-3053
25 Trial head 32 XL/+12 7510-3054 7510-3054
26 Trial head 36 S/+0 7510-3057 7510-3057
27 Trial head 36 M/+4 7510-3058 7510-3058
28 Trial head 36 L/+8 7510-3059 7510-3059
29 Trial head 36 XL/+12 7510-3060 7510-3060
30 Alignment rod ø8 mm 506-015 7500-8164
NANOS™ tray 7510-3166 7510-3166
Lid for tray 7510-2521 7510-2521
*Optional
Figure 22: NANOS™ Instrument Tray
Surgical Instructions – NANOS™ femoral neck prosthesis 13
Figure 23: NANOS Rasp Tray
14 Surgical Instructions – NANOS™ femoral neck prosthesis
NANOS™ Tray Rasps(Set SAP No. 75210420)
Optional Instruments
No. Description Art. No. SAP No.
1 NANOS™ rasp size 1 506-1418 7502-9226
2 NANOS™ rasp size 2 506-1419 7502-9227
3 NANOS™ rasp size 3 506-1420 7502-9228
4 NANOS™ rasp size 4 506-1421 7502-9229
5 NANOS™ rasp size 5 506-1422 7502-9230
6 NANOS™ rasp size 6 506-1423 7502-9231
7 NANOS™ rasp size 7 506-1424 7502-9232
8 NANOS™ rasp size 8 506-1425 7502-9233
9 NANOS™ rasp size 9 506-1426 7502-9234
10 NANOS™ rasp size 10* 7510-3111 7510-3111
NANOS™ tray rasps 7510-3167 7510-3167
Lid for tray 7510-2898 7510-2898
*Optional
Surgical Instructions – NANOS™ femoral neck prosthesis 15
X-ray Templates
Description Art. No. SAP No.
Rasp handle, double offset, right, hook coupling 367-155 7510-3113
Rasp handle, double offset, left, hook coupling 367-156 7510-3114
IMT rasp adapter, double offset, right, hook coupling 367-158 7510-3115
IMT rasp adapter, double offset, left, hook coupling 367-159 7510-3116
NANOS™ IMT rasp adapter, spring hook 506-1210 7509-4887
Rasp handle, spring hook, medial release mechanism 506-1245 7503-0476
NANOS™ rasp handle, spring hook, straight 506-1204 7509-4886
Trial head 22 S/+0 7510-3040 7510-3040
Trial head 22 M/+4 7510-3041 7510-3041
Trial head 22 L/+8 7510-3042 7510-3042
Trial head 22 XL/+12 7510-3043 7510-3043
Trial head 28 XS/-3 7510-3044 7510-3044
Trial head 32 XS/-3 7510-3050 7510-3050
Trial head 36 XS/-3 7510-3056 7510-3056
Description Lit. No.
X-ray template NANOS™ femoral neck prosthesis ø22mm 12330
X-ray template NANOS™ femoral neck prosthesis ø28mm 09777
X-ray template NANOS™ femoral neck prosthesis ø32mm 09778
X-ray template NANOS™ femoral neck prosthesis ø36mm 09779
Distributed by:Smith & Nephew Orthopaedics AGOberneuhofstrasse 10d6340 BaarSwitzerlandPhone: +41 41 766 22 22
ManufacturerOHST Medizintechnik AGGrünauer Fenn 314712 Rathenow, GermanyPhone: +49 (0) 3385 5420 0Fax: +49 (0) 3385 5420 99Email: [email protected]
References:1. Stadler N, Lehner J, Trieb, K. Prospective mid-term results of a consecutive series of a short stem. Acta Orthopædica Belgica. 2016; 82:372-375. 2. Falez F, Casella F, Papalia M. Current Concepts, Classification, and Results in Short Stem Hip Arthroplasty. Healio.com/Orthopedics. 2015; 38(3):6-13. 3. Reinhardt A. Keep it short and simple – Ergebnisse einer Multicenter-Studie. Orthopädie im Profil. 2007;1:6-8. 4. Jerosch J. Kurzschaftendoprothesen an der Hüfte. Springer-Verlag GmbH, Berlin, Deutschland. 2017; 33:161-171. 5. Mai S, Pfeil J, Siebert W, Kutzner KP. Kalkar-geführte Kurzschäfte in der Hüfendoprothetik – eine Übersicht. Deutscher Ärzteverlag OUP. 2016; 5(6):342-347. 6. Falez F, Casella F, Panegrossi G, Favetti F, Barresi C. Perspectives on metaphyseal conservative stems. J Orthop Traumatol. 2008;9:49–54.
©2020 Smith & Nephew11953-en V2 01/20V26.1_50000698-003 REV. 2020-01
We have been assisting and supporting physicians and health care professionals for over 150 years.